K Number
K082462
Date Cleared
2009-02-06

(163 days)

Product Code
Regulation Number
862.1665
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The GemLyte is an automated, microprocessor-controlled analyzer which utilizes ion-selective electrodes for the measurement of sodium, potassium, chloride, calcium and lithium in whole blood, plasma and serum, and prediluted urine samples.

The GemLyte Potassium Assay is intended to measure potassium in whole blood, plasma, serum, and urine on the GemLyte Electrolyte Analyzer. Measurements obtained by this device are used to monitor electrolyte balance in the diagnosis and treatment of diseases conditions characterized by low or high blood potassium levels.

The GemLyte Sodium Assay is intended to measure sodium in whole blood, plasma, serum, and urine on the GemLyte Electrolyte Analyzer. Measurements obtained by this device are used to monitor electrolyte balance in the diagnosis and treatment of aldosteronism (excessive secretion of the hormone aldosterone), diabetes insipidus (chronic excretion of large amounts of dilute urine, accompanied by extreme thirst), adrenal hypertension, Addison's disease (caused by destruction of the adrenal glands), dehydration, inappropriate antidiuretic hormone secretion, or other diseases involving electrolyte imbalance.

The GemLyte Chloride Assay is intended to measure the level of chloride in whole blood, plasma, serum, and urine. Chloride measurements are used in the diagnosis and treatment of electrolyte and metabolic disorders such as cystic fibrosis and diabetic acidosis.

The GemLyte Calcium Assay is intended to measure the ionized calcium level in whole blood, plasma and serum. Calcium measurements are used in the diagnosis and treatment of parathyroid disease, a variety of bone diseases, chronic renal disease and tetany (intermittent muscular contractions or spasms).

The GemLyte Lithium Assay is intended to measure lithium (from the drug lithium carbonate) in whole blood, plasma and serum. Measurements of lithium are used to assure that the proper drug dosage is administered in the treatment of patients with mental disturbances, such as manic-depressive illness (bipolar disorder).

Device Description

The GemLyte is an automated, microprocessor-controlled analyzer which utilizes ion-selective electrodes for the measurement of sodium, potassium, chloride, calcium and lithium in serum, plasma, whole blood and prediluted urine samples. GemLyte analyzer is designed with the user in mind, it is fully automated with simple .Yes' or 'No' commands for menu navigation. This simple interface insures that not only will the analyzer be easy to use for quick analysis, (less than a minute for samples), but also that the testing of samples can be done by even non-skilled operators with relative ease. The analyzer self-calibrates using Diamond Diagnostics Fluid Pack (510(k) 013850) every 4 hours through out the day or on request. Sodium, potassium, chloride and calcium are commonly measured for use in the diagnosis and management of patients with a broad range of renal, metabolic and cardiovascular disorders. Lithium is a drug used to treat mental illness. Mission controls (510 (k) 033063) are the recommended quality control material to be used daily.

AI/ML Overview

The provided text describes the 510(k) summary for the GemLyte Electrolyte Analyzer. Here's a breakdown of the acceptance criteria and the study information as requested:

1. Table of Acceptance Criteria and Reported Device Performance

The acceptance criteria for the GemLyte Electrolyte Analyzer are primarily based on achieving performance similar to or within specified limits of the predicate device (AVL 9180 (K961458)) and meeting internal reproducibility guidelines. The reported device performance metrics are presented as "reproducibility" (precision) and "correlation" (agreement with predicate device).

Acceptance Criteria (Performance Specifications)

MetricParameterMatrixAcceptance Criteria (Internal/Predicate Comparison)Reported Device Performance (Achieved)
Reproducibility (Precision)
Within-RunNa+Blood/Plasma/SerumC.V. ≤1%C.V. ≤0.6%
K+Blood/Plasma/SerumC.V. ≤1.5%C.V. ≤1.5%
Cl-Blood/Plasma/SerumC.V. ≤2%C.V. ≤1%
Ca++Blood/Plasma/SerumSD ≤0.02SD ≤0.02
Li+Blood/Plasma/SerumSD ≤0.03SD ≤0.04 (Note: This is slightly higher than the specified SD ≤0.03 in the "Reproducibility" table but within the "Summary of nonclinical tests" table. The text states "Results are within performance specifications.")
Na+UrineC.V. ≤5%C.V. ≤ 5%
K+UrineC.V. ≤5%C.V. ≤ 5%
Cl-UrineC.V. ≤5%C.V. ≤ 2% (Note: This is better than the specified C.V. ≤ 5% in the "Reproducibility" table but within the "Summary of nonclinical tests" table. The text states "Results are within performance specifications.")
Between-RunNa+Serum/Blood/SerumC.V. ≤2%C.V. ≤1.5%
K+Serum/Blood/SerumC.V. ≤3%C.V. ≤3%
Cl-Serum/Blood/SerumC.V. ≤3%C.V. ≤3%
Ca++Serum/Blood/SerumSD ≤0.06SD ≤0.06
Li+Serum/Blood/SerumSD ≤0.09SD ≤0.09
Na+UrineC.V. ≤5%C.V. ≤4%
K+UrineC.V. ≤5%C.V. ≤5%
Cl-UrineC.V. ≤5%C.V. ≤3%
LinearityAllAll claimed rangeCorrelation coefficients > 0.99 (regression analysis)All correlation coefficients were greater than 0.99.
Correlation to Predicate DeviceAllAll sample typesCorrelation coefficients > 0.99 (regression analysis)"Regression analysis show good correlation to predicate devices for all sample types, whole blood, plasma, serum and urine with correlation coefficients typically greater than 0.99." Specific R² values are provided in the tables above, all of which are ≥0.9687 (Chloride in Serum).

2. Sample Size Used for the Test Set and Data Provenance

The test set for the correlation studies used the following sample sizes:

  • Whole Blood: Sodium (n=128), Potassium (n=112), Chloride (n=123), Calcium (n=124), Lithium (n=123)
  • Plasma: Sodium (n=105), Potassium (n=107), Chloride (n=115), Calcium (n=123), Lithium (n=115)
  • Serum: Sodium (n=122), Potassium (n=127), Chloride (n=125), Calcium (n=120), Lithium (n=131)
  • Urine (Spot): Sodium (n=118), Potassium (n=108), Chloride (n=118)

Data Provenance: The document does not explicitly state the country of origin. It indicates that samples were "collected for testing on the GemLyte as well as another predicate device, the AVL 9180." The language used (e.g., "mmol/L") is standard for clinical chemistry. The study appears to be prospective in nature, as it describes a setup for collecting and testing samples on two devices for comparison.

For precision (reproducibility) studies:

  • Within-Run: 30 replicates for each sample type (blood, serum, urine).
  • Between-Run: 2 runs per day with 2 replicates per run for 10 days for each sample type.
  • Potassium (Whole Blood Specific): Multiple instruments (4) tested over a period of an hour due to instability.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

This type of device (electrolyte analyzer) does not typically rely on "expert" interpretation for ground truth, unlike imaging or diagnostic AI. The "ground truth" for the correlation study is the measurement obtained from the predicate device (AVL 9180), which is an established, legally marketed device for the same purpose. The text states, "Clinical testing was conducted to demonstrate the correlation of Diagnostics GemLyte Analyzer to predicate devices operated by trained personnel." It clarifies that the comparison device was operated by "trained personnel," implying standard laboratory practice, but does not specify "experts" in the sense of clinical specialists like radiologists.

4. Adjudication Method for the Test Set

Not applicable. As explained above, the ground truth is established by the predicate device's readings, not by expert consensus or adjudication. The study design is a direct comparison of measurements between two devices.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

Not applicable. This is not an imaging or diagnostic AI device that involves human readers interpreting cases. It is an automated electrolyte analyzer.

6. Standalone (Algorithm Only Without Human-in-the-Loop Performance) Study

Yes, the studies presented are standalone performance evaluations for the GemLyte Electrolyte Analyzer. The device is an automated system; its performance is measured directly (e.g., reproducibility, linearity) and compared against a predicate device's measurements, not against human performance in a diagnostic reading task.

7. Type of Ground Truth Used

The ground truth for the clinical correlation studies was the measurements obtained from the predicate device, AVL 9180. For within-run precision, the ground truth is implicitly the true stable concentration within the sample, with the device aiming to measure it consistently. For linearity, the ground truth was prepared standard solutions (dilutions from starting stock solutions).

8. Sample Size for the Training Set

The document does not specify a separate "training set" in the context of an AI/ML model for this device. The GemLyte Electrolyte Analyzer is based on ion-selective electrode (ISE) technology, not machine learning. Therefore, it does not typically involve a "training set" in the computational sense. The device is calibrated using internal calibration solutions and Diamond Diagnostics Fluid Pack (510(k) 013850) as described in the calibration section.

9. How the Ground Truth for the Training Set Was Established

As there is no "training set" for an AI/ML model, this question is not applicable in the typical sense.
For the device's internal calibration, the "ground truth" would be the known concentrations of the calibration solutions, which are presumably traceable to established standards. The text states:

  • "The analyzer self-calibrates using Diamond Diagnostics Fluid Pack (510(k) 013850) every 4 hours through out the day or on request."
  • "The GemLyte performs a 2-point calibration (3-point calibration if lithium) every 4 hours or software permits calibration on demand. A 1-point calibration is performed automatically with each measurement."

{0}------------------------------------------------

KOr2462

FEB - 6 2009

5. 510(k) Summary – GemLyte Electrolyte Analyzer

(1) Submitted by:

Diamond Diagnostics 333 Fiske St. Holliston, MA 01746

(2) Contact Person:

Liann Voo, Director of Quality Assurance and Scientific Development 508-429-0450 (x 351) Phone: 508-429-0452 Fax: lvoo@diamonddiagnostics.com E-mail:

(3) Summary Prepared:

August 25, 2008

(4) Device Trade Name:

GemLyte Electrolyte Analyzer Na*/K*/CI'/Ca**/Li*

(5) Regulatory Information

DescriptionCFR SectionDevice ClassProduct Code
Sodium Test System862.1665Class IIJGS
Potassium Test System862.1600Class IICEM
Chloride Test System862.1170Class IICGZ
Calcium Test System862.1145Class IIJFP
Lithium Test System862.3560Class IIJIH

(6) Predicate Devices:

Description510(k)Analytes
AVL9180K961458Sodium, Potassium, Chloride, Calcium and Lithium
IL 943 Flame PhotometerK823480Sodium, Potassium, Chloride
EasyLyteK000926Sodium, Potassium, Chloride
925 ChloridometerK810615Chloride

Statement of Technology Characteristics of the Device Compared to Predicate Device

Operating PrinciplePredicate DeviceGemLyte
PotentiometricNa+, K+, Cl-, Ca++, Li+K000926K961458Same

{1}------------------------------------------------

(7) Device Description and Indications for Use:

The GemLyte is an automated, microprocessor-controlled analyzer which utilizes ion-selective electrodes for the measurement of sodium, potassium, chloride, calcium and lithium in serum, plasma, whole blood and prediluted urine samples. GemLyte analyzer is designed with the user in mind, it is fully automated with simple .Yes' or 'No' commands for menu navigation. This simple interface insures that not only will the analyzer be easy to use for quick analysis, (less than a minute for samples), but also that the testing of samples can be done by even non-skilled operators with relative ease. The analyzer self-calibrates using Diamond Diagnostics Fluid Pack (510(k) 013850) every 4 hours through out the day or on request. Sodium, potassium, chloride and calcium are commonly measured for use in the diagnosis and management of patients with a broad range of renal, metabolic and cardiovascular disorders. Lithium is a drug used to treat mental illness. Mission controls (510 (k) 033063) are the recommended quality control material to be used daily.

The GemLyte Potassium Assay is intended to measure potassium in whole blood, serum, plasma, and urine on the GemLyte Electrolyte Analyzer. Measurements obtained by this device are used to monitor electrolyte balance in the diagnosis and treatment of diseases conditions characterized by low or high blood potassium levels.

The GemLyte Sodium Assay is intended to measure sodium in serum, plasma, and urine on the GemLyte Electrolyte Analyzer. Measurements obtained by this device are used to monitor electrolyte balance in the diagnosis and treatment of aldosteronism (excessive secretion of the hormone aldosterone), diabetes insipidus (chronic excretion of large amounts of dilute urine, accompanied by extreme thirst), adrenal hypertension, Addison's disease (caused by destruction of the adrenal glands), dehydration, inappropriate antidiuretic hormone secretion, or other diseases involving electrolyte imbalance.

The GemLyte Chloride Assay is intended to measure the level of chloride in plasma, serum, sweat, and urine. Chloride measurements are used in the diagnosis and treatment of electrolyte and metabolic disorders such as cystic fibrosis and diabetic acidosis.

The GemLyte Calcium Assay is intended to measure the total calcium level in serum. Calcium measurements are used in the diagnosis and treatment of parathyroid disease, a variety of bone disease and tetany (intermittent muscular contractions or spasms).

The GemLyte Lithium Assay is intended to measure lithium (from the drug lithium carbonate) in serum or plasma. Measurements of lithium are used to assure that the proper drug dosage is administered in the treatment of patients with mental disturbances, such as manic-depressive illness (bipolar disorder).

(8) Technological Characteristics of the Device: Principal of Measurement

The principles of measurements used in the GemLyte Analyzer are identical to those principles existing in the electrolyte analyzers K000926 (EasyLyte) and K961458 (AVL 9180) and are substantially equivalent to the K823480 (IL Flame Photometer) and K810615 (925 Chloridometer)

{2}------------------------------------------------

Calibration:

The GemLyte performs a 2-point calibration (3-point calibration if lithium) every 4 hours or software permits calibration on demand. A 1-point calibration is performed automatically with each measurement.

Analyzer tests samples for Na*/K*/CI/Ca*/Li* Technical Specifications:

Sample Types:Whole Blood, Serum, Plasma or Urine
-----------------------------------------------------------

Sample Size: 95uL blood, plasma, serum 95uL 1:3 dilution of urine

Measurement Range: ParameterMatrixSpecified range
Na+B/P/S/Q40 - 200 mmol/L
U1 - 300 mmol/L
K+B/P/S/Q1.5 - 15 mmol/L
U5 - 120 mmol/L
Cl-B/P/S/Q50 - 200 mmol/L
U1 - 300 mmol/L
iCa++B/P/S/Q0.3 - 5 mmol/L
Li+B/P/S/Q0.2 - 5.5 mmol/L
B = Whole bloodP = PlasmaS = Serum
Q = Aqueous QCU = Urine

Display Resolution: Blood, Plasma, Serum, Aqueous QC Na*: 0.1 mmol/L 0.01 mmol/L K+: 0.1 mmol/L or Cli: 0.1 mmol/L 0.01 mmol/L Ca**: 0.1 mmol/L or 0.01 mmol/L Li : 0.1 mmol/L Or

Urine

Na*: 0 mmol/L / K*: 0.1 mmol/L . Cli: 0 mmol/L

Reproducibility:

Blood, Plasma, Serum
Na+K+Cl-Ca++Li+
Within Range (n=30)C.V. ≤1%C.V. ≤1.5%C.V. ≤ 2%SD ≤0.02SD ≤0.03
Between Run (10 days)C.V. ≤2%C.V. ≤3%C.V. ≤ 3%SD ≤0.06SD ≤0.09

Urine

Na+K+Cl-
Within Range (n=30)C.V. ≤ 5%C.V. ≤ 5%C.V. ≤ 5%
Between Run (10 days)C.V. ≤ 5%C.V. ≤ 5%C.V. ≤ 5%

{3}------------------------------------------------

Analysis Time Calibration

57 seconds Every 4 hours on demand 2 point calibration Na ,K+,CI¯,Ca** 3 point calibration Li*

Power

5 Hz, 6 A or 120 VAC 220-240 VAC 24 Hz, 2 A (Factory set)

Size and Weight 12.4" (31.5cm) W x 13.2" (33.5cm) H x 11.6" (29.5cm) D, 13 lbs. (<6 kg)

(9) Summary of nonclinical tests submitted with the premarket notification for device.

Precision -

Within-Run Precision is determined for each different sample types, blood, serum and urine, by collecting multiple replicates of each sample type, 30, within a single period of time with out re-callbration of the instrument. Samples of blood and serum are measure in standard blood/serum mode while urine samples are measured in 'Urine Mode'. The standard deviation (SD) and Coefficient Variation (%CV) are calculated. Results are within performance specifications which are not different from the predicate device, AVL 9180(K961458).

Blood/Plasma/SerumUrine (1:10 dilution)
Na+C.V. ≤0.6%C.V. ≤5%
K+C.V. ≤1.5%C.V. ≤5%
Cl-C.V. ≤1%C.V. ≤2%
Ca++SD ≤0.02
Li+SD ≤0.04

Between-Run or Total Precision is determined by testing 2 runs per day (AM & PM) with 2 replicates per run for 10 days for each sample type. The standard deviation (SD) and/or Coefficient of Variation (%CV) are calculated. Results are summarized below and are similar to predicate device, AVL 9180(K961458) and with in specifications.

Serum/Blood/SerumUrine (1:3 dilution)
Na+C.V. ≤1.5%C.V. ≤4%
K+C.V. ≤3%C.V. ≤5%
Cl-C.V. ≤3%C.V. ≤3%
Ca++SD ≤0.06
Li+SD ≤0.09

Note: Because potassium in whole blood is unstable, and clinically significant changes are observed within 2 hours of sample collection, a total imprecision test was conducted using multiple instruments (4). Results were collected over a period of an hour.

Linearity

Whole blood, Plasma, Serum and Urine are linear across the claimed performance range. A minimum of 5 levels were tested for each type of sample. Dilutions were made from starting stock solutions and regression analysis done. Correlation coefficients were all greater than 0.99.

{4}------------------------------------------------

(10) Summary of clinical tests submitted with the pre-market notification for the device.

Clinical testing was conducted to demonstrate the correlation of Diagnostics GemLyte Analyzer to predicate devices operated by trained personnel. All sample types, whole blood, plasma, serum and urine were collected for testing on the GemLyte as well as another predicate device, the AVL 9180. Regression analysis show good correlation to predicate devices for all sample types, whole blood, plasma, serum and urine with correlation coefficients typically greater than 0.99.

Correlation Studies

Studies were conducted comparing the Diamond GEMLYTE to the Roche 9180 with whole blood, plasma, serum and spot urine samples.

Whole Blood in mmol/L

ParameterSlopeInterceptRangenSteyx
Sodium0.9881.040.996242.3 - 188.11281.13
Potassium0.9720.2230.99571.51 - 14.461120.16
Chloride1.011-0.490.988254.0 - 191.41231.76
Calcium0.9940.0190.98850.304 - 4.5141240.07
Lithium0.983-0.0490.99190.255 - 5.4101230.079

Plasma in mmol/L

ParameterSlopeInterceptRangenSteyx
Sodium0.9870.870.997743.5 - 194.41051.06
Potassium1.006-0.0140.99731.55 - 13.971070.178
Chloride1.022-3.230.987955.2 - 193.21152.58
Calcium0.9710.0420.99150.363 - 4.3861230.059
Lithium0.9930.0020.98780.298 - 5.0711150.092

Serum in mmol/L

ParameterSlopeInterceptRangenSteyx
Sodium0.990.030.996444.6 - 196.91221.37
Potassium0.9810.0850.99791.58 - 14.641270.118
Chloride1.021-2.570.968750.7 - 185.71253.79
Calcium0.9700.0360.98420.655 - 4.3091200.082
Lithium1.0040.0920.98760.213 - 5.1491310.122

Urine (Spot) in mmol/L

ParameterSlopeInterceptR2RangenStey
Sodium0.983-2.570.98438 - 2961188.42
Potassium0.9650.340.99345.3 - 118.21082.22
Chloride0.976-1.860.980715 - 2721188.99

(11) Conclusions drawn from the clinical and non-clinical testing.

Analysis of the comparative measurement presented in the 510(k) for this device, together with the linearity and precision data collected during these clinical and non-clinical trails demonstrates that the Diamond Diagnostics GemLyte ISE Analyzer (with Na *, K , C , Ca * , Li ) is safe, effective and equivalent to the predicate device to which it is compared.

{5}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/5/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized depiction of an eagle or bird with outstretched wings. The bird is positioned above the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA", which is arranged in a circular fashion around the bird.

Public Health Service

Food and Drug Administration 2098 Gaither Road Rockville MD 20850

FFR - 6 2009

Diamond Diagnostics, Inc. c/o Liann Voo 333 Fiske St. Holliston, MA 01746

K082462 Re:

Trade Name: GemLyte Electrolyte Analyzer Regulation Number: 21 CFR 862.1665 Regulation Name: Sodium Test System Regulatory Class: Class II Product Codes: JGS, CEM, CGZ, JFP, JIH Dated: January 27, 2009 Received: January 28, 2009

Dear Ms. Voo:

1

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820).

{6}------------------------------------------------

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific information about the application of labeling requirements to your device, or questions on the promotion and advertising of your device, please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (301) 594-3084. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html.

Sincerely yours,

C

Courthey C. Harper, Ph.D. Acting Director Division of Chemistry and Toxicology Office of In Vitro Diagnostics Device Evaluation and Safety Center for Devices and Radiological Health

{7}------------------------------------------------

510(k) Number (if known): K082462 Device Name: GemLyte Electrolyte Analyzer Indication For Use:

The GemLyte is an automated, microprocessor-controlled analyzer which utilizes ion-selective electrodes for the measurement of sodium, potassium, chloride, calcium and lithium in whole blood, plasma and serum, and prediluted urine samples.

The GemLyte Potassium Assay is intended to measure potassium in whole blood, plasma, serum, and urine on the GemLyte Electrolyte Analyzer. Measurements obtained by this device are used to monitor electrolyte balance in the diagnosis and treatment of diseases conditions characterized by low or high blood potassium levels.

The GemLyte Sodium Assay is intended to measure sodium in whole blood, plasma, serum, and urine on the GemLyte Electrolyte Analyzer. Measurements obtained by this device are used to monitor electrolyte balance in the diagnosis and treatment of aldosteronism (excessive secretion of the hormone aldosterone), diabetes insipidus (chronic excretion of large amounts of dilute urine, accompanied by extreme thirst), adrenal hypertension, Addison's disease (caused by destruction of the adrenal glands), dehydration, inappropriate antidiuretic hormone secretion, or other diseases involving electrolyte imbalance.

The GemLyte Chloride Assay is intended to measure the level of chloride in whole blood, plasma, serum, and urine. Chloride measurements are used in the diagnosis and treatment of electrolyte and metabolic disorders such as cystic fibrosis and diabetic acidosis.

The GemLyte Calcium Assay is intended to measure the ionized calcium level in whole blood, plasma and serum. Calcium measurements are used in the diagnosis and treatment of parathyroid disease, a variety of bone diseases, chronic renal disease and tetany (intermittent muscular contractions or spasms).

The GemLyte Lithium Assay is intended to measure lithium (from the drug lithium carbonate) in whole blood, plasma and serum. Measurements of lithium are used to assure that the proper drug dosage is administered in the treatment of patients with mental disturbances, such as manic-depressive illness (bipolar disorder).

Prescription Use X (21 CFR Part 801 Subpart D) And/Or

Over the Counter Use (21 CFR Part 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE: CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of In Vitro Diagnostic Device Evaluation and Safety (OVD)

Division Sign-Off Office of In Vitro Diagnostic Device Evaluation and Safety

510(k) K082462

§ 862.1665 Sodium test system.

(a)
Identification. A sodium test system is a device intended to measure sodium in serum, plasma, and urine. Measurements obtained by this device are used in the diagnosis and treatment of aldosteronism (excessive secretion of the hormone aldosterone), diabetes insipidus (chronic excretion of large amounts of dilute urine, accompanied by extreme thirst), adrenal hypertension, Addison's disease (caused by destruction of the adrenal glands), dehydration, inappropriate antidiuretic hormone secretion, or other diseases involving electrolyte imbalance.(b)
Classification. Class II.