(45 days)
Not Found
Yes
The device description explicitly mentions the use of "multivariate pattern recognition methods" for tissue segmentation and classification, which is a common technique used in machine learning.
No.
The device provides information for screening and diagnosis but does not directly treat or cure any condition. Its primary function is to analyze MRI studies and provide measurements and processing functions, and it explicitly states that "Patient management decisions should not be made based solely on the results of CADstream analysis."
Yes
Explanation: The "Intended Use / Indications for Use" section states, "When interpreted by a skilled physician, this device provides information that may be useful in screening and diagnosis." This clearly indicates its role in the diagnostic process.
No
The device description explicitly states that the CADstream system consists of "proprietary software developed by Confirma installed on off-the-shelf personal hardware," and lists specific hardware components like a "PC Server" and an "Archive System" with a CD/DVD burner and printer. This indicates it is a system that includes both software and hardware.
Based on the provided information, this device is NOT an IVD (In Vitro Diagnostic).
Here's why:
- IVD Definition: In vitro diagnostics are tests performed on samples taken from the human body, such as blood, urine, or tissue, to detect diseases, conditions, or infections.
- CADstream's Function: CADstream analyzes medical images (MRI, ultrasound, mammography) that are acquired from the patient, but it does not perform tests on biological samples taken from the patient. It processes and analyzes existing image data to provide information to a physician.
The device is a computer-aided detection (CAD) system for medical imaging, which falls under a different regulatory category than IVDs.
No
The document states "Control Plan Authorized (PCCP) and relevant text: Not Found", which indicates that no PCCP was approved or cleared for this device.
Intended Use / Indications for Use
CADstream is a computer aided detection (CAD) system intended for use in analyzing magnetic resonance imaging (MRI) studies. CADstream automatically registers serial patient image acquisitions (in 2D or 3D) to minimize the impact of patient motion, segments and labels tissue types based on enhancement characteristics (parametric image maps, referred to as angiogenesis maps), and provides quantitative measurements of morphological features of the segmented tissues. CADstream performs other user-defined post-processing functions (image subtractions, multiplanar and oblique reformats, maximum intensity projections, image averaging, removal of cardiac artifact).
CADstream also can be used to provide accurate and reproducible measurements of the segmented tissue volumes of interest (VOI)). These measurements include longest diameter, longest in-plane diameters, volume measurement-(reported in cc), ratio of breast volume to VOI, distance of VOI to anatomical landmarks, and 3D renderings of the VOI.
The system includes an optional remote image viewer (CADalyst™), optimized for viewing breast MR studies processed by CADstream. CADalyst also displays and telecommunicates images from a number of other medical modalities, including ultrasound and digitized mammographic images, processed data from FDA-cleared third party image processing systems, and FDA-approved systems for computer-aided detection and advanced image processing. The ultrasound and digitized mammographic images displayed on the CADalyst remote image viewer must not be used for primary diagnostic interpretation.
CADstream™ includes software to support the use of interventional breast coils and MR stereotactic localization devices to perform MR-guided breast interventional procedures (SureLocTM). Using information from MR images regarding the coordinates of a user-specified region of interest, and fiducial coordinates, the software provides an automatic calculation of the location and depth of the targeted region of interest, such as a lesion or suspected lesion.
CADstream also includes the option to add annotations based on the American College of Radiology's Breast Imaging Reporting and Data System (BI-RADS®) Breast Imaging Atlas.
When interpreted by a skilled physician, this device provides information that may be useful in screening and diagnosis. Patient management decisions should not be made based solely on the results of CADstream analysis.
Product codes (comma separated list FDA assigned to the subject device)
90 LLZ
Device Description
The CADstream device relies on the assumption that pixels having similar MR signal intensities represent similar tissues. The CADstream software simultaneously analyzes the pixel signal intensities from multiple MR sequences and applies multivariate pattern recognition methods to perform tissue segmentation and classification. CADstream is designed to analyze dynamic breast MRI studies.
The CADstream system consists of proprietary software developed by Confirma installed on off-the-shelf personal hardware. The CADstream System consists of the following key components:
- A PC Server: a desk-side or rack-mount PC capable of running the . Server software, Client User Interface software, and Study Viewer software
- Server software: performs CADstream analysis and processing .
- . Client User Interface software: an administrative web page hosted on the Server
- Study Viewer software (CADalyst): an Image Viewer component, . optimized for viewing breast MR studies processed by CADstream
- . Archive System: a PC with a combination CD/DVD burner and printer
To initiate CADstream analysis and processing, images are acquired and pushed to the CADstream system from the MR scanner. CADstream is configured to automatically process the images and create additional series. In this configuration, CADstream forwards the processed and original images to the existing softcopy reading station.
CADstream Version 4.0 includes the following software features:
- CADstream™, which allows the user to select a study to view, define . settings for CADstream processes, process and export CADstream results, and perform administrative tasks.
- CADalyst™ (the study viewer), which allows the user to view . CADstream results, interact with the volumes, curves, and AngioMap overlays, and print and save a CADstream Portfolio. This component can be accessed from a standalone CADstream workstation, a remote computer connected to the same network as the CADstream server, such as a PACS, or a PC.
- . CADalyst includes multi-modality viewing capabilities, allowing the user to view ultrasound and digitized mammographic images. The ultrasound and digitized mammographic images displayed on the CADalyst remote image viewer must not be used for primary diagnostic interpretation.
- SureLoc™ is a guidance tool that assists the user in calculating target ● coordinates for biopsy, wire localization, or other interventional procedures.
- DICOM query/retrieve and print, which allows retrieval of DICOM . data from any DICOM-compliant device that is configured for CADstream query and retrieval.
Mentions image processing
Yes
Mentions AI, DNN, or ML
Not Found
Input Imaging Modality
Magnetic Resonance Imaging (MRI), ultrasound, digitized mammographic images
Anatomical Site
Breast
Indicated Patient Age Range
Not Found
Intended User / Care Setting
Skilled physician / radiology department
Description of the training set, sample size, data source, and annotation protocol
Not Found
Description of the test set, sample size, data source, and annotation protocol
Not Found
Summary of Performance Studies (study type, sample size, AUC, MRMC, standalone performance, key results)
Not Found
Key Metrics (Sensitivity, Specificity, PPV, NPV, etc.)
Not Found
Predicate Device(s): If the device was cleared using the 510(k) pathway, identify the Predicate Device(s) K/DEN number used to claim substantial equivalence and list them here in a comma separated list exactly as they appear in the text. List the primary predicate first in the list.
Reference Device(s): Identify the Reference Device(s) K/DEN number and list them here in a comma separated list exactly as they appear in the text.
Not Found
Predetermined Change Control Plan (PCCP) - All Relevant Information for the subject device only (e.g. presence / absence, what scope was granted / cleared under the PCCP, any restrictions, etc).
Not Found
§ 892.2050 Medical image management and processing system.
(a)
Identification. A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.(b)
Classification. Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).
0
7
510(k) Summary of Safety and Effectiveness
This 510(k) summary of safety and effectiveness information is submitted in accordance with the requirements of 21 CFR Part 807.87 (h).
1. Identification of Submitter:
Submitter: | Confirma, Inc. |
---|---|
Address: | 821 Kirkland Avenue |
Kirkland, WA 98033 | |
Phone: | 425-576-1226 |
Fax: | 425-576-9295 |
Contact: | Patricia A. Milbank |
Title: | Regulatory Consultant |
Phone: | 425-894-9733 |
Fax: | 425-865-9023 |
Date Prepared: | November 19, 2004 |
2. Identification of Product:
Trade Name: | CADstream™ Version 4.0 |
---|---|
Common Name: | Picture Archiving and Communications System |
Classification Name: | Image Processing System (LLZ) 21 CFR 892.2050 |
Manufacturer: | Confirma, Inc. |
---|---|
821 Kirkland Avenue | |
Kirkland, WA 98033 |
3. Marketed Devices
The multi-modality review workstation provided in CADstream Version 4.0 is substantially equivalent to the devices listed below:
Model: | Seno Advantage |
---|---|
Manufacturer: | GE Medical Systems |
510 (k) Number: | K033400 |
Model: | Sectra IDS5 Radiology Workstation |
Manufacturer: | Sectra-Imtec Ab |
510 (k) Number: | K033712. K040376 |
1
4. Device Description:
The CADstream device relies on the assumption that pixels having similar MR signal intensities represent similar tissues. The CADstream software simultaneously analyzes the pixel signal intensities from multiple MR sequences and applies multivariate pattern recognition methods to perform tissue segmentation and classification. CADstream is designed to analyze dynamic breast MRI studies.
The CADstream system consists of proprietary software developed by Confirma installed on off-the-shelf personal hardware. The CADstream System consists of the following key components:
- A PC Server: a desk-side or rack-mount PC capable of running the . Server software, Client User Interface software, and Study Viewer software
- Server software: performs CADstream analysis and processing .
- . Client User Interface software: an administrative web page hosted on the Server
- Study Viewer software (CADalyst): an Image Viewer component, . optimized for viewing breast MR studies processed by CADstream
- . Archive System: a PC with a combination CD/DVD burner and printer
To initiate CADstream analysis and processing, images are acquired and pushed to the CADstream system from the MR scanner. CADstream is configured to automatically process the images and create additional series. In this configuration, CADstream forwards the processed and original images to the existing softcopy reading station.
CADstream Version 4.0 includes the following software features:
- CADstream™, which allows the user to select a study to view, define . settings for CADstream processes, process and export CADstream results, and perform administrative tasks.
- CADalyst™ (the study viewer), which allows the user to view . CADstream results, interact with the volumes, curves, and AngioMap overlays, and print and save a CADstream Portfolio. This component can be accessed from a standalone CADstream workstation, a remote computer connected to the same network as the CADstream server, such as a PACS, or a PC.
- . CADalyst includes multi-modality viewing capabilities, allowing the user to view ultrasound and digitized mammographic images. The ultrasound and digitized mammographic images displayed on the CADalyst remote image viewer must not be used for primary diagnostic interpretation.
2
- SureLoc™ is a guidance tool that assists the user in calculating target ● coordinates for biopsy, wire localization, or other interventional procedures.
- DICOM query/retrieve and print, which allows retrieval of DICOM . data from any DICOM-compliant device that is configured for CADstream query and retrieval.
5. Indications for Use
CADstream is a computer aided detection (CAD) system intended for use in analyzing magnetic resonance imaging (MRI) studies. CADstream automatically registers serial patient image acquisitions (in 2D or 3D) to minimize the impact of patient motion, segments and labels tissue types based on enhancement characteristics (parametric image maps, referred to as angiogenesis maps), and provides quantitative measurements of morphological features of the segmented tissues. CADstream performs other user-defined post-processing functions (image subtractions, multiplanar and oblique reformats, maximum intensity projections, image averaging, removal of cardiac artifact).
CADstream also can be used to provide accurate and reproducible measurements of the segmented tissue volumes of interest (VOI)). These measurements include longest diameter, longest in-plane diameters, volume measurement-(reported in cc), ratio of breast volume to VOI, distance of VOI to anatomical landmarks, and 3D renderings of the VOI.
The system includes an optional remote image viewer (CADalyst™), optimized for viewing breast MR studies processed by CADstream. CADalyst also displays and telecommunicates images from a number of other medical modalities, including ultrasound and digitized mammographic images, processed data from FDA-cleared third party image processing systems, and FDA-approved systems for computer-aided detection and advanced image processing. The ultrasound and digitized mammographic images displayed on the CADalyst remote image viewer must not be used for primary diagnostic interpretation.
CADstream™ includes software to support the use of interventional breast coils and MR stereotactic localization devices to perform MR-guided breast interventional procedures (SureLocTM). Using information from MR images regarding the coordinates of a user-specified region of interest, and fiducial coordinates, the software provides an automatic calculation of the location and depth of the targeted region of interest, such as a lesion or suspected lesion.
3
CADstream also includes the option to add annotations based on the American College of Radiology's Breast Imaging Reporting and Data System (BI-RADS®) Breast Imaging Atlas.
When interpreted by a skilled physician, this device provides information that may be useful in screening and diagnosis. Patient management decisions should not be made based solely on the results of CADstream analysis.
6. Comparison with Predicate Devices
The CADalyst multi-modality review workstation provided with CADstream Version 4.0 is substantially equivalent to the following image processing systems used by radiologists:
Seno Advantage Workstation | |
---|---|
Manufacturer: | GE Medical Systems |
510 (k) Number: | K033400 |
Sectra IDS5 Radiology Workstation Manufacturer: Sectra-Imtec Ab K033712, K040376
Each of these workstations allows easy selection, review, processing, archive, printing and media interchange of multi-modality images from a variety of diagnostic imaging systems.
7. Conclusions
CADstream Version 4.0, with the remote CADalyst multi-modality review workstation, provides additional features to further integrate radiology department workflow. The potential hazards have been studied and controlled as part of the product development process, including risk analysis, test and design considerations, and planned verification and validation testing processes. The CADstream system provides images comparable to the predicate devices.
4
Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo is circular and contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. In the center of the circle is an abstract image of an eagle.
Public Health Service
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
JAN - 6 2005
Ms. Patricia A. Milbank Regulatory Consultant Confirma, Inc. 821 Kirkland Avenue, Suite 100 KIRKLAND WA 98033
Re: K043216 Trade/Device Name: CADstream™ Version 4.0 Regulation Number: 21 CFR 892.2050 Regulation Name: Picture archiving and communications system Regulatory Class: II Product Code: 90 LLZ Dated: November 19, 2004 Received: November 22, 2004
Dear Ms. Milbank:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
5
This letter will allow you to begin marketing your device as described in your Section 510(k) I mis loker will and wys a substantial equivalence of your device of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at one of the following numbers, based on the regulation number at the top of this letter:
21 CFR 876.xxxx | (Gastroenterology/Renal/Urology) | 240-276-0115 |
---|---|---|
21 CFR 884.xxxx | (Obstetrics/Gynecology) | 240-276-0115 |
21 CFR 892.xxxx | (Radiology) | 240-276-0120 |
Other | 240-276-0100 |
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsmamain.html
Sincerely yours,
Nancy C. brogdon
Nancy C. Brogdon Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
6
Indication(s) for Use Statement 6
510(k) Number:
To be assigned by FDA
CADstream™ Version 4.0 Device Name:
Indications for Use:
CADstream is a computer aided detection (CAD) system intended for use in analyzing magnetic resonance imaging (MRI) studies. CADstream automatically registers serial patient image acquisitions (in 2D or 3D) to minimize the impact of patient motion, segments and labels tissue types based on enhancement characteristics (parametric image maps, referred to as angiogenesis maps), and provides quantitative measurements of morphological features of the segmented tissues. CADstream performs other user-defined post-processing functions (image subtractions, multiplanar and oblique reformats, maximum intensity projections, image averaging, removal of cardiac artifact).
CADstream also can be used to provide accurate and reproducible measurements of the segmented tissues volumes (volumes of interest (VOI)). These measurements include longest diameter, longest in-plane diameters, volume measurement (reported in cc), ratio of breast volume to VOI, distance of VOI to anatomical landmarks, and 3D renderings of the VOI.
The system includes an optional remote image viewer (CADalyst™), optimized for viewing breast MR studies processed by CADstream. CADalyst also displays and telecommunicates images from a number of other medical modalities, including ultrasound and digitized mammographic images, processed data from FDA-cleared third party image processing systems, and FDA-approved systems for computer-aided detection and advanced image processing. The ultrasound and digitized mammographic images displayed on the CADalyst remote image viewer must not be used for primary diagnostic interpretation.
CADstream™ includes software to support the use of interventional breast coils and MR stereotactic localization devices to perform MR-guided breast interventional procedures (SureLoc™). Using information from MR images regarding the coordinates of a user-specified region of interest, and fiducial coordinates, the software provides an automatic calculation of the location and depth of the targeted region of interest, such as a lesion or suspected lesion.
7
CADstream also includes the option to add annotations based on the American College of Radiology's Breast Imaging Reporting and Data System (BI-RADS®) Breast Imaging Atlas.
When interpreted by a skilled physician, this device provides information that may be useful in screening and diagnosis. Patient management decisions should not be made based solely on the results of CADstream analysis.
Over-The-Counter Use Prescription Use _ AND/OR (Part 21 CFR 801 Subpart D) (21 CFR 807 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) ーーーーーーーーーーーーーーーーーーーーーーーーーーーーーーーーーーーーーーーーーーーーーーー
Concurrence of CDRH, Office of Device Evaluation (ODE)
Daniel A. Legmann
(Division Sign-Off)
(Division Sign-Off) Division of Reproductive, Abdom and Radiological Devices 510(k) Number