K Number
K102645
Manufacturer
Date Cleared
2010-12-10

(88 days)

Product Code
Regulation Number
870.1250
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The WILLOW PTA Balloon Dilatation Catheter is intended to dilate stenoses in the illac, femoral, ilio-femoral, popliteal, infra-popliteal, and renal arteries, and for the treatment of obstructive lesions of native or synthetic arteriovenous dialysis fistulae.

The ELM PTA Balloon Dilatation Catheter is indicated for use in Percutaneous Transluminal Angioplasty of the femoral, iliac, and renal arteries and for the treatment of obstructive lesions of native or synthetic arteriovenous dialysis fistulae. This catheter is not for use in coronary arteries.

Device Description

Both the WILLOW and ELM are co-axial PTA Balloon Dilatation Catheters with a distal inflatable balloon. The lumen between the inner shaft is used for inflation of the balloon with contrast medium; the inner shaft lumen permits the use of a quide wire to facilitate advancement of the catheter to and through the stenosis to be dilated. Two radiopaque marker bands indicate the dilating section of the balloon and aid in the balloon placement. The marker bands also indicate the nominal length of the balloon. The catheter tip is designed to ease entry into the peripheral arteries and to facilitate the crossing of tight stenoses.

The following materials are used in the WILLOW PTA Balloon Dilatation Catheter: Nylon 11, Polyethylene, Polycarbonate, and 90% Platinum/10% Iridium. The following materials are used in the ELM PTA Balloon Dilatation Catheter: Nylon 11, Pebax, Polycarbonate, Polyethylene, and 90% Platinum/10% Iridium. All materials in both devices were tested per ISO 10993-1 and determined to be biocompatible.

AI/ML Overview

The provided document describes the Creagh Medical WILLOW and ELM PTA Balloon Dilatation Catheters, submitted for 510(k) clearance in 2010. The document focuses on demonstrating substantial equivalence to predicate devices rather than proving de novo safety and effectiveness through clinical trials with specific acceptance criteria related to a disease state or imaging modality.

Therefore, the acceptance criteria and study detailed below relate to device performance characteristics intended to show that the new devices are as safe and effective as their legally marketed predecessors.


1. Table of Acceptance Criteria and Reported Device Performance

The document does not explicitly state pre-defined "acceptance criteria" in a numerical or percentage format for all performance aspects, nor does it provide detailed quantitative results for every test. Instead, it broadly states that "performance testing demonstrated that the device performed as intended and in compliance with relevant consensus standards." However, for specific critical performance metrics, quantitative results are provided to support substantial equivalence.

Performance CharacteristicAcceptance Criteria (Implicit from Predicate Equivalence)Reported Device Performance (WILLOW)Reported Device Performance (ELM)
Rated Burst Pressure99.9% of product meets specification (with 95% confidence)99.9% of product meets specification (with 95% confidence)99.9% of product meets specification (with 95% confidence)
Compliance99.7% of product meets specification (with 95% confidence)99.7% of product meets specification (with 95% confidence)99.7% of product meets specification (with 95% confidence)
BiocompatibilityAll materials tested per ISO 10993-1 and determined biocompatibleAchieved: All materials tested per ISO 10993-1 and determined biocompatibleAchieved: All materials tested per ISO 10993-1 and determined biocompatible
SterilizationValidatedAchieved: Sterilization validatedAchieved: Sterilization validated
Shelf LifeValidatedAchieved: Shelf life validatedAchieved: Shelf life validated
Other V&V Tests*Performed as intended & in compliance with standardsAchieved: Performed as intended & in compliance with standardsAchieved: Performed as intended & in compliance with standards

*Other V&V Tests include: multiple inflation/fatigue and leak tight, balloon lengths and marker band position, inflation and deflation time, wire compatibility, catheter effective length, tensile strength, device profile/sheath compatibility, tip profile, radiopacity, simulated use including pushability and trackability and sheath compatibility, and manifold testing.


2. Sample Size for Test Set and Data Provenance

The document does not specify a "test set" in the context of clinical data for algorithmic performance testing. Instead, it refers to non-clinical design verification/validation tests and analyses.

  • Sample Size: The document indicates that for rated burst pressure and compliance testing, the results are presented "with 95% confidence that 99.9% of the product meet the rated burst pressure specification" and "with 95% confidence that 99.7% of the product meets the compliance specification." While this phrasing suggests statistical sampling was used, the exact sample sizes (n) for each specific test are not reported.
  • Data Provenance: This is a retrospective non-clinical study conducted by the manufacturer, Creagh Medical, whose address is in Ballinasloe, Co. Galway, Ireland. The data is from internal laboratory testing, not from patients or a specific country of origin in the clinical sense.

3. Number of Experts Used to Establish Ground Truth and Qualifications

Not applicable. This is a submission for a medical device (PTA balloon catheter) based on non-clinical engineering and performance testing, not an AI or diagnostic device that requires expert-established ground truth for image interpretation or disease diagnosis. The "ground truth" here is objective physical and mechanical measurements against established engineering and safety standards.


4. Adjudication Method for the Test Set

Not applicable. As described above, this is non-clinical device testing, not a study involving human interpretation or adjudication of results. The "adjudication" would be performed by test engineers against predefined engineering specifications and regulatory standards.


5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

Not applicable. This is a non-clinical submission for a physical medical device. MRMC studies are typically used for evaluating the performance of diagnostic imaging devices or AI algorithms where human readers are involved in interpreting results.


6. Standalone (Algorithm Only Without Human-in-the-Loop) Performance

Not applicable. This is a physical medical device, not an algorithm.


7. Type of Ground Truth Used

The "ground truth" for this submission is based on objective engineering specifications, performance standards (e.g., ISO, internal design documents), and comparison to the characteristics of the predicate devices. For example:

  • Rated Burst Pressure: Ground truth is the specified pressure value the balloon must withstand.
  • Compliance: Ground truth is the specified change in balloon diameter for a given pressure.
  • Biocompatibility: Ground truth is compliance with ISO 10993-1.

8. Sample Size for the Training Set

Not applicable. This is a physical medical device, not an AI algorithm requiring a training set. The "training" for such devices would involve manufacturing process development and optimization, not data-driven machine learning.


9. How the Ground Truth for the Training Set Was Established

Not applicable. There is no training set as this is not an AI/algorithmic device.

{0}------------------------------------------------

DEC 1 0 2010

2.5 510(k) Summary

This 510(k) Summary is being submitted in accordance with the requirements of the SMDA 1990 and 21 CFR 807.92.

1. General Information

Submitter: Creagh Medical IDA Business Park, Ballinasloe, Co. Galway, Ireland Telephone Number: 011 353 909 646300 Fax Number: 011 353 909 646330 Contact Person: Maureen O'Connell O'Connell Regulatory Consultants, Inc. 5 Timber Lane North Reading, MA 01864 Telephone: 978-207-1245 Fax: 978-824-2541 Summary Preparation Date: December 8, 2010 2. Device Information Device Trade Names: WILLOW PTA Balloon Dilatation Catheter ELM PTA Balloon Dilatation Catheter Common Name: PTA Balloon Dilatation Catheter Classification Name: Catheter, Angioplasty, Peripheral, Transluminal (21 CFR 870.1250, Product Code: LIT) 3. Predicate Devices WILLOW PTA Balloon Dilatation Catheter predicate Device Name: AMPHIRION DEEP 0.0 14"OTW PTA Balloon Dilatation Catheter 510(k) Clearance Number: K050073, K052791, K042624, K083919 ELM PTA Balloon Dilatation Catheter predicate Vaccess™ PTA Balloon Dilatation Catheter Device Name:

K073472

510(k) Clearance Number:

{1}------------------------------------------------

4. Device Description

Both the WILLOW and ELM are co-axial PTA Balloon Dilatation Catheters with a distal inflatable balloon. The lumen between the inner shaft is used for inflation of the balloon with contrast medium; the inner shaft lumen permits the use of a quide wire to facilitate advancement of the catheter to and through the stenosis to be dilated. Two radiopaque marker bands indicate the dilating section of the balloon and aid in the balloon placement. The marker bands also indicate the nominal length of the balloon. The catheter tip is designed to ease entry into the peripheral arteries and to facilitate the crossing of tight stenoses.

The following materials are used in the WILLOW PTA Balloon Dilatation Catheter: Nylon 11, Polyethylene, Polycarbonate, and 90% Platinum/10% Iridium. The following materials are used in the ELM PTA Balloon Dilatation Catheter: Nylon 11, Pebax, Polycarbonate, Polyethylene, and 90% Platinum/10% Iridium. All materials in both devices were tested per ISO 10993-1 and determined to be biocompatible.

5. Indications for Use

The WILLOW PTA Balloon Dilatation Catheter is intended to dilate stenoses in the illac, femoral, ilio-femoral, popliteal, infra-popliteal, and renal arteries, and for the treatment of obstructive lesions of native or synthetic arteriovenous dialysis fistulae.

The ELM PTA Balloon Dilatation Catheter is indicated for use in Percutaneous Transluminal Angioplasty of the femoral, iliac, and renal arteries and for the treatment of obstructive lesions of native or synthetic arteriovenous dialysis fistulae. This catheter is not for use in coronary arteries.

6. Substantial Equivalence

This Premarket Notification aims to demonstrate substantial equivalence of the WILLOW PTA Balloon Dilatation Catheter through comparison with the predicate device - the Invatec AMPHIRION DEEP. The substantial equivalence discussion is demonstrated for the intended use of the WILLOW. Where substantial equivalence is not directly demonstrated from the perspective of technology and performance, design verification testing provides evidence of the substantial equivalance of the WILLOW.

The WILLOW has the same technological characteristics as the AMPHIRION DEEP predicate device in that they are both co-axial catheters with a distal, inflatable balloon. They have similar design, however, the exact specifications of the WILLOW and the AMPHIRION DEEP differ in certain aspects. The differences include a longer inflated balloon length for the WILLOW, higher rated burst pressure for the WILLOW, lower nominal pressure for the WILLOW, different strain relief between the WILLOW and the predicate in terms of design and materials, and different marker bands in terms of materials. In each case that there was a difference between the WILLOW and the predicate device, design verification testing was performed to demonstrate substantial equivalence. In all cases, performance testing demonstrated that the device performed as intended and in compliance with relevant consensus standards.

The Premarket Notification also aims to demonstrate substantial equivalence of the ELM PTA Balloon Dilatation Catheter through comparison with the predicate device - the Bard Vaccess™. Substantial equivalence is demonstrated for the intended use of the ELM device. Where substantial equivalence is not directly demonstrated from the perspective of technology and performance, design verification testing provides evidence of the substantial equivalance of the ELM device.

{2}------------------------------------------------

The ELM device has the same technological characteristics as the Vaccess predicate device in that they are both co-axial lumen catheters with a distal, inflatable balloon. They have similar design, however, the exact specifications of ELM and the Vaccess devices differ in certain aspects. The differences include different strain relief, inner shaft and balloon design between the ELM and the predicate in terms of design and materials. different marker bands in terms of materials, different sheath break sizes compared to the predicate, higher rated burst pressure for the ELM across the product matrix compared with the predicate, higher nominal pressure for the ELM, smaller inflated balloon diameter for the ELM, and the usable catheter length varies slightly compared to the predicate device. In each case that there was a difference between the ELM and the predicate device, design verification testing was performed to demonstrate substantial equivalence. In all cases, performance testing demonstrated that the device performed as intended and in compliance with relevant consensus standards.

7. Performance Data

The safety and effectiveness of the WILLOW and ELM PTA Balloon Dilatation Catheters has been demonstrated through data collected from non-clinical design verification/ validation tests and analyses.

Performance testing included generation of compliance data relating balloon diameter to applied pressure to vessel size, sterilization, and validation of shelf life. Additionally, the following verification/validation testing was performed: rated burst pressure, multiple inflation/fatigue and leak tight, balloon lengths and marker band position, inflation and deflation time, wire compatibility, catheter effective length, tensile strength, device profile/sheath compatibility, tip profile, radiopacity, simulated use including pushability and trackability and sheath compatibility, and manifold testing.

8. Conclusions

The substantial equivalence of the ELM and WILLOW PTA Balloon Dilatation Catheters has been demonstrated through comparison with the identified predicate devices. Substantial equivalence is demonstrated for the intended use of both devices. Where substantial equivalence is not directly demonstrated from the perspective of technology and performance, design verification testing provides evidence of the substantial equivalance of the ELM and WILLOW devices. Design verification testing of the WILLOW and ELM meet pre-determined performance characteristics as documented in the Design Output and Product Specification documents. Design validation has been carried out to demonstrate that the WILLOW and ELM designs conform to defined user needs and intended uses as defined in the WILLOW and ELM Design Input and Market Specifications documention.

Additional testing was done to support substantial equivalence to the identified predicate devices. For the WILLOW, rated burst pressure was tested and data demonstrates with 95% confidence that 99.9% of the product meet the rated burst pressure specification and therefore is substantially equivalent to the predicate device. Compliance testing was performed which demonstrated that with 95% confidence that 99.7% of the product meets the compliance specification and therefore satisfies is substantially equivalent to the predicate device.

For the ELM, rated burst pressure was tested and data demonstrates with 95% confidence that 99.9% of the product meet the rated burst pressure specification and therefore is substantially equivalent to the predicate device. Compliance testing was performed which demonstrated that with 95% confidence that 99.7% of the product meets the compliance specification and therefore satisfies is substantially equivalent to the predicate device.

{3}------------------------------------------------

Therefore, the WILLOW and ELM PTA Balloon Dilation Catheters perform as intended and are substantially equivalent with the identified predicate devices.

.

.

.

.

.

{4}------------------------------------------------

Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a stylized eagle with three stripes forming its body and wings. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" is arranged in a circular fashion around the eagle.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002

Creagh Medical Ltd. C/O Maureen O'Connell President 5 Timber Lane North Reading, MA 01864

DEC 1 0 2010

Re: K102645

Trade/Device Name: WILLOW PTA Balloon Dilatation Catheter ELM PTA Balloon Dilatation Catheter Regulation Number: 21 CFR 870.1250 Regulation Name: Percutaneous catheter Regulatory Class: Class II Product Code: LIT, DOY Dated: September 13, 2010 Received: September 13, 2010

Dear Ms. O'Connell:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{5}------------------------------------------------

Page 2 - Ms. Maureen O'Connell

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours.

Duna D. V. Ulmer

Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{6}------------------------------------------------

2.4(a) Indications for Use

510(k) Number. K102645

Device Name: WILLOW PTA Balloon Dilatation Catheter

DEC 1 0 2010

Indications for Use:

The WILLOW PTA Balloon Dilatation Catheter is intended to dilate stenoses in the liia, femoral, ilio-femoral, popliteal, infra-popliteal, and renal arteries, and for the treatment of obstructive lesions of native or synthetic arteriovenous dialysis fistulae.

Prescription Use 区 (Part 21 CFR 801 Subpart D)

And/ Or

Over-The- Counter Use [ (Part 21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IFNEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Dune D. V. Turner

(Division Sign-Off) (Division Olgh Sign of Cardiovascular Devices

V 1021 510(k) Number

{7}------------------------------------------------

2.4(b) Indications for Use

510(k) Number _ K1026 AS

DEC 1 0 2010

Device Name: ELM PTA Balloon Dilatation Catheter

The ELM PTA Balloon Dilatation Catheter is indicated for use in Percutaneous Transluminal Angioplasty of the femoral, iliac, and renal arteries and for the treatment of obstructive lesions of native or synthetic arteriovenous dialysis fistulae. This catheter is not for use in coronary arteries.

Prescription Use [ (Part 21 CFR 801 Subpart D) And/ Or

Over-The- Counter Use (Part 21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IFNEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Dwna R. Vc Anes

(Division Sign-Off) Division of Cardiovascular Devices

510(k) Number_K102

§ 870.1250 Percutaneous catheter.

(a)
Identification. A percutaneous catheter is a device that is introduced into a vein or artery through the skin using a dilator and a sheath (introducer) or guide wire.(b)
Classification. Class II (performance standards).