(56 days)
This Portable Diagnostic Radiographic System is intended for use by a qualified/trained doctor or technician on both adult and pediatric subjects for taking diagnostic radiographic exposures of the skull, spinal column, chest, abdomen, extremities, and other body parts. Applications can be performed with the patient sitting, standing, or lying in the prone or supine position. (Not for mammography)
The Sedecal SPL-HF-4.0 portable x-ray generator consists of: X-ray Unit with: Control Panel with controls and displays for radiographic operations; Power Module containing control and power components; HV Tank that comprises the High Voltage Transformer, the Filament Transformer and the X-ray Tube; and a Collimator with controls to limit the X-ray beam; a Handswitch and a Mobile Column with an Articulated Arm and a Cassette Basket.
This 510(k) submission describes a new mobile X-ray system, the Sedecal SPL-HF-4.0 (and SPL-HF-2.0), which is intended for diagnostic radiographic exposures. The submission aims to demonstrate substantial equivalence to a legally marketed predicate device, the Sedecal Models SP-HF-2.8 and SP-HF-4.0 (K020436).
Based on the provided information, the focus is on demonstrating safety and effectiveness through non-clinical testing rather than a clinical study involving human subjects and diagnostic performance metrics. Therefore, many of the requested elements pertaining to clinical study design, such as sample size, ground truth establishment for a test set, expert qualifications, and comparative effectiveness, are not applicable in this context.
Here's a breakdown of the available information:
1. Table of Acceptance Criteria and Reported Device Performance
| Acceptance Criteria | Reported Device Performance |
|---|---|
| Electrical Safety | Tested to IEC 60601-1 |
| Electromagnetic Compatibility (EMC) | Tested to IEC 60601-1-2 |
| Software Validation | Software validation testing performed |
| Radiation Safety/Performance | Tested to IEC 60601-1-3 and IEC 60601-2-7 (These standards relate to general requirements for radiation protection in diagnostic X-ray equipment and particular requirements for the safety of medical electron accelerators, respectively. While IEC 60601-1-3 is directly relevant to diagnostic X-ray, the inclusion of IEC 60601-2-7 seems slightly out of place for a mobile diagnostic X-ray system and might be a typo or refer to a general safety assessment for components). Implicitly, the device's technological characteristics are "nearly identical" to the predicate, suggesting similar performance. |
| Substantial Equivalence | Based on results of nonclinical tests, the device is "as safe, as effective, and performs as well as or better than the legally marketed devices." Materials and construction methods are "nearly identical" to the predicate. |
2. Sample size used for the test set and the data provenance
Not applicable. The submission describes non-clinical testing (electrical safety, EMC, software, and radiation safety standards) rather than a clinical study with a test set of patient data.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts
Not applicable. No ground truth for image interpretation or diagnostic accuracy was established as this was not a clinical performance study.
4. Adjudication method for the test set
Not applicable. No clinical test set requiring adjudication was used.
5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
Not applicable. No MRMC study was performed as this is a new X-ray generator, not an AI-based interpretation device.
6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done
Not applicable. This is a medical device (X-ray generator), not an algorithm or AI system.
7. The type of ground truth used
Not applicable for a clinical performance study. For the non-clinical testing, the "ground truth" would be the specifications and requirements defined by the referenced IEC standards and internal design specifications.
8. The sample size for the training set
Not applicable. No training set for an algorithm was used.
9. How the ground truth for the training set was established
Not applicable. No training set for an algorithm was used.
Summary of the Study that Proves the Device Meets Acceptance Criteria:
The study proving the device meets its acceptance criteria is primarily a series of non-clinical tests mandated by international standards for medical electrical equipment and X-ray devices. These include:
- Electrical Safety Testing (IEC 60601-1): This standard ensures the device does not pose electrical hazards to patients or operators.
- Electromagnetic Compatibility (EMC) Testing (IEC 60601-1-2): This verifies that the device functions correctly in its electromagnetic environment and does not interfere with other equipment.
- Software Validation Testing: Ensures the software controlling the device functions as intended and is reliable.
- Radiation Safety/Performance Testing (IEC 60601-1-3 and IEC 60601-2-7): IEC 60601-1-3 specifies general requirements for radiation protection in diagnostic X-ray equipment, ensuring appropriate X-ray emission and safety features. As mentioned, IEC 60601-2-7's relevance here is somewhat ambiguous; it typically applies to medical electron accelerators. However, its inclusion suggests a comprehensive approach to safety standards where applicable.
The submission asserts that based on the results of these non-clinical tests, the device is "as safe, as effective, and performs as well as or better than the legally marketed devices" (specifically, the predicate device K020436). The core argument for acceptance and substantial equivalence relies on the fact that the new device shares "nearly identical" technological characteristics, design, materials, chemical composition, and energy source with the predicate device, with this submission representing an "updated design" of the previous model. The conclusion is that because the updated model passes the relevant safety and performance standards, and its technical characteristics are so similar to an already cleared device, it is substantially equivalent.
{0}------------------------------------------------
510(k) Summary
510(k) Number K103 S-22
JAN 2 6 2011
l. Submitter:
SEDECAL SA C/ Pelaya, 9 - 13 Pol. Ind. Río de Janeiro 28110 Algete, Madrid, España (Spain) Tel .- +34 91 6280544 Fax.- +34 91 6280574 Date Prepared: October 27, 2010 Contact: Mª Luisa Gómez de Agüero, Quality and Regulatory Manager
-
- Identification of the Device: Proprietary-Trade Name: Sedecal SPL-HF-4.0 (and SPL-HF-2.0) Classification Name: Mobile x-ray system, Product Code IZL Common/Usual Name: Mobile Diagnostic X-Ray System
-
- Equivalent legally marketed device: K020436 Sedecal Models SP-HF-2.8 and SP-HF-4.0
-
- Description of the Device: The Sedecal SPL-HF-4.0 portable x-ray generator consists of: X-ray Unit with: Control Panel with controls and displays for radiographic operations; Power Module containing control and power components; HV Tank that comprises the High Voltage Transformer, the Filament Transformer and the X-ray Tube; and a Collimator with controls to limit the X-ray beam; a Handswitch and a Mobile Column with an Articulated Arm and a Cassette Basket.
-
- Indications for Use (intended use) Intended for use by a qualified/trained doctor or technician on both adult and pediatric subjects for taking diagnostic radiographic exposures of the skull, spinal column, chest, abdomen, extremities, and other body parts. Applications can be performed with the patient sitting, standing, or lying in the prone or supine position. (Not for mammography)
-
- Technological Characteristics: This device has the same technological characteristics (i.e., design, material, chemical composition, energy source) as the predicate device. Specifications are nearly identical. This submission represents an updated design of our previous model in K020436.
-
- Discussion of the nonclinical tests in the premarket notification submission for a Determination of substantial equivalence: We performed electrical safety (IEC 60601-1), electromagnetic compatibility testing, (IEC 60601-1-2), software validation testing, and testing to IEC 60601-1-3 and IEC 60601-2-7.
-
- Conclusion. Based on the results of the nonclinical tests (that demonstrate that the device is as safe, as effective, and performs as well as or better than the legally marketed devices) we sare, as ofted this modified x-ray system is safe and effective as the predicate identified in paragraph (3). Furthermore, the materials and construction methods are nearly identical to the predicate.
P. 1 of 1
Page 16 of 591
{1}------------------------------------------------
Image /page/1/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three tail feathers. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" are arranged in a circle around the eagle.
DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room - WO66-G609 Silver Spring, MD 20993-0002
Sedecal SA % Mr. Daniel Kamm, P.E. Principal Consultant Kamm & Associates 8870 Ravello Ct NAPLES FL 34114
JAN 2 6 2011
Re: K103522
Trade/Device Name: Sedecal SPL-HF-4.0 (and SPL-HF-2.0) Regulation Number: 21 CFR 892.1720 Regulation Name: Mobile x-ray system Regulatory Class: II Product Code: IZL Dated: November 24, 2010 Received: December 1, 2010
Dear Mr. Kamm:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into class II (Special Controls), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of
{2}------------------------------------------------
medical device-related adverse events) (21 CFR 803); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820). This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Parts 801 and 809), please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (301) 796-5450. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely Yours,
Mary S Postel
Mary Pastel, ScD. Director Division of Radiological Devices Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
Indications for Use
510(k) Number (if known): K1035 22
Device Name: Sedecal SPL-HF-4.0 (and SPL-HF-2.0)
Indications For Use:
This Portable Diagnostic Radiographic System is intended for use by a qualified/trained doctor or technician on both adult and pediatric subjects for taking diagnostic radiographic exposures of the skull, spinal column, chest, abdomen, extremities, and other body parts. Applications can be performed with the patient sitting, standing, or lying in the prone or supine position. (Not for mammography)
Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of In-vitro Diagnostics (OIVD)
Office of In Vitro Diagnostic Devices
(Division Sign-Off)
Division of Radiological Devices
Device Evaluation and Safe
510K K103522
Page 1 of 1
§ 892.1720 Mobile x-ray system.
(a)
Identification. A mobile x-ray system is a transportable device system intended to be used to generate and control x-ray for diagnostic procedures. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.(b)
Classification. Class II.