K Number
K102490
Date Cleared
2011-06-02

(275 days)

Product Code
Regulation Number
866.5660
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The IgX PLEX™ Celiac Qualitative Assay is an in vitro diagnostic test for the qualitative detection of the IgA and IgG immunoglobulin classes of anti-tissue transglutaminase antibody in serum. The test is intended for use in clinical laboratories as an aid in the diagnosis of celiac disease in conjunction with other laboratory and clinical findings, and requires the SQiDworks™ Diagnostics Platform.

Device Description

The IgX PLEX™ Celiac Qualitative Assay is a consumable reagent kit. It is designed to run on the SQiDworks™ Diagnostics Platform. The kit includes a Microarray Plate, Reporter mix, standards, controls, sample diluents, wash buffer concentrates and a CD-ROM.

The Assay Kit detects the presences of the IgG classes of anti-tissue transglutaminase antibody. This is performed in an integrated fashion on the SQiDworks™ Diagnostics Platform (platform) that reports both analytes simultaneously to aid in the diagnosis of Celiac Disease.

The platform automates the entire immunoassay procedure from end-to-end, including calibrator/standards and sample pipetting, sample dilution, incubation, washing, and drying. Once the assay's biochemical reactions have completed, the instrument automatically performs a multi-color fluorescent scan of each well in the microarray, analyzes the data, and generates a report containing qualitative results for both assay markers. Results for each patient sample from the IgX PLEX™ Celiac Qualitative assay are obtained simultaneously for each of the two markers using the results from one well containing one aliquot of the patient's serum. Results are reported independently.

AI/ML Overview

Here's a breakdown of the acceptance criteria and study information for the IgX PLEX™ Celiac Qualitative Assay, based on the provided text:

Acceptance Criteria and Reported Device Performance

Acceptance Criteria CategorySpecific MetricAcceptance Criteria (Implied)Reported Device Performance
ReproducibilityAnti-tTG-IgA ReproducibilityHigh (not explicitly quantified but demonstrated by results)92.4% to 100%
Anti-tTG-IgG ReproducibilityHigh (not explicitly quantified but demonstrated by results)95.8% to 100%
Clinical PerformanceAnti-tTG-IgA Clinical SensitivityHigh (not explicitly quantified but demonstrated by results)98.3%
Anti-tTG-IgG Clinical SensitivityHigh (not explicitly quantified but demonstrated by results)80.9%
Anti-tTG-IgA Clinical SpecificityHigh (not explicitly quantified but demonstrated by results)94.5%
Anti-tTG-IgG Clinical SpecificityHigh (not explicitly quantified but demonstrated by results)89.0%
Agreement with PredicatesAnti-tTG-IgA Overall AgreementHigh (not explicitly quantified but demonstrated by results)89.3%
Anti-tTG-IgG Overall AgreementHigh (not explicitly quantified but demonstrated by results)85.1%
InterferenceEffect of high bilirubin levelsNo significant impactNone affected
Effect of high hemoglobin levelsNo significant impactNone affected
Effect of high triglycerides levelsNo significant impactNone affected
Effect of high human IgG levelsNo significant impactNone affected

Study Information

  1. Sample size used for the test set and the data provenance:

    • The document does not specify the sample size used for the test set.
    • The document does not specify the data provenance (e.g., country of origin, retrospective or prospective).
  2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    • This information is not provided in the document. The study focuses on "clinical sensitivity" and "specificity" and "overall agreement with established predicate test systems," which suggests comparison to existing diagnostic methods rather than a panel of human experts for ground truth in the traditional sense for image-based AI.
  3. Adjudication method (e.g., 2+1, 3+1, none) for the test set:

    • This information is not provided in the document.
  4. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • No, a multi-reader multi-case (MRMC) comparative effectiveness study was not performed. This device is an in vitro diagnostic (IVD) assay that automates the immunoassay procedure and provides qualitative results, intended to aid in diagnosis in conjunction with other clinical findings, rather than an AI system assisting human readers of images.
  5. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:

    • Yes, the performance metrics reported (reproducibility, clinical sensitivity, specificity, agreement) directly reflect the standalone performance of the IgX PLEX™ Celiac Qualitative Assay. The device is an automated system described as running "from end-to-end" and generating reports with "qualitative results." Human "in-the-loop" performance in the sense of an algorithm assisting a human reader is not applicable here as it autonomously provides the diagnostic result.
  6. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

    • The document implies that the ground truth for "clinical sensitivity" and "specificity" and "overall agreement" was established by comparison to established predicate test systems and presumably clinical diagnoses of celiac disease. It does not explicitly state the ultimate "ground truth" method (e.g., small intestinal biopsy for celiac disease diagnosis, which is the gold standard). The phrase "aid in the diagnosis of celiac disease in conjunction with other laboratory and clinical findings" suggests that the "true" diagnosis of celiac disease, against which the assay's performance is measured, would involve a combination of clinical information, and potentially the gold standard of biopsy.
  7. The sample size for the training set:

    • This information is not provided. The description mentions "standards, controls, sample diluents" as part of the kit, which are generally used for calibration and quality control rather than a "training set" in the machine learning sense.
  8. How the ground truth for the training set was established:

    • This information is not provided. As this is an IVD assay rather than a machine learning model, the concept of a "training set" and establishing its ground truth in the typical AI/ML context doesn't directly apply in the same way.

{0}------------------------------------------------

SQI DIAGNOSTICS
TM

JUN - 2 2011

36 Meteor Toronto ON M9W 1A4 [w] 416.674.9500 [f] 416.674.9300 www.sqidiagnostics.com

510(k) Summary

This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92.

Name:SQI Diagnostics Systems Inc.
Address:36 Meteor DriveToronto, OntarioCanadaM9W 1A4
Phone:416-674-9500 x 324
FAX:416-674-9300
Contact:Harumi Kumahara
Email:hkumahara@sqidiagnostics.com
Date Prepared:April 7, 2011
Trade Name:IgX PLEX™ Celiac Qualitative Assay
Common Name:Integrated system for the qualitative detection of IgA and IgG class ofanti-tissue transglutaminase in serum samples
Classification:Class II
Product Code:MVM-866.5660: Autoantibodies, Endomysial (TissueTransglutaminase)
SubstantialEquivalence:Aesku, Inc.- Aeskulisa tTG A and tTG G (K042644)

{1}------------------------------------------------

Image /page/1/Picture/0 description: The image shows the logo for SQI Diagnostics. The logo consists of the letters "SQI" in a bold, sans-serif font, with the word "DIAGNOSTICS" stacked vertically to the right of the "I". To the right of the word "DIAGNOSTICS" is a stylized image of a teardrop or droplet, with a small "TM" symbol in the upper right corner of the droplet.

nto ON M9W 1A4 16 674 9500 16.674.9300 ww.sqidiaqnostics.com

Intended Use

The IgX PLEX™ Celiac Qualitative Assay is an in vitro diagnostic test for the qualitative detection of the IgA and IgG immunoglobulin classes of anti-tissue transglutaminase antibody in serum. The test is intended for use in clinical laboratories as an aid in the diagnosis of celiac disease in conjunction with other laboratory and clinical findings, and requires the SQiDworks™ Diagnostics Platform.

Device Description

The IgX PLEX™ Celiac Qualitative Assay is a consumable reagent kit. It is designed to run on the SQiDworks™ Diagnostics Platform. The kit includes a Microarray Plate, Reporter mix, standards, controls, sample diluents, wash buffer concentrates and a CD-ROM.

The Assay Kit detects the presences of the IgG classes of anti-tissue transglutaminase antibody. This is performed in an integrated fashion on the SQiDworks™ Diagnostics Platform (platform) that reports both analytes simultaneously to aid in the diagnosis of Celiac Disease.

The platform automates the entire immunoassay procedure from end-to-end, including calibrator/standards and sample pipetting, sample dilution, incubation, washing, and drying. Once the assay's biochemical reactions have completed, the instrument automatically performs a multi-color fluorescent scan of each well in the microarray, analyzes the data, and generates a report containing qualitative results for both assay markers. Results for each patient sample from the IgX PLEX™ Celiac Qualitative assay are obtained simultaneously for each of the two markers using the results from one well containing one aliquot of the patient's serum. Results are reported independently.

Comparison to Predicates

The IgX PLEX™ Celiac Qualitative Assay uses the same ELISA assay principles of capture antigen, sample addition/incubation, reporter addition/incubation and fluorescent signal interpretation as the predicates with the improvements of using an automated system and multiplexing immunoassay. The intended use of the IgX PLEX™ Celiac Qualitative Assay and its predicates are equivalent, to aid in the diagnosis of Celiac Disease.

{2}------------------------------------------------

Image /page/2/Picture/0 description: The image shows the logo for SQI Diagnostics. The logo consists of the letters "SQI" in a bold, serif font, followed by the word "DIAGNOSTICS" stacked vertically to the right of the "I". To the right of the word "DIAGNOSTICS" is a stylized image of a teardrop or droplet, with a highlight suggesting a glossy surface. The "TM" symbol is located to the upper right of the droplet.

aidiaanostics.com

Test Studies

A series of test studies were conducted in support of the comparability, safety and effectiveness of the IgX PLEX™ Celiac Qualitative Study and its predicates. The studies completed with the results as follows:

  • Reproducibility ranges were 92.4% to 100% for anti-tTG-IgA and 95.8%-100% . for anti-tTG-lgG.
  • Clinical sensitivity was 98.3% for anti-tTG-IgA and 80.9% for anti-tTG-IgG. . Clinical specificity was 94.5% for anti-tTG-IgA and 89.0% for anti-tTG-IgG.
  • Overall agreement between the analytes in the IgX PLEX™ Celiac Qualitative . Assay and established predicate test systems was 89.3% for anti-tTG-lgA and 85.1% for anti-tTG-IgG.
  • . None of the analytes in the assay were affected by high levels of the following biological substances: bilirubin, hemoglobin, triglycerides and human IgG.

Conclusion

In conclusion, the IgX PLEX™ Celiac Qualitative Assay demonstrated performance, safety and effectiveness equivalent or superior to its predicates.

{3}------------------------------------------------

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Silver Spring, MD 20993

SQI Diagnostics Systems Inc. c/o Dr. Kate Smith Vice President, Regulatory Affairs 36 Meteor Drive Toronto, Ontario, Canada M9W 1A4

JUN 0 2 2011

Re: K102490

Trade/Device Name: IgX PLEXTM Celiac Qualitative Assay Regulation Number: 21 CFR 866.5660 Regulation Name: Multiple autoantibodies immunological test system Regulatory Class: Class II Product Code: MVM Dated: May 27, 2011 Received: May 31, 2011

Dear Dr. Smith:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act, The general controls provisions of the Act include requirements for annual registration. Ilsting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of

{4}------------------------------------------------

Page 2 - Dr. Kate Smith

medical device-related adverse events) (21 CFR 803); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820). This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Parts 801 and 809), please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (301) 796-5450. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely yours,

Reena Philip

City

Maria M. Chan, Ph.D. Director Division of Immunology and Hematology Devices Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

Indications for Use

510(k) Number (if known): K102490

Device Name: IgX PLEX™ Celiac Qualitative Assay

Indications For Use:

The IgX PLEX™ Celiac Qualitative Assay is an in vitro diagnostic test for the qualitative detection of the IgA and IgG immunoglobulin classes of anti-tissue transglutaminase antibody in serum. The test is intended for use in clinical laboratories as an aid in the diagnosis of celiac disease in conjunction with other laboratory and clinical findings, and requires the SQiDworks™ Diagnostics Platform.

Prescription Use (Part 21 CFR 801 Subpart D) AND/OR

Over-The-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of In Vitro Diagnostic Device Evaluation and Safety (OIVD)

Reena Philip

Division Sign-Off Office of In Vitro Diagnostic Device Evaluation and Safety

510(k) K102490

Page 1 of 1

§ 866.5660 Multiple autoantibodies immunological test system.

(a)
Identification. A multiple autoantibodies immunological test system is a device that consists of the reagents used to measure by immunochemical techniques the autoantibodies (antibodies produced against the body's own tissues) in serum and other body fluids. Measurement of multiple autoantibodies aids in the diagnosis of autoimmune disorders (disease produced when the body's own tissues are injured by autoantibodies).(b)
Classification. Class II (performance standards).