K Number
K051435
Date Cleared
2006-01-19

(232 days)

Product Code
Regulation Number
866.5900
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The eSensor® Cystic Fibrosis Carrier Detection System is a device for the detection of carrier status for cystic fibrosis for all adult couples contemplating pregnancy, regardless of ethnicity. It is a qualitative genotyping assay that simultaneously detects mutations currently recommended by the American College of Medical Genetics and American College of Obstetricians and Gynecologists (ACMG/ACOG). The eSensor® CFCD System is not indicated for prenatal screening or to establish a diagnosis of cystic fibrosis.

Device Description

The eSensor® Cystic Fibrosis Carrier Detection System is an in vitro diagnostic test for the detection and genotyping of a selected panel of 23 cystic fibrosis mutations from DNA isolated from human whole blood.

The CFCD System is a clinical multiplex genetic test system which includes reagents for polymerase chain reaction amplification, exonuclease digestion and hybridization of target DNA, instrumentation and software. The CFCD System uses electrochemical detection to determine the carrier status of patient blood specimens for the ACOG/ACMG recommended panel of 23 cystic fibrosis mutations. Sample preparation for genotyping involves converting each blood specimen into purified genomic DNA (gDNA); then using multiplex PCR amplification followed by exonuclease digestion to convert the gDNA into a set of single-stranded targets. The genotyping reaction is set up with the combination of the single-stranded targets with appropriate buffers containing allele-specific signal probes differentially labeled with electrochemical signaling molecules, called ferrocenes. This mixture is then loaded into cartridges that contain single-stranded capture probes bound to an array of electrodes, with each electrode containing capture probes specific for a single mutation. Cartridges are inserted into the eSensor® 4800 Instrument where the single-stranded targets hybridize to the complementary sequences of the capture probes and signal probes. Detection of the target/probe complexes is achieved using alternating current voltammetry that generates specific electrical signals from the hybridized signal probes. The eSensor® DNA Detection System Application Software then classifies the signals from each mutation and generates a report for each specimen that describes the carrier or non-carrier status of each of the cystic fibrosis panel mutations.

AI/ML Overview

The Clinical Micro Sensors, Inc. eSensor® Cystic Fibrosis Carrier Detection System is a qualitative genotyping assay for the detection of carrier status for cystic fibrosis.

1. Table of Acceptance Criteria and Reported Device Performance

Performance CharacteristicAcceptance Criteria (Implicit from Results)Reported Device Performance
Input DNA Requirements
Accuracy (10 ng gDNA)Desired: High accuracy99.0% complete and accurate results (lower one-sided 95% confidence bound of 95.1%)
Accuracy (up to 1,200 ng DNA)Desired: Accurate results with increased DNA quantityAccurate results compared to DNA sequencing
Method Comparison
Per Sample Overall Agreement vs. DNA SequencingNot explicitly stated, but high agreement expected for equivalence.98.8% (479/486)
Per Sample Agreement vs. DNA Sequencing (excluding no-calls)Not explicitly stated, but high agreement expected.99.6% (479/481)
Per Mutation Overall Agreement vs. DNA SequencingNot explicitly stated, but high agreement expected.99.0% (11,061/11,178)
Per Mutation Agreement vs. DNA Sequencing (excluding no-calls)Not explicitly stated, but high agreement expected.99.98% (11,061/11,063)
Interfering SubstancesNo effect on DNA yield, amplification, or genotyping.No effect observed for all listed substances at specified concentrations.
Interfering MutationsCorrect genotyping or "no-call" for specific known interfering mutations.I506V, I507V, F508C genotyped as non-carriers for ΔF508 and ΔI507. F508C with ΔF508 did not affect ΔF508 identification. 2183AA>G genotyped as carriers for 2184delA. R117L gives "no-call" for R117H.
Reproducibility
Overall Per-Sample AgreementHigh agreement across sites and lots.99.8% (after re-testing no-call results)
Per-Sample No-Call RateLow.0.1% (after re-testing)
Per-Sample Contradictory Call RateVery low.0.1% (after re-testing)
Overall Per-Mutation AgreementHigh agreement across sites and lots.99.9%
Per-Mutation No-Call RateLow.0.1%
Per-Mutation Contradictory Call RateVery low.0.004%
System Failure (No-Call Rate)Low first-pass and final no-call rates.3.3% (first-pass), reduced to 1.0% (after up to 2 repeat tests)

2. Sample Sizes Used for the Test Set and Data Provenance

  • Method Comparison (Clinical Trial): 486 samples (freshly collected and banked).
    • Data Provenance: Not explicitly stated, but clinical trial suggests prospective and/or retrospective collection from a clinical setting. Country of origin not specified.
  • Reproducibility: Genomic DNA samples from one non-carrier cell line and 20 carrier cell lines.
  • Interfering Mutations: Not explicitly stated, likely a smaller, targeted set of DNA samples with these specific mutations.
  • Input DNA Requirements: 96 tests for 10 ng gDNA, and samples up to 1,200 ng DNA in method comparison studies.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

The ground truth was established by DNA sequencing, which is a laboratory method, not by experts. Therefore, the concept of "number of experts" and "qualifications of those experts" does not directly apply to the ground truth method.

4. Adjudication Method for the Test Set

Not applicable, as the ground truth was established by DNA sequencing, an objective laboratory method, not by human interpretation requiring adjudication.

5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study Was Done, If So, What Was the Effect Size of How Much Human Readers Improve with AI vs. Without AI Assistance

No, this device is a laboratory-based in vitro diagnostic system for genotyping, not an image-based AI system that would involve human readers or MRMC studies.

6. If a Standalone (i.e. algorithm only without human-in-the loop performance) Was Done

Yes, the "Method Comparison" study demonstrates the standalone performance of the eSensor® Cystic Fibrosis Carrier Detection System by comparing its results directly against DNA sequencing. The device's software classifies signals and generates a report without human interpretative input once the assay is run.

7. The Type of Ground Truth Used

The primary ground truth used for performance evaluation was DNA sequencing. For the "Interfering Mutations" section, the expected genotypes (based on known mutation characteristics) served as a form of ground truth.

8. The Sample Size for the Training Set

The document does not explicitly mention a distinct "training set" or its size. In a traditional genetic assay development context, internal assay optimization and validation would implicitly involve data used to refine parameters, but this is not typically referred to as a "training set" in the same way an AI/machine learning model would have one. The "Method Comparison" and "Reproducibility" data serve as the primary validation of the final device performance rather than a training set for an algorithm.

9. How the Ground Truth for the Training Set Was Established

As no explicit training set is detailed, this question is not fully applicable. However, for any internal development and optimization leading to the final assay, the establishment of ground truth would similarly involve established genetic testing methods, likely including DNA sequencing or synthetic DNA constructs with known mutations.

{0}------------------------------------------------

Clinical Micro Sensors, Inc. eSensor® Cystic Fibrosis Carrier Detection System 510(k) Notification

K 051435

JAN 1 9 2006

eSensor® Cystic Fibrosis Carrier Detection System

510(k) Summary

{1}------------------------------------------------

510(k) Summary

This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SDMA 1990 and CFR 807.92.

1. Submitter name, address, contact person and date prepared:

William A. Coty Clinical Micro Sensors, Inc. 757 South Raymond Ave. Pasadena, CA 91105 Phone: (626) 463-2000, ext. 8009 Fax: (626) 463-2012

Summary Prepared on January 5, 2006

2. Device Name:

eSensor® Cystic Fibrosis Carrier Detection System

Assay Component:

Device Proprietary/Trade Name: eSensor® Cystic Fibrosis Carrier Detection Kit

Common Name: CFCD Kit

Instrument Component:

Device Proprietary/Trade Name: eSensor® 4800 Instrument

Common Name: 4800 Instrument

3. Legally Marketed Equivalent Device Name

The eSensor® Cystic Fibrosis Carrier Detection System is equivalent to the Tm Bioscience Tag-It™ Cystic Fibrosis Kit 510(k) K043011, and to the Affymetrix GeneChip® Microarray Instrumentation System 510(k) K003664.

{2}------------------------------------------------

4. Intended Use of the Device

The eSensor® Cystic Fibrosis Carrier Detection (CFCD) System is a device for the detection of carrier status for cystic fibrosis for all adult couples contemplating pregnancy, regardless of ethnicity. It is a qualitative genotyping assay that simultaneously detects mutations currently recommended by the American College of Medical Genetics and American College of Obstetricians and Gynecologists (ACMG/ACOG). The eSensor® CFCD System is not indicated for prenatal screening or to establish a diagnosis of cystic fibrosis, and is for Rx only professional use within the confines of a licensed laboratory, as defined by the Clinical Laboratory Improvement Amendments (CLIA) of 1988.

5. Device Description

The eSensor® Cystic Fibrosis Carrier Detection System is an in vitro diagnostic test for the detection and genotyping of a selected panel of 23 cystic fibrosis mutations from DNA isolated from human whole blood.

The CFCD System is a clinical multiplex genetic test system which includes reagents for polymerase chain reaction amplification, exonuclease digestion and hybridization of target DNA, instrumentation and software. The CFCD System uses electrochemical detection to determine the carrier status of patient blood specimens for the ACOG/ACMG recommended panel of 23 cystic fibrosis mutations. Sample preparation for genotyping involves converting each blood specimen into purified genomic DNA (gDNA); then using multiplex PCR amplification followed by exonuclease digestion to convert the gDNA into a set of single-stranded targets. The genotyping reaction is set up with the combination of the single-stranded targets with appropriate buffers containing allele-specific signal probes differentially labeled with electrochemical signaling molecules, called ferrocenes. This mixture is then loaded into cartridges that contain single-stranded capture probes bound to an array of electrodes, with each electrode containing capture probes specific for a single mutation. Cartridges are inserted into the eSensor® 4800 Instrument where the single-stranded targets hybridize to the complementary sequences of the capture probes and signal probes. Detection of the target/probe complexes is achieved using alternating current voltammetry that generates specific electrical signals from the hybridized signal probes. The eSensor® DNA Detection System Application Software then classifies the signals from each mutation and generates a report for each specimen that describes the carrier or non-carrier status of each of the cystic fibrosis panel mutations.

{3}------------------------------------------------

6. Performance Characteristics:

Input DNA Requirements:

Samples containing 10 ng gDNA were tested in the CFCD System. Out of the total of 96 tests, 99.0% gave complete and accurate results, with a lower one-sided 95% confidence bound of 95.1%. Samples containing up to 1,200 ng DNA were tested in method comparison studies and gave accurate results in the CFCD System as compared to DNA sequencing.

Method Comparison:

Per sample: In Clinical Trial Method Comparison studies, 486 samples of freshly collected and banked samples were analyzed using the CFCD System and DNA sequencing. The overall agreement of the CFCD System compared to sequencing was 98.8% (479/486). Omitting samples which failed to give a valid call, agreement between the CFCD System and DNA sequencing was 99.6% (479/481).

Per mutation: Using the same data compiled for the per sample calculation above, 11.178 (486 x 23) individual mutations were analyzed. The overall agreement of the CFCD System compared to DNA sequencing was 99.0% (11,061/11,178). Omitting samples which failed to give a valid result, agreement between the CFCD System and DNA sequencing was 99.98% (11,061/11,063)..

Interfering Substances:

The following interfering substances were added separately to whole blood at the concentrations indicated, and no effect was observed on yield of extracted DNA, multiplex amplification of CFTR gene sequences, or genotyping of mutations in the CF carrier panel: triglycerides (3,000 mg/dL), high-density lipoprotein (70 mg/dL), cholesterol (250 mg/dL), bile salts (a mixture of cholate and deoxycholate: 6.4 ug/mL), human albumin (3 g/dL), human immunoglobulin G (3 g/dL), acetaminophen (30 ug/mL), ascorbic acid (30 ug/mL), diphenylhydantoin (phenytoin: 20 ug/mL), gentamicin (12 ug/mL), N-acetylsalicylic acid (200 ug/mL), nicotine (100 ug/mL), theophylline (20 ug/mL), valproic acid (100 ug/mL), vancomycin (100 ug/mL), NaCl (150 mM), KCl (5 mM), CaCl2 (1.08 mM), FeCl3 (9.25 uM).

{4}------------------------------------------------

Interfering Mutations:

AF508 reflex tests: DNA samples containing the non-disease causing polymorphisms I506V. I507V, and F508C, recommended for testing by the ACMG and ACOG in the event of unexpected homozygosity of AF508, have been tested internally with the CFCD System and found to genotype as non-carriers for both ΔF508 and ΔI507. The presence of the F508C polymorphism and the △F508 mutation in the same sample did not affect the identification of this sample as a ΔF508 carrier.

Samples heterozygous for the non-panel mutation 2183AA>G are genotyped as carriers for the panel mutation 2184delA. Samples heterozygous for the non-panel mutation R117L will give a "no-call" result for panel mutation R117H.

Reproducibility:

Reproducibility of the CFCD System was determined by testing genomic DNA samples from one non-carrier cell line and 20 carrier cell lines which together expressed all 23 panel mutations. Testing was performed at three sites with three lots of CFCD Assay Kits. Each panel of 21 cell line DNA samples was tested on separate days (5 days per kit lot per site). After re-testing of no-call results, overall per-sample agreement was 99.8%, with a no-call rate of 0.1% and a contradictory call rate of 0.1%. Overall per-mutation agreement was 99.9%, with a no-call rate of 0.1% and a contradictory call rate of 0.004%.

System Failure:

The first-pass no-call rate observed with the CFCD System in the Clinical Trial Method Comparison study of 486 samples was 3.3%, which was reduced to 1.0% after no more than 2 repeat tests for those samples which gave no-call results.

7. Conclusions

The eSensor® Cystic Fibrosis Carrier Detection System accurately identifies a selected panel of mutations in the CFTR gene from genomic DNA isolated from human whole blood.

{5}------------------------------------------------

Public Health Service

Image /page/5/Picture/2 description: The image shows a logo with a stylized depiction of an eagle's head and neck. The eagle is facing to the right, and its feathers are represented by three thick, curved lines. The logo is surrounded by a circular border with text that is difficult to read due to the image quality. The logo appears to be a symbol or emblem for an organization or institution.

Clinical Micro Sensors, Inc. c/o William A. Coty, Ph.D Director, Product Development 757 South Raymond Ave Pasadena, CA 91105

Food and Drug Administration 2098 Gaither Road Rockville MD 20850

JAN 1 9 2006

Re: K051435

Trade/Device Name: eSensor® Cystic Fibrosis Carrier Detection System Regulation Number: 21 CFR 866.5900 Regulation Name: CFTR (cystic fibrosis transmembrane conductance regulator) gene mutation detection system Regulatory Class: Class II Product Code: NUA Dated: May 31, 2005 Received: June 1, 2005

Dear Dr. Coty:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

{6}------------------------------------------------

Page 2 -

If you desire specific information about the application of labeling requirements to your device, or questions on the promotion and advertising of your device, please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (240) 276-0484. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html

Sincerely yours,

Robert Beckerf

Robert L. Becker, Jr., MD, PK Director Division of Immunology and Hematology Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health

Enclosure

{7}------------------------------------------------

Indications for Use

510(k) Number (if known): K051435

Device Name: eSensor® Cystic Fibrosis Carrier Detection System

Indications For Use:

The eSensor® Cystic Fibrosis Carrier Detection System is a device for the detection of carrier status for cystic fibrosis for all adult couples contemplating pregnancy. regardless of ethnicity. It is a qualitative genotyping assay that simultaneously detects mutations currently recommended by the American College of Medical Genetics and American College of Obstetricians and Gynecologists (ACMG/ACOG). The eSensor® CFCD System is not indicated for prenatal screening or to establish a diagnosis of cystic fibrosis.

Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR

Over-The-Counter Use (21 CFR 807 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD)

Mana Chen
Division Sign-Off

Office of In Vitro Diagnostic Device Evaluation and Safety

510(k) K051435

§ 866.5900 Cystic fibrosis transmembrane conductance regulator (CFTR) gene mutation detection system.

(a)
Identification. The CFTR gene mutation detection system is a device used to simultaneously detect and identify a panel of mutations and variants in the CFTR gene. It is intended as an aid in confirmatory diagnostic testing of individuals with suspected cystic fibrosis (CF), carrier identification, and newborn screening. This device is not intended for stand-alone diagnostic purposes, prenatal diagnostic, pre-implantation, or population screening.(b)
Classification. Class II (special controls). The special control is FDA's guidance document entitled “Class II Special Controls Guidance Document: CFTR Gene Mutation Detection System.” See § 866.1(e) for the availability of this guidance document.