K Number
K073693
Device Name
COSTORSOL
Date Cleared
2008-07-03

(185 days)

Product Code
Regulation Number
876.5880
Reference & Predicate Devices
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

CoStorSol® is intended for the flushing and cold storage of kidney, liver and pancreas organs at the time of organ removal from the donor in preparation for storage, transportation and eventual transplantation into a recipient.

Device Description

Preservation Solutions, Inc. manufacturers CoStorSol® according to a "recipe" pioneered at the University of Wisconsin by Dr. Folkert O. Belzer. The cold storage solution is still referred to as "Belzer UW" solution, even though it has been marketed under the trade name ViaSpan®. The formulation includes soluble colloids, buffers, sodium and postassium salts, redox stabilizers, and phosphoric compounds to aid tissue viability by enabling regeneration of adenosine triphosphate (ATP).

CoStorSol® is a clear to light yellow, sterile, non-pyrogenic solution for hypothermic flushing and storage of organs. The solution is packaged in 1-liter bags.

AI/ML Overview

Here's an analysis of the provided 510(k) summary regarding the CoStorSol® device, focusing on acceptance criteria and supporting studies:

Summary of Acceptance Criteria and Device Performance for CoStorSol®

The provided 510(k) summary for CoStorSol® establishes substantial equivalence by demonstrating that the device is identical in composition, intended use, and performance to a legally marketed predicate device, ViaSpan® (Belzer - UW) Organ Solution (K944866). Therefore, the "acceptance criteria" are implicitly met by demonstrating this identity and confirming essential safety and functional characteristics.

Acceptance Criteria CategorySpecific Criteria/Tests ConductedReported Device Performance (CoStorSol®)
Substantial Equivalence to PredicateIdentical chemical composition"identical solution with exactly the same intended use and chemical composition" as ViaSpan® (Belzer - UW). Fourier transformed infrared scans confirmed stability and chemical identity.
Identical intended use"exactly the same intended use" as ViaSpan® (Belzer - UW), i.e., flushing and cold storage of kidney, liver, and pancreas.
Material/Solution BiocompatibilityISO 10993 series of standardsCoStorSol® itself shown to be a biocompatible solution. Flexible solution administration pouches also tested and shown to be biocompatible.
Sterility & Non-PyrogenicitySterilization validation (ISO 17665 or USP Section <1116>)Supplied sterile and non-pyrogenic. Sterilization processes were validated.
Chemical Stability & IdentityChemical analyses (fresh and aged samples)Chemical analyses on fresh and aged samples show identity to the predicate solution.
Fourier transformed infrared scansConfirmed stability and chemical identity.
PackagingFlexible 1-liter bagsSupplied in 1-liter flexible bags, designed to ease connection to standard administration sets.
Storage RequirementsRefrigerated storage without freezingDesigned for refrigerated storage without freezing.
Use ConditionsUse cold as vasculature flush and cold storage mediumUsed cold as a vasculature flush medium for organs and then as a cold storage medium.

Study Details:

The 510(k) submission for CoStorSol® primarily relies on demonstrating substantial equivalence to a predicate device rather than conducting a de novo clinical study with specific performance endpoints. The "study" here refers to the testing and analysis performed to prove this equivalence.

  1. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective):

    • No specific "test set" in the sense of a patient cohort or a dataset with ground truth labels is described for this type of device. The samples used were samples of the CoStorSol® product itself for various chemical, biological, and physical tests.
    • The provenance of these test samples (e.g., manufacturing location, specific batch numbers) is not detailed. The tests performed are laboratory-based.
    • This is not a retrospective or prospective human clinical study. The data is generated from laboratory testing of the solution and its packaging materials.
  2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience):

    • This is not applicable. The "ground truth" for this type of device is established through validated laboratory methods for chemical composition, sterility, biocompatibility, and physical characteristics, rather than expert interpretation of medical images or clinical outcomes. The experts involved would be laboratory scientists, chemists, and microbiologists adhering to established standards (e.g., ISO, USP).
  3. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

    • Not applicable. Adjudication methods are typically employed in studies where human interpretation or consensus is required (e.g., pathology review, clinical endpoints). Here, objective laboratory measurements are the basis of evaluation.
  4. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • Not applicable. This device is an organ preservation solution, not an AI-powered diagnostic or assistive tool. MRMC studies are irrelevant in this context.
  5. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    • Not applicable. This device is a medical solution, not a software algorithm.
  6. The type of ground truth used (expert concensus, pathology, outcomes data, etc):

    • The "ground truth" here is the established scientific and regulatory standards for chemical identity, sterility, biocompatibility, and stability, as demonstrated through objective laboratory testing. The predicate device's established efficacy and safety, based on its identical formulation, serves as the ultimate benchmark.
  7. The sample size for the training set:

    • Not applicable. There is no "training set" as this is not a machine learning model.
  8. How the ground truth for the training set was established:

    • Not applicable, as there is no training set.

Conclusion from the 510(k) Summary:

The 510(k) submission for CoStorSol® demonstrates substantial equivalence to the predicate device, ViaSpan® (K944866), by showing that it is an identical solution with the same intended use and chemical composition. Supporting evidence includes:

  • Biocompatibility testing against ISO 10993.
  • Validation of sterilization processes against ISO 17665 or USP <1116>.
  • Chemical analyses and Fourier transformed infrared scans confirming identity and stability to the predicate.

This approach is common for devices that are direct copies or minor modifications of existing, legally marketed products, where the safety and effectiveness profile is already well-established by the predicate. Therefore, the "study" demonstrating the device meets acceptance criteria is primarily a series of laboratory tests and analyses confirming identity and adherence to relevant standards.

{0}------------------------------------------------

K073693
182

510(k) SUMMARY

SUBMITTER INFORMATION

  • A. Company Name: Preservation Solutions Inc.
  • B. Company Address: 980 Proctor Drive Elkhorn, Wisconsin 32121
  • C. Company Phone: 262 723 6715
  • D. Company Facsimile: 262 723 4013
  • E. Contact Person: William Wagner Quality Assurance Director

DEVICE IDENTIFICATION

  • A. Device Trade Name: CoStorSol®
  • B. Device Common Name: Organ Storage Solution
  • C. Classification Name: Organ Storage Solution systems and accessories
  • D. Class II (21 CRF 876.5880)
  • E. Device Code: KDL

IDENTIFICATION OF PREDICATE DEVICES

CoStorSol® is a cold storage solution for harvested organs, which is substantially equivalent to a ViaSpan® solution (K944866).

JUL - 3 2008

{1}------------------------------------------------

K073693
2 of 2

DEVICE DESCRIPTION

Preservation Solutions, Inc. manufacturers CoStorSol® according to a "recipe" pioneered at the University of Wisconsin by Dr. Folkert O. Belzer. The cold storage solution is still referred to as "Belzer UW" solution, even though it has been marketed under the trade name ViaSpan®. The formulation includes soluble colloids, buffers, sodium and postassium salts, redox stabilizers, and phosphoric compounds to aid tissue viability by enabling regeneration of adenosine triphosphate (ATP).

CoStorSol® is a clear to light yellow, sterile, non-pyrogenic solution for hypothermic flushing and storage of organs. The solution is packaged in 1-liter bags.

INDICATIONS FOR USE

CoStorSol® is intended for the flushing and cold storage of kidney, liver and pancreas organs at the time of organ removal from the donor in preparation for storage, transportation and eventual transplantation into a recipient.

COMPARISON TO PREDICATE DEVICES

ViaSpan® (Belzer - UW) Organ Solution, cleared under 510(k) K944866 serves as an exact predicate for Preservation Solutions, Inc.'s CoStorSol®, in that it is an identical solution with exactly the same intended use and chemical composition. CoStorSol® and the predicate solution, Viaspan®, are both supplied in 1-liter flexible bags to ease connection to standard administration sets for flushing of harvested organs. The two solutions are designed for refrigerated storage without freezing, and for use cold. Both are used as a vasculature flush medium for an organ then as a cold storage medium.

BIOCOMPATIBILITY, STERILIZATION, PACKAGING, AND BENCH TESTING

Test results have shown CoStorSol® itself to be a biocompatible solution, supplied in flexible solution administration pouches, which have likewise been tested and shown to be biocompatible. The ISO 10993 series of standards were referenced during the planning and execution of all biocompatibility testing. CoStorSol® is supplied sterile and non-pyrogenic in order to assure safety for transplant recipients. Sterilization processes were validated according to either ISO 17665 or USP Section <1116>, as appropriate.

Chemical analyses on both fresh and aged samples of CoStorSol® show it to be identical to the predicate solution. Furthermore, Fourier transformed infrared scans confirmed stability and chemical identity for the CoStorSol® product.

CONCLUSION

The above statements establish substantial equivalence between CoStorSol® and the cited predicate.

{2}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle or bird-like symbol with three curved lines representing its wings or body. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the bird symbol.

Public Health Service

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

Preservation Solutions, Incorporated % Mr. Neil Burris Neil Burris and Associates 4250 Grove Street DENVER CO 80211

Re: K073693

Trade/Device Name: CoStorSol® Regulation Number: 21 CFR 876.5880 Regulation Name: Isolated kidney perfusion and transport system and accessories Regulatory Class: II Product Code: KDN Dated: June 13, 2008 Received: June 17, 2008

Dear Mr. Burris:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{3}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. Your of t comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements is set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act), 21 CFR 1000-1050.

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permotto your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance to one of the following numbers, based on the regulation number at the top of this letter.

21 CFR 876.xxxx(Gastroenterology/Renal/Urology)240-276-0115
21 CFR 884.xxxx(Obstetrics/Gynecology)240-276-0115
21 CFR 892.xxxx(Radiology)240-276-0120
Other240-276-0100

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRF's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at 24/02/6 3474. For questions regarding the reporting of device adverse events (Medical Device Leporting (MDR)), please contact the Division of Surveillance Systems at 240-276-3464. You may other other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) or 38-2011 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.2011 0.1

Sincerely yours.

Nancy C. Brogdon

Nancy C. Brogdon . Director, Division of Reproductive. Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

K073693

4. Statement of Indication for Use

Device Name: CoStorSol®

Indications for Use

CoStorSol® is intended for the flushing and cold storage of kidney, liver and pancreas organs at the time of organ removal from the donor in preparation for storage, transportation and eventual transplantation into a recipient.

Prescription Use XXXX (Part 21 CFR 801 Subpart D)

AND/OR

Over-The-Counter Use (21 CFR 807 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Helentlemen

(Division Sign-Off) Division of Reproductive, Abdominal, and Radiological Devices

510(k) Number __

§ 876.5880 Isolated kidney perfusion and transport system and accessories.

(a)
Identification. An isolated kidney perfusion and transport system and accessories is a device that is used to support a donated or a cadaver kidney and to maintain the organ in a near-normal physiologic state until it is transplanted into a recipient patient. This generic type of device may include tubing, catheters, connectors, an ice storage or freezing container with or without bag or preservatives, pulsatile or nonpulsatile hypothermic isolated organ perfusion apparatus with or without oxygenator, and disposable perfusion set.(b)
Classification. Class II (performance standards).