(289 days)
No
The document describes a mechanical and chemical system for hemodialysis and does not mention any AI or ML components or functionalities.
Yes
The device is described as a "Sorbent System" intended for "adult acute and chronic uremic patients where hemodialysis is prescribed," which directly treats a medical condition.
No
Explanation: The device is described as a hemodialysis system for treating uremic patients, which functions to purify and regenerate dialysate. Its purpose is therapeutic treatment, not diagnosis. Although it is stated that it operates to meet performance specifications and has passed verification tests, this is in reference to its functional operation as a treatment device.
No
The device description clearly states it consists of two distinct hardware components: the 2008 Machine and the SORB™ Module, which is a sorbent dialysate regenerative system. It also describes the physical process of dialysate purification and regeneration.
Based on the provided information, the 2008 Sorbent System is not an In Vitro Diagnostic (IVD) device.
Here's why:
- Intended Use: The intended use is for treating uremic patients through hemodialysis. This is a therapeutic procedure performed on the patient's body, not a test performed on a sample taken from the body.
- Device Description: The description details a system for purifying and regenerating dialysate used in hemodialysis. This is a life support and treatment system, not a device for analyzing biological samples.
- Lack of IVD Characteristics: There is no mention of analyzing biological samples (like blood, urine, etc.), detecting analytes, or providing diagnostic information based on such analysis.
IVD devices are specifically designed to examine specimens derived from the human body to provide information for diagnostic, monitoring, or compatibility purposes. The 2008 Sorbent System's function is to facilitate the process of removing waste products from the blood, which is a treatment, not a diagnostic test.
N/A
Intended Use / Indications for Use
The 2008 Sorbent System is intended for adult acute and chronic uremic patients in the presence of a healthcare practitioner where hemodialysis is prescribed on the order of a physician.
Product codes (comma separated list FDA assigned to the subject device)
FKT
Device Description
The 2008 Sorbent System is intended for adult acute and chronic uremic patients in the presence of a healthcare practitioner where hemodialysis is prescribed on the order of a physician. The 2008 Sorbent System operates in a similar manner to standard hemodialysis systems currently approved for use; it is substantially equivalent to the Fresenius 2008T hemodialysis machine (K080964) and the Renal Solutions Allient® Sorbent Hemodialysis System (K070739). The significant difference between the 2008 Sorbent System and standard single-pass hemodialysis machines is that the 2008 Sorbent System utilizes a sorbent cartridge to purify dialysate made from potable tap water and then regenerates fresh dialysate from spent dialysate (like the Allient Sorbent Hemodialysis System). The 2008 Sorbent System consists of two distinct components: the 2008 Machine, and the SORB™ Module. The SORB Module is a sorbent dialysate regenerative system that attaches to the 2008 Machine and replaces the machine's existing single-pass dialysate delivery system. During treatment, used dialysate is chemically reprocessed into fresh, new dialysate and sent back to the dialyzer instead of being sent down a drain. Bv recirculating and regenerating the dialysate, the 2008 Sorbent System uses less than 15 liters of potable tap water per treatment compared to a standard single-pass system that uses at least 120 liters of purified water during a standard 4-hour dialysis treatment.
Mentions image processing
Not Found
Mentions AI, DNN, or ML
Not Found
Input Imaging Modality
Not Found
Anatomical Site
Not Found
Indicated Patient Age Range
adult
Intended User / Care Setting
healthcare practitioner
Description of the training set, sample size, data source, and annotation protocol
Not Found
Description of the test set, sample size, data source, and annotation protocol
Not Found
Summary of Performance Studies (study type, sample size, AUC, MRMC, standalone performance, key results)
The 510(k) verification (non-clinical) testing information consists of performance, safety, and software testing. Functional testing demonstrated that the device performs as designed and expected, including specific verification tests, protocol descriptions (objective, test articles, methods, acceptance criteria), and system-level hazard analysis. Safety testing per FDA Consensus Standards included Electrical safety testing (IEC 60601-1), Electromagnetic compatibility (EMC) testing (IEC 60601-2-16), Biocompatibility testing (AAMI / ANSI / ISO 10993-1) on new patient-fluid contacting materials, and Risk Analysis (ISO 14971). Software testing confirmed the device software meets design input requirements, including structural chart/flow chart, summary of development procedures, software requirements specification with traceability, verification and validation test plans, system level test results, and current software version information.
Key Metrics (Sensitivity, Specificity, PPV, NPV, etc.)
Not Found
Predicate Device(s): If the device was cleared using the 510(k) pathway, identify the Predicate Device(s) K/DEN number used to claim substantial equivalence and list them here in a comma separated list exactly as they appear in the text. List the primary predicate first in the list.
Allient Sorbent Hemodialysis System, Fresenius 2008T Hemodialysis System
Reference Device(s): Identify the Reference Device(s) K/DEN number and list them here in a comma separated list exactly as they appear in the text.
Not Found
Predetermined Change Control Plan (PCCP) - All Relevant Information for the subject device only (e.g. presence / absence, what scope was granted / cleared under the PCCP, any restrictions, etc).
Not Found
§ 876.5600 Sorbent regenerated dialysate delivery system for hemodialysis.
(a)
Identification. A sorbent regenerated dialysate delivery system for hemodialysis is a device that is part of an artificial kidney system for the treatment of patients with renal failure or toxemic conditions, and that consists of a sorbent cartridge and the means to circulate dialysate through this cartridge and the dialysate compartment of the dialyzer. The device is used with the extracorporeal blood system and the dialyzer of the hemodialysis system and accessories (§ 876.5820). The device includes the means to maintain the temperature, conductivity, electrolyte balance, flow rate and pressure of the dialysate, and alarms to indicate abnormal dialysate conditions. The sorbent cartridge may include absorbent, ion exchange and catalytic materials.(b)
Classification. Class II (performance standards).
0
K093362 Page 1 of 4
2008 Sorbent Hemodialysis System 510(k) Premarket Notification Section 5: 510(k) Summary
AUG 1 3 2010
Official Contact David J. Vanella Senior Vice President, Quality Systems Renal Solutions, Inc. 770 Commonwealth Drive Warrendale, Pa, 15086 Phone: (724) 720-2840 FAX: (724) 772-6925 Classification Name System, Dialysate Delivery, Sorbent Regenerated Regulation Number 876.5600 Product Code FKT Common/Usual Name Sorbent Hemodialysis System Proprietary Name 2008 Sorbent Hemodialysis System Predicate Devices Allient Sorbent Hemodialysis System Fresenius 2008T Hemodialysis System Reason for submission New Device
Section 5: 510(k) Summary K093362
Substantial Equivalence
The 2008 Sorbent Hemodialysis System device has the following key similarities to the predicate devices:
- Intended use .
- Environment of use .
- . Operating principle
- Technology .
Renal Solutions has determined that the differences from the predicate devices have no impact on the safety and effectiveness of the device and that the technological characteristics of the 2008 Sorbent System do not raise new types of safety or effectiveness questions. The information provided including design verification, risk management, safety, biocompatibility and comparative testing were performed on the 2008 Sorbent Hemodialysis System to demonstrate the device meets the design requirements and to support substantial equivalence. In summary, the 2008 Sorbent Hemodialysis System is substantially equivalent to the predicate devices.
1
K093362 Page 2 of 4
2008 Sorbent Hemodialysis System 510(k) Premarket Notification Section 5: 510(k) Summary
The 2008 Sorbent System incorporates the addition of a module that regenerates spent dialysate and delivers clean dialysate to the dialysis system and required modifications to the 2008 dialysis system. A risk analysis has been established for the 2008 Sorbent System and has includes the potential hazards associated with the device where all hazards have been identified and mitigated to acceptable risk levels.
2008 Sorbent System 510(k) Verification (Non-Clinical) Testing Summary
The verification (non-clinical) testing information consists of performance, safety, and software testing that was performed to verify the 2008 Sorbent System meets its performance specifications and to demonstrate the device is substantially equivalent to the established predicate devices.
The following table summarizes the 510(k) verification testing activities performed. These include performance, safety and software testing, which demonstrates by technical examination that the 2008 Sorbent System meets its performance specifications, the designated (FDA Consensus) standard requirements, and the software design input requirements.
510(k) Verification Testing | 510(K) Verification Testing Activities |
---|---|
2008 Sorbent System Performance Testing | |
The 510(k) contains details of the performance | Functional testing that demonstrates that the device |
performs as designed and expected, includes the | |
following: | |
• The specific verification tests conducted | |
(verification) testing and includes the results | • A description of all test protocol including: |
that support the performance characteristics of | • objective of the tests |
the 2008 Sorbent System. | • test articles used in the tests |
• test methods and procedures | |
• pre-defined acceptance or pass/fail criteria. | |
• System-level hazard analysis evaluation that | |
confirms that the device does not perform in an | |
unexpected and/or unsafe manner | |
2008 Sorbent System Safety Testing | |
The 510(k) contains details of the safety | Product safety testing that demonstrates that the device |
performs per the FDA Consensus Standards, as | |
identified below: | |
(standards) verification testing and includes the | |
results that support the safety characteristics of | |
the 2008 Sorbent System. | • Electrical safety testing (IEC 60601-1:1988 + |
A1:1991 + A2:1995 Medical electrical equipment- | |
Part 1:General requirements for safety) | |
Biocompatibility testing was performed on all | |
new materials that are patient-fluid contacting | • Electromagnetic compatibility (EMC) testing (IEC |
60601-2-16: 1998-02 Medical electrical equipment | |
Part 2-16: Particular requirements for the safety of | |
Hemodialysis, haemodiafiltration and haemofiltration | |
equipment) | |
• Biocompatibility Testing (AAMI / ANSI / ISO 10993- | |
1:2003 Biological evaluation of medical devices Part | |
1: Evaluation & testing) | |
• Risk Analysis (ISO 14971:2007 Medical devices - | |
Application of risk management to medical device) |
2
K093362 Page 3 of 4
2008 Sorbent Hemodialysis System 510(k) Premarket Notification Section 5: 510(k) Summary
510(k) Verification Testing | 510(K) Verification Testing Activities |
---|---|
2008 Sorbent System Software Testing | |
The 510(k) contains details of the software | |
testing and includes the required | |
documentation as required per the document | |
titled "Guidance for the Content of Premarket | |
Submissions for Software". | Software testing that demonstrates the device software |
meets the design input requirements. The device was | |
tested and includes the following | |
Structure chart of flow chart describing software | |
architecture Summary of software development procedures, | |
including changes made to the software Software requirements specification with traceability | |
back to the hazard analysis Verification and validation test plans, including | |
pass/fail criteria and traceability back to the | |
requirements; System level test results Current software version number and date of latest | |
revision. |
The conclusions drawn from this testing demonstrates that the 2008 Sorbent System is as safe, as effective, and performs at least as safely and effectively as the legally marketed devices identified as predicate devices to which it was compared.
Indications of Use
The 2008 Sorbent System is intended for adult acute and chronic uremic patients in the presence of a healthcare practitioner where hemodialysis is prescribed on the order of a physician.
General Safety and Effectiveness
The intended use for the 2008 Sorbent System is for treatment of adult acute and chronic uremic patients in the presence of a healthcare practitioner where hemodialysis is prescribed on the order of a physician. The operator must be knowledgeable of hemodialysis methodology and relevant physiology, proficient in healthcare procedures, thoroughly familiar with the contents of the Operator's Manual and fully trained and qualified to operate this device.
Device Description
The 2008 Sorbent System is intended for adult acute and chronic uremic patients in the presence of a healthcare practitioner where hemodialysis is prescribed on the order of a physician. The 2008 Sorbent System operates in a similar manner to standard hemodialysis systems currently approved for use; it is substantially equivalent to the Fresenius 2008T hemodialysis machine (K080964) and the Renal Solutions Allient® Sorbent Hemodialysis System (K070739).
3
K093362 Page 4 of 4
2008 Sorbent Hemodialysis System 510(k) Premarket Notification
Section 5: 510(k) Summary
The significant difference between the 2008 Sorbent System and standard single-pass hemodialysis machines is that the 2008 Sorbent System utilizes a sorbent cartridge to purify dialysate made from potable tap water and then regenerates fresh dialysate from spent dialysate (like the Allient Sorbent Hemodialysis System).
The 2008 Sorbent System consists of two distinct components:
the 2008 Machine, and
the SORB™ Module.
The SORB Module is a sorbent dialysate regenerative system that attaches to the 2008 Machine and replaces the machine's existing single-pass dialysate delivery system.
During treatment, used dialysate is chemically reprocessed into fresh, new dialysate and sent back to the dialyzer instead of being sent down a drain. Bv recirculating and regenerating the dialysate, the 2008 Sorbent System uses less than 15 liters of potable tap water per treatment compared to a standard single-pass system that uses at least 120 liters of purified water during a standard 4-hour dialysis treatment.
4
Image /page/4/Picture/0 description: The image shows a logo for the Department of Health & Human Services. The logo features a stylized abstract symbol resembling an eagle or bird in flight, composed of three curved lines. The text "DEPARTMENT OF HEALTH & HUMAN" is arranged in a circular fashion around the symbol.
DEPARTMENT OF HEALTH & HUMAN SERVICES
Food and Drug Administration 10903 New Hampshire Avenue Document Mail Center - WO66-G6 Silver Spring, MD 20993-0002
Mr. David J. Vanella Senior Vice President, Quality Systems Renal Solutions, Inc. 770 Commonwealth Drive, Suite 101 WARRENDALE PA 15086
ന്നു
പ്രാ 2010 AUG
. Re: K093362
Trade/Device Name: 2008 Sorbent System Regulation Number: 21 CFR §876.5600
Regulation Number: 21 CFR §70.500
Regulation Name: Sorbent regenerated dialysate delivery system for hemodialysis Regulatory Class: II
Product Code: FKT
Dated: August 10, 2010
Received: August 11, 2010
Dear Mr. Vanella:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indivitions for use stated in the enclosure) to legally marketed predicate devices marked in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract indici warranties. We remind you, however, that device labeling must be truthful and not nisleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related
5
Page 2
adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Huker Lemur us
ferbert P. Lerner, M.D., Director (Acting) Division of Reproductive, Gastro-Renal and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
6
2008 Sorbent Hemodialysis System 510(k) Premarket Notification Section 4: Indications for Use
AUG 1 3 2010
Section 4: Indications for Use Statement
510(k) Number (if known):-N/A K093362
Device Name: 2008 Sorbent System
Indications for Use:
The 2008 Sorbent System is intended for adult acute and chronic uremic patients in the presence of a healthcare practitioner where hemodialysis is prescribed on the order of a physician.
Prescription Use × (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE -CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
signature
(Division Sign-Off) (Division of Reproductive, Abdominal, and Radiological Devices 510(k) Number
Page 10