K Number
K102713
Date Cleared
2011-01-25

(127 days)

Product Code
Regulation Number
878.4040
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Tiger SFM0001 surgical mask is intended for single use by operating room personnel and other health care workers to protect both the patients and the health care workers from transfer of microorganisms, blood and body fluids, and airborne particulate materials.

Device Description

The Tiger surgical mask SFM0001 (colors: blue) is flat pleated by 3-ply masks with outer layer and inner layers (spunbonded polypropylene) that sanwich a meltbown polypropylene filter material; Ear-loops are made of urethane elastic fiber for free elastic loops. The nose piece is white aluminum strip with PP covering. All of the materials used in the construction of the Tiger flat surgical Masks are being used in currently marketed devices (see predicate information). All items are nosterilize and only for single use.

AI/ML Overview

This submission describes the acceptance criteria and the study that proves the device meets the acceptance criteria for a Tiger Surgical Mask SFM0001.
The study is a series of non-clinical bench tests performed by accredited laboratories to demonstrate substantial equivalence to predicate devices.

1. A table of acceptance criteria and the reported device performance:

Performance CharacteristicsTest methodAcceptance criteria/result (Tiger Surgical Mask SFM0001)Predicate K061716 resultsPredicate K022256 results
Fluid resistanceASTM 186231 of 32 Pass at 120mmHg29 of 32 passNo visual penetration
Particulate Filtration EfficiencyASTM F2299>99.9% at 2.7µm2.0 microns98.5% at 0.1 microns
Bacterial Filtration EfficiencyASTM F210199.86%99.6%97.9%
Flammability Class16CFR 1610112
Delta-PMil M36954C3.782.61.8
BiocompatibilityISO10993-1Cytotoxicity: no cytotoxic potentialN/AN/A
Sensitization: negligibleN/AN/A
Irritation: No reactionsN/AN/A

2. Sample size used for the test set and the data provenance:

  • Fluid Resistance (ASTM F1862): 32 samples were tested. The data provenance is from Nelson Laboratories.
  • Particulate Filtration Efficiency (ASTM F2299): Specific sample size not explicitly stated in the provided text, but testing was performed by Nelson Laboratories.
  • Bacterial Filtration Efficiency (ASTM F2101): Specific sample size not explicitly stated in the provided text, but testing was performed by Nelson Laboratories.
  • Flammability Class (16CFR 1610): Specific sample size not explicitly stated, but testing was performed by Nelson Laboratories.
  • Delta-P (Mil M36954C): Specific sample size not explicitly stated, but testing was performed by Nelson Laboratories.
  • Biocompatibility (ISO 10993-1): Specific sample sizes for cytotoxicity, sensitization, and irritation tests are not explicitly stated, but tests were performed by Toxikon Corporation.

The data provenance is retrospective as these are non-clinical bench tests of an already manufactured device, and the testing was conducted by independent laboratories (Nelson Laboratories and Toxikon Corporation). The country of origin for the device manufacturer is China.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

This is a non-clinical bench study on device performance characteristics, not a diagnostic or clinical study requiring expert consensus for ground truth. The "ground truth" is based on the objective measurements and established standards of the specified ASTM, CFR, and ISO test methods. The laboratories performing these tests are accredited and their expertise lies in conducting these specific performance evaluations.

4. Adjudication method for the test set:

Not applicable. This is not a clinical study involving subjective interpretation or multiple reviewers. The results are based on objective measurements according to standardized test protocols.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

Not applicable. This is a study for a surgical mask, not an AI-powered diagnostic device, and therefore, no MRMC study or AI assistance evaluation was performed.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

Not applicable. This is a physical device (surgical mask), not an algorithm.

7. The type of ground truth used:

The ground truth used for this study is based on objective measurements from standardized test methods. For example:

  • Fluid resistance is measured by the number of passes at a specific blood pressure.
  • Filtration efficiencies are measured as percentages of particles/bacteria blocked.
  • Flammability is categorized into classes.
  • Biocompatibility is assessed against established criteria for cytotoxicity, sensitization, and irritation.

8. The sample size for the training set:

Not applicable. There is no "training set" for physical device performance testing of this nature. The tests are directly evaluating the product's performance against pre-defined standards.

9. How the ground truth for the training set was established:

Not applicable, as there is no training set.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo for Tiger Medical Group. The logo consists of a circular emblem on the left and the text "Tiger Medical" on the right. Below "Tiger Medical" is the word "Group" in a smaller font size. The emblem appears to be a stylized image, possibly representing a tiger or a medical symbol.

Page 1 of 2

510(k) Summary

JAN 2 5 2011

This summary of 510K safety and effectiveness information is being submitted in accrodance with the requirements of 21 CFR 807.92.

The assigned 510K number is

  1. Submitter's Identification:

Tiger Medical Products Ltd

Liulin Tower, STE 1910

1 Huai Hai Zhong Road

Shanghai 200021

P.R.China

Contact:

Maggie Zhong

Tel:+86-21-33907930

Fax:+86-21-33907932

Mail: maggie-zhong2001@vip.sina.com

Date of Summary: March 8th, 2010

    1. Device Name:
      Tiger Surgical Mask
    1. Classification Name: Surgical Mask
  • 4.. Device Description

The Tiger surgical mask SFM0001 (colors: blue) is flat pleated by 3-ply masks with outer layer and inner layers (spunbonded polypropylene) that sanwich a meltbown polypropylene filter material; Ear-loops are made of urethane elastic fiber for free elastic loops. The nose piece is white aluminum strip with PP covering. All of the materials used in the construction of the Tiger flat surgical Masks are being used in currently marketed devices (see predicate information). All items are nosterilize and only for single use.

5. Intended Use:

Tiger surgical mask SFM0001 is intended for single use by operating room personnel or general health care workers for protection against microscopic organisms, body fluids and particulates. These would include use procedure mask, isolation mask or dental

6. Comparison to Predicate Devices

Tiger surgical mask SFM0001(color: blue) substantially equivalent is safety and effectiveness to the predicate devices.

Tucker & Associates company-K022256 Surgical face masks white, yellow, pink, blue and green

Prestige Ameritech-K061716 Prestige Ameritech Face Mask (multiple labels)

{1}------------------------------------------------

K102713

Image /page/1/Picture/1 description: The image shows the logo for Tiger Medical Group. The logo consists of a circular emblem on the left and the text "Tiger Medical" on the right. Below "Tiger Medical" is the word "Group". The emblem appears to contain the letters TMG.

Page 2 of 2

Performance CharacteristicsTest methodAcceptance criteria/resultPredicateresultsPredicate devicePredicate device results
Tiger surgical maskSFM0001K061716K022256
Fluid resistanceASTM 186231 of 32 Pass at 120mmHg29 of 32 passNo visual penetration
ParticulateFiltrationEfficiencyASTM F2299>99.9% at 2.7µm2.0 microns98.5% at 0.1 microns
Bacterial Filtration EfficiencyASTM F210199.86%99.6%97.9%
Flammability Class16CFR 1610112
Delta-PMil M36954C3.782.61.8
BiocompatibilityISO10993-1Cytotoxictiy: no cyctotoxitialpotentialN/AN/A
Sensitization: negligibleN/AN/A
Irritation: No reactionsN/AN/A

Discussion of Non-clinical Test Performed for Determination of Substantial Equivalence are as follows:

NELSON Fluid Resistance-Resistance of Liquid (Synthetic Blood Penetration I. Resistance) ASTM F 1862

NELSON Bacterial Filtration Efficiency (BFE) test ASTMF2101 II.

NELSON Particulate Filtration Efficiency (Latex Particule Challenge ASTM F2299) III.

NELSON Flammability, Complied with 16 CFR 1610 Class L IV.

V. TOXIKION Biocompatibility per ISO 10993

It is our conclusion that performance testing meet all relevant requirements of the aforementioned test standard.

Discussion of Clinical Tests Performed.

Not Applicable

  1. Conclusions

Tiger surgical mask SFM0001 has the same intended use and technology characteristics as the predicate devices (K022256,K061716). Moreover, the bench testing contained in this submission supplied demonstrate that the technological characteristics do not raise any new question of safety or effectiveness. Therefore, the Tiger SFM0001 flat surgical mask is substantially equivalent to the predicate devices.

Confidential - Property of Tiger Medical Group - Not To Be Used Without Written Authorization by Tiger Medical Group

{2}------------------------------------------------

Standards & Regulations Declaration of Conformity

As manufacturerTiger Medical Products Ltd
Liulin Tower, STE 1910
1 Huai Hai Zhong Road
Shanghai 200021
P.R.China

We declare that SFM0001 flat surgical masks are conforms to standards/ regulations listed herein according to FDA requirements.

Standards & regulations declaration of conformity

ReferenceDescriptionTest LaboratoriesComments
ASTM F1862Synthetic Blood PenetrationResistanceNelson LaboratoriesRefer to test report
ASTM F2101-01Bacterial FiltrationEfficiency(BFE) and Delta-PNelson LaboratoriesRefer to test report
ASTM F2299Particulate Filtration EfficiencyNelson LaboratoriesRefer to test report
16 CFR 1610Flammability 16Nelson LaboratoriesRefer to test report
ISO10993-5CytotoxicityToxikon CorporationRefer to test report
ISO10993-10Irritation & SensitizationToxikon CorporationRefer to test report

(Signature)

Elaine M. Young

(Typed Name) Elaine M Fong

(Date) March 8, 2010 503

{3}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room - WO66-G609 Silver Spring. MD 20993-0002

Ms. Maggie Zhong Consultant Tiger Medical Products, Limited Liulin Tower, STE 1910 1 Huai Hai Zhong Road Shanghai 200021 P.R. China

JAN 2 5 2011

Re: K102713

Trade/Device Name: Tiger Surgical Mask for single use Regulation Number: 21 CFR 878.4040 Regulation Name: Surgical Apparel Regulatory Class: II Product Code: FXX Dated: December 23, 2010 Received: January 10, 2011

Dear Ms. Zhong:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Bearing of the country of the county of the county of 11.11

{4}------------------------------------------------

Page 2- Ms. Zhong

Enclosure

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to

http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the . MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours.

Anthony D. Watson, B.S., M.S., M.B.A. Director Division of Anesthesiology, General Hospital,

Infection Control and Dental Devices Office of Device Evaluation Center for Devices and

Radiological Health

{5}------------------------------------------------

INDICATIONS FOR USE

510(k) NUMBER (IF KNOWN): K 1027 / 3

APPLICANT: Tiger Medical Products Ltd

DEVICE NAME: Tiger Surgical Mask for single use

INDICATION FOR USE:

The Tiger SFM0001 surgical mask is intended for single use by operating room personnel and other health care workers to protect both the patients and the health care workers from transfer of microorganisms, blood and body fluids, and airborne particulate materials.

Prescription Use _____________________________________________________________________________________________________________________________________________________________ (Part 21 CFR 801 Subpart D)

AND/OR Over-The-Counter Use X (21 CFR 807 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Page 1 of ____________________________________________________________________________________________________________________________________________________________________

Elizabeth F. Camine-Willee
(Division Sign-Off)

al Profon Olgh-Olf)
Division of Anesthesiology, General Hospital and Control of AneSthesiology, General Devices

510(k) Number: K102713

4

§ 878.4040 Surgical apparel.

(a)
Identification. Surgical apparel are devices that are intended to be worn by operating room personnel during surgical procedures to protect both the surgical patient and the operating room personnel from transfer of microorganisms, body fluids, and particulate material. Examples include surgical caps, hoods, masks, gowns, operating room shoes and shoe covers, and isolation masks and gowns. Surgical suits and dresses, commonly known as scrub suits, are excluded.(b)
Classification. (1) Class II (special controls) for surgical gowns and surgical masks. A surgical N95 respirator or N95 filtering facepiece respirator is not exempt if it is intended to prevent specific diseases or infections, or it is labeled or otherwise represented as filtering surgical smoke or plumes, filtering specific amounts of viruses or bacteria, reducing the amount of and/or killing viruses, bacteria, or fungi, or affecting allergenicity, or it contains coating technologies unrelated to filtration (e.g., to reduce and or kill microorganisms). Surgical N95 respirators and N95 filtering facepiece respirators are exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 878.9, and the following conditions for exemption:(i) The user contacting components of the device must be demonstrated to be biocompatible.
(ii) Analysis and nonclinical testing must:
(A) Characterize flammability and be demonstrated to be appropriate for the intended environment of use; and
(B) Demonstrate the ability of the device to resist penetration by fluids, such as blood and body fluids, at a velocity consistent with the intended use of the device.
(iii) NIOSH approved under its regulation.
(2) Class I (general controls) for surgical apparel other than surgical gowns and surgical masks. The class I device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 878.9.