K Number
K090629
Date Cleared
2009-06-04

(87 days)

Product Code
Regulation Number
862.1345
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Easy Step / DIACHEX PRO Blood Glucose Monitoring System is intended for use in the quantitatively measurement of glucose (sugar) in fresh capillary whole blood from the fingertip and the alternative sites: the palm and the forearm. The Easy Step / DIACHEX* PRO Blood Glucose Test Strips are for testing outside the body (in vitro . diagnostic use). The Easy Step / DIACHEX* PRO Blood Glucose Monitoring System is intended for use at home (over the counter [OTC]) by persons with diabetes, or in clinical setting by healthcare professionals as an aid in monitoring the effectiveness of diabetes control. It is not intended for the diagnosis of or screening for diabetes mellitus. The alternative site testing in the systems can be used only during steady-state blood glucose conditions. It is not intended for neonatal testing.

Device Description

The Easy Step / DIACHEX* PRO Blood Glucose Monitoring System designed by Tyson Bioresearch Inc., an amperometric biosensor, is adopted for its ease of use, its ability to process accurate results utilizing only a small volume of blood, and its quick response time. Easy Step / DIACHEX* PRO provide a convenient and safe monitoring system for diabetes health care professionals, hospitals and most importantly, people with diabetes.

The Easy Step / DIACHEX* PRO Blood Glucose Monitoring System is intended for use in the quantitatively measurement of glucose (sugar) in fresh capillary whole blood from the fingertip and the alternative sites: the palm and the forearm. The Easy Step / DIACHEX* PRO Blood Glucose Test Strips are for testing outside the body (in vitro diagnostic use). When the edge of the test strip is touched to a drop of blood, the test strip draws the blood into the sample chamber and the glucose reading is displayed on the meter after 5 seconds. The test measures glucose from 20 mg/dL (1.1mmol/L) to 600 mg/dL (33.3 mmol/L). The test strip is calibrated to display the equivalent of plasma glucose values to allow the comparison of results with laboratory methods.

AI/ML Overview

This submission pertains to the Easy Step / DIACHEX* PRO Blood Glucose Monitoring System.

1. Table of Acceptance Criteria and Reported Device Performance

The acceptance criteria for the device are based on the ISO 15197 standard for clinical accuracy.

Acceptance Criteria (ISO 15197)Reported Device Performance (Easy Step)Reported Device Performance (DIACHEX* PRO)Reported Device Performance (DIACHEX* INFINITY - Predicate)
95% of individual glucose results falling within ±15 mg/dL at glucose concentration for samples <75 mg/dL
95% of individual glucose results falling within ±20% at glucose concentrations ≥75 mg/dL (when compared to a laboratory reference)Finger: 98.4% (127/129) Palm: 100.0% (113/113) Forearm: 98.2% (111/113)Finger: 99.2% (128/129) Palm: 98.2% (111/113) Forearm: 96.5% (109/113)Finger: 100% (129/129) Palm: 98.2% (111/113) Forearm: 97.3% (110/113)

Additionally, the following
performance characteristics were evaluated:

  • Within-Day Precision:
    • Easy Step: CV% ranged from 2.08% to 5.80% across 5 glucose levels. Bias% ranged from -3.1% to 2.8%.
    • DIACHEX PRO:* CV% ranged from 2.17% to 5.08% across 5 glucose levels. Bias% ranged from -2.0% to 2.1%.
    • Predicate Device (DIACHEX INFINITY):* CV% ranged from 5.45% to 2.45% (two data points missing for the highest and lowest glucose levels). Bias% ranged from -3.4% to 2.6%.
  • Linearity (20 to 600 mg/dL):
    • Easy Step: Slope 1.01, Intercept -1.11, R$^2$ 0.9997, sy.x 5.45
    • DIACHEX PRO:* Slope 1.00, Intercept 0.41, R$^2$ 0.9996, sy.x 6.66
    • Predicate Device (DIACHEX INFINITY):* Slope 1.01, Intercept -0.23, R$^2$ 0.9998, sy.x 6.77
  • Clinical Accuracy (Fingertip vs. YSI):
    • Easy Step: N=129, Slope 1.03, Intercept 2.18, R$^2$ 0.9887
    • DIACHEX PRO:* N=129, Slope 1.01, Intercept 3.10, R$^2$ 0.9890
    • Predicate Device (DIACHEX INFINITY):* N=129, Slope 1.02, Intercept 1.44, R$^2$ 0.9871
  • Alternate Site Testing (Palm vs. YSI, Forearm vs. YSI, Palm vs. Finger, Forearm vs. Finger): Results consistently showed slopes close to 1, intercepts close to 0, and high R$^2$ values (>0.96) for all proposed and predicate devices.

2. Sample Size Used for the Test Set and Data Provenance

  • Clinical Accuracy (Primary Test Set):
    • Sample Size: 129 participants for fingertip testing. 113 participants for alternate site testing (palm and forearm).
    • Data Provenance: The study was an "in-house study" performed by technicians, implying it was conducted by the manufacturer or a contracted lab. The document does not specify the country of origin of the data beyond the manufacturer being in Taiwan. The study appears to be prospective, as it involved collecting samples from participants specifically for this evaluation.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

The ground truth for the test set was established using a YSI glucose analyzer (likely a YSI 2300 STAT Plus Glucose & Lactate Analyzer or similar laboratory reference method). There is no mention of experts (e.g., radiologists, pathologists) being used to establish the ground truth for this type of glucose monitoring device. The YSI system itself serves as the reference standard.

4. Adjudication Method for the Test Set

Not applicable. For blood glucose monitoring systems, the ground truth is typically a direct measurement from a highly accurate laboratory reference instrument (like the YSI analyzer), rather than a subjective assessment requiring expert adjudication.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

No, a multi-reader multi-case (MRMC) comparative effectiveness study was not presented. This type of study is usually conducted for imaging devices where human readers interpret results, and the AI's impact on human performance is evaluated. For blood glucose meters, the performance is assessed against a laboratory reference.

6. Standalone Performance Study

Yes, a standalone (algorithm only without human-in-the-loop performance) study was primarily conducted. The "Performance Characteristic Summary" section details the device's accuracy (precision, linearity, clinical accuracy) directly compared to the YSI reference method. The "clinical accuracy" study involved technicians taking readings, but the core evaluation is of the device's measurement against a golden standard.

7. Type of Ground Truth Used

The type of ground truth used was laboratory reference method data, specifically results obtained from a YSI glucose analyzer. This is considered the "gold standard" for glucose measurement in many clinical and research settings.

8. Sample Size for the Training Set

The document does not provide a specific sample size for a "training set." The submission describes performance characteristics of the finalized device. For IVD devices like blood glucose meters, the development process involves internal calibration and validation steps, but a distinct "training set" in the context of machine learning is not typically disclosed in 510(k) summaries, nor is it always applicable in the same way it would be for complex AI algorithms in image analysis.

9. How the Ground Truth for the Training Set Was Established

As mentioned above, information on a distinct "training set" is not provided. For the initial calibration and development of such a device, similar laboratory reference methods (e.g., YSI) would be used to establish the correct response for different glucose concentrations. The chemical formula and enzyme used in the test strip are identical to a previously cleared device (K062829), suggesting that the fundamental detection mechanism, which would have been rigorously characterized, remains unchanged.

{0}------------------------------------------------

K090629

510(k) SUMMARY

1. DATE PREPARED

February 20, 2009

JUN - 4 2009 2. SPONSOR INFORMATION

Address

TYSON BIORESEARCH. INC.

5 F., # 22, KE E. ROAD III., SCIENCE BASED INDUSTRIAL PARK CHUN-NAN, MIAO-LI COUNTY, CHINA (TAIWAN) 350

Contact Person: WEN-HAI TSAI

BHONE: 886-37-585988 FACSIMILE: 886-37-585996

3. NAME OF DEVICE:

Trade Name:

Easy Step Blood Glucose Monitoring System DIACHEX* PRO Blood Glucose Monitoring System

Common Names/Descriptions: Blood Glucose Monitoring System Classification Names: Glucose test system, product code 75CGA and "System, test, blood glucose, over the counter", product code 75NBW, 21 CFR 862.1345

4. DEVICE DESCRIPTION:

The Easy Step / DIACHEX* PRO Blood Glucose Monitoring System designed by Tyson Bioresearch Inc., an amperometric biosensor, is adopted for its ease of use, its ability to process accurate results utilizing only a small volume of blood, and its quick response time. Easy Step / DIACHEX* PRO provide a convenient and safe monitoring system for diabetes health care professionals, hospitals and most importantly, people with diabetes.

The Easy Step / DIACHEX* PRO Blood Glucose Monitoring System is intended for use in the quantitatively measurement of glucose (sugar) in fresh capillary whole blood from the fingertip and the alternative sites: the palm and the forearm. The Easy Step / DIACHEX* PRO Blood Glucose Test Strips are for testing outside the

{1}------------------------------------------------

body (in vitro diagnostic use). When the edge of the test strip is touched to a drop of blood, the test strip draws the blood into the sample chamber and the glucose reading is displayed on the meter after 5 seconds. The test measures glucose from 20 mg/dL (1.1mmol/L) to 600 mg/dL (33.3 mmol/L). The test strip is calibrated to display the equivalent of plasma glucose values to allow the comparison of results with laboratory methods.

5. INTENDED USE:

The Easy Step / DIACHEX* PRO Blood Glucose Monitoring System is intended for use in the quantitatively measurement of glucose (sugar) in fresh capillary whole blood from the fingertip and the alternative sites: the palm and the forearm. The Easy Step / DIACHEX* PRO Blood Glucose Test Strips are for testing outside the body (in vitro diagnostic use). The Easy Step / DIACHEX* PRO Blood Glucose Monitoring System is intended for use at home (over the counter [OTC]) by persons with diabetes, or in clinical setting by healthcare professionals as an aid in monitoring the effectiveness of diabetes control. It is not intended for the diagnosis of or screening for diabetes mellitus. The alternative site testing in the systems can be used only during steady-state blood glucose conditions. It is not intended for neonatal testing.

6. TEST PRINCIPLE

The test principle is based on electrochemical biosensor technology using glucose oxidase. Glucose is oxidized to gluconic acid and electrons are produced from the reaction. The electrons are then trapped by a chemical mediator, potassium ferricyanide. Once the enzymatic reaction is complete, a potential is provided by the meter for a further electrochemical reaction in order to generate a current from the release of trapped electrons. This current is then measured and correlated to the glucose concentration in the whole-blood sample. The test strip is calibrated to display the equivalent of plasma glucose values to allow easy comparison of results with laboratory methods.

7. PREDICATE DEVICE:

Predicate device name(s): DIACHEX* INFINITY Blood Glucose Monitoring System

Predicate 510(k) number(s): K073492

Comparison with predicate:

This 510(K) amendment addresses the changes of the DIACHEX* INFINITY Blood

{2}------------------------------------------------

Glucose Monitoring System (K073492). The modifications encompass the change for operating process, addition of selectable backlight on/off function, addition of graphic indicator for pre-meal and post-meal test results on the LCD result display of Easy Step / DIACHEX* PRO meter and the removal of hypoglycemic and hyperglycemic alarm setting function and the change for meter outside looking of Easy Step Glucose meter only. All the main internal electronic component, function and detection algorithm of proposed devices remain the same as DIACHEX* INFINITY Blood Glucose Meter. The chemical formula and used enzyme of the Easy Step / DIACHEX* PRO test strip is identical to the original cleared device test strip (K062829) without ant change. The table below lists the differences between the Easy Step and DIACHEX* PRO device from the predicate DIACHEX* INFINITY device.

ItemPredicate Device (K073492)Proposed Device
DIACHEX* INFINITYEasy StepDIACHEX* PRO
Hypoglycemic andhyperglycemic alarm2 user setting alarmsnone2 user setting alarms
BacklightnoneSelectable backlight on/off by user
LCD DisplayImage: DIACHEX INFINITY LCD DisplayIncrease pre/pro meal test result indicatorImage: DIACHEX PRO LCD Display
CodingFixed codeFixed codeGlucode chip
PC link modeEnter PC link mode by setting meterAutomatically enter PC link mode whenplugging in the USB cable
Memory modePress M button to read average and300 test result memoriesPress M button to read 300 test resultmemories, press S button to readaverage results
Meter AppearanceImage: DIACHEX INFINITY Meter AppearanceImage: Easy Step Meter AppearanceImage: DIACHEX PRO Meter Appearance
Meter Size92 x 58 x 19 (mm)79 x 60 x 17 (mm)92 x 58 x 19 (mm)
Meter WeightAppx. 60 gramsAppx. 55 gramsAppx. 60 grams

Differences:

{3}------------------------------------------------

8. PERFORMANCE CHARACTERISTIC SUMMARY

Within day precision test was performed with 5 levels of spiked whole blood each with one lot test strip. Samples were tested with 100 measurements obtained from 10 meters with each level of blood sample. Results are summarized below.

Unit:mg/dLYSIAVGBias %StdCV %
ProposedDeviceEasy Step43.842.6-2.82.475.80
87.885.1-3.12.923.43
132135.72.84.653.42
206209.21.54.982.38
320323.41.06.722.08
DIACHEX*PRO45.645.3-0.62.305.08
82.183.82.13.043.62
126124.5-1.23.763.02
224219.4-2.05.632.56
337334.3-0.87.252.17
PredicateDeviceDIACHEX*INFINITY39.740.72.62.225.45
92.489.3-3.42.45
136139.22.44.50
243240.0-1.25.49
354349.0-1.47.09

Linearity test was performed with 10 levels of spiked whole blood. Samples were tested with 10 measurements obtained from 10 meters with each level of blood sample using one lot test strip. The linear response range of proposed meter was defined from 20 to 600 mg/dL. and the actual studies were performed from ~25 to ~566 mg/dL. Linear regression results compared to YSI are summarized below.

DeviceProposed DevicePredicate Device
Easy StepDIACHEX* PRODIACHEX* INFINITY
Slope1.011.001.01
Intercept-1.110.41-0.23
R20.99970.99960.9998
sy.x5.456.666.77

To ensure that Easy Step, DIACHEX* PRO and DIACHEX* INFINITY system

{4}------------------------------------------------

similarly, clinical accuracy was assessed in an in-house study performing by technician.129 participants both males and females ranged from age with sample ranged from 33.2 to 577mg/dL and hematocrit ranged from 34% to 53%. Values obtained from meters were compared to YSI results; linear results regression analysis yielded the following results:

FingertipProposed DevicesPredicate Device
Easy StepDIACHEX* PRODIACHEX* INFINITY
N129129129
Slope1.031.011.02
Intercept2.183.101.44
R20.98870.98900.9871

The alternate site testing was also been evaluated.

Easy StepFinger vs YSIPalm vs YSIPalm vs FingerForearm vs YSIForearm vs Finger
N113113113113113
Slope1.031.010.981.010.97
Intercept2.993.061.920.810.43
R20.97970.97670.98710.97340.9753
DIACHEX*PROFinger vs YSIPalm vs YSIPalm vs FingerForearm vs YSIForearm vs Finger
N113113113113113
Slope1.021.010.991.031.00
Intercept4.923.020.58-2.72-3.97
R20.98150.97690.98560.97050.9659
DIACHEX*INFINITYFinger vs YSIPalm vs YSIPalm vs FingerForearm vs YSIForearm vs Finger
N113113113113113
Slope1.021.020.981.020.99
Intercept2.53-0.40-0.45-1.08-1.42
R20.97700.97630.97390.97590.9766

The acceptable criteria is the ISO 15197 requirement of 95% of individual glucose results falling within ±15mg/dL at glucose concentration for samples <75mg/dL and within ±20% at glucose concentrations≥75mg/dL. The samples that met the ISO 15197 requirement were summarized in the table below.

{5}------------------------------------------------

DeviceProposed devicePredicate device
Sample SiteEasy StepDIACHEX* PRODIACHEX* INFINITY
Finger98.4%(127 / 129)99.2%(128 / 129 )100%(129 / 129)
Palm100.0%(113 / 113)98.2%(111 / 113 )98.2%(111 / 113)
Forearm98.2%(111 / 113)96.5%(109 / 113 )97.3%(110 / 113)

From the above test results, we conclude that the modification of the DIACHEX* INFINITY Meter do not affect the effectiveness and safety of the device. The proposed device Easy Step / DIACHEX* PRO Blood Glucose Monitoring System is substantial equivalent to the original cleared device. The performance characteristics of the proposed system under various conditions including temperature effect, altitude effect, hematocrit levels, interference study, sensitivity and linearity has been established in the previously device cleared under K073492. Easy Step / DIACHEX* PRO system are suitable for its intended use. The submitted information in this premarket notification is complete and supports a substantial equivalence decision.

{6}------------------------------------------------

Image /page/6/Picture/0 description: The image shows the logo for the Department of Health & Human Services - USA. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized emblem featuring a symbol that resembles a caduceus or a staff with a snake winding around it.

DEPARTMENT OF HEALTH & HUMAN SERVICES

Food and Drug Administration 2098 Gaither Road Rockville MD 20850

Dr. Wen-Hai Tsai Tyson Bioresearch, Inc. 5F #22 Ke E. Road, III Science-Based Industrial Park Chun-Nan, Miao-Li County China (Taiwan) 350

K090629 Re:

Trade/Device Name: Easy Step/DIACHEX PRO Blood Glucose Monitoring System Regulation Number: 21 CFR 862.1345 Regulation Name: Glucose test system Regulatory Class: Class II Product Code: NBW, CGA Dated: May 4, 2009 Received: May 5, 2009

Dear Dr. Tsai:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act . or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820).

JUN - 4 2009

{7}------------------------------------------------

Page - 2

This letter will allow you to begin marketing your device as described in your Section 510/k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (240) 276-0450. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at (240) 276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at (240) 276-3464. For more information regarding the reporting of adverse events, please go to http://www.fda.gov/cdrh/mdr.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely yours,

At. H.

Courtney C. Harper, Ph.D. Acting Director Division of Chemistry and Toxicology Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health

Enclosure

{8}------------------------------------------------

Indication for Use

11090629 510(k) Number (if known):

Device Name: Easy Step / DIACHEX* PRO Blood Glucose Monitoring System

Indication for Use:

The Easy Step / DIACHEX PRO Blood Glucose Monitoring System is intended for use in the quantitatively measurement of glucose (sugar) in fresh capillary whole blood from the fingertip and the alternative sites: the palm and the forearm. The Easy Step / DIACHEX* PRO Blood Glucose Test Strips are for testing outside the body (in vitro . diagnostic use). The Easy Step / DIACHEX* PRO Blood Glucose Monitoring System is intended for use at home (over the counter [OTC]) by persons with diabetes, or in clinical setting by healthcare professionals as an aid in monitoring the effectiveness of diabetes control. It is not intended for the diagnosis of or screening for diabetes mellitus. The alternative site testing in the systems can be used only during steady-state blood glucose conditions. It is not intended for neonatal testing.

Prescription Use (21 CFR Part 801 Subpart D) And/Or Over the Counter Use (21 CFR Part 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE; CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of In Vitro Diagnostic Device Evaluation and Safety (OIVD)

signature

Division Sign-Off Office of In Vitro Diagnostic Device Evaluation and Safety

510(k) K090629

§ 862.1345 Glucose test system.

(a)
Identification. A glucose test system is a device intended to measure glucose quantitatively in blood and other body fluids. Glucose measurements are used in the diagnosis and treatment of carbohydrate metabolism disorders including diabetes mellitus, neonatal hypoglycemia, and idiopathic hypoglycemia, and of pancreatic islet cell carcinoma.(b)
Classification. Class II (special controls). The device, when it is solely intended for use as a drink to test glucose tolerance, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9.