(251 days)
The CELL-DYN Emerald System is an automated hematology analyzer designed for in vitro diagnostic use in clinical laboratories.
The CELL-DYN Emerald System is a bench top analyzer consisting of the main analyzer with data module, display station, and printer. The main analyzer, data module, and display station are housed in a single chassis. The printer is a stand-alone module. The CELL-DYN Emerald open sampler is equipped to aspirate blood from a collection tube that has been opened and is held under the open sample aspiration probe.
The CELL-DYN Emerald™ System is an automated hematology analyzer.
Here's an analysis of the acceptance criteria and the study that proves the device meets them:
1. Table of Acceptance Criteria and Reported Device Performance
The provided 510(k) summary does not explicitly list quantitative acceptance criteria in a structured table. Instead, it states that the CELL-DYN Emerald System's performance was compared to the predicate device (CELL-DYN 1800 System) and found to be substantially equivalent in terms of:
- Background: Implies that the baseline noise or interference from reagents/instrumentation is acceptable and comparable.
- Correlation: Suggests that the measurements from the Emerald system correlate well with those from the 1800 system across various parameters (WBC, RBC, HGB, HCT, MCV, PLT, etc.).
- Precision: Indicates that the Emerald system provides reproductible results, with variability comparable to the predicate device.
- Linearity: Means the system accurately measures analytes over their clinically relevant range.
- Carryover: Refers to the absence of significant contamination from one sample to the next.
Since specific quantitative targets are not provided in this public summary, we can infer that the acceptance criteria for these performance characteristics were that the CELL-DYN Emerald System performed comparably to, or within acceptable predefined ranges relative to, the established performance of the CELL-DYN 1800 System.
2. Sample Size Used for the Test Set and Data Provenance
The document mentions that the CELL-DYN Emerald System was compared "in both external as well as an internal clinical trial" to the CELL-DYN 1800 System. However, it does not specify the sample size used for these clinical trials.
The data provenance (country of origin, retrospective/prospective) is also not explicitly stated in this summary. Such details would typically be found in the Medical Clinical Summary Report (Attachment F) which is referenced but not provided here.
3. Number of Experts and Qualifications for Ground Truth
The provided text does not contain information about the number of experts used or their qualifications to establish ground truth for the test set. For hematology analyzers, ground truth often involves comparison to reference methods or manual microscopic differentials performed by trained medical technologists or hematopathologists.
4. Adjudication Method for the Test Set
The document does not specify any adjudication method used for the test set. In clinical studies, adjudication (e.g., by multiple experts) is often used to resolve discrepancies or ensure accuracy of ground truth.
5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study
A Multi-Reader Multi-Case (MRMC) comparative effectiveness study, which typically evaluates how human readers' performance changes with and without AI assistance, is not applicable and therefore not mentioned for this type of device. The CELL-DYN Emerald System is an automated analyzer, not an AI-assisted diagnostic tool that supports human interpretation of images or other complex data. Its equivalency is assessed against another automated device.
6. Standalone Performance Study
Yes, a standalone performance study was implicitly done. The "Equivalency Data Summary" section describes a study where the CELL-DYN Emerald System's results were directly compared to those of the predicate device, the CELL-DYN 1800 System. This comparison covered background, correlation, precision, linearity, and carryover, indicating that the algorithm's (device's) performance was evaluated independently.
7. Type of Ground Truth Used
The ground truth for this device's performance evaluation was established through comparison to a legally marketed predicate device (Abbott CELL-DYN® 1800 System). This is a common approach for 510(k) submissions, where substantial equivalence is demonstrated by showing that the new device performs comparably to an existing, cleared device, which itself would have established its performance against reference methods or clinical outcomes. The underlying "ground truth" for the predicate likely involved standard reference methods for cell counting and differentials.
8. Sample Size for the Training Set
The document does not provide information regarding a separate "training set" sample size. For traditional automated analyzers like this, the development process might involve internal datasets for calibration and algorithm optimization, but these are typically not documented as explicitly as training sets for machine learning models in 510(k) summaries. The clinical trials mentioned (external and internal) would primarily serve as validation/test sets to demonstrate performance.
9. How Ground Truth for the Training Set Was Established
As no specific training set information is provided, there is no explicit description of how ground truth for such a set was established. For development of automated hematology analyzers, the initial ground truth typically comes from:
- Reference materials: Calibrated standards and controls.
- Reference methods: Manual microscopy for differentials, or other established laboratory methods.
- Validated clinical samples: Samples with well-characterized results from established methods.
The algorithms and parameters of the CELL-DYN Emerald would have been developed and optimized using such data to ensure accurate and reliable measurements.
{0}------------------------------------------------
510(k) Summary
CELL-DYN Emerald™ System
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92.
Contact Person
The assigned 510(k) number is:
f1495
Abbott Laboratories 5440 Patrick Henry Drive Santa Clara, CA 95054
Michelle Roeding Section Manager, Regulatory Affairs Phone: (408) 567-3781 Fax: (408) 982-4863
Date Prepared:
Submitted by
December 23, 2008
CELL-DYN Emerald™ System
Automated Hematology Analyzer
Proprietary Name:
Common Name:
Automated Differential Cell Counter (21 CFR 864.5220)
Predicate Device:
Classification Name:
Abbott CELL-DYN® 1800 System, K030513
Device Description:
The CELL-DYN Emerald System is a bench top analyzer consisting of the main analyzer with data module, display station, and printer. The main analyzer, data module, and display station are housed in a single chassis. The printer is a stand-alone module.
The CELL-DYN Emerald open sampler is equipped to aspiratc blood from a collection tube that has been opened and is held under the open sample aspiration probe.
CELL-DYN Emerald™ System Traditional 510(k) Submission Page 1 of 6
510(k) Summary
{1}------------------------------------------------
Intended Use
The CELL-DYN Emerald System is an automated hematology analyzer designed for in vitro diagnostic use in clinical laboratories.
Similarities and Differences
The CELL-DYN Emerald System and the CELL-DYN 1800 System are similar in that:
- Both systems are bench top 3-part differential hematology with built-in monitors and data station.
Both systems analyze 16 parameters (measurands).
Both systems accept open specimens presented manually by the operator.
- Both systems automatically aspirate the specimen and present it for automated processing.
- Both systems use microprocessors for systems control, data acquisition, and data analysis.
Both systems provide RS232 interface to an on-line LIS.
- Both systems provide Histograms, Dispersional Data Alerts, Suspect Parameter Messages, and Critical Limit Flagging.
- Both systems use electrical impedance technology for counting and sizing cells.
Both systems use LED Hemoglobin analysis.
Both systems use a Cyanide-Free differential lyse reagent.
- Both systems are capable of inputting specimen information from a bar code through a hand held bar code scanner and both systems use alphanumeric specimen ID's.
Both systems are capable of data output to a dot matrix or inkjet printer.
The CELL-DYN Emerald System and the CELL-DYN 1800 System are different in that:
The CELL-DYN Emerald analyzer is smaller and more compact than the CELL-DYN 1800.
{2}------------------------------------------------
- The CELL-DYN Emerald System has a touch screen and built-in keypad while the CELL-DYN 1800 has an external keyboard.
- The CELL-DYN Emerald System has two USB ports and one serial port while the CELL-DYN 1800 System has a floppy drive and serial port for printing and data storage options.
The CELL-DYN Emerald System uses different boards and microprocessors.
- The CELL-DYN Emerald System has a password protection to secure software fields while the CELL-DYN 1800 does not have password protection.
- The CELL-DYN Emerald Cleaner reagent contains an enzymatic agent while the CELL-DYN 1800 has a separate enzymatic cleaner.
- The CELL-DYN Emerald aspirates K2EDTA-anticoagulated human whole blood specimens while the CELL-DYN 1800 uses K3EDTA-anticoagulated human whole blood specimens.
- The CELL-DYN Emerald sample size is approximately 9.8 micro liters while the CELL-DYN 1800 is approximately 30 micro liters.
Similarities and Differences Table:
| Predicate DeviceAbbott CELL-DYN® 1800 | Submission DeviceCELL-DYN Emerald™ | |
|---|---|---|
| Device Description | Bench top analyzer with built-inmonitor and data station | Same |
| Instrument Size | Height: 17 inches (44 cm)Width: 26 inches (66 cm)Depth: 21 inches (53 cm) | Height: 13.8 inches (35 cm)Width: 9.8 inches (25 cm)Depth: 13.8 inches (35 cm) |
| Intended Use | The CELL-DYN 1800 is amulti-parameter, automatedhematology analyzer designedfor in vitro diagnostic use inclinical laboratories. | The CELL-DYN Emerald is anautomated hematology analyzerdesigned for in vitro diagnosticuse in clinical laboratories. |
| Predicate DeviceAbbott CELL-DYN® 1800 | Submission DeviceCELL-DYN Emerald™ | |
| WBC Differential | 3-Part Differential | Same |
| Parameters | White Blood Cells:• WBC • GRAN• LYM • %GRAN• %LYM• MID• %MID | Same |
| Red Blood Cells:• RBC • MCV• HCT • RDW | ||
| Hemoglobin:• HGB• MCH• MCHC | ||
| Platelets:• PLT• MPV | ||
| Technology | • Electrical impedance | Same |
| • LED Hemoglobin Analysis | Same | |
| • Lysate modified differential | Same | |
| • Modified Methemoglobin forhemoglobin measurement | Same | |
| • Boards: Single BoardComputed | • Boards: Pre-Amp, CPU,IHM | |
| • Microprocessor: Celeron 850MHz | • Microprocessor: FreescaleColdfire | |
| • No Password Protection | • Password Protection | |
| Time to First Result | Approximately 60 seconds | Same |
| Sampling / Dilution | Open Mode Analysis:Manual presentation andaspiration of a well-mixed wholeblood specimen for automated | Open Mode Analysis:Manual presentation andaspiration of a well-mixed wholeblood specimen for automated |
| Predicate Device | Submission Device | |
| Abbott CELL-DYN® 1800 | CELL-DYN Emerald™ | |
| analysis; automatic dilution ofthe aspirated sample andautomatic presentation of eachdilution for measurement. | analysis; automatic dilution ofthe aspirated sample andautomatic presentation of eachdilution for measurement. | |
| Pre-Dilute Mode Analysis:Off line dilution of whole bloodspecimen and manualpresentation of each dilution formeasurement; manualpresentation and aspiration ofpre-diluted sample for automatedanalysis. | ||
| Specimen Type | K3EDTA-anticoagulated humanwhole blood for all parameters | K2EDTA-anticoagulated humanwhole blood for all parameters |
| Sample Size | Open Mode Analysis:30 µLPre-Dilute Mode Analysis:40 µL | Open Mode Analysis:9.8µL |
| Reagents | • Detergent• Diluent• CN-Free Lyse Reagent | • Cleaner (with enzymes)• Diluent• CN-Free Lyse Reagent |
| Histograms | RBC, WBC, PLT | Same |
| Data Output / DataInput Interface | Data Output:• Host RS232• Color Monitor• Dot Matrix Printer• Inkjet Printer• DiskettesData Input:• Keyboard (external)• Bar code reader (optional) | Data Output:• Host RS232• Ethernet• Color Monitor• Dot Matrix Printer• Inkjet Printer• USB PrinterData Input:• Keypad (internal)• Bar code reader (standard) |
| AlphanumericSpecimen ID | Yes | Same |
| Predicate DeviceAbbott CELL-DYN® 1800 | Submission DeviceCELL-DYN Emerald™ | |
| Electrical, Mechanical,and Thermal Safety | System certification to US andCanadian product safetystandards and EU low voltagesafety standards | Same |
CELL-DYN Emerald™ System Traditional 510(k) Submission
510(k) Summary
{3}------------------------------------------------
:
.
.
510(k) Summary
: ·
{4}------------------------------------------------
:
100 - 100 - 100
.
510(k) Summary
. .
{5}------------------------------------------------
Equivalency Data Summary
The CELL-DYN Emerald System is an automated hematology analyzer for in vitro diagnostic use in clinical laboratories. The CELL-DYN Emerald System, which includes reagents and software, was compared in both external as well as an internal clinical trial to the CELL-DYN 1800 System. The data compiled supports the claim that the CELL-DYN Emerald System is substantially equivalent to the CELL-DYN 1800 System and includes data for background, correlation, precision, linearity and carryover. Refer to Medical Clinical Summary Report (Attachment F) for the data supporting the substantially equivalent claim between the CELL-DYN Emerald and the CELL-DYN® 1800.
Conclusion
The CELL-DYN Emerald System is substantially equivalent to the CELL-DYN® 1800 (predicate device). The differences noted do not pose new questions of safety and effectiveness.
{6}------------------------------------------------
DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/6/Picture/1 description: The image shows the seal of the Department of Health & Human Services (HHS) of the United States. The seal features a stylized eagle with its wings spread, and the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the eagle. The seal is black and white.
Public Health Service
Food and Drug Administration 2098 Gaither Road Rockville MD 20850
FEB - 4 2009
Abbott Laboratories c/o Ms Michelle B. Roeding Section Manager, Regulatory Affairs 5440 Patrick Henry Drive Santa Clara, CA 95054
Re: K081495
Trade/Device Name: Cee-Dyn Emerald™ System Regulation Number: 21 CFR 864.5220 Regulation Name: Automated Differential Cell Counter Regulatory Class: Class II Product Code: GKZ Dated: January 16, 2009 Received: January 21, 2009
Dear Ms. Roeding:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820). This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
{7}------------------------------------------------
Page 2 - Ms. Michelle B. Roeding
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (240) 276-0450. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at (240) 276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at (240) 276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
Monroe M Chapp.
Maria M. Chan, Ph.D. Acting Division Director Division of Immunology and Hematology Devices Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health
Enclosure
{8}------------------------------------------------
Indication for Use
510(k) Number (if known): K081495
Device Name:
CELL-DYN Emerald™ System
Indication For Use:
The CELL-DYN Emerald is an automated hematology analyzer designed for in-vitro diagnostic use in clinical laboratories.
Prescription Use _____________________________________________________________________________________________________________________________________________________________ (21 CFR Part 801 Subpart D)
And/Or
Over the Counter Use (21 CFR Part 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE; CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Ja Vitry Diggnostic Device Evaluation and Safety (OVD)
Doughine Bautisti
Division Sign-Off Office of In Vitro Diagnostic Device Evaluation and Safety
510(k) K081495
§ 864.5220 Automated differential cell counter.
(a)
Identification. An automated differential cell counter is a device used to identify one or more of the formed elements of the blood. The device may also have the capability to flag, count, or classify immature or abnormal hematopoietic cells of the blood, bone marrow, or other body fluids. These devices may combine an electronic particle counting method, optical method, or a flow cytometric method utilizing monoclonal CD (cluster designation) markers. The device includes accessory CD markers.(b)
Classification. Class II (special controls). The special control for this device is the FDA document entitled “Class II Special Controls Guidance Document: Premarket Notifications for Automated Differential Cell Counters for Immature or Abnormal Blood Cells; Final Guidance for Industry and FDA.”