K Number
K100775
Device Name
BD INTIMA II CLOSED IV CATHETER
Date Cleared
2010-06-03

(77 days)

Product Code
Regulation Number
880.5200
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP Authorized
Intended Use
The BD Intima II IV Catheter is inserted into a patient's vascular system for short-term use to monitor blood pressure or administer fluids intravascularly.
Device Description
The Intima II IV Catheter consists of a Vialon® material catheter, a notched needle to enhance flashback visualization, septum, extension tubing and either a straight or Y Luer connector. A needle-accessed PRN is attached to the Luer connector. An end cap is attached to the second Luer connector on the Y design. Blood is contained within the device. A slide clamp is provided on the extension tube. A notched needle resides within the catheter until it is withdrawn through the septum after the catheter has been threaded into the vein. The needle unit includes a paddle that is held during the insertion process. This device does not include a needle-shielding mechanism.
More Information

Not Found

No
The device description and performance studies focus on the physical components and nonclinical testing of an IV catheter, with no mention of AI or ML.

No
The device is used to monitor blood pressure or administer fluids intravascularly, which are diagnostic or supportive functions rather than therapeutic.

No

The device is used to monitor blood pressure or administer fluids, which are therapeutic or monitoring functions, not diagnostic. While monitoring blood pressure provides information, it's a general physiological measurement rather than a specific diagnostic test for a disease.

No

The device description clearly outlines physical components like a catheter, needle, tubing, and connectors, indicating it is a hardware medical device.

Based on the provided information, this device is not an IVD (In Vitro Diagnostic).

Here's why:

  • Intended Use: The intended use is to be "inserted into a patient's vascular system for short-term use to monitor blood pressure or administer fluids intravascularly." This describes a device used in vivo (within the living body) for direct patient care.
  • Device Description: The description details a catheter, needle, tubing, and connectors designed for insertion into a vein. This is consistent with a medical device used for accessing the vascular system.
  • Lack of IVD Characteristics: There is no mention of the device being used to examine specimens in vitro (outside the living body), such as blood, urine, or tissue, to provide information for diagnosis, monitoring, or screening.

IVD devices are specifically designed to perform tests on samples taken from the body. This device is designed to be inserted into the body.

N/A

Intended Use / Indications for Use

The BD Intima II™ IV Catheter is inserted into a patient's vascular system for short-term use to monitor blood pressure or administer fluids intravascularly.

Product codes (comma separated list FDA assigned to the subject device)

80FOZ

Device Description

The Intima II IV Catheter consists of a Vialon® material catheter, a notched needle to enhance flashback visualization, septum, extension tubing and either a straight or Y Luer connector. A needle-accessed PRN is attached to the Luer connector. An end cap is attached to the second Luer connector on the Y design. Blood is contained within the device. A slide clamp is provided on the extension tube. A notched needle resides within the catheter until it is withdrawn through the septum after the catheter has been threaded into the vein. The needle unit includes a paddle that is held during the insertion process. This device does not include a needle-shielding mechanism.

Mentions image processing

Not Found

Mentions AI, DNN, or ML

Not Found

Input Imaging Modality

Not Found

Anatomical Site

vascular system

Indicated Patient Age Range

Not Found

Intended User / Care Setting

Not Found

Description of the training set, sample size, data source, and annotation protocol

Not Found

Description of the test set, sample size, data source, and annotation protocol

Not Found

Summary of Performance Studies (study type, sample size, AUC, MRMC, standalone performance, key results)

Performance testing was conducted in accordance with consensus standards and design control requirements. Nonclinical test results and technological and dosign control require size catheters were shown to be equivalent between the subject device and the two predicate devices. The differences among the subject do not raise any new issues of safety or effectiveness. Refer to the two actious tables for the testing conducted to determine substantial equivalence.

Key Metrics (Sensitivity, Specificity, PPV, NPV, etc.)

Not Found

Predicate Device(s): If the device was cleared using the 510(k) pathway, identify the Predicate Device(s) K/DEN number used to claim substantial equivalence and list them here in a comma separated list exactly as they appear in the text. List the primary predicate first in the list.

BD Intima™ Catheter System K833657, BD Insvte-N ™ IV Catheter K843033

Reference Device(s): Identify the Reference Device(s) K/DEN number and list them here in a comma separated list exactly as they appear in the text.

Not Found

Predetermined Change Control Plan (PCCP) - All Relevant Information for the subject device only (e.g. presence / absence, what scope was granted / cleared under the PCCP, any restrictions, etc).

Not Found

§ 880.5200 Intravascular catheter.

(a)
Identification. An intravascular catheter is a device that consists of a slender tube and any necessary connecting fittings and that is inserted into the patient's vascular system for short term use (less than 30 days) to sample blood, monitor blood pressure, or administer fluids intravenously. The device may be constructed of metal, rubber, plastic, or a combination of these materials.(b)
Classification. Class II (performance standards).

0

K100775 p. lot 3

Becton Dickinson Infusion Therapy Systems Inc. Sandy UT 84070

| | BD Intima II™ IV Catheter
510(k) Summary | |
|-------------------------|----------------------------------------------------------------------------------------------|--------------|
| Submitter: | Becton Dickinson Infusion Therapy Systems Inc.
9450 South State Street
Sandy, UT 84070 | JUN - 3 2010 |
| Contact person: | Jane Stickel, Director, Regulatory Affairs | |
| Telephone number: | (801) 565-2686 | |
| FAX number: | (801) 565-2749 | |
| Date summary prepared: | April 9, 2010 | |
| Trade name: | BD Intima II™ IV Catheter | |
| Common name: | Peripheral Intravascular Catheter or IV Catheter | |
| Classification name: | Intravascular Catheter 21 CFR § 880.5200 | |
| Panel and product code: | 80FOZ | |
| Predicate devices: | BD Intima™ Catheter System K833657
BD Insvte-N ™ IV Catheter K843033 | |

Product description:

The Intima II IV Catheter consists of a Vialon® material catheter, a notched needle to enhance flashback visualization, septum, extension tubing and either a straight or Y Luer connector. A needle-accessed PRN is attached to the Luer connector. An end cap is attached to the second Luer connector on the Y design. Blood is contained within the device. A slide clamp is provided on the extension tube. A notched needle resides within the catheter until it is withdrawn through the septum after the catheter has been threaded into the vein. The needle unit includes a paddle that is held during the insertion process. This device does not include a needle-shielding mechanism.

Catalog NumberProduct Description
38340018G x 1.16" (1.3mm x 30mm) Straight luer connection site
38340120G x 1.16" (1.1mm x 30mm) Straight luer connection site
38340222G x 1.00" (0.9mm x 25mm) Straight luer connection site
38340324G x 0.75" (0.7mm x 19mm) Straight luer connection site
38340422G x 0.75" (0.9mm x 19mm) Straight luer connection site
38340518G x 1.16" (1.3mm x 30mm) Y luer connection site
38340620G x 1.16" (1.1mm x 30mm) Y luer connection site
38340722G x 1.00" (0.9mm x 25mm) Y luer connection site
38340824G x 0.75" (0.7mm x 19mm) Y luer connection site
38340922G x 0.75" (0.9mm x 19mm) Y luer connection site

Indications for Use:

510(k) Summary 4/9/10 rev 3

1

K100775 p. 2 of 3

Becton Dickinson Infusion Therapy Systems Inc. Sandy UT 84070

The BD Intima II™ IV Catheter is inserted into a patient's vascular system for short-term use to monitor blood pressure or administer fluids intravascularly.

Nonclinical test results and technological characteristic comparisons submitted to determine substantial equivalence:

The following technological characteristics of the BD Intima II IV Catheter are substantially equivalent per gauge size to the two predicate devices, the BD Insyte IV Catheter and BD Intima Catheter System:

  • Biocompatibility ISO 10993 Intended use 이
  • Distal end configuration l according to ISO 10555-5
  • . Minimum catheter/adapter bond strength
  • Vialon™ material catheter .
  • No needle-shielding mechanism
  • r Gauge sizes color coded
  • . Maximum cannula penetration force
  • l Maximum catheter tip penetration
  • l Maximum catheter drag
  • . EtO sterilization

Due to product design and/or customer requirements, the following technological characteristics of the BD Intima II IV Catheter differ from the two predicate devices:

  • 1 Effective catheter lengths and gauge sizes
  • 1 Integrated PVC extension tubing with PRN
  • 프 Nominal water flow rate
  • 트 Flexible wings on catheter adapter
  • 1 Notched needle
  • I Flexible wings on catheter adapter
  • 1 Minimum cannula/cannula hub bond strength
  • Catheter lie distance ■
  • Component materials 트

In determining substantial equivalence between the subject and predicate devices, testing was done per in-house protocol and according with the following standards, when applicable:

  • ISO 10555-1:1995/AMD2:2004 Sterile, single-use intravascular catheters -. Part 1 General Requirements: Testing conformed with all sections of this standard.
  • ISO 10555-5 Amd 1:1999, Corr 1:2002 Sterile, single-use intravascular 다 catheters - Part 5: Testing conformed to all sections of this standard with the exception of Sections 4.3 and 4.4.4. Section 4.3 relates to the identification

2

K100775 p. 3 of 3

Becton Dickinson Infusion Therapy Systems Inc. Sandy UT 84070

of lumens on a multilumen catheter and does not apply to the Intima II device since it is a single lumen catheter. Section 4.4.4 states that a vent fitting shall be provided. This requirement does not apply to the Intima II device since it is vented via a notch in the needle cannula, not a vent fitting.

  • ISO 594/2:1998 Conical fittings with 6% (Luer) taper for syringes, needles, and certain other medical equipment - Part 2 Lock fittings: Testing conformed with all sections of this standard.
    The initial biocompatibility testing of the intima II device conformed to ISO 10993 Part 1:1992 and revisions (1997), and FDA Blue Book Memorandum #G95-1 pertaining to the selection of tests and overall assessment. Various sub-parts of the ISO 10993 standard are updated on independent cycles. To address these changes to the ISO standards, all internal protocols and methods are reviewed and updated as necessary to ensure continued compliance with the standard. Additional testing is performed in the case of significant changes to the standard which may have the potential to impact the test results. In the case of the lintima Which may nave nave parts which were re-issued in 2009 (e.g. 10993 Part 1, Part 3, and Part 5) have been addressed and the product remains in compliance with these requirements.
TestTest MethodResults
CytotoxicityISO 10993-5Passed
SensitizationISO 10993-10Passed
Irritation or Intracutaneous Reactivity -ISO 10993-10Passed
Acute Systemic ToxicityISO 10993-11Passed
Subacute or Subchronic ToxicityISO 10993-11Passed
GenotoxicityISO 10993-3Passed
ImplantationISO 10993-6Passed
HemocompatibilityISO 10993-4Passed

The stability test report included in the 510(k) submission supports a shelf life of 3 years.

Clinical tests submitted:

No clinical test results were included in this submission.

Conclusions drawn from nonclinical tests to demonstrate that the device is safe and effective as the legally marketed devices:

Performance testing was conducted in accordance with consensus standards and design control requirements. Nonclinical test results and technological and dosign control require size catheters were shown to be equivalent between the subject device and the two predicate devices. The differences among the subject do not raise any new issues of safety or effectiveness. Refer to the two actious tables for the testing conducted to determine substantial equivalence.

3

Image /page/3/Picture/1 description: The image shows the seal of the Department of Health & Human Services - USA. The seal is circular and contains the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. In the center of the seal is a stylized symbol that appears to represent a person or family.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002

Becton Dickinson Infusion Therapy Systems Incorporated C/O Mr. Mark Job Responsible Third Party Official Regulatory Technology Services, LLC 1394 25th Street, NW Buffalo, Minnesota 55313

JUN - 3 2010

Re: K100775

Trade/Device Name: BD Intima IITM Closed IV Catheter Regulation Number: 21 CFR 880.5200 Regulation Name: Intravascular Catheter Regulatory Class: II Product Code: FOZ Dated: May 21, 2010 Received: May 24, 2010

Dear Mr. Job:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

: : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : :

4

Page 2-Mr. Mark Job

" Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to

http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers. International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely yours,

f

timper WDEr

Anthony D. Watson, B.S., M.S., M.B.A. Director

Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

5

Becton Dickinson Infusion Therapy Systems Inc.
Sandy, UT 84070

INDICATIONS FOR USE

Device Name:Intima II™ IV Catheter
Indications for Use:

The BD Intima II IV Catheter is inserted into a patient's vascular system for short-term use to monitor blood pressure or administer fluids intravascularly.

Prescription Use:XAND/OROver-the-Counter Use:
-------------------------------------------------------

(PLEASE DO NOT WRITE BELOW THIS LINE – CONTINUE ON ANOTHER PAGE IF NEEDED.)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Page 1 of _______

(Division Sign-Off)

Division of Anesthesiology, General Hospital,

Infection Control, Dental Devices

510(k) NumberK100775
------------------------

のお気になるということです。