(217 days)
The Genuine First Aid CPR Face (with oxygen port) is single use designed for mouth to mask ventilation to health emergency victims requiring oxygen support (inhalation) or cardiopulmonary resuscitation (CPR) rescue techniques of a non-breathing adult. It is also used as a barrier that will direct expired air from the patient away from the user. This device should only be used by persons who have received adequate training.
Genuine First Aid CPR Face Mask and oxygen port, made up of medical grade PVC and one-way valve made up of medical grade K-resin. The mask is used for mouth-to-mask breathing. There is a shield between the person who gives respiration and the victim. Single use. CPR mask includes: One-way filter valve PVC mask with oxygen port Elastic Strap Packaged for easy portability and quick access
This document describes the performance testing of the Genuine First Aid CPR Mask.
1. Table of Acceptance Criteria and Reported Device Performance
| Performance Metric | Acceptance Criteria | Reported Device Performance |
|---|---|---|
| Expiratory resistance | < 5 cmH2O (at 50 L/min) | 3 cmH2O (at 50 L/min) |
| Inspiratory resistance | < 5 cmH2O (at 50 L/min) | 2 cmH2O (at 50 L/min) |
| In vitro cytotoxicity | Required (passes ISO10993-1,-5,-10) | Passes |
| Skin irritation | Required (passes ISO10993-1,-5,-10) | Passes |
| Delayed-type hypersensitivity | Required (passes ISO10993-1,-5,-10) | Passes |
2. Sample Size and Data Provenance for Test Set
The provided document does not specify a distinct "test set" in the context of a typical AI/software device evaluation. The performance metrics listed (expiratory and inspiratory resistance) refer to a single device (the Genuine First Aid CPR mask) being tested against established industry standards. Biocompatibility testing results are also presented for the device.
There is no information on:
- The sample size used for the resistance and biocompatibility tests.
- The country of origin of the data.
- Whether the data was retrospective or prospective.
3. Number of Experts and Qualifications for Ground Truth Establishment (Test Set)
Not applicable. This device is a physical medical device (CPR mask), not an AI or software device that requires expert-established ground truth for a test set. The performance is assessed against recognized engineering and biological standards.
4. Adjudication Method for Test Set
Not applicable. As noted above, this is a physical device, not an AI/software device requiring an adjudication process for ground truth.
5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study
Not applicable. This device is a CPR mask and not an AI or imaging device for which MRMC studies are typically conducted to assess the impact of AI assistance on human readers.
6. Standalone Performance
Yes, standalone performance was done. The document directly reports the performance of the Genuine First Aid CPR mask for expiratory and inspiratory resistance, and biocompatibility in isolation, without involving human users for comparative analysis in these specific tests.
7. Type of Ground Truth Used
The "ground truth" for the performance criteria is based on:
- Established industry standards: Specifically, ISO10651-4:2002 for lung ventilators (for resistance measurements) and ISO10993-1, -5, -10 for biological evaluation of medical devices (for biocompatibility). These international standards define the acceptable limits for the device's physical and biological properties.
8. Sample Size for Training Set
Not applicable. This is a physical medical device, not an AI/machine learning algorithm, and therefore does not have a "training set" in the sense of data used to train a model.
9. How Ground Truth for Training Set was Established
Not applicable. As this device does not involve an AI/machine learning model, there is no training set or associated ground truth establishment process.
{0}------------------------------------------------
510(k) SUMMARY
KI 8846
12-2261
MAR 1 8 2011
Submitter: Genuine First Aid LLC 600 Cleveland Street Suite 400 Clearwater FL 33755, USA Office: (727) 449 2150 (727) 449 2145 Fax: Contact: Desiree Jacques
Contact/Consultant: Gary Lehnus Lehnus & Associates Consulting 150 Cherry Lane Rd East Stroudsburg, PA 18301 USA (570) 620-0198 (570) 620-0199
Trade Name: Genuine First Aid CPR Mask
Common Name: Emergency CPR Mask
Classification Name: Valve, Non-Rebreathing Product Code: СВР Classification: Class II Panel: Anesthesiology 21CFR 868.5870 Requlatory Classification:
Medisource CPR Mask with Oxygen Port - K081516 Predicate Device:
Device Description:
Genuine First Aid CPR Face Mask and oxygen port, made up of medical grade PVC and one-way valve made up of medical grade K-resin. The mask is used for mouth-to-mask breathing. There is a shield between the person who gives respiration and the victim.
Single use. CPR mask includes: One-way filter valve PVC mask with oxygen port Elastic Strap Packaged for easy portability and quick access
Specifications:
Dimensions: (122mm x 100mm x 90mm) mask connector 22mm ID oxygen port 6mm OD Inspiratory resistance: < 5 cmH2O (at 50 L/min) Expiratory resistance: < < 5 cmH2O (at 50 L/min)
{1}------------------------------------------------
Operation Temperature: -18°C~ 40°C
Storage Temperature: -30℃ ~ 50℃
Storage relative humidity: <85%
Indications For Use:
The Genuine First Aid CPR Face (with oxygen port) is single use designed for mouth to mask ventilation to health emergency victims requiring oxygen support (inhalation) or cardiopulmonary resuscitation (CPR) rescue techniques of a non-breathing adult. It is also used as a barrier that will direct expired air from the patient away from the user. This device should only be used by persons who have received adequate training.
Substantial Equivalency Summary
| Features | Medisource | Genuine First Aid CPR mask |
|---|---|---|
| Configuration | One piece | One piece |
| Use | Single | Single |
| Size | Adult | Adult |
| Material | PVC、silicone、Non-Woven | PVC、silicone、Non-Woven |
| Intended Use | Mouth to mask ventilation | mouth to mask ventilation |
| Ventilation | manual | manual |
| Oxygen | Oxygen port optional use | Oxygen port optional use |
| Oxygen port | standard 6mm nipple | standard 6mm nipple |
| Connector | Standard 22 mm | Standard 22 mm |
| Expiratory resistance | $1.94 cmH2O$ ( at 50 L/min) | $3 cmH2O$ ( at 50 L/min) |
| Inspiratory resistance | $2.04 cmH2O$ (at 50 L/min) | $2 cmH2O$ ( at 50 L/min) |
| Includes | Universal breathing tube | Universal breathing tube |
| One-way filtered valve | One-way filtered valve | |
| Head strap | Head strap | |
| With or without oxygen port | With or without oxygen port |
PERFORMANCE TESTING
Biocompatibility testing:
| Biocompatibility test | ISO10993-1,-5,-10requirement | Testing results |
|---|---|---|
| vitro cytotoxicity | required | passes |
| skin irritation | required | passes |
| delayed-type | required | passes |
{2}------------------------------------------------
| -------------------- | ||
|---|---|---|
| nvnersensitivity100 AM And | Property of Children Warrant of the Property | STATUTE OF CONSULTER COLLECTION OF CLAIM |
Conclusion:
The potential for causing irritation is remote for the follow reasons:
- . The raw material and mask have been tested beyond the requirements of ISO10993 for skin contact.
- The time of contact and percentage of body exposure is very low. .
- The injection molding process does not significantly alter the raw material . characteristics
| Device component | Material ofconstruction | Patient /rescuercontact | Contact time |
|---|---|---|---|
| One way valve | K-resin/silicone | mouth (rescuer) | 3-60minutes |
| Face mask | PVC | Skin(Patient) | 3-60minutes |
| Elastic strap | Non-woven | Hair/skin(Patient) | 3-60minutes |
Patient contact components description
Expiratory resistance and Inspiratory resistance performance testing was done using the test methods described in ISO10651-4:2002, Lung ventilators-Part 4:Particular requirements for operator-powered resuscitators. Some minor modifications were made to the methods
| Device tested | Standard requirement | Genuine First Aid CPR mask |
|---|---|---|
| Expiratory resistance | <5 cmH2O ( at 50 L/min) | 3 cmH2O ( at 50 L/min) |
| Inspiratory resistance | <5 cmH2O ( at 50 L/min) | 2 cmH2O ( at 50 L/min) |
Recognized consensus standards
| ISO5356-1:2004 | Anaesthetic and respiratory equipment-Conicalconnectors - Part1: Cones and sockets |
|---|---|
| ISO10651-4:2002 | Lung ventilators-Part 4: Particular requirements foroperator-powered resuscitators |
| ISO10993-1:2003 | Biological Evaluation of Medical Device-Part 1:evaluation and testing |
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized depiction of an eagle or bird-like figure with three curved lines representing its wings or feathers. To the left of the emblem, the text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" is arranged in a circular pattern around the emblem.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
Genuine First Aid LLC C/O Mr. Gary Lehnus Lehnus & Associates Consulting 150 Cherry Lane Road East Stroudsburg, Pennsylvania 18301
MAR 1 8 201
Re: K102299
Trade/Device Name: Genuine First Aid CPR Face Mask Regulation Number: 21 CFR 868.5870 Regulation Name: Nonbreathing Valve Regulatory Class: II Product Code: CBP Dated: March 11, 2011 Received: March 16, 2011
Dear Mr. Lehnus:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{4}------------------------------------------------
Page 2- Mr. Lehnus
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801). please go to
http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportalProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
dam L Rosotad
for
Anthony D. Watson, B.S., M.S., M.B.A. Director
Division of Anesthesiology, General Hospital, Infection Control and Dental Devices
Office of Device Evaluation
Center for Devices and
Radiological Health
Enclosure
{5}------------------------------------------------
Indications for Use
510(k) Number (if known):
Device Name: _________________________________________________________________________________________________________________________________________________________________ Genuine First Aid CPR Face Mask
Indications For Use:
The Genuine First Aid CPR Face (with oxygen port) is single use designed for mouth to mask ventilation to health emergency victims requiring oxygen support (inhalation) or cardiopulmonary resuscitation (CPR) rescue techniques of a non-breathing adult. It is also used as a barrier that will direct expired air from the patient away from the user. This device should only be used by persons who have received adequate training..
Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (OIVD)
Y. Schutte
(Division Sign-Off) Division of Anesthesiology, General Hospital Infection Control, Dental Devices
510(k) Number: K102299
Page 1_of 1
§ 868.5870 Nonrebreathing valve.
(a)
Identification. A nonrebreathing valve is a one-way valve that directs breathing gas flow to the patient and vents exhaled gases into the atmosphere.(b)
Classification. Class II (performance standards).