AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The eSensor® CF Genotyping Test is an in vitro diagnostic device used to simultaneously detect and identify a panel of mutations and variants in the cystic fibrosis transmembrane conductance regulator (CFTR) gene in genomic DNA samples isolated from human peripheral whole blood specimens. The panel includes mutations and variants recommended by the 2004 American College of Medical Genetics (ACMG). The eSensor® CF Genotyping Test is a qualitative genotyping test that provides information intended to be used for cystic fibrosis carrier screening as recommended by ACMG and the 2005 American College of Obstetricians and Gynecologists (ACOG) for adults of reproductive age, as an aid in newborn screening for cystic fibrosis, and in confirmatory diagnostic testing for cystic fibrosis in newborns and children. The test is not indicated for use in fetal diagnostic or pre-implantation testing. This test is also not indicated for stand-alone diagnostic purposes and results should be used in conjunction with other available laboratory and clinical information.

The eSensor® CF Genotyping Test is intended for use on the eSensor® XT-8 System.

Device Description

The eSensor® CF Genotyping Test on the eSensor® XT-8 System is an in vitro diagnostic device for performing hybridization and genotyping of multiple mutations and/or polymorphisms in an amplified DNA sample. A single-use, disposable test cartridge is used to perform hybridization and genotyping. The cartridge contains an EEPROM chip which transmits the cartridge lot number, expiration date and protocol identity to the XT-8 instrument.

The analysis process for each sample consists of three steps: 1) Genomic DNA isolated from whole blood obtained using EDTA as anti-coagulant is combined with PCR Mix and Taq polymerase enzyme and is subjected to amplification of target sequences by PCR using a thermal cycler. 2) Amplified DNA is treated with exomuclease enzyme to generate single-stranded target DNA. 3) Single-stranded, amplified target DNA is mixed with hybridization and genotyping reagents and transferred to an eSensor® CF Genotyping Tet cartinge, and the cartridge is inserted in the eSensor® XT-8 Instrument controls the circulation of the sample inside the cartridge to allow hybridization at a controlled temperature and then detects and genotypes the sample by voltammetry.

Genotyping of the test panel polymorphisms is achieved by a sandwich assay principle: 1) Each pair of electrodes contains a different synthetic oligonucleotide capture probe which is complementary to one of the target DNA fragments. 2) The hybridization reagents contain pairs of ferrocenc-labeled synthetic oligonucleotide signal probes; one member of each pair is complementary to the major allele sequence of the target polymorphism, while the second member of the pair is complementary to the minor allele sequence. Each member of the probe pair has a ferrocene label with a different oxidation potential for each allele. 3) Single-stranded, amplificd target DNA hybridizes to its specific capture probe, and in turn hybridizes to the allele-specific, ferrocenc-labeled signal probe. 4) Each electrode of the array is analyzed by voltammetry; the target polymorphism is determined by the location of the electrode containing the capture probe, and the genotype is identified by the ratio of signals from the allclespecific ferrocene labels. The array also includes positive and negative confirm the hybridization reaction and detect non-specific signals.

Upon completion of the test, the EEPROM chip on the cartridge contains information that prevents its re-use with a new sample. The eSensor® XT-8 instrument analyzes the results and provides a report of the test results.

AI/ML Overview

The eSensor® CF Genotyping Test on the XT-8 System is an in vitro diagnostic device intended for genotyping multiple mutations or polymorphisms in an amplified DNA sample utilizing electrochemical detection technology.

Here's an analysis of the acceptance criteria and the study that proves the device meets them:

1. Table of Acceptance Criteria and Reported Device Performance

The acceptance criteria are implied through the successful demonstration of 100% agreement with DNA sequencing after repeat testing for initial no-calls across various studies. The primary performance metric is the agreement with the gold standard (DNA sequencing).

Acceptance Criteria (Implied)Reported Device Performance
Overall Agreement100% (after repeat testing for initial no-calls) for all studies
First-pass Correct CallsVaries by study/site, generally high (e.g., 96.9% to 100%)
First-pass No-callsPresent in some initial runs, resolved to 100% correct after repeats
First-pass Miscalls0% (across all studies at first-pass and final)

2. Sample Sizes and Data Provenance

  • Test Set Sample Sizes:
    • Reproducibility Study: 22 gDNA samples containing positive calls for all ACOG/ACMG panel mutations and the 5/7/9T polymorphism. Each sample was run in duplicate, generating 1320 total replicates across multiple sites and operators.
    • Lot-to-Lot Reproducibility: 21 genomic DNA samples covering all possible genotypes.
    • Genomic DNA Extraction Reproducibility: 20 whole blood samples of different genotypes.
    • Method Comparison: 112 gDNA samples extracted from whole blood.
  • Data Provenance: The document does not explicitly state the country of origin for the data or whether it was retrospective or prospective. Given the nature of a 510(k) submission for a diagnostic test, it is highly likely that samples were collected prospectively or obtained from biobanks with appropriate ethical approvals, and the studies were conducted in a controlled, prospective manner to validate device performance. The reproducibility study involved an "Internal Site" and "External Sites," suggesting multi-center data collection, likely within the USA where the manufacturer is located.

3. Number of Experts and Qualifications for Ground Truth Establishment

The document does not explicitly state the number of experts or their qualifications for establishing the ground truth. However, the ground truth was established using DNA sequencing, which is a widely accepted gold standard in genetic testing. The interpretation of DNA sequencing results for CFTR mutations is a specialized task typically performed by molecular geneticists or clinical laboratory directors with expertise in sequencing analysis.

4. Adjudication Method for the Test Set

The document does not describe a formal adjudication method (e.g., 2+1, 3+1). Instead, discrepancies (initial "no-calls") in the device's output were resolved by re-running the tests, leading to "Final Correct Calls." Miscalls were not observed.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

No multi-reader multi-case (MRMC) comparative effectiveness study was done. This device is an automated genotyping test, not an imaging device that would typically involve human readers for interpretation. The comparison is between the automated device's output and the established DNA sequencing results, and not aimed at quantifying human reader improvement with AI assistance.

6. Standalone Performance

Yes, standalone performance (algorithm only without human-in-the-loop) was performed. The eSensor® CF Genotyping Test is an automated system that provides results via a software program on the eSensor® XT-8 instrument. The performance data presented (reproducibility, method comparison, limit of detection, interfering substances) all reflect the standalone analytical performance of the device without explicit human intervention in the result interpretation process (beyond standard laboratory procedures for running the test and reviewing the final report).

7. Type of Ground Truth Used

The type of ground truth used was DNA sequencing. This is explicitly stated as the reference method for comparison throughout the performance characteristics section (e.g., "All samples gave 100% correct calls when compared with DNA sequencing," and "All samples gave 100% agreement with DNA sequencing").

8. Sample Size for the Training Set

The document does not specify a separate training set size for the device's development or algorithm. This is common for predicate-based 510(k) submissions of in vitro diagnostic devices where the focus is on analytical and clinical validation of the final product, rather than the development process of a machine learning algorithm that typically requires a distinct training phase. The device uses established electrochemical detection technology and a predefined panel of mutations, not a learning algorithm that would necessitate a large training dataset as seen in modern AI/ML submissions.

9. How Ground Truth for the Training Set Was Established

As no separate training set is explicitly mentioned or seems applicable in the context of this device's technology, the method for establishing ground truth for a training set is not provided. The development of the device likely relied on established scientific knowledge of CFTR mutations and electrochemical detection principles, rather than an iterative machine learning training process. The validation studies (using DNA sequencing as ground truth) confirm the performance of the final device.

{0}------------------------------------------------

KU90901

510(k) SUMMARY: eSensor® CF Genotyping Test on XT-8 System

Preparation Date: June 12, 2009

Submitted By:

Osmetech Molecular Diagnostics 757 South Raymond Avenue Pasadena, CA 91105 USA Fax: 626 463-2012 Phone: 626 463-2000

JUL - 6 2009

Contacts:

Robert Dicheck, Vice President - Quality & Regulatory Affairs (Official Correspondent) Aviva Jacobs, Ph.D., Associate Director - Product Development (Project Manager)

Proprietary Names and Classifications:

For the assav: eSensor® CF Genotyping Test (Kit) Regulation: 21CFR 866.5900 Panel: Immunology (82) Classification: II Product Code: NUA - Cystic Fibrosis Transmembrane Conductance Regulator (CFTR) Gene Mutation Detection System

For the instrument: eSensor® XT-8 Instrument (System) Regulation: 21CFR 862.2570 Panel: Clinical Chemistry (75) Classification: II Product Code: NSU - Instrument for Clinical Multiplex Test Systems

Common name:

Cystic Fibrosis Transmembrance Regulator (CFTR) Gene Mutation Detection System

Intended uses:

The eSensor® CF Genotyping Test is an in vitro diagnostic device used to simultaneously detect and identify a panel of mutations and variants in the cystic fibrosis transmembrance regulator (CFTR) gene in genomic DNA samples isolated from human peripheral whole blood specimens. The panel includes mutations and variants recommended by the 2004 American College of Medical Genetics (ACMG). The eSensor® CF Genotyping Test is a qualitative genotyping test that provides information intended to be used for cystic fibrosis carrier screening as recommended by ACMG and the 2005 American College of Obstetricians and Gynecologists (ACOG) for adults of reproductive age, as an aid in newborn screening for cystic fibrosis, and in confirmatory diagnostic testing for cystic fibrosis in newborns and children. The test is not in fittal diagnostic or preimplantation testing. This test is also not indicated for stand-alone diagnostic purposes and results should be used in conjunction with other available laboratory and clinical information.

Special conditions for use statement(s):

For Prescription Use Only

The cScasor® CF Genotyping Test is an in vitro diagnostic device intended for genotyping multiple mutations or polymorphisms in an amplified DNA sample utilizing electrochemical detection technology, for use on the eSensor® XT-8 Instrument.

{1}------------------------------------------------

Predicate devices:

eSensor® CFCD System, K060543 and K051435 InPlex™ CF Molecular Test, K063787 eSensor® XT-8 Instrument, K073720

Device Description:

The eSensor® CF Genotyping Test on the eSensor® XT-8 System is an in vitro diagnostic device for performing hybridization and genotyping of multiple mutations and/or polymorphisms in an amplified DNA sample. A singleuse, disposable test cartridge is used to perform hybridization and genotyping. The cartridge contains an EEPROM chip which transmits the cartridge lot number, expiration date and protocol identity to the XT-8 instrument.

The analysis process for each sample consists of three steps: 1) Genomic DNA isolated from whole blood obtained using EDTA as anti-coagulant is combined with PCR Mix and Taq polymerase enzyme and is subjected to amplification of target sequences by PCR using a thermal cycler. 2) Amplified DNA is treated with exomuclease enzyme to generate single-stranded target DNA. 3) Single-stranded, amplified target DNA is mixed with hybridization and genotyping reagents and transferred to an eSensor® CF Genotyping Tet cartinge, and the cartridge is inserted in the eSensor® XT-8 Instrument controls the circulation of the sample inside the cartridge to allow hybridization at a controlled temperature and then detects and genotypes the sample by voltammetry

Genotyping of the test panel polymorphisms is achieved by a sandwich assay principle: 1) Each pair of electrodes contains a different synthetic oligonucleotide capture probe which is complementary to one of the target DNA fragments. 2) The hybridization reagents contain pairs of ferrocenc-labeled synthetic oligonucleotide signal probes; one member of each pair is complementary to the major allele sequence of the target polymorphism, while the second member of the pair is complementary to the minor allele sequence. Each member of the probe pair has a ferrocene label with a different oxidation potential for each allele. 3) Single-stranded, amplificd target DNA hybridizes to its specific capture probe, and in turn hybridizes to the allele-specific, ferrocenc-labeled signal probe. 4) Each electrode of the array is analyzed by voltammetry; the target polymorphism is determined by the location of the electrode containing the capture probe, and the genotype is identified by the ratio of signals from the allclespecific ferrocene labels. The array also includes positive and negative confirm the hybridization reaction and detect non-specific signals.

Upon completion of the test, the EEPROM chip on the cartridge contains information that prevents its re-use with a new sample. The eSensor® XT-8 instrument analyzes the results and provides a report of the test results.

Comparison to technological features of the predicate devices: The following is a comparison of the Osmetech Molecular Diagnostics eSensor® CF Genotyping Test on the XT-8 System to the predicates.

CharacteristiceSensor® CFCD System(Predicate 1: K060543 and K051435)InPlex™ CF Molecular Test (Predicate2: K063787)eSensor® CF Genotyping Test
Test typeQualitative genetic test for single nucleotidepolymorphism detectionQualitative genetic test for singlenucleotide polymorphism detectionSame as predicates 1 and 2
Sample TypeGenomic DNA obtained from a human wholeblood sampleGenomic DNA obtained from a humanwhole blood sampleSame as predicates 1 and 2
Target of detectionSingle-nucleotide polymorphismSingle-nucleotide polymorphismSame as predicates 1 and 2
DNA extractionPerformed off-linePerformed off-lineSame as predicates 1 and 2
GenesCFTRCFTRSame as predicates 1 and 2
Number of Loci genotyped23 mutations and 1 variant as recommended byACMG (2004)/ACOG (2005).23 mutations and 4 variantsrecommended by ACMG(2004)/ACOG (2005).Same as predicates 1
Genotyping reactionlocationTest cartridgeMicrofluidic cardSame as predicate 1
Genotyping principleSandwich hybridization testFluorometric resonanceenergy transfer (FRET)Same as predicate 1
Instrument operating systemeSensor® Instrument Model 4800Multi-well fluorometereSensor® Instrument Model XT-8Random access compatible with multiplesimultaneous test types.
Assay resultsAssay signal results are interpreted by asoftware program and are assigned result thatis presented to the end-user in a report formatInPlex™ CF Molecular TestCall Reporting SoftwareAssay signal results are interpreted by asoftware program and are assigned aresult that is presented to the end-user ina report format

{2}------------------------------------------------

Performance Characteristics:

Site to Site, Operator to Operator, Day to Day, Run to Run and sample reproducibility

A reproducibility study was performed over 5 non consecutive days at 3 different sites and 1 internal site). Each site performed the testing twice each day, using two different operators and the same testing materials. Twenty two (22) gDNA samples containing positive calls for all ACOG/ACMG panel mutations in addition to the 5/7/9T polymorphism were used. For practical considerations, the 22 samples were divided into 2 set of 11 samples, so that the sample set was run in duplicate to evaluated intra assay reproducibility using a single kit. Only one lot of materials was used for this study.

By SiteByOperatorNumber ofSampleReplicatesFirst-passCorrectcallsFirst-passNo-callsFirst-passMiscallsFinalCorrectcallsFinalincorrectcalls%Agreement
Site A(Internal)Operator 12202081202200100%
Operator 2220216402200100%
Site B:(External)Operator 1220215502200100%
Operator 2220211902200100%
Site C(External)Operator 1220216402200100%
Operator 2220218202200100%
Total1320128436013200100%

Summary of Inter-laboratory, Inter-operator, Reproducibility Results

Summary of Reproducibility Results by Sample Genotype and Site

Sample Genotype bySequencing5/7/9TNumber of Sample Replicatestested byeSensor® CFGenotyping TestSite ASite ASite BSite BSite CSite CNumber of CF Sample Calls Before Repeat TestingSite ACorrectCallsSite BCorrectCallsSite CCorrectCallsNoCallsMiscallsNumber of CF Sample Calls After Repeat TestingSite ACorrectcallsSite BCorrectCallsSite CCorrectcallsNo Calls
1717-1G>A7T/7T606060595959306060600
1898+1G>A/ΔF5087T/9T606060596058306060600
2184delA/ΔF5087T/9T606060586058406060600
3120+1G>A/621+1G>T7T/9T606060596059206060600
2789+5G>A/2789+5G>A7T/7T606060596059206060600
3659delC/ΔF5087T/9T606060595960206060600
3849+10KbC>T/ΔF5087T/9T606060605958306060600
621+1G>T/G85E7T/9T606060586060206060600
711+1G>T/621+1G>T7T/9T606060596060106060600
A455E/621+1G>T9T/9T606060596059206060600
Δ15077T/7T606060606060006060600
G542X7T/9T606060595960206060600
G551D/R347P7T/7T606060595960206060600
N1303K7T/9T606060606060006060600
R1162X7T/7T606060606059106060600
R117H/ΔF5085T/9T606060606059106060600
R334W7T/7T606060605959206060600
R553X/G551D7T/7T606060596060106060600
R560T/ΔF5087T/9T606060596060106060600
W1282X5T/7T606060596059206060600
WT7T/7T606060606060006060600
R117H/ΔF5085T/9T606060606060006060600

{3}------------------------------------------------

Lot to Lot Reproducibility

A total of 21 genomic DNA samples covering all possible genotypes were tested using three different kit lots and tested using the eSensor® CF Genotyping Test. The data were evaluated after first-pass results and following additional runs for no-calls. All samples gave 100% correct calls when compared with DNA sequencing. No impact of kit lot observed in this study. The following table summarizes the results of lot to lot reproducibility study.

LOTSamplesTestedFirst passcorrectcallsFirst passFinalFinalcorrectcalls%Agreement
No CallsMiscallsNo CallsMiscalls
12121000021100%
22121000021100%
32120100021100%
Total6362100063100%

Genomic DNA Extraction Reproducibility

A total of 20 whole blood samples of different genotypes were extracted by three commonly used extraction methods and tested using the eSensor® CF Genotyping Test.. The data were evaluated after first-pass results and following additional runs for no-calls. All samples gave 100 correct calls when compared with DNA sequencing. There was no impact of extraction method observed in this study. The following table summarizes the results of extraction reproducibility study.

Extraction Method# Samples Tested/GenotypeFirst pass correct callsFirst pass No CallsMiscallsFinal correct callsFinal Agreement
A20173020100
B20182020100
C20*191020100

{4}------------------------------------------------

Method Comparison

In a method comparison study, a total of 112 gDNA samples extracted from whole blood with A260-280 ratios of 1.2-2.5 were genotyped using the eSensor® CF Genotyping Test and DNA sequencing.

All samples gave 100% agreement with DNA sequencing. The following table summarizes the results of the method comparison study.

SequencingCalls1st Pass CF GT Calls% AgreementFINAL CF GT Calls% Agreement
Genotype bysequencingCalls permutationPosNegPosNegNoCallsOverallPosNegPosNegNoCallsOverallPosNeg
AF50811247ર્ણ ર466599.197.9100476501001001 (AQ
G542X1127! OS61 05ਰੇਰੇ I85.7100710501 001001 00
WI 282X112б10061 0601001 000100610601 00100I (X)
GSSID112810481040100100100,810401 00100100
621+1G>T1127રે OS71 0501001001007105D1 001001 (ਮ)
N1303K11281 047104ਉਹ87.5ﺍ 0081040100100100
RSSX1124108410801 (A01001 00410801 001001 00
AI2071123I 093I tra01 (X)1 00100310901001001 00
3120+1G>A112د1102I I Q01001001 00211001001001 00
3849+10kbC>T11251075107010010010051070! (10I ()()100
RITTH1128104810401001 00100810401001 00
1717-1G>A1125107510701 00100100510701 00100100
2789+5G>A11251 07510701 00100ા રીવ51070100100100
R334W11241 084ા 080100100ﺍ 0041 080100100100
K347P11241 0841080100100ﺍ 004108010000100
711+1G>T11241 68410801001001 0041 0801001 00100
R 560 T11241 08ﻟﺪ108-ਰੇਰੇ75.010041 080100100100
RITE2X11281048104010010010081040100100100
3659dclC11241084108010010010041080100100100
A455E1122110211001001 60100211001001001 00
C82EII271057દ્દ OS0100100100710501001001 00
2184dc1A112211021100100100100211001001001 00
1898+1G>V112211021100100100100211001001001 00
Reflex Polymorphism
IVS8--8 5T/7T/9TVariant (S)1126 I1066 1119601 001 00100611)S01001001 00
Polymorphisms not specifically genotyped
1507V112-1111110100100100111101001 001 00
FS08C1121l l1110ﺍ 00100ા ೧೧l1 I0100100100
Potentially interfering mutations not part of assay panel
2183AA>G1122110211001001001002l 1001 001 00100
RITL112l 11l1110100100100l ] I01001 00100
Grand Total26881622526ા રેક્ષે2526499,997.5100162252630100100100

s For the purpose of the IVSS-577779T Variant, "Positive" samples are east one opy of the ST allele while "Negative" samples are regarded as having only the 7T and/or 9T allele. $1 sample is 5T/5T Mutant.

The number of positive sequencing calls is greater than the number of inclusion of compond heleruzyeus sumples.

The Grand Total consists of the total number of sells for mutations, the 1 57/57 sample included, polymophisms not specifically genoryed, and the potentially interfeing not part of assay panel. 1507V, F508C, 2183AA>G, and R117L are non-panel polymorphism containing samples correctly called as Wild-Type, which are not included in the grand total for calls. No samples with 1506V were tested.

{5}------------------------------------------------

Other Characteristics of the eSensor® CF Genotyping Test:

CharacteristicResult
Limit of detectionTwo genomic DNA samples of different genotypes were extracted from whole blood stored in EDTA and serially diluted andtested 20 times each at input amounts of 0.01, 0.1, 1, 10, 100, 500 and 1000 ng/PCR. using the eSensor® CF Genotyping Test.An additional run was performed for tests that gave a first pass no call result. All input amounts from 0.01ng to 1000ng for bothsamples gave equivalent final performance. (100% Agreement with 98.76%, 95% LCB), (99.95% LCB on a per SNP basis).The limit of detection was established as the lowest concentration at which no-calls or mis-calls were obtained.The lower detection limit was determined to be 0.1 ng of purified DNA per reaction and the upper detection limit wasdetermined to be 1000 ng of purified DNA per reaction. The recommended range of DNA input amounts for the eSensor® CFGenotyping Test is from 10 to 500 ng.
Interfering substancesTest performance was not affected by addition of the following substances seven whole blood samples of different genotypesprior to extraction:Bilurubin (30 mg /dL whole blood). Triglycerides (500 mg/dL whole blood). Hemoglobin (~20g /dL whole blood). EDTA (at a concentration equivalent to 5-fold higher than that provided by a standard EDTA blood collection tube)
Test LimitationsThe eSensor® CF Genotyping Test does not identify all possible mutations in the CFTR gene, for example E60X, 1148T,1078delT, V520F, 2143delT, 3199del6, D1152H, 3876delA, 2183AA>G, R560K, R117L, R347H, G551S, 711+5G>A,394delTT, and 3905insT. A negative result for all the mutations in this panel does not necessarily indicate that the individual isnegative for cystic fibrosis (carrier or affected status). The mutations included in this test represent >80% of the mutationscarried by Caucasian American adults.The eSensor® CF Genotyping Test does not identify the 1506V, 1507V, and F508C polymorphisms, thus in the case of AF508homozygosity, reflex testing by bi-directional DNA sequencing is recommended
Interferingmutations and polymorphismsMore than 1,000 mutations have been identified in the CFTR gene with varying confidence that they are CF disease causing.When present in the same region as a panel mutation, they may interfere with genotyping resultsThe following mutations have been evaluated and demonstrated not to impact the results of the eSensor® CF Genotyping TestNon-Panel Mutation or Polymorphism ACOG/ACMG Panel Mutation 2183AA>GR117L 2184delAR117H

{6}------------------------------------------------

Kit Stability:

eSensor® CF Genotyping Test kit components should be stored under the appropriate conditions until the expiration date printed on the label:

  • · PCR Box containing CF Genotyping Test PCR Mix and Taq Polymerase: Store at -20℃ in a designated pre-PCR area.
  • · Cartridges: Store at 10° to 25°C
  • · Genotyping Box containing Exonuclease, CF Genotyping Test Signal Buffer, XT-Buffer 2: Store at -20℃ in a designated post-PCR area.

In-process stability has been established for the following components, working reagents and samples:

  • · Cartides can be stored for up to 14 days after opening the foil pouches. If stored, cartridges should be kept in their original foil pouch at room temperature in a dry place with the zip-loc closure sealed.
  • · Once open, reagents can be stored at -20°C for up to 30 days.
  • · Reagents can be thawed up to 3 times.
  • Whole blood stored in EDTA can be stored for up to 4 weeks after collection of gDNA for use in the eSensor® CF Genotyping Test.
  • · PCR product can be stored at 4°C or -20°C for up to 7 days.
  • · Exonuclease-digested PCR product can be stored store at 4℃ or -20℃ for up to 7 days.
  • · After combining the exonuclease-digested PCR with hybridization reaction can be loaded on the cartridge and held at ambient temperature for up to 8 hours before initiating hybridization of the cartridge on the XT-8 instrument.

Conclusion:

The above internal and clinical test results support the safety and effectiveness of the eSensor® XT-8 System, and demonstrate substantial equivalence to the predicate device.

eSensor® is a registered trademark of Osmetech and its subsidiaries

{7}------------------------------------------------

Image /page/7/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES · USA" around the perimeter. Inside the circle is a stylized symbol featuring three abstract human figures connected by flowing lines, representing the department's focus on health and human well-being.

DEPARTMENT OF HEALTH & HUMAN SERVICES

Food and Drug Administration 10903 New Hampshire Avenue Building 66 Silver Spring, MD 20993

Osmetech Molecular Diagnostics c/o Mr. Robert S. Dicheck, Vice President - Quality & Regulatory Affairs 757 S. Raymond Avenue Pasadena, CA 91105

JUL - 6 2009

Re: K090901

Trade/Device Name: eSensor® CF Genotyping Test Regulation Number: 21 CFR 866.5900 Regulation Name: CFTR (cystic fibrosis transmembrane conductance regulator) gene mutation detection system Regulatory Class: Class II Product Code: NUA, NSU Dated: June 9, 2009 Received: June 10, 2009

Dear Mr. Dicheck:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Ilsting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); and good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820). This letter

{8}------------------------------------------------

Page 2 - Mr. Robert S. Dicheck

will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (301) 796-5451. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Maria mchan

Maria M. Chan, Ph.D. Director Division of Immunology and Hematology Devices Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health

Enclosure

{9}------------------------------------------------

Indication(s) for Use

510(k) Number (if known): K090901

Device Name: eSensor® CF Genotyping Test

Intended Use/Indication(s) For Use:

The eSensor® CF Genotyping Test is an in vitro diagnostic device used to simultaneously detect and identify a panel of mutations and variants in the cystic fibrosis transmembrane conductance regulator (CFTR) gene in genomic DNA samples isolated from human peripheral whole blood specimens. The panel includes mutations and variants recommended by the 2004 American College of Medical Genetics (ACMG). The eSensor® CF Genotyping Test is a qualitative genotyping test that provides information intended to be used for cystic fibrosis carrier screening as recommended by ACMG and the 2005 American College of Obstetricians and Gynecologists (ACOG) for adults of reproductive age, as an aid in newborn screening for cystic fibrosis, and in confirmatory diagnostic testing for cystic fibrosis in newborns and children. The test is not indicated for use in fetal diagnostic or pre-implantation testing. This test is also not indicated for stand-alone diagnostic purposes and results should be used in conjunction with other available laboratory and clinical information.

The eSensor® CF Genotyping Test is intended for use on the eSensor® XT-8 System.

Prescription Use X. (21 CFR Part 801 Subpart D) And/Or

Over the Counter Use (21 CFR Part 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE; CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of In Vitro Diagnostic Device Evaluation and Safety (OIVD)

Nu Ru

Division Sign-Off Office of In Vitro Diagnostic Device Evaluation and Safety

510(k) K090901

§ 866.5900 Cystic fibrosis transmembrane conductance regulator (CFTR) gene mutation detection system.

(a)
Identification. The CFTR gene mutation detection system is a device used to simultaneously detect and identify a panel of mutations and variants in the CFTR gene. It is intended as an aid in confirmatory diagnostic testing of individuals with suspected cystic fibrosis (CF), carrier identification, and newborn screening. This device is not intended for stand-alone diagnostic purposes, prenatal diagnostic, pre-implantation, or population screening.(b)
Classification. Class II (special controls). The special control is FDA's guidance document entitled “Class II Special Controls Guidance Document: CFTR Gene Mutation Detection System.” See § 866.1(e) for the availability of this guidance document.