K Number
K090625
Device Name
WIRELESS PORTABLE DETECTOR FD-W17
Date Cleared
2009-03-24

(15 days)

Product Code
Regulation Number
892.1680
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP Authorized
Intended Use
As a part of a radiographic system, the Wireless Portable Detector FD-W17 is intended to acquire digital radiographic images. The Wireless Portable Detector FD-W17 is suitable for all routine radiography exams, including specialist areas like intensive care, trauma, or pediatric work, excluding fluoroscopy, angiography and mammography.
Device Description
As a part of a radiographic system, the Wireless Portable Detector FD-W17 is intended to acquire digital radiographic images. The Detector is combined with a Philips XD-S workstation (K063781) which consists of a computer, keyboard, display, mouse. The complete X-ray system would further include other Philips subsystems and components, like patient table, X-ray control(s), X-ray high voltage generator, X-ray tube(s), collimator(s), accessories, etc. The XD-S workstation and the complete X-ray systems are not changed other than by replacing an Xray receptor with the Wireless Portable Detector FD-W17.
More Information

No
The summary describes a digital radiographic detector and associated workstation, focusing on image acquisition and basic pre-processing (grid suppression). There is no mention of AI, ML, or any features that would typically involve such technologies for image analysis, interpretation, or workflow optimization. The comparison to predicate and reference devices also points to standard digital radiography technology.

No
The device is described as a detector intended to acquire digital radiographic images, which is a diagnostic function, not a therapeutic one.

No.
The device acquires digital radiographic images, which are then used by other components of the radiographic system for diagnosis. The device itself is not performing the diagnosis.

No

The device description explicitly states it is a "Wireless Portable Detector FD-W17" and is a physical component of a radiographic system, not solely software.

Based on the provided information, this device is not an IVD (In Vitro Diagnostic).

Here's why:

  • Intended Use: The intended use clearly states that the device is for acquiring digital radiographic images as part of a radiographic system. This is a diagnostic imaging function, not an in vitro diagnostic function.
  • Device Description: The description reinforces that it's a component of an X-ray system used for acquiring images.
  • Nature of IVDs: In vitro diagnostics are tests performed on samples taken from the human body (like blood, urine, tissue) to detect diseases, conditions, or infections. This device does not interact with biological samples in this way.

The device is a component of a medical imaging system used for diagnostic purposes, but it is not an in vitro diagnostic device.

N/A

Intended Use / Indications for Use

As a part of a radiographic system, the Wireless Portable Detector FD-W17 is intended to acquire digital radiographic images. The Wireless Portable Detector FD-W17 is suitable for all routine radiography exams, including specialist areas like intensive care, trauma, or pediatric work, excluding fluoroscopy, angiography and mammography.

Product codes

MQB, MOB

Device Description

As a part of a radiographic system, the Wireless Portable Detector FD-W17 is intended to acquire digital radiographic images. The Detector is combined with a Philips XD-S workstation (K063781) which consists of a computer, keyboard, display, mouse. The complete X-ray system would further include other Philips subsystems and components, like patient table, X-ray control(s), X-ray high voltage generator, X-ray tube(s), collimator(s), accessories, etc. The XD-S workstation and the complete X-ray systems are not changed other than by replacing an Xray receptor with the Wireless Portable Detector FD-W17.

Mentions image processing

Yes

Mentions AI, DNN, or ML

Not Found

Input Imaging Modality

Digital Radiography

Anatomical Site

Not Found

Indicated Patient Age Range

Not Found

Intended User / Care Setting

Not Found

Description of the training set, sample size, data source, and annotation protocol

Not Found

Description of the test set, sample size, data source, and annotation protocol

Not Found

Summary of Performance Studies (study type, sample size, AUC, MRMC, standalone performance, key results)

Not Found

Key Metrics (Sensitivity, Specificity, PPV, NPV, etc.)

Not Found

Predicate Device(s)

K982795

Reference Device(s)

K051523, K063781

Predetermined Change Control Plan (PCCP) - All Relevant Information

Not Found

§ 892.1680 Stationary x-ray system.

(a)
Identification. A stationary x-ray system is a permanently installed diagnostic system intended to generate and control x-rays for examination of various anatomical regions. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.(b)
Classification. Class II (special controls). A radiographic contrast tray or radiology diagnostic kit intended for use with a stationary x-ray system only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.

0

Wireless Portable Detector FD-W17 510(k) Submission

MAR 2 4 2009

March Of

5 - 510(k) Summary of Safety and Effectiveness

In accordance with the requirements of the Safe Medical Device Act, Philips Medical Systems North America Company herewith submits a Summary of Safety and Effectiveness.

| MANUFACTURER: | Philips Medical Systems DMC GmbH
Roentgenstrasse 24
22335 Hamburg
Germany |
|---------------------------------------------|--------------------------------------------------------------------------------------------|
| Establishment Registration No.: | 3003768251 |
| SUBMITTER: | Philips Medical Systems
22100 Bothell Everett Highway
Bothell, WA 98041-3003 |
| Establishment Registration No.:
Contact: | 1217116
Lynn Harmer
425-487-7312 |
| Date Prepared: | March 06, 2009 |
| CLASSIFICATION NAME: | Solid State X-Ray Imager (Flat Panel/Digital Imager)
Class II
MQB
21 CFR 892.1650 |
| COMMON/USUAL NAME: | Solid State X-Ray Imager (Flat Panel/Digital Imager) |
| TRADE/PROPRIETARY NAME: | Wireless Portable Detector FD-W17 |

PERFORMANCE STANDARDS:

This device complies with the federal X-Ray performance standards (CFR 1020.30, .31)

SYSTEM DESCRIPTION:

As a part of a radiographic system, the Wireless Portable Detector FD-W17 is intended to acquire digital radiographic images.

1

K090625
Page 2 of 3

Wireless Portable Detector FD-W17 510(k) Submission

October 20, 2008

The Detector is combined with a Philips XD-S workstation (K063781) which consists of a computer, keyboard, display, mouse.

The complete X-ray system would further include other Philips subsystems and components, like patient table, X-ray control(s), X-ray high voltage generator, X-ray tube(s), collimator(s), accessories, etc.

The XD-S workstation and the complete X-ray systems are not changed other than by replacing an Xray receptor with the Wireless Portable Detector FD-W17.

INTENDED USE:

As a part of a radiographic system, the Wireless Portable Detector FD-W17 is intended to acquire digital radiographic images. The Wireless Portable Detector FD-W17 is suitable for all routine radiography exams, including specialist areas like intensive care, trauma, or pediatric work, excluding fluoroscopy, angiography and mammography.

EQUIVALENCE INFORMATION:

The Wireless Portable Detector FD-W17 is considered substantially equivalent to the Pixium 4600 detector, which received FDA clearance on November 24, 1998, under the name Philips Bucky Vision in 510(k) Number K982795.

The grid supression pre-processing is equivalent to the pre-processing used by the Konica Minolta Regius Console CS 2000/CS-3000 (K051523).

The other parts of the workstation are equal to the Philips XD-S Direct Radiography Workstation/Package, which received FDA clearance on January 05, 2007, under 510(k) Number K063781.

SAFETY INFORMATION:

The Wireless Portable Detector FD-W17 uses mature technology. It is designed to be in compliance with relevant national and international standards for electrical safety (UL 60601-1, IEC 60601-1) , radiation protection (IEC 60601-1-3) and Electromagnetic Compatibility (IEC-60601-1-2).

The wireless transmission of data is evaluated in risk management and tested under worst case scenarios. The Center for Devices and Radiological Health (CDRH) draft guidance "Radio-Frequency Wireless Technology in Medical Devices" from January 3, 2007 was used in preparing tests and documentation for the wireless functionality.

2

K09C625
Page 3 of 3

Wireless Portable Detector FD-W17 510(k) Submission

October 20, 2008

A product risk mangement is executed according to ISO 14971 and all risks are reduced to an acceptable level by implementation and verification of appropriate measures.

CONCLUSION:

Philips Medical Systems believes that the Wireless Portable Detector FD-W17 is substantially equivalent to the currently legally marketed devices. It does not introduce new indications for use, has the same technological characteristics and does not introduce new potential hazards or safety risks.

3

Image /page/3/Picture/0 description: The image shows the logo of the Department of Health & Human Services. The logo includes the department's name in blue, with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES" written in a clear, sans-serif font. To the left of the text is a circular emblem featuring an abstract design, also in blue, which represents the department's mission and values.

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room - WO66-G609 Silver Spring, MD 20993-0002

Philips Ultrasound, Inc. % Mr. Mark Job Responsible Third Party Official Regulatory Technology Services LLC 1394 25th Street NW BUFFALO MN 55313

Re: K090625

Trade/Device Name: Wireless Portable Detector FD-W17 Regulation Number: 21 CFR 892.1680 Regulation Name: Stationary x-ray system Regulatory Class: II Product Code: MOB Dated: March 7, 2009 Received: March 9. 2009

AUG 2 3 2013

Dear Mr. Job:

This letter corrects our substantially equivalent letter of March 24, 2009.

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The r ou may, more, mains of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into class II (Special Controls), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must or any I cieral statutes and equirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of

4

medical device-related adverse events) (21 CFR 803); and good manufacturing practice mouitements as set forth in the quality systems (QS) regulation (21 CFR Part 820). This letter will allow you to begin marketing your device as described in your Section 510(k) premarket whit anow you to ocgan mailsoming your available quivalence of your device to a legally marketed notication. THU PDF imaning of Scation for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Parts 801 and If you desire specific advice of In Vitro Diagnostic Device Evaluation and Safety at (301) 796-007), promote note the regulation entitled, "Misbranding by reference to premarket 5 150. The stime not 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Tou may of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely Yours,

Janine M. Morris

Acting Director Division of Radiological Devices Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health

Enclosure

5

Wireless Portable Detector FD-W17 51Q(k) Submission Section 4

October 20, 2008

: 六

"

Indications for Use

K090625 510(k) Number (if known):

Device Name: Wireless Portable Detector FD-W17

Indications For Use:

As a part of a radiographic system, the Wireless Portable Detector FD-W17 is intended to acquire digital radiographic images. The Wireless Portable Detector FD-W17 is suitable for all routine radiography exams, including specialist areas like intensive care, trauma, or pediatric work, excluding fluoroscopy, angiography and mammography.

Prescription Use × (Part 21 CFR 801 Subpart D)

AND/OR

Over-The-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IFNEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)


(Division Sign-Off)

Division of Reproductive, Abdominal and Radiological Devices 510(k) Number

Page 1 of 1

ત્ત્વ