K Number
K093517
Manufacturer
Date Cleared
2010-06-29

(228 days)

Product Code
Regulation Number
880.6250
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The nitrile examination gloves are a disposable device intended for medical and dental purposes that is worn on the examiner's hand to prevent contamination between the patient and examiner.

Device Description

Powder-Free Nitrile Medical Examination Gloves meets the current specifications listed under the ASTM Specification D6319-00a, Standard Specification for Nitrile Examination Gloves for Medical Application.

AI/ML Overview

Here's an analysis of the provided text, focusing on the acceptance criteria and study information for the NitriTech® Non-Sterile Powder-Free Nitrile Examination Gloves:

This document is a 510(k) summary for a medical device (examination gloves). As such, the "acceptance criteria" and "study" described are for the device itself and its physical properties and biocompatibility, not for an AI algorithm. Therefore, many of the questions related to AI-specific study design (like MRMC, standalone AI performance, training sets, expert qualifications for ground truth in AI) are not applicable.


Acceptance Criteria and Device Performance

1. Table of Acceptance Criteria and Reported Device Performance

TestAcceptance Criteria (ASTM 6319-00a)Reported Device Performance (Powder-Free Nitrile Examination Gloves)
Detection of Holes (ASTM 5151)Multiple Normal G1, AQL = 2.5Pass G1 AQL = 2.5*
Length (mm)
- Size XSMin 220240 mm minimum for all sizes*
- Size SMin 220
- Size MMin 230
- Size LMin 230
- Size XLMin 230
Palm Width (mm)
- Size XS70 ± 1073-78*
- Size S80 ± 1083-88*
- Size M95 ± 1093-98*
- Size L111 ± 10103-107*
- Size XL120 ± 10110-118*
Thickness (mm)
- FingerMin 0.05Min 0.06*
- PalmMin 0.05Min 0.05*
Physical Properties - Before Aging
- Tensile Strength (Mpa)Min 1415 - 21*
- Ultimate Elongation (%)Min 500550 - 630*
Physical Properties - After Aging
- Tensile Strength (Mpa)Min 1414 - 22*
- Ultimate Elongation (%)Min 400520 - 610*
Powder ContentMax 2.0 mg/gloveBelow 2.0 mg/glove*
Primary Skin Irritation (CPSC Title 16 Chapter II Part 1500)PII ≥ 5.0 indicates an irritate (irritant threshold)PII was "0.04"*
Dermal Sensitization Test (ISO 10993-10)Score ≥ 1 indicates irritation/hypersensitivity (sensitizer threshold)Score was "0"*

* Indicates data points explicitly marked with an asterisk in the source document, referencing "Third Party test data is available in Appendix A" or "Amendment F" (for biocompatibility).


Study Details (Applicable to Device Performance)

2. Sample size used for the test set and the data provenance

The document provides acceptance criteria from ASTM D6319-00a and CPSC/ISO standards. It states that the device "meets the ASTM D6319-00a Standard" and "meets... CPSC Title 16 Chapter II Part 1500 and ISO 10993-10 requirements."

  • Test Sets: The document refers to various ASTM test methods (D5151 for holes, D412/D573 for tensile properties, D6124-06 for powder content) and specific biocompatibility tests (CPSC Title 16 Chapter II Part 1500 for Primary Skin Irritation and ISO 10993-10 for Dermal Sensitization).
  • Sample Sizes: The exact sample sizes for each specific test (e.g., how many gloves were tested for holes, how many animals for biocompatibility) are not explicitly specified in this summary document. It only provides the performance results.
  • Data Provenance: The document states "Third Party test data is available in Appendix A" for the general performance tests and "Third Party test data is available in Amendment F" for biocompatibility. The biocompatibility tests were completed by "Mediamal Testing and Consultancy Services." The country of origin for the data is not specified beyond originating from this third-party testing entity. The studies are prospective in nature, as they involve actively testing new device samples.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

This question is not applicable to this type of device and study. The "ground truth" here is established by standardized physical and chemical tests (e.g., measuring glove dimensions, tensile strength, chemical content, biological reactions) rather than expert interpretation of complex data (like medical images). The "experts" would be the technicians and scientists performing the ASTM, CPSC, and ISO standardized tests, ensuring they follow the protocols correctly. Their qualifications are inherent in the accreditation of the third-party testing facility.

4. Adjudication method for the test set

Not applicable. The tests are objective measurements or standardized biological assays, not subjective assessments requiring adjudication among multiple human readers.

5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

Not applicable. This is a physical medical device (examination gloves), not an AI-powered diagnostic or assistive tool for human readers.

6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done

Not applicable. No AI algorithm is involved.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

The ground truth is established by:

  • Standardized Physical Measurements: Length, width, thickness, tensile strength, elongation are objective physical properties measured according to ASTM standards.
  • Chemical Analysis: Powder content is determined by a chemical analysis method (ASTM D6124-06).
  • Standardized Biological Assays: Primary Skin Irritation and Dermal Sensitization tests are conducted according to CPSC and ISO standards on biological models (animal testing as implied by the standards for such biocompatibility tests). The "ground truth" for irritation/sensitization is based on observing specific, quantified biological responses compared to established thresholds.

8. The sample size for the training set

Not applicable. There is no AI model or algorithm that requires a "training set" for these examination gloves. The device is evaluated based on its inherent physical and chemical properties.

9. How the ground truth for the training set was established

Not applicable, as no training set exists for this type of device.

{0}------------------------------------------------

510K Summary

. / -

510K: 093517

1. Information
Submitter's
Name:Summit Glove Inc.
Address:310 South Grant Street
Minerva, Ohio 44657
Phone:330-868-0066
Contact Person:James Moore
E-mail:jmoore@summitglove.com
2. Name of Device
Trade or Proprietary Name:NitriTech®, Non-Sterile Powder-Free NitrileExamination Glove in the color blue
Common or Usual Name:Synthetic Rubber Examination Gloves
Classification Name:Patient Examination Glove, Powder-Free
3. Predicate Device:
Non-Sterile Nitrile Powder-Free Medical ExaminationGloves Submission Number: K090828. Based on the belowtest results, it is concluded the subject gloves are asubstantial equivalent to the predicate device.

4. Identification of the Legally Marketed Devices

Powder-Free Nitrile Examination Gloves as described in this 510K Notification is substantially equivalent to the Class 1 patient examination glove bearing the product code 80LZA (21CFR 880.6250). It meets all the current specifications listed under the ASTM Specification D6319-00a, Standard for Nitrile Gloves for Medical Application.

5. Description of Device

Powder-Free Nitrile Medical Examination Gloves meets the current specifications listed under the ASTM Specification D6319-00a, Standard Specification for Nitrile Examination Gloves for Medical Application.

6. Intended Use of Gloves

This is a disposable device intended for medical purposes: that is worn on the examiner hand, to prevent contamination between the patient and examiner.

{1}------------------------------------------------

7. Summary of Performance Data:

Performance data of gloves based on ASTM D6319-00a. ASTM D6319-00a Section 9 is inapplicable for non-sterile examination grade nitrile gloves.

TestASTM 6319-00aPowder-Free NitrileExamination Gloves
1. Detection of Holes (ASTM5151)Multiple NormalG1AQL = 2.5Pass G1 AQL = 2.5*
2. Length (mm)Size XSSMLXLMin 220 -Min 220 -Min 230 -Min 230 -Min 230240 mm minimum for allsizes*
3. Palm Width (mm)Size XSSMLXL70+ 1080+ 1095+ 10111+ 10120 + 1073-7883-8893-98103-107110 - 118*
4. Thickness (mm)FingerPalmMin 0.05Min 0.05Min 0.06Min 0.05*
5. Physical PropertiesBefore AgingTensile Strength (Mpa)Ultimate Elongation (%)After AgingTensile Strength (Mpa)Ultimate Elongation (%)Min 14Min 500Min 14Min 40015 - 21550 - 630*14 - 22520 - 610*
6. Powder ContentMax 2.0mg/gloveBelow 2.0mg/glove
7. Primary Skin Irritation CPSCTitle 16 Chapter II Part 1500PII Greater or equal to 5.0indicates an irritatePII was "0.04"*
8. Dermal Sensitization Test ISO10993-10Score greater or equal to 1indicatesirritation/hypersensitivityScore was "0"*

*Third Party test data is available in Appendix A

The performance data of the gloves as shown above meet the ASTM D6319-00a Standard using the following test methods:

{2}------------------------------------------------

  • ASTM D412 Tensile Properties Unaged Test Method A ●
  • ASTM D573 Tensile Properties Aged .
  • ASTM D5151 Detection of Holes .
  • ASTM D6124-06 Section 7 Procedure I for Powder-free Gloves .
    1. The Biocompatibility Test consists of Primary Skin Irritation (CPSC Title 16 Chapter II Part 1500) and Dermal Sensitization Tests (ISO 10993-10).

*Third Party test data is available in Amendment F

The conclusion of the Biocompatibility Testing would indicate that there is no irritation or hypersensitivity to the subject gloves (NitriTech® Non-Sterile Powder-Free Nitrile Examination Gloves in the color blue).

9. Conclusion:

It is concluded that the NitriTech® Non-Sterile Powder-Free Nitrile Examination Gloves in the color blue for this submission meets ASTM D6319-00a, CPSC Title 16 Chapter II Part 1500 and ISO 10993-10 requirements.

{3}------------------------------------------------

Statement concerning the Biocompatibility Testing

The gloves tested in the Biocompatibility Testing were the subject glove (NitriTech®, Non-Sterile Powdered Nitrile Examination Gloves in the color blue) of this submission. Mediamal Testing and Consultancy Services had the Primary Skin Irritation (study report # MB-PSI-27-09) and Dermal Sensitization Test (study report # MB-DSAb-27-09) completed to CPSC Title 16 Chapter II Part 1500 and ISO 10993-10 standards.

18

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized graphic of an eagle or bird-like figure with three curved lines representing its wings or body. The graphic is positioned to the right of a circular arrangement of text that reads "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA".

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002

JUN 2 9 2010

Mr. James Moore Project Coordinator Summit Glove, Incorporated 310 South Grant Street Minerva, Ohio 44657

Re: K093517

Trade/Device Name: NitriTech® Non-Sterile Powdered Nitrile Medical Examination Gloves in the color blue Regulation Number: 21 CFR 880.6250 Regulation Name: Patient Examination Glove Regulatory Class: I Product Code: LZA Dated: June 10, 2010 Received: June 15, 2010

Dear Mr. Moore:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{5}------------------------------------------------

Page 2- Mr. Moore

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reportung (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to

http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.goy/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

f

Sincerely yours

Susan Runner

Anthony D. Watson, B.S., M.S., M.B.A. Director

Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure .

{6}------------------------------------------------

Indications for Use

510(k) Number (if known): K093517

Device Name: NitriTech® Non-Sterile Powder Free Nitrile Medical Examination Gloves in the color blue.

Indications For Use:

The nitrile examination gloves are a disposable device intended for medical and dental purposes that is worn on the examiner's hand to prevent contamination between the patient and examiner.

AND/OR Over-The-Counter Use Prescription Use × (Part 21 CFR 801 Subpart D) (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

signature

(Division-Sign-Off) (Division of Anesthesiology, General Hospital Infection Control, Dental Devices

K093517 510(k) Number: _______________________________________________________________________________________________________________________________________________________________

§ 880.6250 Non-powdered patient examination glove.

(a)
Identification. A non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. A non-powdered patient examination glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.(b)
Classification. Class I (general controls). The device, when it is a finger cot, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 880.9.