(323 days)
The Arstasis ITG Vascular Access System is intended to provide access for the percutaneous introduction of devices into the peripheral vasculature and to promote hemostasis at the arteriotomy site as an adjunct to manual compression. The ITG Vascular Access System is indicated for use in patients undergoing diagnostic femoral artery catheterization procedures using 5F or 6F introducer sheaths.
The ITG Vascular Access System is a sterile, single use system that consists of the following components: ITG Vascular Access Device, ITG Guide Wire (0.016'' x 80 cm) and ITG Introducer Needle (21 gauge). The ITG Vascular Access System is designed to be used with the Merit Medical Prelude 5F or 6F Sheath Introducer (Merit Catalog Number PSI-6F-11-018 and PSI-5F-11-018; 510(k) K050962, K070159 and K073035; CE Mark CE023). The ITG Vascular Access Device with integrated needle is designed to traverse the femoral attery wall at a shallow, oblique angle as it enters the arterial lumen. The ITG Vascular Access System is used at the beginning of the procedure and creates a shallow vasular access pathway that, in conjunction with manual compression, helps reduce hemostasis time as compared to manual compression alone after removal of the procedural introducer sheath.
The ITG Vascular Access Device is a sterile, single use device and is comprised of four primary components: (1) guide cannula, (2) anchor mechanism, (3) shaft, and (4) handle with control features.
The guide cannula consists of a polymeric, single concentric lumen catheter that extends from the distal tip of the guide cannula to the anchor mechanism and has an access port that is compatible with a 0.016" Guide Wire. An anchor mechanism is located between the guide cannula and the catheter shaft. The anchor mechanism is used to position the device relative to the vessel wall prior to advancement of the integrated needle, and is also used during guide wire placement. The catheter shaft extends from the anchor mechanism proximally to the distal end of the device handle. At the distal end of the shaft there is a flexible, deflectable section that is controlled by a lever on the device handle. The integrated needle is housed at the distal end of the deflectable section and is attached to the plunger on the handle.
The handle contains controls that operate the different features of the ITG Vascular Access Device. The plunger at the proximal end of the handle controls deployment of the heel and integrated needle. The handle lever articulates the flexible, deflectable section of the shaft from a "path formation position" to a "deflected position" in order to control the path of the integrated needle. The lever is pressed against the handle to deflect the flexible section of the device in order to redirect the needle into the vessel lumen. Pressing the lever also releases the internal stop mechanism to allow full deployment of the integrated needle. Full deployment of the integrated needle into the vessel lumen is accomplished by completely depressing the plunger.
The 0.016" Guide Wire is inserted into the plunger port and advanced through the ITG Vascular Access Device into the arterial lumen. Plunger retraction automatically releases the anchor heel and retracts the integrated needle for device removal. The 0.016" Guide Wire remains in the arterial lumen as the ITG Vascular Access Device is removed. An introducer sheath is advanced over the 0.016" Guide Wire and the user proceeds with the percutaneous procedure.
The provided text describes the Arstasis ITG Vascular Access System and its regulatory submission (K091006). However, it does not include specific acceptance criteria or a detailed study proving the device meets said criteria. The document focuses on demonstrating substantial equivalence to predicate devices.
Here's a breakdown of the requested information based on the provided text:
1. A table of acceptance criteria and the reported device performance
The document states that "Bench performance testing of the ITG Vascular Access System demonstrates that the ITG System meets or exceeds the performance requirements for the intended clinical use of the system." However, it does not list specific acceptance criteria (e.g., success rates, hemostasis times, complication rates) or provide a table comparing them to reported device performance. It only makes a general statement of compliance.
2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)
- Sample Size for Test Set: The document mentions "Multiple clinical evaluations of the ITG Vascular Access System were conducted." It also states that "The long term safety, as well as the ability to access and re-access, was retrospectively studied in a smaller cohort of patients." However, specific sample sizes for these clinical evaluations are not provided.
- Data Provenance: The document does not specify the country of origin for the clinical data. It clearly states that the long-term safety and re-access study was retrospective. The "short term safety and clinical performance" studies are implied to be prospective, but this is not explicitly stated.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)
This information is not provided in the document. The filing focuses on the device itself and its equivalence to predicates, not on detailed expert-driven evaluations of a test set.
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set
This information is not provided in the document.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
The Arstasis ITG Vascular Access System is a medical device for vascular access and hemostasis, not an AI or imaging diagnostic tool. Therefore, a multi-reader multi-case (MRMC) comparative effectiveness study is not applicable in this context, and no such study is mentioned.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
This is not applicable as the device is a physical medical instrument, not an algorithm.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)
For the clinical evaluations, the "ground truth" would likely be based on clinical outcomes data related to successful vascular access, hemostasis time, and complication rates directly observed during the procedures. However, the document does not explicitly define how this "ground truth" was established or what specific metrics were used beyond general statements about "safety and clinical performance."
8. The sample size for the training set
The document does not refer to a "training set" in the context of device development or clinical evaluation. For a physical medical device, development often involves engineering testing, bench testing, and animal/cadaver studies, rather than a "training set" in the machine learning sense. The clinical studies mentioned are for verification and validation, not model training.
9. How the ground truth for the training set was established
As there is no mention of a "training set" in the machine learning context, this information is not applicable. For the preliminary animal and cadaver studies, the "ground truth" would be direct observation of the device's performance in these non-human models.
{0}------------------------------------------------
K0911006
: :
FEB 2 5 2010
510(k) SUMMARY
Arstasis, Inc.
ITG Vascular Access System
510(k) Notification
GENERAL INFORMATION
| Manufacturer: | Arstasis, Inc.1021 Howard Avenue, Suite CSan Carlos, CA 94070Phone: (650) 508-1549Facsimile: (650) 594-4326Establishment Registration Number: Pending |
|---|---|
| Contact Person: | Dominique FillouxVice President Research & Development, Clinical and Regulatory Affairs |
| Date Prepared: | February 24, 2010 |
| DEVICE INFORMATION | |
| Trade name: | ITG Vascular Access System |
| Classification Names: | Catheter, Introducer §870.1340 (DYB) |
|---|---|
| ----------------------- | -------------------------------------- |
Class II
Classification:
PREDICATE DEVICES
Cardiva Medical, Inc. Vasostasis / Boomerang System (K041486 / K061075) Boston Scientific Super Sheath Introducer Sheath (K060190) Merit Medical Prelude Introducer Sheath / Kit (K071059)
INTENDED USE/INDICATIONS FOR USE
The Arstasis ITG Vascular Access System is intended to provide access for the percutaneous introduction of devices into the peripheral vasculature and to promote hemostasis at the arteriotomy site as an adjunct to manual compression. The ITG Vascular Access System is indicated for use in patients undergoing diagnostic femoral artery catheterization procedures using 5F or 6F introducer sheaths.
DEVICE DESCRIPTION
The ITG Vascular Access System is a sterile, single use system that consists of the following components: ITG Vascular Access Device, ITG Guide Wire (0.016'' x 80 cm) and ITG Introducer Needle (21 gauge). The ITG Vascular Access System is designed to be used with the Merit Medical Prelude 5F or 6F Sheath Introducer (Merit Catalog Number PSI-6F-11-018 and PSI-5F-11-018; 510(k) K050962, K070159 and K073035; CE Mark CE023). The ITG Vascular Access Device with integrated needle is designed to traverse the femoral attery wall at a shallow, oblique angle as it enters the arterial lumen. The ITG Vascular Access System is used at the beginning of the procedure and creates a shallow vasular access pathway that, in conjunction with manual compression, helps reduce hemostasis time as compared to manual compression alone after removal of the procedural introducer sheath.
The ITG Vascular Access Device is a sterile, single use device and is comprised of four primary components: (1) guide cannula, (2) anchor mechanism, (3) shaft, and (4) handle with control features.
{1}------------------------------------------------
The guide cannula consists of a polymeric, single concentric lumen catheter that extends from the distal tip of the guide cannula to the anchor mechanism and has an access port that is compatible with a 0.016" Guide Wire. An anchor mechanism is located between the guide cannula and the catheter shaft. The anchor mechanism is used to position the device relative to the vessel wall prior to advancement of the integrated needle, and is also used during guide wire placement. The catheter shaft extends from the anchor mechanism proximally to the distal end of the device handle. At the distal end of the shaft there is a flexible, deflectable section that is controlled by a lever on the device handle. The integrated needle is housed at the distal end of the deflectable section and is attached to the plunger on the handle.
The handle contains controls that operate the different features of the ITG Vascular Access Device. The plunger at the proximal end of the handle controls deployment of the heel and integrated needle. The handle lever articulates the flexible, deflectable section of the shaft from a "path formation position" to a "deflected position" in order to control the path of the integrated needle. The lever is pressed against the handle to deflect the flexible section of the device in order to redirect the needle into the vessel lumen. Pressing the lever also releases the internal stop mechanism to allow full deployment of the integrated needle. Full deployment of the integrated needle into the vessel lumen is accomplished by completely depressing the plunger.
The 0.016" Guide Wire is inserted into the plunger port and advanced through the ITG Vascular Access Device into the arterial lumen. Plunger retraction automatically releases the anchor heel and retracts the integrated needle for device removal. The 0.016" Guide Wire remains in the arterial lumen as the ITG Vascular Access Device is removed. An introducer sheath is advanced over the 0.016" Guide Wire and the user proceeds with the percutaneous procedure.
DATA DEMONSTRATING SUBSTANTIAL EQUIVALENCE
Test data provided in the subject Premarket Notification include: a) bench performance b) biocompatibility c) preliminary animal (in non GLP studies) and cadaver assessments', and d) the results of clinical investigations for the ITG Vascular Access System. Bench performance testing of the ITG Vascular Access System demonstrates that the ITG System meets or exceeds the performance requirements for the intended clinical use of the system. Biocompatibility testing of ITG Vascular Access System was conducted pursuant to FDA's Guidance Document (#G95-1), Use of International Standard ISO-10993-1. "Biological Evaluation of Medical Devices Part 1: Evaluation and Testing (1995)". Based on the test results, the ITG Vascular Access System is biocompatible.
Multiple clinical evaluations of the ITG Vascular Access System were conducted. The short term safety and clinical performance of ITG Vascular Access System were established. The long term safety, as well as the ability to access and re-access, was retrospectively studied in a smaller cohort of patients. In summary, the cumulative clinical results demonstrate that the ITG Vascular Access System is substantially equivalent to its predicates in providing access to the arterial lumen and facilitating the introduction and placement of devices into the peripheral vasculature and achievement of hemostasis.
1 The Preliminary Animal Studies and Cadaver Assessments were conducted using prototypes of a similar design and configuration.
{2}------------------------------------------------
Image /page/2/Picture/2 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular border with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" around the edge. Inside the circle is a stylized image of an eagle or bird-like figure with its wings spread.
FEB 2 5 2010
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room W-066-0609 Silver Spring, MD 20993-0002
Arstasis, Inc. c/o Mr. Dominique Filloux Vice President. Research and Development and RA/CA 1021 Howard Avenue, Suite C San Carlos, CA 94070
Re: K091006
Arstasis ITG Vascular Access System Regulation Number: 21 CFR §870.4450 Regulation Name: Catheter, Introducer Regulatory Class: Class II (Two) Product Code: DYB Dated: December 30, 2009 Received: December 31, 2009
Dear Mr. Filloux:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{3}------------------------------------------------
Page 2 - Mr. Dominique Filloux
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807): labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely vours.
Bram D. Zuckerman, MD Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
Indications for Use
510(k) Number (if known):
K091006
Device Name:
ITG Vascular Access System
Indications For Use:
The ITG Vascular Access System is intended to provide access for the percutaneous introduction of devices into the peripheral vasculature and to promote hemostasis at the arteriotomy site as an adjunct to manual compression. The ITG Vascular Access System is indicated for use in patients undergoing diagnostic femoral artery catheterization procedures using 5F of 6F introducer sheaths.
Prescription Use J (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
sumer R. V. Mmmmmm
(Division Sign-Off) (Division Oigh Sign Scular Devices
510(k) Number_KO910 GG
Page 1 of 1
§ 870.1340 Catheter introducer.
(a)
Identification. A catheter introducer is a sheath used to facilitate placing a catheter through the skin into a vein or artery.(b)
Classification. Class II (performance standards).