K Number
K061716
Manufacturer
Date Cleared
2006-08-14

(56 days)

Product Code
Regulation Number
878.4040
Reference & Predicate Devices
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The following Prestige Ameritech Surgical Masks are single use disposable devices intended to be worn in the operating room as well as dental, isolation and other medical procedures to protect both the patient and healthcare personnel from transfer of microorganisms, body fluids, and particulate material.

Prestige Ameritech Earloop Surgical Mask-Blue, Pink, Green, White, Yellow, Peach Prestige Ameritech Earloop Surgical Mask with splash shield-Blue Green, White, Peach Prestige Ameritech Earloop Surgical Mask Tissue-Blue, Pink, White, Yellow Prestige Ameritech Tie on Surgical Mask-Blue, Pink, Green, White, Peach Prestige Ameritech Tie on Surgical Mask with splash shield-Blue, Green, White, Peach

Device Description

The Prestige Ameritech Surgical Mask device is composed of three layers of material flat folded and pleated to form the Mask. The inner layer is composed of cellulose or spunbonded polypropylene, the middle layer is a meltblown polypropylene filter material and the outer layer is a medical grade tissue or spunbonded polypropylene. Masks with splash shields have anti-fog plastic shields attached to masks. Masks are held in place on wearer with tie strips or latex free elastic loops and contain a malleable aluminum nosepiece strip. All of the materials used in this device are typical materials commonly used in the construction of Surgical Face Masks and are being used in current legally marketed devices.

AI/ML Overview

The provided 510(k) summary describes the Prestige Ameritech Face Mask, a surgical face mask. The acceptance criteria and device performance are based on bench testing of physical properties rather than clinical studies with human participants. Therefore, several of the requested sections (e.g., sample size for test set, number of experts, adjudication method, MRMC study, training set information) are not applicable or cannot be extracted from this document.

1. Table of Acceptance Criteria and Reported Device Performance

The acceptance criteria are implicitly defined by the results of the predicate devices, as the Prestige Ameritech device aims to demonstrate substantial equivalence by performing similarly or better.

Performance CharacteristicsTest MethodPredicate Device K022256 Results (Implied Acceptance)Predicate Device K060776 Results (Implied Acceptance)Prestige Ameritech Device Performance
Fluid ResistanceASTM F1862No Visual Penetration (Implied: Fluid Resistant)Fluid ResistantFluid Resistant
Particulate FiltrationASTM F22992.0 microns (Implicit: Efficiency at 2.0 microns)96.8% at 0.1 microns98.5% at 0.1 Microns
Bacterial FiltrationASTM F210197.9%99.9%99.6%
Differential PressureMil M36954C1.82.342.6
Flammability Class16CFR 1610211

2. Sample size used for the test set and the data provenance

  • Sample Size: Not specified in the document. These are bench tests, so sample sizes would refer to the number of mask samples tested for each characteristic.
  • Data Provenance: The data is generated from laboratory bench testing (non-clinical tests) of the device and predicate devices. No information about the country of origin or whether it's retrospective or prospective is provided, but it's understood to be prospective testing for the submission.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

Not applicable. The ground truth for these performance characteristics is established through standardized laboratory test methods (e.g., ASTM F1862, ASTM F2299, ASTM F2101, MIL M36954C, 16CFR 1610). Experts are involved in designing and executing these tests, but "ground truth" isn't established through expert consensus in the typical sense for medical imaging or diagnostic devices.

4. Adjudication method for the test set

Not applicable. This is not a study involving human interpretation or subjective assessments that would require adjudication.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

Not applicable. This is a surgical mask, not an AI-powered diagnostic device, so MRMC studies and AI assistance are not relevant.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

Not applicable. This is a surgical mask, not an algorithm.

7. The type of ground truth used

The "ground truth" for the device's performance is objective measurement against established industry standards and test methodologies. For example:

  • Fluid Resistance: Visual observation of penetration after exposure to synthetic blood.
  • Particulate Filtration Efficiency: Quantitative measurement of particle removal at a specific micron size.
  • Bacterial Filtration Efficiency: Quantitative measurement of bacterial removal.
  • Differential Pressure: Measurement of airflow resistance.
  • Flammability: Classification based on burn rate.

8. The sample size for the training set

Not applicable. This is a physical device, and the submission does not describe a training set in the context of machine learning or AI.

9. How the ground truth for the training set was established

Not applicable. No training set is described.

{0}------------------------------------------------

510(k) SUMMARY

This summary of 510(k) safety and effectiveness information is being submitted in accordance with requirements of SMDA 1990.

The assigned 510(k) number is: K061716

  • Submitted for: Prestige Ameritech 7425 Airport Frwy. Fort Worth, TX. 76118 817 595-1131 Establishment Registration number: 3005022483 Dan Reese, President Contact Person: 7425 Airport Frwy. Fort Worth, TX. 76118 817 595-1131 Date Prepared: July 19, 2006 Proprietary Name: Prestige Ameritech Face Mask (multiple labels); Prestige Ameritech Earloop Surgical Mask-Blue, Pink, Green, White, Yellow, Peach Prestige Ameritech Earloop Surgical Mask with splash shield-Blue Green, White, Peach Prestige Ameritech Earloop Surgical Mask Tissue-Blue, Pink, White, Yellow Prestige Ameritech Tie on Surgical Mask-Blue, Pink, Green, White, Peach Prestige Ameritech Tie on Surgical Mask with splash shield-Blue, Green, White, Peach Common Name: Surgical Face Mask Classification Name: Mask, Surgical Classification Product Code: FXX Regulation Number: 878.4040 Predicate Devices: Tucker and Associates Surgical Face Mask K022256 Hong Ya Non-Woven Products Co. Ltd. Surgical Face Mask K060776

{1}------------------------------------------------

KOG1716

Device Description: The Prestige Ameritech Surgical Mask device is composed of three layers of material flat folded and pleated to form the Mask. The inner layer is composed of cellulose or spunbonded polypropylene, the middle layer is a meltblown polypropylene filter material and the outer layer is a medical grade tissue or spunbonded polypropylene. Masks with splash shields have anti-fog plastic shields attached to masks. Masks are held in place on wearer with tie strips or latex free elastic loops and contain a malleable aluminum nosepiece strip. All of the materials used in this device are typical materials commonly used in the construction of Surgical Face Masks and are being used in current legally marketed devices.

Intended Use: The following Prestige Ameritech Surgical Masks are single use disposable devices intended to be worn in the operating room as well as dental, isolation and other medical procedures to protect both the patient and healthcare personnel from transfer of microorganisms, body fluids, and particulate material.

Prestige Ameritech Earloop Surgical Mask-Blue, Pink, Green, White, Yellow, Peach Prestige Ameritech Earloop Surgical Mask with splash shield-Blue Green, White, Peach Prestige Ameritech Earloop Surgical Mask Tissue-Blue, Pink, White, Yellow Prestige Ameritech Tie on Surgical Mask-Blue, Pink, Green, White, Peach Prestige Ameritech Tie on Surgical Mask with splash shield-Blue. Green, White, Peach

PerformanceCharacteristicsTest MethodResultsK022256ResultsK060776ResultsPrestigeAmeritechDevice
Fluid ResistanceASTMF1862No VisualPenetrationFluidResistantFluidResistant
Particulate FiltrationEfficiencyASTMF22992.0microns96.8% at 0.1microns98.5% at 0.1Microns
Bacterial FiltrationEfficiencyASTMF210197.9%99.9%99.6%
Differential PressureMilM36954C1.82.342.6
Flammability Class16CFR 1610211

Comparison To Predicate Devices:

{2}------------------------------------------------

Discussion of Non-Clinical tests Performed for Determination of Substantial Equivalence:

    1. Fluid Resistance-ASTM F1862
    1. Particulate Filtration Efficiency-ASTM F2299
    1. Bacterial Filtration Efficiency-ASTM F2101
    1. Differential Pressure-MIL M36954C
    1. Flammability Testing-16CFR 1610
    1. Irritation and Sensitization-ISO 10993-10

Conclusions: The Prestige Ameritech Face Masks have the same intended use and technological characteristics as the predicate devices K022256 and K060776. Moreover, bench testing contained in this submission demonstrates that the technological characteristics do not raise any new question of safety or effectiveness. Therefore, the Prestige Face Mask (multiple labels) is substantially equivalent to the predicate devices.

{3}------------------------------------------------

Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo features a stylized eagle with three lines forming its body and wings. The eagle is encircled by the text "DEPARTMENT OF HEALTH AND HUMAN SERVICES - USA" in a circular arrangement.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

AUG 1 4 2006

Mr. Dan Reese President Prestige Ameritech 7425 Airport Freeway Fort Worth, Texas 76118

Rc: K061716

Trade/Device Name: Prestige Ameritech Face Mask (Multiple Labels) Regulation Number: 878.4040 Regulation Name: Surgical Apparel Regulatory Class: II Product Code: FXX Dated: July 17, 2006 Received: July 20, 2006

Dear Mr. Reese:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting , your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Lederal Register.

{4}------------------------------------------------

Page 2 - Mr. Reesc

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0115. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely yours,

Sutte Y. Michael Dms.

Chiu Lin, Ph.D. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

K061716

Indications for Use

510(k) Number (if known):

Device Name:

Device Name;

Indications For Use:

The following Prestige Ameritech Surgical Masks are single use disposable devices intended to be worn in the operating room as well as dental, isolation and other medical procedures to protect both the patient and healthcare personnel from transfer of microorganisms, body fluids, and particulate material.

Prestige Ameritech Earloop Surgical Mask-Blue, Pink, Green, White. Yellow, Peach Prestige Ameritech Earloop Surgical Mask with splash shield-Blue Green, White, Peach Prestige Ameritech Earloop Surgical Mask Tissue-Blue, Pink, White, Yellow Prestige Ameritech Tie on Surgical Mask-Blue, Pink, Green, White, Peach Prestige Ameritech Tie on Surgical Mask with splash shield-Blue, Green, White, Peach

Prescription Use (Part 21 CFR 801 Subpart D) AND/OR

Over-The-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Shale H. Munday, R.D. 8/9/04
(Sign-Off)

ion of Anesthesiology, General Hospital,
ion Control, Dental Devices

) Number_

Page 1 of __

§ 878.4040 Surgical apparel.

(a)
Identification. Surgical apparel are devices that are intended to be worn by operating room personnel during surgical procedures to protect both the surgical patient and the operating room personnel from transfer of microorganisms, body fluids, and particulate material. Examples include surgical caps, hoods, masks, gowns, operating room shoes and shoe covers, and isolation masks and gowns. Surgical suits and dresses, commonly known as scrub suits, are excluded.(b)
Classification. (1) Class II (special controls) for surgical gowns and surgical masks. A surgical N95 respirator or N95 filtering facepiece respirator is not exempt if it is intended to prevent specific diseases or infections, or it is labeled or otherwise represented as filtering surgical smoke or plumes, filtering specific amounts of viruses or bacteria, reducing the amount of and/or killing viruses, bacteria, or fungi, or affecting allergenicity, or it contains coating technologies unrelated to filtration (e.g., to reduce and or kill microorganisms). Surgical N95 respirators and N95 filtering facepiece respirators are exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 878.9, and the following conditions for exemption:(i) The user contacting components of the device must be demonstrated to be biocompatible.
(ii) Analysis and nonclinical testing must:
(A) Characterize flammability and be demonstrated to be appropriate for the intended environment of use; and
(B) Demonstrate the ability of the device to resist penetration by fluids, such as blood and body fluids, at a velocity consistent with the intended use of the device.
(iii) NIOSH approved under its regulation.
(2) Class I (general controls) for surgical apparel other than surgical gowns and surgical masks. The class I device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 878.9.