(70 days)
The Wireless DR Imaging Option -WDR1, when used with a radiographic imaging system, is indicated for use in generating radiographic images of human anatomy. It is intended for use in all routine radiography exams, and specialized areas including pediatric work, intensive care and trauma, wherever conventional screen-film systems or CR systems maybe used.
The device is not intended for mammography or fluoroscopy applications.
The Wireless DR Imaging Option - WDR1 consists of a wireless detector, system interface box, computer, display, keyboard and mouse. It is designed to acquire digital radiographic images when used with existing radiographic x-ray systems. Images captured can be communicated to the system via wireless signal, tethered cable or direct connection (docked). The Wireless detector is very similar to the digital detector currently in use as the GE Digital Radiographic Detector (K982196), and the Philips Medical Systems Wireless Portable Detector FD-W17 (K090625) comprised of amorphous silicon with a cesium iodide scintillator.
It is used on radiographic X-ray systems as a substitute for film/screen systems or CR systems. The Radiographic X-ray system remains unchanged except for the replacement of the systems receptor with the Wireless WDR 1 detector.
The provided text describes a 510(k) premarket notification for the GE Wireless DR Imaging Option - WDR1. This is largely a declaration of substantial equivalence to existing devices rather than a detailed study demonstrating acceptance criteria for the device's diagnostic performance. Therefore, many of the requested points related to a diagnostic performance study are not explicitly present in the provided document.
However, based on the information provided, here's what can be extracted and inferred:
1. A table of acceptance criteria and the reported device performance
The document does not explicitly present a table of acceptance criteria for diagnostic accuracy or clinical performance in terms of disease detection. Instead, the acceptance criteria are focused on safety and technological equivalence to predicate devices. The "reported device performance" is essentially that it meets these safety and equivalence standards.
| Acceptance Criteria Category | Specific Criteria | Reported Device Performance |
|---|---|---|
| Technological Equivalence | - No fundamental change in X-ray detection technology or image creation. - Uses virtually identical operating principles to predicate device (GE Digital Radiographic Detector K982196). - Same technological characteristics related to safety and effectiveness as predicate devices. | - The Wireless DR Imaging Option - WDR1 is an enhanced wireless version of and substantially equivalent to the GE Digital Radiographic Detector (K982196) and Philips Medical Systems wireless detector FD-W17 (K090625). - Changes involve wireless functionality, reduced weight, power, and design modifications (hardware, software, firmware), but not fundamental X-ray detection technology. |
| Safety | - No new potential safety risks compared to predicate devices. - Complies with X-ray requirements of 21 CFR. - Complies with safety requirements of IEC 60601-1 and associated collateral and particular standards. - Device evaluated for electrical, mechanical, and radiation safety. - Controlled potential hazards through risk management (hazard analysis), system verification/validation, adherence to industry/international standards, and compliance to CDRH 21 CFR subchapter J. - Designed and manufactured under 21 CFR 820 Quality System Regulations. | - "No new safety concerns and is as safe an effective as the predicate devices." - Certified to comply with 21 CFR X-ray requirements and IEC 60601-1 standards. - Evaluated for electrical, mechanical, and radiation safety. - Conforms to applicable medical device safety standards. |
| Performance (Bench) | - Performs as well as devices currently on the market. | - "Performs as well as the devices currently on the market." (Based on standards testing and bench data). |
| Indications for Use | - Similar indications for use as predicate devices. | - Indicated for use "in generating radiographic images of human anatomy" and "all routine radiography exams, and specialized areas including pediatric work, intensive care and trauma." - Not intended for mammography or fluoroscopy. (Consistent with general radiography detectors). |
2. Sample size used for the test set and the data provenance
The document refers to "bench and standards testing" and "bench data," but does not specify a sample size for a clinical test set or images. It also does not provide details on the data provenance (e.g., country of origin, retrospective or prospective) for any image-based testing. This is typical for a 510(k) focused on technological equivalence rather than new clinical claims.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts
This information is not provided in the document. As no clinical performance study with image interpretation is detailed, there's no mention of experts or ground truth establishment in that context.
4. Adjudication method for the test set
This information is not provided in the document, as no clinical performance study involving multiple readers or complex diagnostic interpretation requiring adjudication is described.
5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
No MRMC comparative effectiveness study is described. This device is a digital radiography detector, not an AI-powered diagnostic tool. Therefore, questions about human reader improvement with AI assistance are not applicable.
6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done
This device is hardware (a detector), not a standalone algorithm. Therefore, "standalone" algorithm performance is not applicable in this context. The document refers to its performance in acquiring digital radiographic images.
7. The type of ground truth used
For the safety and technological equivalence claims, the "ground truth" implicitly refers to:
- Predicate device characteristics and performance: The known technological aspects, safety profile, and output of the GE Digital Radiographic Detector (K982196) and Philips FD-W17 (K090625).
- Industry and regulatory standards: IEC 60601-1, 21 CFR X-ray requirements, and 21 CFR 820 Quality System Regulations.
- Bench test results: Outputs from various technical tests that confirm the device's electrical, mechanical, and radiation safety properties, and its ability to produce digital images.
There is no mention of "expert consensus, pathology, or outcomes data" being used as ground truth for a clinical diagnostic performance study.
8. The sample size for the training set
This is not applicable as the device is hardware and not an AI algorithm that undergoes training.
9. How the ground truth for the training set was established
This is not applicable for the same reason as point 8.
{0}------------------------------------------------
K102615
P. 1 of 3
NOV 1 8 2010
GE Medical Systems, LLC 510(k) Premarket Notification Wireless DR Imaging Option WDRI
510(K) Summary of Safety and Effectiveness
Prepared in accordance with 21 CFR Part 807.92.
| Date Prepared: | September 3, 2010 |
|---|---|
| Submitter: | GE Healthcare, (GE Medical Systems, LLC)3000 N. Grandview Blvd.Waukesha, WI 53188 |
| Primary Contact : | John L. SchmidtRegulatory Affairs Leader, X-RayTelephone: 262-548-4964; Fax: 262-548-2032e-mail: John.L.Schmidt@ge.com |
| Secondary Contact : | David Blonski.Regulatory Affairs Director, X-RayGE Healthcare, (GE Medical Systems, LLC)Telephone: 262-513-4072; Fax: 262-548-2032e-mail: David.Blonski@ge.com |
DEVICE INDENTIFICATION
| Trade Name: | GE Wireless DR Imaging Option – WDR1 |
|---|---|
| Common/Usual Name: | Solid State X-Ray Imager (Flat Panel/Digital Imager) |
| Classification Name: | Solid State X-ray Imager (flat panel/digital imager)Class II, MQB, 21CFR 892.1650 |
DEVICE DESCRIPTION:
The Wireless DR Imaging Option - WDR1 consists of a wireless detector, system interface box, computer, display, keyboard and mouse. It is designed to acquire digital radiographic images when used with existing radiographic x-ray systems. Images captured can be communicated to the system via wireless signal, tethered cable or direct connection (docked). The Wireless detector is very similar to the digital detector currently in use as the GE Digital Radiographic Detector (K982196), and the Philips
{1}------------------------------------------------
K102615
P. 2 of 3
GE Medical Systems, LLC 510(k) Premarket Notification Wireless DR Imaging Option WDRI
Medical Systems Wireless Portable Detector FD-W17 (K090625) comprised of amorphous silicon with a cesium iodide scintillator.
It is used on radiographic X-ray systems as a substitute for film/screen systems or CR systems. The Radiographic X-ray system remains unchanged except for the replacement of the systems receptor with the Wireless WDR 1 detector.
INDICATIONS FOR USE:
The Wireless DR Imaging Option -WDRI, when used with a radiographic imaging system, is indicated for use in generating radiographic images of human anatomy. It is intended for use in all routine radiography exams, and specialized areas including pediatric work, intensive care and trauma, wherever conventional screen-film systems or CR systems maybe used.
The device is not intended for mammography or fluoroscopy applications.
COMPARISON WITH PREDICATE DEVICES:
The Wireless DR Imaging Option - WDR1 is an enhanced wireless version of and substantially equivalent to, the GE Digital Radiographic Detector cleared under 510(k) number K982196 and Philips Medical Systems wireless detector FD-W17 (K090625). It involves changes from the predicate device (K982196) to add wireless functionality, reduced weight and power, and modifications in design that involve hardware, software, and firmware. The Wireless DR Imaging Option WDR1 and the predicate devices are both for use in acquiring digital radiographic images and have similar indications for use.
Changes in electronics, hardware and firmware have been made to affect these power and weight reductions, but there is no fundamental change in X-ray detection technology, or image creation, and uses virtually identical operating principles to the GE Digital Radiographic Detector currently marketed system (K982196). It has the same technological characteristics related to safety and effectiveness as the predicate devices. A review of all bench and standards testing indicate that the new device provides no new safety concerns and is as safe an effective as the predicate devices.
The Wireless DR Imaging Option - WDR1 is certified to comply with the X-ray requirements of 21 CFR, as well as safety requirements of IEC 60601-1 and associated collateral and particular standards.
{2}------------------------------------------------
K102615
P. 3 of 3
Image /page/2/Picture/1 description: The image shows the General Electric (GE) logo. The logo consists of the letters "G" and "E" intertwined within a circular border. The logo is black and white and has a vintage look.
GE Medical Systems, LLC 510(k) Premarket Notification Wireless DR Imaging Option WDR 1
ADVERSE EFFECTS ON HEALTH:
The device has been evaluated for electrical, mechanical, and radiation safety, and conforms to applicable medical device safety standards, as confirmed by a Nationally Recognized Test Laboratory.
The potential hazards of electrical and mechanical are identified in a risk management summary (hazard analysis) and controlled by:
- . System verification and validation to ensure performance to specifications, Federal Regulations, and user requirements.
- Adherence and certification to industry and international standards. .
- Compliance to applicable CDRH 21 CFR subchapter J requirements. t
The device is designed and manufactured under the Quality System Regulations of 21 CFR 820.
CONCLUSION:
The Wireless DR Imaging Option - WDR1 is an evolutionary modification to the GE Digital Radiographic Detector. It does not result in any new potential safety risks, has the same technological characteristics, and performs as well as the devices currently on the market. The Wireless DR Imaging Option WDR1 will be certified to comply with the X-ray requirements of 21 CFR, as well as safety requirements of IEC 60601-1 and associated collateral and particular standards.
After analyzing standards testing and bench data, it is the conclusion of GE Healthcare that the Wireless DR Imaging Option - WDR1 is substantially equivalent to other marketed devices with similar indications for use and meeting the same standards.
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image is a black and white logo for the Department of Health & Human Services USA. The logo consists of a circular border with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" written around it. Inside the circle is a stylized image of an eagle or bird-like figure with outstretched wings.
DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room - WO66-G609 Silver Spring, MD 20993-0002
Mr. John J. Schmidt Regulatory Affairs Leader, X-Ray GE Medical Systems, LLC 3000 N. Grandview Blvd. WAUKESHO WI 53188
AUG 2 3 2013
Re: K102615
Trade/Device Name: Wireless DR Imaging Option-WDR1 Regulation Number: 21 CFR 892.1680 Regulation Name: Stationary x-ray system Regulatory Class: II Product Code: MOB Dated: September 2, 2010 Received: September 9, 2010
Dear Mr. Schmidt:
This letter corrects our substantially equivalent letter of November 18, 2010.
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate 10 rase surved in the encreativent of the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The I ou may, arones provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into class II (Special Controls), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of
{4}------------------------------------------------
medical device-related adverse events) (21 CFR 803); and good manufacturing pactice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820). This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Parts 801 and If you desire specific acritic for your as not is not is not is Evaluation and Safety at (301) 796-5450. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Tod inay of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely Yours,
Janine M. Morris
Acting Director Division of Radiological Devices Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
K102615
P. 1 of 1
GE Medical Systems, LLC 510(k) Premarket Notification Wireless DR Imaging Option WDR I
Indications for Use
NOV 1 8 2010
510(k) Number (if Known): K | 0 26 15 Device Name: Wireless DR Imaging Option - WDR I
Indications for Use:
The Wireless DR Imaging Option -WDR1, when used with a radiographic imaging system. is indicated for use in generating radiographic images of human anatomy. It is intended for use in all routine radiography exams, and specialized areas including pediatric work, intensive care and trauma, wherever conventional screen-film systems or CR systems maybe used.
The device is not intended for mammography or fluoroscopy applications.
Prescription Use_X (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Usc (Part 21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRA, Office of In Vitro Diagnostic Devices (OIVD)
Division Sign-Off
Office of In Vitro Diagnostic Device Evaluation and Safety
510(k) K102615
Page 1 of 1
§ 892.1680 Stationary x-ray system.
(a)
Identification. A stationary x-ray system is a permanently installed diagnostic system intended to generate and control x-rays for examination of various anatomical regions. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.(b)
Classification. Class II (special controls). A radiographic contrast tray or radiology diagnostic kit intended for use with a stationary x-ray system only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.