(120 days)
The ARCHITECT Homocysteine assay is a chemiluminescent microparticle immunoassay (CMIA) for the quantitative determination of total L-homocysteine in human serum or plasma on the ARCHITECT i System. Homocysteine values can assist in the diagnosis and treatment of patients suspected of having hyperhomocysteinemia and homocystinuria.
The ARCHITECT Homocysteine Calibrators are for the calibration of the ARCHITECT i System when used for the quantitative determination of total L-homocysteine in human serum or plasma.
The ARCHITECT Homocysteine Controls are for the estimation of test precision and the detection of systematic analytical deviations of the ARCHITECT i System (reagents, calibrators and instrument), when used for the quantitative determination of total L-homocysteine in human serum or plasma.
For in vitro diagnostic use.
The ARCHITECT Homocysteine assay is a one-step immunoassay for the quantitative determination of total L-homocysteine in human serum or plasma using CMIA technology, with flexible assay protocols, referred to as Chemiflex. Bound or dimerised homocysteine (oxidized form) is reduced by dithiothreitol (DTT) to free homocysteine, which is then converted to Sadenosyl homocysteine (SAH) by the action of the recombinant enzyme S-adenosyl homocysteine hydrolase (rSAHHase) in the presence of excess adenosine. The SAH then competes with acridinium-labeled S-adenosyl cysteine for particle-bound monoclonal antibody. Following a wash stage and magnetic separation, pre-trigger and trigger solutions are added to the reaction mixture and the resulting chemiluminescence is measured as relative light units (RLUs). An indirect relationship exists between the amount of homocysteine in the sample and the RLUs detected by the ARCHITECT i System optics.
The ARCHITECT Homocysteine assay is compared to the AxSYM Homocysteine assay. The acceptance criteria and the study results proving the device meets these criteria are as follows:
- Table of Acceptance Criteria and Reported Device Performance:
| Performance Metric | Acceptance Criteria (Implicit) | Reported Device Performance |
|---|---|---|
| Method Comparison (Regression) | Slope between 0.95 and 1.05 | 0.98 |
| Intercept close to 0 | -0.74 | |
| Correlation Coefficient (r) > 0.95 | 0.98 | |
| Precision | Substantially equivalent to predicate device | Substantially equivalent to predicate device |
| Linearity | Substantially equivalent to predicate device | Substantially equivalent to predicate device |
| Interferences | Substantially equivalent to predicate device | Substantially equivalent to predicate device |
Note: The acceptance criteria are implicit based on the statement of "substantially equivalent performance" and typical ranges for method comparison studies in clinical chemistry. The provided 510(k) summary does not explicitly state numerical acceptance criteria, but rather demonstrates equivalence to a predicate device.
-
Sample size used for the test set and the data provenance:
- Sample Size: 456 plasma samples.
- Data Provenance: Not specified (country of origin, retrospective/prospective). Since it's a method comparison study for an in vitro diagnostic, it would typically involve clinical samples, likely prospective or a well-characterized retrospective cohort.
-
Number of experts used to establish the ground truth for the test set and the qualifications of those experts:
- Not applicable. For an immunoassay, the "ground truth" is typically established by the reference method or comparative method, which in this case is the predicate device (AxSYM Homocysteine assay). The performance of the predicate device is assumed to be accurate.
-
Adjudication method for the test set:
- Not applicable. This is a method comparison study for an in vitro diagnostic device, not an image-based diagnostic or clinical trial requiring expert adjudication of diagnoses. The predicate device's results serve as the comparison point.
-
If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
- Not applicable. This is not an AI-based device for image interpretation or diagnosis by human readers. It is an in vitro diagnostic immunoassay.
-
If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:
- Yes, the study performed is a standalone performance assessment of the ARCHITECT Homocysteine assay against a predicate device. The assay itself is automated and does not involve a human in the loop for its direct analytical performance once a sample is loaded.
-
The type of ground truth used:
- Comparative data with a legally marketed predicate device: The AxSYM Homocysteine assay results were used as the comparison "ground truth" to establish substantial equivalence.
-
The sample size for the training set:
- Not applicable. This device is an immunoassay, not a machine learning or AI algorithm that requires a separate training set. Its chemical and mechanical principles are fixed during development.
-
How the ground truth for the training set was established:
- Not applicable (see point 8).
{0}------------------------------------------------
510(k) Summary
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92.
The assigned 510(k) number is: K073640
Submission correspondent:
Dr Claire Dora Regulatory Affairs Officer Axis-Shield Diagnostics, Ltd. The Technology Park Dundee DD2 1XA, UK Telephone: +44 13 82 422000 Date prepared: December 19, 2007
Device Name: ARCHITECT Homocysteine
Reagents:
Classification Name: Urinary Homocystine (Nonquantitative) Test System Trade Name: ARCHITECT Homocysteine Common Name: Homocysteine Enzyme Immunoassay Governing Regulation: 21 CFR 862.1377 Device Classification: Class II Classification Panel: Clinical Chemistry Product Code: LPS
Calibrators:
Classification Name: Calibrator, Secondary Trade Name: ARCHITECT Homocysteine Calibrators Common Name: Calibrator Governing Regulation: 862.1150 Device Classification: Class II Classification Panel: Clinical Chemistry Product Code: JIT
Controls:
Classification Name: Single (specified) analyte controls (assayed and unassayed) Trade Name: ARCHITECT Homocysteine Control Common Name: Control Governing Regulation: 862.1660 Device Classification: Class I Classification Panel: Clinical Chemistry Product Code: JJX
Legally marketed device to which equivalency is claimed:
】【 2014/08/2019 08:00:00 PM ID: 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0
AxSYM Homocysteine Assay; K992858.
{1}------------------------------------------------
Intended Use of Device:
The ARCHITECT Homocysteine assay is a chemiluminescent microparticle immunoassay (CMIA) for the quantitative determination of total L-homocysteine in human serum or plasma on the ARCHITECT i System.
Homocysteine values can assist in the diagnosis and treatment of patients suspected of having hyperhomocysteinemia and homocystinuria.
Description of Device:
The ARCHITECT Homocysteine assay is a one-step immunoassay for the quantitative determination of total L-homocysteine in human serum or plasma using CMIA technology, with flexible assay protocols, referred to as Chemiflex. Bound or dimerised homocysteine (oxidized form) is reduced by dithiothreitol (DTT) to free homocysteine, which is then converted to Sadenosyl homocysteine (SAH) by the action of the recombinant enzyme S-adenosyl homocysteine hydrolase (rSAHHase) in the presence of excess adenosine. The SAH then competes with acridinium-labeled S-adenosyl cysteine for particle-bound monoclonal antibody. Following a wash stage and magnetic separation, pre-trigger and trigger solutions are added to the reaction mixture and the resulting chemiluminescence is measured as relative light units (RLUs). An indirect relationship exists between the amount of homocysteine in the sample and the RLUs detected by the ARCHITECT i System optics.
Comparison of Technological Characteristics:
ARCHITECT Homocysteine and AxSYM Homocysteine are both automated immunoassays for the quantitative determination of total L-homocysteine in human serum or plasma.
The ARCHITECT and AxSYM systems differ in their detection methods; the ARCHITECT is a chemiluminescent microparticle immunoassay (CMIA) whereas the AxSYM is a fluorescence polarization immunoassay (FPIA).
Summary of Non-Clinical Performance:
The ARCHITECT Homocysteine assay is substantially equivalent to the AxSYM Homocysteine assay in terms of precision, linearity and interferences as demonstrated in non-clinical performance data in this 510(k) submission.
Summary of Clinical Performance:
The ARCHITECT Homocysteine assay demonstrated substantially equivalent performance to the AxSYM Homocysteine indicated by a method comparison study.
Passing-Bablok linear regression method comparison was performed on 456 plasma samples in the range of 3.70 to 49.94 umol/L. The ARCHITECT Homocysteine versus AxSYM Homocysteine gave a slope of 0.98 (95% Confidence interval 0.97 to 1.00) and an intercept of -0.74 (95% Confidence interval -0.99 to -0.54). ARCHITECT Homocysteine versus AxSYM
Homocysteine gave an r value of 0.98 (95% Confidence interval 0.98 to 0.99).
{2}------------------------------------------------
DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle with its wings spread, symbolizing protection and care. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the eagle.
Food and Drug Administration 2098 Gaither Road Rockville MD 20850
APR 2 4 2008
Axis-Shield Diagnostics, Ltd. c/o Dr. Claire Dora Acting Regulatory Affairs Manager Luna Place. The Technology Park Dundee, DD2 1XA, Scotland United Kingdom
Re: K073640
Trade/Device Name: ARCHITECT Homocysteine Reagents, ARCHITECT Homocysteine Calibrators, and ARCHITECT Homocysteine Controls Regulation Number: 21 CFR 862.1377 Regulation Name: Urinary homocystine (non-quantitative) test system. Regulatory Class: II Product Code: LPS, JIT, JJX Dated: March 13, 2008 Received: March 17, 2008
Dear Dr. Dora:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21
{3}------------------------------------------------
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific information about the application of labeling requirements to your device, or questions on the promotion and advertising of your device, please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (240) 276-0490. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address at http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
Jean M. Cooper, M.S., D.v.M.
Jean M. Cooper, M.S., D.V.M. Director Division of Chemistry and Toxicology Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
Indication for Use
510(k) Number (if known): K073640
Device Name: AR CHITECT Homocysteine Reagents, ARCHITECT Homocysteine Calibrators and ARCHITECT Homocysteine Controls
Indication For Use:
Reagents:
Homocysteine assay is a chemiluminescent microparticle The ARCHITECT immunoassay (CMIA) for the quantitative determination of total 1-homocysteine in human serum or plasma on the ARCHITECT i System. Homocysteine values can assist in the diagnosis and treatment of patients suspected of having hyperhomocysteinemia and homocystinuria.
Calibrators:
The ARCHITECT Homocysteine Calibrators are for the calibration of the ARCHITECT i System when used for the quantitative determination of total 1-homocysteine in human serum or plasma.
Controls:
The ARCHITECT Homocysteine Controls are for the estimation of test precision and the detection of systematic analytical deviations of the ARCHITECT i System (reagents, calibrators and instrument), when used for the quantitative determination of total r-homocysteine in human serum or plasma.
For in vitro diagnostic use.
Prescription Use X (21 CFR Part 801 Subpart D) And/Or
Over the Counter Use __ (21 CFR Part 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE; CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of In Vitro Diagnostic Device Evaluation and Safety (OIVD)
Carol C. Benson
Division Sign-Off Office of In Vitro Diagnostic Device Evaluation and Safety
510(k) K073640
Page 1 of 1
§ 862.1377 Urinary homocystine (nonquantitative) test system.
(a)
Identification. A urinary homocystine (nonquantitative) test system is a device intended to identify homocystine (an analogue of the amino acid cystine) in urine. The identification of urinary homocystine is used in the diagnosis and treatment of homocystinuria (homosystine in urine), a heritable metabolic disorder which may cause mental retardation.(b)
Classification. Class II.