K Number
K020436
Manufacturer
Date Cleared
2002-02-22

(14 days)

Product Code
Regulation Number
892.1720
Panel
RA
Reference & Predicate Devices
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Models SP-HF-2.8 and SP-HF-4.0 are intended for use by a qualified/trained doctor or technician on both adult and pediatric subjects for taking diagnostic radiographic exposures of the skull, spinal column, chest, abdomen, extremities, and other body parts. Applications can be performed with the patient sitting, standing, or lying in the prone or supine position.

Device Description

Models SP-HF-2.8 and SP-HF-4.0 are a portable units which operate from 120 V 50-60~ AC. The unit utilizes a newly designed high frequency inverter and can be either mounted to a tripod or support arm or can be hand held.

AI/ML Overview

The provided document describes the safety and effectiveness summary for two portable X-ray units, Models SP-HF-2.8 and SP-HF-4.0. The study's acceptance criteria are implicitly defined by demonstrating substantial equivalence to a predicate device, the MinXray HF100H.

Here's an analysis of the acceptance criteria and study information:

1. Table of Acceptance Criteria and Reported Device Performance

The acceptance criteria are generally understood to be that the new devices (Models SP-HF-2.8 and SP-HF-4.0) perform as safely and effectively as the predicate device (MinXray HF100H) and present no new indications for use or significant technological differences that would raise new questions of safety or effectiveness. The document achieves this by comparing technical specifications and intended use.

CharacteristicAcceptance Criteria (Predicate Device: MinXray HF100H)Reported Device Performance (Models SP-HF-2.8 & SP-HF-4.0)Meets Acceptance Criteria?
Intended UsePortable general purpose diagnostic X-ray unitSAME (Models SP-HF-2.8 and SP-HF-4.0 are intended for use by a qualified/trained doctor or technician on both adult and pediatric subjects for taking diagnostic radiographic exposures of the skull, spinal column, chest, abdomen, extremities, and other body parts. Applications can be performed with the patient sitting, standing, or lying in the prone or supine position.)Yes
Physical Characteristics:
Size8.75"H x 9.5" W x 15.35" D.8.7"H x 10.4" W x 16.5" DYes*
Weight40.9 lbs33 lbYes*
Energy Source120 V 50-60~ AC90 to 285 VAC (50-60 Hz)Yes*
User InterfaceUp-Down pushbuttons for three kVp selections and exposure time selections with LED indicatorsUp-Down pushbuttons for kVp and mAs. kVp adjustable in 1 kVp stepsYes*
Exposure times0.08-2.00 Sec. In 192 stepsSP-HF-2.8: 0.002-10 sec 38 steps
SP-HF-4.0: 0.001-10 sec 41 stepsYes*
Ma.20 mASP-HF-2.8: 5, 6.4, 8, 10, 12.5, 16, 20, 25, 32, 40, 50
SP-HF-4.0: 5, 6.4, 8, 10, 12.5, 16, 20, 25, 32, 40, 50, 64, 80, 100Yes*
KvP100 KvPSP-HF-2.8: 110 KvP
SP-HF-4.0: 115 KvPYes*
Standards & Safety Characteristics:
Performance Standard21 CFR 1020.30SAMEYes
Electrical safetyUL 2601, IEC 60601-1SAMEYes
Overall Safety and Effectiveness"as safe and effective as the predicate device""the Models SP-HF-2.8 and SP-HF-4.0 are as safe and effective as the predicate device, have few technological differences, and has no new indications for use, thus rendering it substantially equivalent to the predicate device."Yes

Note: For the physical and operational characteristics (Size, Weight, Energy Source, User Interface, Exposure times, Ma., KvP), the acceptance criteria are not explicitly stated as strict numerical equivalence. Instead, "substantial equivalence" is being claimed, meaning the differences are not considered to raise new questions of safety or effectiveness. The new devices often show improvements (e.g., wider range of mAs/kVp, lower weight) or minor variations that are deemed acceptable and within the scope of the predicate device's intended use.

2. Sample size used for the test set and the data provenance

The document states that "bench and user testing" was performed. However, there is no specific sample size provided for the test set. The data provenance (country of origin, retrospective/prospective) is also not specified.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

The document mentions "user testing" but does not specify the number or qualifications of experts used to establish ground truth. Given the nature of the device (portable X-ray unit) and the era of the submission, "user testing" likely refers to evaluations by technicians or qualified medical personnel regarding the device's operational aspects and image acquisition capabilities rather than diagnostic interpretation.

4. Adjudication method (e.g., 2+1, 3+1, none) for the test set

The document does not describe any adjudication method for establishing ground truth from the "user testing."

5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

No MRMC comparative effectiveness study was done or reported. This device is a portable X-ray unit, not an AI-powered diagnostic tool. The comparison is between the new X-ray unit and a predicate X-ray unit, focusing on technical specifications and overall safety/effectiveness in producing images, not on human reader performance or AI assistance.

6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

No standalone algorithm performance study was done or reported. This is a hardware device for image acquisition, not a software algorithm.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

The concept of a "ground truth" as typically applied to diagnostic performance studies (e.g., pathology, expert consensus on disease presence) is not applicable or described in this 510(k) summary. The "ground truth" for this submission revolves around the technical performance metrics of the X-ray unit (e.g., kVp, mA, exposure times, safety standards) and whether it can produce diagnostic-quality images, implicitly judged during "user testing." The "ground truth" for substantial equivalence is the established safety and effectiveness of the predicate device based on its technical specifications and intended use.

8. The sample size for the training set

Not applicable. This submission is for a medical device (X-ray unit) that does not involve machine learning or AI algorithms requiring a training set in the typical sense.

9. How the ground truth for the training set was established

Not applicable. As stated above, this device does not utilize a training set.

§ 892.1720 Mobile x-ray system.

(a)
Identification. A mobile x-ray system is a transportable device system intended to be used to generate and control x-ray for diagnostic procedures. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.(b)
Classification. Class II.