K Number
K081556

Validate with FDA (Live)

Manufacturer
Date Cleared
2009-02-02

(244 days)

Product Code
Regulation Number
892.2050
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

CADstream is intended to be used in the visualization, analysis, and reporting of magnetic resonance imaging (MRI) studies. CADstream supports evaluation of dynamic MR data acquired during contrast administration. CADstream performs other user selected processing functions (such as image registration, subtractions, measurements, 3D renderings, and reformats).

CADstream also includes user-configurable features for reporting on findings in breast or general MRI studies. Additionally, CADstream assists users in planning MRI guided interventional procedures.

When interpreted by a skilled physician, this device provides information that may be used for screening, diagnosis, and interventional planning. Patient management decisions should not be made based solely on the results of CADstream.

CADstream may also be used as an image viewer of multi-modality, digital images, including ultrasound and mammography. CADstream is not intended for primary interpretation of digital mammography images.

Device Description

CADstream is an image processing system designed to assist in the visualization, analysis, and reporting of magnetic resonance imaging (MRI) studies. CADstream also is intended to provide workflow efficiency and interventional planning tools.

CADstream receives DICOM magnetic resonance images from a PACS or directly from the MRI scanner. As they are received, CADstream processes and displays the results on the CADstream server or a client personal computer.

Available features support:

  • Visualization (standard image viewing tools, MIPs, and reformats)
  • Analysis (registration, subtractions, coil inhomogeneity correction, kinetic curves, parametric image maps, and 3D volume rendering)
  • Reporting of user-selected findings and assessment
  • Interventional planning
  • Workflow efficiency
  • Communication and storage (DICOM import/export, query/retrieve, and study storage)

The CADstream system consists of proprietary software developed by Confirma installed on an off-the-shelf computer.

AI/ML Overview

Here's a breakdown of the acceptance criteria and study information for the CADstream Version 5 device, based on the provided text:

1. Table of Acceptance Criteria and Reported Device Performance:

The provided text does not explicitly state specific numerical acceptance criteria for performance metrics (e.g., sensitivity, specificity, accuracy, or quantitative measures of visualization/analysis improvement). Instead, the performance testing is described in a more general sense, focusing on verification and validation against requirements and substantial equivalence to predicate devices.

Therefore, the table will reflect the general statements made about performance rather than specific, measurable criteria.

Acceptance Criteria (Implied from text)Reported Device Performance
Device meets input requirements (Verification Testing)CADstream version 5 has successfully undergone extensive verification testing to verify the device meets input requirements.
Product modifications do not raise new safety or effectiveness concernsBench validation testing was also performed to demonstrate the product modifications from version 4 resulted in a substantially equivalent product and did not raise any new safety or effectiveness concerns.
Device conforms to user needs and intended uses (Clinical Validation)Additionally, the completed clinical validation testing demonstrates the device conforms to user needs and intended uses per 21CFR820.30(g).
Device is safe and effective"The successful completion of verification and validation testing has demonstrated conformance to design controls, user needs, and intended use, and that the device is safe and effective." "Based on the information supplied in this 510(k), we conclude that the subject device is safe, effective, and substantially equivalent to the predicate devices."
Substantially equivalent to predicate devicesBench validation testing demonstrated substantial equivalence to version 4 and other predicate devices in terms of features (as shown in Table 1 and Table 2 comparison) and overall safety and effectiveness. The 510(k) summary explicitly concludes substantial equivalence to predicate devices based on the submitted information.
Provides features for visualization, analysis, and reporting of MR imagesDevice description and feature comparison tables illustrate the capability to perform: - Visualization (standard image viewing tools, MIPs, reformats, 3D volume rendering) - Analysis (registration, subtractions, coil inhomogeneity correction, kinetic curves, parametric image maps) - Reporting of user-selected findings and assessment - Interventional planning - Workflow efficiency - Communication and storage (DICOM import/export, query/retrieve, and study storage)

2. Sample Size Used for the Test Set and Data Provenance:

  • Sample Size for Test Set: The document does not specify the sample size (number of patients or cases) used for the "clinical validation testing."
  • Data Provenance: The document does not specify the country of origin of the data or whether it was retrospective or prospective.

3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications:

  • Number of Experts: The document does not specify the number of experts used to establish ground truth for the test set.
  • Qualifications of Experts: The document does not specify the qualifications of any experts involved in establishing ground truth. The "Indications for Use" section mentions interpretation "by a skilled physician," but this is a general statement about clinical use, not a description of ground truth establishment during testing.

4. Adjudication Method for the Test Set:

  • The document does not mention any specific adjudication method (e.g., 2+1, 3+1, none) used for the test set.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done:

  • The document does not mention a multi-reader multi-case (MRMC) comparative effectiveness study, nor does it discuss the effect size of human readers' improvement with or without AI assistance. The focus is on device features and general validation, not comparative reader performance.

6. If a Standalone (Algorithm Only Without Human-in-the-Loop Performance) Was Done:

  • The document describes "Performance Testing" including "extensive verification testing," "bench validation testing," and "clinical validation testing." While these tests would involve evaluating the algorithm's functionality, the document does not explicitly state that a standalone performance study comparing the algorithm's diagnostic accuracy against a ground truth was conducted without a human in the loop. The device is described as an "image processing system designed to assist," implying a human-in-the-loop context for its intended use and evaluation.

7. The Type of Ground Truth Used:

  • The document does not explicitly state the type of ground truth used for any testing (e.g., expert consensus, pathology, outcomes data). It broadly refers to "user needs and intended uses" for clinical validation.

8. The Sample Size for the Training Set:

  • The document does not mention a training set or its sample size. This is typical for 510(k) submissions of this era for image processing systems, as they often leverage established algorithms or feature sets rather than reporting on deep learning model training.

9. How the Ground Truth for the Training Set Was Established:

  • Since a training set is not mentioned, the document does not describe how ground truth for a training set was established.

{0}------------------------------------------------

K08 1556

FEB -2 2009

510(k) Summary 5

This 510(k) summary is submitted as part of the PreMarket Notification in accordance with the requirements of SMDA 1990, 21 CFR 807.87(h), and 21 CFR 807.92.

1.Date PreparedDecember 16, 2008
2.510(k) SubmitterConfirma, Inc.11040 Main Street, Suite 100Bellevue, WA 98004Phone: 425-691-1400Fax: 425-691-1599
3.510(k) Contact PersonBrent Lewis, Director RegulatoryAffairs and Quality AssuranceEmail: blewis@confirma.comPhone: 206-455-5398Fax: 425-691-1599
4.Device Common NameImage Processing System
5.Device Trade NameCADstream Version 5
6.Classification Regulation21 CFR 892.2050
7.Class2
8.PanelRadiology Panel
9.Product CodeLLZ

10. Comparison with Legally Marketed Devices

As with the predicate devices listed in Table 1, CADstream version 5 is indicated for use in the visualization, analysis, reporting, interventional planning, and improved workflow efficiency of magnetic resonance imaging (MRI) studies and supports the evaluation of dynamic MR data acquired during contrast administration. All these devices are intended to provide information that may be useful in screening or diagnosis.

Table 1: Predicate Devices

ManufacturerProductCleared 510(k)Product CodeClassification
ConfirmaCADstream 5PendingLLZ
CADstream 4K043216LLZ
CADstream 2K031779LLZ
InvivoDynaCAD 1.0K041286LZ
CADsciences3TPK031350. K050862LIZ
GE MedicalAW & FunctoolK96026517
ClarioZ3D Contrast AcuityK080196LL7

{1}------------------------------------------------

Table 2 presents a detailed comparison of features in CADstream version 5 to the predicate devices.

CADstream 5 FeaturesPredicate Devices
CADstream2 & 4DynaCAD1.0AW &Functool3TPZ3D
General Body MR Functions
Standard image viewing toolsXXXXX
View/play clips from ultrasoundimagesX0000
MIPsXX0XX
ReformatsXX00X
RegistrationXX0XX
Subtraction seriesXX0XX
Coil inhomogeneity correctionX000
View 3D volume renderingXXX0X
Kinetic curvesXXXXX
Parametric image mapsXXXXX
Portfolio of user-selected findingsXXXXX
DICOM import/export andquery/retrieveXXX00
User-Customizable ReportingXX0XX
Breast Specific: Module Features
Reporting for Breast MR StudiesXX0X0
View/invert mammography imagesXX000
View/edit finding volumeXX0X0
View/edit finding locationXX0X0
View/edit finding sizeXX0X0
View/edit kinetic curve with highestuptakeXX0X0
View images and findings from twostudies from same patientXX0X0
Interventional planningXX000

Table 2: Predicate Device Feature Comparison

Key

Substantially equivalent feature cleared in predicate device. X

0 Not known from literature.

11. Device Description

CADstream is an image processing system designed to assist in the visualization, analysis, and reporting of magnetic resonance imaging

{2}------------------------------------------------

(MRI) studies. CADstream also is intended to provide workflow efficiency and interventional planning tools.

CADstream receives DICOM magnetic resonance images from a PACS or directly from the MRI scanner. As they are received, CADstream processes and displays the results on the CADstream server or a client personal computer.

Available features support:

  • Visualization (standard image viewing tools, MIPs, and 미 reformats)
  • Analysis (registration, subtractions, coil inhomogeneity B correction, kinetic curves, parametric image maps, and 3D volume rendering)
  • Reporting of user-selected findings and assessment ■
  • Interventional planning 트
  • Workflow efficiency l
  • Communication and storage (DICOM import/export, query/retrieve, and study storage)

The CADstream system consists of proprietary software developed by Confirma installed on an off-the-shelf computer.

12. Indications for Use

CADstream is intended to be used in the visualization, analysis, and reporting of magnetic resonance imaging (MRI) studies. CADstream supports evaluation of dynamic MR data acquired during contrast administration. CADstream performs other user selected processing functions (such as image registration, subtractions, measurements, 3D renderings, and reformats).

CADstream also includes user-configurable features for reporting on findings in breast or general MRI studies. Additionally, CADstream assists users in planning MRI guided interventional procedures.

When interpreted by a skilled physician, this device provides information that may be used for screening, diagnosis, and interventional planning. Patient management decisions should not be made based solely on the results of CADstream.

{3}------------------------------------------------

CADstream may also be used as an image viewer of multi-modality, digital images, including ultrasound and mammography. CADstream is not intended for primary interpretation of digital mammography images.

13. Performance Testing

CADstream version 5 has successfully undergone extensive verification testing to verify the device meets input requirements. Bench validation testing was also performed to demonstrate the product modifications from version 4 resulted in a substantially equivalent product and did not raise any new safety or effectiveness concerns. Additionally, the completed clinical validation testing demonstrates the device conforms to user needs and intended uses per 21CFR820.30(g).

14. Conclusion

CADstream Version 5 provides features to integrate radiology department workflow by facilitating the visualization, analysis, and reporting of MR images. The potential hazards have been studied and controlled as part of the product development process, including risk analysis and design considerations. The successful completion of verification and validation testing has demonstrated conformance to design controls, user needs, and intended use, and that the device is safe and effective.

Based on the information supplied in this 510(k), we conclude that the subject device is safe, effective, and substantially equivalent to the predicate devices.

{4}------------------------------------------------

Truthful and Accurate Statement

PREMARKET NOTIFICATION

TRUTHFUL AND ACCURATE STATEMENT

[As required by 21 CFR 807.87(k)]

I certify that, in my capacity as Acting Director of Regulatory Affairs at Confirma, Inc., I believe to the best of my knowledge, that all data and information submitted in this 510(k) PreMarket Notification are truthful and accurate and that no material fact has been omitted.

ZZZ.

Brent Lewis Acting Director, Regulatory Affairs Confirma, Inc.

02 Jun 2008

Date

02Jun2008

6

{5}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Public Health Service

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

FEB - 2 2009

Mr. Brent Lewis Director, Regulatory Affairs and Quality Assurance Confirma, Incorporated 11040 Main Street. Suite 100 BELLEVUE WA 98004

Re: K081556

Trade/Device Name: CADstream@ Version 5 Regulation Number: 21 CFR §892.2050 Regulation Name: Picture archiving and communications systems Regulatory Class: II Product Code: LLZ

Dated: December 23, 2008 Received: December 23, 2008

Dear Mr. Lewis:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Ilsting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{6}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801; good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act): 21 CFR 1000-1050.

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding os substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at one of the following numbers, based on the regulation number at the top of this letter.

21 CFR 876.xxx(Gastroenterology/Renal/Urology)(240) 276-0115
21 CFR 884.xxx(Obstetrics/Gynecology)(240) 276-0115
21 CFR 892.xxx(Radiology)(240) 276-0120
Other(240) 276-0100

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometrics' (OSB's) Division of Postmarket Surveillance at 240-276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at 240-276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufactures, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrb/industry.suppot/index.html.

Sincerely yours,

Janine M. Morris

cting Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{7}------------------------------------------------

Indication(s) for Use Statement

510(k) Number: K081556

Device Name: CADstream@ Version 5

Indications for Use:

CADstream is intended to be used in the visualization, analysis, and reporting of magnetic resonance imaging (MRI) studies. CADstream supports evaluation of dynamic MR data acquired during contrast administration. CADstream performs other user selected processing functions (such as image registration, subtractions, measurements, 3D renderings, and reformats).

CADstream also includes user-configurable features for reporting on findings in breast or general MRI studies. Additionally, CADstream assists users in planning MRI guided interventional procedures.

When interpreted by a skilled physician, this device provides information that may be used for screening, diagnosis, and interventional planning. Patient management decisions should not be made based solely on the results of CADstream.

CADstream may also be used as an image viewer of multi-modality, digital images, including ultrasound and mammography. CADstream is not intended for primary interpretation of digital mammography images.

Prescription Use X (Part 21 CFR 801 Subpart D)

Over-The-Counter Use AND/OR (21 CFR 807 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)
(Division Sign-Off)
Division of Reproductive, Abdominal and
Radiological Devices
510(k) NumberK081656

4

§ 892.2050 Medical image management and processing system.

(a)
Identification. A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.(b)
Classification. Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).