K Number
K082724

Validate with FDA (Live)

Date Cleared
2008-12-16

(90 days)

Product Code
Regulation Number
868.5895
Age Range
All
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The device is intended to provide continuous or intermittent mechanical ventilator support for the care of individuals who require mechanical ventilation. Specifically, the NEWPORT HT50 is applicable for adult and pediatric (i.e. infant, child and adolescent) patients, greater than or equal to 10 kg (22 lbs), who require the following general types of ventilatory support: positive pressure ventilation with Assist/Control, SIMV and SPONT/CPAP modes of ventilation.

The NEWPORT HT50 is a restricted medical device intended for use by qualified, trained personnel under the direction of a physician; it is suitable for use in hospital, sub-acute, emergency room, and home care environments, as well as for transport and emergency response applications.

Device Description

The modified NEWPORT HT50 Ventilator is an electrically powered, microprocessor controlled, variable flow generating ventilator which provides Assist Control, SIMV and SPONT/CPAP modes of ventilation in combination with servo-solenoidcontrolled, built-in PEEP.

The modified NEWPORT HT50 may be powered by external power (100-240 VAC or 12-30 VDC) or internal battery system power. Any time external power is connected to the ventilator, the internal battery system is charging.

AI/ML Overview

Here's an analysis of the acceptance criteria and study information for the NEWPORT HT50® Ventilator based on the provided text:

1. Table of Acceptance Criteria and Reported Device Performance

The submission describes acceptance criteria and reported device performance primarily in terms of battery life and functional integrity following the addition of a new battery system.

Acceptance CriteriaReported Device Performance
Dual PAC™ Internal Battery System Functionality:
Correct switching to backup battery at target voltage.Verified (tested)
Proper charging circuit function.Verified (tested)
Ventilator operation on main and backup battery.Verified (tested)
Proper warnings regarding battery status.Verified (tested)
Proper switching to backup battery if main battery is defective.Verified (tested)
Proper warning to user if backup battery is defective.Verified (tested)
Low Battery alarm activation (message, LED, audible).Confirmed, indicates minimum 30 minutes remaining.
Minimum 30 minutes of combined battery power after low battery alarm before shutdown.Confirmed, provides minimum 30 minutes.
Battery Empty alarm activation (message, LED, audible).Confirmed, indicates minimum 15 minutes remaining.
Minimum 15 minutes of combined battery power after battery empty alarm before shutdown.Confirmed, provides minimum 15 minutes.
Re-Verification of All Functions and Alarms:
Proper functionality and alarming for all ventilator operations (to ensure no inadvertent impact from software revisions).Confirmed (complete re-verification performed).
EMC and Safety Testing:
Meets electrical safety requirements with new battery pack.Verified (tested).
Meets environmental requirements with new battery pack.Verified (tested).
Meets EMC requirements with new battery pack.Verified (tested).
Transport Use Standards:
Compliance with MIL-STD-810E Environmental Test Methods.Verified (testing based on this standard).
Compliance with IEC 68-2 Environmental Testing.Verified (testing based on this standard).

2. Sample Size Used for the Test Set and Data Provenance

The document does not specify a distinct "test set" in the context of patient data or clinical trials. Instead, the testing described appears to be primarily engineering verification and validation of the device itself and its new battery system.

  • Sample Size: Not applicable in the traditional sense of patient data. The "sample" would be the device itself (presumably one or more units).
  • Data Provenance: The testing was integral to the device development and modification, suggesting it was conducted prospectively by the manufacturer (Newport Medical Instruments, Inc.). There is no mention of country of origin for external data as no such data was used.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

This section is not applicable as the described testing did not involve establishing ground truth from patient data with expert review. The "ground truth" here would be the defined engineering specifications and regulatory standards that the device performance was measured against. The experts involved would be the engineers and quality assurance personnel at Newport Medical Instruments responsible for designing, testing, and verifying the device's compliance. Their qualifications are not explicitly stated in the submission but would include expertise in ventilator design, electrical engineering, software engineering, and regulatory compliance.

4. Adjudication Method for the Test Set

This is not applicable as there was no expert review or consensus-based ground truth establishment using patient data. The "adjudication" in this context would be the internal review and approval processes by the manufacturer's engineering and quality teams to confirm that the test results met the predefined specifications and standards.

5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study Was Done, If So, What Was the Effect Size of How Much Human Readers Improve with AI vs Without AI Assistance

No, an MRMC comparative effectiveness study was not done. This submission is for a continuous ventilator, not an AI-assisted diagnostic or interpretative device that would typically involve human readers.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

This is not applicable as the device is a mechanical ventilator, not an algorithm, and its operation inherently involves a human operator (qualified, trained personnel under the direction of a physician). The performance testing was for the device's physical and functional capabilities, not an algorithm's standalone accuracy.

7. The Type of Ground Truth Used

The ground truth used for this submission is based on:

  • Engineering Specifications: Designed performance parameters for the ventilator and its DUAL PAC™ Internal Battery system (e.g., specific voltage switching points, charging circuit functionality, alarm thresholds, and minimum battery run times).
  • Regulatory Standards: Compliance with relevant national and international standards for medical devices, electrical safety, environmental testing, and EMC (e.g., MIL-STD-810E, IEC 68-2).
  • Predicate Device Performance: The modified device's performance was also assessed against the established safety and effectiveness of its predicate devices.

8. The Sample Size for the Training Set

This is not applicable. The device is a mechanical ventilator, not a machine learning algorithm that requires a "training set" of data.

9. How the Ground Truth for the Training Set Was Established

This is not applicable for the same reason as above; there is no training set mentioned or implied for a mechanical ventilator.

{0}------------------------------------------------

K082724
page 1 of 3

Newport
MEDICAL

DEC 1 6 2008

510(k) Summary NEWPORT HT50® Ventilator

This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92.

I.Submitter Information:Newport Medical Instruments, Inc.1620 Sunflower AvenueCosta Mesa, CA 92626
Contact Person:Dana RodriguezVice President, RAQA
Summary Date:16 December 2008
II.Device Name
Proprietary:NEWPORT HT50® VentilatorModels HT50-H, HT50-HB, HT50-H1, and HT50-H1B
Common:Continuous Ventilator
Classification:II
Product Code:CBK
CFR Section:868.5895

Predicate Device III.

The predicate devices for modified NEWPORT HT50 Ventilator are:

IV. Device Description

The modified NEWPORT HT50 Ventilator is an electrically powered, microprocessor controlled, variable flow generating ventilator which provides Assist Control, SIMV and SPONT/CPAP modes of ventilation in combination with servo-solenoidcontrolled, built-in PEEP.

{1}------------------------------------------------

The modified NEWPORT HT50 may be powered by external power (100-240 VAC or 12-30 VDC) or internal battery system power. Any time external power is connected to the ventilator, the internal battery system is charging.

V. Intended Use

This modified NEWPORT HT50 is intended to provide continuous or intermittent mechanical ventilator support for the care of individuals who require mechanical ventilation. Specifically, the NEWPORT HT50 is applicable for adult and pediatric (i.e. infant, child and adolescent) patients, greater than or equal to 10 kg (22 lbs), who require the following general types of ventilatory support: positive pressure ventilation with Assist/Control, SIMV and SPONT/CPAP modes of ventilation.

The modified NEWPORT HT50 is a restricted medical device intended for use by qualified, trained personnel under the direction of a physician; it is suitable for use in hospital, sub-acute, emergency room, and home care environments, as well as for transport and emergency response applications.

Technological Characteristics VI.

The modified NEWPORT HT50 combines a secondary, internal, back-up Nickel-Metal Hydride (NiMH) battery with an independent charging system with the existing Lead Acid (LA) internal battery and charging system, thereby creating the DUAL PAC™ Internal Battery system. When the primary internal LA battery charge level reaches the low battery threshold, the Low Battery alarm activates (message, LED indicator and audible alarm). The unit will automatically add in the power of the secondary NiMH battery and both batteries will continue to power the ventilator and the alarms for a minimum of 30 minutes before shut down. This alarm message (Low Battery), the blinking LED indicator and audible alarm give notice that a minimum of 30 minutes remains to find an alternate power source before the unit will shutdown. When the combined battery power reaches the battery empty threshold, a Battery Empty alarm is activated (message, LED indicator and audible alarm) alerting the user that there is a minimum of 15 minutes to shut down. Ensuring these defined time periods for battery life increases the margin of safety for battery usage.

VII. Device Testing

Testing supporting transport use of the device was based on the following standards:

  • MIL-STD-810E Environmental Test Methods and Engineering Guidelines .
  • IEC 68-2 Environmental Testing .

Testing for the DUAL PAC Internal Battery design encompassed validation and verification of the new design, re-verification of all functional and alarm aspects, and EMC and safety testing; this testing is summarized following:

  • A. Validation and Verification of New Design
    The battery pack and charging circuits were tested to verify functionality, including:

  • Correct switching to back-up battery at target voltage; .

  • Proper charging circuit function; .

  • Ventilator operation on main battery and back up battery; .

{2}------------------------------------------------

  • Verification of proper warnings; .
  • Verification of proper switching to back up battery if main battery is defective; and
  • Verification of proper warning to user in the event that back up battery . is defective.

Re-Verification of All Functions and Alarms B.

The NEWPORT HT50 was tested to ensure proper functionality and alarming for ventilator operations. This complete re-verification of all functions/alarms assured that software revisions associated with the DUAL PAC Internal Battery system did not inadvertently affect any ventilator function or alarm.

C. EMC and Safety Testing

Testing was also performed to assure that the ventilator with the new battery pack configuration meets electrical safety, environmental, and EMC requirements.

Clinical and Animal Testing VIII.

No clinical or animal data were included in this submission.

IX. Conclusions

All testing demonstrates that the modified NEWPORT HT50 Ventilator performs as intended and is therefore suitable for its intended use.

{3}------------------------------------------------

Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle symbol on the right and the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged in a circular fashion around the left side. The eagle symbol is composed of three thick, curved lines representing the body and wings, with a smaller curved line at the bottom representing the tail.

Public Health Service

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

DEC 1 6 2008

Ms. Dana Rodriguez Vice President of Regulatory Affairs and Quality Assurance Newport Medical Instruments, Incorporated 1620 Sunflower Avenue Costa Mesa. California 92626

Re: K082724

Trade/Device Name: NEWPORT HT50® Ventilator Models HT50-H, HT50-HB, HT50-H1, and HT50-H1B

Regulation Number: 21 CFR 868.5895 Regulation Name: Continuous Ventilator Regulatory Class: II Product Code: CBK, NOU Dated: September 15, 2008

Received: September 17, 2008

Dear Ms. Rodriguez:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{4}------------------------------------------------

Page 2 - Ms. Rodriguez

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807): labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at 240-276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at 240-276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely vours.

Snyte Y. Mckee Omd.

Chiu Lin. Ph.D. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

{5}------------------------------------------------

Indications for Use

510(k) Number (if known): K082724

NEWPORT HT50® Ventilator Device Name: Models HT50-H, HT50-HB, HT50-H1, and HT50-H1B

Indications for Use:

The device is intended to provide continuous or intermittent mechanical ventilator support for the care of individuals who require mechanical ventilation. Specifically, the NEWPORT HT50 is applicable for adult and pediatric (i.e. infant, child and adolescent) patients, greater than or equal to 10 kg (22 lbs), who require the following general types of ventilatory support: positive pressure ventilation with Assist/Control, SIMV and SPONT/CPAP modes of ventilation.

The NEWPORT HT50 is a restricted medical device intended for use by qualified, trained personnel under the direction of a physician; it is suitable for use in hospital, sub-acute, emergency room, and home care environments, as well as for transport and emergency response applications.

Prescription Use (Part 21 CFR 801 Subpart D)

AND/OR

Over-The-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Zu. M. Zi.

(Division Sign-Off) Division of Anesthesiology, General Hospital Infection Control, Dental Devices

Page 1 of

KD82774 510(k) Number:

§ 868.5895 Continuous ventilator.

(a)
Identification. A continuous ventilator (respirator) is a device intended to mechanically control or assist patient breathing by delivering a predetermined percentage of oxygen in the breathing gas. Adult, pediatric, and neonatal ventilators are included in this generic type of device.(b)
Classification. Class II (performance standards).