K Number
K081329
Date Cleared
2008-07-09

(58 days)

Product Code
Regulation Number
870.1220
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Reprocessed Electrophysiology (EP) Catheters are indicated for temporary intracardiac sensing, recording, stimulation and electrophysiological mapping of cardiac structures.

Device Description

Diagnostic Electrophysiology (EP) Catheters are specially designed electrode catheters that transmit electrical impulses and can be positioned for endocardial recording or stimulation. Diagnostic EP catheters incorporate a hand piece, a flexible shaft and a distal tip section containing diagnostic electrodes. The distal tip of deflectable catheters can be deflected into a curve by manipulating the hand piece. Reprocessed Electrophysiology Catheters are intended for temporary intracardiac sensing, recording, stimulation, and electrophysiology mapping of cardiac structures.

AI/ML Overview

The provided document is a 510(k) Summary for a reprocessed medical device, specifically Electrophysiology Catheters. This type of submission focuses on proving "substantial equivalence" to existing legally marketed devices (predicates) rather than establishing new performance claims or conducting clinical efficacy studies.

Therefore, the acceptance criteria and study information requested, which are typically associated with demonstrating novel performance or clinical superiority of a device, are not directly applicable in the context of this 510(k) submission.

Here's an analysis based on the available information:

1. Table of Acceptance Criteria and Reported Device Performance

This document does not present specific quantitative acceptance criteria or detailed device performance metrics in the way one would see for a new device's clinical trial or performance study. Instead, the performance demonstration focuses on showing that the reprocessed devices function identically to their predicate devices and meet safety standards after reprocessing.

The "Performance data" section states:

  • "Bench and laboratory testing was conducted to demonstrate performance (safety and effectiveness) of Reprocessed Electrophysiology Catheters."
  • This included:
    • Biocompatibility
    • Validation of reprocessing
    • Sterilization Validation
    • Function test(s)
    • Packaging Validation
  • "Performance testing demonstrates that Reprocessed Electrophysiology Catheters perform as originally intended."

Since there are no specific numeric acceptance criteria or quantitative performance results provided in the summary, a table cannot be populated as requested. The "acceptance criteria" here are implied to be adherence to the original device specifications and established reprocessing standards. The "reported device performance" is a general statement that they "perform as originally intended."

2. Sample Size Used for the Test Set and Data Provenance

The document does not specify the sample size used for the bench and laboratory testing. It also does not explicitly state the data provenance (e.g., country of origin, retrospective or prospective) for these tests. This information is typically found in the detailed test protocols and reports, which are not part of this summary.

3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications of those Experts

This is not applicable to the type of testing described (bench and laboratory). Bench testing for medical devices typically relies on engineering specifications and measurement against those specifications, not expert consensus on interpretations of qualitative data.

4. Adjudication Method for the Test Set

This is not applicable. Adjudication methods (e.g., 2+1, 3+1) are used for resolving disagreements among human readers or experts when establishing ground truth, typically in image interpretation or diagnostic studies. The tests mentioned (biocompatibility, sterilization, function, packaging) are objective, quantitative, or qualitative assessments against predefined standards, not subjective interpretations requiring adjudication.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

This is not applicable. The device described is a reprocessed electrophysiology catheter, which is a physical medical device, not an AI or software algorithm. Therefore, an MRMC study related to AI assistance for human readers is irrelevant.

6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) was done

This is not applicable as the device is not an algorithm or AI.

7. The Type of Ground Truth Used

For the bench and laboratory testing:

  • Biocompatibility: Ground truth would be adherence to established biocompatibility standards (e.g., ISO 10993).
  • Validation of reprocessing: Ground truth would be the successful removal of soil, decontamination, and functionality post-reprocessing, verified by established protocols and measurable criteria.
  • Sterilization Validation: Ground truth would be a Sterility Assurance Level (SAL) meeting an accepted standard (e.g., 10^-6), confirmed by established sterilization validation methods.
  • Function test(s): Ground truth would be that the reprocessed catheter performs according to the original manufacturer's specifications for function (e.g., electrical conductivity, deflection, structural integrity).
  • Packaging Validation: Ground truth would be that the packaging maintains sterility and integrity, confirmed by established packaging validation methods.

8. The Sample Size for the Training Set

This is not applicable. The device is a reprocessed physical medical device, not a machine learning model that requires a training set.

9. How the Ground Truth for the Training Set Was Established

This is not applicable, as there is no training set for a physical reprocessed device.

In Summary:

The provided document describes a 510(k) submission for a reprocessed electrophysiology catheter. The "acceptance criteria" and "study" described are fundamental safety and performance validations for a reprocessed medical device, ensuring it remains equivalent to the original, legally marketed device. It does not involve AI, clinical trials, or expert consensus studies for diagnostic interpretation. The focus is on demonstrating that the reprocessing does not compromise the device's original intended use, safety, or effectiveness.

{0}------------------------------------------------

JUL - 9 2008

SECTION 5: 510(k) SUMMARY

12/2 Ascent Healthcare Solutions Submitter: 10232 South 51st Street Phoenix, Arizona 85044 Katie Bray Contact: Regulatory Affairs Manager (480) 763-5300 (o) (480) 763-6089 (f) mbarton@ascenths.com Date of preparation: May 9, 2008 Trade/Proprietary Name: Reprocessed Electrophysiology Name of device: Catheters Classification Name: Electrode recording catheter or electrode recording probe Manufacturer 510{k) Title Predicate Device Inquiry™ AFocus ™, Inquiry ™ Irvine Biomedical K042775 AFocusII™, or Inquiry™ Optima™ Steerable Electrophysiology Catheter Irvine Biomedical Inquiry™ AFocus™ Steerable K010471 Electrophysiology Catheter Irvine Biomedical Orthogonal™ Steerable K990958 Electrophysiology Catheter Irvine Biomedical 1BI-1100 Bi-directional Steerable K982232 Electrophysiology Catheter IBI-1100 Steerable Electrophysiology Irvine Biomedical K961924 Catheter Irvine Biomedical IBI-1000™ Electrophysiology K946333 Catheter Diagnostic Electrophysiology (EP) Catheters are specially Device description: designed electrode catheters that transmit electrical impulses and can be positioned for endocardial recording or stimulation. Diagnostic EP catheters incorporate a hand piece, a flexible shaft and a distal tip section containing diagnostic electrodes. The distal tip of deflectable catheters can be deflected into a curve by manipulating the hand piece. Reprocessed Electrophysiology Catheters are intended for Indications for Use: temporary intracardiac sensing, recording, stimulation, and electrophysiology mapping of cardiac structures. The design, materials, and intended use of Reprocessed Technological Electrophysiology Catheters are identical to the predicate characteristics: devices. The mechanism of action of Reprocessed Electrophysiology Catheters is identical to the predicate devices

Ascent Healthcare Solutions Reprocessed Electrophysiology Catheters Traditional 510(k)

Page 13

K081329

{1}------------------------------------------------

K081329

in that the same standard mechanical design, materials, and in that the cultilized. There are no changes to the claims, intended use, clinical applications, patient population, performance specifications, or method of operation. In addition, Ascent Healthcare Solutions' reprocessing of Electrophysiology Catheters includes removal of adherent visible soil and decontamination. Each individual Electrophysiology Catheter is tested for appropriate function of its components prior to packaging and labeling operations.

Performance data:

Bench and laboratory testing was conducted to demonstrate performance (safety and effectiveness) of Reprocessed Electrophysiology Catheters. This included the following tests:

  • Biocompatibility .
  • Validation of reprocessing .
  • Sterilization Validation .
  • Function test(s) .
  • Packaging Validation .

Performance testing demonstrates that Reprocessed Electrophysiology Catheters perform as originally intended.

Conclusion:

Ascent Healthcare Solutions concludes that the modified devices (Reprocessed Electrophysiology Catheters) are safe, effective, and substantially equivalent to the predicate devices as described herein.

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" arranged around the perimeter. Inside the circle is an abstract symbol resembling a stylized human figure with three horizontal lines extending from its arm.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

JUL - 9 2008

Ascent Healthcare Solutions c/o Ms. Katie Bray Regulatory Affairs Manager, 10232 South 51st Street Phoenix, AZ 85044

Re: K081329

Reprocessed Electrophysiology Catheters (See Enclosed List) Regulation Number: 21 CFR 870.1220 Regulation Name: Electrode Recording Catheter or Electrode Recording Probe Regulatory Class: Class II (two) Product Codc: NLH Dated: May 9, 2008 Received: May 12, 2008

Dear Ms. Bray:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{3}------------------------------------------------

Pagc 2 - Ms. Katie Bray

Please be advised that FDA's issuance of a substantial cquivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0120. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely yours,

Bram D. Zuckerman, M

Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

Page 3 - Ms. Katie Bray

List of Models Found SE:

ItemNo.DescriptionTrade NameElectrodesFrenchSizeElectrodeSpacingCurve Type
800011010-6-25-JInquiry Fixed Curve106F2-5-2 mmJosephson
800021010-6-25-CInquiry Fixed Curve106F2-5-2 mmCournand
800031010-6-28-SCInquiry Fixed Curve106F2-8-2 mmSpecial
800511010-5-2-JInquiry Fixed Curve105F2 mmJosephson
800521010-5-25-JInquiry Fixed Curve105F2-5-2 mmJosephson
800551010-5-25-CInquiry Fixed Curve105F2-5-2 mmCournand
800631010-5-28-CInquiry Fixed Curve105F2-8-2 mmSpecial Curve
800641010-5-28-SC(60)Inquiry Fixed Curve105F2-8-2 mmSpecial Curve
800651010-5-25-SC(60)Inquiry Fixed Curve105F2-5-2 mmSpecial Curve
801161010-6-2-J-TE4BE4Inquiry Fixed Curve106F2 mmJosephson
801311010-4-5-CInquiry Fixed Curve104F5 mmCournand
801321010-4-5(10)5-J1Inquiry Fixed Curve104F2-5-2 mmCournand
801331010-4-2-JInquiry Fixed Curve104F2 mmJosephson
801341010-4-5-JInquiry Fixed Curve104F5 mmJosephson
801351010-4-25-JInquiry Fixed Curve104F2-5-2 mmJosephson
801371010-4-5(50)-J1Inquiry Fixed Curve104F5(50)5 mmJosephson 1
801381010-4-5(10)-J1Inquiry Fixed Curve104F5(50)5 mmJosephson 1
801391010-6-28-SC(SOFT)Inquiry Fixed Curve106F2-8-2 mmSpecial
804051004-6-25-JInquiry Fixed Curve46F2-5-2 mmJosephson
804061004-6-5-CInquiry Fixed Curve46F5 mmJosephson
804071004-6-2-7Inquiry Fixed Curve46F2 mmCournand
804081004-6-5-JInquiry Fixed Curve46F2-5-2 mmCournand
804091004-6-5-CInquiry Fixed Curve46F5 mmCournand
804111004-6-25-DInquiry Fixed Curve46F2-5-2 mmDomato
804121004-6-5-DInquiry Fixed Curve46F5 mmDomato
804131004-6-10-JInquiry Fixed Curve46F10 mmJosephson
804141004-6-10-CInquiry Fixed Curve46F10 mmCournand
804151004-6-10-DInquiry Fixed Curve46F10 mmDamato
804401004-5-5-C(HIS)(SOFT)Inquiry HIS FixedCurve45F5 mmCournand
804511004-5-2-JInquiry Fixed Curve45F2 mmJosephson
804521004-5-25-JInquiry Fixed Curve45F2-5-2 mmJosephson
804531004-5-5-JInquiry Fixed Curve45F5 mmJosephson
804551004-5-25-CInquiry Fixed Curve45F2-5-2 mmCournand
804561004-5-5-CInquiry Fixed Curve45F5 mmCournand
804581004-5-25-CInquiry Fixed Curve45F2-5-2 mmDamato
ItemNo.DescriptionTrade NameElectrodesFrenchSizeElectrodeSpacingCurve Type
804591004-5-5-DInquiry Fixed Curve45F5 mmDomato
804631004-5-10-CInquiry Fixed Curve45F10 mmCournand
804641004-4-2-CInquiry Fixed Curve44F2 mmCournand
804651004-4-5-CInquiry Fixed Curve44F5 mmCournand
804661004-4-25-CInquiry Fixed Curve44F2-5-2 mmCournand
804671004-4-2-JInquiry Fixed Curve44F2 mmJosephson
804681004-4-5-JInquiry Fixed Curve44F5 mmJosephson
804691004-4-25-JInquiry Fixed Curve44F2-5-2 mmJosephson
804731004-5-5J(SOFT)Inquiry Fixed Curve45F5 mmJosephson
804761004-5-5-C (SOFT)Inquiry Fixed Curve45F5 mmCournand
804791004-5-5-D(SOFT)Inquiry Fixed Curve45F5 mmDomato
804841004-5-5-J1-TE1BE2Inquiry Fixed Curve45F5 mmJosephson 1
804851004-4-5-J1Inquiry Fixed Curve44F5 mmJosephson 1
805011005-6-25-CInquiry Fixed Curve56F2-5-2 mmCournand
805081005-6-25-JInquiry Fixed Curve56F2-5-2 mmJosephson
805161004-6-2-C(SOFT)Inquiry Fixed Curve46F2 mmCournand
805171004-6-25-J(SOFT)Inquiry Fixed Curve46F2-5-2 mmJosephson
805181004-6-5-J1(SOFT)Inquiry Fixed Curve46F10 mmJosephson 1
805191004-6-10-D(SOFT)Inquiry Fixed Curve46F10 mmDamato
805201004-6-10-C(SOFT)Inquiry Fixed Curve46F10 mmCournand
805331010-5-28-SC(SOFT)Inquiry Fixed Curve105F2-8-2 mmSpecial
805341010-5-5(30)5-C(SOFT)Inquiry Fixed Curve105F5(30)5Cournand
805351004-4-25-J1Inquiry Fixed Curve44F2-5-2 mmJosephson 1
805361004-4-5-DInquiry Fixed Curve44F5 mmDamato
805371004-6-10-J(SOFT)Inquiry Fixed Curve46F10 mmJosephson
805441010-5-28-C(SOFT)Inquiry Fixed Curve105F2-8-2 mmCournand
805671004-4-5-C(HIS)Inquiry HIS FixedCurve44F5 mmCournand
806021006-6-25-JInquiry Fixed Curve66F2-5-2 mmJosephson
806031006-6-5-JInquiry Fixed Curve66F5 mmJosephson
806041006-6-2-CInquiry Fixed Curve66F2 mmCournand
806061006-6-5-CInquiry Fixed Curve66F5 mmCournand
806271006-5-5-J1 (SOFT)Inquiry Fixed Curve65F5 mmJosephson 1
808031008-6-5-JInquiry Fixed Curve86F5 mmJosephson
808041008-6-2-CInquiry Fixed Curve86F2 mmCournand
808061008-6-5-DInquiry Fixed Curve86F5 mmDamato
808101008-6-5-CInquiry Fixed Curve86F5 mmCournand
808201008-5-5-(20)5-C(HIS)Inquiry HIS FixedCurve85F20(5) mmCournand
809001014-7-3(70)-TE2BE2Inquiry Fixed Curve3-3-3-3-3-3-70-3-3-3-3-3-3mmCournand
Item No.DescriptionTrade Name14Electrodes7FFrench SizeElectrode SpacingCurve Type
811011110-6-2-FInquiry Steerable Curves106F2 mmMed 3.2 cm
811021110-6-25-MInquiry Steerable Curves106F2-5-2 mmMed 3.2 cm
811041110-6-25-LInquiry Steerable Curves106F2-5-2 mmLarge 4.1 cm
811051110-6-25-XLInquiry Steerable Curves106F2-5-2 mmX-Large 5.0 cm
811061110-6-25-FInquiry Steerable Curves106F2-5-2 mmFar Reach
811071110-6-5-LInquiry Steerable Curves106F5 mmLarge
811081110-6-2-LInquiry Steerable Curves106F2 mmLarge
811091110-6-2-L-TE2BE2Inquiry Steerable Curves106F2 mmLarge
811101110-6-17-HInquiry H Steerable Atrial106F1 & 7 mmH
811201120-7-17-HInquiry H Steerable Atrial207F1 & 7 mmH
811211121-7-17-H-SCInquiry H Steerable Atrial217F1-7-1 mmH-SC
811241124-7-271-HInquiry H Steerable Atrial247F2-7-1 mmH
811251106-6-27-HInquiry H Steerable Atrial66F2-7-2 mmH
811261110-6-291-HLInquiry H SteerableAtrial106F2-9-1 mmHL
811281121-7-5-H-UPInquiry H Steerable Atrial217F5 mmH
811301120-7-19-HLInquiry H Steerable Atrial207F1 & 9 mmHL
811311121-7-19-HL-SCInquiry H Steerable Atrial217F1-9-1 mmHL-SC
811341124-7-291-HLInquiry H Steerable Atrial247F2-9-1 mmHL
811361120-7-5-HL-UPInquiry H Steerable Atrial207F5 mmHL
811421106-6-27-HLInquiry H Steerable Atrial66F2-7-2 mmHL
811501124-7-271-H-SCEInquiry H Steerable Atrial247F2-7-1 mmH-SCE
811711110-5-2-MInquiry Steerable Curves105F2 mmMed 3.2 cm
811721110-5-25-MInquiry SteerableCurves105F2-5-2 mmMed 3.2 cm
811741110-5-25-LInquiry Steerable Curves105F2-5-2 mmLarge 4.1 cm
811761110-5-25-LInquiry Steerable Curves105F2-5-2 mmFar Reach
Item No.DescriptionTrade NameElectrodesFrench SizeElectrode SpacingCurve Type
811771110-5-5-LInquiry Steerable Curves105F5 mmLarge 4.1 cm
811781110-5-2-LInquiry Steerable Curves105F2 mmLarge 4.1 cm
811791110-5-2-EInquiry Steerable Curves105F2 mmExtended Reach
812021120-7-2-10-XXLTen-Ten Duodecapolar207F2-10 mmXX Large 4.8 cm
812031110-5-25-M(80)SOFTInquiry Soft Tip Steerable105F2-5-2 mmMed 3.2 cm
812071120-7-5-SLTen-Ten Duodecapolar207F5 mmSuper Lrg 5.1 cm
812091120-7-25-SLTen-Ten Duodecapolar207F2-5 mmSuper Lrg 5.1 cm
812111120-7-2(20)2(25)2-SLTen-Ten Duodecapolar207F2(20)2(25)2 mmSuper Lrg 5.1 cm
812231110-5-2(20)3-XLInquiry Steerable Curves105F2(50)3x-large
814011104-6-5-SInquiry Steerable Curves46F5 mmSmall 2.7 cm
814021104-6-25-MInquiry Steerable Curves46F2-5-2 mmMedium 3.2 cm
814031104-6-5-MInquiry Steerable Curves46F5 mmMedium 3.2 cm
814041104-6-25-LInquiry Steerable Curves46F2-5-2 mmLarge 4.2 cm
814051104-6-5-LInquiry Steerable Curves46F5 mmLarge 4.2 cm
814061104-6-25-EInquiry Steerable Curves46F2-5-2 mmExtended Reach
814071104-6-25-FInquiry Steerable Curves46F2-5-2 mmFar Reach
814121104-6-2-EInquiry Steerable Curves46F2 mmExtended Reach
814171104-6-5-XLInquiry Steerable Curves46F5 mmX-Large 5.0 cm
814181104-6-25-XLInquiry Steerable Curves46F2-5-2 mmX-Large 5.0 cm
814711104-5-5-SInquiry Steerable Curves45F5 mmSmall 2.7 cm
814721104-5-25-MInquiry Steerable Curves45F2-5-2 mmMed 3.2 cm
814731104-5-5-MInquiry Steerable Curves45F5 mmMed 3.2 cm
814741104-5-25-LInquiry SteerableCurves45F2-5-2 mmLarge 4.1 cm
814751104-5-5-LInquiry SteerableCurves45F5 mmLarge 4.1 cm
814781104-5-2-SInquiry SteerableCurves45F2 mmSmall 2.7 cm
ItemNo.DescriptionTrade NameElectrodesFrenchSizeElectrodeSpacingCurve Type
814791104-5-25-SInquiry SteerableCurves45F2-5-2 mmSmall 2.7 cm
814831104-5-5-E(HIS)(SOFT)Inquiry IIIS SteerableCurve45F5 mmExtended
814841104-5-5-L(EX-HIS)(SOFT)Inquiry HIS SteerableCurve45F5 mmLarge
814851104-5-5-E(EX-HIS)(SOFT)Inquiry HIS SteerableCurve45F5 mmExtended
815031104-6-5-L-TE2BE2-BDInquiry Bi-Directional46F5 mmLarge 4.1 cm
815041110-6-5-M-TE2BE2-BDInquiry Bi-Directional106F5 mmMed 3.2 cm
815111105-6-25-MInquiry SteerableCurves66F2-5-2 mmMed 3.2 cm
815201110-6-2-XL-TE4BE4Inquiry SteerableCurves106F2 mmX-Large 5.0cm
815211105-5-5(170)-M/L-TE1BE2(SOFT)Inquiry Soft TipSteerable55F5(170)Med/Large
815241110-6-2-L-TE4BE4Inquiry SteerableCurves106F2 mmLarge 4.1 cm
815301110-4-2-MInquiry SteerableCurves104F2 mmMed 3.2 cm
815311110-4-25-MInquiry SteerableCurves104F2-5-2 mmMed 3.2 cm
815321110-4-025-LInquiry SteerableCurves104F2-5-2 mmLarge 4.1 cm
815341110-4-5-LInquiry SteerableCurves104F5 mmLarge 4.1 cm
815361110-4-25-M(80)Inquiry SteerableCurves104F2-5-2 mmMed 3.2 cm
815371110-4-25-M(SC)(60)Inquiry SteerableCurves104F2-5-2 mmMed 3.2 cm
815401104-4-25-MInquiry SteerableCurves44F2-5-2 mmMed 3.2 cm
815411104-4-2-MInquiry SteerableCurves44F2 mmMed 3.2 cm
815421104-4-5-MInquiry SteerableCurves44F5 mmMed 3.2 cm
815431104-4-25-LInquiry SteerableCurves44F2-5-2 mmLarge 4.1 cm
815451104-4-5-LInquiry SteerableCurves44F5 mmLarge 4.1 cm
815871120-7-1(2.5)-SM-AF20A-Focus II Steerable Atrial207F1(2.5) mm20 mm diameter
815891114-7-15-SM-AF20A-Focus II Steerable Atrial147F1-5-1 mm20 mm diameter
815911114-7-13-SM-AF15A-Focus II Steerable Atrial147F1-3 mm15 mm diameter
Item No.DescriptionTrade NameElectrodesFrench SizeElectrode SpacingCurve Type
815941110-7-3.5-SM-AF15A-Focus II Steerable Atrial107F3.5 mm15 mm diameter
815951110-7-5-SM-AFA-Focus II Steerable Atrial107F5 mm20 mm diameter
815961110-7-27-SM-AF20A-Focus II Steerable Atrial107F2-7 mm20 mm diameter
815971110-7-2(11)-SM-AF25A-Focus II Steerable Atrial107F2(11) mm25 mm diameter
815981120-7-1(4.5)-SM-AF25A-Focus II Steerable Atrial207F1(4.5) mm25 mm diameter
815991110-7-8-SM-AF25A-Focus II Steerable Atrial107F8 mm25 mm diameter
816011106-6-5-MInquiry Steerable Curves66F5 mmMed 3.2 cm
816021106-6-5-LInquiry Steerable Curves66F5 mmLarge 4.1 cm
816031106-6-5-EInquiry Steerable Curves66F5 mmExtended Reach
816041106-6-5-FInquiry Steerable Curves66F5 mmFar Reach
816051106-6-5-XLInquiry Steerable Curves66F5 mmX-Large 5.0 cm
816701112-5-2-(10)-AF20A-Focus I Steerable Atrial125F2(10) mm25 mm diameter
816711114-5-2(9)-AF30A-Focus I Steerable Atrial145F2(9) mm30 mm diameter
816721110-5-27-SM-AF20A-Focus II Steerable Atrial105F2-7 mm20 mm diameter
816731110-5-5-SM-AF20A-Focus II Steerable Atrial105F5 mm20 mm diameter
816741110-5-5-AF20A-Focus I Steerable Atrial105F5 mm20 mm diameter
816761110-5-3.5-AF15A-Focus I Steerable Atrial105F3.5 mm15 mm diameter
816801110-5-3.5-SM-AF15A-Focus II Steerable Atrial105F3.5 mm15 mm diameter
816831120-7-1(4.5)-SM-OPT25Optima Steerable Lasso205F1(4.5) mm25 mm - 15 mm
Item No.DescriptionTrade NameElectrodesFrench SizeElectrode SpacingCurve Type
816871110-7-10-SM-OPT25Optima Steerable Lasso107F10mm25 mm - 15 mm
817031110-5-28-M/L(SC)(XS)(SOFT)Inquiry Soft Tip Steerable105F2-8-2 mmMed 3.2 cm/Large 4.1 cm
817041110-5-5-M/L(SC)(XS)(SOFT)Inquiry Soft Tip Steerable105F5 mmMed 3.2 cm/Large SC 4.1 cm
817051110-5-28-M(SC)(XS)(SOFT)Inquiry Soft Tip Steerable105F2-8-2 mmM(SC)(XS)
817061110-5-5-M/L(XS)(SOFT)Inquiry Soft Tip Steerable105F5 mmM/L(XS)
817111108-5-5(20)5-E(HIS)(SOFT)Inquiry HIS Steerable Curve85F5(20)5 mmExtended
817171124-7-1(4.5)(20)(3)-SM-OPT25Optima Steerable Lasso247F4(4.5)25 mm - 15 mm
817211110-5-25-M(SC)(SOFT)Inquiry Soft Tip Steerable105F2-5-2 mmMed (SC) 3.2 cm
817301110-5-28-M/L(SOFT)Inquiry Soft Tip Steerable105F2-8-2 mmM/L
817321110-5-25-M(SOFT)Inquiry Soft Tip Steerable105F2-5-2 mmMed 3.2 cm
817331110-5-2-L(SOFT)Inquiry Soft Tip Steerable105F2 mmLarge 4.1 cm
817341110-5-25-L(SOFT)Inquiry Soft Tip Steerable105F2-5-2 mmLarge 4.1 cm
817351110-5-5-L(SOFT)Inquiry Soft Tip Steerable105F5 mmLarge 4.1 cm
817381110-5-1-L(SOFT)Inquiry Soft Tip Steerable105F1 mmLarge 4.1 cm
817421104-5-25-M(SOFT)Inquiry Soft Tip Steerable45F2-5-2 mmMed 3.2 cm
817431104-5-5-M(SOFT)Inquiry Soft Tip Steerable45F5 mmMed 3.2 cm
817451104-5-25-L(SOFT)Inquiry Soft Tip Steerable45F2-5-2 mmLarge 4.1 cm
817501110-5-25-L(60cm)(SOFT)Inquiry Soft Tip Steerable105F2-5-2 mmLarge 4.1 cm
817511108-5-2-M(SOFT)Inquiry Soft Tip Steerable85F2 mmMed 3.2 cm
818011108-6-2-MInquiry Steerable Curves86F2 mmMed 3.2 cm
818021108-6-25-MInquiry Steerable Curves86F2-5-2 mmMed 3.2 cm
818071108-6-2-LInquiry Steerable Curves86F2 mmLarge 4.1 cm
818091108-6-25-LInquiry Steerable Curves86F2-5-2 mmLarge 4.1 cm
818231108-6-2-M(SOFT)Inquiry Soft TipSteerable86F2 mmMed 3.2 cm
818711108-5-2-MInquiry SteerableCurves85F2 mmMed 3.2 cm
818721108-5-25-MInquiry SteerableCurves85F2-5-2 mmMed 3.2 cm
818731108-5-5-MInquiry SteerableCurves85F5 mmMed 3.2 cm
818771108-5-2-LInquiry SteerableCurves85F2 mmLarge 4.1 cm
818791108-5-25-LInquiry SteerableCurves85F2-5-2 mmLarge 4.1 cm
ItemNo.DescriptionTrade NameElectrodesFrenchSizeElectrodeSpacingCurve Type
819011120-7-13-MTen-Ten Duodecapolar207F1-3 mmMedium 3.2cm
819021120-7-13-LTen-Ten Duodecapolar207F1-3 mmLarge 4.2 cm
819401110-6-2-M(SOFT)Inquiry Soft TipSteerable106F2 mmMed 3.2 cm
819431110-6-2-L(SOFT)Inquiry Soft TipSteerable106F2 mmLarge 4.1 cm
819451110-6-25-L(SOFT)Inquiry Soft TipSteerable106F2-5-2 mmLarge 4.1 cm
819471110-6-5-M/L(SOFT)Inquiry Soft TipSteerable106F5 mmM/L
819511104-6-5-M(SOFT)Inquiry Soft TipSteerable46F5 mmMed 3.2 cm
819541104-6-5-L(SOFT)Inquiry Soft TipSteerable46F5 mmLarge 4.1 cm
859311924-SInquiry DiagnosticConnecting Cablesn/an/a1.5m lengthn/a
859531910-SAInquiry DiagnosticConnecting Cablesn/an/a1.5m lengthn/a
859541904-SAInquiry DiagnosticConnecting Cablesn/an/a1.5m lengthn/a
859551914-SAInquiry DiagnosticConnecting Cablesn/an/a1.5m lengthn/a
859301910-SInquiry DiagnosticConnecting Cablesn/an/a2.5m lengthn/a

.

{5}------------------------------------------------

:

{6}------------------------------------------------

·

{7}------------------------------------------------

Page 6 -- Ms. Katie Bray

{8}------------------------------------------------

Page 7 – Ms. Katie Bray

{9}------------------------------------------------

Page 8 – Ms. Katic Bray

:

:

{10}------------------------------------------------

Page 9 -- Ms. Katie Bray

{11}------------------------------------------------

:

:

Page 10 ·· Ms. Katie Bray

.

{12}------------------------------------------------

SECTION 4: INDICATIONS FOR USE STATEMENT

510(k) Number (if known):

Device Name: Reprocessed Electrophysiology Catheters

Indications For Use: Reprocessed Electrophysiology (EP) Catheters are indicated for temporary intracardiac sensing, recording, stimulation and electrophysiological mapping of cardiac structures.

Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR

Over-The-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

eamall

Division of Cardiovascular Devices

Kos1329 510(k) Number

Ascent Healthcare Solutions Reprocessed Electrophysiology Catheters Traditional 510(k)

§ 870.1220 Electrode recording catheter or electrode recording probe.

(a)
Identification. An electrode recording catheter or an electrode recording probe is a device used to detect an intracardiac electrocardiogram, or to detect cardiac output or left-to-right heart shunts. The device may be unipolar or multipolar for electrocardiogram detection, or may be a platinum-tipped catheter which senses the presence of a special indicator for cardiac output or left-to-right heart shunt determinations.(b)
Classification. Class II (performance standards).