(204 days)
Not Found
No
The summary describes a standard mobile X-ray system and does not mention any AI or ML capabilities, image processing beyond basic digital radiography, or performance metrics typically associated with AI/ML algorithms.
No.
The device is clearly indicated for "general-purpose diagnostic radiographic examinations and procedures" and is designed to "generate diagnostic radiographic images." It is an imaging system used to detect disease or injury, not to provide therapy.
Yes
The device is described as "designed to generate diagnostic radiographic images (medical x-rays)" and is indicated for "general-purpose diagnostic radiographic examinations and procedures."
No
The device description clearly states it is a "mobile radiographic imaging system" and mentions hardware components like a "GE Wireless Detector" and "flat panel detector." This indicates it is a hardware device with integrated software, not a software-only medical device.
Based on the provided text, this device is not an IVD (In Vitro Diagnostic).
Here's why:
- IVD Definition: In vitro diagnostics are tests performed on samples taken from the human body, such as blood, urine, or tissue, to detect diseases, conditions, or infections.
- Device Function: The GE Automatic Mobile X-Ray (AMX) Series is a radiographic imaging system. It generates medical x-rays to create images of the internal structures of the body. This is an in vivo diagnostic method (performed on a living organism), not an in vitro method.
- Intended Use: The intended use clearly states that the device is for generating diagnostic radiographic images to detect disease or injury. It describes taking exposures of various body parts. This aligns with in vivo imaging.
Therefore, the GE Automatic Mobile X-Ray (AMX) Series is a medical imaging device, not an in vitro diagnostic device.
N/A
Intended Use / Indications for Use
The GE Automatic Mobile X-Ray (AMX) Series: Brivo XR285amx, Optima XR200amx, Optima XR220amx are intended to take exposures utilizing film or computed radiography (CR), however the Optima XR220amx utilizes the GE Wireless Detector, which is intended to replace radiographic film screen systems in all general purpose diagnostic procedures, for digital . radiography (DR).
Brivo XR285amx, Optima XR200amx, Optima XR220amx are self-contained: battery operated mobile radiographic imaging systems designed to generate diagnostic radiographic images (medical x-rays) that may increase the ability to detect disease or injury early enough for a medical problem to be managed, treated, or cured. Medical x-rays are used in many types of examinations and procedures, some examples include: x-ray radiography (to find orthopedic damage, tumors, pneumonias, foreign objects).
The series are indicated for use on adult and pediatric patients for general-purpose diagnostic radiographic examinations and procedures. Its mobility enables general-purpose radiographic procedures throughout the clinical environment, or as needed within the emergency, intensive care, premature birth ward, cardiac and operating departments, for patients that may not be able to be moved or in cases where it is unsafe or impractical to move them to a traditional RAD room.
The systems are indicated for taking radiographic exposures of the skull, spinal column, chest, abdomen, extremities, and other body parts with the patient sitting, standing, or lying in the prone or supine position.
These devices are not intended for mammographic applications.
Product codes (comma separated list FDA assigned to the subject device)
IZL
Device Description
The GE Automatic Mobile X-Ray (AMX) Series: Brivo XR285amx, Optima XR200amx, Optima XR220amx are intended to take exposures utilizing film or computed radiography (CR), however the Optima XR220amx utilizes the GE Wireless Detector, which is intended to replace radiographic film screen systems in all general purpose diagnostic procedures, for digital radiography (DR).
Brivo XR285amx, Optima XR200amx, Optima XR220amx are self-contained; battery operated mobile radiographic imaging systems designed to generate diagnostic radiographic images (medical x-rays) that may increase the ability to detect disease or injury early enough for a medical problem to be managed, treated, or cured. Medical x-rays are used in many types of examinations and procedures, some examples include: x-ray radiography (to find orthopedic damage, tumors, pneumonias, foreign objects).
The series are indicated for use on adult and pediatric patients for general-purpose diagnostic radiographic examinations and procedures. Its mobility enables general-purpose radiographic procedures throughout the clinical environment, or as needed within the emergency, intensive care, premature birth ward, cardiac and operating departments, for patients that may not be able to be moved or in cases where it is unsafe or impractical to move them to a traditional RAD room.
The flat panel detector provides increased functionality to enable images of patients of all sizes, and can produce comparable quality images with as little as half the dose of traditional computer radiography (CR), cassettes and other flat panel detectors with lower DQE. The digital detector is designed to withstand a distributed load of 352Lbs to accommodate certain larger patients.
The systems are indicated for taking radiographic exposures of the skull, spinal column, chest, abdomen, extremities, and other body parts with the patient sitting, standing, or lying in the prone or supine position.
These devices are not intended for mammographic applications.
Mentions image processing
Not Found
Mentions AI, DNN, or ML
Not Found
Input Imaging Modality
X-ray
Anatomical Site
skull, spinal column, chest, abdomen, extremities, and other body parts
Indicated Patient Age Range
adult and pediatric patients
Intended User / Care Setting
Not Found
Description of the training set, sample size, data source, and annotation protocol
Not Found
Description of the test set, sample size, data source, and annotation protocol
Not Found
Summary of Performance Studies (study type, sample size, AUC, MRMC, standalone performance, key results)
The subject of this premarket submission, [Brivo XR285amx, Optima XR200amx, Optima XR220amx], did not require clinical studies to support substantial equivalence.
Key Metrics (Sensitivity, Specificity, PPV, NPV, etc.)
Not Found
Predicate Device(s): If the device was cleared using the 510(k) pathway, identify the Predicate Device(s) K/DEN number used to claim substantial equivalence and list them here in a comma separated list exactly as they appear in the text. List the primary predicate first in the list.
Reference Device(s): Identify the Reference Device(s) K/DEN number and list them here in a comma separated list exactly as they appear in the text.
Not Found
Predetermined Change Control Plan (PCCP) - All Relevant Information for the subject device only (e.g. presence / absence, what scope was granted / cleared under the PCCP, any restrictions, etc).
Not Found
§ 892.1720 Mobile x-ray system.
(a)
Identification. A mobile x-ray system is a transportable device system intended to be used to generate and control x-ray for diagnostic procedures. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.(b)
Classification. Class II.
0
JUN 1 6 2011
GE Healthcare
510(k) Premarket Notification Submission Brivo XR285amx, Optima XR200amx, Optima XR220amx
510(k) Summary י
In accordance with 21 CFR 807.92 the following summary of information is provided: | |
---|---|
Date: | November 23, 2010 |
Submitter: | GE Healthcare, (GE Medical Systems, LLC) |
3000 N. Grandview Blvd. | |
Waukesha, WI 53188 | |
Primary Contact Person: | Nidhi Chaudhary |
Regulatory Affairs Leader, X-Ray | |
GE Healthcare, (GE Medical Systems, LLC) | |
Telephone: 414-721-2899; Fax: 414-918-8184 | |
e-mail: Nidhi.Chaudhary@ge.com | |
Secondary Contact Person: | David Blonski |
Regulatory Affairs Director, X-Ray | |
GE Healthcare, (GE Medical Systems, LLC) | |
Telephone: 262-513-4072; Fax: 262-364-2509 | |
e-mail: David.Blonski@ge.com | |
Device Identification: | |
Trade Name: | Brivo XR285amx ; Optima XR200amx ; Optima XR220amx |
Common/Usual Name: | Brivo XR285amx – Analog Mobile |
Optima XR200amx - Digital Ready Mobile | |
Optima XR220amx - Digital Mobile | |
Classification Names: | Brivo XR285amx ; Optima XR200amx ; Optima XR220amx |
Product Code: | Class II, IZL, System, X-ray, Mobile, 21 CFR 892.1720 |
Predicate Device(s): | AMX-4 Plus Mobile X-ray system K021016 |
GE Definium AMX700, Model AMX 700 K052897 | |
WIRELESS DR IMAGING OPTION WDR1 K102615 (pending) | |
Image /page/0/Picture/6 description: The image shows the General Electric (GE) logo. The logo consists of the letters "GE" intertwined within a circular border. The letters are stylized with curved lines and a vintage aesthetic. The image is in black and white, with the logo appearing in black against a white background.
1
Image /page/1/Picture/0 description: The image shows the General Electric (GE) logo. The logo consists of the letters 'G' and 'E' intertwined in a stylized, cursive font. The letters are enclosed within a circular border that has a textured or patterned design. The logo is black and white.
GE Healthcare 510(k) Premarket Notification Submission Brivo XR285amx, Optima XR200amx, Optima XR220amx
Device Description:
The GE Automatic Mobile X-Ray (AMX) Series: Brivo XR285amx, Optima XR200amx, Optima XR220amx are intended to take exposures utilizing film or computed radiography (CR), however the Optima XR220amx utilizes the GE Wireless Detector, which is intended to replace radiographic film screen systems in all general purpose diagnostic procedures, for digital radiography (DR).
Brivo XR285amx, Optima XR200amx, Optima XR220amx are self-contained; battery operated mobile radiographic imaging systems designed to generate diagnostic radiographic images (medical x-rays) that may increase the ability to detect disease or injury early enough for a medical problem to be managed, treated, or cured. Medical x-rays are used in many types of examinations and procedures, some examples include: x-ray radiography (to find orthopedic damage, tumors, pneumonias, foreign objects).
The series are indicated for use on adult and pediatric patients for general-purpose diagnostic radiographic examinations and procedures. Its mobility enables general-purpose radiographic procedures throughout the clinical environment, or as needed within the emergency, intensive care, premature birth ward, cardiac and operating departments, for patients that may not be able to be moved or in cases where it is unsafe or impractical to move them to a traditional RAD room.
The flat panel detector provides increased functionality to enable images of patients of all sizes, and can produce comparable quality images with as little as half the dose of traditional computer radiography (CR), cassettes and other flat panel detectors with lower DQE. The digital detector is designed to withstand a distributed load of 352Lbs to accommodate certain larger patients.
The systems are indicated for taking radiographic exposures of the skull, spinal column, chest, abdomen, extremities, and other body parts with the patient sitting, standing, or lying in the prone or supine position.
These devices are not intended for mammographic applications.
2
Image /page/2/Picture/0 description: The image shows the General Electric (GE) logo. The logo consists of the letters 'G' and 'E' intertwined in a stylized, cursive font. The letters are surrounded by a circular border with a swirling pattern inside.
GE Healthcare 510(k) Premarket Notification Submission Brivo XR285amx, Optima XR200amx, Optima XR220amx
Indications for Use:
The GE Automatic Mobile X-Ray (AMX) Series: Brivo XR285amx, Optima XR200amx, Optima XR220amx are intended to take exposures utilizing film or computed radiography (CR), however the Optima XR220amx utilizes the GE Wireless Detector, which is intended to replace radiographic film screen systems in all general purpose diagnostic procedures, for digital radiography (DR).
Brivo XR285amx, Optima XR200amx, Optima XR220amx are self-contained; battery operated mobile radiographic imaging systems designed to generate diagnostic radiographic images (medical x-rays) that may increase the ability to detect disease or injury early enough for a medical problem to be managed, treated, or cured. Medical x-rays are used in many types of examinations and procedures, some examples include: x-ray radiography (to find orthopedic damage, tumors, pneumonias, foreign objects).
The series are indicated for use on adult and pediatric patients for general-purpose diagnostic radiographic examinations and procedures. Its mobility enables general-purpose radiographic procedures throughout the clinical environment, or as needed within the emergency, intensive care, premature birth ward, cardiac and operating departments, for patients that may not be able to be moved or in cases where it is unsafe or impractical to move them to a traditional RAD room.
The systems are indicated for taking radiographic exposures of the skull, spinal column, chest, abdomen, extremities, and other body parts with the patient sitting, standing, or lying in the prone or supine position.
These devices are not intended for mammographic applications.
The systems are indicated for taking radiographic exposures of the skull, spinal column, chest, abdomen, extremities, and other body parts with the patient sitting, standing, or lying in the prone or supine position.
These devices are not intended for mammographic applications.
Technology:
The Brivo XR285amx, Optima XR200amx, Optima XR220amx employs the same fundamental scientific technology as its predicate devices.
3
Image /page/3/Picture/0 description: The image shows the General Electric (GE) logo. The logo consists of the letters 'GE' intertwined within a circular border. The letters are stylized with a flowing, cursive design, and the entire logo has a textured, slightly blurred appearance.
GE Healthcare
. 510(k) Premarket Notification Submission Brivo XR285amx, Optima XR200amx, Optima XR220amx
Determination of Substantial Equivalence:
Summary of Non-Clinical Tests:
The Brivo XR285amx, Optima XR200amx, Optima XR220amx and its applications comply with voluntary standards as detailed in Section 9, 11 and 17 of this premarket submission. The following quality assurance measures were applied to the development of the system:
- . Risk Analysis
- . Requirements Reviews
- Design Reviews
- . Testing on unit level (Module verification)
- Integration testing (System verification) .
- . Performance testing (Verification)
- . Safety testing (Verification)
- Simulated use testing (Validation) �
Summary of Clinical Tests:
The subject of this premarket submission, [Brivo XR285amx, Optima XR200amx, Optima XR220amx], did not require clinical studies to support substantial equivalence.
Comparison with Predicate Devices:
The Brivo and Optima Systems are updated versions of the AMX4+ and Definium AMX700 products. The primary changes from the predicates are due to improved technology Ainze the AMX4+ and AMX700 products were designed. Changes involve reduced overall size and weight of the product, improved charging efficiency, implementation of a dual focal sportune, and higher power generator. Additionally in comparison to the AMX4+ product, the Frivo system has an integrated touch screen display for customer input. Modifications to implement these updates involve hardware, software and firmware changes. The Brivo and Optima systems are all for use in mobile x-ray radiology and have similar indications for use. Changes made to achieve the new system designs were made in hardware, software and firmware, but there is no fundamental change in the use of the product for mobile x-ray, nor in technology. It has the same technological characteristics related to safety and effectiveness as the bredicate devices. A review of all bench and standards testing indicate that the new device provides no new safety concerns and is as safe an effective as the predicate devices. The Brivo and Ordima systems are certified to comply with the X-ray requirements of 21 CFR, arreautic spafety requirements of IEC 60601-1 and associated collateral and particular standards.
4
Image /page/4/Picture/0 description: The image shows the General Electric (GE) logo. The logo consists of the letters 'G' and 'E' intertwined in a stylized, circular design. The letters are surrounded by a circular border with a decorative, swirling pattern.
GE Healthcare
510(k) Premarket Notification Submission Brivo XR285amx, Optima XR200amx, Optima XR220amx
Adverse Effects on Health:
The device has been evaluated for electrical, mechanical, and radiation safety, and conforms to applicable medical device safety standards, as confirmed by a Nationally Recognized Test Laboratory.
The potential hazards of electrical and mechanical are identified in a risk management summary (hazard analysis) and controlled by:
-
System verification and validation to ensure performance to specifications,
Federal Regulations, and user requirements. -
Adherence and certification to industry and international standards.
-
Compliance to applicable CDRH 21 CFR subchapter J requirements.
The device is designed and manufactured under the Quality System Regulations of 21CFR 820.
Conclusion:
GE Healthcare considers the Brivo XR285amx, Optima XR200amx, Optima XR220amx to be as safe, as effective, and performance is substantially equivalent to the predicate device(s).
The Brivo and Optima systems are an improvement in design technology, and customer needs. It does not result in any new potential safety risks, has the same technological characteristics, and performs as well or better then devices currently on the market. The Brivo and Optima Systems will be certified to comply with the X-ray requirements of 21 CFR, as well as safety requirements of IEC 60601-1 and associated collateral and particular standards.
After analyzing standards testing and bench data, it is the conclusion of GE Healthcare that the Brivo and Optima Systems are substantially equivalent to other marketed devices with similar indications for use and meeting the same standards.
5
Image /page/5/Picture/0 description: The image is a black and white logo for the U.S. Department of Health and Human Services. The logo features a stylized abstract symbol resembling a bird or a person with outstretched arms. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES-USA" is arranged in a circular fashion around the symbol.
DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room - WO66-G609 Silver Spring, MD 20993-0002
Ms. Nidhi Chaudhary Regulatory Affairs Leader, X-Ray GE Medical Systems, LLC 3200 N. Grandview Blvd. WAUKESHA WI 51388
JUN 1 6 201
Re: K103476
Trade/Device Name: Brivo XR285amx, Optima XR200amx, Optima XR220amx Regulation Number: 21 CFR 892.1720 Regulation Name: Mobile x-ray system Regulatory Class: II Product Code: IZL Dated: May 1, 2011 Received: May 13, 2011
Dear Ms. Chaudhary:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into class II (Special Controls), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of
6
medical device-related adverse events) (21 CFR 803); and good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820). This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Parts 801 and 809), please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (301) 796-5450. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely Yours.
Mary Pastel
Mary S. Pastel, Sc.D. Director Division of Radiological Devices Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health
Enclosure
7
Image /page/7/Picture/0 description: The image shows the General Electric (GE) logo. The logo consists of the letters "G" and "E" intertwined in a stylized, circular design. The letters are surrounded by a textured, circular border.
GE Healthcare
510(k) Premarket Notification Submission Brivo XR285anx, Optima XR200amx, Optima XR220amx
510(k) Number (if known):
Device Name: Family:
Brivo XR285amx, Optima XR200amx, Optima XR220amx Brivo XR285amx - Analog Mobile Optima XR200amx - Digital Ready Mobile Optima XR220amx - Digital Mobile
Indications for Use:
The GE Automatic Mobile X-Ray (AMX) Series: Brivo XR285amx, Optima XR200amx, Optima XR220amx are intended to take exposures utilizing film or computed radiography (CR), however the Optima XR220amx utilizes the GE Wireless Detector, which is intended to replace radiographic film screen systems in all general purpose diagnostic procedures, for digital . radiography (DR).
Brivo XR285amx, Optima XR200amx, Optima XR220amx are self-contained: battery operated mobile radiographic imaging systems designed to generate diagnostic radiographic images (medical x-rays) that may increase the ability to detect disease or injury early enough for a medical problem to be managed, treated, or cured. Medical x-rays are used in many types of examinations and procedures, some examples include: x-ray radiography (to find orthopedic damage, tumors, pneumonias, foreign objects).
The series are indicated for use on adult and pediatric patients for general-purpose diagnostic radiographic examinations and procedures. Its mobility enables general-purpose radiographic procedures throughout the clinical environment, or as needed within the emergency, intensive care, premature birth ward, cardiac and operating departments, for patients that may not be able to be moved or in cases where it is unsafe or impractical to move them to a traditional RAD room.
The systems are indicated for taking radiographic exposures of the skull, spinal column, chest, abdomen, extremities, and other body parts with the patient sitting, standing, or lying in the prone or supine position.
These devices are not intended for mammographic applications.
Prescription Use X AND/OR Over-The-Counter Use (Part 21 CFR 801 Subpart D) (Part 21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD)
mun Division Sign-Off Office of In Vitro Diagnostic Device Evaluation and Safety 510(k) 1102476