K Number
K092570
Device Name
ABX PENTRA TPU CAL, MODEL A11A01898
Manufacturer
Date Cleared
2009-10-16

(56 days)

Product Code
Regulation Number
862.1150
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP Authorized
Intended Use
The ABX PENTRA Urine Cal is a calibrator for use in the calibration of the quantitative method : ABX PENTRA Urinary Proteins CP on ABX PENTRA 400 clinical chemistry analyzer as specified on the vials.
Device Description
The calibrator included in this submission is for use on the ABX PENTRA 400 (K052007), which is a discrete photometric benchtop clinical chemistry analyzer. The ABX PENTRA TPU Cal is a liquid calibrator based on a buffered aqueous solution, containing human serum. The assigned values of the calibrator components are given on the calibrator vials, ensuring optimal calibration of the appropriate HORIBA ABX methods on the ABX PENTRA 400 analyzer. This calibrator is provided in three vials of 3 ml.
More Information

No
The document describes a calibrator for a clinical chemistry analyzer, which is a standard laboratory reagent. There is no mention of AI or ML in the intended use, device description, or performance studies.

No.
The device is a calibrator for a clinical chemistry analyzer, used for diagnostic purposes, not for treating or preventing disease.

No

The device, ABX PENTRA Urine Cal, is described as a calibrator used to ensure the accuracy of a quantitative method on a clinical chemistry analyzer (ABX PENTRA 400). It does not directly diagnose a condition but rather helps in the proper functioning of a diagnostic instrument.

No

The device description clearly states it is a liquid calibrator based on a buffered aqueous solution containing human serum, which is a physical substance, not software.

Yes, this device is an IVD (In Vitro Diagnostic).

Here's why:

  • Intended Use: The intended use explicitly states it's a "calibrator for use in the calibration of the quantitative method: ABX PENTRA Urinary Proteins CP on ABX PENTRA 400 clinical chemistry analyzer". Calibrators are essential components used in in vitro diagnostic testing to ensure the accuracy of the results.
  • Device Description: It describes a liquid calibrator used on a "clinical chemistry analyzer". Clinical chemistry analyzers are devices used to perform in vitro diagnostic tests on biological samples.
  • Context: The entire description revolves around its use in a laboratory setting for calibrating a method that measures substances in urine (urinary proteins), which is a common type of in vitro diagnostic test.

The fact that it's a calibrator for a clinical chemistry analyzer used to measure substances in a biological sample (urine) firmly places it within the definition of an In Vitro Diagnostic device.

N/A

Intended Use / Indications for Use

ABX Pentra TPU Cal is used to calibrate total proteins in urine measurement with reagent ABX Pentra Urinary Proteins CP on ABX PENTRA 400 analyzer.

The ABX PENTRA Urine Cal is a calibrator for use in the calibration of the quantitative method : ABX PENTRA Urinary Proteins CP on ABX PENTRA 400 clinical chemistry analyzer as specified on the vials.

Product codes

JIT

Device Description

The calibrator included in this submission is for use on the ABX PENTRA 400 (K052007), which is a discrete photometric benchtop clinical chemistry analyzer.

The ABX PENTRA TPU Cal is a liquid calibrator based on a buffered aqueous solution, containing human serum. The assigned values of the calibrator components are given on the calibrator vials, ensuring optimal calibration of the appropriate HORIBA ABX methods on the ABX PENTRA 400 analyzer. This calibrator is provided in three vials of 3 ml.

Mentions image processing

Not Found

Mentions AI, DNN, or ML

Not Found

Input Imaging Modality

Not Found

Anatomical Site

Not Found

Indicated Patient Age Range

Not Found

Intended User / Care Setting

Not Found

Description of the training set, sample size, data source, and annotation protocol

Not Found

Description of the test set, sample size, data source, and annotation protocol

Not Found

Summary of Performance Studies

The performance testing data conclude that the safety and effectiveness of the device are not compromised, and that they met all acceptance criteria, demonstrating that the device is substantially equivalent to the predicate device.

Key Metrics

Not Found

Predicate Device(s)

K071779

Reference Device(s)

K052007

Predetermined Change Control Plan (PCCP) - All Relevant Information

Not Found

§ 862.1150 Calibrator.

(a)
Identification. A calibrator is a device intended for medical purposes for use in a test system to establish points of reference that are used in the determination of values in the measurement of substances in human specimens. (See also § 862.2 in this part.)(b)
Classification. Class II (special controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9.

0

Premarket Notification [510(k)] Summary

OCT 1 6 2009

This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92

The assigned 510(k) number is : 1092570

Company: Horiba ABX SAS Parc Euromédecine Rue du Caducée - BP 7290 34184 Montpellier cedex 4 FRANCE + (33) 4 67 14 18 43 Telephone: Fax: + (33) 4 67 14 18 75

Contact Person: Caroline Ferrer (caroline.ferrer(@horiba.com)

Date Prepared: 11th August 2009

Device Name:

.

The following calibrator is for use in conjunction with the ABX PENTRA 400, cleared to market under K052007.

CALIBRATOR:

Trade/Proprictary Name: Common or Usual Name: Device Class Classification Name: Product Code:

ABX PENTRA TPU Cal Urine calibrator Class II $862.1150 : Calibrator JIT ; Calibrator, Secondary

Substantial Equivalence:

The data and information supplied in this submission demonstrates substantial equivalence to the predicate device:

| Submission device | Substantially equivalent
Predicate device |
|--------------------|----------------------------------------------|
| ABX PENTRA TPU Cal | ABX PENTRA Urine Cal |
| Horiba ABX SAS | Horiba ABX SAS
K071779 |

1

Description:

The calibrator included in this submission is for use on the ABX PENTRA 400 (K052007), which is a discrete photometric benchtop clinical chemistry analyzer.

The ABX PENTRA TPU Cal is a liquid calibrator based on a buffered aqueous solution, containing human serum. The assigned values of the calibrator components are given on the calibrator vials, ensuring optimal calibration of the appropriate HORIBA ABX methods on the ABX PENTRA 400 analyzer. This calibrator is provided in three vials of 3 ml.

Intended Use:

ABX Pentra TPU Cal is used to calibrate total proteins in urine measurement with reagent ABX Pentra Urinary Proteins CP on ABX PENTRA 400 analyzer.

CALIBRATOR

ABX PENTRA TPU Cal:
AnalytesTotal Proteins in urine
FormatLiquid with chemical additives and materials of biological
origin
StabilityClosed stability: 24 months at 2-8°C
Open stability: 9 weeks at 2-8°C

Conclusions for Performance Testing :

The performance testing data conclude that the safety and effectiveness of the device are not compromised, and that they met all acceptance criteria, demonstrating that the device is substantially equivalent to the predicate device.

p.2/2

2

Image /page/2/Picture/0 description: The image is a seal for the Department of Health & Human Services (HHS). The seal is circular and contains the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. In the center of the seal is an abstract image of an eagle.

DEPARTMENT OF HEALTH & HUMAN SERVICES

Food and Drug Administration 10903 New Hampshire Avenue Document Mail Center - WO66-0609 Silver Spring, MD 20993-0002

Horiba ABX SAS, Inc. c/o Ms. Caroline Ferrer Regulatory Affairs Specialist Parc Euromedecine Rue Du Caducee, BP 7290 34184 Montpellier Cedex 4. France

OCT 1 6 2009

Re: K092570

Trade Name: ABX PENTRA TPU Cal Regulation Number: 21 CFR §862.1150 Regulation Name: Calibrator Regulatory Class: Class II Product Codes: JIT Dated: August 18, 2009 Received: August 21, 2009

Dear Ms. Ferrer:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820).

3

Page 2

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (301) 796-5450. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at (301) 796-5760. For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or ( 301 ) 796-5680 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Courtney C. Harner, Ph.D. Director Division of Chemistry and Toxicology Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health

Enclosure

4

Indication for Use

510(k) Number (if known): K0935 70

Device Name: ABX PENTRA TPU Cal

Indication For Use:

The ABX PENTRA Urine Cal is a calibrator for use in the calibration of the quantitative method : ABX PENTRA Urinary Proteins CP on ABX PENTRA 400 clinical chemistry analyzer as specified on the vials.

Prescription Use X (21 CFR Part 801 Subpart D)

And/Or

Over the Counter Use _ (21 CFR Part 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE; CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of In Vitro Diagnostic Device Evaluation and Safety (OVD)

Carol Bens

Division Sign-Off Office of In Vitro Diagnostic Device Evaluation and Safety

510(k) K 092570

Page 1 of 1