Search Filters

Search Results

Found 31 results

510(k) Data Aggregation

    K Number
    K241366
    Date Cleared
    2024-10-29

    (168 days)

    Product Code
    Regulation Number
    868.5870
    Reference & Predicate Devices
    Why did this record match?
    510k Summary Text (Full-text Search) :

    device trade name: QualityFlow O2 Demand Valve Common name: Valve, Non-rebreathing Classification name: 868.5870
    |
    | Regulation Number | 21 CFR 868.5870
    | 21 CFR 868.5870

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The demand valve is used for the application of 100 vol% oxygen during manual ventilation with the resuscitator bag, as well as during direct non-invasive inhalation with the resuscitator mask in spontaneously breathing patients.

    Device Description

    The valve, non - rebreathing QualityFlow O2 demand valve is a medical device to deliver medical oxygen or a medical air - oxygen mixture to a patient on demand. Unlike continuous oxygen systems, the demand valve only supplies medical gas when the patient inhales, reducing oxygen consumption and ensuring the patient receives gas only when needed. The QualityFlow O2 Demand Valve from DEHAS consists of a membrane that is activated by negative pressure during inspiration and a valve mechanism that controls gas flow. During inspiration, the valve opens due to the movement of the membrane, while during expiration it is closed by a return spring. The function of the demand valve is based on providing gas only when a negative pressure is applied to the membrane, e.g. by the patient inhaling with a connected mask or suction of the resuscitator when connected (expansion after previous compression of the resuscitator). The valve therefore ensures that gas only flows when it is needed.

    AI/ML Overview

    The provided document is a 510(k) Premarket Notification for a medical device called "QualityFlow O2 Demand Valve". This type of submission aims to demonstrate that a new device is substantially equivalent to a legally marketed predicate device, and thus does not require clinical studies in many cases.

    Based on the provided text, the device in question (QualityFlow O2 Demand Valve) is a mechanical device, not an AI/software-driven one. Therefore, the questions related to AI/MRMC studies, training/test sets, ground truth establishment by experts, and adjudication methods are not applicable to this submission. The "acceptance criteria" discussed in this document refer to the device's ability to meet various engineering, safety, and performance standards, rather than diagnostic accuracy metrics common for AI/ML devices.

    Here's an analysis based on the information relevant to this type of medical device submission:

    1. Table of Acceptance Criteria (Relevant Standards) and Reported Device Performance:

    The document summarizes non-clinical tests performed according to international standards to demonstrate substantial equivalence and that the device performs as intended.

    Acceptance Criteria (Standard Used)Description of TestReported Device Performance
    ISO 5356-1 Third edition 2004-05-15: Anaesthetic and respiratory equipment - Conical connectors: Part 1: Cones and socketsSpecifies the dimension and dimensional requirements for cones and bushings intended for the connection of anesthesia and ventilator equipment (e.g., in ventilation systems, anesthetic gas transport systems and vaporizers).Result: pass
    IEC 62366-1 Edition 1.1 2020-06 CONSOLIDATED VERSION: Medical devices - Part 1: Application of usability engineering to medical devicesUsed to verify and validate the usability of the QualityFlow O2 demand valve.Result: pass
    ISO 14971 Third Edition 2019-12: Medical devices - Application of risk management to medical devicesProcess for identification and assessment of hazards and associated risks, control of these risks, and monitoring effectiveness of risk control measures.Result: pass
    ISO 15001:2010: Anesthetic and respiratory equipment - Compatibility with oxygenUsed to ensure the oxygen compatibility of the used materials.Result: pass
    ISO 10993-1 Fifth edition 2018-08: Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management processUsed for biological evaluation of medical devices and testing within a risk management process.Result: pass
    ISO 18562-1 First edition 2017-03: Biocompatibility evaluation of breathing gas pathways in healthcare applications - Part 1: Evaluation and testing within a risk management processEvaluated biocompatibility of breathing gas pathways, specifically for indirect patient contact via medical oxygen (gas pathway contact, limited exposure
    Ask a Question

    Ask a specific question about this device

    K Number
    K234053
    Date Cleared
    2024-08-09

    (231 days)

    Product Code
    Regulation Number
    868.5870
    Reference & Predicate Devices
    Why did this record match?
    510k Summary Text (Full-text Search) :

    2013 New Zealand

    Re: K234053

    Trade/Device Name: F&P Optiflow Flow Diverter Regulation Number: 21 CFR 868.5870
    |
    | Regulatory Class | Class II (21 CFR §868.5870

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    This product is intended for use as an in-line pressure relief device, designed to limit the system pressure within the limits of its stated technical specifications.
    This product is indicated for patients in a hospital for use by appropriately qualified healthcare professionals.

    Device Description

    The F&P Optiflow Flow Diverter is intended for use as an in-line pressure relief device, designed to limit the system pressure when used with the compatible F&P Optiflow Oxygen Kits and nasal interfaces.
    The F&P Flow Diverter is placed between the flow meter and the dryline during use. It is a multi-patient use prescription-only device, provided in a non-sterile state. It operates at flow rates between 5 to 70 L/min and is intended to be used by appropriately qualified healthcare professionals in hospitals.
    When an anesthesia mask is applied over the Optiflow Switch interface, the respiratory gas no longer has a pathway to flow through the nasal interface to the patient. This will trigger respiratory gas flow to be diverted into the atmosphere.

    AI/ML Overview

    The medical device in question is the "F&P Optiflow™ Flow Diverter", a nonrebreathing valve (Product Code: CBP).

    Here's an analysis of the acceptance criteria and study information provided:

    1. Table of Acceptance Criteria and Reported Device Performance:

    The document does not explicitly provide a table of acceptance (performance) criteria with numerical targets. Instead, it states that "Performance testing of the Flow Diverter was completed and confirms the subject device does not raise new questions of safety and effectiveness." It implies that the device met the safety and performance standards outlined in the listed ISO and IEC standards.

    However, based on the comparison table with the predicate device, we can infer some key performance characteristics. The document emphasizes equivalence to the predicate device, the "OJR215 Pressure Relief Manifold (K173770)".

    Characteristic / Acceptance Criteria (Inferred)F&P Optiflow™ Flow Diverter (Subject Device) PerformanceOJR215 Pressure Relief Manifold (Predicate Device) PerformanceComments (for equivalence)
    Indications for UseIn-line pressure relief, limits system pressure within stated specs, for hospital use by qualified professionals.Protects neonates, infants, children from excessive inspiratory pressure in downstream occlusion in continuous flow systems via nasal cannula. For flow rates 0.5-36 L/min. Upstream of patient, single use, prescription only.Equivalent (despite slight wording differences, the core function of pressure relief in a respiratory circuit is comparable).
    AvailabilityPrescription use (Part 21 CFR 801 Subpart D)Prescription use (Part 21 CFR 801 Subpart D)Identical
    Intended Use EnvironmentHospitalHospital/InstitutionalIdentical
    UsersQualified healthcare professionalsQualified healthcare professionalsIdentical
    Ambient Operating Temperature18 to 26 °C (64.4 to 78.8°F)18 to 26 °C (64.4 to 78.8°F)Identical
    Shelf LifeUp to 7 years3 yearsDifferent (Subject device has longer shelf life, which is generally a performance improvement from a user's perspective, but would have required stability testing).
    Storage Temperature-10°C to +50 °C (14 to 122 °F)-10°C to +50 °C (14 to 122 °F)Identical
    SterilityDevice not provided sterile.Device not provided sterile.Identical
    Operating Pressure - Maximum PressureMaximum system pressure of 60cmH2O at maximum rated flow rate of 70L/minMaximum relief pressure is 75 cm H2O at maximum rated flow of 36 L/min.Equivalent (Both devices are designed to provide pressure relief; the specific maximum pressures and flow rates differ but are within acceptable ranges for their respective claims. The subject device operates at a lower maximum pressure but at a higher maximum flow).
    Flow Range5-70 L/min0.5-36 L/minEquivalent (Both cover a range of flow rates, the subject device covers a higher range).
    Application/TherapyNasal High Flow TherapyNasal High Flow TherapyIdentical
    Principles of OperationValve seals, allows gas flow; if relief pressure exceeded, shaft moves, seal opens, pressure relieved, flow diverted to atmosphere.Valve seals, allows gas flow; if relief pressure exceeded, shaft moves, seal opens, pressure relieved, flow diverted to atmosphere.Equivalent
    Inlet ConnectionDISS 1240Inlet barbDifferent (Requires compatibility verification)
    Outlet ConnectionF&P female safety connection22mm female taper connection as per ISO 5356-1Different (Requires compatibility verification)
    Housing MaterialPolycarbonateABSEquivalent (Both are suitable medical-grade plastics).
    Plunger Seal FunctionRests on valve body, accurately relieves pressure by opening when relief pressure is reached.Seals manifold valve during normal use, accurately relieves pressure by opening when relief pressure is reached.Equivalent
    Spring Pressure Relief Valve MechanismShaft moved by valve seal that moves when set pressure is exceeded.Shaft moved by a spring pressure relief valve that moves when set pressure is exceeded.Equivalent
    Manifold Shroud / Tamper PreventionRear and front caps prevent user modification to preset pressure relief.Prevents access to pre-set operating pressures and tampering.Equivalent

    2. Sample Size Used for the Test Set and Data Provenance:

    The document does not explicitly state a sample size for a "test set" in the context of a clinical performance study. The testing described is non-clinical performance testing against standards. Therefore, concepts like sample size for clinical test data, country of origin, or retrospective/prospective data are not applicable here. The testing would have involved multiple units of the device under various conditions described by the cited standards.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications:

    Not applicable. This is not a clinical study involving expert interpretation or "ground truth" establishment in a diagnostic context. The performance is assessed against engineering and safety standards.

    4. Adjudication Method for the Test Set:

    Not applicable. There's no human interpretation or diagnostic assessment that would require an adjudication method.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, and the effect size of how much human readers improve with AI vs without AI assistance:

    Not applicable. This device is a mechanical pressure relief valve, not an AI-powered diagnostic tool.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    Not applicable. This is not an algorithm-based device.

    7. The type of ground truth used:

    The "ground truth" for this device would be its performance against the objective criteria defined in the referenced industry standards (e.g., ISO 80601-2-74:2017 for respiratory humidifying equipment, IEC 60601-1 for basic safety and essential performance, ISO 80601-2-90:2021 for respiratory high-flow therapy equipment). This includes measurements such as maximum pressure relief, flow rates, temperature tolerances, and mechanical integrity, as well as adherence to design principles and safety features.

    8. The sample size for the training set:

    Not applicable. This is a medical device, not a machine learning model requiring a training set.

    9. How the ground truth for the training set was established:

    Not applicable.

    In summary:

    The document describes a 510(k) submission for a non-AI, mechanical medical device. The "acceptance criteria" are primarily established by compliance with recognized international and national standards for medical devices and respiratory equipment (ISO, IEC, CGA), as well as demonstrating substantial equivalence to a predicate device. The "study" proving this is a series of non-clinical performance and safety tests conducted in accordance with these standards. The concepts of clinical test sets, ground truth by experts, and training sets for algorithms are not relevant to this type of device submission.

    Ask a Question

    Ask a specific question about this device

    K Number
    K210992
    Date Cleared
    2021-12-23

    (265 days)

    Product Code
    Regulation Number
    868.5870
    Reference & Predicate Devices
    Why did this record match?
    510k Summary Text (Full-text Search) :

    Re: K210992

    Trade/Device Name: Exhalation Valve (Single Limb, Dual Limb) Regulation Number: 21 CFR 868.5870
    |
    | Regulation Number: | 21 CFR 868.5870
    |
    | Regulation Number: | 21 CFR 868.5870
    |
    | | CFR 868.5870
    | CFR 868.5870

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The exhalation valves are intended to be used in ventilator circuits consistent with the indicated use of the ventilator to which they are attached.

    The single limb model directs flow and allows for pressure monitoring and may be used within hospitals and for home care use.

    The dual limb model regulates flow and is for use in hospitals and for home care use.

    The exhalation valve are used with adults, pediatrics, and neonates.

    The exhalation valves are for single patient use and can be used for a maximum of 30 days.

    Device Description

    The Altech® Exhalation Valves (Single Limb) are available in 2 configurations intended for use in Single and Dual Limb ventilator circuits. The difference between the configurations is location of exhaust of the expired gas to the room. In the Single limb, the valve is integrated to the limb and expired gas exhausts through the opening below the valve whereas in the valve is attached to the end of the exhalation limb and the expired gas exhausts through device end of the valve.

    AI/ML Overview

    The Altech® Exhalation Valves (Single Limb and Dual Limb) underwent non-clinical performance testing to demonstrate substantial equivalence to the predicate device.

    1. Table of Acceptance Criteria and Reported Device Performance:

    Test MethodologyAcceptance CriteriaReported Device Performance
    ISO 5356-1 (Conical fittings)Meet performance criteriaFittings were tested according to ISO 5356-1 and meet the performance criteria
    ISO 5367 (Leakage)1Toxicological Risk Assessment resulted in Margin of Safety >1
    Leakage in main bodyNot explicitly stated as acceptance criteria, but compared to predicate (predicate had
    Ask a Question

    Ask a specific question about this device

    K Number
    K192285
    Device Name
    CPR Face Shield
    Date Cleared
    2020-11-25

    (461 days)

    Product Code
    Regulation Number
    868.5870
    Reference & Predicate Devices
    N/A
    Why did this record match?
    510k Summary Text (Full-text Search) :

    Pudong Shanghai, 200120 Cn

    Re: K192285

    Trade/Device Name: CPR Face Shield Regulation Number: 21 CFR 868.5870

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    This device is used as a physical barrier for mouth to mouth resuscitation. The use of this device does not guarantee complete protection from disease transmission.

    Device Description

    Not Found

    AI/ML Overview

    I am sorry, but the provided text from the FDA 510(k) clearance letter for the "CPR Face Shield" (K192285) does not contain any information about a study proving the device meets acceptance criteria.

    The document is a standard FDA clearance letter, which confirms that the device is "substantially equivalent" to legally marketed predicate devices. It discusses regulatory matters such as registration, listing, manufacturing practices, and adverse event reporting.

    Therefore, I cannot fulfill your request to describe the acceptance criteria and the study that proves the device meets the acceptance criteria, as this information is not present in the provided text.

    The document details:

    • The device name: CPR Face Shield
    • Regulation Number: 21 CFR 868.5870 (Nonrebreathing Valve)
    • Regulatory Class: Class II
    • Product Code: CBP
    • Indications for Use: "This device is used as a physical barrier for mouth to mouth resuscitation. The use of this device does not guarantee complete protection from disease transmission."

    None of these sections imply or describe performance studies with acceptance criteria as one would expect for an AI/software as a medical device (SaMD) or other complex diagnostic/therapeutic device. For a simple physical barrier like a CPR face shield, the "acceptance criteria" for FDA clearance primarily revolve around substantial equivalence to existing devices, safety considerations inherent to its material and design, and its stated intended use. A formal "study" with the characteristics you outlined (test sets, experts, MRMC, etc.) would not typically be required or relevant for this type of device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K173770
    Date Cleared
    2019-04-05

    (480 days)

    Product Code
    Regulation Number
    868.5870
    Reference & Predicate Devices
    N/A
    Why did this record match?
    510k Summary Text (Full-text Search) :

    2013 NZ

    Re: K173770

    Trade/Device Name: OJR215 Pressure Relief Manifold Regulation Number: 21 CFR 868.5870

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The OJR215 Pressure Relief Manifold is designed to protect neonates, infants and children from excessive inspiratory pressure in the event of a downstream occlusion occurring in continuous flow breathing gas delivery systems via nasal cannula.

    The device is intended for use with flow rates greater than 0.5 L/min up to, and including. 36 L/min. The OJR215 is fitted upstream of the patient. The OJR215 is single use only and is prescription only.

    Device Description

    Not Found

    AI/ML Overview

    I am sorry, but I cannot answer the question with the provided context. The document is an FDA 510(k) clearance letter and Indications for Use statement for a medical device called the "OJR215 Pressure Relief Manifold." This document explains the device's intended use and regulatory classification but does not include information about acceptance criteria, device performance studies, sample sizes, expert qualifications, or ground truth establishment. These details would typically be found in the 510(k) submission itself or related study reports, which are not part of this document.

    Ask a Question

    Ask a specific question about this device

    K Number
    DEN170044
    Device Name
    ClearMate
    Date Cleared
    2019-03-14

    (573 days)

    Product Code
    Regulation Number
    868.5480
    Type
    Direct
    Reference & Predicate Devices
    Why did this record match?
    510k Summary Text (Full-text Search) :

    bags (21 CFR 868.5320), oxygen cannulas (21 CFR 868.5340), masks (21 CFR 868.5550), valves (21 CFR 868.5870
    cannulas (21 CFR 868.5340, Class I), masks (21 CFR 868.5550, cleared under K953107), valves (21 CFR 868.5870

    Device Description :

    cannulas (21 CFR 868.5340, Class I), masks (21 CFR 868.5550, cleared under K953107), valves (21 CFR 868.5870

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    ClearMate™ is intended to be used by emergency department medical professionals as an adjunctive treatment for patients suffering from carbon monoxide poisoning. The use of ClearMate" enables accelerated elimination of carbon monoxide from the body by allowing isocapnic hyperventilation through simulated partial rebreathing.

    Device Description

    This device is intended to induce isocapnic hyperventilation in patients to speed up elimination of carbon monoxide (CO). Isocapnic hyperventilation can be defined as large increases in patient minute volume with minimal changes in arterial partial pressure of carbon dioxide (CO2). This device replaces CO2 levels in the airway, thereby maintaining CO2 levels in the blood that ultimately causes hyperventilation. This pneumatic device initially provides 100% supplemental oxygen (O2) at minute volumes selected based on patient weight. If the patient minute volume demand is more than the preset supplement O2 volume, this device supplies a mixture of 94%/6% (O2/CO2), which maintains CO2 levels in the airway to enable isocapnic breathing by partial simulated rebreathing (of CO2). This device consists of:

    1. The subject of this De Novo, the Control unit ("briefcase"), connects to sources of O2 and CO2 (neither gas is supplied with this device). The unit includes pressure gauges to read the source gas pressures. Internal components control supplemental gas flowrates, gas concentrations, and CO2 diversion away from the gas delivery pathway should O2 pressures be insufficient. This unit weighs about 2 kg and is pneumatically driven (i.e., no electronics).
    2. Two breathing circuits, which are not the subject of this De Novo, can attach to the gas outlet ports of the control unit. These circuits are constructed of reservoir bags (21 CFR 868.5320, Class I), oxygen cannulas (21 CFR 868.5340, Class I), masks (21 CFR 868.5550, cleared under K953107), valves (21 CFR 868.5870, cleared under K142402), resuscitation bags (21 CFR 868.5915, cleared under K912203), and/or tubing (21 CFR 868.5925, cleared under K161420).
    3. Hoses for source gas connections and a device stand for steadying the device, which are a subject of this De Novo.
    AI/ML Overview

    Here's a breakdown of the acceptance criteria and the study information for the ClearMate device, based on the provided text:

    Acceptance Criteria and Device Performance

    Acceptance Criteria (Special Controls - 21 CFR 868.5480)Reported Device Performance
    Non-clinical performance testing demonstrates device performs as intended under anticipated conditions of use, including:Verified through bench testing.
    A. Gas concentration accuracy testing for the range of intended concentrations;Control Unit Performance: Ten control units were tested. "Operation of supplemental O2 concentration, blended gas concentrations... All predefined specifications were met."
    B. Airway pressure delivery accuracy testing;Breathing Circuit Performance: "One-way valves (e.g., relief valves, demand valves) were individually tested for minimum pressure responses. Testing validated the worst case range of pressures experienced in the circuits, leakage rate of the circuit... All predefined specifications were met." Control Unit Performance: Not directly mentioned, but implied through the overall performance of the circuits.
    C. Supplemental O2 flowrate accuracy testing;Breathing Circuit Performance: "Supplemental O2 flowrate accuracy specifications. Circuits were tested to facilitate the correct gas concentrations output by the control unit... All predefined specifications were met." Control Unit Performance: "Supplemental O2 flowrate accuracy. All predefined specifications were met."
    D. Alarm testing;Control Unit Performance: Ten control units were tested. "Low O2 pressures (upper and lower limits) that trigger the CO2-driven alarm. CO2 diversion to drive the alarm... All predefined specifications were met."
    E. Use life testing.Use life Performance: Challenged 5 device samples at double the typical maximum breaths per minute and 1.4 times the typical adult tidal volume for 10 hours (simulating 8 patients/year over 5 years). "The results confirm supplemental O2 output and O2/CO2 demand valve operating specifications, which are the components subjected to most repeated use over the use life."
    The patient-contacting components of the device must be demonstrated to be biocompatible.Biocompatibility/Materials: A biological risk assessment was performed in accordance with ISO 10993-1. Components (face mask, adapters, resuscitation bag, leaflet valve) were either previously cleared with similar biocompatibility concerns or tested for dry gas pathway concerns (particulate matter and VOCs). "Based on the submitted testing and evaluations, the applicant has demonstrated device biocompatibility for this intended use."
    Labeling must include:Labeling is required to inform proper use.
    A. Instructions for use;Required.
    B. A precaution that monitoring of capnography is necessary during treatment with nonspontaneously breathing patients; andRequired. "When providing treatment to a non-spontaneously breathing patient using the ClearMate™ non-spontaneous breathing patient circuit, CO2 monitoring equipment for the measurement of expiratory carbon dioxide concentration must be used." and "A precaution that monitoring of capnography is necessary during treatment with nonspontaneously breathing patients."
    C. Use life specification.Required. "The applicant proposed a 5-year use life."

    Study Information

    2. Sample Size Used for the Test Set and Data Provenance:

    The document primarily discusses preclinical (animal and bench) and clinical literature review, rather than a single prospective "test set" for the device itself in the traditional sense of an AI/algorithm.

    • Bench Testing:

      • Breathing Circuit Performance: 10 samples of spontaneously breathing circuits and 10 samples of non-spontaneously breathing circuits (20 samples total).
      • Control Unit Performance: 10 control units.
      • Use Life Performance: 5 device samples.
      • Data Provenance: Retrospective, conducted by the applicant (Thornhill Research, Inc.) presumably in Canada (given the contact address is in Scottsdale, AZ, but the company is Thornhill Research, Inc., which is Canadian). This is inferred as internal testing presented to the FDA.
    • Animal Studies:

      • Fisher JA et al. (1999): Dog model (specific N not provided in the summary). Provenance: Published U.S. study.
      • Krech T et al. (2001): Mechanically ventilated sheep (specific N not provided in the summary). Provenance: Published U.S. study.
    • Human Clinical Literature Review (considered as evidence of effectiveness):

      • Anand et al. (2017): 13 healthy volunteers (chronic smokers). Provenance: Published study (journal name "PLOS One" suggests an international publication, DOI:10.1371/journal.pone.0170621).
      • Rucker et al. (2002): 14 healthy volunteers. Provenance: Published U.S. study.
      • Takeuchi et al. (2000): Healthy human volunteers (specific N not provided in the summary). Provenance: Published U.S. study.
      • Katznelson et al. (2008): Specific N not provided in the summary (anesthetic agents study). Provenance: Published study.
      • Katznelson et al. (2011): 44 obese elective surgical patients (anesthetic agents study). Provenance: Published study.
      • Wu et al. (2015) - external study, not used for efficacy but for safety: 319 patients in treatment group, 320 in control group. Provenance: Chinese study, published in "Chinese Journal of Clinicians."

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications:

    This device is not an AI/algorithm requiring expert-established ground truth for a test set in the diagnostic sense. The "ground truth" for its performance is derived from:

    • Bench Testing: Engineering specifications and physical measurements. Experts involved would be engineering and quality control personnel. Their specific qualifications are not detailed, but they would be presumed to be qualified engineers/technicians.
    • Animal Studies: Scientific observation and measurement by researchers/veterinarians.
    • Human Clinical Studies: Clinical endpoints (e.g., COHb levels, elimination half-life, cerebral blood flow) measured by medical professionals and researchers. The peer-review process for these published studies implicitly involves expert clinicians and scientists.

    4. Adjudication Method for the Test Set:

    Not applicable, as this is not an AI/diagnostic algorithm using expert review for a "test set." For the literature review, the FDA's internal review team served as the adjudicators of the existing scientific evidence.

    5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study was done:

    No, an MRMC comparative effectiveness study, in the context of human reader performance with/without AI assistance, was not performed. The device is a physical therapeutic device, not a diagnostic AI system intended to assist human readers.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done:

    Yes, the device itself is the "standalone" component. Its performance was evaluated independently through bench testing (as detailed in item #2) and animal studies. The human clinical studies also assess the device's effect on physiological markers when used therapeutically.

    7. The Type of Ground Truth Used:

    • Bench Testing: Device specifications, physical properties, engineering standards, and direct measurement of gas concentrations, flow rates, and pressures.
    • Animal Studies: Physiological measurements (e.g., COHb levels, cardiac output, oxygen delivery) directly measured from the animal subjects.
    • Human Clinical Studies (for effectiveness): Physiological measurements (e.g., COHb levels, elimination half-life, cerebral blood flow) obtained from human volunteers or patients using established medical measurement techniques.
    • Human Clinical Study (Wu et al. for safety, but with limitations on efficacy): Clinical outcomes adjudicated by medical personnel based on consciousness, continence, disappearance of CO poisoning signs, EEG findings, and BI (Barthel Index) scores.

    8. The Sample Size for the Training Set:

    Not applicable. This is a physical medical device, not an AI model requiring a training set in the machine learning sense. The device's design is based on scientific principles of isocapnic hyperventilation, not data training.

    9. How the Ground Truth for the Training Set was Established:

    Not applicable (as above). The design and performance targets for the ClearMate device are based on established physiological principles and engineering requirements, not on a "ground truth" derived from a training dataset for an AI algorithm.

    Ask a Question

    Ask a specific question about this device

    Why did this record match?
    510k Summary Text (Full-text Search) :

    Oxygen Port, Non-Rebreathing Valve (15mmOD), Non-Rebreathing Valve (18.5mm) Regulation Number: 21 CFR 868.5870
    Non-Rebreathing Valve Classification Name: Non-Rebreathing Valve Regulation Class: II Regulation Number: 868.5870
    |
    | FDA Classification | 868.5870
    | 868.5870
    | 868.5870

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The CPR Mask with Oxygen Port is designed to assist in providing immediate life support (mouth to mask ventilation) to health emergency victims requiring oxygen support (inhalation) or cardiopulmonary resuscitation (CPR) rescue techniques. The CPR Mask with Oxygen Port is intended for prescription use.

    The CPR Mask without Oxygen Port is designed to assist in providing immediate life support (mouth to mask ventilation) to health emergency victims requiring cardiopulmonary resuscitation (CPR) rescue techniques. The CPR Mask without Oxygen Port is intended for over-the-counter use.

    The Infant CPR Mask without Oxygen Port is designed to provide immediate life support (mouth to mask ventilation) to health emergency victims requiring cardiopulmonary resuscitation (CPR) rescue techniques. The Infant CPR mask is indicated for use on infants with body weight up to 10 kg (22 lbs). The Infant CPR Mask is intended for over-the-counter use.

    The Non-Rebreathing Valve is a single-use, non-sterile device intended to be used with a ventilation mask to provide mouth-to-mask ventilation to health emergency victims requiring cardiopulmonary resuscitation (CPR) rescue techniques. It is intended for over-the-counter use.

    Device Description

    Compower Emergency CPR masks with non-rebreathing valves are designed to assist in providing immediate life support to health emergency victims requiring oxygen support and cardiopulmonary resuscitation rescue techniques. The devices subject to this 510(k) include the following: CPR Mask with Oxygen Port, CPR Mask without Oxygen Port, Infant CPR Mask without Oxygen Port, Non-Rebreathing Valve (15mmOD), Non-Rebreathing Valve (18.5mm).

    CPR Mask with Oxygen Port: comprised of an air cushion, foldable mask cover with oxygen port, a non-rebreathing valve and an elastic head strap. Constructed of transparent polyvinyl chloride. Provided non-sterile and is intended for single use only.

    CPR Mask without Oxygen Port: comprised of an air cushion, foldable mask cover, a non-rebreathing valve and an elastic head strap. Constructed of transparent polyvinyl chloride. Provided non-sterile and is intended for single use only.

    Infant CPR Mask without Oxygen Port: comprised of an air cushion, mask cover and non-rebreathing valve. Constructed of transparent polyvinyl chloride. Provided non-sterile and is intended for single use only.

    Non-Rebreathing Valve: a non-sterile, single-use device designed with a rigid styrene-butadiene copolymer housing, a silicone valve plate and polypropylene filter. Compatible for use with 22mm connectors used in standard ventilation masks. Designed with 15mmOD and 18.5mm ports.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and study information based on the provided document:

    1. Table of Acceptance Criteria and Reported Device Performance

    The document presents performance data primarily in the context of comparison to predicate devices, rather than explicit acceptance criteria with pass/fail values. The "acceptance criteria" are implied by meeting or falling within ranges comparable to the predicate devices and adherence to relevant standards.

    Comparison of Compower CPR Mask (with/without Oxygen Port) to KYOLING CPR Mask (Predicate)

    MetricAcceptance Criteria (Implied by Predicate & Standard)Reported Compower CPR Mask PerformanceReported KYOLING CPR Mask Performance (Predicate)
    Inspiratory Resistance Specification≤0.5 kPa at 50 L/min (from AS 4259)0.40 - 0.49 kPa at 50 L/min0.30 - 0.31 kPa at 50 L/min
    Expiratory Resistance Specification≤0.5 kPa at 50 L/min (from AS 4259)0.20 - 0.30 kPa at 50 L/min0.31 - 0.32 kPa at 50 L/min
    Biocompatibility - CytotoxicityPass (ISO 10993-5)PassSame (Implied by Predicate Comparison)
    Biocompatibility - SensitizationPass (ISO 10993-10)PassSame (Implied by Predicate Comparison)
    Biocompatibility - IrritationPass (ISO 10993-10)PassN/A (Not explicitly listed for predicate)
    Standards UsedConformance to ISO 5356-1 and AS 4259-1995Conformance demonstratedConformance demonstrated
    Visual InspectionImplied to meet specificationsPassedN/A
    Dimensional MeasurementImplied to meet specificationsPassedN/A
    TransparencyImplied to meet specificationsPassedN/A
    Function after Contamination/ImmersionImplied to meet specificationsPassedN/A
    Performance After Storage and AgingImplied to meet specificationsPassedN/A
    Mean Delivered Oxygen ConcentrationImplied to meet specificationsPassedN/A

    Comparison of Compower Infant CPR Mask to EMS Child/Infant Mask (Primary Predicate) & Laerdal Paediatric Pocket Mask (Reference)

    MetricAcceptance Criteria (Implied by Predicate/Reference & Standard)Reported Compower Infant CPR Mask PerformanceReported EMS Child/Infant Mask Performance (Predicate)Reported Laerdal Paediatric Pocket Mask Performance (Reference)
    Inspiratory Resistance Specification
    Ask a Question

    Ask a specific question about this device

    K Number
    K152521
    Date Cleared
    2016-06-03

    (274 days)

    Product Code
    Regulation Number
    868.5870
    Reference & Predicate Devices
    Why did this record match?
    510k Summary Text (Full-text Search) :

    Wisconsin 53703

    Re: K152521

    Trade/Device Name: Seal Rite Non-Rebreathing Valve Regulation Number: 21 CFR 868.5870
    |
    | Regulatory Classification | 21 CFR 868.5870

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Seal Rite Non-Rebreathing Valve is a single-use, non-sterile device intended to be used with a ventilation mask (without oxygen port) to provide mouth-to-mask ventilation to health emergency victims requiring cardiopulmonary resuscitation ("CPR") rescue techniques. It is intended for over-the-counter use.

    Device Description

    The Seal Rite™ Non-Rebreathing Valve ("Seal Rite Valve") is a resuscitation valve designed for resuscitation using expired air for ventilation. It has fittings on the grooved patient end that will adapt to standard masks (22 mm ID) used for resuscitation. The Seal Rite Valve allows the rescuer to blow air into a mask through its silicone one-way valve and directs the patient's exhaled air away from the rescuer via a rigid one-way valve and multiple exhaust ports. The Seal Rite Valve helps protect the rescuer from potential patient contamination. The Seal Rite Valve does not have an oxygen port.

    AI/ML Overview

    The provided text is a 510(k) summary for the Seal Rite Non-Rebreathing Valve (K152521). It describes the device, its intended use, and the testing conducted to establish its substantial equivalence to a predicate device.

    Here's an analysis of the acceptance criteria and study information based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance

    The device was tested to ensure conformance with the FDA Recognized AS 4259-1995 standard ("Ancillary devices for expired air resuscitation") and passed all testing requirements. The table below summarizes the key performance parameters tested and the reported results. The explicit acceptance criteria (e.g., maximum allowable resistance values per the standard) are not fully detailed in the provided text, but the text states that the device "passed all testing requirements" and "performs as intended and meets the requirements of the AS 4259-1995 standard." I've extracted specific quantitative performance values where available and indicated "Pass" for qualitative tests.

    Acceptance Criteria (Implied by standard AS 4259-1995)Reported Device Performance (Seal Rite™ Non-Rebreathing Valve)Predicate Device Performance (Respironics Rescue Valve)
    Biocompatibility:
    Meet Cytotoxicity requirements (ISO 10993-5)PassNot explicitly stated for predicate in this document
    Meet Irritation requirements (ISO 10993-10)PassNot explicitly stated for predicate in this document
    Meet Sensitization requirements (ISO 10993-12)PassNot explicitly stated for predicate in this document
    Bench Testing (Functional Performance/Conformity to AS 4259-1995):
    Ventilation Performance requirementsPassNot explicitly stated for predicate in this document
    Expiratory Resistance for the Patient requirementsPassNot explicitly stated for predicate in this document
    Expiratory Resistance for the Rescuer requirementsPassNot explicitly stated for predicate in this document
    Inspiratory Resistance for a Spontaneously Breathing Patient requirementsPassNot explicitly stated for predicate in this document
    Function After Contamination with Stomach Contents requirementsPassNot explicitly stated for predicate in this document
    Function After Immersion in Water requirementsPassNot explicitly stated for predicate in this document
    Measurement of Dead Space requirementsPassNot explicitly stated for predicate in this document
    Resistance to Disengagement of Parts requirementsPassNot explicitly stated for predicate in this document
    High and Low Storage Conditions requirementsPassNot explicitly stated for predicate in this document
    High Temperature Operation requirementsPassNot explicitly stated for predicate in this document
    Low Temperature Operation requirementsPassNot explicitly stated for predicate in this document
    Attempted Displacement of Valve by Finger, Function After Dropping, Function After Potentially Deforming Weight requirementsPassNot explicitly stated for predicate in this document
    Conformance of Connectors requirementsPassNot explicitly stated for predicate in this document
    ISTA Procedure 2A Shipping Testing requirementsPassNot explicitly stated for predicate in this document
    Expiratory Resistance at 50 LPM0.327 cm H2O (0.0321 kPa)0.25 cm H2O at 50 LPM (per K142402 510(k) Summary)
    Inspiratory Resistance at 50 LPM1.15 cm H2O (0.113 kPa)1.85 cm H2O at 50 LPM (per K142402 510(k) Summary)
    Inlet ConnectorStandard 22 mm IDStandard 22 mm ID
    Outlet ConnectorStandard 22 mm OD/15 mm IDStandard 22 mm OD/15 mm ID

    2. Sample Size Used for the Test Set and Data Provenance

    The document does not explicitly state the sample sizes for the non-clinical tests (biocompatibility and bench testing). It mentions that the testing was conducted according to recognized standards (FDA Blue Book Memo G95-1, ISO 10993 series, AS 4259-1995, ISTA Procedure 2A). Typically, these standards specify the number of samples required for each test. The data provenance is implied to be from laboratory testing as mandated by these standards, not from human subjects or real-world use. It is retrospective in the sense that the testing was performed to support the 510(k) submission.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    This information is not applicable as the evaluation relies on a non-clinical, benchtop and laboratory testing against established engineering and biocompatibility standards, not on expert interpretation of medical data.

    4. Adjudication Method for the Test Set

    This information is not applicable as the evaluation relies on a non-clinical, benchtop and laboratory testing against established engineering and biocompatibility standards. There is no human adjudication process described.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, If So, What Was the Effect Size of How Much Human Readers Improve with AI vs. Without AI Assistance

    This information is not applicable. The Seal Rite Non-Rebreathing Valve is a mechanical medical device, not an AI or imaging diagnostic tool. Therefore, an MRMC study with human readers (or AI assistance) would not be relevant.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

    This information is not applicable as the device is a mechanical non-rebreathing valve, not an algorithm or AI system. The performance evaluated is the standalone performance of the physical device according to engineering and biocompatibility standards.

    7. The Type of Ground Truth Used

    The "ground truth" for this device's performance is based on established engineering standards and biocompatibility standards. Specifically:

    • AS 4259-1995: "Ancillary devices for expired air resuscitation"
    • FDA Blue Book Memo G95-1: Guidance for biocompatibility testing.
    • ISO 10993-5: Biocompatibility testing (Cytotoxicity)
    • ISO 10993-10: Biocompatibility testing (Irritation)
    • ISO 10993-12: Biocompatibility testing (Sensitization)
    • ISTA Procedure 2A: Shipping testing.

    These standards define the acceptable performance parameters and test methods for such devices.

    8. The Sample Size for the Training Set

    This information is not applicable. The device is a mechanical non-rebreathing valve, not a machine learning model. Therefore, there is no "training set."

    9. How the Ground Truth for the Training Set Was Established

    This information is not applicable as there is no training set for this type of device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K142764
    Date Cleared
    2015-07-09

    (287 days)

    Product Code
    Regulation Number
    868.5870
    Reference & Predicate Devices
    Why did this record match?
    510k Summary Text (Full-text Search) :

    : KYOLING CPR Mask with Oxygen Port, KYOLING CPR Mask without Oxygen Port Regulation Number: 21 CFR 868.5870
    | Valve, Non-Rebreathing |
    | Regulation classification: | 21 CFR 868.5870

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The CPR Mask with Oxygen Port is designed to assist in providing immediate life support (mouth to health emergency victims requiring oxygen support (inhalation) or cardiopulmonary resuscitation (CPR) rescue techniques. The CPR Mask with Oxygen Port is for prescription use.

    The CPR Mask without Oxygen Port is designed to assist in providing immediate life support (mouth to health emergency victims requiring cardiopulmonary resuscitation (CPR) resuscitator without oxygen port is for overthe-counter use.

    Device Description

    The KYOLING CPR mask is used during cardiopulmonary resuscitation procedures, which are used in emergency situations to supply oxygen and produce blood flow in the heart and lungs.

    The CPR mask is made of PVC and one-way valve made of medical grade K-resin, the mask is used for mouth-to-mask breathing, it provides a physical barrier between the rescuer and victim, eliminating direct contact of the rescuer's lips with the unknown subject; and also it promotes an airtight seal to the face allowing ventilation through both the mouth and nose simultaneously. The KYOLING CPR Mask includes transparent dome, universal breathing tube, one-way filtered valve, head strap, and with Oxygen Port or without Oxygen Port.

    AI/ML Overview

    The provided document describes the acceptance criteria and a study to prove that the KYOLING CPR mask with and without an oxygen port meets these criteria.

    Here's a breakdown of the requested information:

    1. A table of acceptance criteria and the reported device performance

    ParameterAcceptance CriteriaReported Device Performance (KYOLING CPR Mask with Oxygen Port)Reported Device Performance (KYOLING CPR Mask without Oxygen Port)
    BiocompatibilityMeeting ISO 10993 standards for cytotoxicity, sensitization, and irritationAll tests conducted in accordance with FDA Blue Book Memorandum #G95-1 and ISO 10993-1, with tests for cytotoxicity, sensitization, and irritation completed. Cushion and valve considered surface contacting for
    Ask a Question

    Ask a specific question about this device

    K Number
    K142402
    Date Cleared
    2015-03-23

    (208 days)

    Product Code
    Regulation Number
    868.5870
    Reference & Predicate Devices
    Why did this record match?
    510k Summary Text (Full-text Search) :

    CA 92841

    Re: K142402

    Trade/Device Name: BigEasy™ Non-Rebreathing Valve Regulation Number: 21 CFR 868.5870
    |
    | Regulatory Classification | 21 CFR 868.5870

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The BigEasy™ Non-Rebreathing Valve is intended to assist in providing immediate life support (mouth to mask ventilation) to health emergency victims requiring oxygen support or cardiopulmonary resuscitation (CPR) rescue techniques.

    Device Description

    The BigEasy™ Non-Rebreathing Valve is a resuscitation valve designed for resuscitation using expired air for ventilation. It has fittings on the patient end that will adapt to standard masks (22mm ID) or endotracheal tubes (15mm OD) used for resuscitation. It provides a path for a rescuer to blow expired air into a mask or endotracheal tube through a silicone valve in the device and has a rigid one-way valve that directs exhaled air from the patient away from the rescuer. It also incorporates an oxygen inlet port on the rescuer's side of the valve for the adding of supplemental oxygen without requiring a separate adapter.

    AI/ML Overview

    The provided text describes the acceptance criteria and the study that proves the device meets those criteria for the BigEasy™ Non-Rebreathing Valve.

    Here's the breakdown of the information requested:

    1. A table of acceptance criteria and the reported device performance

    Acceptance Criteria CategoryAcceptance CriteriaReported Device Performance
    BiocompatibilityMeeting ISO 10993-1 requirements for Cytotoxicity, Irritation, and Sensitization.The BigEasy™ Non-Rebreathing Valve passed all required biocompatibility tests (Cytotoxicity, Irritation, Sensitization) based on ISO10993-1:2009/2010 and the 2013 FDA Guidance Document.
    Packaging and Assembly / ShippingMeeting packaging and assembly / shipping requirements.The BigEasy™ Non-Rebreathing Valve met the packaging and assembly / shipping testing acceptance criteria.
    EnvironmentalMeeting environmental requirements for storage and operation. (Specific ranges provided: Operating Temperature: -18°C to 50°C, Storage Temperature: -40°C to 60°C, Storage Relative Humidity: 10% to 95%)The BigEasy™ Non-Rebreathing Valve met the environmental requirements acceptance criteria. (Specific values: Operating Temperature: -18°C to 50°C, Storage Temperature: -40°C to 60°C, Storage Relative Humidity: 10% to 95% were also listed as specifications).
    Volatile Organic Compounds (VOCs), Ozone, CO, CO2 and Fine Particle DischargeMeeting TO-15 standards for VOCs and requirements for the discharge of other gases and particles below the standards' thresholds.The BigEasy™ Non-Rebreathing Valve met the requirements for VOCs, Ozone, CO, CO2, and Fine Particles.
    Materials SpecificationsMeeting materials specifications requirements.The BigEasy™ Non-Rebreathing Valve met the materials specifications acceptance criteria.
    Labeling VerificationMeeting labeling verification requirements.The BigEasy™ Non-Rebreathing Valve met the labeling verification acceptance criteria.
    Inlet and Outlet FittingsMeeting inlet and outlet fitting requirements (Standard 22mm ID for inlet, Standard 22mm OD/15mm ID for outlet).The BigEasy™ Non-Rebreathing Valve met the fittings acceptance criteria.
    Resistance to Inhalation and ExhalationMeeting resistance requirements after and during environmental exposure, vomitus contamination, water submersion, and mechanical displacement.The BigEasy™ Non-Rebreathing Valve met the resistance criteria. (Specific values: Inspiratory Resistance: 1.9 cm H2O at 50 LPM, Expiratory Resistance: 0.25 cm H2O at 50 LPM were also listed as specifications).
    Drop TestWithstanding a drop on a concrete floor from 1 meter.The BigEasy™ Non-Rebreathing Valve met the drop test requirements.
    Mean Concentration of Oxygen at 15 LPM and Circuit Backpressure at 30 LPMDetermining the mean oxygen concentration and backpressure during oxygen delivery.The BigEasy™ Non-Rebreathing Valve met the requirements for oxygen delivery.
    Assembly and ApplicationValidation that the device can be assembled and applied according to the Instructions For Use (IFU).Participants were able to perform the intended actions while following the IFU.

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    The document does not explicitly state the sample sizes for each specific test or the data provenance (country of origin, retrospective/prospective). The studies are described as "Non-clinical test results" and "Bench Tests," which typically implies laboratory testing rather than human subject data.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    This information is not provided. Given that the tests are primarily "Bench Tests" and "Biocompatibility Testing," they typically involve objective measurements against established standards, rather than expert interpretation for ground truth.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    This information is not provided. The tests appear to be objective measurements against predefined standards, thus not requiring an adjudication method by human experts.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    No MRMC comparative effectiveness study was done. This device is a non-rebreathing valve, which is a physical medical device, not an AI-powered diagnostic tool. The "performance testing" focuses on physical and material properties, not diagnostic accuracy with or without AI.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    A standalone performance evaluation was done, but it was for the physical device itself, not an algorithm. The device was tested independently against various physical, material, and operational standards.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    The "ground truth" for the performance tests was based on established industry standards and regulations, such as:

    • AS 4259-1995 (Ancillary devices for expired air resuscitation)
    • ISO 13544-2:2002 (Respiratory Therapy Equipment – Part 2: Tubing and Connectors)
    • ISO 5356-1:2004 (Anaesthetic and Respiratory Equipment – Conical Connectors - Part 1: Cones and Sockets)
    • BS EN ISO 10651-4:2009 (Lung Ventilators – Part 4: Particular requirements for operator-powered non-rebreathing valves)
    • ISTA-2A:2011 (Packaged-Products weighing 150 lbs (68 kg) or Less)
    • ISO 10993-1:2009 COR 1 2010 (Biological Evaluation of Medical Devices – Part 1: Evaluation and Testing within a Risk Management Process)
    • FDA Guidance Document (Draft) (Reviewer Guidance on Face Masks and Shield for CPR)
    • BS EN ISO 15223-1:2012 (Medical Devices – Symbols to be used with Medical Devices Labels, Labelling and Information to be supplied– Part 1: General Requirements)
    • EPA-453/R-98-008B (Method TO-15, Determination Of Volatile Organic Compounds (VOCs) In Air)

    8. The sample size for the training set

    Not applicable. This is a physical medical device, not an AI algorithm requiring a 'training set.'

    9. How the ground truth for the training set was established

    Not applicable. There is no training set for this type of device.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 4