K Number
K090346

Validate with FDA (Live)

Date Cleared
2010-07-28

(532 days)

Product Code
Regulation Number
864.5220
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The ADVIA 2120/2120i auto-analyzer Body Fluid Application is an in vitro diagnostic test for the enumeration of the total nucleated cell (TNC) count and RBC count for pleural, and peritoneal dialysis (PD) specimens collected in K2 or K3 EDTA.

Device Description

The ADVIA 2120/2120i Body Fluid Application uses the Basophil/Lobularity and RBC/Platelet channels to enumerate the TNC and RBC counts. The TNC count is derived from the Basophil/Lobularity channel. The ADVIA 2120/2120i BASO reagent contains surfactant and phthalic acid which, in the presence of low heat in the Baso channel reaction chamber, lyses RBCs and strips the cytoplasmic membrane from all leukocytes except basophils. This cell suspension is subsequently passed through the flow cell. The cell suspension is intercented by light from the laser diode where the low-angle light scatter (2° to 3°) and high-angle light scatter (5° to 15°) signals of each cell are counted.

The RBC count is derived from the RBC/Platelet channel. The ADVIA 2120/2120i RBC/Platelet reagent uses sodium dodecyl sulfate and glutaraldehyde to sphere and fix the RBCs. This cell suspension uses the same flow cell and low-angle and high-angle light scatter signals as the Baso channel to count the RBCs. The system automatically reports the TNC and RBC counts in conventional or SI units appropriate for body fluid samples.

AI/ML Overview

Here's an analysis of the acceptance criteria and the study information based on the provided text for the ADVIA® 2120 systems Body Fluids Application:

1. Table of Acceptance Criteria and Reported Device Performance:

The provided document does not explicitly state specific acceptance criteria in terms of numerical thresholds (e.g., minimum accuracy percentages, correlation coefficients, or specific agreement ranges). Instead, it relies on a statement of substantial equivalence to a predicate device.

The document describes the device, its intended use, and compares its characteristics to the predicate device. The conclusion states that the device is "substantially equivalent" to the predicate. From a regulatory perspective, "substantial equivalence" implies that the new device performs as safely and effectively as a legally marketed predicate device without raising new questions of safety or effectiveness.

Therefore, the implied acceptance criterion is that the ADVIA 2120/2120i Body Fluids Application's performance for TNC and RBC enumeration in the specified body fluids should be comparable to the COULTER® LH 750 Body Fluids Application.

No specific performance metrics are reported in this summary, other than the general statement of substantial equivalence. The detailed performance data would typically be found in the full 510(k) submission, which this document summarizes.

Acceptance Criterion (Implied)Reported Device Performance
Substantial equivalence to the predicate device (COULTER® LH 750 Body Fluids Application) for enumeration of RBC and TNC (WBC) in body fluids.The ADVIA 2120/2120i Body Fluids Application is substantially equivalent to the COULTER® LH 750 Body Fluids Application for enumeration of RBC and TNC (WBC) in body fluids. (Stated in Section J. Conclusion)

2. Sample Size Used for the Test Set and Data Provenance:

The provided text does not specify the sample size used for the test set or the data provenance (e.g., country of origin, retrospective or prospective nature of the data). This information would be crucial for understanding the rigor of the testing and is typically found in the detailed study report within the 510(k) submission.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts:

The document does not mention the number of experts used to establish ground truth or their qualifications. The method for establishing ground truth for the test set is not described.

4. Adjudication Method for the Test Set:

The document does not provide any information regarding an adjudication method for the test set.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, and Its Effect Size:

The provided text does not describe a Multi-Reader Multi-Case (MRMC) comparative effectiveness study involving human readers with and without AI assistance. This device is an automated hematology analyzer, implying it aims to replace manual counting or provide automated results.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Study Was Done:

Yes, a standalone study was implicitly performed, as the device is an automated hematology analyzer designed to enumerate TNC and RBC counts directly. The entire summary focuses on the device's ability to perform these counts independently. The "Principles of the Procedure" describe the automated process by which the device enumerates cells using light scatter technology.

7. The Type of Ground Truth Used:

The document implies that the ground truth for evaluating the device's performance (though not explicitly detailed in this summary) would be based on comparison to a reference method, likely a manual microscopic count or another validated automated method, which would be considered the "expert consensus" or established laboratory practice for cell enumeration. The goal is to show the device performs comparably to such established methods. The predicate device's performance would also serve as a benchmark for "ground truth" in the comparative assessment for substantial equivalence.

8. The Sample Size for the Training Set:

The provided text does not specify the sample size used for the training set.

9. How the Ground Truth for the Training Set Was Established:

The document does not provide information on how ground truth was established for any training set. As an automated analyzer, the "training" (calibration, algorithm development) would typically involve internal validation against known standards, reference methods, and potentially a large set of characterized samples. However, this level of detail is not in the summary.

{0}------------------------------------------------

JUL 2 8 2040

510(k) Summary of Safety and Effectiveness for the

ADVIA® 2120 systems Body Fluids Application

This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92.

A. 510(k) Number: K090346

B. Date of Preparation: July 15, 2010

C. Proprietary and Established Names:

ADVIA® 2120 and 2120i Hematology Analyzer, Body Fluids Application

D. Applicant:

Siemens Healthcare Diagnostics Inc., 511 Benedict Ave, Tarrytown, NY 10591

Ernest Joseph, Senior Regulatory Affairs Specialist

Office: (914) 524-2431 Fax: (914) 524-2500

E. Regulatory Information:

ADVIA® 2120/2120i Body Fluids Application

    1. Regulation section: 21 CFR § 864.5220 Automated Differential Cell Counter
    1. Classification: Class II
    1. Product Code: GKZ
    1. Panel: Hematology

F. Predicate Device:

COULTER® LH 750 Body Fluids Application 510(K) # K030606

G. Device Description:

Summary and Explanation: Peritoneal and pleural fluids found in the serous membranes that line the peritoneal and pleural body cavities. Under normal conditions the amount of fluid in the serous membranes is very small. An effusion, the accumulation of fluid within the serous membranes, is a symptom of a pathologic process that may be transudative or exudative in origin.

{1}------------------------------------------------

Peritoneal Dialysis (PD) is a procedure that can be used to treat patients with renal disease. PD works on the principle that the mesothelial cells of the serous membrane lining the peritoneum function as a filter to remove the toxic substances that accumulate in blood. Peritonitis is a serious complication in patients who undergo PD.

Total Nucleated Cell (TNC) and RBC enumeration is a part of the clinical laboratory assessment of pleural, peritoneal, and PD fluids. The ADVIA 2120/2120i Body Fluid application TNC and RBC counts are to be used in conjunction with other diagnostic information and the attending healthcare professional's evaluation of the patient's condition.

Principles of the Procedure: The ADVIA 2120/2120i Body Fluid Application uses the Basophil/Lobularity and RBC/Platelet channels to enumerate the TNC and RBC counts. The TNC count is derived from the Basophil/Lobularity channel. The ADVIA 2120/2120i BASO reagent contains surfactant and phthalic acid which, in the presence of low heat in the Baso channel reaction chamber, lyses RBCs and strips the cytoplasmic membrane from all leukocytes except basophils. This cell suspension is subsequently passed through the flow cell. The cell suspension is intercented by light from the laser diode where the low-angle light scatter (2° to 3°) and high-angle light scatter (5° to 15°) signals of each cell are counted.

The RBC count is derived from the RBC/Platelet channel. The ADVIA 2120/2120i RBC/Platelet reagent uses sodium dodecyl sulfate and glutaraldehyde to sphere and fix the RBCs. This cell suspension uses the same flow cell and low-angle and high-angle light scatter signals as the Baso channel to count the RBCs. The system automatically reports the TNC and RBC counts in conventional or SI units appropriate for body fluid samples.

H. Intended Use:

The ADVIA 2120/2120i auto-analyzer Body Fluid Application is an in vitro diagnostic test for the enumeration of the total nucleated cell (TNC) count and RBC count for pleural, and peritoneal dialysis (PD) specimens collected in K2 or K3 EDTA.

I. Substantial Equivalence Information:

A comparison of the important similarities and differences between the device and the predicate is provided in the following tables:

Similarities
CharacteristicsCOULTER® LH 750 Body FluidsApplication (Predicate)Siemens ADVIA 2120/2120i BodyFluids Application
InstrumentAutomated Hematology AnalyzerSame as predicate device.
Intended Useenumeration of the total nucleatedcell (TNC) count and RBC countSame as predicate device.
DilutionAutomated dilutionSame as predicate device
CountsAutomated calculation of countsSame as predicate device
Differences
CharacteristicsCOULTER® LH 750 Body FluidsApplication (Predicate)Siemens ADVIA 2120/2120i BodyFluids Application

{2}------------------------------------------------

Sample TypePleural, peritoneal, and PeritonealDialysis fluids.Cerebrospinal fluid, serous fluid, andsynovial fluid.
DetectionImpedance technologyLight scatter technology

J. Conclusion:

The ADVIA 2120/2120i Body Fluids Application is substantially equivalent to the COULTER® LH 750 Body Fluids Application (predicate method) for enumeration of RBC and TNC (WBC) in body fluids.

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the seal of the Department of Health & Human Services - USA. The seal is circular and contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. In the center of the seal is an image of an eagle with its wings spread.

Food and Drug Administration 10903 New Hampshire Avenue Document Mail Center - WO66-0609 Silver Spring, MD 20993-0002

Siemens Healthcare Diagnostics, Inc c/o Mr. Ernest Joseph Senior Regulatory Affairs Specialist 511 Benedict Avenue Tarrytown, NY 10591

JUL 2 8 2010

Re: K090346

Trade/Device Name: Body Fluid Application for ADVIA 2120 and the ADVIA 2120i Regulation Number: 21 CFR 864.5220 Regulation Name: Automated differential cell counter Regulatory Class: Class II Product Code: GKZ Dated: June 25, 2010 Received: June 28, 2010

Dear Mr. Joseph:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). · You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into class II (Special Controls), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); and good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820). This letter will allow you to begin marketing your device as described in your Section 510(k) premarket

{4}------------------------------------------------

Page 2 - Mr. Ernest Joseph

notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Parts 801 and 809), please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (301) 796-5450. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely yours,

Reena Philip

Toy

Maria M. Chan, Ph.D. Director Division of Immunology and Hematology Devices Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

Indications for Use Form

510(k) Number (if known) __ K090346

Device Name: _Body Fluid Application for ADVIA 2120 and the ADVIA 2120i.

Indications for Use:

The ADVIA 2120/2120i auto-analyzer Body Fluid Application is an in vitro diagnostic test for the enumeration of the total nucleated cell (TNC) count and RBC count for pleural, peritoneal, and peritoneal dialysis (PD) specimens collected in K2 or K3 EDTA.

Prescription Use × (Part 21 CFR 801 Subpart D) AND/OR

Over-The-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD)

Reaching for Off

Division Sign-Off Office of In Vitro Diagnostic Device Evaluation and Safety

510(k) K090346

Page 1 of 1

§ 864.5220 Automated differential cell counter.

(a)
Identification. An automated differential cell counter is a device used to identify one or more of the formed elements of the blood. The device may also have the capability to flag, count, or classify immature or abnormal hematopoietic cells of the blood, bone marrow, or other body fluids. These devices may combine an electronic particle counting method, optical method, or a flow cytometric method utilizing monoclonal CD (cluster designation) markers. The device includes accessory CD markers.(b)
Classification. Class II (special controls). The special control for this device is the FDA document entitled “Class II Special Controls Guidance Document: Premarket Notifications for Automated Differential Cell Counters for Immature or Abnormal Blood Cells; Final Guidance for Industry and FDA.”