(188 days)
Not Found
No
The description focuses on the mechanical and optical components of a surgical CO2 laser system and its scanning accessories. There is no mention of AI, ML, or any computational processing that would suggest the use of such technologies for tasks like image analysis, pattern recognition, or automated decision-making. The scanning systems are described as having an "electronic controller" and "touch screen settings," which are standard features for controlling mechanical movement and parameters, not indicative of AI/ML.
Yes
This device is clearly identified as a "medical device indicated for the incision, excision, ablation, vaporization and coagulation of body soft tissues" for surgical purposes, which falls under the definition of a therapeutic device.
No
The device description and intended use clearly state that it is a surgical CO2 laser used for incision, excision, ablation, vaporization, and coagulation of body soft tissues. These are therapeutic, not diagnostic, actions.
No
The device description clearly details a physical CO2 laser system with hardware components such as a laser source, articulated arm, handpiece, footswitch, and control panel. While it mentions integrated electronic controllers and scanning systems, these are part of a larger hardware device, not a standalone software application.
Based on the provided information, this device is not an IVD (In Vitro Diagnostic).
Here's why:
- Intended Use: The intended use clearly states that the device is for the "incision, excision, ablation, vaporization and coagulation of body soft tissues". This describes a surgical procedure performed directly on the patient's body.
- Device Description: The description details a surgical CO2 laser system used for cutting and manipulating tissue. It focuses on the physical delivery of laser energy to the treatment area.
- Lack of IVD Characteristics: An IVD device is used to examine specimens derived from the human body (like blood, urine, tissue samples) to provide information for diagnosis, monitoring, or screening. This device does not perform any such analysis of biological samples.
The device is a surgical laser, which is a different category of medical device than an IVD.
N/A
Intended Use / Indications for Use
The SmartXide 50 HS/MS system with its delivery accessories is a medical device indicated for the incision, excision, ablation, vaporization and coagulation of body soft tissues including intraoral tissues, in medical specialties including aesthetic (dermatology and plastic surgery), otolaryngology (ENT), gynaecology, neurosurgery and genitourinary surgery.
Product codes
GEX
Device Description
SmartXide 50 HS/MS is a surgical CO2 laser with 50W of maximum power in continuous mode. The laser system is produced and sold with an internal predisposition for scanning system (HiScan Surgical for SmartXide HS, MiniScan Plus for SmartXide MS).
Laser activation is by footswitch. Overall weight of the laser is 43 kg, and the size is 160 cm x 48 cm x 55 cm (H x W x D).
Electrical requirement is 115VAC, 6A, 50-60 Hz, single phase.
The wavelength of this laser is 10.600nm. This wavelength is mostly absorbed by water and that makes this laser particularly suitable for soft tissue surgery.
The CO2 laser radiation is delivered to the treatment area through the handpiece. The handpiece is attached to the distal end of the articulated arm, which is a permanently mounted laser delivery system of the system.
The articulated arm is an optical assembly that delivers free beam laser radiation. It is made up of seven mirrors placed on rotating knuckles: the mechanical accuracy of the articulated arm allows the CO2 laser beam to travel inside it and along its axis however the arm is oriented.
The field of action of the articulated arm covers a radius of approximately 80 cm, the transfer efficiency of power is greater than 85%. The loss of 15% is balanced by a suitable calibration of the internal power meter.
An air flow is provided by an internal pump in order to avoid dust and particles deposition on the optics during laser operations.
Emission parameters are selected on the front panel while emission activation is by the footswitch. The on-off switch and emergency switch are also located on the front panel of the system.
A warning light is located on the top cover, close to the control panel. Light ON state indicates that the system is enabled and ready.
The CO2 laser microsurgery is done with microspot micromanipulators and scanning systems. The CO2 laser focalized on very little spots (140-250um) and moved by scanning systems is useful to fasten the surgical procedures and limit the thermal damage to the tissues surrounding the ablation. In the SmartXide 50 HS/MS series, the electronic controller of the scanner has been integrated in the laser system with touch screen settings.
The scanning systems move the beam on the tissue with controlled velocity and defined patterns to optimize the laser ablation. The high Power Density reached though little spots and typical of the cutting, are thus well controlled by the operator, and the tissue destruction happens in a very quick, delicate and precise way, reducing drastically the surgery time and limiting the lateral thermal damage and the negative phenomenon of carbonisation, with evident advantage for the patient and for the surgeon as well.
Both scanning units can be used together with Deka micromanipulator EasySpot.
Easyspot has a single ring nut rapid focalization system that allows to focalize the beam to the same focal length of the microscope and fix the position with a mechanical block. In this way the micromanipulator "remembers" the focus position, still allowing eventually the surgeon to defocus the beam from the same ring nut.
Thanks to its joystick, it is possible to regulate the mechanical tension and the maximum work field in order to easily control and never "loose" the beam even inside small size faryngoscopes.
On top of the joystick Easyspot can mount a remote control specially conceived to command top level scanning systems (HiScan Surgical).
It allows the surgeon to have under direct control the more useful electronic scanning functions (rotation and dimension of the figures, scan off-scan on, centering) without moving his eyes from the microscope.
Mentions image processing
Not Found
Mentions AI, DNN, or ML
Not Found
Input Imaging Modality
Not Found
Anatomical Site
body soft tissues including intraoral tissues
Indicated Patient Age Range
Not Found
Intended User / Care Setting
Not Found
Description of the training set, sample size, data source, and annotation protocol
Not Found
Description of the test set, sample size, data source, and annotation protocol
Not Found
Summary of Performance Studies
Nonclinical Performance Data: None
Clinical Performance Data: None
Key Metrics
Not Found
Predicate Device(s)
Reference Device(s)
Not Found
Predetermined Change Control Plan (PCCP) - All Relevant Information
Not Found
§ 878.4810 Laser surgical instrument for use in general and plastic surgery and in dermatology.
(a)
Identification. (1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
(b)
Classification. (1) Class II.(2) Class I for special laser gas mixtures used as a lasing medium for this class of lasers. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807 of this chapter, subject to the limitations in § 878.9.
0
K101904 pg 1 of 3
510(K) Summary
JAN 1 2 2011
Submitter:
El.En. S.p.A. via Baldanzese. 17 50041 Calenzano (FI), Italy
Contact:
Paolo Peruzzi Senior Regulatory Affairs Engineer Phone: +39.055.8826807 E-mail: standards@elen.it
Date Summary Prepared:
December, 14, 2010
CO2 Laser System
Smartxide 50 HS, Smartxide 50 MS
79-GEX, 21 CFR 878.4810
Instrument, surgical, powered, laser
Lumenis Acupulse30 ST/40 ST (K082809)
Common Name:
Device Trade Name:
Classification Name:
Equivalent Devices:
Device Description:
DEKA Smart CO2 with DOT scanner (K072159) Lumenis UltraPulse Surgitouch (K030147) SmartXide 50 HS/MS is a surgical CO2 laser with 50W of maximum
power in continuous mode. The laser system is produced and sold with an internal predisposition for scanning system (HiScan Surgical for SmartXide HS, MiniScan Plus for SmartXide MS).
Laser activation is by footswitch. Overall weight of the laser is 43 kg, and the size is 160 cm x 48 cm x 55 cm (H x W x D).
Electrical requirement is 115VAC, 6A, 50-60 Hz, single phase.
The wavelength of this laser is 10.600nm. This wavelength is mostly absorbed by water and that makes this laser particularly suitable for soft tissue surgery.
The CO2 laser radiation is delivered to the treatment area through the handpiece. The handpiece is attached to the distal end of the articulated arm, which is a permanently mounted laser delivery system of the system.
The articulated arm is an optical assembly that delivers free beam laser radiation. It is made up of seven mirrors placed on rotating knuckles: the mechanical accuracy of the articulated arm allows the CO2 laser beam to travel inside it and along its axis however the arm is oriented.
The field of action of the articulated arm covers a radius of approximately 80 cm, the transfer efficiency of power is greater than 85%. The loss of 15% is balanced by a suitable calibration of the internal power meter.
An air flow is provided by an internal pump in order to avoid dust and particles deposition on the optics during laser operations.
1
Emission parameters are selected on the front panel while emission activation is by the footswitch. The on-off switch and emergency switch are also located on the front panel of the system.
A warning light is located on the top cover, close to the control panel. Light ON state indicates that the system is enabled and ready.
is The CO2 laser microsurgery done with microspot micromanipulators and scanning systems. The CO2 laser focalized on very little spots (140-250um) and moved by scanning systems is useful to fasten the surgical procedures and limit the thermal damage to the tissues surrounding the ablation. In the SmartXide 50 HS/MS series, the electronic controller of the scanner has been integrated in the laser system with touch screen settings.
The scanning systems move the beam on the tissue with controlled velocity and defined patterns to optimize the laser ablation. The high Power Density reached though little spots and typical of the cutting, are thus well controlled by the operator, and the tissue destruction happens in a very quick, delicate and precise way, reducing drastically the surgery time and limiting the lateral thermal damage and the negative phenomenon of carbonisation, with evident advantage for the patient and for the surgeon as well.
Both scanning units can be used together with Deka micromanipulator EasySpot.
Easyspot has a single ring nut rapid focalization system that allows to focalize the beam to the same focal length of the microscope and fix the position with a mechanical block. In this way the micromanipulator "remembers" the focus position, still allowing eventually the surgeon to defocus the beam from the same ring nut.
Thanks to its joystick, it is possible to regulate the mechanical tension and the maximum work field in order to easily control and never "loose" the beam even inside small size faryngoscopes.
On top of the joystick Easyspot can mount a remote control specially conceived to command top level scanning systems (HiScan Surgical).
It allows the surgeon to have under direct control the more useful electronic scanning functions (rotation and dimension of the figures, scan off-scan on, centering) without moving his eyes from the microscope.
The Smartxide 50 HS/MS system with its accessories is a medical Intended Use: device indicated for the incision, excision, ablation, vaporization and coagulation of body soft tissues including intraoral tissues, in medical specialties including aesthetic (dermatology and plastic surgery), otolaryngology (ENT), gynaecology, neurosurgery and genitourinary surgery.
The Smartxide 50 HS/MS is substantially equivalent to the Lumenis Comparison: Acupulse30 ST/40 ST (K082809), to the DEKA Smart CO2 (SmartXide, SmartUS20D) with DOT scanner (K072159) and to the Lumenis UltraPulse Surgitouch (K030147).
2
K101904 pg 3 of 3
It has the same indication for use, the same principle of operation, the same wavelength and essentially the same power and pulse energy range as the predicate devices.
Nonclinical Performance Data: | None |
---|---|
Clinical Performance Data: | None |
Conclusion: | The Smartxide 50 HS, Smartxide 50 MS lasers with delivery |
accessories are safe and effective devices for incision, excision, | |
ablation, vaporization and coagulation of body soft tissue, including | |
intraoral tissue. | |
Additional Information: | None |
3
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle with three stripes extending from its body, representing the department's commitment to health, well-being, and human services. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" is arranged in a circular fashion around the eagle symbol.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
El.En. S.p.A % Mr. Paolo Peruzzi Senior Regulatory Affairs Engineer via Baldanzese, 17 50041 Calenzano (FI), Italy
JAN 1 2 200
Re: K101904
Trade/Device Name: SmartXide 50 HS/MS Laser with delivery accessories Regulation Number: 21 CFR 878.4810 Regulation Name: Laser surgical instrument for use in general and plastic surgery and in dermatology Regulatory Class: Class II Product Code: GEX Dated: December 15, 2010 Received: December 17, 2010
Dear Mr. Peruzzi:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
4
Page 2 - Mr. Paolo Peruzzi
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportalProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
Mark N. Melkerson
Director Division of Surgical, Orthopedic And Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
5
JAN 1 2 2011
8 ปี 2000 (1988) 1999)
1979 (1999)
1979 (1999)
1979 (1999)
1979 (1999) 1999 (1999) 1999 (1999)
1999 (1999)
1999 (1999)
1999 (1999)
1999 (1999) 1999 (1999)
1999 ( 510(K) Number (if known):_K (0) 904
Device Name: __ SmartXide 50 HS/MS Laser with delivery accessories
Indications For Use:
The SmartXide 50 HS/MS system with its delivery accessories is a medical device indicated for the incision, excision, ablation, vaporization and coagulation of body soft tissues including intraoral tissues, in medical specialties including aesthetic (dermatology and plastic surgery), otolaryngology (ENT), gynaecology, neurosurgery and genitourinary surgery.
Prescriptive Use (Part 21 CFR 801 Subpart D)
OR
Over-the-Counter Use (Part 21 CFR 801 Subpart C)
(PLEASE NO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
(Optional Format 1-2-96)
Nathr ashw for mkm
(Division Sign-Off) Division of Surgical. Orthopedic, and Restorative Devices
510(k) Number K101904