K Number
K100285

Validate with FDA (Live)

Device Name
LITETOUCH MASK
Date Cleared
2010-06-29

(148 days)

Product Code
Regulation Number
868.5630
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The LiteTouch medium mask for ages 1-5 is to be used with OptiChamber Advantage Valved Holding Chambers. The mask is intended to be used to help administer aerosolized medication from most pressurized metered dose inhalers used with valved holding chambers as prescribed by a physician or health care professional. The intended environments for use include homes, hospitals, and clinics.

Device Description

The LiteTouch mask is a Class II device. It is a mask for use with Respironics' OptiChamber Valved Holding Chamber (K962822). The LiteTouch mask is a mask that utilizes the same operating principles as the AeroChamber Max Valved Holding Chamber Small and Medium Mask (K032972). The difference between the LiteTouch mask and the AeroChamber Max Valved Holding Chamber Small and Medium Mask is that, the LiteTouch mask does not come assembled to the chamber and the exhalation valve is located in different positions and is not of the same shape. The LiteTouch mask comprises of a Flap Valve and a Cushion plate.

AI/ML Overview

Here's an analysis of the provided text regarding acceptance criteria and the study that proves the device meets them:

1. Table of Acceptance Criteria and Reported Device Performance

The document does not explicitly state formal "acceptance criteria" with numerical thresholds. Instead, it frames the evaluation as a comparison to a predicate device to demonstrate "substantial equivalence." The performance is reported in terms of Fine Particle Dose and Leak Test results.

Study TypeUnstated Acceptance Criteria (Implied)Reported LiteTouch Mask Performance (with OptiChamber)Predicate Device (AeroChamber Max) Performance (or baseline)
Fine Particle Dose (0.4-4.7 µg)Maintain comparable Fine Particle Dose to predicate device, ensuring effective drug delivery.Ventolin: 21.1 µgFlovent: 18.5 µgAtrovent: 3.7 µgVentolin: 29.5 µgFlovent: 27.1 µgAtrovent: 5.4 µg
Fine Particle Dose Percentage (Drug mass recovered)Maintain a high percentage of drug mass recovered compared to the chamber without the mask. (No explicit threshold, but values in the 70% range are considered acceptable to support substantial equivalence)Ventolin: 71.5%Flovent: 68.3%Atrovent: 68.5%(OptiChamber without LiteTouch is the baseline for this percentage)
Leak Test Results (with Tilt Angles and Applied Force)Achieve comparable or superior seal effectiveness (lower leaks) to other facemasks, including the predicate device, under various conditions (tilt angles and applied force). (Implicitly, the lower the leak percentage, the better the seal.)See Table below for specific angles and forces. (The LiteTouch mask generally shows comparable or lower leak percentages compared to other masks at various conditions, indicating an effective seal.)See Table below for specific angles and forces of predicate and other masks.

Detailed Leak Test Results (approximated from the image, as text parsing was difficult)

Device0°/0.4lb0°/1lb10°/0.4lb10°/1lb5°/0.4lb5°/1lb
LiteTouch13.8%6.1%79.6%7.7%75.7%7.9%
AeroChamber Plus14.2%2.8%89.5%7.6%74.2%6.5%
OptiChamber49.8%7.7%66.8%4.8%67.8%6.1%
AeroChamber Max4.4%7.3%93.3%7.7%73.6%8.4%
PocketChamber0.8%9.9%97.0%9.8%97.0%7.1%

Note: The leak test results in the provided text are difficult to parse accurately due to formatting issues and seemingly non-English characters. The table above is an interpretation based on the discernible numbers and column headers. The document's conclusion states that the LiteTouch mask demonstrates substantial equivalence in terms of seal, dead space volume, and resistance to inhalation.

2. Sample Size Used for the Test Set and Data Provenance

  • Sample Size: Not explicitly stated for either the Fine Particle Dose or the Leak Test. The tables present aggregated results (e.g., average values) rather than individual sample measurements.
  • Data Provenance: Not explicitly stated (e.g., country of origin). The testing would be considered retrospective as it was conducted to support the 510(k) submission for an already developed device.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

Not applicable. This device is a medical device accessory (mask) for drug delivery. The "ground truth" is established through engineering and performance testing (Fine Particle Dose, Leak Tests) against measurable physical parameters and comparison with predicate devices, not through expert consensus on diagnostic images or clinical outcomes.

4. Adjudication Method for the Test Set

Not applicable. As noted above, this is an objective performance study rather than an evaluation requiring human adjudication (like in a diagnostic image study).

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

No, an MRMC comparative effectiveness study was not done. This type of study is typically used for diagnostic devices where human readers interpret medical images or data, and the AI's impact on their performance is being evaluated. The LiteTouch mask is a physical accessory for drug delivery, not an AI-powered diagnostic tool.

6. If a Standalone (Algorithm Only Without Human-in-the-Loop Performance) Was Done

Yes, the studies conducted (Fine Particle Dose and Leak Tests) represent "standalone" performance in the context of device function. The device's physical properties and interaction with the OptiChamber were tested without direct human interaction as part of the primary measurement, other than technicians operating the testing equipment. The goal was to assess the mask's functional efficacy and integrity.

7. The Type of Ground Truth Used

The ground truth used for these tests is based on measurable physical performance characteristics and comparison to established predicate devices.

  • Fine Particle Dose: Measured efficacy of drug delivery, quantified in micrograms (µg) and percentages. This is an objective, quantitative measure.
  • Leak Test Results: Measured seal effectiveness under controlled conditions (tilt angles, applied force), quantified as a percentage of leak. This is also an objective, quantitative measure.
  • The overall "ground truth" for submission is demonstrating substantial equivalence to an already marketed predicate device based on these engineering and performance tests.

8. The Sample Size for the Training Set

Not applicable. This device is a physical medical device accessory, not an AI or machine learning algorithm that requires a "training set." The design and manufacturing processes are validated, but there isn't a data-driven training phase in the context of artificial intelligence.

9. How the Ground Truth for the Training Set Was Established

Not applicable, as there is no training set for this type of device.

{0}------------------------------------------------

JUN 2 9 2010

Section 5

510(K) SUMMARY

Prepared: October 15, 2009

This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR Part 807.92.

The assigned 510(k) number is: K100285

1. Submitter's Identification:

Respironics New Jersey, Inc. 5 Wood Hollow Road Parsippany, New Jersey 07054 Phone: 973-581-6008 Fax: 973-599-5650 Contact Name: Lauren Ziegler e-mail: lauren.ziegler@respironics.com

Establishment Registration Number : 2243193

Official Correspondent:

Lauren Ziegler Director, QA, RA, and Clinical Affairs 5 Wood Hollow Road Parsippany, New Jersey 07054 Phone: 973-581-6008 Fax: 973-599-5650 e-mail: Lauren.ziegler@respironics.com

  1. Name of the Device:

LiteTouch mask Common Name or Classification Name (21 CFR Part 807.87) of Device: Mask, 21 CFR Part 868.5630. Device Classification: CAF

{1}------------------------------------------------

Predicate Device Information: 3.

Identification of legally marketed device which we claim substantial equivalence to: AeroChamber Max Valved Holding Chamber Small and Medium Mask K032972 AeroChamber Plus Valved Holding Chamber (K992917) Trudell Medical International. 725 Third Street London, Ontario Canada, N5V 5G4

Device Description: 4.

The LiteTouch mask is a Class II device. It is a mask for use with Respironics' OptiChamber Valved Holding Chamber (K962822).

The LiteTouch mask is a mask that utilizes the same operating principles as the AeroChamber Max Valved Holding Chamber Small and Medium Mask (K032972). The difference between the LiteTouch mask and the AeroChamber Max Valved Holding Chamber Small and Medium Mask is that, the LiteTouch mask does not come assembled to the chamber and the exhalation valve is located in different positions and is not of the same shape. (See picture below).

Image /page/1/Figure/5 description: The image is a diagram of a face mask with two labeled parts. The first labeled part, "Face Plate Cushion Assembly", is labeled with the number 1. The second labeled part, "Flap Valve", is labeled with the number 2.

{2}------------------------------------------------

The objective in using a mask is to have patients who are unable to use a spacer mouthpiece or have difficulty with hand-breath coordination get their medication through normal breathing pattern.

To use the LiteTouch mask, the user needs to unpack the mask and attach it to the mouthpiece end of the spacer. The MDI (including the actuator) is inserted into the opening in the MDI adaptor at the distal end of the spacer. The MDI and the spacer combination should be shaken vigorously just prior to inhalation. The patient should exhale normally through the mask. The valve from the mask prevents room air from entering the mask as the patient inhales. The valve's opening and closing movement also indicates that there is an effective mask/face seal; that the patient is breathing and how fast the patient is breathing.

The LiteTouch mask comprises of a Flap Valve and a Cushion plate.

5. Intended Use:

The LiteTouch medium mask for ages 1-5 is to be used with OptiChamber Advantage Valved Holding Chambers. The mask is intended to be used to help administer aerosolized medication from most pressurized metered dose inhalers used with valved holding chambers as prescribed by a physician or health care professional. The intended environments for use include homes, hospitals, and clinics.

6. Comparison to Predicate Devices:

The subject LiteTouch mask and the predicate device (AeroChamber Max Valved Holding Chamber Small and Medium Mask (K032972) are indicated for the same intended use. The difference between the LiteTouch mask and the AeroChamber Max Valved Holding Chamber Small and Medium Mask is that, the LiteTouch mask does not come assembled to the chamber and the exhalation valve is located in different locations and is not of the same shape.

7. Discussion of Non-Clinical Tests Performed for Determination of Substantial Equivalence are as follows:

The LiteTouch medium mask was tested while attached to the OptiChamber Valved Holding Chamber in comparison with the OptiChamber Valved Holding Chamber (K962822) for Fine particle Dose (Reference table 1 below).

{3}------------------------------------------------

FINE PARTICLE DOSE COMPARISON (0.4-4.7µg)

DevicesDrugs
VentolinFloventAtrovent
OptiChamber with LiteTouch21.118.53.7
OptiChamber without LiteTouch29.527.15.4

FINE PARTICLE DOSE PERCENTAGE (Drug mass recovered) OCA with LiteTouch vs OCA without LiteTouch (%)

DevicesDrugs
(OptiChamber with LiteTouch$\div$Optichamber without LiteTouch)*100VentolinFloventAtrovent
71.568.368.5

Table 1

The LiteTouch medium mask was also tested during inhalation in comparison with other facemasks to determine leaks (Reference table 2 below).

{4}------------------------------------------------

DCAICCS
rceLitetouch (softSealmaskAeroChamber Plus masl)ptiChamber mask(PediatriceroChamberMaxmasPocketChamber mask
- డిం-10°00501001 0-50-10°noಕೊ100-ಕಿಂ-10000Cട്രം100-50-10°------------------------------------------------------------------------------------------------------------------------------------------------------------------------------00నం100-50-10°00ਦੇ ਹ1001 0
10-----------------------------13.879.60 776.1 W2381.475 7i7974.289.507665.214.2tર 878.196.3રતું ર77 74.473.693.3C97.98.8C970899.41 00
------------------------------------------------------------------------------------------------------------------------------------------------------------------------------612ﻟﻤﺴﺎﻋﺪ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮVouir73.8TTP://www.firet.re76 1/ 0 . 170.986.494.6I61.50i7749.877 1iતેર/ 55773.868.284.6વેરે તે89.483.8/1.67197ووووووووووووووووووووووووووووووووووووووووووووووووووووووووووووووووووووووووووووووووووووووووووووووووووووووووووووووووووووووووووووووووووووووووووووووووووووووووووووووووووووووووووووو
2.9I46.567.576.66082.4में कि91રે રે66.848 9678うવેત્ત્વના વિદ્યારત તેમ જ દૂધની ડેરી જેવી સવલતો પ્રાપ્ય થયેલી છે. આ ગામનાં લોકોનો મુખ્ય વ્યવસાય ખેતી, ખેતમજૂરી તેમ જ પશુપાલન છે. આ ગામનાં મુખ્યત્વે ખેતP57ಗ74ಗરવಗ7407475y7250C02------------------------------------------------------------------------------------------------------------------------------------------------------------------------------00

LEAK TEST RESULTS WITH TILT ANGLES AND APPLIED FORG

Devices

Table 2

Force:

.

  • , 41b are forces applied to the facemasks while on the face to create a sea
  • gele (0° is the baseline)
    " is the angle of the chine is the baseline creating a gap between the mask and the nose bridge.
    " is the angle of the chin tilting 10° upward from . . . . . .

  • the angle of the chin tilting 5° downward from the baseline creating a gap between the mask and the ch

{5}------------------------------------------------

8. Discussion of Clinical Tests Performed:

Not Applicable

9. Conclusions:

We have demonstrated that the LiteTouch Mask is as safe and effective as a predicate device, the AeroChamber Max Valved Holding Chamber Small and Medium mask, presently on the market, based on substantially equivalent of effectiveness of the seal to the face, Dead space volume, and resistance to inhalation testing. The performance of the LiteTouch before and after washing is also substantially equivalent. The two devices are also substantially equivalent in terms of their operating principle and indications for use.

{6}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/6/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo features a stylized eagle with three lines forming its body and head. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the eagle.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002

Respironics New Jersey, Incorporated C/O Mr. Ned Devine Senior Staff Engineer Underwriters Laboratories, Incorporated 333 Pfingsten Road Northbrook, Illinois 60062

JUN 2 9 2010

Re: K100285

Trade/Device Name: LiteTouch Mask Regulation Number: 21 CFR 868.5630 Regulation Name: Nebulizer Regulatory Class: II Product Code: CAF Dated: June 10, 2010 Received: June 14, 2010

Dear Mr. Devine:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{7}------------------------------------------------

Page 2- Mr. Devine

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to

http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely yours.

Wh for

Anthony D. Watson, B.S., M.S., M.B.A. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{8}------------------------------------------------

Indications for Use

510(k) Number (if known): K100285

Device Name: LiteTouch mask

Indications for Use:

The LiteTouch medium mask for ages 1-5 is to be used with OptiChamber Advantage Valved Holding Chambers. The mask is intended to be used to help administer aerosolized medication from most pressurized metered dose inhalers used with valved holding chambers as prescribed by a physician or health care professional. The intended environments for use include homes, hospitals, and clinics.

Prescription Use X (Part 21 CFR 801 Subpart D)

AND/OR

Over-The-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

$\qquad$ Concurrence of CDRH, Office of Device Evaluation (ODE)

Division Sign-Off) Jivision of Anesthesiology, General Hospital infection Control, Dental Devices

510(k) Number: K100285 4-2

i

§ 868.5630 Nebulizer.

(a)
Identification. A nebulizer is a device intended to spray liquids in aerosol form into gases that are delivered directly to the patient for breathing. Heated, ultrasonic, gas, venturi, and refillable nebulizers are included in this generic type of device.(b)
Classification. Class II (performance standards).