K Number
K091305
Device Name
CLEARPATH
Manufacturer
Date Cleared
2009-10-23

(172 days)

Product Code
Regulation Number
876.1500
Reference & Predicate Devices
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The ClearPath is intended to connect to standard colonoscopes to help facilitate intraprocedural cleaning of a poorly prepared colon by irrigating or cleaning the colon and evacuating the irrigation fluid (water) and feces.

It is for use only by trained medical personnel located in hospitals, clinics, and doctors' offices.

Device Description

The ClearPath's main purpose is to improve procedure reliability and decrease colonoscopy procedures cancellation rate by enabling intra-procedural cleaning of a poorly prepared colon.

The ClearPath utilizes a suction / irrigation head for the purpose of irrigating or cleaning the colon and enabling the rapid evacuation of large amounts of water and feces.

The ClearPath is composed of two major units:

  • . The disposable Irrigator and
  • . the reusable Control cabinet.

The disposable Irrigator is attached to a standard colonoscope by means of silicon bands and does not hinder colonoscope functionality nor affects the procedure sequence in any way.

The Control cabinet supplies the water flow and vacuum control using a peristaltic pump. The standard medical facility vacuum system is the vacuum source.

AI/ML Overview

The provided document describes the ClearPath device, an external uterine monitor for intra-procedural colon cleaning. The information primarily focuses on substantial equivalence to predicate devices based on technical specifications and indications for use, rather than a clinical study evaluating its effectiveness.

However, based on the information provided, here's a breakdown regarding acceptance criteria and the "study" that proves the device meets them, recognizing that this "study" is more accurately described as performance testing for substantial equivalence:

1. Table of Acceptance Criteria and Reported Device Performance:

The acceptance criteria are implicitly defined by the performance specifications of the predicate device, specifically the Olympus (KeyMed) OFP-1 endoscopic flushing pump (K000948). The ClearPath device aims to be "equal or less than" these specifications, which indicates non-inferiority in performance for certain parameters.

Performance ParameterAcceptance Criteria (Predicate: Olympus OFP-1, K000948)Reported Device Performance (ClearPath ACE-1000)Meets Acceptance Criteria?
Air / Water pressure52 PsiUp to 30 PsiYes (Equal or less than)
Flow rateUp to 300 ml/minUp to 300 ml/minYes (Equal or less than)
Suction specifications (Vacuum)Approximately 0.5 Bar (Standard wall suction)Approximately 0.5 Bar (Standard wall suction)Yes (Equal or less than)

2. Sample Size Used for the Test Set and Data Provenance:

The document mentions "testing to demonstrate that the operating performance specifications of the ClearPath is equivalent to the predicates for flow rate, irrigation and vacuum pressures."

  • Sample Size: The document does not specify the sample size (e.g., number of devices tested, number of measurements taken). It refers to "testing" in a general sense for technical specifications.
  • Data Provenance: This testing appears to be internal engineering or lab testing. There is no information provided regarding country of origin, retrospective or prospective nature, or if it involved human subjects or real-world clinical data. It is likely bench testing data.

3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications:

  • Not applicable for this type of performance testing. The "ground truth" for the technical specifications (pressure, flow rate, suction) would be established by calibrated measurement devices and engineering standards, not expert consensus.

4. Adjudication Method for the Test Set:

  • Not applicable as this was technical performance testing against established specifications, not a subjective evaluation requiring adjudication.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study:

  • No, an MRMC comparative effectiveness study was not conducted for this device. The purpose of the submission was to demonstrate substantial equivalence based on technical performance and indications for use, not to prove clinical effectiveness or improvement in human reader performance. This device is an instrument for cleaning during a procedure, not a diagnostic or interpretation aid.

6. Standalone (Algorithm Only Without Human-in-the-Loop Performance) Study:

  • Yes, a standalone "study" (i.e., standalone performance testing) was performed. The device's performance (pressure, flow, suction) was measured independently without human intervention directly influencing the measurement of these parameters. This testing focused on the device's technical capabilities as described in the table above.

7. Type of Ground Truth Used:

  • The ground truth for the performance testing was based on established engineering specifications and measurements from the predicate device (Olympus OFP-1 endoscopic flushing pump) and general industry standards for such devices (e.g., standard wall suction levels).

8. Sample Size for the Training Set:

  • Not applicable. This device is a mechanical instrument, not an AI or machine learning algorithm that requires a "training set."

9. How the Ground Truth for the Training Set Was Established:

  • Not applicable, as there is no training set for this device.

{0}------------------------------------------------

K09/305
ClearPath

510(k) Summary Page 1 of 2 30-Apr-09

EasyGlide Ltd.30 Ha'Ella St.Kfar Truman, 73150IsraelOCT 23 2009
Tel - 011-972-528565644Fax - 011-972-776201003
Official Contact:Izhak Fabian - CEO
Proprietary or Trade Name:ClearPath
Common/Usual Name:External uterine monitor
Classification Name/Code:FDF - colonoscope and accessories, flexible/rigidCFR 876.1500
Device:ClearPath
Predicate Devices:K032688 - Sightline - Colonsight 510BK031773 - Bryne Medical - EndoGatorK000948 Olympus (KeyMed) - OFP-1 endoscopic flushing pump

Device Description:

The ClearPath's main purpose is to improve procedure reliability and decrease colonoscopy procedures cancellation rate by enabling intra-procedural cleaning of a poorly prepared colon.

The ClearPath utilizes a suction / irrigation head for the purpose of irrigating or cleaning the colon and enabling the rapid evacuation of large amounts of water and feces.

The ClearPath is composed of two major units:

  • . The disposable Irrigator and
  • . the reusable Control cabinet.

The disposable Irrigator is attached to a standard colonoscope by means of silicon bands and does not hinder colonoscope functionality nor affects the procedure sequence in any way.

The Control cabinet supplies the water flow and vacuum control using a peristaltic pump. The standard medical facility vacuum system is the vacuum source.

Indications for Use:

The ClearPath is intended to connect to standard colonoscopes to help facilitate intra-procedural cleaning of a poorly prepared colon by irrigating or cleaning the colon and evacuating the irrigation fluid (water) and feces.

It is for use only by trained medical personnel located in hospitals, clinics, and doctors' offices.

{1}------------------------------------------------

K091305
ClearPath

510(k) Summary Page 2 of 2 30-Apr-09

Patient Population:

Individuals undergoing procedures where a colonoscope is used.

Environment of Use:

Hospitals, clinics, and doctors' offices.

Summary of substantial equivalence:

We have performed testing to demonstrate that the operating performance specifications of the ClearPath is equivalent to the predicates for flow rate, irrigation and vacuum pressures.

ProposedClearPath (ACE-1000)Olympus (KeyMed)OFP-1 endoscopicflushing pump
510(k) numberK000948
Air / Water pressurespecificationsUp to 30 Psi52 Psi
Flow rateUp to 300 ml/minUp to 300 ml/min
SuctionspecificationsStandard wall suctionApproximately 0.5 BarStandard wall suctionApproximately 0.5Bar

In addition we have preformed testing under the following standards:

  • IEC 60601 .
  • IEC 60601-1 .
  • . IEC 60601-2-18

The ClearPath is viewed as substantially equivalent to the predicate devices because:

Indications -

  • Identical to predicate K032688 Sightline Colonsight 510B .

Technology -

  • Similar technology used -.
    • K032688 Sightline Colonsight 510B o
    • K031773 Bryne Medical EndoGator 0

Environment of Use -

  • . Identical to predicate - K032688 - Sightline Colonsight 510B

Performance specifications -

  • Irrigation pressures, flow rate and suction pressures are equal or less than the predicates . - K000948 Olympus (KeyMed) OFP-1 endoscopic flushing pump

{2}------------------------------------------------

Image /page/2/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three stripes forming its body and wing. The eagle is facing right. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the eagle.

DEPARTMENT OF HEALTH & HUMAN SERVICES

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room - WO66-G609 Silver Spring, MD 20993-0002

OCT 2 8 2009

EasyGlide Ltd. c/o Mr. Paul E. Dryden President Regulatory Consultant ProMedic, Inc. 24301 Woodsage Drive BONITA SPRINGS FL 34134-2958

Re: K091305

Trade/Device Name: ClearPath Regulation Number: 21 CFR §876.1500 Regulation Name: Endoscope and accessories Regulatory Class: II Product Code: FDF Dated: October 12, 2009 Received: October 14, 2009

Dear Mr. Dryden:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical

{3}------------------------------------------------

Page 2 -

device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely yours,

Janine M. Morris

Janine M. Morris Acting Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

K09/305
181/

Indications for Use Statement

Page 1 of 1

510(k) Number:

K09/305(To be assigned)

Device Name:

ClearPath

Indications for Use:

The ClearPath is intended to connect to standard colonoscopes to help facilitate intraprocedural cleaning of a poorly prepared colon by irrigating or cleaning the colon and evacuating the irrigation fluid (water) and feces.

It is for use only by trained medical personnel located in hospitals, clinics, and doctors' offices.

Prescription Use XX (Part 21 CFR 801 Subpart D) or

Over-the-counter use (21 CFR 807 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Hulutleum

(Division Sign-Off) Division of Reproductive, Abdominal and Radiological Devices 510(k) Number _

§ 876.1500 Endoscope and accessories.

(a)
Identification. An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.(b)
Classification —(1)Class II (special controls). The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.