(58 days)
Reprocessed Electrophysiology (EP) Catheters are indicated for temporary intracardiac sensing, recording, stimulation and electrophysiological mapping of cardiac structures.
Diagnostic Electrophysiology (EP) Catheters are specially designed electrode catheters that transmit electrical impulses and can be positioned for endocardial recording or stimulation. Diagnostic EP catheters incorporate a hand piece, a flexible shaft and a distal tip section containing diagnostic electrodes. The distal tip of deflectable catheters can be deflected into a curve by manipulating the hand piece. Reprocessed Electrophysiology Catheters are intended for temporary intracardiac sensing, recording, stimulation, and electrophysiology mapping of cardiac structures.
The provided document is a 510(k) Summary for a reprocessed medical device, specifically Electrophysiology Catheters. This type of submission focuses on proving "substantial equivalence" to existing legally marketed devices (predicates) rather than establishing new performance claims or conducting clinical efficacy studies.
Therefore, the acceptance criteria and study information requested, which are typically associated with demonstrating novel performance or clinical superiority of a device, are not directly applicable in the context of this 510(k) submission.
Here's an analysis based on the available information:
1. Table of Acceptance Criteria and Reported Device Performance
This document does not present specific quantitative acceptance criteria or detailed device performance metrics in the way one would see for a new device's clinical trial or performance study. Instead, the performance demonstration focuses on showing that the reprocessed devices function identically to their predicate devices and meet safety standards after reprocessing.
The "Performance data" section states:
- "Bench and laboratory testing was conducted to demonstrate performance (safety and effectiveness) of Reprocessed Electrophysiology Catheters."
- This included:
- Biocompatibility
- Validation of reprocessing
- Sterilization Validation
- Function test(s)
- Packaging Validation
- "Performance testing demonstrates that Reprocessed Electrophysiology Catheters perform as originally intended."
Since there are no specific numeric acceptance criteria or quantitative performance results provided in the summary, a table cannot be populated as requested. The "acceptance criteria" here are implied to be adherence to the original device specifications and established reprocessing standards. The "reported device performance" is a general statement that they "perform as originally intended."
2. Sample Size Used for the Test Set and Data Provenance
The document does not specify the sample size used for the bench and laboratory testing. It also does not explicitly state the data provenance (e.g., country of origin, retrospective or prospective) for these tests. This information is typically found in the detailed test protocols and reports, which are not part of this summary.
3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications of those Experts
This is not applicable to the type of testing described (bench and laboratory). Bench testing for medical devices typically relies on engineering specifications and measurement against those specifications, not expert consensus on interpretations of qualitative data.
4. Adjudication Method for the Test Set
This is not applicable. Adjudication methods (e.g., 2+1, 3+1) are used for resolving disagreements among human readers or experts when establishing ground truth, typically in image interpretation or diagnostic studies. The tests mentioned (biocompatibility, sterilization, function, packaging) are objective, quantitative, or qualitative assessments against predefined standards, not subjective interpretations requiring adjudication.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
This is not applicable. The device described is a reprocessed electrophysiology catheter, which is a physical medical device, not an AI or software algorithm. Therefore, an MRMC study related to AI assistance for human readers is irrelevant.
6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) was done
This is not applicable as the device is not an algorithm or AI.
7. The Type of Ground Truth Used
For the bench and laboratory testing:
- Biocompatibility: Ground truth would be adherence to established biocompatibility standards (e.g., ISO 10993).
- Validation of reprocessing: Ground truth would be the successful removal of soil, decontamination, and functionality post-reprocessing, verified by established protocols and measurable criteria.
- Sterilization Validation: Ground truth would be a Sterility Assurance Level (SAL) meeting an accepted standard (e.g., 10^-6), confirmed by established sterilization validation methods.
- Function test(s): Ground truth would be that the reprocessed catheter performs according to the original manufacturer's specifications for function (e.g., electrical conductivity, deflection, structural integrity).
- Packaging Validation: Ground truth would be that the packaging maintains sterility and integrity, confirmed by established packaging validation methods.
8. The Sample Size for the Training Set
This is not applicable. The device is a reprocessed physical medical device, not a machine learning model that requires a training set.
9. How the Ground Truth for the Training Set Was Established
This is not applicable, as there is no training set for a physical reprocessed device.
In Summary:
The provided document describes a 510(k) submission for a reprocessed electrophysiology catheter. The "acceptance criteria" and "study" described are fundamental safety and performance validations for a reprocessed medical device, ensuring it remains equivalent to the original, legally marketed device. It does not involve AI, clinical trials, or expert consensus studies for diagnostic interpretation. The focus is on demonstrating that the reprocessing does not compromise the device's original intended use, safety, or effectiveness.
{0}------------------------------------------------
JUL - 9 2008
SECTION 5: 510(k) SUMMARY
12/2 Ascent Healthcare Solutions Submitter: 10232 South 51st Street Phoenix, Arizona 85044 Katie Bray Contact: Regulatory Affairs Manager (480) 763-5300 (o) (480) 763-6089 (f) mbarton@ascenths.com Date of preparation: May 9, 2008 Trade/Proprietary Name: Reprocessed Electrophysiology Name of device: Catheters Classification Name: Electrode recording catheter or electrode recording probe Manufacturer 510{k) Title Predicate Device Inquiry™ AFocus ™, Inquiry ™ Irvine Biomedical K042775 AFocusII™, or Inquiry™ Optima™ Steerable Electrophysiology Catheter Irvine Biomedical Inquiry™ AFocus™ Steerable K010471 Electrophysiology Catheter Irvine Biomedical Orthogonal™ Steerable K990958 Electrophysiology Catheter Irvine Biomedical 1BI-1100 Bi-directional Steerable K982232 Electrophysiology Catheter IBI-1100 Steerable Electrophysiology Irvine Biomedical K961924 Catheter Irvine Biomedical IBI-1000™ Electrophysiology K946333 Catheter Diagnostic Electrophysiology (EP) Catheters are specially Device description: designed electrode catheters that transmit electrical impulses and can be positioned for endocardial recording or stimulation. Diagnostic EP catheters incorporate a hand piece, a flexible shaft and a distal tip section containing diagnostic electrodes. The distal tip of deflectable catheters can be deflected into a curve by manipulating the hand piece. Reprocessed Electrophysiology Catheters are intended for Indications for Use: temporary intracardiac sensing, recording, stimulation, and electrophysiology mapping of cardiac structures. The design, materials, and intended use of Reprocessed Technological Electrophysiology Catheters are identical to the predicate characteristics: devices. The mechanism of action of Reprocessed Electrophysiology Catheters is identical to the predicate devices
Ascent Healthcare Solutions Reprocessed Electrophysiology Catheters Traditional 510(k)
Page 13
{1}------------------------------------------------
in that the same standard mechanical design, materials, and in that the cultilized. There are no changes to the claims, intended use, clinical applications, patient population, performance specifications, or method of operation. In addition, Ascent Healthcare Solutions' reprocessing of Electrophysiology Catheters includes removal of adherent visible soil and decontamination. Each individual Electrophysiology Catheter is tested for appropriate function of its components prior to packaging and labeling operations.
Performance data:
Bench and laboratory testing was conducted to demonstrate performance (safety and effectiveness) of Reprocessed Electrophysiology Catheters. This included the following tests:
- Biocompatibility .
- Validation of reprocessing .
- Sterilization Validation .
- Function test(s) .
- Packaging Validation .
Performance testing demonstrates that Reprocessed Electrophysiology Catheters perform as originally intended.
Conclusion:
Ascent Healthcare Solutions concludes that the modified devices (Reprocessed Electrophysiology Catheters) are safe, effective, and substantially equivalent to the predicate devices as described herein.
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" arranged around the perimeter. Inside the circle is an abstract symbol resembling a stylized human figure with three horizontal lines extending from its arm.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
JUL - 9 2008
Ascent Healthcare Solutions c/o Ms. Katie Bray Regulatory Affairs Manager, 10232 South 51st Street Phoenix, AZ 85044
Re: K081329
Reprocessed Electrophysiology Catheters (See Enclosed List) Regulation Number: 21 CFR 870.1220 Regulation Name: Electrode Recording Catheter or Electrode Recording Probe Regulatory Class: Class II (two) Product Codc: NLH Dated: May 9, 2008 Received: May 12, 2008
Dear Ms. Bray:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{3}------------------------------------------------
Pagc 2 - Ms. Katie Bray
Please be advised that FDA's issuance of a substantial cquivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0120. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
Bram D. Zuckerman, M
Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
Page 3 - Ms. Katie Bray
List of Models Found SE:
| ItemNo. | Description | Trade Name | Electrodes | FrenchSize | ElectrodeSpacing | Curve Type |
|---|---|---|---|---|---|---|
| 80001 | 1010-6-25-J | Inquiry Fixed Curve | 10 | 6F | 2-5-2 mm | Josephson |
| 80002 | 1010-6-25-C | Inquiry Fixed Curve | 10 | 6F | 2-5-2 mm | Cournand |
| 80003 | 1010-6-28-SC | Inquiry Fixed Curve | 10 | 6F | 2-8-2 mm | Special |
| 80051 | 1010-5-2-J | Inquiry Fixed Curve | 10 | 5F | 2 mm | Josephson |
| 80052 | 1010-5-25-J | Inquiry Fixed Curve | 10 | 5F | 2-5-2 mm | Josephson |
| 80055 | 1010-5-25-C | Inquiry Fixed Curve | 10 | 5F | 2-5-2 mm | Cournand |
| 80063 | 1010-5-28-C | Inquiry Fixed Curve | 10 | 5F | 2-8-2 mm | Special Curve |
| 80064 | 1010-5-28-SC(60) | Inquiry Fixed Curve | 10 | 5F | 2-8-2 mm | Special Curve |
| 80065 | 1010-5-25-SC(60) | Inquiry Fixed Curve | 10 | 5F | 2-5-2 mm | Special Curve |
| 80116 | 1010-6-2-J-TE4BE4 | Inquiry Fixed Curve | 10 | 6F | 2 mm | Josephson |
| 80131 | 1010-4-5-C | Inquiry Fixed Curve | 10 | 4F | 5 mm | Cournand |
| 80132 | 1010-4-5(10)5-J1 | Inquiry Fixed Curve | 10 | 4F | 2-5-2 mm | Cournand |
| 80133 | 1010-4-2-J | Inquiry Fixed Curve | 10 | 4F | 2 mm | Josephson |
| 80134 | 1010-4-5-J | Inquiry Fixed Curve | 10 | 4F | 5 mm | Josephson |
| 80135 | 1010-4-25-J | Inquiry Fixed Curve | 10 | 4F | 2-5-2 mm | Josephson |
| 80137 | 1010-4-5(50)-J1 | Inquiry Fixed Curve | 10 | 4F | 5(50)5 mm | Josephson 1 |
| 80138 | 1010-4-5(10)-J1 | Inquiry Fixed Curve | 10 | 4F | 5(50)5 mm | Josephson 1 |
| 80139 | 1010-6-28-SC(SOFT) | Inquiry Fixed Curve | 10 | 6F | 2-8-2 mm | Special |
| 80405 | 1004-6-25-J | Inquiry Fixed Curve | 4 | 6F | 2-5-2 mm | Josephson |
| 80406 | 1004-6-5-C | Inquiry Fixed Curve | 4 | 6F | 5 mm | Josephson |
| 80407 | 1004-6-2-7 | Inquiry Fixed Curve | 4 | 6F | 2 mm | Cournand |
| 80408 | 1004-6-5-J | Inquiry Fixed Curve | 4 | 6F | 2-5-2 mm | Cournand |
| 80409 | 1004-6-5-C | Inquiry Fixed Curve | 4 | 6F | 5 mm | Cournand |
| 80411 | 1004-6-25-D | Inquiry Fixed Curve | 4 | 6F | 2-5-2 mm | Domato |
| 80412 | 1004-6-5-D | Inquiry Fixed Curve | 4 | 6F | 5 mm | Domato |
| 80413 | 1004-6-10-J | Inquiry Fixed Curve | 4 | 6F | 10 mm | Josephson |
| 80414 | 1004-6-10-C | Inquiry Fixed Curve | 4 | 6F | 10 mm | Cournand |
| 80415 | 1004-6-10-D | Inquiry Fixed Curve | 4 | 6F | 10 mm | Damato |
| 80440 | 1004-5-5-C(HIS)(SOFT) | Inquiry HIS FixedCurve | 4 | 5F | 5 mm | Cournand |
| 80451 | 1004-5-2-J | Inquiry Fixed Curve | 4 | 5F | 2 mm | Josephson |
| 80452 | 1004-5-25-J | Inquiry Fixed Curve | 4 | 5F | 2-5-2 mm | Josephson |
| 80453 | 1004-5-5-J | Inquiry Fixed Curve | 4 | 5F | 5 mm | Josephson |
| 80455 | 1004-5-25-C | Inquiry Fixed Curve | 4 | 5F | 2-5-2 mm | Cournand |
| 80456 | 1004-5-5-C | Inquiry Fixed Curve | 4 | 5F | 5 mm | Cournand |
| 80458 | 1004-5-25-C | Inquiry Fixed Curve | 4 | 5F | 2-5-2 mm | Damato |
| ItemNo. | Description | Trade Name | Electrodes | FrenchSize | ElectrodeSpacing | Curve Type |
| 80459 | 1004-5-5-D | Inquiry Fixed Curve | 4 | 5F | 5 mm | Domato |
| 80463 | 1004-5-10-C | Inquiry Fixed Curve | 4 | 5F | 10 mm | Cournand |
| 80464 | 1004-4-2-C | Inquiry Fixed Curve | 4 | 4F | 2 mm | Cournand |
| 80465 | 1004-4-5-C | Inquiry Fixed Curve | 4 | 4F | 5 mm | Cournand |
| 80466 | 1004-4-25-C | Inquiry Fixed Curve | 4 | 4F | 2-5-2 mm | Cournand |
| 80467 | 1004-4-2-J | Inquiry Fixed Curve | 4 | 4F | 2 mm | Josephson |
| 80468 | 1004-4-5-J | Inquiry Fixed Curve | 4 | 4F | 5 mm | Josephson |
| 80469 | 1004-4-25-J | Inquiry Fixed Curve | 4 | 4F | 2-5-2 mm | Josephson |
| 80473 | 1004-5-5J(SOFT) | Inquiry Fixed Curve | 4 | 5F | 5 mm | Josephson |
| 80476 | 1004-5-5-C (SOFT) | Inquiry Fixed Curve | 4 | 5F | 5 mm | Cournand |
| 80479 | 1004-5-5-D(SOFT) | Inquiry Fixed Curve | 4 | 5F | 5 mm | Domato |
| 80484 | 1004-5-5-J1-TE1BE2 | Inquiry Fixed Curve | 4 | 5F | 5 mm | Josephson 1 |
| 80485 | 1004-4-5-J1 | Inquiry Fixed Curve | 4 | 4F | 5 mm | Josephson 1 |
| 80501 | 1005-6-25-C | Inquiry Fixed Curve | 5 | 6F | 2-5-2 mm | Cournand |
| 80508 | 1005-6-25-J | Inquiry Fixed Curve | 5 | 6F | 2-5-2 mm | Josephson |
| 80516 | 1004-6-2-C(SOFT) | Inquiry Fixed Curve | 4 | 6F | 2 mm | Cournand |
| 80517 | 1004-6-25-J(SOFT) | Inquiry Fixed Curve | 4 | 6F | 2-5-2 mm | Josephson |
| 80518 | 1004-6-5-J1(SOFT) | Inquiry Fixed Curve | 4 | 6F | 10 mm | Josephson 1 |
| 80519 | 1004-6-10-D(SOFT) | Inquiry Fixed Curve | 4 | 6F | 10 mm | Damato |
| 80520 | 1004-6-10-C(SOFT) | Inquiry Fixed Curve | 4 | 6F | 10 mm | Cournand |
| 80533 | 1010-5-28-SC(SOFT) | Inquiry Fixed Curve | 10 | 5F | 2-8-2 mm | Special |
| 80534 | 1010-5-5(30)5-C(SOFT) | Inquiry Fixed Curve | 10 | 5F | 5(30)5 | Cournand |
| 80535 | 1004-4-25-J1 | Inquiry Fixed Curve | 4 | 4F | 2-5-2 mm | Josephson 1 |
| 80536 | 1004-4-5-D | Inquiry Fixed Curve | 4 | 4F | 5 mm | Damato |
| 80537 | 1004-6-10-J(SOFT) | Inquiry Fixed Curve | 4 | 6F | 10 mm | Josephson |
| 80544 | 1010-5-28-C(SOFT) | Inquiry Fixed Curve | 10 | 5F | 2-8-2 mm | Cournand |
| 80567 | 1004-4-5-C(HIS) | Inquiry HIS FixedCurve | 4 | 4F | 5 mm | Cournand |
| 80602 | 1006-6-25-J | Inquiry Fixed Curve | 6 | 6F | 2-5-2 mm | Josephson |
| 80603 | 1006-6-5-J | Inquiry Fixed Curve | 6 | 6F | 5 mm | Josephson |
| 80604 | 1006-6-2-C | Inquiry Fixed Curve | 6 | 6F | 2 mm | Cournand |
| 80606 | 1006-6-5-C | Inquiry Fixed Curve | 6 | 6F | 5 mm | Cournand |
| 80627 | 1006-5-5-J1 (SOFT) | Inquiry Fixed Curve | 6 | 5F | 5 mm | Josephson 1 |
| 80803 | 1008-6-5-J | Inquiry Fixed Curve | 8 | 6F | 5 mm | Josephson |
| 80804 | 1008-6-2-C | Inquiry Fixed Curve | 8 | 6F | 2 mm | Cournand |
| 80806 | 1008-6-5-D | Inquiry Fixed Curve | 8 | 6F | 5 mm | Damato |
| 80810 | 1008-6-5-C | Inquiry Fixed Curve | 8 | 6F | 5 mm | Cournand |
| 80820 | 1008-5-5-(20)5-C(HIS) | Inquiry HIS FixedCurve | 8 | 5F | 20(5) mm | Cournand |
| 80900 | 1014-7-3(70)-TE2BE2 | Inquiry Fixed Curve | 3-3-3-3-3-3-70-3-3-3-3-3-3mm | Cournand | ||
| Item No. | Description | Trade Name | 14Electrodes | 7FFrench Size | Electrode Spacing | Curve Type |
| 81101 | 1110-6-2-F | Inquiry Steerable Curves | 10 | 6F | 2 mm | Med 3.2 cm |
| 81102 | 1110-6-25-M | Inquiry Steerable Curves | 10 | 6F | 2-5-2 mm | Med 3.2 cm |
| 81104 | 1110-6-25-L | Inquiry Steerable Curves | 10 | 6F | 2-5-2 mm | Large 4.1 cm |
| 81105 | 1110-6-25-XL | Inquiry Steerable Curves | 10 | 6F | 2-5-2 mm | X-Large 5.0 cm |
| 81106 | 1110-6-25-F | Inquiry Steerable Curves | 10 | 6F | 2-5-2 mm | Far Reach |
| 81107 | 1110-6-5-L | Inquiry Steerable Curves | 10 | 6F | 5 mm | Large |
| 81108 | 1110-6-2-L | Inquiry Steerable Curves | 10 | 6F | 2 mm | Large |
| 81109 | 1110-6-2-L-TE2BE2 | Inquiry Steerable Curves | 10 | 6F | 2 mm | Large |
| 81110 | 1110-6-17-H | Inquiry H Steerable Atrial | 10 | 6F | 1 & 7 mm | H |
| 81120 | 1120-7-17-H | Inquiry H Steerable Atrial | 20 | 7F | 1 & 7 mm | H |
| 81121 | 1121-7-17-H-SC | Inquiry H Steerable Atrial | 21 | 7F | 1-7-1 mm | H-SC |
| 81124 | 1124-7-271-H | Inquiry H Steerable Atrial | 24 | 7F | 2-7-1 mm | H |
| 81125 | 1106-6-27-H | Inquiry H Steerable Atrial | 6 | 6F | 2-7-2 mm | H |
| 81126 | 1110-6-291-HL | Inquiry H SteerableAtrial | 10 | 6F | 2-9-1 mm | HL |
| 81128 | 1121-7-5-H-UP | Inquiry H Steerable Atrial | 21 | 7F | 5 mm | H |
| 81130 | 1120-7-19-HL | Inquiry H Steerable Atrial | 20 | 7F | 1 & 9 mm | HL |
| 81131 | 1121-7-19-HL-SC | Inquiry H Steerable Atrial | 21 | 7F | 1-9-1 mm | HL-SC |
| 81134 | 1124-7-291-HL | Inquiry H Steerable Atrial | 24 | 7F | 2-9-1 mm | HL |
| 81136 | 1120-7-5-HL-UP | Inquiry H Steerable Atrial | 20 | 7F | 5 mm | HL |
| 81142 | 1106-6-27-HL | Inquiry H Steerable Atrial | 6 | 6F | 2-7-2 mm | HL |
| 81150 | 1124-7-271-H-SCE | Inquiry H Steerable Atrial | 24 | 7F | 2-7-1 mm | H-SCE |
| 81171 | 1110-5-2-M | Inquiry Steerable Curves | 10 | 5F | 2 mm | Med 3.2 cm |
| 81172 | 1110-5-25-M | Inquiry SteerableCurves | 10 | 5F | 2-5-2 mm | Med 3.2 cm |
| 81174 | 1110-5-25-L | Inquiry Steerable Curves | 10 | 5F | 2-5-2 mm | Large 4.1 cm |
| 81176 | 1110-5-25-L | Inquiry Steerable Curves | 10 | 5F | 2-5-2 mm | Far Reach |
| Item No. | Description | Trade Name | Electrodes | French Size | Electrode Spacing | Curve Type |
| 81177 | 1110-5-5-L | Inquiry Steerable Curves | 10 | 5F | 5 mm | Large 4.1 cm |
| 81178 | 1110-5-2-L | Inquiry Steerable Curves | 10 | 5F | 2 mm | Large 4.1 cm |
| 81179 | 1110-5-2-E | Inquiry Steerable Curves | 10 | 5F | 2 mm | Extended Reach |
| 81202 | 1120-7-2-10-XXL | Ten-Ten Duodecapolar | 20 | 7F | 2-10 mm | XX Large 4.8 cm |
| 81203 | 1110-5-25-M(80)SOFT | Inquiry Soft Tip Steerable | 10 | 5F | 2-5-2 mm | Med 3.2 cm |
| 81207 | 1120-7-5-SL | Ten-Ten Duodecapolar | 20 | 7F | 5 mm | Super Lrg 5.1 cm |
| 81209 | 1120-7-25-SL | Ten-Ten Duodecapolar | 20 | 7F | 2-5 mm | Super Lrg 5.1 cm |
| 81211 | 1120-7-2(20)2(25)2-SL | Ten-Ten Duodecapolar | 20 | 7F | 2(20)2(25)2 mm | Super Lrg 5.1 cm |
| 81223 | 1110-5-2(20)3-XL | Inquiry Steerable Curves | 10 | 5F | 2(50)3 | x-large |
| 81401 | 1104-6-5-S | Inquiry Steerable Curves | 4 | 6F | 5 mm | Small 2.7 cm |
| 81402 | 1104-6-25-M | Inquiry Steerable Curves | 4 | 6F | 2-5-2 mm | Medium 3.2 cm |
| 81403 | 1104-6-5-M | Inquiry Steerable Curves | 4 | 6F | 5 mm | Medium 3.2 cm |
| 81404 | 1104-6-25-L | Inquiry Steerable Curves | 4 | 6F | 2-5-2 mm | Large 4.2 cm |
| 81405 | 1104-6-5-L | Inquiry Steerable Curves | 4 | 6F | 5 mm | Large 4.2 cm |
| 81406 | 1104-6-25-E | Inquiry Steerable Curves | 4 | 6F | 2-5-2 mm | Extended Reach |
| 81407 | 1104-6-25-F | Inquiry Steerable Curves | 4 | 6F | 2-5-2 mm | Far Reach |
| 81412 | 1104-6-2-E | Inquiry Steerable Curves | 4 | 6F | 2 mm | Extended Reach |
| 81417 | 1104-6-5-XL | Inquiry Steerable Curves | 4 | 6F | 5 mm | X-Large 5.0 cm |
| 81418 | 1104-6-25-XL | Inquiry Steerable Curves | 4 | 6F | 2-5-2 mm | X-Large 5.0 cm |
| 81471 | 1104-5-5-S | Inquiry Steerable Curves | 4 | 5F | 5 mm | Small 2.7 cm |
| 81472 | 1104-5-25-M | Inquiry Steerable Curves | 4 | 5F | 2-5-2 mm | Med 3.2 cm |
| 81473 | 1104-5-5-M | Inquiry Steerable Curves | 4 | 5F | 5 mm | Med 3.2 cm |
| 81474 | 1104-5-25-L | Inquiry SteerableCurves | 4 | 5F | 2-5-2 mm | Large 4.1 cm |
| 81475 | 1104-5-5-L | Inquiry SteerableCurves | 4 | 5F | 5 mm | Large 4.1 cm |
| 81478 | 1104-5-2-S | Inquiry SteerableCurves | 4 | 5F | 2 mm | Small 2.7 cm |
| ItemNo. | Description | Trade Name | Electrodes | FrenchSize | ElectrodeSpacing | Curve Type |
| 81479 | 1104-5-25-S | Inquiry SteerableCurves | 4 | 5F | 2-5-2 mm | Small 2.7 cm |
| 81483 | 1104-5-5-E(HIS)(SOFT) | Inquiry IIIS SteerableCurve | 4 | 5F | 5 mm | Extended |
| 81484 | 1104-5-5-L(EX-HIS)(SOFT) | Inquiry HIS SteerableCurve | 4 | 5F | 5 mm | Large |
| 81485 | 1104-5-5-E(EX-HIS)(SOFT) | Inquiry HIS SteerableCurve | 4 | 5F | 5 mm | Extended |
| 81503 | 1104-6-5-L-TE2BE2-BD | Inquiry Bi-Directional | 4 | 6F | 5 mm | Large 4.1 cm |
| 81504 | 1110-6-5-M-TE2BE2-BD | Inquiry Bi-Directional | 10 | 6F | 5 mm | Med 3.2 cm |
| 81511 | 1105-6-25-M | Inquiry SteerableCurves | 6 | 6F | 2-5-2 mm | Med 3.2 cm |
| 81520 | 1110-6-2-XL-TE4BE4 | Inquiry SteerableCurves | 10 | 6F | 2 mm | X-Large 5.0cm |
| 81521 | 1105-5-5(170)-M/L-TE1BE2(SOFT) | Inquiry Soft TipSteerable | 5 | 5F | 5(170) | Med/Large |
| 81524 | 1110-6-2-L-TE4BE4 | Inquiry SteerableCurves | 10 | 6F | 2 mm | Large 4.1 cm |
| 81530 | 1110-4-2-M | Inquiry SteerableCurves | 10 | 4F | 2 mm | Med 3.2 cm |
| 81531 | 1110-4-25-M | Inquiry SteerableCurves | 10 | 4F | 2-5-2 mm | Med 3.2 cm |
| 81532 | 1110-4-025-L | Inquiry SteerableCurves | 10 | 4F | 2-5-2 mm | Large 4.1 cm |
| 81534 | 1110-4-5-L | Inquiry SteerableCurves | 10 | 4F | 5 mm | Large 4.1 cm |
| 81536 | 1110-4-25-M(80) | Inquiry SteerableCurves | 10 | 4F | 2-5-2 mm | Med 3.2 cm |
| 81537 | 1110-4-25-M(SC)(60) | Inquiry SteerableCurves | 10 | 4F | 2-5-2 mm | Med 3.2 cm |
| 81540 | 1104-4-25-M | Inquiry SteerableCurves | 4 | 4F | 2-5-2 mm | Med 3.2 cm |
| 81541 | 1104-4-2-M | Inquiry SteerableCurves | 4 | 4F | 2 mm | Med 3.2 cm |
| 81542 | 1104-4-5-M | Inquiry SteerableCurves | 4 | 4F | 5 mm | Med 3.2 cm |
| 81543 | 1104-4-25-L | Inquiry SteerableCurves | 4 | 4F | 2-5-2 mm | Large 4.1 cm |
| 81545 | 1104-4-5-L | Inquiry SteerableCurves | 4 | 4F | 5 mm | Large 4.1 cm |
| 81587 | 1120-7-1(2.5)-SM-AF20 | A-Focus II Steerable Atrial | 20 | 7F | 1(2.5) mm | 20 mm diameter |
| 81589 | 1114-7-15-SM-AF20 | A-Focus II Steerable Atrial | 14 | 7F | 1-5-1 mm | 20 mm diameter |
| 81591 | 1114-7-13-SM-AF15 | A-Focus II Steerable Atrial | 14 | 7F | 1-3 mm | 15 mm diameter |
| Item No. | Description | Trade Name | Electrodes | French Size | Electrode Spacing | Curve Type |
| 81594 | 1110-7-3.5-SM-AF15 | A-Focus II Steerable Atrial | 10 | 7F | 3.5 mm | 15 mm diameter |
| 81595 | 1110-7-5-SM-AF | A-Focus II Steerable Atrial | 10 | 7F | 5 mm | 20 mm diameter |
| 81596 | 1110-7-27-SM-AF20 | A-Focus II Steerable Atrial | 10 | 7F | 2-7 mm | 20 mm diameter |
| 81597 | 1110-7-2(11)-SM-AF25 | A-Focus II Steerable Atrial | 10 | 7F | 2(11) mm | 25 mm diameter |
| 81598 | 1120-7-1(4.5)-SM-AF25 | A-Focus II Steerable Atrial | 20 | 7F | 1(4.5) mm | 25 mm diameter |
| 81599 | 1110-7-8-SM-AF25 | A-Focus II Steerable Atrial | 10 | 7F | 8 mm | 25 mm diameter |
| 81601 | 1106-6-5-M | Inquiry Steerable Curves | 6 | 6F | 5 mm | Med 3.2 cm |
| 81602 | 1106-6-5-L | Inquiry Steerable Curves | 6 | 6F | 5 mm | Large 4.1 cm |
| 81603 | 1106-6-5-E | Inquiry Steerable Curves | 6 | 6F | 5 mm | Extended Reach |
| 81604 | 1106-6-5-F | Inquiry Steerable Curves | 6 | 6F | 5 mm | Far Reach |
| 81605 | 1106-6-5-XL | Inquiry Steerable Curves | 6 | 6F | 5 mm | X-Large 5.0 cm |
| 81670 | 1112-5-2-(10)-AF20 | A-Focus I Steerable Atrial | 12 | 5F | 2(10) mm | 25 mm diameter |
| 81671 | 1114-5-2(9)-AF30 | A-Focus I Steerable Atrial | 14 | 5F | 2(9) mm | 30 mm diameter |
| 81672 | 1110-5-27-SM-AF20 | A-Focus II Steerable Atrial | 10 | 5F | 2-7 mm | 20 mm diameter |
| 81673 | 1110-5-5-SM-AF20 | A-Focus II Steerable Atrial | 10 | 5F | 5 mm | 20 mm diameter |
| 81674 | 1110-5-5-AF20 | A-Focus I Steerable Atrial | 10 | 5F | 5 mm | 20 mm diameter |
| 81676 | 1110-5-3.5-AF15 | A-Focus I Steerable Atrial | 10 | 5F | 3.5 mm | 15 mm diameter |
| 81680 | 1110-5-3.5-SM-AF15 | A-Focus II Steerable Atrial | 10 | 5F | 3.5 mm | 15 mm diameter |
| 81683 | 1120-7-1(4.5)-SM-OPT25 | Optima Steerable Lasso | 20 | 5F | 1(4.5) mm | 25 mm - 15 mm |
| Item No. | Description | Trade Name | Electrodes | French Size | Electrode Spacing | Curve Type |
| 81687 | 1110-7-10-SM-OPT25 | Optima Steerable Lasso | 10 | 7F | 10mm | 25 mm - 15 mm |
| 81703 | 1110-5-28-M/L(SC)(XS)(SOFT) | Inquiry Soft Tip Steerable | 10 | 5F | 2-8-2 mm | Med 3.2 cm/Large 4.1 cm |
| 81704 | 1110-5-5-M/L(SC)(XS)(SOFT) | Inquiry Soft Tip Steerable | 10 | 5F | 5 mm | Med 3.2 cm/Large SC 4.1 cm |
| 81705 | 1110-5-28-M(SC)(XS)(SOFT) | Inquiry Soft Tip Steerable | 10 | 5F | 2-8-2 mm | M(SC)(XS) |
| 81706 | 1110-5-5-M/L(XS)(SOFT) | Inquiry Soft Tip Steerable | 10 | 5F | 5 mm | M/L(XS) |
| 81711 | 1108-5-5(20)5-E(HIS)(SOFT) | Inquiry HIS Steerable Curve | 8 | 5F | 5(20)5 mm | Extended |
| 81717 | 1124-7-1(4.5)(20)(3)-SM-OPT25 | Optima Steerable Lasso | 24 | 7F | 4(4.5) | 25 mm - 15 mm |
| 81721 | 1110-5-25-M(SC)(SOFT) | Inquiry Soft Tip Steerable | 10 | 5F | 2-5-2 mm | Med (SC) 3.2 cm |
| 81730 | 1110-5-28-M/L(SOFT) | Inquiry Soft Tip Steerable | 10 | 5F | 2-8-2 mm | M/L |
| 81732 | 1110-5-25-M(SOFT) | Inquiry Soft Tip Steerable | 10 | 5F | 2-5-2 mm | Med 3.2 cm |
| 81733 | 1110-5-2-L(SOFT) | Inquiry Soft Tip Steerable | 10 | 5F | 2 mm | Large 4.1 cm |
| 81734 | 1110-5-25-L(SOFT) | Inquiry Soft Tip Steerable | 10 | 5F | 2-5-2 mm | Large 4.1 cm |
| 81735 | 1110-5-5-L(SOFT) | Inquiry Soft Tip Steerable | 10 | 5F | 5 mm | Large 4.1 cm |
| 81738 | 1110-5-1-L(SOFT) | Inquiry Soft Tip Steerable | 10 | 5F | 1 mm | Large 4.1 cm |
| 81742 | 1104-5-25-M(SOFT) | Inquiry Soft Tip Steerable | 4 | 5F | 2-5-2 mm | Med 3.2 cm |
| 81743 | 1104-5-5-M(SOFT) | Inquiry Soft Tip Steerable | 4 | 5F | 5 mm | Med 3.2 cm |
| 81745 | 1104-5-25-L(SOFT) | Inquiry Soft Tip Steerable | 4 | 5F | 2-5-2 mm | Large 4.1 cm |
| 81750 | 1110-5-25-L(60cm)(SOFT) | Inquiry Soft Tip Steerable | 10 | 5F | 2-5-2 mm | Large 4.1 cm |
| 81751 | 1108-5-2-M(SOFT) | Inquiry Soft Tip Steerable | 8 | 5F | 2 mm | Med 3.2 cm |
| 81801 | 1108-6-2-M | Inquiry Steerable Curves | 8 | 6F | 2 mm | Med 3.2 cm |
| 81802 | 1108-6-25-M | Inquiry Steerable Curves | 8 | 6F | 2-5-2 mm | Med 3.2 cm |
| 81807 | 1108-6-2-L | Inquiry Steerable Curves | 8 | 6F | 2 mm | Large 4.1 cm |
| 81809 | 1108-6-25-L | Inquiry Steerable Curves | 8 | 6F | 2-5-2 mm | Large 4.1 cm |
| 81823 | 1108-6-2-M(SOFT) | Inquiry Soft TipSteerable | 8 | 6F | 2 mm | Med 3.2 cm |
| 81871 | 1108-5-2-M | Inquiry SteerableCurves | 8 | 5F | 2 mm | Med 3.2 cm |
| 81872 | 1108-5-25-M | Inquiry SteerableCurves | 8 | 5F | 2-5-2 mm | Med 3.2 cm |
| 81873 | 1108-5-5-M | Inquiry SteerableCurves | 8 | 5F | 5 mm | Med 3.2 cm |
| 81877 | 1108-5-2-L | Inquiry SteerableCurves | 8 | 5F | 2 mm | Large 4.1 cm |
| 81879 | 1108-5-25-L | Inquiry SteerableCurves | 8 | 5F | 2-5-2 mm | Large 4.1 cm |
| ItemNo. | Description | Trade Name | Electrodes | FrenchSize | ElectrodeSpacing | Curve Type |
| 81901 | 1120-7-13-M | Ten-Ten Duodecapolar | 20 | 7F | 1-3 mm | Medium 3.2cm |
| 81902 | 1120-7-13-L | Ten-Ten Duodecapolar | 20 | 7F | 1-3 mm | Large 4.2 cm |
| 81940 | 1110-6-2-M(SOFT) | Inquiry Soft TipSteerable | 10 | 6F | 2 mm | Med 3.2 cm |
| 81943 | 1110-6-2-L(SOFT) | Inquiry Soft TipSteerable | 10 | 6F | 2 mm | Large 4.1 cm |
| 81945 | 1110-6-25-L(SOFT) | Inquiry Soft TipSteerable | 10 | 6F | 2-5-2 mm | Large 4.1 cm |
| 81947 | 1110-6-5-M/L(SOFT) | Inquiry Soft TipSteerable | 10 | 6F | 5 mm | M/L |
| 81951 | 1104-6-5-M(SOFT) | Inquiry Soft TipSteerable | 4 | 6F | 5 mm | Med 3.2 cm |
| 81954 | 1104-6-5-L(SOFT) | Inquiry Soft TipSteerable | 4 | 6F | 5 mm | Large 4.1 cm |
| 85931 | 1924-S | Inquiry DiagnosticConnecting Cables | n/a | n/a | 1.5m length | n/a |
| 85953 | 1910-SA | Inquiry DiagnosticConnecting Cables | n/a | n/a | 1.5m length | n/a |
| 85954 | 1904-SA | Inquiry DiagnosticConnecting Cables | n/a | n/a | 1.5m length | n/a |
| 85955 | 1914-SA | Inquiry DiagnosticConnecting Cables | n/a | n/a | 1.5m length | n/a |
| 85930 | 1910-S | Inquiry DiagnosticConnecting Cables | n/a | n/a | 2.5m length | n/a |
.
{5}------------------------------------------------
:
{6}------------------------------------------------
·
{7}------------------------------------------------
Page 6 -- Ms. Katie Bray
{8}------------------------------------------------
Page 7 – Ms. Katie Bray
{9}------------------------------------------------
Page 8 – Ms. Katic Bray
:
:
{10}------------------------------------------------
Page 9 -- Ms. Katie Bray
{11}------------------------------------------------
:
:
Page 10 ·· Ms. Katie Bray
.
{12}------------------------------------------------
SECTION 4: INDICATIONS FOR USE STATEMENT
510(k) Number (if known):
Device Name: Reprocessed Electrophysiology Catheters
Indications For Use: Reprocessed Electrophysiology (EP) Catheters are indicated for temporary intracardiac sensing, recording, stimulation and electrophysiological mapping of cardiac structures.
Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
eamall
Division of Cardiovascular Devices
Kos1329 510(k) Number
Ascent Healthcare Solutions Reprocessed Electrophysiology Catheters Traditional 510(k)
§ 870.1220 Electrode recording catheter or electrode recording probe.
(a)
Identification. An electrode recording catheter or an electrode recording probe is a device used to detect an intracardiac electrocardiogram, or to detect cardiac output or left-to-right heart shunts. The device may be unipolar or multipolar for electrocardiogram detection, or may be a platinum-tipped catheter which senses the presence of a special indicator for cardiac output or left-to-right heart shunt determinations.(b)
Classification. Class II (performance standards).