K Number
K100807
Date Cleared
2010-06-07

(77 days)

Product Code
Regulation Number
878.3300
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The MiniArc Precise™ Single-Incision Sling System is intended for the placement of a sub-urethral sling for the treatment of female stress urinary incontinence (SUI) resulting from urethral hypermobility and/or intrinsic sphincter deficiency (ISD).

Device Description

The MiniArc Precise Single-Incision Sling System (MiniArc Precise) is a modification of the currently commercialized device, MiniArc Single-Incision Sling System (MiniArc), which consist of one sterile mesh sling and one sterile surgical instrument · used for sub-urethral sling placement. The minor modifications to MiniArc for the development of the MiniArc Precise are as follows: Self-fixating tips on the mesh sling are for a "secure connection" (snap-fit) to 0 the delivery tool needle tip. Delivery tool has an actuating needle tip release mechanism for the mesh sling. ● Mesh slings center-length dimensions are modified and the mesh sling arms 0 are knit with reinforcement fibers. The MiniArc Precise is available in two configurations; each configuration contains one sterile mesh sling and one sterile delivery tool. The identical delivery tool is used with both configurations. The configurations are identified by the mesh sling option, either the short center-length mesh or the long center-length mesh. The mesh sling and mesh arms material for MiniAre Precise is the same knitted polypropylene monofilament material used in the predicate, MiniArc. MiniArc Precise uses the same surgical approach and the same mesh placement procedures as the predicate device. MiniArc. The MiniArc Precise device is for the treatment of female stress urinary incontinence (SUI) resulting from urethral hypermobility and/or intrinsic sphincter deficiency (ISD). MiniArc Precise device is for single use only and is not to be re-sterilized.

AI/ML Overview

The provided text describes a 510(k) premarket notification for a medical device called the MiniArc Precise™ Single-Incision Sling System. This submission focuses on demonstrating substantial equivalence to a predicate device (MiniArc® Sling System) rather than presenting a de novo study with acceptance criteria in the traditional sense of a performance study.

Therefore, the information regarding acceptance criteria, reported device performance, sample sizes for test/training sets, expert qualifications, adjudication methods, MRMC studies, or standalone algorithm performance, as typically found in AI/diagnostic device submissions, is not applicable here.

This submission relies on non-clinical testing to show that the minor modifications to the MiniArc Precise device do not change its fundamental scientific technology, intended use, or product performance specifications compared to the predicate device.

Here's an breakdown of the information that is available based on your request, with an emphasis on what is not present:


1. Table of Acceptance Criteria and Reported Device Performance

Acceptance CriterionReported Device PerformanceStudy Type
Equivalence to Predicate Device for Intended UseConfirmed product performance suitable for intended use, which is the same as the predicate device.Bench testing
Tissue Fixation and Tissue ResponseConfirmed equivalence to the predicate device.Animal testing
Surgical Procedure (using "secure connection" self-fixating tips)Confirmed equivalence to the predicate device.Cadaver lab testing
BiocompatibilityEnsured biocompatibility based on similarity of materials to the predicate device.Biocompatibility assessment
Design VerificationAdherence to Quality System Requirements (FDA 21 CFR Part 820.20, EN ISO 13485:2003); Design Controls (FDA 21 CFR Part 820.30).Non-clinical testing
Device PerformanceAdherence to Quality System Requirements (FDA 21 CFR Part 820.20, EN ISO 13485:2003); Design Controls (FDA 21 CFR Part 820.30).Non-clinical testing
PackagingAdherence to ISO 11607:2006 Part 1 & 2.Non-clinical testing
Shelf LifeAdherence to ISO 11607:2006 Part 1, 6.4 Stability Testing.Non-clinical testing
SterilizationAdherence to ISO 11135 and EN ISO 13485.Non-clinical testing

Explanation: The "acceptance criteria" here are effectively the demonstrations of equivalence or adherence to standards, rather than specific performance metrics like sensitivity/specificity for a diagnostic device. The "reported device performance" is a confirmation that these criteria were met.


2. Sample size used for the test set and the data provenance

  • Sample Size for Test Set: Not explicitly stated as a numerical sample size for human subjects or data sets in the context of device performance testing. The "test sets" refer to the subjects/materials used in bench testing, animal testing, and cadaver lab testing.
  • Data Provenance:
    • Bench testing: In-house laboratory testing (implied).
    • Animal testing: Performed as per FDA 21 CRR: Part 58 Good Laboratory Practice Regulations (location not specified, but typically conducted in regulated lab environments).
    • Cadaver lab testing: Performed using human cadavers (location not specified).

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

  • Not Applicable. This submission does not involve a diagnostic AI algorithm or a study requiring expert readers to establish ground truth for a test set. The validation is based on engineering principles, material science, and biological response assessments (animal/cadaver models) to demonstrate physical and functional equivalence to a predicate device.

4. Adjudication method for the test set

  • Not Applicable. As there are no expert readers establishing ground truth for a test set in statistical terms (e.g., image interpretation), no adjudication method is relevant.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

  • Not Applicable. This submission is for a surgical implant (suburethral sling system), not a diagnostic device assisted by AI. Therefore, an MRMC study or AI assistance effect size is irrelevant.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

  • Not Applicable. This is a hardware medical device (surgical sling) and does not involve any algorithms or AI for standalone performance.

7. The type of ground truth used

  • For bench testing: Engineering specifications and performance standards.
  • For animal testing: Histopathological and physiological assessment of tissue response and fixation.
  • For cadaver lab testing: Observational assessment of surgical procedure and device placement/fixation, likely by surgeons/medical professionals.

8. The sample size for the training set

  • Not Applicable. This submission does not involve machine learning or AI, and therefore, there is no "training set."

9. How the ground truth for the training set was established

  • Not Applicable. As there is no training set, this question is not relevant.

{0}------------------------------------------------

p. 1 of 4

510(k) Summary As required by 21 CFR 807.92 (b)

510(k) Number: K100807

Date Revised:June 2, 2010JUN - 7 2010
Submitter Information:
Submitter's Name/AddressAmerican Medical Systems, Inc.10700 Bren Road WestMinnetonka, MN 55434-USA
Owner/Operator:American Medical Systems, Inc.10700 Bren Road WestMinnetonka, MN 55434-USA
Manufacturing Sites:American Medical Systems, Inc.10700 Bren Road WestMinnetonka, MN 55434-USAFDA Establishment Registration Number: 2183959
Sterigenics US, Inc.7775 South Quincy StreetWillowbrook, IL 60527FDA Establishment Registration Number: 1450293

Submission Contact Information:

Primary Contact:Donna M. SemlakSenior Regulatory Affairs SpecialistTelephone: 952-939-7082Fax: 952-939-5785Donna.Semlak@AmericanMedicalSystems.com
Alternative Contact:Melanie HessSenior Manager Regulatory AffairsTelephone: 952-930-6490Fax: 952-939-5785Melanie.Hess@AmericanMedicalSystems.com

{1}------------------------------------------------

Device(s) Information:

Trade Name:MiniArc Precise™ Single-Incision Sling
Common Name:Surgical Mesh
Classification Name:Surgical Mesh, Polymeric
Class:Class II / 21 CFR § 878.3300
Product Code:PAH

Predicate Device:

Device NameSubmission NumberClearance Date
MiniArc® Sling SystemK070065March 1, 2007
MiniArc® Sling SystemK071902August 24, 2007
MiniArc® Sling SystemK073703January 20, 2008

Device Description:

The MiniArc Precise Single-Incision Sling System (MiniArc Precise) is a modification of the currently commercialized device, MiniArc Single-Incision Sling System (MiniArc), which consist of one sterile mesh sling and one sterile surgical instrument · used for sub-urethral sling placement.

The minor modifications to MiniArc for the development of the MiniArc Precise are as follows:

  • Self-fixating tips on the mesh sling are for a "secure connection" (snap-fit) to 0 the delivery tool needle tip.
  • Delivery tool has an actuating needle tip release mechanism for the mesh sling. ●
  • Mesh slings center-length dimensions are modified and the mesh sling arms 0 are knit with reinforcement fibers.

The MiniArc Precise is available in two configurations; each configuration contains one sterile mesh sling and one sterile delivery tool. The identical delivery tool is used with both configurations. The configurations are identified by the mesh sling option, either the short center-length mesh or the long center-length mesh. The mesh sling and mesh arms material for MiniAre Precise is the same knitted polypropylene monofilament material used in the predicate, MiniArc.

MiniArc Precise uses the same surgical approach and the same mesh placement procedures as the predicate device. MiniArc. The MiniArc Precise device is for the treatment of female stress urinary incontinence (SUI) resulting from urethral hypermobility and/or intrinsic sphincter deficiency (ISD). MiniArc Precise device is for single use only and is not to be re-sterilized.

{2}------------------------------------------------

Indications for Use:

Existing Indications for Use:

The MiniArc Sling System (MiniArc) is intended for the placement of a suburethral sling for the treatment of female stress urinary incontinence (SUI) resulting from urethral hypermobility and/or intrinsic sphincter deficiency (ISD).

Proposed Indications for Use:

There are no changes to the existing indications for use with the exception of the device name, as shown below:

The MiniArc Precise™ Single-Incision Sling System is intended for the placement of a suburethral sling for the treatment of female stress urinary incontinence (SUI) resulting from urethral hypermobility and/or intrinsic sphincter deficiency(ISD).

Comparison to Predicate Device:

The proposed MiniArc Precise and the predicate device, MiniArc, consist of one sterile mesh sling and one sterile surgical instrument (needle passer or deliver tool).

The MiniArc Precise and MiniArc devices are intended for the placement of a suburethral mesh sling under the female urethra, with self-fixating tips in the right and left obturator. The mesh sling is placed with the use of a surgical instrument. The surgical instrument is a stainless steel curved needle with an attached plastic handle. The tip portion of surgical instrument is configured to allow for secure connection and placement of the mesh sling. The mesh sling is a piece of knitted polypropylene mesh with self-fixating tips at each end. The mesh sling is intended to remain in the body as a permanent implant and is not absorbed or degraded by the action of tissue in-growth or tissue enzymes.

These device systems are for the treatment of female stress urinary incontinence (SUI) resulting from urethral hypermobility and/or intrinsic sphincter deficiency (1SD). These devices are for single use and are not to be re-sterilized.

  • The MiniArc Precise has the same intended use, same implant materials, and same . • sterilization methods. The MiniArc Precise also uses similar manufacturing processes and similar delivery tool materials/characteristics to the predicate device, MiniArc.

Summary of Non-Clinical Testing / Statement of Substantial Equivalence

The non-clinical testing included the assessment of the MiniArc Precise physical properties; ability to achieve its intended use; and substantial equivalence to the MiniArc predicate device.

{3}------------------------------------------------

  • Bench testing confirmed the product performance of the MiniArc Precise to be . suitable for its intended use: the same intended use as the predicate device.
  • Animal testing confirmed the tissue fixation and tissue response is equivalent ● to the predicate device.
  • Cadaver lab testing confirmed the surgical procedure, using the "secure ● connection" self-fixating tips to the delivery tool needle, is the equivalent to the predicate device.
  • Biocompatibility assessment ensured the MiniArc Precise materials are ● biocompatible based on the similarity of the materials to the predicate device.

The following table is a summary of non-clinical testing conducted in accordance to industry standards/regulations:

TestIndustry Standards
Design VerificationQuality System Requirements FDA 21 CFR Part 820.20 EN ISO13485: 2003; Design Controls FDA 21 CFR Part 820.30
Device PerformanceQuality System Requirements FDA 21 CFR Part 820.20 EN ISO13485: 2003; Design Controls FDA 21 CFR Part 820.30
Animal ModelFDA 21 CRR: Part 58 Good Laboratory Practice Regulations
PackagingISO 11607: 2006 Part 1 and Part 2: Packaging for TerminalSterilized Medical Devices
Shelf LifeISO 11607: 2006 Part 1, 6.4 Stability Testing
BiocompatibilityISO 10993-1: 2009 Biological Evaluation of Medical Devices.
SterilizationISO 11135: Medical Devices - Validation and Routine Control ofEthylene Oxide Sterilization and EN ISO 13485.

The proposed MiniArc Precise device performance, intended use and fundamental scientific technology remain unchanged from the predicate device. The proposed modifications do not change the mesh fixation mechanism found in the predicate device, MiniArc. The tissue fixation element fixates the mesh until tissue in-growth occurs through the pores of the mesh. The mesh sling will still be used as a supporting framework in areas where the connective tissue has weakened or ruptured. The mesh sling, with self-fixating tips, is intended to remain in the body as a permanent implant.

American Medical Systems considers the MiniArc Precise product performance to be significantly equivalent to the predicate device, MiniArc, because there are no changes to the product performance specifications; device intended use; or device functional scientific technology.

Conclusion

The above non-clinical testing for the MiniArc Precise demonstrates product performance to be substantially equivalent and demonstrates that it is as safe and effective as the predicate device. MiniArc.

{4}------------------------------------------------

Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three stripes forming its body and wing. The eagle is encircled by the text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" in a circular arrangement.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002

American Medical Systems, Inc. % Ms. Donna Semlak Senior Regulatory Affairs Specialist 10700 Bren Road West MINNETONKA MN 55434

SEP 2 8 2012

Re: K100807

Trade/Device Name: MiniArc Precise™ Single-Incision Sling System Regulation Number: 21 CFR 878.3300 Regulation Name: Surgical mesh Regulatory Class: II Product Code: PAH Dated: April 23, 2010 Received: April 26, 2010

Dear Ms. Semlak:

This letter corrects our substantially equivalent letter of June 7, 2010.

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must

{5}------------------------------------------------

Page 2 -

comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely yours,

Benjamin R. Perkins

Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{6}------------------------------------------------

K100807
p. 1. f 1

Indications for Use Statement

INDICATIONS FOR USE STATEMENT

510(k) Number (if known):

Name of Devices:

Indications for Use:

MiniArc Precise™ Single-Incision Sling System

The MiniArc Precise™ Single-Incision Sling System is intended for the placement of a sub-urethral sling for the treatment of female stress urinary incontinence (SUI) resulting from urethral hypermobility and/or intrinsic sphincter deficiency (ISD).

Prescription Use X (Part 21 CFR 801 Subpart D)

AND/OR

Over-The-Counter Use (Part 21 CFR 801 Subpart C)

Daniel Kione for MKCM

(Division Slyn-Off) Division of Surgical, Orthopedic. and Restorative Devices

510(k) Number K100807

Trade Secret/Confidential

American Medical Systems

§ 878.3300 Surgical mesh.

(a)
Identification. Surgical mesh is a metallic or polymeric screen intended to be implanted to reinforce soft tissue or bone where weakness exists. Examples of surgical mesh are metallic and polymeric mesh for hernia repair, and acetabular and cement restrictor mesh used during orthopedic surgery.(b)
Classification. Class II.