K Number
K073258
Date Cleared
2008-02-14

(86 days)

Product Code
Regulation Number
872.3630
Panel
Dental
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Atlantis Abutment is intended for use with an endosseous implant to support a prosthetic device in a partially or completely edentulous patient. It is intended for use to support single and multiple tooth prosthesis, in the mandible or maxilla. The prosthesis can be cement retained to the abutment. The abutment screw is intended to secure the abutment to the endosseous implant.

This device is compatible with the following manufacturers' implant systems: BioHorizons The Prodigy System™ Dental Implants 3.5mm, 4.0mm, 5.0mm and 6.0mm Implants.

Please note: This device may be used in an early load situation, but is dependent on the specific implant system and protocol used by the dental professional.

Highly angled abutments (i.e. 30 degrees) on implants with diameters less than 4 mm are intended for the anterior region of the mouth and are not intended for the posterior region due to limited strength of the implant.

Device Description

The devices covered in this submission are abutments which are placed into the dental implant to provide support for a prosthetic restoration. The subject abutments are indicated for cemented restorations.

The Atlantis™ Abutments for BioHorizons Implant and abutment screws are made from Titanium grade Ti-6A1-4V ELI (Meets ASTM Standard F-136). The abutment is placed over the implant shoulder and is mounted into the implant with a screw. The abutments are compatible with BioHorizons Prodigy System™ 3.5mm, 4.0mm, 5.0mm and 6.0mm Implants and BioHorizons

AI/ML Overview

This submission, K073258, is for the Atlantis Abutment for BioHorizons Implant. It's a pre-market notification (510(k)) that claims substantial equivalence to previously cleared devices, not an AI/ML device, and therefore does not have the acceptance criteria or study details requested in the prompt. The document details the device's name, description, intended use, and the basis for substantial equivalence to predicate devices, but does not contain any information about acceptance criteria, device performance metrics, sample sizes, expert ground truth, adjudication methods, MRMC studies, standalone performance, or training set details as it is not an AI/ML product.

The device is an endosseous dental implant abutment made from Titanium grade Ti-6A1-4V ELI, intended to support prosthetic reconstructions in partially or completely edentulous patients.

Since this request is about an AI/ML device and the provided document describes a dental implant abutment, it does not contain the information required to answer the specific questions posed.

{0}------------------------------------------------

K073258

Pre-market Notification Page - 000009 - - - - - - - - - - - - - - - - -

VII. SECTION 10 - 510(K) SUMMARY

This summary of 510(k) substantial equivalence information is being submitted in accordance with the requirements of 21 CFR Part 807.92.

Applicant's Name and Address 1 .

Atlantis Components Inc. 25 First Street Cambridge, Massachusetts 02141 Telephone Number: 617-871-2783 Fax Number: 617-871-6607 Contact Person: Franklin Uyleman Manager of Quality and Regulatory Affairs

2. Name of Device

Trade Name:Atlantis™ Abutment for BioHorizons Implant
Common Name:Endosseous dental implant abutment
Classification Name:Endosseous dental implant abutment
21 CFR 872.3630 Product code NHA

Legally Marketed Device to which Equivalence is claimed (Predicate Device) 3.

ManufacturerDevice510(k)Number
Atlantis ComponentsInc.Atlantis™ Abutment for ZimmerInterfaceK053373
BioHorizons ImplantSystems Inc.The Prodigy System™ DentalImplantsK042429

4. Description of the Device

The devices covered in this submission are abutments which are placed into the dental implant to provide support for a prosthetic restoration. The subject abutments are indicated for cemented restorations.

{1}------------------------------------------------

Pre-market Page - 000010 - - - - - - - - - - - - - - - - -

4. Description of the Device (continued)

The Atlantis™ Abutments for BioHorizons Implant and abutment screws are made from Titanium grade Ti-6A1-4V ELI (Meets ASTM Standard F-136). The abutment is placed over the implant shoulder and is mounted into the implant with a screw. The abutments are compatible with BioHorizons Prodigy System™ 3.5mm, 4.0mm, 5.0mm and 6.0mm Implants and BioHorizons

5. Intended Use of the Device

The devices covered by this submission are abutments which are placed into a dental implant to provide support for a prosthetic reconstruction. The Atlantis Abutment is intended for use with an endosseous implant to support a prosthetic device in a partially or completely edentulous patient. It is intended for use to support single and multiple tooth prosthesis, in the mandible or maxilla. The prosthesis can be cement retained to the abutment. The abutment screw is intended to secure the abutment to the endosseous implant.

Basis for Substantial Equivalence 6.

The Atlantis™ Abutments for BioHorizons Implants are substantially equivalent in intended use, material, design and performance to the Atlantis Abutments for Zimmer Interface cleared under K053373 and BioHorizons Prodigy System™ Dental Implants cleared under K042429.

{2}------------------------------------------------

Image /page/2/Picture/2 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle-like symbol with three overlapping, curved lines forming the body and wings. The symbol is enclosed within a circular border, and the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged around the upper portion of the circle.

FEB 1 4 2008

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

Atlantis Components, Incorporated C/O Ms. Betsy A. Brown Consultant B.A. Brown & Associates 8944 Tamaroa Terracc Skokie, Illinois 60076

Re: K073258

Trade/Device Name; AtlantisTM Abutment for BioHorizons Implant Regulation Number: 21 CFR 872.3630 Regulation Name: Endosseous Dental Implant Abutment Regulatory Class: II Product Code: NHA Dated: November 16, 2007 Received: November 20, 2007

Dear Ms. Brown:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{3}------------------------------------------------

Page 2 - Ms. Brown

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely yours,

Sutte K. Michie Davis

Chiu Lin, Ph.D. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

Pre-market Notification Page - 000227

Indications for Use

KO73258 510(k) Number (if Known)_

Device Name: Atlantis ™ Abutment for BioHorizons Implant

Indication for Use:

The Atlantis Abutment is intended for use with an endosseous implant to support a prosthetic device in a partially or completely edentulous patient. It is intended for use to support single and multiple tooth prosthesis, in the mandible or maxilla. The prosthesis can be cement retained to the abutment. The abutment screw is intended to secure the abutment to the endosseous implant.

This device is compatible with the following manufacturers' implant systems: BioHorizons The Prodigy System™ Dental Implants 3.5mm, 4.0mm, 5.0mm and 6.0mm Implants.

Please note: This device may be used in an early load situation, but is dependent on the specific implant system and protocol used by the dental professional.

Highly angled abutments (i.e. 30 degrees) on implants with diameters less than 4 mm are intended for the anterior region of the mouth and are not intended for the posterior region due to limited strength of the implant.

Prescription Use 12 (Part 21 CFR 801 SubpartD)

AND/OR

Over-The-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Susan Runner

(Division Sign-Off) Division of Anesthesiology, General Hospital Infection Control, Dental Devices

510(k) Number: K073258

§ 872.3630 Endosseous dental implant abutment.

(a)
Identification. An endosseous dental implant abutment is a premanufactured prosthetic component directly connected to the endosseous dental implant and is intended for use as an aid in prosthetic rehabilitation.(b)
Classification. Class II (special controls). The guidance document entitled “Class II Special Controls Guidance Document: Root-Form Endosseous Dental Implants and Endosseous Dental Implant Abutments” will serve as the special control. (See § 872.1(e) for the availability of this guidance document.)