K Number
K103476
Date Cleared
2011-06-16

(204 days)

Product Code
Regulation Number
892.1720
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The GE Automatic Mobile X-Ray (AMX) Series: Brivo XR285amx, Optima XR200amx, Optima XR220amx are intended to take exposures utilizing film or computed radiography (CR), however the Optima XR220amx utilizes the GE Wireless Detector, which is intended to replace radiographic film screen systems in all general purpose diagnostic procedures, for digital radiography (DR).

Brivo XR285amx, Optima XR200amx, Optima XR220amx are self-contained; battery operated mobile radiographic imaging systems designed to generate diagnostic radiographic images (medical x-rays) that may increase the ability to detect disease or injury early enough for a medical problem to be managed, treated, or cured. Medical x-rays are used in many types of examinations and procedures, some examples include: x-ray radiography (to find orthopedic damage, tumors, pneumonias, foreign objects).

The series are indicated for use on adult and pediatric patients for general-purpose diagnostic radiographic examinations and procedures. Its mobility enables general-purpose radiographic procedures throughout the clinical environment, or as needed within the emergency, intensive care, premature birth ward, cardiac and operating departments, for patients that may not be able to be moved or in cases where it is unsafe or impractical to move them to a traditional RAD room.

The systems are indicated for taking radiographic exposures of the skull, spinal column, chest, abdomen, extremities, and other body parts with the patient sitting, standing, or lying in the prone or supine position.

These devices are not intended for mammographic applications.

Device Description

The GE Automatic Mobile X-Ray (AMX) Series: Brivo XR285amx, Optima XR200amx, Optima XR220amx are intended to take exposures utilizing film or computed radiography (CR), however the Optima XR220amx utilizes the GE Wireless Detector, which is intended to replace radiographic film screen systems in all general purpose diagnostic procedures, for digital radiography (DR).

Brivo XR285amx, Optima XR200amx, Optima XR220amx are self-contained; battery operated mobile radiographic imaging systems designed to generate diagnostic radiographic images (medical x-rays) that may increase the ability to detect disease or injury early enough for a medical problem to be managed, treated, or cured. Medical x-rays are used in many types of examinations and procedures, some examples include: x-ray radiography (to find orthopedic damage, tumors, pneumonias, foreign objects).

The series are indicated for use on adult and pediatric patients for general-purpose diagnostic radiographic examinations and procedures. Its mobility enables general-purpose radiographic procedures throughout the clinical environment, or as needed within the emergency, intensive care, premature birth ward, cardiac and operating departments, for patients that may not be able to be moved or in cases where it is unsafe or impractical to move them to a traditional RAD room.

The flat panel detector provides increased functionality to enable images of patients of all sizes, and can produce comparable quality images with as little as half the dose of traditional computer radiography (CR), cassettes and other flat panel detectors with lower DQE. The digital detector is designed to withstand a distributed load of 352Lbs to accommodate certain larger patients.

The systems are indicated for taking radiographic exposures of the skull, spinal column, chest, abdomen, extremities, and other body parts with the patient sitting, standing, or lying in the prone or supine position.

These devices are not intended for mammographic applications.

AI/ML Overview

1. A table of acceptance criteria and the reported device performance

The provided document describes the Brivo XR285amx, Optima XR200amx, Optima XR220amx mobile X-ray systems. It outlines the safety and performance evaluations conducted, but it does not explicitly state specific quantitative acceptance criteria or corresponding reported device performance metrics in a readily extractable table format.

Instead, the document states:

  • "The Brivo and Optima Systems are certified to comply with the X-ray requirements of 21 CFR, arreautic spafety requirements of IEC 60601-1 and associated collateral and particular standards." (Page 3)
  • "A review of all bench and standards testing indicate that the new device provides no new safety concerns and is as safe an effective as the predicate devices." (Page 3)
  • "The device has been evaluated for electrical, mechanical, and radiation safety, and conforms to applicable medical device safety standards, as confirmed by a Nationally Recognized Test Laboratory." (Page 4)
  • "GE Healthcare considers the Brivo XR285amx, Optima XR200amx, Optima XR220amx to be as safe, as effective, and performance is substantially equivalent to the predicate device(s)." (Page 4)
  • "It does not result in any new potential safety risks, has the same technological characteristics, and performs as well or better then devices currently on the market." (Page 4)

These statements indicate that the acceptance criteria were based on compliance with relevant regulatory standards (21 CFR, IEC 60601-1) and demonstrating substantial equivalence in safety and effectiveness to predicate devices. The "reported device performance" is implicitly stated as meeting these standards and being comparable or superior to predicate devices.

2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

The document explicitly states: "The subject of this premarket submission, [Brivo XR285amx, Optima XR200amx, Optima XR220amx], did not require clinical studies to support substantial equivalence." (Page 3)

Therefore, there is no test set of clinical images or patient data described in this submission. The evaluation was based on non-clinical tests, bench data, and compliance with standards.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

As no clinical studies or test sets with patient images were conducted (as stated in point 2), there was no ground truth established by experts for a test set in this submission.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

As no clinical studies or test sets were conducted (as stated in point 2), there was no adjudication method used for a test set in this submission.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

No MRMC comparative effectiveness study was done. The submission explicitly states no clinical studies were required or performed. There is no mention of AI assistance in this document, which predates widespread clinical AI applications in radiology.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

No standalone algorithm performance study was done. This submission is for X-ray imaging systems, not image analysis algorithms. The device itself is the primary subject of evaluation, not a standalone algorithm.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

Since no clinical studies were performed, no clinical ground truth (expert consensus, pathology, outcomes data) was used. The "ground truth" for the device's performance was established through compliance with technical standards and bench testing, demonstrating that the device meets specified electrical, mechanical, and radiation safety parameters, and its performance is substantially equivalent to predicate devices.

8. The sample size for the training set

Not applicable. This submission is for an X-ray imaging system, not an AI or machine learning algorithm that requires a training set.

9. How the ground truth for the training set was established

Not applicable. As per point 8, there was no training set.

{0}------------------------------------------------

K103476

JUN 1 6 2011

GE Healthcare

510(k) Premarket Notification Submission Brivo XR285amx, Optima XR200amx, Optima XR220amx

510(k) Summary י

In accordance with 21 CFR 807.92 the following summary of information is provided:
Date:November 23, 2010
Submitter:GE Healthcare, (GE Medical Systems, LLC)
3000 N. Grandview Blvd.
Waukesha, WI 53188
Primary Contact Person:Nidhi Chaudhary
Regulatory Affairs Leader, X-Ray
GE Healthcare, (GE Medical Systems, LLC)
Telephone: 414-721-2899; Fax: 414-918-8184
e-mail: Nidhi.Chaudhary@ge.com
Secondary Contact Person:David Blonski
Regulatory Affairs Director, X-Ray
GE Healthcare, (GE Medical Systems, LLC)
Telephone: 262-513-4072; Fax: 262-364-2509
e-mail: David.Blonski@ge.com
Device Identification:
Trade Name:Brivo XR285amx ; Optima XR200amx ; Optima XR220amx
Common/Usual Name:Brivo XR285amx – Analog Mobile
Optima XR200amx - Digital Ready Mobile
Optima XR220amx - Digital Mobile
Classification Names:Brivo XR285amx ; Optima XR200amx ; Optima XR220amx
Product Code:Class II, IZL, System, X-ray, Mobile, 21 CFR 892.1720
Predicate Device(s):AMX-4 Plus Mobile X-ray system K021016
GE Definium AMX700, Model AMX 700 K052897
WIRELESS DR IMAGING OPTION WDR1 K102615 (pending)

Image /page/0/Picture/6 description: The image shows the General Electric (GE) logo. The logo consists of the letters "GE" intertwined within a circular border. The letters are stylized with curved lines and a vintage aesthetic. The image is in black and white, with the logo appearing in black against a white background.

{1}------------------------------------------------

Image /page/1/Picture/0 description: The image shows the General Electric (GE) logo. The logo consists of the letters 'G' and 'E' intertwined in a stylized, cursive font. The letters are enclosed within a circular border that has a textured or patterned design. The logo is black and white.

GE Healthcare 510(k) Premarket Notification Submission Brivo XR285amx, Optima XR200amx, Optima XR220amx

Device Description:

The GE Automatic Mobile X-Ray (AMX) Series: Brivo XR285amx, Optima XR200amx, Optima XR220amx are intended to take exposures utilizing film or computed radiography (CR), however the Optima XR220amx utilizes the GE Wireless Detector, which is intended to replace radiographic film screen systems in all general purpose diagnostic procedures, for digital radiography (DR).

Brivo XR285amx, Optima XR200amx, Optima XR220amx are self-contained; battery operated mobile radiographic imaging systems designed to generate diagnostic radiographic images (medical x-rays) that may increase the ability to detect disease or injury early enough for a medical problem to be managed, treated, or cured. Medical x-rays are used in many types of examinations and procedures, some examples include: x-ray radiography (to find orthopedic damage, tumors, pneumonias, foreign objects).

The series are indicated for use on adult and pediatric patients for general-purpose diagnostic radiographic examinations and procedures. Its mobility enables general-purpose radiographic procedures throughout the clinical environment, or as needed within the emergency, intensive care, premature birth ward, cardiac and operating departments, for patients that may not be able to be moved or in cases where it is unsafe or impractical to move them to a traditional RAD room.

The flat panel detector provides increased functionality to enable images of patients of all sizes, and can produce comparable quality images with as little as half the dose of traditional computer radiography (CR), cassettes and other flat panel detectors with lower DQE. The digital detector is designed to withstand a distributed load of 352Lbs to accommodate certain larger patients.

The systems are indicated for taking radiographic exposures of the skull, spinal column, chest, abdomen, extremities, and other body parts with the patient sitting, standing, or lying in the prone or supine position.

These devices are not intended for mammographic applications.

{2}------------------------------------------------

Image /page/2/Picture/0 description: The image shows the General Electric (GE) logo. The logo consists of the letters 'G' and 'E' intertwined in a stylized, cursive font. The letters are surrounded by a circular border with a swirling pattern inside.

GE Healthcare 510(k) Premarket Notification Submission Brivo XR285amx, Optima XR200amx, Optima XR220amx

Indications for Use:

The GE Automatic Mobile X-Ray (AMX) Series: Brivo XR285amx, Optima XR200amx, Optima XR220amx are intended to take exposures utilizing film or computed radiography (CR), however the Optima XR220amx utilizes the GE Wireless Detector, which is intended to replace radiographic film screen systems in all general purpose diagnostic procedures, for digital radiography (DR).

Brivo XR285amx, Optima XR200amx, Optima XR220amx are self-contained; battery operated mobile radiographic imaging systems designed to generate diagnostic radiographic images (medical x-rays) that may increase the ability to detect disease or injury early enough for a medical problem to be managed, treated, or cured. Medical x-rays are used in many types of examinations and procedures, some examples include: x-ray radiography (to find orthopedic damage, tumors, pneumonias, foreign objects).

The series are indicated for use on adult and pediatric patients for general-purpose diagnostic radiographic examinations and procedures. Its mobility enables general-purpose radiographic procedures throughout the clinical environment, or as needed within the emergency, intensive care, premature birth ward, cardiac and operating departments, for patients that may not be able to be moved or in cases where it is unsafe or impractical to move them to a traditional RAD room.

The systems are indicated for taking radiographic exposures of the skull, spinal column, chest, abdomen, extremities, and other body parts with the patient sitting, standing, or lying in the prone or supine position.

These devices are not intended for mammographic applications.

The systems are indicated for taking radiographic exposures of the skull, spinal column, chest, abdomen, extremities, and other body parts with the patient sitting, standing, or lying in the prone or supine position.

These devices are not intended for mammographic applications.

Technology:

The Brivo XR285amx, Optima XR200amx, Optima XR220amx employs the same fundamental scientific technology as its predicate devices.

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the General Electric (GE) logo. The logo consists of the letters 'GE' intertwined within a circular border. The letters are stylized with a flowing, cursive design, and the entire logo has a textured, slightly blurred appearance.

GE Healthcare

. 510(k) Premarket Notification Submission Brivo XR285amx, Optima XR200amx, Optima XR220amx

Determination of Substantial Equivalence:

Summary of Non-Clinical Tests:

The Brivo XR285amx, Optima XR200amx, Optima XR220amx and its applications comply with voluntary standards as detailed in Section 9, 11 and 17 of this premarket submission. The following quality assurance measures were applied to the development of the system:

  • . Risk Analysis
  • . Requirements Reviews
  • Design Reviews
  • . Testing on unit level (Module verification)
  • Integration testing (System verification) .
  • . Performance testing (Verification)
  • . Safety testing (Verification)
  • Simulated use testing (Validation) �

Summary of Clinical Tests:

The subject of this premarket submission, [Brivo XR285amx, Optima XR200amx, Optima XR220amx], did not require clinical studies to support substantial equivalence.

Comparison with Predicate Devices:

The Brivo and Optima Systems are updated versions of the AMX4+ and Definium AMX700 products. The primary changes from the predicates are due to improved technology Ainze the AMX4+ and AMX700 products were designed. Changes involve reduced overall size and weight of the product, improved charging efficiency, implementation of a dual focal sportune, and higher power generator. Additionally in comparison to the AMX4+ product, the Frivo system has an integrated touch screen display for customer input. Modifications to implement these updates involve hardware, software and firmware changes. The Brivo and Optima systems are all for use in mobile x-ray radiology and have similar indications for use. Changes made to achieve the new system designs were made in hardware, software and firmware, but there is no fundamental change in the use of the product for mobile x-ray, nor in technology. It has the same technological characteristics related to safety and effectiveness as the bredicate devices. A review of all bench and standards testing indicate that the new device provides no new safety concerns and is as safe an effective as the predicate devices. The Brivo and Ordima systems are certified to comply with the X-ray requirements of 21 CFR, arreautic spafety requirements of IEC 60601-1 and associated collateral and particular standards.

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows the General Electric (GE) logo. The logo consists of the letters 'G' and 'E' intertwined in a stylized, circular design. The letters are surrounded by a circular border with a decorative, swirling pattern.

GE Healthcare

510(k) Premarket Notification Submission Brivo XR285amx, Optima XR200amx, Optima XR220amx

Adverse Effects on Health:

The device has been evaluated for electrical, mechanical, and radiation safety, and conforms to applicable medical device safety standards, as confirmed by a Nationally Recognized Test Laboratory.

The potential hazards of electrical and mechanical are identified in a risk management summary (hazard analysis) and controlled by:

  • System verification and validation to ensure performance to specifications,
    Federal Regulations, and user requirements.

  • Adherence and certification to industry and international standards.

  • Compliance to applicable CDRH 21 CFR subchapter J requirements.

The device is designed and manufactured under the Quality System Regulations of 21CFR 820.

Conclusion:

GE Healthcare considers the Brivo XR285amx, Optima XR200amx, Optima XR220amx to be as safe, as effective, and performance is substantially equivalent to the predicate device(s).

The Brivo and Optima systems are an improvement in design technology, and customer needs. It does not result in any new potential safety risks, has the same technological characteristics, and performs as well or better then devices currently on the market. The Brivo and Optima Systems will be certified to comply with the X-ray requirements of 21 CFR, as well as safety requirements of IEC 60601-1 and associated collateral and particular standards.

After analyzing standards testing and bench data, it is the conclusion of GE Healthcare that the Brivo and Optima Systems are substantially equivalent to other marketed devices with similar indications for use and meeting the same standards.

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image is a black and white logo for the U.S. Department of Health and Human Services. The logo features a stylized abstract symbol resembling a bird or a person with outstretched arms. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES-USA" is arranged in a circular fashion around the symbol.

DEPARTMENT OF HEALTH & HUMAN SERVICES

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room - WO66-G609 Silver Spring, MD 20993-0002

Ms. Nidhi Chaudhary Regulatory Affairs Leader, X-Ray GE Medical Systems, LLC 3200 N. Grandview Blvd. WAUKESHA WI 51388

JUN 1 6 201

Re: K103476

Trade/Device Name: Brivo XR285amx, Optima XR200amx, Optima XR220amx Regulation Number: 21 CFR 892.1720 Regulation Name: Mobile x-ray system Regulatory Class: II Product Code: IZL Dated: May 1, 2011 Received: May 13, 2011

Dear Ms. Chaudhary:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into class II (Special Controls), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of

{6}------------------------------------------------

medical device-related adverse events) (21 CFR 803); and good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820). This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Parts 801 and 809), please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (301) 796-5450. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely Yours.

Mary Pastel

Mary S. Pastel, Sc.D. Director Division of Radiological Devices Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health

Enclosure

{7}------------------------------------------------

Image /page/7/Picture/0 description: The image shows the General Electric (GE) logo. The logo consists of the letters "G" and "E" intertwined in a stylized, circular design. The letters are surrounded by a textured, circular border.

GE Healthcare

510(k) Premarket Notification Submission Brivo XR285anx, Optima XR200amx, Optima XR220amx

510(k) Number (if known):

Device Name: Family:

Brivo XR285amx, Optima XR200amx, Optima XR220amx Brivo XR285amx - Analog Mobile Optima XR200amx - Digital Ready Mobile Optima XR220amx - Digital Mobile

Indications for Use:

The GE Automatic Mobile X-Ray (AMX) Series: Brivo XR285amx, Optima XR200amx, Optima XR220amx are intended to take exposures utilizing film or computed radiography (CR), however the Optima XR220amx utilizes the GE Wireless Detector, which is intended to replace radiographic film screen systems in all general purpose diagnostic procedures, for digital . radiography (DR).

Brivo XR285amx, Optima XR200amx, Optima XR220amx are self-contained: battery operated mobile radiographic imaging systems designed to generate diagnostic radiographic images (medical x-rays) that may increase the ability to detect disease or injury early enough for a medical problem to be managed, treated, or cured. Medical x-rays are used in many types of examinations and procedures, some examples include: x-ray radiography (to find orthopedic damage, tumors, pneumonias, foreign objects).

The series are indicated for use on adult and pediatric patients for general-purpose diagnostic radiographic examinations and procedures. Its mobility enables general-purpose radiographic procedures throughout the clinical environment, or as needed within the emergency, intensive care, premature birth ward, cardiac and operating departments, for patients that may not be able to be moved or in cases where it is unsafe or impractical to move them to a traditional RAD room.

The systems are indicated for taking radiographic exposures of the skull, spinal column, chest, abdomen, extremities, and other body parts with the patient sitting, standing, or lying in the prone or supine position.

These devices are not intended for mammographic applications.

Prescription Use X AND/OR Over-The-Counter Use (Part 21 CFR 801 Subpart D) (Part 21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD)

mun Division Sign-Off Office of In Vitro Diagnostic Device Evaluation and Safety 510(k) 1102476

§ 892.1720 Mobile x-ray system.

(a)
Identification. A mobile x-ray system is a transportable device system intended to be used to generate and control x-ray for diagnostic procedures. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.(b)
Classification. Class II.