K Number
K092566
Manufacturer
Date Cleared
2009-10-09

(50 days)

Product Code
Regulation Number
866.3980
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Verigene® Respiratory Virus Nucleic Acid Test on the Verigene SP System (RVNATsg) is a qualitative multiplex in vitro diagnostic test for the detection and identification of Influenza A Virus, Influenza B Virus, and Respiratory Syncytial Virus (RSV) nucleic acids purified from nasopharyngeal swab specimens obtained from patients symptomatic for viral upper respiratory infection. The test is intended to be used on the Verigene® SP System as an aid in the differential diagnosis of Influenza A, Influenza B, and RSV infections. The test is not intended to detect Influenza C virus.

Negative results do not preclude influenza virus or RSV infection and should not be used as the sole basis for treatment or other management decisions. It is recommended that negative test results be confirmed by culture.

Performance characteristics for Influenza A Virus were established when Influenza A/H3 and A/H1 were the predominant Influenza A viruses in circulation. When other Influenza A viruses are emerging, performance characteristics may vary.

If infection with a novel Influenza A virus is suspected based on current clinical and epidemiological screening criteria recommended by public health authorities, specimens should be collected with appropriate infection control precautions for novel virulent Influenza viruses and sent to state or local health department for testing. Viral culture should not be attempted in these cases unless a BSL 3+ facility is available to receive and culture specimens.

Device Description

The entire RVNATSP is performed on the Verigene® SP System, which is a bench-top molecular diagnostics workstation that consists of two instruments, the Verigene SP Processor and the Verigene Reader. The Verigene SP Processor performs the assay steps on each sample by using a robotic pipettor to transfer and mix reagents within and between separate testing modules designed for nucleic acid extraction, target amplification, and the Verigene Hybridization test. The Verigene hybridization test module is the same as in the previous Verigene System with added modules for nucleic acid extraction and RT-PCR target amplification. Key functions of the Verigene SP Processor include:

  1. Reading of the barcode identification label on inserted Test Consumables to maintain positive identification of patient samples throughout processing.
  2. Facilitation of nucleic acid extraction, multiplex RT-PCR target amplification, and the Verigene Hybridization Test.
  3. Real-time communication of test processing status to the Reader.

The Verigene Reader is the same instrument as in the FDA-cleared VRNAT. It is a free-standing instrument with a touch screen control panel and a wand-based barcode scanner. It utilizes a graphical user interface to guide the user through the process of ordering tests and reporting results. There are no serviceable parts and no user calibration is required. Interaction with the touch screen is minimized through barcode use. This instrument also serves as the reader of the Test Cartridges using advanced optics. The key functions of the Verigene Reader include:

  1. Entry and tracking of specimen identification numbers via manual keyboard input or via barcode-reader wand.
  2. Test selection for each specimen.
  3. Automated transfer of specimen processing instructions on Test Cartridge-specific basis to linked Processor unit(s). A single Reader unit can control up to 32 Processor units.
  4. Automated imaging and analysis of Test Cartridges.
  5. Results display.
  6. Results report generation.

RVNATSP consumables within each single-use disposable test kit include: (i) Tip Holder Assembly; (ii) Extraction Tray; (iii) Amplification Tray; and (iv) RV Test Cartridge. The kit components are inserted into the corresponding module of the Verigene SP Processor prior to each test, and the sample is added to the Extraction Tray. Patient information is entered into the Reader to initiate the test procedure.

  1. Tip Holder Assembly – The robotic pipettor picks up pipettes from the Tip Holder Assembly. The pipettes are used for mixing and transferring reagents within the test procedure.
  2. Extraction Tray – Nucleic acids are extracted from the sample by using magnetic bead-based methods within the Extraction Tray. Each Tray contains reagents for a single extraction procedure. A robotic pipette transfers reagents to designated wells within the Extraction Tray to affect the steps of lysis, capture of nucleic acids onto the magnetic beads, washing, and eluting the isolated nucleic acids from the magnetic beads.
  3. Amplification Tray – The isolated nucleic acids are amplified by using multiplex RT-PCR within the Amplification Tray. Each Tray contains reagents for a single multiplex RT-PCR procedure. A robotic pipette transfers the reagents to a specific well within the Amplification Tray. A set thermal profile is then initiated to perform all of the amplification related steps including UDG-based decontamination, reverse transcription, and multiplex PCR in a single tube. Upon completion, an aliquot of the amplified sample is mixed with hybridization buffer containing the virus specific mediator probes. The sample is then transferred to the Test Cartridge.
  4. RV Test Cartridge for Verigene Hybridization Test - The virus-specific amplicons are detected and identified within a Test Cartridge by using specific nucleic acid probes in conjunction with gold nanoparticle probe-based detection technology. Each Test Cartridge is a self-contained, laboratory consumable that consists of two parts. The upper housing of each cartridge is called the "reagent pack" and contains reservoirs filled with the detection reagents. When in place with the 'substrate holder', the reagent pack creates an air-tight hybridization chamber surrounding the region of the substrate containing a target-specific capture array. As each step of the test is completed, old reagents are moved out of the hybridization chamber and new reagents are added from the reagent pack via microfluidic channels and pumps. Once the test is complete, the Test Cartridge is removed from the Verigene SP Processor unit and the reagent pack is snapped off and discarded. The remaining slide is now ready for imaging and analysis in the Verigene Reader.
  5. End-point detection on the Verigene Reader: The test slide is inserted into the Verigene Reader wherein it is illuminated along its side. The gold-silver aggregates at the test sites scatter the light, which is in turn captured by a photosensor. The relative intensity arising from each arrayed test site is tabulated. Net signals, defined as the absolute signal intensities with background signals subfracted, are compared with thresholds determined by negative controls within the slide in order to arrive at a decision regarding the presence or absence of target. These results are linked to the test and patient information entered at the beginning of each test session to provide a comprehensive results file.
AI/ML Overview

The provided document describes the Verigene® Respiratory Virus Nucleic Acid Test on the Verigene® SP System (RVNATSP), a qualitative multiplex in vitro diagnostic test for the detection and identification of Influenza A Virus, Influenza B Virus, and Respiratory Syncytial Virus (RSV) nucleic acids. The study aims to demonstrate substantial equivalence to a previously cleared predicate device, the Verigene® Respiratory Virus Nucleic Acid Test (VRNAT, K083088).

1. Table of Acceptance Criteria and Reported Device Performance:

The primary acceptance criteria for the RVNATSP were established by demonstrating equivalence to the predicate VRNAT device (K083088) in terms of analytical sensitivity (Limit of Detection), lack of carryover/crossover contamination, precision/reproducibility, and method comparison (percent agreement). Based on the provided information, the acceptance criteria are implicitly that the performance of the RVNATSP should be comparable or identical to that of the cleared VRNAT.

Acceptance Criteria CategorySpecific MetricAcceptance Threshold (Implicit)RVNATSP Performance (Reported)
Analytical SensitivityLimit of Detection (LOD) for Influenza B, RSV A, RSV B, and Influenza A.Identical to the LOD of the cleared VRNAT (K083088).Identical to the LOD observed with the same strains on the cleared VRNAT (K083088) for Influenza B (50 TCID50/mL), RSV A (10 TCID50/mL), RSV B (2 TCID50/mL), and Influenza A (2 TCID50/mL).
ContaminationEvidence of carryover and/or crossover contamination.No evidence of cross-contamination.No evidence of cross-contamination from any test steps (sample extraction, multiplex RT-PCR, Verigene Hybridization Test).
Precision/ReproducibilityAgreement of 'Observed Results' to 'Expected Results' across High Negative, Low Positive, and Moderate Positive samples for Influenza A, Influenza B, RSV A, and RSV B. Combined % Agreement and 95% Confidence Interval.Clinically and statistically equivalent to the cleared VRNAT.Clinically and statistically equivalent to the cleared VRNAT. For all panel members, the RVNATSP showed high agreement (e.g., 100% for most moderate positive and high negative samples, 98.7% for low positives, 99% for RSV A High Negative) with comparable 95% CIs to the VRNAT.
Method ComparisonPositive Percent Agreement (PPA) and Negative Percent Agreement (NPA) between RVNATSP and cleared VRNAT for all viruses combined.Collective lower bound 95% CI for PPA and NPA greater than 90%. Clinically and statistically equivalent.PPA: 97.9% (95% CI: 94.8% - 99.0%) (all sites combined). NPA: 100.0% (95% CI: 99.0% - 100.0%) (all sites combined). All reported discordant samples were low-positive Influenza A, which were positive on repeat testing.

2. Sample Size Used for the Test Set and Data Provenance:

  • Analytical Sensitivity (LOD) Study: For each virus strain, 20 replicates were tested at the LOD concentration after an initial quadruplicate testing of serially diluted stocks. This was done for Influenza B, RSV A, RSV B, and Influenza A. The samples were prepared from strains with established titers serially diluted into Universal Transport Media (Copan). The country of origin of the data is not explicitly stated but implies internal laboratory studies. The data is prospective for the purpose of this study.
  • Carryover and Crossover Contamination Studies: High positive samples of Influenza A, Influenza B, and RSV B were alternated with high negative samples. The exact number of samples or alternations is not specified but the "collective data" demonstrated no cross-contamination. This appears to be a prospective internal study.
  • Precision/Reproducibility Studies: Eight unique sample panels (viral strains at high negative, low positive, and moderate positive concentrations) were tested.
    • Site 1 (Precision Study): The sample set was tested over 12 non-consecutive days with two operators performing the test in duplicate (total of 48 tests per panel member for Influenza A, Influenza B, RSV A, RSV B).
    • Sites 2 and 3 (Reproducibility Study): The sample set was tested in triplicate daily by 2 operators on each of five non-consecutive days (total of 15 tests per panel member per site for Influenza A, Influenza B, RSV A, RSV B).
    • The total number of individual results used for precision/reproducibility comparison across all sites for each panel member was 78 (48 from Site 1 + 15 from Site 2 + 15 from Site 3). This was a prospective, multi-site study. Data provenance is not specified beyond "three sites."
  • Method Comparison Studies: A total of 62 unique samples (from culture positive and negative nasopharyngeal swab samples) were prepared, representing Influenza A (15 positive, 47 negative), Influenza B (16 positive, 46 negative), and RSV A/B (34 positive, 28 negative). These were diluted to yield viral load levels close to low positive. Each sample provided a decision for all three viruses, resulting in 186 decisions per test.
    • The sample set was tested at one internal site (Site 1) using the cleared VRNAT.
    • The same sample set was tested at all three sites (Site 1, Site 2, and Site 3) using the RVNATSP.
    • "Total of 62x4 = 248 unique tests" (implying 62 samples tested once on VRNAT at Site 1, and once on RVNATSP at each of the 3 sites). The comparison data in Tables 7-9 use the VRNAT (Old System) results (presumably from Site 1) as the reference for each RVNATSP site's results. The combined data in Table 10 then aggregates results from all three RVNATSP sites against the VRNAT reference.
    • Data provenance is not explicitly stated, but "culture positive nasopharyngeal swab samples" suggests patient samples. This was a prospective, multi-site study.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications:

This device is a molecular diagnostic test. The "ground truth" for the test set is established by the known concentration of viral strains (for analytical sensitivity/LOD and precision/reproducibility) or by the results from the predicate device (cleared VRNAT) for the method comparison study. There are no human "experts" required to establish a ground truth in the way a radiologist would for an imaging study. The reference method for comparison is a previously cleared, similar molecular diagnostic test.

4. Adjudication Method for the Test Set:

Not applicable in the context of this molecular diagnostic device comparison. The "ground truth" for method comparison is the result obtained from the predicate device (cleared VRNAT). For discordant results in the method comparison study (e.g., in Site 2 and Site 3 performance for Influenza A), "Repeat tests were positive and gave the expected result," suggesting internal re-testing or confirmation processes were employed for discrepancies, but not necessarily a formal adjudication by a panel of human experts.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

Not applicable. This is a molecular diagnostic device, not an imaging device requiring human readers or AI assistance in interpretation. The output is a qualitative "Detected" or "Not Detected" result for specific viral nucleic acids.

6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) was done:

The device (RVNATSP) operates as a standalone system once the sample is loaded. The "algorithm" in this context refers to the molecular diagnostic assay steps and the instrument's internal decision algorithm for generating "Detected" or "Not Detected" results. The performance studies described (analytical sensitivity, contamination, precision, method comparison) inherently evaluate the standalone performance of the RVNATSP system without human intervention in the result interpretation after the run is initiated. The comparison against the predicate device also assesses standalone performance.

7. The Type of Ground Truth Used (expert consensus, pathology, outcomes data, etc.):

  • For Analytical Sensitivity (LOD) and Precision/Reproducibility: The ground truth was based on known concentrations of specific viral strains (e.g., TCID50/mL values) that were serially diluted into a sample matrix.
  • For Method Comparison: The ground truth was established by the results obtained from the FDA-cleared predicate device (VRNAT, K083088). This is a "device-to-device" comparison where the cleared device serves as the reference standard.

8. The Sample Size for the Training Set:

This document describes a 510(k) submission for a molecular diagnostic device, not a machine learning or AI-based system that typically uses a "training set." The RVNATSP is built upon established molecular biology principles and reagents. Therefore, the concept of a "training set" for an algorithm, as described in AI development, is not directly applicable to this device. The development of such assays involves extensive R&D and analytical optimization, which could be considered an iterative "training" process for the assay design itself, but not a distinct dataset for algorithm training in the computational sense.

9. How the Ground Truth for the Training Set Was Established:

As noted above, the concept of a "training set" with established ground truth as understood in AI/ML development does not directly apply to this molecular diagnostic device. The assay design and optimization would rely on established molecular biology techniques, purified viral samples, and potentially a range of clinical samples, but these are part of the assay development and validation rather than a formally defined "training set" for an algorithm.

{0}------------------------------------------------

K092566

:

Image /page/0/Picture/1 description: The image shows the word "Nanosphere" in a sans-serif font. To the left of the word is a circular logo that is divided into two halves. The left half of the logo is filled with a pattern of small dots, while the right half is filled with a pattern of horizontal lines.

OCT - 9 2009

Summary of 510(k)

510(k)numbers:K092566: Verigene® Respiratory Virus Nucleic Acid Test on the Verigene® SP System (RVNATSP)
Summarypreparationdate:October 7, 2009
Submittedby:Nanosphere, Inc.4088 Commercial AvenueNorthbrook, IL 60062Phone: 847-400-9000 Fax: 847-400-9199
Contact:Gregory W. Shipp, M.D.Chief Medical OfficerVP, Medical and Regulatory Affairs and Quality Assurance
Proprietarynames:For instrument:Verigene® SP SystemFor the assay:Verigene® Respiratory Virus Nucleic Acid Test on the Verigene® SP System (RVNATSP)
Commonnames:For the instrument:Bench-top molecular diagnostics workstationFor the assays:Respiratory panelRespiratory virus panelRespiratory virusesInfluenza A assayInfluenza B assayRSV assayInfluenza A/B and RSV assay
ComparisonVRNAT andRVNATSP:Additional claim to the 510(k) application, K083088 is described. The Verigene Respiratory Virus NucleicAcid Test (VRNAT) will add the Verigene SP System to the labeling as a cleared system. The safety andeffectiveness of the Verigene® Respiratory Virus Nucleic Acid Test on the Verigene SP System(RVNATSP) is demonstrated with analytical and method comparison studies.The FDA-cleared VRNAT (K083088) is a qualitative test based on identifying virus-specific nucleic acidsfor Influenza A virus, Influenza B virus, and respiratory syncytial virus (RSV). Within this test, the steps ofsample preparation (or nucleic acid extraction), target amplification, and hybridization test and analysistake place on three separate instruments (Table 1). The RVNATSP is the same assay with additionalsystem claims for the Verigene SP System, a modified Verigene® System that allows sample preparation,target amplification, and hybridization test and analysis using a single system (Table 1).

{1}------------------------------------------------

Table 1.
Sample PreparationTarget AmplificationHybridization Test and Analysis
VRNAT(cleared, K083088)NucliSENS EasyMAG(bioMerieux)ThermocyclerVerigene System
RVNATSPVerigene SP SystemVerigene SP SystemVerigene SP System
The RVNATSP is designed to identify virus-specific nucleic acids for Influenza A virus, Influenza B virus, and respiratory syncytial virus (RSV). The RVNATSP involves:
1)Sample Preparation – magnetic bead-based viral RNA extraction from nasopharyngeal swab specimens obtained from symptomatic patients;
2)Target Amplification – Multiplex RT-PCR-based amplification of the eluted viral RNA targets to generate virus-specific amplicons;
3)Verigene Hybridization Test and Analysis - detection and identification of virus-specific amplicons by using gold nanoparticle probe-based technology.
The entire RVNATSP is performed on the Verigene® SP System, which is a bench-top molecular diagnostics workstation that consists of two instruments, the Verigene SP Processor and the Verigene Reader. The Verigene SP Processor performs the assay steps on each sample by using a robotic pipettor to transfer and mix reagents within and between separate testing modules designed for nucleic acid extraction, target amplification, and the Verigene Hybridization test. The Verigene hybridization test module is the same as in the previous Verigene System with added modules for nucleic acid extraction and RT-PCR target amplification. Key functions of the Verigene SP Processor include:
1)Reading of the barcode identification label on inserted Test Consumables to maintain positive identification of patient samples throughout processing.
2)Facilitation of nucleic acid extraction, multiplex RT-PCR target amplification, and the Verigene Hybridization Test.
Device description:3)Real-time communication of test processing status to the Reader.
The Verigene Reader is the same instrument as in the FDA-cleared VRNAT. It is a free-standing instrument with a touch screen control panel and a wand-based barcode scanner. It utilizes a graphical user interface to guide the user through the process of ordering tests and reporting results. There are no serviceable parts and no user calibration is required. Interaction with the touch screen is minimized through barcode use. This instrument also serves as the reader of the Test Cartridges using advanced optics. The key functions of the Verigene Reader include:
1)Entry and tracking of specimen identification numbers via manual keyboard input or via barcode-reader wand.
2)Test selection for each specimen.
3)Automated transfer of specimen processing instructions on Test Cartridge-specific basis to linked Processor unit(s). A single Reader unit can control up to 32 Processor units.
4)Automated imaging and analysis of Test Cartridges.
5)Results display.
6)Results report generation.
RVNATSP consumables within each single-use disposable test kit include: (i) Tip Holder Assembly; (ii) Extraction Tray; (iii) Amplification Tray; and (iv) RV Test Cartridge. The kit components are inserted into the corresponding module of the Verigene SP Processor prior to each test, and the sample is added to the Extraction Tray. Patient information is entered into the Reader to initiate the test procedure.
1)Tip Holder Assembly – The robotic pipettor picks up pipettes from the Tip Holder Assembly. The pipettes are used for mixing and transferring reagents within the test procedure.

.

{2}------------------------------------------------

  1. Extraction Tray – Nucleic acids are extracted from the sample by using magnetic bead-based methods within the Extraction Tray. Each Tray contains reagents for a single extraction procedure. A robotic pipette transfers reagents to designated wells within the Extraction Tray to affect the steps of lysis, capture of nucleic acids onto the magnetic beads, washing, and eluting the isolated nucleic acids from the magnetic beads. 3) Amplification Tray – The isolated nucleic acids are amplified by using multiplex RT-PCR within the Amplification Tray. Each Tray contains reagents for a single multiplex RT-PCR procedure. A robotic pipette transfers the reagents to a specific well within the Amplification Tray. A set thermal profile is then initiated to perform all of the amplification related steps including UDG-based decontamination, reverse transcription, and multiplex PCR in a single tube. Upon completion, an aliquot of the amplified sample is mixed with hybridization buffer containing the virus specific mediator probes. The sample is then transferred to the Test Cartridge. 4) RV Test Cartridge for Verigene Hybridization Test - The virus-specific amplicons are detected and identified within a Test Cartridge by using specific nucleic acid probes in conjunction with gold nanoparticle probe-based detection technology. Each Test Cartridge is a self-contained, laboratory consumable that consists of two parts. The upper housing of each cartridge is called the "reagent pack" and contains reservoirs filled with the detection reagents. When in place with the 'substrate holder', the reagent pack creates an air-tight hybridization chamber surrounding the region of the substrate containing a target-specific capture array. As each step of the test is completed, old reagents are moved out of the hybridization chamber and new reagents are added from the reagent pack via microfluidic channels and pumps. Once the test is complete, the Test Cartridge is removed from the Verigene SP Processor unit and the reagent pack is snapped off and discarded. The remaining slide is now ready for imaging and analysis in the Verigene Reader. 5) End-point detection on the Verigene Reader: The test slide is inserted into the Verigene Reader wherein it is illuminated along its side. The gold-silver aggregates at the test sites scatter the light, which is in turn captured by a photosensor. The relative intensity arising from each arrayed test site is tabulated. Net signals, defined as the absolute signal intensities with background signals subfracted, are compared with thresholds determined by negative controls within the slide in order to arrive at a decision regarding the presence or absence of target. These results are linked to the test and patient information entered at the beginning of each test session to provide a comprehensive results file. The RVNAT so is a qualitative multiplex in vitro diagnostic test for the detection and identification of Influenza A Virus, Influenza B Virus, and Respiratory Syncytial Virus (RSV) nucleic acids purified from nasopharyngeal swab specimens obtained from patients symptomatic for viral upper respiratory infection. The test is intended to be used on the Verigene® SP System as an aid in the differential diagnosis of Influenza A, Influenza B, and RSV infections. The test is not intended to detect Influenza C virus. Neqative results do not preclude influenza virus or RSV infection and should not be used as the sole basis for treatment or other management decisions. It is recommended that negative test results be confirmed by culture. Intended uses: Performance characteristics for Influenza A Virus were established when Influenza A/H3 and A/H1 were the predominant Influenza A viruses in circulation. When other Influenza A viruses are emerging, performance characteristics may vary. If infections with a novel Influenza A virus is suspected based on current clinical and epidemiological screening criteria recommended by public health authorities, specimens should be collected with appropriate infection control precautions for novel virulent influenza viruses and sent to state or local health department for testing. Viral culture should not be attempted in these cases unless a BSL 3+ facility is available to receive and culture specimens. The cleared Verigene® Respiratory Virus Nucleic Acid Test (VRNAT), K083088 is claimed as the predicate Predicate device (Table 2). It is a multiplex in vitro diagnostic test for the rapid and qualifative detection and device: discrimination of Influenza A Virus, Influenza B Virus, and Respiratory Syncytial Virus (RSV).

{3}------------------------------------------------

FeatureVRNAT (Predicate, K083088)Verigene RVNATSP
Intended useThe Verigene® Respiratory Virus Nucleic Acid Teston Verigene SP System (RVNAT gp) is a qualitativemultiplex in vitro diagnostic test for the detection andidentification of Influenza A Virus, Influenza B Virus,and Respiratory Syncytial Virus (RSV) nucleic acidspurified from nasopharyngeal swab specimensobtained from patients symptomatic for viral upperrespiratory infection. The test is intended to be usedon the Verigene® SP System as an aid in thedifferential diagnosis of Influenza A, Influenza B, andRSV infections. The test is not intended to detectInfluenza C virus.Negative results do not preclude influenza virus orRSV infection and should not be used as the solebasis for treatment or other management decisions. Itis recommended that negative results be confirmed byculture.Performance characteristics for Influenza A Virus wereestablished when Influenza A/H3 and A/H1 were thepredominant Influenza A viruses in circulation. Whenother Influenza A viruses are emerging, performancecharacteristics may vary.If infection with a novel Influenza A virus is suspectedbased on current clinical and epidemiologicalscreening criteria recommended by public healthauthorities, specimens should be collected withappropriate infection control precautions for novelvirulent influenza viruses and sent to state or localhealth department for testing. Viral culture should notbe attempted in these cases unless a BSL 3+ facility isavailable to receive and culture specimens.Same test on Verigene SP System
l argetsIntluenza A, Influenza B, RSVldentical
SpecimenNasopharyngeal swabs in sample matrixIdentical
SamplepreparationAutomated extraction of nucleic acids performedexternally on the NucliSENS EasyMAG (bioMerieux)using silica coated magnetic beads and chaotropicsalts using chemistry based on U.S. Patent 5,234,809.Automated extraction of nucleic acidsperformed on the Verigene SP Processorusing silica coated magnetic beads andchaotropic salts licensing the identicalpatent chemistry.
Sample size200 µLIdentical
QualitycontrolInternal procedural quality control, external qualitycontrol solutionsIdentical
AmplificationmethodMultiplex RT-PCR: Target Amplification by RT-PCR isperformed on an external thermocyclerMultiplex RT-PCR: Identical reagents;performed within the added AmplificationModule on the Verigene SP Processor
M-MLV Reverse TranscriptaseIdentical
Pipetting
PipettingPipetting between the three steps is performed by the user.Robotic pipettor added to automate fluid transfer steps.
DetectionMethodVerigene Hybridization Test is performed in the hybridization module housed in the Processor of the Verigene® System by using single-use Test CartridgesIdentical
DecisionalgorithmTarget-specific signal intensities are compared to a signal threshold and ratioed against positive and negative controls for a decision.Identical
ResultsPositive or negative qualitative resultsIdentical
ReaderProvides the user interface, controls the Processor, performs image analysis, and provides results.Identical
SoftwareA custom embedded software application running under the Micro-C/OS real-time operating system.Architecture is the same. Additional software programming to control the Extraction and Amplification Modules.
ReagentstorageTest Cartridge: 2 - 8 °CAmplification Kit: -20 °CTest Cartridge and Extraction Tray: 2 - 8 °C.Amplification Tray: -20 °C
AssayPerformanceAs in the cleared VRNAT (K083088).Identical Limits of Detection as the cleared VRNAT.Precision/Reproducibility - Clinically and statistically equivalent.Method Comparison - Clinically and statistically equivalent

Table 2. Similarities and Differences between the Cleared and the New System.

.

·

{4}------------------------------------------------

.

. .

{5}------------------------------------------------

Performance Characteristics of the RVNATse

Analytical and method comparison studies to establish the performance of the test on the Verigene SP System are described.

  • A. Comparison of Analytical Sensitivity
    The analytical sensitivity of the RVNAT o was compared to the cleared VRNAT (K083088) by determining the Limit of Detection (LOD) of Influenza B, RSV A, and RSV B viruses. Strains with established titers were used for each virus stock was serially diluted into a sample matrix (Universal Transport Media, Copan), and each concentration was tested in quadruplicate using the RVNAT ge. The LOD was confirmed by performing an additional 20 replicates for each strain in order to demonstrate that the virus was detected ≥95% of the time. The LOD for each virus was identical to the LOD observed with the same strains on the cleared VRNAT (K083088) (Table 3).
Limits of DetectionConcentration
Influenza A (A/Wisconsin/67/05)2 TCID50/mL
Influenza B (B/Florida/04/2006)50 TCID50/mL
RSV A (Strain Long)10 TCID50/mL
RSV B (B-1 Wild Type (B WV/14617/85))2 TCID50/mL
Table 3. Limit of Detection
  • B. Carryover and Crossover Contamination Studies
    High positive samples of Influenza A, Influenza B, RSV B were alternated with high neqative samples for all four viruses. Based on the collective data, there was no evidence of crosscontamination from any of the test steps including sample extraction, multiplex RT-PCR step, and the Verigene Hybridization Test.

  • C. Precision/Reproducibility Studies Comparison between the RVNATso and the cleared VRNAT (K083088)
    The Precision/Reproducibility Studies were performed at each of three sites. At Site 1, the Reproducibility Study was part of a larger precision/Reproducibility Studies for the RVNAT go were conducted exactly as for the previously cleared VRNAT (K083088) to allow equivalency comparisons. As before, eight unique samples were created by diluting known concentrations of viral particles with Viral Transport Medium (Table 4). Since the Analytical Sensitivity of the RVNATs was identical to the cleared VRNAT, the same strains and levels were used in the studies. Each strain was represented at 3 distinct concentrations: high negative, low positive, and moderate positive.

{6}------------------------------------------------

Unique SamplesViral Strains and Levels
1Influenza A - High Negative; Influenza B - High Negative
2RSV A - High Negative; RSV B - High Negative
3Influenza A - Low Positive
4Influenza B - Low Positive
5RSV A - Low Positive
6RSV B - Low Positive
7Influenza A - Moderate Positive; RSV A - Moderate Positive
8Influenza B - Moderate Positive; RSV B - Moderate Positive
Table 4. Sample panel for the Precision/Reproducibility Studies.
----------------------------------------------------------------------------

The Precision Study (Site 1) tested the sample set over 12 non-consecutive days. On each test day, two operators performed the RVNAT و in duplicate for each sample sets per day total). In the reproducibility study performed by sites 2 and 3, the sample set was tested in triplicate daily by 2 operators on each of five non-consecutive days.

Performance characteristics of the RVNAT go in the Precision/Reproducibility Studies were equivalent to those for the cleared VRNAT (Table 5).

Agreement of 'Observed Results' to 'Expected Results'
New RVNATSPCleared VRNAT
Panel MemberSite 1Site 2Site 3AllSites%Agreement95% ScoreCISite 1Site 2Site 3AllSites%Agreement95% ScoreCI
Influenza AHighNegative48/4815/1515/1578/78100%95.3 - 100%46/4814/1515/1575/7896%89.3 - 98.7%
LowPositive48/4815/1515/1578/78100%95.3 - 100%48/4814/1515/1577/7898.7%93.1 - 99.8%
ModeratePositive48/4815/1515/1578/78100%95.3 - 100%48/4815/1515/1578/78100.0%95.3 - 100%
Influenza BHighNegative48/4815/1515/1578/78100%95.3 - 100%48/4815/1515/1578/78100%95.3 - 100%
LowPositive47/4815/1515/1577/7898.7%93.1 - 99.8%47/4815/1515/1577/7898.7%93.1 - 99.8%
ModeratePositive48/4815/1515/1578/78100%95.3 - 100%48/4815/1515/1578/78100%95.3 - 100%
RSV AHighNegative48/4815/1514/1577/7899%93.1 - 99.8%48/4813/1515/1576/7897%91.1 - 99.3%
LowPositive47/4815/1515/1577/7898.7%93.1 - 99.8%47/4815/1515/1577/7898.7%93.1 - 99.8%
ModeratePositive48/4815/1515/1578/78100.0%95.3 - 100%48/4815/1515/1578/78100.0%95.3 - 100%
RSV BHighNegative48/4815/1515/1578/78100%95.3 - 100%46/4814/1514/1574/7895%87.5 - 98.0%
LowPositive48/4815/1515/1578/78100%95.3 - 100%48/4815/1515/1578/78100%95.3 - 100%
ModeratePositive48/4815/1515/1578/78100%95.3 - 100%48/4815/1515/1578/78100.0%95.3 - 100%

Table 5. Comparison of Precision/Reproducibility data from the new RVNAT م to the cleared VRNAT

{7}------------------------------------------------

D. Method Comparison Studies for RVNAT sp and the cleared VRNAT (K083088)

A sample set representing Influenza B, and RSV was prepared by diluting culture positive nasopharyngeal swab samples with negative samples (Table 6). Dilutions were aimed to yield viral load levels close to the low positive levels for each virus type. A total of 62 unique samples were diluted, aliquoted, and frozen. The sample set was tested at the internal site (Site 1) using the cleared VRNAT, and at all three sites using the RVNATs for a total of 62x4=248 unique tests.

Each sample set yielded a total of 186 decisions from 62 unique samples as each test provides a decision of 'Detected' or 'Not Detected' for each of the 3 viruses, Influenza B, and RSV. Of the 62 samples, 3 samples had dual infections where 2 viruses were present.

Sample SetPositivesNegativesTotals
Influenza A154762
Influenza B164662
RSV A/B342862
Total65*121186

Table 6. Method Comparison Study Sample Set.

*3 samples had 2 viruses raising the total number of positives from 62 to 65

Decisions on the RVNAT of or each site were compared to the cleared VRNAT. The Positive Percent Agreement (PPA) and Negative Percent Agreement (NPA) are provided in Tables 7 - 10. The two systems are equivalent based on the collected data.

All VirusesVRNAT (Old System)
PositiveNegativeTotal
RVNATSP(New System)Positive65065PPA 100.0 %(95%CI=94.4% - 100.0%)
Negative0121121NPA 100.0%(95%CI=96.9% - 100.0%)
Total65121186

Table 8. RVNAT se Method Comparison Data collected at Site 2

All VirusesVRNAT (Old System)
PositiveNegativeTotal
RVNATSP(New System)Positive64064PPA 98.5 %(95%CI=91.8% - 99.7%)
Negative1a121122NPA 100.0%(95%CI=96.9% - 100.0%)
Total65121186

3Low positive discordant Influenza A. Repeat tests were positive and gave the expected result.

{8}------------------------------------------------

All VirusesVRNAT (Old System)
PositiveNegativeTotal
RVNATSP(New System)Positive62062PPA 95.4 %(95%CI=87.3% - 98.4%)
Negative3a121124NPA 100.0%(95%CI=96.9% - 100.0%)
Total65121186

Table 9. RVNAT sp Method Comparison Data collected at Site 3

All discordant samples were Influenza A low positive Repeat tests were positive and gave the expected result.

Table 10. RVNAT و Method Comparison Data - Combined data from all 3 sites
All Viruses/All SitesVRNAT (Old System)
PositiveNegativeTotal
RVNATSP(New System)Positive1910191PPA 97.9 %(95%CI=94.8% - 99.0%)
Negative4a363367NPA 100.0%(95%CI=99.0% - 100.0%)
Total195363558

8 All discordant samples were Influenza A low positive. Repeat tests were positive and gave the expected result.

Conclusions Concerning Claim of Substantial Equivalence

The FDA-cleared VRNAT (K083088) is a multiplexed, qualitative test, based on identifying virus-specific nucleic acids for Influenza A virus. Influenza B virus, and Respiratory Syncytial Virus (RSV). Nucleic acids are externally extracted and amplified from viruses and then detected within the Verigene® System by using single-use disposable Test Cartridges.

The RVNAT go is the identical test as the cleared VRNAT with regard to sample extraction chemistry, RT-PCR amplification reagents, and assay detection methods, but it is performed on the Verigene SP System, a fully automated 'sample-to-result' device. As with the Verigene System, the Verigene SP System comprises two units: Reader and SP Processor. The SP Processor has the same hybridization module as the Verigene Processor plus additional modules for sample extraction and multiplex RT-PCR - steps that are performed externally on two separate instruments in the cleared VRNAT.

Assay comparison studies were performed for the RVNAT sp. The studies included:

    1. Analytical Sensitivity (Limits of Detection) determination of the RVNAT sp
    • The analytical sensitivities or Limits of Detection (LOD) of RVNAT p were identical to the o cleared VRNAT for the Influenza A, Influenza B, RSV A, and RSV B strains tested, demonstrating that the performance characteristics of the two devices were equivalent.
    1. Carryover and Cross-contamination Studies on the RVNAT se
    • The carryover and cross-contamination studies showed no evidence of contamination in the o RVNATSP.
    1. Precision/Reproducibility studies on the RVNATsp and comparison to the corresponding VRNAT studies
    • The Precision/Reproducibility studies performed across three clinical test sites replicated the o study that was performed previously for the cleared VRNAT. Comparison of the clinical outcomes and the statistical results for the Precision/Reproducibility studies showed equivalence between the RVNAT sp and the cleared VRNAT.

{9}------------------------------------------------

    1. Method Comparison Studies on the RVNAT (New) and VRNAT (Cleared).
    • In a comparison study conducted across three sites, the same set of samples was compared o between the RVNAT op and the cleared VRNAT. The collective lower bound 95% Cl for the positive and neqative percent agreements was greater than 90%. Moreover, comparisons of signal intensities between the two devices showed equivalent performance across sites and across the individual tests. Collectively, the data showed that the RVNATsp and the cleared VRNAT performance were equivalent.

The performance data support the claims of equivalence between the RVNAT ۾ and the cleared VRNAT. Thus, the performance characteristics in the package insert for the cleared VRNAT with respect to accuracy, sensitivity, specificity, and additional attributes are applicable to the RVNAT sp.

{10}------------------------------------------------

Image /page/10/Picture/0 description: The image shows the seal of the Department of Health & Human Services (HHS). The seal is circular and contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. In the center of the seal is an abstract image of a bird-like figure.

DEPARTMENT OF HEALTH & HUMAN SERVICES

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Building 66 Silver Spring, MD 20993

Gregory W. Shipp, M.D. Vice President, Medical and Regulatory Affairs Nanosphere, Inc. 4088 Commercial Avenue Northbrook, IL 60062

OCT - 9 2009

Re: K092566 Trade/Device Name: Verigene® Respiratory Virus Nucleic Acid Test on the Verigene® SP System Regulation Number: 21 CFR 866.3980 Regulation Name: Respiratory Viral Panel Multiplex Nucleic Acid Assay Regulatory Class: Class II Product Code: OCC, NSU Dated: September 29, 2009 Received: September 30, 2009

Dear Dr. Shipp:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{11}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820). This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific information about the application of labeling requirements to your device, or questions on the promotion and advertising of your device, please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (301) 594-3084. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html.

Sincerely yours.

Sally A. Hoivat, M.Sc., Ph.D. Director Division of Microbiology Devices Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health

Enclosure

{12}------------------------------------------------

Indication for Use

510(k) Number (if known): K092566

Device Name: Verigene® Respiratory Virus Nucleic Acid Test on the Verigene® SP System (RVNATsp)

Indication for Use:

The Verigene® Respiratory Virus Nucleic Acid Test on the Verigene SP System (RVNATsg) is a qualitative multiplex in vitro diagnostic test for the detection and identification of Influenza A Virus, Influenza B Virus, and Respiratory Syncytial Virus (RSV) nucleic acids purified from nasopharyngeal swab specimens obtained from patients symptomatic for viral upper respiratory infection. The test is intended to be used on the Verigene® SP System as an aid in the differential diagnosis of Influenza A, Influenza B, and RSV infections. The test is not intended to detect Influenza C virus.

Negative results do not preclude influenza virus or RSV infection and should not be used as the sole basis for treatment or other management decisions. It is recommended that negative test results be confirmed by culture.

Performance characteristics for Influenza A Virus were established when Influenza A/H3 and A/H1 were the predominant Influenza A viruses in circulation. When other Influenza A viruses are emerging, performance characteristics may vary.

If infection with a novel Influenza A virus is suspected based on current clinical and epidemiological screening criteria recommended by public health authorities, specimens should be collected with appropriate infection control precautions for novel virulent Influenza viruses and sent to state or local health department for testing. Viral culture should not be attempted in these cases unless a BSL 3+ facility is available to receive and culture specimens.

Prescription Use X (Part 21 CFR 801 Subpart D) And/Or

Over-The-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE; CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of In Vitro Diagnostic Device Evaluation and Safety (OIVD)

Division Sign-Off Office of In Vitro Diagnostic Device Evaluation and Safety

Uve Sch
Division Sign-Off

Office of In Vitro Diagnostic Device Evaluation and Safety

510(k) K092566

§ 866.3980 Respiratory viral panel multiplex nucleic acid assay.

(a)
Identification. A respiratory viral panel multiplex nucleic acid assay is a qualitative in vitro diagnostic device intended to simultaneously detect and identify multiple viral nucleic acids extracted from human respiratory specimens or viral culture. The detection and identification of a specific viral nucleic acid from individuals exhibiting signs and symptoms of respiratory infection aids in the diagnosis of respiratory viral infection when used in conjunction with other clinical and laboratory findings. The device is intended for detection and identification of a combination of the following viruses:(1) Influenza A and Influenza B;
(2) Influenza A subtype H1 and Influenza A subtype H3;
(3) Respiratory Syncytial Virus subtype A and Respiratory Syncytial Virus subtype B;
(4) Parainfluenza 1, Parainfluenza 2, and Parainfluenza 3 virus;
(5) Human Metapneumovirus;
(6) Rhinovirus; and
(7) Adenovirus.
(b)
Classification. Class II (special controls). The special controls are:(1) FDA's guidance document entitled “Class II Special Controls Guidance Document: Respiratory Viral Panel Multiplex Nucleic Acid Assay;”
(2) For a device that detects and identifies Human Metapneumovirus, FDA's guidance document entitled “Class II Special Controls Guidance Document: Testing for Human Metapneumovirus (hMPV) Using Nucleic Acid Assays;” and
(3) For a device that detects and differentiates Influenza A subtype H1 and subtype H3, FDA's guidance document entitled “Class II Special Controls Guidance Document: Testing for Detection and Differentiation of Influenza A Virus Subtypes Using Multiplex Nucleic Acid Assays.” See § 866.1(e) for the availability of these guidance documents.