K Number
K080003
Manufacturer
Date Cleared
2008-10-23

(295 days)

Product Code
Regulation Number
864.7360
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The BinaxNOW® G6PD (Glucose-6-Phosphate Dehydrogenase) Test is an in vitro enzyme chromatographic test for the qualitative detection of G6PD enzyme activity in human venous whole blood, collected in heparin or ethylenediaminetetraacetic acid (EDTA). The BinaxNOW® G6PD Test is a visual screening test used for differentiating normal from deficient G6PD activity levels in whole blood and is intended to aid in the identification of people with G6PD deficiency. Samples which generate deficient results should be assayed using a quantitative G6PD test method to verify the deficiency.

Device Description

The BinaxNOW® G6PD test device consists of a lateral flow test strip comprised of a white sample pad and a reaction pad, which is located at the top of the strip (2 U.S. patents pending). The reaction pad contains the reagents necessary for the G6PD enzymatic reaction and the subsequent reduction of a nitro blue tetrazolium dye into its concomitant blue formazan product. The resulting color change on the strip indicates enough G6PD activity is present to presume the sample is not deficient.

To perform the test, a whole blood sample is mixed with red blood cell (RBC) lysing reagent in a sample preparation vial and then transferred to the test device sample pad. The lysed blood sample migrates up the test strip, reconstituting reagents in the reaction pad. When the sample front (or liquid migration) covers the entire reaction pad, the device is closed.

Test results are read visually. If no change in the red color of the sample front is observed at the test read time, the sample is presumed to be deficient in G6PD enzyme activity. Samples normal in G6PD activity produce a distinct color change - the red sample color changes to a brown / black color on the upper half of the reaction pad.

AI/ML Overview

Here's a breakdown of the acceptance criteria and the study details for the BinaxNOW® G6PD Test, based on the provided text:

Acceptance Criteria and Device Performance

The core acceptance criteria revolve around the agreement of the BinaxNOW® G6PD Test results with a commercially available quantitative G6PD test, particularly for identifying G6PD deficiency and normal activity.

Acceptance Criteria (Implicit)Reported Device Performance (BinaxNOW® G6PD Test)
Agreement with Quantitative G6PD Test for Deficient Results (Heparin Samples): High percentage of agreement for deficient results compared to the quantitative method.Deficient result percent agreement (Heparin Samples): 98.0% (CI = 89.3 - 99.6%)
Agreement with Quantitative G6PD Test for Normal Results (Heparin Samples): High percentage of agreement for normal results compared to the quantitative method.Normal result percent agreement (Heparin Samples): 97.9% (CI = 94.8 - 99.2%)
Overall Agreement with Quantitative G6PD Test (Heparin Samples): High overall percentage of agreement.Overall percent agreement (Heparin Samples): 97.9% (CI = 95.3 - 99.1%)
Agreement with Quantitative G6PD Test for Deficient Results (EDTA Samples): High percentage of agreement for deficient results compared to the quantitative method.Deficient result percent agreement (EDTA Samples): 98.0% (CI = 89.5 - 99.6%)
Agreement with Quantitative G6PD Test for Normal Results (EDTA Samples): High percentage of agreement for normal results compared to the quantitative method.Normal result percent agreement (EDTA Samples): 97.4% (CI = 94.2 - 98.9%)
Overall Agreement with Quantitative G6PD Test (EDTA Samples): High overall percentage of agreement.Overall percent agreement (EDTA Samples): 97.6% (CI = 94.8 - 98.9%)
Consistency between Heparin and EDTA Samples: High agreement between the two sample types.Percent agreement between heparin and EDTA samples: 99% (for 240/243 subjects)
Reproducibility Across Operators and Sites: High agreement with expected test results across multiple operators, days, and sites.Reproducibility Agreement: 98.4% (123/125) with no significant differences within run, between run, between sites, or between operators.
Precision (Single Operator): Consistent results by a single operator over multiple days.Demonstrated 100% consistent results for both normal and deficient samples over ten successive days by a single operator.
Absence of Interference from Endogenous Substances: Test performance not affected by high levels of various endogenous blood components.Interference Testing: None of the tested endogenous blood components (bilirubin, triglycerides, total cholesterol, lactic acid, lactate dehydrogenase, glucose, copper sulfate) affected test performance.
Performance with Abnormally Low/High Hematocrit Levels: Performance maintained within defined limits (though limitations existed as per package insert).Test performance was affected by abnormally low and high hematocrit levels (17-18% and 54-65%), as described in the Limitations section of the package insert (details not provided in this summary).

Study Details:

  1. Sample Size used for the test set and the data provenance:

    • Sample Size: 246 subjects (for the initial clinical sample performance study).
    • Data Provenance: Prospective study conducted in 2007-2008 in the U.S.
  2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    • The document does not specify the number or qualifications of experts used to establish the ground truth. The "ground truth" was established by a "commercially available quantitative G6PD test," implying a quantitative laboratory assay rather than human expert interpretation of the BinaxNOW® test results.
  3. Adjudication method for the test set:

    • Adjudication method for the test set results is not explicitly mentioned. The BinaxNOW® G6PD Test is a visual screening test. The comparison was made against a "commercially available quantitative G6PD test," implying a direct numerical comparison against the quantitative assay's results.
  4. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • No, an MRMC comparative effectiveness study was not done. This device is a rapid diagnostic test (lateral flow, enzyme chromatographic) for G6PD deficiency, read visually, not an AI-powered diagnostic imaging tool. It involves a single visual reading of the color change on the strip.
    • The closest "multi-reader" equivalent mentioned is the Reproducibility Study, which involved "multiple operators" (6 operators) across "3 separate sites." However, this was to assess the consistency of the device's output and interpretation, not to compare human performance with/without AI assistance.
  5. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    • Yes, in effect, a standalone performance was done for comparing the device's output against the quantitative method. The BinaxNOW® G6PD Test is designed to be a standalone visual screening test. The "performance summary" directly compares its qualitative output (deficient/normal) to the quantitative method's results, indicating its performance as a standalone diagnostic tool. Human readers are involved in the interpretation of the visual result of the BinaxNOW® test, but the device itself is a qualitative biochemical assay, not an AI algorithm.
  6. The type of ground truth used:

    • Quantitative Assay Results: The ground truth was established using results from a "commercially available quantitative G6PD test." This serves as the reference standard. A specific cut-off value of 4.2 U/gHb from the comparative method was used to define "deficient" and "normal" for comparison purposes.
  7. The sample size for the training set:

    • The document does not explicitly mention a separate "training set" as would be typical for machine learning models. For this type of in-vitro diagnostic device, development and optimization often involve internal studies, but specific details on a "training set" size for a machine learning context are not provided because it's not an AI device. The clinical performance study used 246 subjects.
  8. How the ground truth for the training set was established:

    • As noted above, a distinct "training set" with established ground truth in the context of an AI algorithm is not applicable here. The development and validation of the BinaxNOW® G6PD Test would have involved establishing performance characteristics against known G6PD activity levels, likely using quantitative methods, but this is a product development process rather than an AI model training process.

{0}------------------------------------------------

ATTACHMENT 3

510(k) SUMMARY (Updated October 21, 2008)

This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92.

The assigned 510(k) number is: K080003

SUBMITTER

Binax, Inc., d/b/a Inverness Medical 10 Southgate Road Scarborough, Maine 04074 (207) 730-5739 (Office) (207) 730-5717 (FAX) Establishment Registration Number: 1221359

CONTACT PERSON

Anne Jepson anne.jepson@invmed.com (email)

DATE PREPARED

December 31, 2007 (revised October 21, 2008)

TRADE NAME BinaxNOW® G6PD Test

COMMON NAME

BinaxNOW® G6PD Test, BinaxNOW® G6PD, Binax NOW® G6PD Test, Binax NOW® G6PD, NOW® G6PD Test, NOW® G6PD

CLASSIFICATION NAME

Glucose-6-Phosphate Dehydrogenase (Erythrocytic), Screening (per 21 CFR 864.7360)

PREDICATE DEVICE

Trinity Biotech Glucose-6-Phosphate Dehydrogenase (G-6-PDH) Deficiency reagent set. #K933934.

DEVICE DESCRIPTION

The BinaxNOW® G6PD test device consists of a lateral flow test strip comprised of a white sample pad and a reaction pad, which is located at the top of the strip (2 U.S. patents pending). The reaction pad contains the reagents necessary for the G6PD enzymatic reaction and the subsequent reduction of a nitro blue tetrazolium dye into its concomitant blue formazan product. The resulting color change on the strip indicates enough G6PD activity is present to presume the sample is not deficient.

To perform the test, a whole blood sample is mixed with red blood cell (RBC) lysing reagent in a sample preparation vial and then transferred to the test device sample pad. The lysed blood sample migrates up the test strip, reconstituting reagents in the reaction pad. When the sample front (or liquid migration) covers the entire reaction pad, the device is closed.

{1}------------------------------------------------

Test results are read visually. If no change in the red color of the sample front is observed at the test read time, the sample is presumed to be deficient in G6PD enzyme activity. Samples normal in G6PD activity produce a distinct color change - the red sample color changes to a brown / black color on the upper half of the reaction pad.

INTENDED USE

The BinaxNOW® G6PD (Glucose-6-Phosphate Dehydrogenase) Test is an in vitro enzyme chromatographic test for the qualitative detection of G6PD enzyme activity in human venous whole blood, collected in heparin or ethylenediaminetetraacetic acid (EDTA). The BinaxNOW® G6PD Test is a visual screening test used for differentiating normal from deficient G6PD activity levels in whole blood and is intended to aid in the identification of people with G6PD deficiency. Samples which generate deficient results should be assayed using a quantitative G6PD test method to verify the deficiency.

TECHNOLOGICAL CHARACTERISTICS

The BinaxNOW® G6PD Test and the predicate Trinity Biotech Glucose-6-Phosphate Dehydrogenase (G-6-PDH) Deficiency reagent set utilize different technologies. The BinaxNOW® test uses lateral flow, enzyme chromatographic technology, while the Trinity Biotech G-6-PDH Deficiency reagent set is a liquid system fluorescing "spot" assay. Both tests are qualitative tests for the visual determination of G6PD deficiency in whole blood.

PERFORMANCE SUMMARY

Clinical Sample Performance – BinaxNOW® G6PD Test Method Comparison

The performance of the BinaxNOW® test was compared to a commercially available quantitative G6PD test in a prospective study conducted in 2007-2008 in the U.S. Both heparinized and EDTA whole blood specimens were collected from 246 subjects and were evaluated on the BinaxNOW® test.

All of the samples that generated a value less than or equal to 2.0 U/gHb on the comparative method generated deficient results on the BinaxNOW G6PD test.

The percent agreement analyses and 95% confidence intervals for the BinaxNOW® G6PD test results for detection of G6PD enzyme activity deficiency, as compared to the comparative method, on both heparinized and EDTA blood samples, is provided below. For both sample types, a comparative method cut-off value of 4.2 U/gHb was used in the analysis. For all results less than or equal to 4.2 U/qHb on the comparative method, a deficient BinaxNOW result was considered a correct result. Likewise, for all comparative method results greater than 4.2 U/gHb, a normal BinaxNOW result was considered a correct result.

{2}------------------------------------------------

% AGREEMENT WITH HEPARIN SAMPLES:

ComparativeMethod →
DeficientNormal
BinaxNOWDeficient484
Test →Normal1190

"Deficient result" percent agreement = 48 / 49 = 98.0% (CI = 89.3 - 99.6%) "Normal result" percent agreement = 190 / 194 = 97.9% (Cl = 94.8 - 99.2%) Overall percent agreement = 238 / 243* = 97.9% (CI = 95.3 - 99.1%) (* 3 invalid tests)

% AGREEMENT WITH EDTA SAMPLES:

ComparativeMethod →
DeficientNormal
BinaxNOWDeficient495
Test →Normal1191

"Deficient result" percent agreement = 49 / 50 = 98.0% (Cl = 89.5 - 99.6%) "Normal result" percent agreement = 191 / 196 = 97.4% (Cl = 94.2 - 98.9%) Overall percent agreement = 240 / 246 = 97.6% (CI = 94.8 - 98.9%)

Additionally, BinaxNOW® test results on the heparin samples were the same as the results on the EDTA samples for 240 of the 243 subjects, whose samples generated valid results on both sample types, yielding a percent agreement of 99%.

Interfering Substances

The BinaxNOW® G6PD test was evaluated for possible interference from high levels of endogenous blood components. Whole blood samples were tested that contained bilirubin (conjugated and unconjugated), triglycerides, total cholesterol, lactic acid, lactate dehydrogenase, or glucose at concentrations above physiological levels. None of the endogenous blood components affected test performance. The presence of an elevated level of copper sulfate, which is known to inhibit G6PD enzyme activity, was also evaluated and did not affect test performance.

Blood samples with abnormally low and high hematocrit levels (17-18% and 54-65% respectively) were evaluated, and test performance was affected as described in the Limitations section of the package insert.

Reproducibility Study - Multiple Operators

A blind study of the BinaxNOW® G6PD Test was conducted at 3 separate sites using panels of blind coded specimens, which included G6PD normal and deficient samples. Participants tested each sample multible times on 3 different days. There was 98.4% (123/125) agreement with expected test results, with no significant differences within run (replicates tested by one operator), between run (3 different days), between sites (3 sites), or between operators (6 operators).

Precision Study - Single Operator

{3}------------------------------------------------

Blood samples from two individuals were drawn into both EDTA and heparin collection tubes, and all 4 samples were tested in duplicate on the BinaxNOW test on ten successive days by a single operator. The samples collected from one individual were interpreted as normal 100% of the time. The samples collected from the other individual were interpreted as deficient 100% of the time.

{4}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Public Health Service

Food and Drug Administration 2098 Gaither Road Rockville MD 20850

Binax, Inc. Ms. Anne Jepson 10 Southgate Road Scarborough, Maine 04074

OCT 23 2008

Re: K080003

Trade/Device Name: BinaxNOW® G6PD Test Regulation Number: 21 CFR 864.7360 Regulation Name: Glucose-6-Phosphate Dehydrogenase (Erythrocytic), Screening Regulatory Class: Class II Product Code: JBF Dated: September 19, 2008 Received: Oct 15, 2008

Dear Ms. Jepson:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820). This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed

{5}------------------------------------------------

Page 2 - Ms. Jepson

predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at 240-276-0450. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at 240-276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at 240-276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely yours.

Maria mchen

Maria Chan, Ph.D Acting Director Division of Immunology and Hematology Devices Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health

Enclosure

{6}------------------------------------------------

INDICATIONS FOR USE STATEMENT

510(k) Number (if known)

510(k) Number (if known): K080003

Device Name: BinaxNOW® G6PD Test

Indications for Use:

The BinaxNOW® G6PD (Glucose-6-Phosphate Dehydrogenase) Test is an in vitro enzyme chromatographic test for the qualitative detection of G6PD enzyme activity in human venous whole blood, collected in heparin or ethylenediaminetetraacetic acid (EDTA). The BinaxNOW® G6PD Test is a visual screening test used for differentiating normal from deficient G6PD activity levels in whole blood and is intended to aid in the identification of people with G6PD deficiency. Samples which generate deficient results should be assayed using a quantitative G6PD test method to verify the deficiency.

Prescription Use(Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use(21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)

Concurrence of CDRH, Office of Division Evaluation (ODE)

Office of In Vitro Diagnostic Device
Evaluation and Safety

510(k) K080003

Revision 12/28/07

§ 864.7360 Erythrocytic glucose-6-phosphate dehydrogenase assay.

(a)
Identification. An erythrocytic glucose-6-phosphate dehydrogenase assay is a device used to measure the activity of the enzyme glucose-6-phosphate dehydrogenase or of glucose-6-phosphate dehydrogenase isoenzymes. The results of this assay are used in the diagnosis and treatment of nonspherocytic congenital hemolytic anemia or drug-induced hemolytic anemia associated with a glucose-6-phosphate dehydrogenase deficiency. This generic device includes assays based on fluorescence, electrophoresis, methemoglobin reduction, catalase inhibition, and ultraviolet kinetics.(b)
Classification. Class II (performance standards).