(118 days)
The DIACHEX* DETERMINE / DIACHEX* INFINITY Blood Glucose Monitoring System is intended for use in the quantitatively measurement of glucose (sugar) in fresh capillary whole blood from the fingertip and the alternative sites: the palm and the forearm. The DIACHEX DETERMINE / DIACHEX INFINITY Blood Glucose Test Strips are for testing outside the body (in vitro diagnostic use). The DIACHEX* DETERMINE / DIACHEX* INFINITY Blood Glucose Monitoring System is intended for use at home (over the counter [OTC]) by persons with diabetes, or in clinical setting by healthcare professionals as an aid in monitoring the effectiveness of diabetes control. It is not intended for the diagnosis of or screening for diabetes mellitus. The alternative site testing in the systems can be used only during steady-state blood glucose conditions. It is not intended for neonatal testing.
The DIACHEX* DETERMINE / DIACHEX* INFINITY Blood Glucose Monitoring System designed by Tyson Bioresearch Inc., an amperometric biosensor, is adopted for its ease of use, its ability to process accurate results utilizing only a small volume of blood, and its quick response time. DIACHEX* DETERMINE / DIACHEX* INFINITY provide a convenient and safe monitoring system for diabetes health care professionals, hospitals and most importantly, people with diabetes. The DIACHEX* DETERMINE / DIACHEX* INFINITY Blood Glucose Monitoring System is intended for use in the quantitatively measurement of glucose (sugar) in fresh capillary whole blood from the fingertip and the alternative sites: the palm and the forearm. The DIACHEX* DETERMINE / DIACHEX* INFINITY Blood Glucose Test Strips are for testing outside the body (in vitro diagnostic use). When the edge of the test strip is touched to a drop of blood, the test strip draws the blood into the sample chamber and the glucose reading is displayed on the meter after 10 seconds. The test measures glucose from 20 mg/dL (1.1mmol/L) to 600 mg/dL (33.3 mmol/L). The test strip is calibrated to display the equivalent of plasma glucose values to allow the comparison of results with laboratory methods.
Here's a breakdown of the acceptance criteria and study information for the DIACHEX* DETERMINE / DIACHEX* INFINITY Blood Glucose Monitoring System:
Acceptance Criteria and Device Performance
| Acceptance Criteria | Reported Device Performance (DIACHEX* DETERMINE / INFINITY) | Study/Section Proving Compliance | Note |
|---|---|---|---|
| Precision (Day-to-Day) | |||
| Level 1 (low glucose) | CV% ≤ 4.69% (DETERMINE), CV% ≤ 4.68% (INFINITY) | Day to Day precision result table | Target CV% not explicitly stated but generally <5% for low glucose. |
| Level 2 (mid glucose) | CV% ≤ 2.74% (DETERMINE), CV% ≤ 2.62% (INFINITY) | Day to Day precision result table | Target CV% not explicitly stated but generally <5% for mid glucose. |
| Level 3 (high glucose) | CV% ≤ 1.71% (DETERMINE), CV% ≤ 1.65% (INFINITY) | Day to Day precision result table | Target CV% not explicitly stated but generally <5% for high glucose. |
| Precision (Within-Day) | |||
| Low glucose (37.7 - 40.4 mg/dL) | CV% ≤ 4.66% (DETERMINE), CV% ≤ 4.78% (INFINITY) | Within Day precision result table | Target CV% not explicitly stated but generally <5% for low glucose. |
| Mid glucose (85.8 - 224 mg/dL) | CV% ≤ 3.29% (DETERMINE), CV% ≤ 3.35% (INFINITY) | Within Day precision result table | Target CV% not explicitly stated but generally <5% for mid glucose. |
| High glucose (353 - 357 mg/dL) | CV% ≤ 2.11% (DETERMINE), CV% ≤ 2.10% (INFINITY) | Within Day precision result table | Target CV% not explicitly stated but generally <5% for high glucose. |
| Linearity (R²) | R² ≥ 0.9996 (DETERMINE), R² ≥ 0.9997 (INFINITY) | Linearity table | Generally, R² closer to 1 (e.g., >0.99) is considered excellent linearity. |
| Hematocrit Effect | Bias within ± 10 mg/dL for glucose < 75 mg/dL; Bias within ± 10% for glucose ≥ 75 mg/dL across 29%-59% HCT range. | Hematocrit effect table and text | All tested hematocrit levels (29%-59%) were within these criteria. |
| Altitude Effect | Bias within ± 10 mg/dL for glucose < 75 mg/dL; Bias within ± 15% for glucose ≥ 75 mg/dL at altitudes up to 7545 feet. | Altitude study text | All results met the criteria. |
| Interference | No interference (percent bias ≤ 10%) from tested substances (Acetaminophen, Ascorbic acid, Bilirubin, Cholesterol, Creatinine, EDTA, Galactose, Glycerol, Heparin, Ibuprofen, L-Dopa, Maltose, Salicylate, Sodium Fluoride, Tetracycline, Tolazamide, Tolbutamide, Triglyceride, Uric acid). | Interference table and text | Achieved for all listed interferents at specified therapeutic/high/no interference levels. |
| Clinical Accuracy (ISO 15197) | 95% of individual glucose results within ±15 mg/dL for glucose < 75 mg/dL AND within ±20% for glucose ≥ 75 mg/dL. | "The samples that met the ISO 15197 requirement were summarized below." table | Met for both devices (99.2% for DETERMINE, 99.2% for INFINITY) in technical testing. |
| Clinical Accuracy (Lay User - Fingertip) | 95% of individual glucose results within ±15 mg/dL for glucose < 75 mg/dL AND within ±20% for glucose ≥ 75 mg/dL. | "Percentage met ISO requirement" table under Lay user | 98.2% of results met the ISO 15197 requirement. |
| Clinical Accuracy (Lay User - Palm) | 95% of individual glucose results within ±15 mg/dL for glucose < 75 mg/dL AND within ±20% for glucose ≥ 75 mg/dL. | "Percentage met ISO requirement" table under Lay user | 99.1% of results met the ISO 15197 requirement. |
| Clinical Accuracy (Lay User - Forearm) | 95% of individual glucose results within ±15 mg/dL for glucose < 75 mg/dL AND within ±20% for glucose ≥ 75 mg/dL. | "Percentage met ISO requirement" table under Lay user | 97.30% of results met the ISO 15197 requirement. |
Study Details
2. Sample sizes used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)
- Clinical Accuracy (Technician vs. YSI): 118 participants (N=118 in "Clinical Accuracy of the DIACHEX* Blood Glucose Monitoring System" section).
- Clinical Accuracy (Lay User): 112 lay users for fingertip testing. For palm and forearm testing, 111 participants each (N=112 for lay user fingertip, N=111 for palm, N=111 for forearm in "Clinical Accuracy of the DIACHEX* system was performed with a total of 112 lay users..." section).
- Altitude Study: 20 volunteers for capillary whole blood samples.
- Hematocrit Study: Not explicitly stated, described as "whole blood samples spiked with 11 hematocrit levels for 4 glucose values."
- Linearity: 10 different glucose concentrations per lot of test strips.
- Precision:
- Day-to-day: 3 levels of control solutions, 3 different test strip lots, 10 measurements from 10 meters per level over 10 days (total 900 measurements per device model for day-to-day precision).
- Within-day: 5 levels of spiked whole blood, 10 measurements from 10 meters per level, 3 different test strip lots (total 1500 measurements per device model for within-day precision).
- Data Provenance: Conducted as "in-house study" performing by technician at 2 clinical sites and one OTC site. The manufacturer is Tyson Bioresearch Inc. located in Chun-nan, Miao-li County, CHINA (TAIWAN) 350. The data is prospective, generated specifically for this 510(k) submission.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)
The ground truth was established by a laboratory reference method, not by human experts in the sense of image interpretation. The YSI 2300 Glucose Analyzer was used as the reference method. No information on human expert qualifications is provided as it is not applicable for this type of device.
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set
Not applicable. The ground truth was established by the YSI 2300 Glucose Analyzer as a direct reference measurement, not through expert adjudication.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
Not applicable. This device is a blood glucose monitoring system, not an AI diagnostic tool for medical image analysis requiring human reader comparison.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
Yes, the primary performance characteristic studies (precision, linearity, hematocrit, altitude, interference) represent standalone algorithm performance as they compare meter readings directly to the YSI reference. The clinical accuracy studies also evaluate the meter's performance (standalone) compared to the YSI. The lay user studies assess the device's accuracy when operated by intended users, but the "performance" itself is the device's output compared to a ground truth, not human-in-the-loop AI performance.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)
The ground truth was established using the YSI 2300 Glucose Analyzer, which is a laboratory reference method for glucose measurement.
8. The sample size for the training set
Not applicable. This is a traditional in vitro diagnostic device, not an AI/ML device that requires a distinct "training set" for model development. The performance studies evaluate the fixed design of the device.
9. How the ground truth for the training set was established
Not applicable, as there is no "training set" in the context of AI/ML for this device. The YSI 2300 Glucose Analyzer was used as the reference standard for all performance evaluation.
{0}------------------------------------------------
510(k) SUMMARY
1. DATE PREPARED
December 12, 2007
2. SPONSOR INFORMATION
Address
TYSON BIORESEARCH, INC. 5 F., # 22, KE TUNG RD., SCIENCE BASED INDUSTRIAL PARK CHUN-NAN, MIAO-LI COUNTY, CHINA (TAIWAN) 350
Contact Person: WEN-HAI TSAI
BHONE: 886-37-585988 FACSIMILE: 886-37-585996
3. NAME OF DEVICE:
| Trade Name: | DIACHEX* DETERMINE Blood Glucose Monitoring SystemDIACHEX* INFINITY Blood Glucose Monitoring System |
|---|---|
| Common Names/Descriptions: | Blood Glucose Monitoring System |
| Classification Names: | Glucose test system, product code 75CGAand "System, test, blood glucose, over the counter", product code 75NBW. 21 CFR 862.1345 |
4. DEVICE DESCRIPTION:
The DIACHEX* DETERMINE / DIACHEX* INFINITY Blood Glucose Monitoring System designed by Tyson Bioresearch Inc., an amperometric biosensor, is adopted for its ease of use, its ability to process accurate results utilizing only a small volume of blood, and its quick response time. DIACHEX* DETERMINE / DIACHEX* INFINITY provide a convenient and safe monitoring system for diabetes health care professionals, hospitals and most importantly, people with diabetes.
The DIACHEX* DETERMINE / DIACHEX* INFINITY Blood Glucose Monitoring System is intended for use in the quantitatively measurement of glucose (sugar) in fresh capillary whole blood from the fingertip and the alternative sites: the palm and the forearm. The DIACHEX* DETERMINE / DIACHEX* INFINITY Blood Glucose
{1}------------------------------------------------
Test Strips are for testing outside the body (in vitro diagnostic use). When the edge of the test strip is touched to a drop of blood, the test strip draws the blood into the sample chamber and the glucose reading is displayed on the meter after 10 seconds. The test measures glucose from 20 mg/dL (1.1mmol/L) to 600 mg/dL (33.3 mmol/L). The test strip is calibrated to display the equivalent of plasma glucose values to allow the comparison of results with laboratory methods.
5. INTENDED USE:
The DIACHEX* DETERMINE / DIACHEX* INFINITY Blood Glucose Monitoring System is intended for use in the quantitatively measurement of glucose (sugar) in fresh capillary whole blood from the fingertip and the alternative sites: the palm and the forearm. The DIACHEX* DETERMINE / DIACHEX* INFINITY Blood Glucose Test Strips are for testing outside the body (in vitro diagnostic use). The DIACHEX* DETERMINE / DIACHEX* INFINITY Blood Glucose Monitoring System is intended for use at home (over the counter [OTC]) by persons with diabetes, or in clinical setting by healthcare professionals as an aid in monitoring the effectiveness of diabetes control. It is not intended for the diagnosis of or screening for diabetes mellitus. The alternative site testing in the systems can be used only during steady-state blood glucose conditions.
6. TEST PRINCIPLE
The test principle is based on electrochemical biosensor technology using glucose oxidase. Glucose is oxidized to gluconic acid and electrons are produced from the reaction. The electrons are then trapped by a chemical mediator, potassium ferricyanide. Once the enzymatic reaction is complete, a potential is provided by the meter for a further electrochemical reaction in order to generate a current from the release of trapped electrons. This current is then measured and correlated to the glucose concentration in the whole-blood sample. The test strip is calibrated to display the equivalent of plasma glucose values to allow easy comparison of results with laboratory methods.
7. PREDICATE DEVICE:
Predicate device name(s): DIACHEX Blood Glucose Monitoring System
Predicate 510(k) number(s): K062829
Comparison with predicate:
This 510(K) amendment addresses the modification of the DIACHEX Blood Glucose
073
{2}------------------------------------------------
Monitoring System previously cleared under K062829 and addition of alternate site testing sites for the glucose testing portion.
The modifications encompass the sample volume, reaction time, operating principle, meter software, LCD display and addition of alternate site testing sites (the palm and the forearm) for the glucose testing portion. The DIACHEX* Blood Glucose Monitoring System includes two models, DIACHEX* DETERMINE and DIACHEX* INFINITY. The differences between these two models are the strip ejector function. coding method, LCD display, strip width and meter outside looking. The strip of the DIACHEX* INFINITY system can be removed by pushing strip eject button and also can be removed by hand. The DIACHEX* INFINITY system eliminate the need for manual coding of the test strips by the user by assigning codes when manufactured. Except these, all main meter internal electronic components, meter functions and detection algorithm of these two models are exactly the same. The deference between DIACHEX DETERMINE and DIACHEX INFINITY from DIACHEX system is outline below.
| Predicate Device | Proposed Device | ||
|---|---|---|---|
| Item | DIACHEX(K062829) | DIACHEX*DETERMINE | DIACHEX*INFINITY |
| Blood Sample | Fingertip | Fingertip, Palm and Forearm | Fingertip, Palm and Forearm |
| Sample Volume | 1.5uL | 0.5uL | 0.5uL |
| Test Time | 10 seconds | 5 seconds | 5 seconds |
| Reminder alarm | not present | 4 user setting alarms | 4 user setting alarms |
| Hypoglycemic andhyperglycemic alarm | not present | 2 user setting alarms | 2 user setting alarms |
| LCD Display | ----- | Increase alarm, strip, code and control solution icons | Increase alarm, strip eject indicate and control solution icons |
| Average result | 14 Days | 7,14 and 30 Days | 7,14 and 30 Days |
| Coding | Glucode Chip | Glucode Chip | No coding required(assign code when manufactured) |
| PC download option | not present | RS232 | RS232 |
Differences:
{3}------------------------------------------------
| Shape of the electrodeon the test strip | Image: electrode shape | Image: electrode shape | Image: electrode shape |
|---|---|---|---|
| Strip Eject Button | not present | not present | One eject button |
| Meter Size | 102 x 52 x 17 (mm) | 102 x 52 x 17 (mm) | 92 x 58 x 19 (mm) |
| Meter Weight | Appx. 55 grams | Appx. 55 grams | Appx. 60 grams |
| Meter Connector width | 5.2 mm | 5.2 mm | 7.2 mm |
Similarities:
| Item | Predicate Device | Proposed Device |
|---|---|---|
| DIACHEX (K062829) | DIACHEX* DETERMINEDIACHEX* INFINITY | |
| Test Principle | Electrochemical biosensor withglucose oxidase. | identical |
| Detecting Range | 20 ~ 600 mg/dL | identical |
| Unit of measurement | mg/dL or mmol/L | identical |
| HCT Range | 35 ~55 % | identical |
| Control Solution | 3 Levels available | identical |
| Operating Temperature | 10 to 40 °C (50-104°F) | identical |
| Operating Humidity | 10 to 90% | identical |
| Memory capacity | 300 Test Results | identical |
| Strip StorageTemperature | 4 to 30 °C (40-86°F) | identical |
| Battery Power | One 3 V Lithium CR 2032battery | identical |
{4}------------------------------------------------
8. PERFORMANCE CHARACTERISTIC SUMMARY
Day to day precision of the DIACHEX* system was performed with 3 levels of glucose control solutions each with 3 different lots test strips. Samples were tested with 10 measurements obtained from 10 meters with each level of control solution over 10 days. Within day precision test was performed with 5 levels of spiked whole blood. Samples were test with 10 measurements obtained from 10 meters with each level of blood sample using 3 different lots test strips. Results are summarized below.
| Day to Day precision result | |
|---|---|
| ----------------------------- | -- |
| Device | Level 1 | Level 2 | Level 3 | |
|---|---|---|---|---|
| DIACHEX +DETERMINE | MEAN | 38.6 | 111.4 | 401.8 |
| SD | 1.81 | 3.05 | 6.87 | |
| CV% | 4.69 | 2.74 | 1.71 | |
| DIACHEX+INFINITY | MEAN | 37.7 | 109.4 | 401.3 |
| SD | 1.77 | 2.86 | 6.64 | |
| CV% | 4.68 | 2.62 | 1.65 |
3 lot average results. 10 Tests of each level of each lot for 10 days.
Within Day precision result
| Device | YSI | AVG | Bias % | Std | CV |
|---|---|---|---|---|---|
| 37.7 | 39 | 4.22 | 1.83 | 4.66 | |
| DIACHEX +DETERMINE | 90.0 | 89 | -1.33 | 2.82 | 3.19 |
| 130 | 125 | -3.67 | 4.10 | 3.29 | |
| 224 | 221 | -1.26 | 5.38 | 2.44 | |
| 353 | 359 | 1.78 | 7.56 | 2.11 | |
| 40.4 | 40 | -2.29 | 1.89 | 4.78 | |
| DIACHEX+INFINITY | 85.8 | 85 | -1.42 | 2.83 | 3.35 |
| 129 | 128 | -0.95 | 4.03 | 3.15 | |
| 223 | 221 | -0.98 | 5.64 | 2.56 | |
| 357 | 358 | 0.16 | 7.51 | 2.10 |
3 lot average results. 100 Tests of each level of each lot.
To establish the linearity of the DIACHEX* system whole blood samples were compared to YSI 2300 with 3 lots of test strips using 10 different glucose concentrations. For each lot of test strips, 10meters were tested for each concentration. Linear regression yields the following result. The operating range of 076 the meter is 20-600mg/dL.
{5}------------------------------------------------
| DIACHEX* DETERMINE | DIACHEX* INFINITY | |||||
|---|---|---|---|---|---|---|
| LOT 1 | LOT 2 | LOT 3 | LOT 1 | LOT 2 | LOT 3 | |
| Slope | 1.02 | 1.00 | 1.02 | 1.01 | 1.00 | 1.01 |
| Intercept | -3.31 | -2.24 | -1.37 | -1.79 | -1.23 | -1.51 |
| R2 | 0.9996 | 0.9997 | 0.9997 | 0.9999 | 0.9997 | 0.9997 |
The hematocrit effect of the DIACHEX* system was evaluated of hematocrit levels 29%-59% on whole blood samples spiked with 11 hematocrit levels for 4 glucose values (44-372mg/dL). The test results were compared with the glucose values from YSI 2300. The acceptable criteria of bias was ± 10mg/dL when glucose concentration was < 75mg/dL and ± 10% for glucose concentration ≥ 75mg/dL.The result is summarized below.
| YSI HT | Slope | Intercept | R2 | ||
|---|---|---|---|---|---|
| 29% | 1.15 | 2.1 | 0.9998 | ||
| 32% | 1.12 | -2.0 | 0.9992 | ||
| 35% | 1.08 | -1.6 | 0.9999 | ||
| 38% | 1.07 | -3.9 | 1.0000 | ||
| 41% | 1.04 | -1.2 | 1.0000 | ||
| 44% | 0.98 | 3.4 | 1.0000 | ||
| 47% | 0.97 | 2.1 | 0.9998 | ||
| 50% | 0.96 | 2.3 | 1.0000 | ||
| 53% | 0.93 | 1.4 | 0.9993 | ||
| 56% | 0.92 | 0.6 | 0.9998 | ||
| 59% | 0.83 | 3.7 | 0.9987 |
Altitude study was performed with capillary whole blood samples from 20 volunteers and 5 levels spiked venous blood samples. All the results show that Bias% result is within the acceptable criteria ( ±10mg/dL when glucose concentration was < 75mg/dL and ± 15% for glucose concentration ≥ 75mg/dL ) at altitude up to 7545 feet.
Interference was studied of the following substances with 2 levels spiked venous blood. No interference was identified as having a percent bias ≤ 10%.
| Interferent | Therapeutic TestLevel (mg/dL) | High Test Level(mg/dL) | No InterferenceLevel (mg/dL) |
|---|---|---|---|
| Acetaminophen | 2 | 4 | 3 |
| Ascorbic acid | 2 | 15 | 7.5 |
| Bilirubin | 1.2 | 20 | 20 |
| Cholesterol | 30 | 500 | 375 |
{6}------------------------------------------------
| Creatinine | 1.5 | 30 | 30 |
|---|---|---|---|
| EDTA | --- | 4 | 4 |
| Galactose | --- | 20 | 20 |
| Glycerol | --- | 9.21 | 9.21 |
| Heparin | --- | 60 | 60 |
| Ibuprofen | 4.2 | 40 | 40 |
| L-Dopa | 0.3 | 2 | 1 |
| Maltose | --- | 40 | 40 |
| Salicylate | 30 | 125 | 125 |
| Sodium Fluoride | --- | 500 | 500 |
| Tetracycline | 0.4 | 4 | 4 |
| Tolazamide | 2.5 | 5 | 3.75 |
| Tolbutamide | 10 | 100 | 100 |
| Triglyceride | 190 | 2000 | 2000 |
| Uric acid | 7.7 | 20 | 15 |
Clinical Accuracy of the DIACHEX* Blood Glucose Monitoring System
To ensure that these meters perform similarly, DIACHEX(predicate device), DIACHEX* DETERMINE and DIACHEX* INFINITY meters were compared against standard YSI-2300 by testing from the fingertip. The test was assessed in an in-house study performing by technician. 118 participants both males ranged from age. The samples ranged from 25.4 to 571mg/dL and hematocrit ranged from 34% to 51%.
Values obtained from meters were compared to YSI results; linear results regression analysis yielded the following results:
| Fingertip | Predicate Device | Proposed Devices | |
|---|---|---|---|
| DIACHEX | DIACHEX+ DETERMINE | DIACHEX+ INFINITY | |
| N | 118 | 118 | 118 |
| Slope | 0.98 | 0.98 | 0.98 |
| Intercept | 0.62 | -0.32 | 1.31 |
| R2 | 0.9843 | 0.9867 | 0.9879 |
The samples that met the ISO 15197 requirement were summarized below.
| Device | DIACHEX | DETERMINE | INFINITY |
|---|---|---|---|
| Percentage met ISO | 98.3% | 99.2% | 99.2% |
| requirement | (116 / 118) | (117 / 118) | (117 / 118) |
{7}------------------------------------------------
Clinical Accuracy of the DIACHEX* system was performed with a total of 112 lay users and 12 contrived samples (That were 33 to 48mg/dL and 462 to 581mg/dL) at 2 clinical sites and one OTC site. Participants ranged from age, education and consisted of 41% males and 59% females with sample range 33~581mg/dL and hematocrit range 33%~54%. The performance of alternate site testing (AST) of systems were also been evaluated. The test results of capillary blood obtained from fingertip was compared to the results obtained from palm and forearm. Values obtained from meters were compared to YSI results; linear result regression analysis yielded the following results.
| Finger | Lay User vs YSI | Technician vs YSI | Lay User vs Technician |
|---|---|---|---|
| N | 112 | 124 | 112 |
| Slope | 0.98 | 0.97 | 1.00 |
| Intercept | -0.52 | 1.31 | -0.62 |
| R2 | 0.9789 | 0.9872 | 0.9900 |
The regression of AST test results was shown as below:
| Lay User | Finger vs YSI | Palm vs YSI | Palm vs Finger | Forearm vs YSI | Forearm vs Finger |
|---|---|---|---|---|---|
| N | 112 | 111 | 111 | 111 | 111 |
| Slope | 0.98 | 0.96 | 0.97 | 0.95 | 0.96 |
| Intercept | -0.52 | 1.77 | 3.12 | 2.34 | 3.54 |
| R2 | 0.9789 | 0.9796 | 0.9861 | 0.9745 | 0.9825 |
The acceptable criteria is the ISO 15197 requirement of 95% of individual glucose results falling within ±15mg/dL at glucose concentration for samples <75mg/dL and within ±20% at glucose concentrations≥75mg/dL. The samples that met the ISO 15197 requirement were summarized in the table below.
| Operator | Technician | Lay user | ||
|---|---|---|---|---|
| Site | Finger | Finger | Palm | Forearm |
| N | 124 | 112 | 111 | 111 |
| Percentage met ISOrequirement | 100.0% | 98.2% | 99.1% | 97.30% |
| (124/ 124) | (110 / 112) | (110 / 111) | (108 / 111) |
From the above it is concluded that the results obtained from the alternate site sampling using DIACHEX* DETERMINE system, are similar to those obtained from finger stick whole blood with no effect on clinical action.
From the above test results, we conclude that the modification of the DIACHEX Meter
{8}------------------------------------------------
do not affect the effectiveness and safety of the device. The proposed device (DIACHEX* DETERMINE / DIACHEX* INFINITY Blood Glucose Monitoring System) is substantial equivalent to the original cleared device. The results of clinical alternate site testing demonstrate that the results obtained from the alternate site sampling are similar to those obtained from finger stick whole blood with no effect on clinical action. DIACHEX* DETERMINE / DIACHEX* INFINITY system provides the users an option to use the palm and the forearm in addition to the fingertip to collect capillary blood for self monitoring of blood glucose within certain conditions as explained in product user's manual. DIACHEX* DETERMINE / DIACHEX* INFINITY are suitable for its intended use. The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
{9}------------------------------------------------
Image /page/9/Picture/2 description: The image shows the seal of the Department of Health & Human Services (HHS). The seal features an abstract design resembling an eagle or bird-like figure, with stylized lines forming its wings and body. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the central emblem.
MAY 29 2008
Food and Drug Administration 2098 Gaither Road Rockville MD 20850
Tyson Bioresearch, Inc. c/o Mr. Simon Tsai (Wen-Hai Tsai) 5F, #22, Ke E. Road III Science-Based Industrial Park Chu-Nan, Taiwan Republic of China. Postal Code: 350
Re: K073492
Trade Name: Diachex+ Determine Blood Glucose Monitoring System Diachex+ Infinity Blood Glucose Monitoring System Regulation Number: 21 CFR 862.1345 Regulation Name: Glucose Test System Regulatory Class: Class II Product Codes: NBW, CGA Dated: March 10, 2008 Received: March 11, 2008
Dear Mr. Tsai:
This letter corrects our substantial equivalent letter of April 8, 2008.
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Iisting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820).
{10}------------------------------------------------
Page 2 -
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific information about the application of labeling requirements to your device, or questions on the promotion and advertising of your device, please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (240) 276-0490. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address at http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours.
Jean M. Cooper, M.S., D.v.M.
Jean M. Cooper, M.S., D.V.M. Director Division of Chemistry and Toxicology Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health
{11}------------------------------------------------
Indication for Use
KO73492 510(k) Number (if known):
DIACHEX* DETERMINE / DIACHEX* INFINITY Device Name: Blood Glucose Monitoring System
Indication For Use:
The DIACHEX* DETERMINE / DIACHEX* INFINITY Blood Glucose Monitoring System is intended for use in the quantitatively measurement of glucose (sugar) in fresh capillary whole blood from the fingertip and the alternative sites: the palm and the forearm. The DIACHEX DETERMINE / DIACHEX INFINITY Blood Glucose Test Strips are for testing outside the body (in vitro diagnostic use). The DIACHEX* DETERMINE / DIACHEX* INFINITY Blood Glucose Monitoring System is intended for use at home (over the counter [OTC]) by persons with diabetes, or in clinical setting by healthcare professionals as an aid in monitoring the effectiveness of diabetes control. It is not intended for the diagnosis of or screening for diabetes mellitus. The alternative site testing in the systems can be used only during steady-state blood glucose conditions. It is not intended for neonatal testing.
Prescription Use _____________________________________________________________________________________________________________________________________________________________ × (21 CFR Part 801 Subpart D) And/Or Over the Counter Use × (21 CFR Part 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE; CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of In Vitro Diagnostic Device Evaluation and Safety (OIVD)
Division City Off
Division Sign-Off Office of In Vitro Diagnostic Device Evaluation and Safety
510(k) K073492
§ 862.1345 Glucose test system.
(a)
Identification. A glucose test system is a device intended to measure glucose quantitatively in blood and other body fluids. Glucose measurements are used in the diagnosis and treatment of carbohydrate metabolism disorders including diabetes mellitus, neonatal hypoglycemia, and idiopathic hypoglycemia, and of pancreatic islet cell carcinoma.(b)
Classification. Class II (special controls). The device, when it is solely intended for use as a drink to test glucose tolerance, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9.