K Number
K102778
Manufacturer
Date Cleared
2011-09-16

(357 days)

Product Code
Regulation Number
864.5260
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

DM1200 is an automated system intended for in-vitro diagnostic use.

The body fluid application is intended for differential count of white blood cells. The system automatically locates and presents images of cells on cytocentrifuged body fluid preparations. The operator identifies and verifies the suggested classification of each cell according to type.

DM1200 is intended to be used by skilled operators, trained in the use of the device and in recognition of blood cells.

Device Description

CellaVision DM1200 with the body fluid application automatically locates and presents images of nucleated cells on cytocentrifuged body fluid preparations. The system suggests a classification for each cell and the operator verifies the classification and has the opportunity to change the suggested classification of any cell.
The system preclassifies to the following WBC classes: Unidentified, Neutrophils, Eosinophils, Lymphocytes, Macrophages (including Monocytes) and Other. Cells preclassified as Basophils, Lymphoma cells, Atypical lymphocytes, Blasts and Tumor cells are automatically forwarded to the cell class Other.
Unidentified is a class for cells and objects which the system has pre-classified with a low confidence level.

AI/ML Overview

Here's an analysis of the provided text, outlining the acceptance criteria and study details for the CellaVision DM1200 with the body fluid application:


Acceptance Criteria and Device Performance Study for CellaVision DM1200 with Body Fluid Application

The CellaVision DM1200 with body fluid application is an automated cell-locating device intended for in-vitro diagnostic use, specifically for the differential count of white blood cells in cytocentrifuged body fluid preparations. The system automatically locates and presents cell images, suggests a classification, and requires operator verification.

1. Table of Acceptance Criteria and Reported Device Performance

The provided document does not explicitly state pre-defined "acceptance criteria" as pass/fail thresholds for accuracy or precision. Instead, it presents the results of a method comparison study between the CellaVision DM1200 (Test Method) and its predicate device, CellaVision DM96 (Reference Method), for various leukocyte classifications. The implied acceptance criteria are that the DM1200 demonstrates comparable accuracy and precision to the predicate device.

Cell ClassAcceptance Criteria (Implied: Comparable to Predicate)Reported Device Performance (CellaVision DM1200 vs. DM96)
Accuracy (Regression Analysis: DM1200 = Slope * DM96 + Intercept)
NeutrophilsRegression where slope is close to 1 and intercept close to 0, with high R²y = 0.9969x + 0.0050, R² = 0.9932 (95% CI Slope: 0.9868-1.0070, 95% CI Intercept: 0.0004-0.0096)
LymphocytesRegression where slope is close to 1 and intercept close to 0, with high R²y = 0.9815x + 0.0016, R² = 0.9829 (95% CI Slope: 0.9656-0.9973, 95% CI Intercept: -0.0049-0.0081)
EosinophilsRegression where slope is close to 1 and intercept close to 0, with high R²y = 1.1048x - 0.0002, R² = 0.9629 (95% CI Slope: 1.0782-1.1314, 95% CI Intercept: -0.0007-0.0003)
MacrophagesRegression where slope is close to 1 and intercept close to 0, with high R²y = 1.0067x - 0.0050, R² = 0.9823 (95% CI Slope: 0.9901-1.0232, 95% CI Intercept: -0.0125-0.0024)
Other cellsRegression where slope is close to 1 and intercept close to 0, with high R²y = 0.9534 + 0.0032, R² = 0.9273 (95% CI Slope: 0.9207-0.9861, 95% CI Intercept: -0.0002-0.0065)
Precision/Reproducibility (Short-term Imprecision)
NeutrophilsSD % comparable between test and reference methodTest Method: Mean % 32.0, SD % 3.2; Reference Method: Mean % 31.6, SD % 3.4
LymphocytesSD % comparable between test and reference methodTest Method: Mean % 30.1, SD % 5.6; Reference Method: Mean % 30.5, SD % 5.7
EosinophilsSD % comparable between test and reference methodTest Method: Mean % 0.6, SD % 0.7; Reference Method: Mean % 0.5, SD % 0.6
MacrophagesSD % comparable between test and reference methodTest Method: Mean % 35.3, SD % 5.8; Reference Method: Mean % 35.5, SD % 6.2
Other cellsSD % comparable between test and reference methodTest Method: Mean % 2.1, SD % 1.7; Reference Method: Mean % 1.9, SD % 2.5

The conclusion states that the short-term imprecision was found to be equivalent for the test method and the reference method, and the accuracy results (high R-squared values, slopes close to 1, and intercepts close to 0) demonstrate substantial equivalence.

2. Sample Size Used for the Test Set and Data Provenance

  • Sample Size for Test Set: 260 samples.
    • CSF: 62 samples
    • Serous fluid: 151 samples
    • Synovial fluid: 47 samples
  • Data Provenance:
    • Country of Origin: Not explicitly stated, but samples were collected from "two sites." Given the submitter is in Sweden, and the regulatory contact is in the USA, it's unclear if these sites were in Sweden, the USA, or elsewhere.
    • Retrospective or Prospective: Not explicitly stated, but the description "collected from two sites" and then analyzed suggests a prospective collection or at least fresh samples for the study.

3. Number of Experts and Qualifications for Ground Truth

  • Number of Experts: Not explicitly stated for the establishment of ground truth for the test set's initial classifications. However, the document mentions:
    • "The results were then verified by skilled human operators." This indicates human review post-analysis by both the test and reference methods.
    • The "Intended Use" section states: "DM1200 is intended to be used by skilled operators, trained in the use of the device and in recognition of blood cells."
  • Qualifications of Experts: "Skilled human operators, trained in the use of the device and in recognition of blood cells." No specific professional qualifications (e.g., "radiologist with 10 years of experience") are provided.

4. Adjudication Method for the Test Set

The document states: "The results were then verified by skilled human operators." It does not specify a multi-reader adjudication method like 2+1 or 3+1. It implies a single operator verification for each result generated by both the test and reference methods.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

No, a multi-reader multi-case (MRMC) comparative effectiveness study designed to measure the effect size of how much human readers improve with AI vs. without AI assistance was not explicitly described. This study was a method comparison between two devices (one with the new application, one being the predicate) with human verification. The device's function is to suggest classifications, which implies an assistive role, but the study design was not an MRMC study comparing human performance with and without AI.

6. Standalone (Algorithm Only) Performance Study

Yes, a standalone performance was implicitly done. The "Test Method" (CellaVision DM1200) "suggests a classification for each cell," meaning the algorithm performs an initial classification without human intervention. The reported accuracy metrics ($R^2$, slope, intercept) compare these suggested classifications to those obtained by the reference method (DM96, which also involves algorithmic preclassification). However, the study concludes with human verification of these results, suggesting the standalone performance without the "human-in-the-loop" step described in the intended use is not the final reported performance. The "Accuracy results" table (Table 3.3) and "Precision/Reproducibility" table (Table 3.4) reflect the device's performance before the final human verification step that might change classifications.

7. Type of Ground Truth Used

The ground truth used for the comparison was established by the predicate device (CellaVision DM96) with its own human verification, after undergoing "a 200-cell differential count... with both the test method and the reference method. The results were then verified by skilled human operators." Therefore, it's a form of expert-verified reference measurement. It is not pathology, or outcomes data.

8. Sample Size for the Training Set

The document does not explicitly state the sample size for the training set used to develop the CellaVision DM1200's classification algorithms. It mentions "deterministic artificial neural networks (ANN's) trained to distinguish between classes of white blood cells," but no details about the training data are provided within this summary.

9. How the Ground Truth for the Training Set Was Established

The document does not explicitly describe how the ground truth for the training set was established. It states that the ANNs were "trained to distinguish between classes of white blood cells," implying that a labeled dataset was used for training, but the process of creating these labels (e.g., expert consensus, manual review) is not detailed in this 510(k) summary.

{0}------------------------------------------------

K102778

1 510(k) Summary

،

100 - 100 -

Submitter:CellaVision ABIdeon Science ParkSE-223 70 LundSwedenPhone: +46 46 286 44 00Fax: +46 46 286 44 70SEP 16 2011
ContactInformation:C. G. Bundy Associates, Inc.435 Rice Creek Terrace NEFridley, MN 55432USAPhone: 763-574-1976Fax: 763-571-2437
Submission Date:September 20, 2010
Device Name:Proprietary name: CellaVision® DM1200 with the body fluidapplicationCommon/Classification name: Automated cell-locating device
ClassificationRegulation:21 CFR 864.5220 and 21 CFR 864.5260 Class II medical devices.
Equivalent DeviceIdentification:CellaVision AB believes that CellaVision DM1200 with the body fluidapplication is substantially equivalent to the legally marketedCellaVision DM96 with the body fluid application (K080595).
DeviceDescription:CellaVision DM1200 with the body fluid application automaticallylocates and presents images of nucleated cells on cytocentrifuged bodyfluid preparations. The system suggests a classification for each celland the operator verifies the classification and has the opportunity tochange the suggested classification of any cell.The system preclassifies to the following WBC classes: Unidentified,Neutrophils, Eosinophils, Lymphocytes, Macrophages (includingMonocytes) and Other. Cells preclassified as Basophils, Lymphomacells, Atypical lymphocytes, Blasts and Tumor cells areautomatically forwarded to the cell class Other.
Unidentified is a class for cells and objects which the system has pre-classified with a low confidence level.
Intended Use:DM1200 is an automated system intended for in vitro diagnostic use.The body fluid application is intended for differential count of whiteblood cells. The system automatically locates and presents images ofcells on cytocentrifuged body fluid preparations. The operator identifies

{1}------------------------------------------------

and verifies the suggested classification of each cell according to type.

DM1200 is intended to be used by skilled operators, trained in the use of the device and in recognition of blood cells.

{2}------------------------------------------------

CharacteristicDM96 with the body fluidDM1200 with the bodyEquivalent
application (predicate - )fluid application
device)
Intended useAutomated cell-locatingAutomated cell-locating
device for cell-location anddevice for cell-location and
identification of white bloodidentification of white blood
cells in cytocentrifuged bodycells in cytocentrifuged body
fluid preparations, for in-vitrofluid preparations, for in-
diagnostic use.vitro diagnostic use.
Verification of results byVerification of results by
skilled human operator.skilled human operator.
Specimen typeBody fluids.Body fluids.
SampleBody fluid samples areBody fluid samples areV
preparationprepared using aprepared using a
cytocentrifuge and stainedcytocentrifuge and stained
with Romanowsky stain.with Romanowsky stain.
AnalysisWhite blood cells:White blood cells:
techniqueNucleated cells areNucleated cells are
located/counted by movinglocated/counted by moving
according to the battlementaccording to the battlement
track pattern. Cell images aretrack pattern. Cell images are
analyzed using standardanalyzed using standard
mathematical methods,mathematical methods,
including deterministicincluding deterministic
artificial neural networksartificial neural networks
(ANN's) trained to(ANN's) trained to
distinguish between classesdistinguish between classes
of white blood cells.of white blood cells.
The cell images are
preclassified by the systemThe cell images arepreclassified by the system
and the operator verifies theand the operator verifies the
suggested classification bysuggested classification by
either accepting oreither accepting or
reclassifying.reclassifying.
OverviewThe device presents anThe device presents an
imageoverview image. The imageoverview image. The image
gives the operatorgives the operator
possibilities to get anpossibilities to get an
overview on parts of or theoverview on parts of or the
whole slide in differentwhole slide in different
magnifications.magnifications.

:

·

.

Table 3.1: Comparative features of CellaVision DM1200 with the body fluid application and the predicate device

{3}------------------------------------------------

Performance characteristics:

A method comparison was conducted to compare CellaVision DM1200 with the body fluid application (Test Method) to CellaVision DM96 with the body fluid application (Reference Method).

Body fluid samples were collected from two sites. All samples were initially analyzed on a cell counter or counted in a hemocytometer to get the leukocyte concentration. From each sample two cytocentrifuged preparations were made. A 200-cell differential count was performed for each slide (400 cells/sample) with both the test method and the reference method. The results were then verified by skilled human operators. The comparison was based on the approved guideline, CLSI document EP9-A2.

Accuracy:

The accuracy was tested through scatter plots for each cell class.

TypeNumber of samples
CSF62
Serous fluid151
Synovial fluid47
Total260

Table 3.2: Samples included in the study.

Accuracy results for all samples included were as follows:

Cell classAccuracy95% CI Slope95% CI InterceptNumber ofsamples
Neutrophils$y = 0.9969x + 0.0050$$R^2=0.9932$0.9868-1.00700.0004-0.0096260
Lymphocytes$y = 0.9815x + 0.0016$$R^2 = 0.9829$0.9656-0.9973-0.0049-0.0081260
Eosinophils$y = 1.1048x - 0.0002$$R^2 = 0.9629$1.0782-1.1314-0.0007-0.0003260
Macrophages$y = 1.0067x - 0.0050$$R^2 = 0.9823$0.9901-1.0232-0.0125-0.0024260
Other cells$y = 0.9534 + 0.0032$$R^2 = 0.9273$0.9207-0.9861-0.0002-0.0065260

Table 3.3; Accuracy results. All samples included (nooled data).

{4}------------------------------------------------

Precision/Reproducibility:

The table below shows the short-term imprecision results found for all 260 samples included in the clinical evaluation. The short-term imprecision was found to be equivalent for the test method and the reference method.

Test MethodReference Method
Mean % *SD %Mean % *SD %
Neutrophils32.03.231.63.4
Lymphocytes30.15.630.55.7
Eosinophils0.60.70.50.6
Macrophages35.35.835.56.2
Other cells2.11.71.92.5

Table 3.4: Short-term imprecision (pooled data).

$$\text{\textbullet \textquotesingle} \text{\textquotesingle} \text{\textquotesingle} \text{\textquotesingle} \text{\textquotesingle} \text{\textquotesingle} \text{\textquotesingle} \text{\textquotesingle} \text{\textquotesingle} \text{\textquotesingle} \text{\textquotesingle} \text{\textquotesingle} \text{\textquotesingle} \text{\textquotesingle} \text{\textquotesingle} \text{\textquotesingle} \text{\textquotesingle} \text{\textquotesingle} \text{\textquotesingle} \text{\textquotesingle} \text{\textquotesingle} \text{\textquotesingle} \text{\textquotesingle} \text{\textquotesingle} \text{\textquotesingle} \text{\textquotesingle} \text{\textquotesingle} \text{\textquotesingle} \text{\textquotesingle} \text{\textquotesingle} \text{\textquotesingle} \text{\textquotesingle} \text{\textquotesingle} \text{\textquotesingle} \text{\textquotesingle} \text{\textquotesingle} \text{\textquotesingle} \text{\textquotesingle} \text{\textquotesingle} \text{\textquotesingle} \text{\textquotesingle} \text{\textquotesingle} \text{\textquotesingle} \text{\textquotesingle} \text{\textquotesingle} \text{\textquotesingle} \text{\textquotesingle} \text{\textquotesingle} \text{\textquotesingle} \text{\textquotesingle} \text{\textquotesingle} \text{\textquotesingle} \text{\textquotesingle} \text{\textquotesingle} \text{\textquotesingle} \text{\textquotesingle} \text{\textquotesingle} \text{\textquotesingle} \text{\textquotesingle} \text{\textquotesingle} \text{\textquotesingle} \text{\textquotesingle} \text{\textquotesingle} \text{\textquotesingle} \text{\textquotesingle} \text{\textquotesingle} \text{\textquotesingle} \text{\textquotesingle}} \text{\textquotesing$$

Conclusion:

The body fluid application is a modification of the intended use of CellaVision DM1200 Automated Hematology Analyzer, cleared by FDA in 2009 (K092868). The intended use has been modified to also include presentation of white blood cells on cytocentrifuged body fluid preparations.

Based on extensive performance testing, including comparison to the predicate device, it is the conclusion of CellaVision AB that CellaVision DM1200 with the body fluid application is substantially equivalent to devices already on the market (cleared by the 510(k) process) and presents no new concerns regarding safety and effectiveness.

{5}------------------------------------------------

Image /page/5/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a stylized eagle with three tail feathers, representing the three levels of government: federal, state, and local. The eagle is encircled by the words "DEPARTMENT OF HEALTH & HUMAN SERVICES. USA" in a circular arrangement.

Food and Drug Administration 10903 New Hampshire Avenue Silver Spring, MD 20993

CellaVision AB c/o Ms. Constance G. Bundy C. G. Bundy Associates, Inc. 435 Rice Creek Terrace NE Fridley. MN 55432

P 1 6 2011

Re: K102778

Trade/Device Name: CellaVision® DM 1200 with body fluid application Regulation Number: 21 CFR 864.5260 Regulation Name: Automated cell-locating device Regulatory Class: Class II Product Code: JOY Dated: September 7, 2011 Received: September 9, 2011

Dear Ms. Bundy.

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into class II (Special Controls), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must a comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820). This letter

{6}------------------------------------------------

Page 2 - Ms. Constance G. Bundy

will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Parts 801 and 809), please contact the Office of In Viro Diagnostic Device Evaluation and Safety at (301) 796-5450. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportalProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely yours.

Reena Philip

on Maria M. Chan. Ph.D. Director Division of Immunology and Hematology Devices Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health

Enclosure

{7}------------------------------------------------

1 Indications for Use Statement

510(k) Number (if known): K102778

Device Name: CellaVision® DM1200 with the body fluid application

Indications for Use:

DM1200 is an automated system intended for in-vitro diagnostic use.

The body fluid application is intended for differential count of white blood cells. The system automatically locates and presents images of cells on cytocentrifuged body fluid preparations. The operator identifies and verifies the suggested classification of each cell according to type.

DM1200 is intended to be used by skilled operators, trained in the use of the device and in recognition of blood cells.

Prescription Use X (Part 21 CFR 801.109) OR

Over the Counter Use

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Page 1 of _l

Reena Philip

Division Sign-Off

Office of In Vitro Diagnos Device Evaluation and S

S10K K102778

§ 864.5260 Automated cell-locating device.

(a)
Identification. An automated cell-locating device is a device used to locate blood cells on a peripheral blood smear, allowing the operator to identify and classify each cell according to type. (Peripheral blood is blood circulating in one of the body's extremities, such as the arm.)(b)
Classification. Class II (performance standards).