K Number
K083080
Date Cleared
2009-10-29

(378 days)

Product Code
Regulation Number
866.5775
Reference & Predicate Devices
N/A
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The IgX PLEX™ Rheumatoid Arthritis Qualitative Assay and SQiDworks™ Diagnostics Platform is an in vitro diagnostic test system for the qualitative detection of the IgA and IgM classes of Rheumatoid Factors, and the IgG class of anti-cyclic citrullinated peptide antibodies (CCPproprietary third generation equivalent formulation) in human serum specimens.

The IgX PLEXTM Rheumatoid Arthritis Qualitative Assay is intended for use in clinical laboratories as an aid in the diagnosis of Rheumatoid Arthritis in conjunction with other laboratory and clinical findings, and requires the SQiDworks™ Diagnostics Platform.

Device Description

The device consists of the IgX PLEX™ Rheumatoid Arthritis Qualitative Assay (RL1 kit) and the SQiDworks™ Diagnostics Platform (the Platform); the Platform incorporates the SQiDworks™ Integrated Software (the Software). The Platform is a multiplex immunoassay instrument that fully automates the process of a specific IgX PLEX™ Assay from serum transfer to reporting of all assay markers for each individual patient sample. Once the assay's biochemical reactions have completed, the instrument automatically performs a multi-color fluorescent scan of each well in the microarray, analyzes the data, and generates a report containing qualitative results for all assay markers. The SQiDworks Diagnostics Platform also includes numerous internal quality checks and user safety features with fail-safe and interlock mechanisms.

The instrument integrates an automated pipetting station, a fluorescent scanner, washing and drying stations, and other ancillary hardware components using dedicated instrument control. In addition, the software provides scheduling, self-verification, data acquisition, data management, analysis algorithms and reporting software.

Results for each patient sample from the IgX PLEXIM Rheumatoid Arthritis Qualitative Assay and the SQiDworks™ Diagnostics Platform are obtained simultaneously for cach of the three assay markers: RF IgM, RF IgA and CCP IgG using the results from one well containing one aliquot of the patient's serum. Results are reported independently.

The IgX PLEXIM Rheumatoid Arthritis Qualitative Assay (RL1) kit consists of two boxes (with different temperature requirements) of components as described below.

AI/ML Overview

The provided document describes the IgX PLEX™ Rheumatoid Arthritis Qualitative Assay and SQiDworks™ Diagnostics Platform. Here's a breakdown of the acceptance criteria and study details based on the available information:

1. Table of Acceptance Criteria and Reported Device Performance:

The document lists performance characteristics from nonclinical (in-house) studies. It describes ranges for reproducibility, clinical sensitivity, clinical specificity, and overall agreement with predicate devices rather than pre-defined acceptance criteria with specific thresholds. It implies these ranges demonstrate the safety and effectiveness for intended use.

Performance MetricAcceptance Criteria (Implied)Reported Device Performance
Reproducibility (RF IgA)Expected to be high for consistent results95.0%-100%
Reproducibility (RF IgM)Expected to be high for consistent results96.3%-100%
Reproducibility (CCP IgG)Expected to be high for consistent results83.3%-100%
Clinical Sensitivity (RF IgA)(Not explicitly stated, but consistent with literature references)Not explicitly stated for RF IgA, but overall ranged from 77.7% to 93.3%
Clinical Sensitivity (RF IgM)(Not explicitly stated, but consistent with literature references)93.3%
Clinical Sensitivity (CCP IgG)(Not explicitly stated, but consistent with literature references)77.7%
Clinical Specificity (RF IgA)(Not explicitly stated, but consistent with literature references)92.7%
Clinical Specificity (RF IgM)(Not explicitly stated, but consistent with literature references)Not explicitly stated for RF IgM, but overall ranged from 92.7% to 96.0%
Clinical Specificity (CCP IgG)(Not explicitly stated, but consistent with literature references)96.0%
Overall Agreement (RF IgA) with predicateExpected to show substantial equivalence85%
Overall Agreement (RF IgM) with predicateExpected to show substantial equivalenceNot explicitly stated for RF IgM, but overall ranged from 85% to 95%
Overall Agreement (CCP IgG) with predicateExpected to show substantial equivalence95%
InterferenceNo significant interference from common biological substancesNone of the analytes affected by high levels of bilirubin, hemoglobin, triglycerides, and human IgG

2. Sample Size Used for the Test Set and Data Provenance:

The document mentions "A series of nonclinical (in-house) studies were conducted," but it does not specify the sample size for the test set used in these studies. The data provenance is stated as "in-house" studies, implying it was conducted by the manufacturer, SQI Diagnostics Systems, Inc., which is based in Toronto, Ontario, Canada. It does not explicitly state whether the studies were retrospective or prospective.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts:

The document states that "clinical diagnosis (RA or non-RA, including normals) was used as the reference result" for clinical sensitivity and specificity. However, it does not specify the number or qualifications of experts who established this clinical diagnosis (ground truth). It also does not mention if experts were used for interpreting the device's results in the study.

4. Adjudication Method for the Test Set:

The document does not provide information on any adjudication method used for the test set, or for establishing the clinical diagnosis that served as ground truth.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

This device is an in vitro diagnostic test system that automates the process from serum transfer to reporting of assay markers. It does not involve human "readers" in the context of interpreting images or complex data that would typically necessitate an MRMC comparative effectiveness study with AI assistance. Therefore, no MRMC comparative effectiveness study involving human readers with and without AI assistance was done.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done:

Yes, the studies described are for the "IgX PLEX™ Rheumatoid Arthritis Qualitative Assay and SQiDworks™ Diagnostics Platform." The platform "fully automates the process... from serum transfer to reporting of all assay markers." The performance characteristics described (reproducibility, clinical sensitivity, specificity, agreement) refer to the performance of this automated system, i.e., the algorithm/device primarily in a standalone capacity. While the intended use requires it in conjunction with other laboratory and clinical findings, the reported performance metrics are for the device's direct output.

7. The Type of Ground Truth Used:

For clinical sensitivity and specificity, the ground truth was clinical diagnosis (RA or non-RA, including normals). For the study assessing overall agreement, the predicate test systems (Quanta LITE™ RF IgA ELISA, Quanta LITE™ RF IgM ELISA, and Quanta LITE™ CCP3 IgG ELISA) served as the reference for comparison.

8. The Sample Size for the Training Set:

The document does not provide information on the sample size used for any training set. Given the nature of a multiplex immunoassay instrument and assay, "training set" might refer to data used for establishing assay parameters, cutoff values, or calibration curves rather than machine learning model training in the typical sense. However, no specifics are provided.

9. How the Ground Truth for the Training Set Was Established:

The document does not mention a distinct "training set" or how its ground truth would have been established. It only discusses the ground truth for the clinical validation referenced in the performance characteristics section. For calibrators, it mentions they are "eight dilutions of a sample derived from human sera containing an appropriate representation of each of the analytes to be reported" and that "The assay standards (secondary standards) for RF (IgA and IgM) are traceable to the WHO/First British Standard 64/2 (primary standards)." For CCP IgG, it states "results are internally calculated in U/mL and are comparable to other assays on the market." This suggests a process of using established standards and internal calculations to define quantitative values, which are then converted to qualitative results based on assay-specific cutoff values.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows a logo for SQI Diagnostics. The logo consists of the letters "SQI" in a stylized font, with the word "DIAGNOSTICS" written vertically to the right of the "I". To the right of the word "DIAGNOSTICS" is a drawing of a drop of liquid, possibly blood.

36 Meteor Drive Toronto ON M9W 1A4 [w] 416.674.9500 [f] 416.674.9300 www.sqidiagnostics.com

SECTION 8 510(K) SUMMARY

OCT 2 9 2009

This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92.

The assigned 510(k) number is K083080.

807.92 (a)(1):

Name:SQI Diagnostics Systems, Inc.
Address:36 Meteor Drive
Toronto, Ontario
Canada M9W 1A4
Phone:416-674-9500 x239
FAX:416-674-9300
Contact:Ms. Kate Smith
Email:ksmith@sqidiagnostics.com
807.92 (a)(2):Device name- trade name and common name, and classification
Trade name:IgX PLEXTM Rheumatoid Arthritis Qualitative Assay andSQiDworksTM Diagnostics Platform
Common Name:Integrated system for the qualitative detection of IgA and IgMRheumatoid Factors and IgG antibodies to cyclic citrullinatedpeptides in serum samples
Classification:Class II
21 CFR 866.5775- Rheumatoid Factor Immunological System
21 CFR 862.2570- Instrumentation for Clinical Multiplex Test

{1}------------------------------------------------

Image /page/1/Picture/0 description: The image shows the logo for SQI Diagnostics. The logo consists of the letters "SQI" in a stylized font, with the word "DIAGNOSTICS" written vertically to the right of the "I". To the right of the text is a drawing of a water droplet.

36 Meteor Drive oronto ON M9W 1A4 416.674.9500 416.674.9300 www.sqidiagnostics.com

Identification of the legally marketed predicate device 807.92 (a)(3):

The test system is substantially equivalent to other ELISA test kits and microarray systems, specifically the Quanta LITE™ line of reagents (Inova Diagnostics, Inc., San Diego, CA), namely Quanta LITE™ RF IgA ELISA, Quanta LITE™ RF IgM ELISA, and Quanta LITE™ CCP3 IgG ELISA, and the Bioplex™ 2200 Multi-Analyte Detection System (the instrument, Bio-Rad Laboratories, Redmond, WA) respectively.

807.92 (a)(4): Device Description

The device consists of the IgX PLEX™ Rheumatoid Arthritis Qualitative Assay (RL1 kit) and the SQiDworks™ Diagnostics Platform (the Platform); the Platform incorporates the SQiDworks™ Integrated Software (the Software). The Platform is a multiplex immunoassay instrument that fully automates the process of a specific IgX PLEX™ Assay from serum transfer to reporting of all assay markers for each individual patient sample. Once the assay's biochemical reactions have completed, the instrument automatically performs a multi-color fluorescent scan of each well in the microarray, analyzes the data, and generates a report containing qualitative results for all assay markers. The SQiDworks Diagnostics Platform also includes numerous internal quality checks and user safety features with fail-safe and interlock mechanisms.

The instrument integrates an automated pipetting station, a fluorescent scanner, washing and drying stations, and other ancillary hardware components using dedicated instrument control. In addition, the software provides scheduling, self-verification, data acquisition, data management, analysis algorithms and reporting software.

Results for each patient sample from the IgX PLEXIM Rheumatoid Arthritis Qualitative Assay and the SQiDworks™ Diagnostics Platform are obtained simultaneously for cach of the three assay markers: RF IgM, RF IgA and CCP IgG using the results from one well containing one aliquot of the patient's serum. Results are reported independently.

The IgX PLEXIM Rheumatoid Arthritis Qualitative Assay (RL1) kit consists of two boxes (with different temperature requirements) of components as described below.

QuantityContainerVolume (each)Storage
Microarray plateOne (1)Foil package2-8°C
Reporter mixOne (1)Amber bottle15 mL≤-15°C
CalibratorsEight (8)2.0 mL tube400 μL≤-15°C
Positive Control #1One (1)2.0 mL tube80 μL≤-15°C
Positive Control #2One (1)2.0 mL tube80 μL≤-15°C

RL1 Kit Components

{2}------------------------------------------------

Image /page/2/Picture/0 description: The image shows the logo for SQI Diagnostics. The logo consists of the letters "SQI" in a large, bold font, followed by the word "DIAGNOSTICS" written vertically in a smaller font. To the right of the text is a drawing of a teardrop.

36 Meteor Drive ronto ON M9W 1A4 1416.674.9500 [] 416.674.9300 ww.saidiaanostics.com

Negative controlOne (1)2.0 mL tube80 μL≤-15°C
Sample diluentOne (1)Translucent bottle41 mL≤-15°C
Wash BufferConcentratesThree (3)Translucent bottle42 mL2-8°C
Documentation andbatch-specificcalibrationinformationOne (1) pershipmentCD-ROM

IgX PLEX RL1 Microarray Plate

The RL1 microarray plate consists of an array of protein and antibody replicate spots, covalently bound to the surface of coated glass within each well of a standard 96-well assay plate. All wells have the identical configuration of microarray elements, containing:

  • . Capture spots for Rheumatoid Factor Fc and CCP.
  • Internal normalization curve and control spots for internal consistency . confirmation.

The outer plate holder is labeled with a barcode indicating the assay type and lot number which is read by the platform.

Reporter Mix

The reporter mix is comprised of diluted fluorescently labeled marker antibodies. There is a marker antibody which is specific to each antibody of interest. In the case of the RL1 assay, the labeled markers are anti-human immunoglobulin antibodies (for example mouse o-hIgM). Each type of marker antibody is labeled with a dye in a different spectral range.

Calibrators/Standards

The external calibrators (standards) are eight dilutions of a sample derived from human sera containing an appropriate representation of each of the analytes to be reported. The assay standards (secondary standards) for RF (IgA and IgM) are traceable to the WHO/First British Standard 64/2 (primary standards) and internally calculated in IU/mL. The CCP IgG results are internally calculated in U/mL and are comparable to other assays on the market. The quantitative results are converted to qualitative results (positive and negative), based on assay-specific cutoff values. No. dilutions or reconstitutions are needed.

Controls

The positive and negative control samples are derived from human sera. The platform treats the controls as samples and does not perform any quality assessment based on the

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the logo for SQI Diagnostics. The logo consists of the letters "SQI" in a bold, serif font, with the word "DIAGNOSTICS" written vertically to the right of the letters. To the right of the word "DIAGNOSTICS" is a drawing of a drop of liquid.

w.sqidiagnostics.com

results. Although expected results for each applicable analyte of each control are provided, each lab is expected to follow their own quality procedures for assessing controls.

Sample Diluent

The sample diluent is a proprietary buffer mix optimized to the requirements of the platform.

Wash Buffer Concentrates

Specific buffer compositions for each wash procedure. The concentrate must be diluted prior to use.

CDROM

One CD is included in each shipment. The contents are a database of lot specific calibration data for each lot in the shipment and the documentation for the user's convenience.

Intended Use 807.92 (a)(5):

The IgX PLEX™ Rheumatoid Arthritis Qualitative Assay and SQiDWorks™M Diagnostics Platform is an in vitro diagnostic test system for the qualitative detection of the IgA and IgM classes of Rheumatoid Factors and the IgG class of anti-cyclic citrullinated peptide antibodies (CCP-proprietary third generation equivalent formulation) in human serum specimens.

The IgX PLEX™ Rheumatoid Arthritis Qualitative Assay is intended for use in clinical laboratories as an aid in the diagnosis of Rheumatoid Arthritis in conjunction with other laboratory and clinical findings, and requires the SQiDworks™ Diagnostics Platform.

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows the logo for SQI Diagnostics. The logo consists of the letters "SQI" in a stylized font, followed by the word "DIAGNOSTICS" written vertically. To the right of the text is an image of a blood droplet. The logo is black and white.

:

36 Meteor Drive
Toronto ON M9W 1A4
[w] 416.674.9500
[w] 416.674.9500 {f} 416.674.9300 www.sqidiagnostics.com

.

807.92 (a)(6): Technological Similarities and Differences to the Predicate

IgX PLEXTM Rheumatoid Arthritis Qualitative AssayPredicate(s)
Aspect/Feature/CharacteristicIgX PLEX RF IgAQUANTA LiteTM RF IgAK983084
Intended UseAs aid in diagnosis of rheumatoidarthritisSame
Sample MatrixSerumSame
MethodologyMicroarray-based fluorescentdetectionELISA
CalibrationOn each plateSame
Concentration DeterminationQualitativeSemi-Quantitative
Expected ValueCut-off is 12.0 IU/mLCut-off is 6 U/mL
Assay Substrate96-well microarray plates96-well microtiter plates
Multiplexed AssayYesNo
Aspect/Feature/CharacteristicIgX PLEX RF IgMQUANTA LiteTM RF IgMK971614
Intended UseAs aid in diagnosis of rheumatoidarthritisSame
Sample MatrixSerumSame
MethodologyMicroarray-based fluorescentdetectionELISA
CalibrationOn each plateSame
Concentration DeterminationQualitativeSemi-Quantitative
Expected ValueCut-off is 18.2 IU/mLCut-off is 6 U/mL
Assay Substrate96-well microarray plates96-well microtiter plates
Multiplexed AssayYesNo
Aspect/Feature/CharacteristicIgX PLEX Anti-CCP IgGQUANTA LiteTM CCP3 IgGK052264
Intended UseAs aid in diagnosis of rheumatoidarthritisSame
Sample MatrixSerumSame
MethodologyMicroarray-based fluorescentdetectionELISA
CalibrationOn each plateSame
Concentration DeterminationQualitativeSemi-Quantitative
Expected ValueCut-off is 11.7 U/mLCut-off is 20 U/mL
Assay Substrate96-well microarray plates96-well microtiter plates
Multiplexed AssayYesNo

. A R A B T T T T T T T T T T M T R . F . F . F . F . F . F . F . F . F . F . F . F . F . F . F . F . F . F . . . . . . . . . . . . . . . . .

And States Callery

13.80

..............................................................................................................................................................................

And Warry Contractions

2008

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image shows the logo for SQI Diagnostics. The logo features the letters "SQI" in a bold, serif typeface, with the word "DIAGNOSTICS" written vertically to the right of the "I". To the right of the text is a detailed illustration of a single drop of liquid, likely representing blood, which is commonly associated with medical diagnostics.

36 Meteor Drive Toronto ON M9W 1A4 [w] 416.674.9500 [f] 416.674.9300 www.sqidiagnostics.com

177 - 13

a

·

Aspect/Feature/CharacteristicSQiDworks MicroarrayDiagnostic PlatformBioPlex™ 2200 Multi-AnalyteDetection SystemK041658
Multi-analyteYesSame
Multiplexing MethodBased on multiplex microarraybased technologyBased on multiplex,bead-based technology
Detection TypeFluorescenceSame
Target of DetectionMicroarray SpotBead
DetectorMultichannel fluorescence CCDcamera scannerLaser monochromatic bead flowcytometry-like reader
Sample Handling and ProcessingAutomatedSame
Reagent StorageNo reagent storageOn-board, refrigeratedreagent storage
Internal ControlsBlank, serum verificationSame
Internal ControlsReaction condition verificationsNo
Assay Substrate96-well microarray platesIndividual cuvettes
CalibrationExternal calibratorsSame
Laboratory EnvironmentTraditional laboratoryenvironmentSame

.

·

்களை முறையும் பயன்படுத்தில் வெளியிட்டுள்ள விளைவிட்டு விடிக்கும் குறிக்கும் புக்கும் புக்கும் புக்கும் புக

{6}------------------------------------------------

Image /page/6/Picture/0 description: The image shows the logo for SQI Diagnostics. The logo consists of the letters "SQI" in a bold, serif font, with the word "DIAGNOSTICS" written vertically to the right of the "I". To the right of the text is a stylized image of a teardrop.

6 Meteor Drive onto ON M9W 1A4 416 674 9500 w.sqidiagnostics.com

Brief Description of Nonclinical and Clinical Data 807.92 (b)(1) and (b)(2):

A series of nonclinical (in-house) studies were conducted in support of the performance characteristics for the IgX PLEXTM Rheumatoid Arthritis Qualitative Assay and SQiDworks™ Diagnostics Platform. Those characteristics are defined as follows:

  • . Reproducibility ranges were 95.0%-100% for RF IgA, 96.3%-100% for RF IgM and 83.3%-100% for CCP IgG.
  • . Clinical sensitivity ranged from 77.7% (CCP IgG) to 93.3% (RF IgM) and clinical specificity ranged from 92.7% (RF IgA) to 96.0% (CCP IgG). This performance is consistent with literature references. In this study, clinical diagnosis (RA or non-RA, including normals) was used as the reference result.
  • Overall agreement between the analytes in the IgX PLEX™ Rheumatoid Arthritis . Qualitative Assay and established predicate test systems ranged from 85% (RF IgA) to 95% (CCP IgG).
  • . None of the analytes in the assay were affected by high levels of the following biological substances: bilirubin, hemoglobin, triglycerides and human IgG.

807.92 (b)(3): Conclusions from Nonclinical and Clinical Testing

Nonclinical and clinical testing was performed for the IgX PLEX™ Rheumatoid Arthritis Qualitative Assay and SQiDworks™ Diagnostics Platform. The test system was shown to be safe and effective for its intended use.

{7}------------------------------------------------

Image /page/7/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of the department's name arranged in a circular fashion around a symbol. The symbol is a stylized representation of a human figure with three flowing lines emanating from it, possibly representing health and well-being.

DEPARTMENT OF HEALTH & HUMAN SERVICES

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Mail Center - WO66-G609 Silver Spring, MD 20993-0002

SOI Diagnostics Systems c/o Ms. Kate Smith 36 Meteor Drive Toronto, ON Canada M9W 1A4

OCT 2 9 2009

Re: K083080

Trade/Device Name: IgX PLEX™ Rheumatoid Arthritis Qualitative Assay and SQiDworks™ Diagnostics Platform

Regulation Number: 21 CFR §866.5775 Regulation Name: Rheumatoid Factor Immunological System Regulatory Class: Class II Product Code: DHR, NHX, NHX, NSU Dated: October 9, 2009 Received: October 13, 2009

Dear Ms. Smith:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into class II (Special Controls), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or

{8}------------------------------------------------

Page 2 - Ms. Kate Smith

any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely yours,

ia m chen

Maria M. Chan, Ph.D. Director Division of Immunology and Hematology Devices Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health

Enclosure

{9}------------------------------------------------

INDICATIONS FOR USE

510(K) Number (if known): K083080

Device Name:

IgX Plex™ Rheumatoid Arthritis Qualitative Assay and SQiDworks™ Diagnostics Platform

Indications for Use:

The IgX PLEX™ Rheumatoid Arthritis Qualitative Assay and SQiDworks™ Diagnostics Platform is an in vitro diagnostic test system for the qualitative detection of the IgA and IgM classes of Rheumatoid Factors, and the IgG class of anti-cyclic citrullinated peptide antibodies (CCPproprietary third generation equivalent formulation) in human serum specimens.

The IgX PLEXTM Rheumatoid Arthritis Qualitative Assay is intended for use in clinical laboratories as an aid in the diagnosis of Rheumatoid Arthritis in conjunction with other laboratory and clinical findings, and requires the SQiDworks™ Diagnostics Platform.

Prescription Use (21 CFR Part 801 Subpart D)

And/Or

Over the Counter Use (21 CFR Part 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE; CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of In Vitro Diagnostic Device Evaluation and Safety (OIVD)

Reena Philip

Division Sign-Off Office of In Vitro Diagnostic Device Evaluation and Safety

510(k) K083080

§ 866.5775 Rheumatoid factor immunological test system.

(a)
Identification. A rheumatoid factor immunological test system is a device that consists of the reagents used to measure by immunochemical techniques the rheumatoid factor (antibodies to immunoglobulins) in serum, other body fluids, and tissues. Measurement of rheumatoid factor may aid in the diagnosis of rheumatoid arthritis.(b)
Classification. Class II (performance standards).