K Number
K092937
Date Cleared
2010-08-02

(312 days)

Product Code
Regulation Number
862.1150
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The STA® - Hybrid Hep Calibrator is a set of calibrator plasmas intended for use with analyzers of the STA® line suitable to these reagents, for the calibration of heparin (UFH and LMWH) activity assay by measuring the anti-Xa activity using the chromogenic method, STA® - Rotachrom® Heparin.

Device Description

The STA® Hybrid Hep Calibrator is a set of lyophilized human plasmas used to create the calibration curve on the STA® line of IVD instruments performing the chromogenic method for heparin (UFH and LMWH) assays. Each STA® Hybrid Hep Calibrator available contains: 4 x 1-ml vials of Reagent 1: STA® Hybrid Hep Calibrator O lyophilized human plasma free of heparin. 4 x 1-ml vials of Reagent 2: STA® Hybrid Hep Calibrator 3 lyophilized human plasma containing a well-defined quantity of UFH. 4 x 1-ml vials of Reagent 3: STA® Hybrid Hep Calibrator 6 lyophilized human plasma containing a well-defined quantity of UFH that is greater than that of Reagent 2. 4 x 1-ml vials of Reagent 4: STA® - Hybrid Hep Calibrator 9 lyophilized human plasma containing a well-defined quantity of LMWH. 4 x 1-ml vials of Reagent 5: STA® Hybrid Hep Calibrator 18 lyophilized human plasma containing a well-defined quantity of LMWH that is greater than that of Reagent 4.

AI/ML Overview

The provided text describes a 510(k) summary for the STA® - Hybrid Hep Calibrator, a medical device. This submission seeks to bundle two previously cleared devices (STA® - Hepanorm® H and STA® - Calibrator HBPM/LMWH Kit) into a single kit. The document focuses on demonstrating substantial equivalence to these predicate devices for regulatory purposes.

Therefore, the document does not contain information about:

  • Specific acceptance criteria for device performance (e.g., accuracy, precision) as typically defined in a clinical study.
  • A formal study that proves the device meets specific acceptance criteria in terms of measured performance metrics.
  • Sample sizes for test sets, data provenance, number of experts for ground truth, adjudication methods, MRMC comparative effectiveness studies, standalone performance, or grand truth types.
  • Training set sample sizes or how their ground truth was established, as this is a calibrator device and not an AI/ML diagnostic system.

The document states that the STA® - Hybrid Hep Calibrator and its predicate devices are "identical products regarding indication/intended use, formulation or materials of construction and design, technology, and safety." The primary difference is the bundling of two existing calibrators into one kit. The substantial equivalence argument relies on the fact that no new questions of safety, effectiveness, or technology are raised due to this bundling.

Essentially, the "study" proving the device meets acceptance criteria is implied by its substantial equivalence to previously cleared devices. The acceptance criteria, in this regulatory context, are primarily that the bundled calibrator performs equivalently to the two separate predicate calibrators, which have already been deemed safe and effective.

Summary of available information related to device performance and regulatory acceptance:

Acceptance Criteria (Implied by Substantial Equivalence)Reported Device Performance (Implied by Substantial Equivalence)
Performs calibration for heparin (UFH and LMWH) activity assays by measuring anti-Xa activity using the chromogenic method (STA® - Rotachrom® Heparin).The STA® - Hybrid Hep Calibrator is "identical" to the predicate devices (STA® - Calibrator HBPM/LMWH Kit and STA® - Hepanorm® H) regarding indication/intended use, technology, and principles of operation.
Operates with STA® line of IVD instruments.Designed to operate utilizing the STA® product line of IVD coagulation analyzers for creating calibration curves for heparin assays.
Ensures safety and effectiveness for its intended use.Yields "no new questions in safety, effectiveness, or technology" compared to the predicate devices.

Information not available in the provided text:

  • Sample sized used for the test set and the data provenance: Not applicable in this context as this is a calibrator, not a diagnostic algorithm. Performance is assessed through equivalence to existing calibrators.
  • Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not applicable.
  • Adjudication method for the test set: Not applicable.
  • If a multi reader multi case (MRMC) comparative effectiveness study was done: Not applicable.
  • If a standalone (i.e. algorithm only without human-in-the-loop performance) was done: Not applicable as this is a calibrator device, not an algorithm.
  • The type of ground truth used: For calibrators, ground truth typically refers to the accurately assigned values of the analytes (UFH and LMWH in this case) within the calibrator plasmas. The text states the plasmas contain "well-defined quantity of UFH" and "well-defined quantity of LMWH", implying these values are established and verified during manufacturing of the calibrator materials.
  • The sample size for the training set: Not applicable as this is not an AI/ML device.
  • How the ground truth for the training set was established: Not applicable.

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the word "Stago" in a serif font, with the letters evenly spaced and of uniform size. Above the word, there is a stylized graphic element that appears to be composed of three curved shapes arranged in a circular fashion, resembling a stylized flower or abstract design. The graphic is solid black, contrasting with the white background.

510(k) Summary

Submitter's Name, Address, Telephone Number, Contact Person and Date Prepared:

Umberto V. Parrotta, Jr. Diagnostica Stago, Inc. Five Century Drive Parsippany, New Jersey 07054

K092937

AUG 0 2 2010

Phone: (973) 631-1200, x-2044 Facsimile: (973) 695-0095

Contact Person: Umberto V. Parrotta, Jr.

Date Prepared: September 23, 2009

Name of Device and Name/Address of Sponsor:

STA® - Hybrid Hep Calibrator

Diagnostica Stago, Inc. Five Century Drive Parsippany, New Jersey 07054

Common or Usual Name:

  • IVD Coagulation Calibration Device. ●
  • IVD Calibrator. .

Classification Name:

  • Name: Calibrator, Secondary. .
  • Description: Calibrator .

Predicate Device:

  • STA® Calibrator HBPM/LMWH Kit (K010350). ●
  • STA® Hepanorm® H (K854762) { formerly cleared as Hepanorm® . Calibration Plasma Set}.

Purpose of the Traditional 510(k) Notice:

STA® - Hybrid Hep Calibrator is a new device that bundles two of the Company's previously FDA cleared devices, STA® - Hepanorm® H and STA® - Calibrator

DSI-STA-HHC-Attachment-04v1-510(k)Summary.doc

{1}------------------------------------------------

Image /page/1/Picture/1 description: The image shows the word "Stago" in a serif font. Above the word is a graphic consisting of two curved shapes that resemble stylized leaves or petals. The shapes are arranged in a circular fashion, creating a sense of movement or rotation above the word.

510(k) Summary

HBPM/LMWH into a single kit. Thus, the primary difference between the subject product and predicate devices is that it contains a set of calibration plasmas for calibration of assays of unfractionated heparin (UFH) and Low Molecular Weight Heparin (LMWH) bundled together in a single kit.

Indication/Intended Use:

The STA® - Hybrid Hep Calibrator is a set of calibrator plasmas intended for use with analyzers of the STA® line suitable to these reagents, for the calibration of heparin (UFH and LMWH) activity assay by measuring the anti-Xa activity using the chromogenic method, STA® - Rotachrom® Heparin.

Technological Characteristics:

  • The STA® Hybrid Hep Calibrator is a set of lyophilized human plasmas . used to create the calibration curve on the STA® line of IVD instruments performing the chromogenic method for heparin (UFH and LMWH) assays.
  • Each STA® Hybrid Hep Calibrator available contains: .
    • 0 4 x 1-ml vials of Reagent 1: STA® Hybrid Hep Calibrator O lyophilized human plasma free of heparin.
    • 4 x 1-ml vials of Reagent 2: STA® Hybrid Hep Calibrator 3 o lyophilized human plasma containing a well-defined quantity of UFH.
    • 4 x 1-ml vials of Reagent 3: STA® Hybrid Hep Calibrator 6 o lyophilized human plasma containing a well-defined quantity of UFH that is greater than that of Reagent 2.
    • (9) 4 x 1-ml vials of Reagent 4: STA® - Hybrid Hep Calibrator o lyophilized human plasma containing a well-defined quantity of LMWH.
    • 4 x 1-ml vials of Reagent 5: STA® Hybrid Hep Calibrator (18 o lyophilized human plasma containing a well-defined quantity of LMWH that is greater than that of Reagent 4.

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image shows the logo for Stago. The logo consists of the word "Stago" in a serif font, with a stylized graphic above it. The graphic appears to be an abstract shape, possibly representing a stylized eye or a leaf. The logo is simple and modern in design.

510(k) Summary

Principles of Operation

The STA® - Hybrid Hep Calibrator is designed to operate utilizing the STA® product line of IVD coagulation analyzers for the purpose of creating calibration curves for assays of heparin by measuring the anti-Xa activity. While the subject device performs calibration of assays based on chromogenic method, the predicate devices perform calibration of assays based on chromogenic method and on clotting method, both being anti-Xa method.

Substantial Equivalence:

The STA® - Hybrid Hep Calibrator, the subject product of this submission, and the predicate devices, STA® - Calibrator HBPM/LMWH Kit and STA® - Hepanorm® H are identical products regarding indication/intended use, formulation or materials of construction and design, technology, and safety. The primary difference between the subject product and predicate devices is that the product contains a set of calibration plasmas for assays of both unfractionated heparin (UFH) and Low Molecular Weight Heparin (LMWH) bundled in a single kit whereas the predicate devices are marketed as two (2) separate and distinct products - one kit for UFH and the other kit for LMWH. In summary, the predicate devices are bundled together as a single kit. The STA® - Hybrid Hep Calibrator like the predicate devices, is a calibrator used in creating the calibration curve on IVD devices for the purpose of Heparin activity assay by measuring the anti-Xa activity.

The product, STA® - Hybrid Hep Calibrator and the predicate devices. STA® -Hepanorm® H and STA®- Calibrator HBPM/LMWH are similar in indication/intended use, technology, principles of operation, and application of use (with IVD medical devices) thus yielding no new questions in safety, effectiveness, or technology. Therefore, this concludes the product STA® - Hybrid Hep Calibrator is substantially equivalent to the predicate devices, STA® - Calibrator HBPM/LMWH Kit (K010350) and STA® - Hepanorm® H {previously cleared as Hepanorm® Calibration Plasma Set} (K854762).

{3}------------------------------------------------

Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized depiction of an eagle with three curved lines representing its wings. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" is arranged in a circular pattern around the eagle.

Food and Drug Administration 10903 New Hampshire Avenue Document Mail Center - WO66-0609 Silver Spring. MD 20993-0002

Diagnostica Stago, Inc. c/o Mr. Umberto V. Parrotta Director of Regulatory Affairs and Quality Assurance Five Century Drive Parsippany, New Jersey 07054

AUG 0 2 2010

Re: K092937

Trade/Device Name: STA® - Hybrid Hep Calibrator Regulation Number: 21 CFR 864.7525 Regulation Name: Heparin Assay Regulatory Class: Class II Product Code: KFF, JIT Dated: July 21, 2010 Received: July 22, 2010

Dear Mr. Parrotta:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into class II (Special Controls), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820). This letter will allow you to begin marketing your device as described in your Section 510(k) premarket

{4}------------------------------------------------

Page 2 - Mr. Umberto Parrotta

notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Parts 801 and 809), please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (301) 796-5450. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office

of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html

Sincerely yours.

m. chan

Maria M. Chan, Ph.D. Director Division of Immunology and Hematology Devices Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

Indications for Use

AUG 0 2 2010

510(k) Number (if known): 《092937

Device Name: STA®- Hybrid Hep Calibrator.

Indication for Use:

The STA® - Hybrid Hep Calibrator is a set of calibrator plasmas intended for use with analyzers of the STA® line suitable to these reagents, for the calibration of heparin (UFH and LMWH) assay by measuring the anti-Xa activity using the chromogenic method, STA® - Rotachrom® Heparin.

Prescription Use X (21 CFR Part 801 Subpart D)

And/Or

Over the Counter Use (21 CFR Part 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE; CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of In Vitro Diagnostic Device Evaluation and Safety (OIVD)

Lenck R. Si.

Division Sign-Off Office of In Vitro Diagnostic Device Evaluation and Safety 510(k) K092937

DSI-STA-HHC-Attachment-01-IndicationsForUseStatementForm.doc

Page 1 of 1

§ 862.1150 Calibrator.

(a)
Identification. A calibrator is a device intended for medical purposes for use in a test system to establish points of reference that are used in the determination of values in the measurement of substances in human specimens. (See also § 862.2 in this part.)(b)
Classification. Class II (special controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9.