(78 days)
A patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.
The tested chemotherapy drugs are as follows: Amethopterin Hydrate (Methotrexate), Cisplatin, Cyclophosphamide (Cytoxan), Dacarbazine (DTIC), Doxorubicin Hydrochloride, Etoposide (Toposar), 5-Fluorouracil, Paclitaxel (Taxol), and Vincristine Sulfate.
Non sterile, Powder free Nitrile Examination Glove, Lavender Blue with Polymer coated, Tested for use with chemotherapy drugs.
The provided text describes the acceptance criteria and performance of a Non-sterile, Powder-free Nitrile Examination Glove, Lavender Blue with Polymer coated, Tested for use with chemotherapy drugs. This document is a 510(k) Summary.
Here's an analysis of the provided information, addressing your specific points:
1. A table of acceptance criteria and the reported device performance
| CHARACTERISTICS | STANDARDS Used for Acceptance Criteria | DEVICE PERFORMANCE |
|---|---|---|
| Dimensions | ASTM D 6319-00a-05 | Meets |
| Physical Properties | ASTM D 6319-00a-05 | Meets |
| Freedom from pinholes | ASTM D 6319-00a-05 | Meets |
| Powder Free Residue | ASTM D 6319-00a-05 | Meets |
| Biocompatibility: Primary Skin Irritation | Rabbits (method not specified in detail) | Passes |
| Biocompatibility: Guinea Pig Sensitization | Guinea Pig (method not specified in detail) | Passes |
| Permeation by Chemotherapy Drugs | ASTM D6978-05 | Passes for all 9 specified chemotherapy drugs |
Comment: The document outlines the specific ASTM standards used for various physical and chemical properties and indicates that the device met these standards or "passed" the tests.
Regarding the study that proves the device meets the acceptance criteria, and further points:
The provided document is a 510(k) summary, which focuses on demonstrating substantial equivalence to a predicate device based on non-clinical performance data. It does not describe a clinical study in the typical sense (e.g., involving human subjects, patient data, or a complex AI model evaluation). Instead, it presents laboratory test results against established standards.
Therefore, many of your requested points regarding clinical study design (sample size for test set, data provenance, expert ground truth, adjudication, MRMC, standalone performance, training set) are not applicable to this type of device submission and the information provided.
Let's address the relevant points based on the document's content:
2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)
- Sample Size: The document does not specify the exact number of gloves or samples tested for each characteristic (e.g., how many gloves were tested for pinholes, or how many animals were used for biocompatibility). The ASTM standards themselves would specify the minimum sample sizes for their respective tests, but this detail is not present in the 510(k) summary.
- Data Provenance: The testing was non-clinical (laboratory and animal studies). The document does not explicitly state the country where the tests were conducted, but the submitter is "Siam Sempermed Corp., Ltd" from Thailand, and the official correspondent is Sempermed USA Inc.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)
- Not Applicable. This device evaluation relies on objective measurements against engineering and biocompatibility standards, not on expert interpretation of medical images or clinical outcomes. The "ground truth" is defined by the technical specifications and test methods outlined in the ASTM standards.
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set
- Not Applicable. This concept is relevant for studies involving human interpretation or subjective assessments, often in clinical imaging. The testing here is based on objective laboratory measurements.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
- Not Applicable. This is a medical device, but not an AI-powered diagnostic or assistive tool for human readers. No MRMC study was conducted or is relevant here.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
- Not Applicable. This is not an algorithmic device. The device itself is the glove, and its performance is measured against physical and chemical standards.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)
- The "ground truth" for this device's performance is established by validated technical standards and test methods (e.g., ASTM D 6319-00a-05 for physical properties, ASTM D6978-05 for chemotherapy drug permeation, and established animal testing protocols for biocompatibility). The acceptance criteria are specified within these standards.
8. The sample size for the training set
- Not Applicable. This device does not involve a "training set" as it is not a machine learning or AI-based device.
9. How the ground truth for the training set was established
- Not Applicable. As there is no training set, this question is not relevant.
In summary, the provided 510(k) document details how the glove device meets established industry standards for physical properties, biocompatibility, and resistance to chemotherapy drug permeation through laboratory and animal testing. It's a non-clinical submission demonstrating substantial equivalence to a predicate device, and thus, many of the questions related to clinical trials, human expert evaluation, and AI/ML algorithms are not applicable.
{0}------------------------------------------------
510 (k) Summary
MAR 5 2009
As Required by 21 section 807.92 (c)
-
- Submitter Name: Siam Sempermed Corp., Ltd
110 Kanjanavanit Road. Pahtong Hatyai 2. Address: Songkhla. Thailand 90230
- Submitter Name: Siam Sempermed Corp., Ltd
-
(466) 74 291 648 3. Phone:
-
(466) 74 291 650 4. Fax:
-
న్. Contract Person: Mrs. Sureerat Choosri (Product Manager)
-
Date summary prepared: December 01,2008 6.
-
- Official Correspondent: Sempermed USA Inc.
13900 49th Street North 8. Address: Clearwater, USA , FL 33762
-
- Phone: 727 787 7250
- 727 787 7558 10. Fax:
- Mr. William E. Harris 11. Contact person:
-
- Device Trade or Proprietary Name: Non sterile, Powder free Nitrile Examination Glove, Lavender Blue with Polymer coated, Tested for use with chemotherapy drugs.
-
- Device Common or usual name: Examination glove
-
- Device Classification Name: Nitrile Patient Examination Glove (Powder Free(Polymer coated),Lavender Blue color).
15. Description of the Device:
Non sterile, Powder free Nitrile Examination Glove, Lavender Blue with Polymer coated, Tested for use with chemotherapy drugs.
16. Intended use of the device:
This device is a disposable device intended for medical purpose that is worn on the examiner 's hand to prevent contamination between patient and examiner.
The chemotherapy drugs tested are as follows: Amethopterin Hydrate (Methotrexate), Cisplatin, Cyclophosphamide (Cytoxan), Dacarbazine (DTIC), Doxorubicin Hydrochloride, Etoposide (Toposar), 5-Fluorouracil, Paclitaxel (Taxol), and Vincristine Sulfate.
17. Summary of The Technological Characteristics of The devices:
Non sterile, Powder free Nitrile Examination Glove, Lavender Blue with Polymer coated, Tested for use with chemotherapy drugs are summarized with the following technological characteristics compared to ASTM or equivalent standards.
| CHARACTERISTICS | STANDARDS | DEVICE PERFORMANCE |
|---|---|---|
| Dimensions | ASTM D 6319-00a-05 | Meets |
| Physical Properties | ASTM D 6319-00a-05 | Meets |
| Freedom from pinholes | ASTM D 6319-00a-05 | Meets |
| Powder Free Residue | ASTM D 6319-00a-05 | Meets |
| Biocompatibility | Primary Skin Irritation in Rabbits | Passes |
| Guinea Pig Sensitization | Passes |
{1}------------------------------------------------
| CHARACTERISTICS | STANDARDS | DEVICE PERFORMANCE |
|---|---|---|
| Permeation by ChemotherapyDrugs (ASTM D6978-05) | - | Passes for all chemicals as below :1.Amethopterin Hydrate (Methotrexate)2.Cisplatin3.Cyclophosphamide (Cytoxan)4.Dacarbazine (DTIC)5.Doxorubicin Hydrochloride6.Etoposide (Toposar)7.5-Fluorouracil8.Paclitaxel (Taxol)9. Vincristine Sulfate |
18. Substantial Equivalents Based on Assessment of Non-Clinical Performance Data
The performance test data of the non-clinical tests that support a determination of substantial equivalence is the same as mentioned immediately above.
19. Conclusion
It can be concluded that the Non sterile, Powder free Nitrile Examination Glove, Lavender Blue with Polymer coated, Tested for use with chemotherapy drugs will perform according to the glove performance standards referenced in section 17 above and meet ASTM standards, and FDA requirements. Consequently, this device is substantially equivalent to currently marketed device is safe and effective as the predicate device Siam Sempermed Nitrile, Examination Glove, Blue, Powder free(Polymer coated) ,Tested for use with chemotherapy drugs. Indeed, it is equivalent. This is better expressed in the tabulated comparison as below.
Technical comparison of specific elements is attached in the main submission.
| FDA file reference number | 510k number: K080520 |
|---|---|
| Attachments inside notification submission file | REFER TO APPENDIX 1 |
| TECHNOLOGICALCHARACTERISTICS | Comparison resultREFER TO ADDITIONAL TECHNICALCOMPARATIVE TABLE WITHIN 510KSUBMISSION |
| Indications for use | Identical |
| Target population | Identical |
| Design | Similar |
| Materials | Similar |
| Performance | Identical |
| Sterility | Identical (Not applicable) |
| Biocompatibility | Identical |
| Mechanical safety | Identical |
| Chemical safety | Identical |
| Anatomical sites | Identical |
| Human factors | Identical |
| Energy used and/or delivered | Identical (Not applicable) |
| Compatibility with environment andother devices | Identical |
| Where used | Identical |
| Standards met | Identical |
| Electrical safety | Identical (not applicable) |
| Thermal safety | Identical (not applicable) |
| Radiation safety | Identical (not applicable) |
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract symbol that resembles an eagle or bird in flight. The symbol is stylized and composed of curved lines.
5 2009 MAR
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Siam Sempermed Corporation, Limited C/o Mr. William E. Harris President & CEO Sempermed USA, Incorporated 13900 49th Street North Clearwater, Florida 33762
Re: K083755
Trade/Device Name: Non-Sterile, Powder-Free Nitrile Examination Glove, Lavender Blue with Polymer Coating, Tested for Use with Chemotherapy Drugs Regulation Number: 21 CFR 880.2650 Regulation Name: Patient Examination Glove Regulatory Class: I Product Code: LZA, LZC Dated: February 12, 2009 Received: February 13, 2009
Dear Mr. Harris:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{3}------------------------------------------------
Page 2 - Mr. Harris
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at 240-276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at 240-276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours.
Anthony D. Watson ba
Ginette Y. Michaud, M.D. Acting Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
{4}------------------------------------------------
Indications for Use
510(k) Number (if known): KO83755
Device Name: Non-Sterile, Powder-Free Nitrile Examination Glove, Lavender Blue with Polymer Coating, Tested for Use With Chemotherapy Drugs
Indications For Use: A patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.
The tested chemotherapy drugs are as follows: Amethopterin Hydrate (Methotrexate), Cisplatin, Cyclophosphamide (Cytoxan), Dacarbazine (DTIC), Doxorubicin Hydrochloride, Etoposide (Toposar), 5-Fluorouracil, Paclitaxel (Taxol), and Vincristine Sulfate.
Prescription Use (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Shule H. M. Mynhey, M.D.
Concurrence of CDRH, Office of Device Evaluation (ODE) (Division Sign-Off) Division of Anesthesiology, General Hospital Infection Control, Dental Devices
510(k) Number:
Page 1 of
c2
§ 880.6250 Non-powdered patient examination glove.
(a)
Identification. A non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. A non-powdered patient examination glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.(b)
Classification. Class I (general controls). The device, when it is a finger cot, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 880.9.