K Number
K072798
Device Name
VIATRAC 14 PLUS PERIPHERAL DILATION CATHETER
Date Cleared
2008-01-31

(122 days)

Product Code
Regulation Number
870.1250
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP Authorized
Intended Use
The Viatrac® 14 PLUS Peripheral Dilatation Catheter is indicated to dilate stenosis in the peripheral vasculature (iliac, femoral, ilio-femoral, popliteal, infra-popliteal, renal arteries, and carotid arties) and for the treatment of obstructive lesions of native or synthetic arteriovenous dialysis fistulae. This device is also indicated for post-dilatation of balloon-expandable and self-expanding stents in the peripheral vasculature.
Device Description
The Viatrac 14 PLUS Pcripheral Dilatation Catheter has an integrated shaft system and a balloon near the distal tip. The shaft has a combination of single lumen and dual lumen tubing. One lumen is used for inflation of the balloon with contrast medium. The second lumen in the distal shaft permits the use of a guide wire to facilitate the advancement of the catheter to and through the stenosis to be dilated. The balloon has radiopaque marker(s) to aid in positioning the balloon in the stenosis, and is designed to provide an expandable segment of known diameter and length at a specific pressure. On the 135 cm catheter length, there are two proximal shaft markers (95 cm and 105 cm from the distal tip). On the 80 cm catheter length, there is a single proximal marker (55 cm from the distal tip). Both indicate the relative position of the catheter to the end of a brachial, femoral or renal guiding catheter. An additional marker is located at the guide wire exit notch and aids in locating the guide wire exit notch. The design of this catheter does not incorporate a lumen for distal dye injections and distal pressure measurements.
More Information

Not Found

No
The description focuses on the mechanical components and function of a balloon catheter for dilation, with no mention of AI or ML capabilities.

Yes
The device is used to dilate stenosis in arteries and for the treatment of obstructive lesions, which are therapeutic actions.

No

The device is a dilatation catheter used to treat stenosis in peripheral vasculature and obstructive lesions. Its function is interventional (therapeutic), not diagnostic.

No

The device description clearly details a physical catheter with a balloon, lumens, and markers, indicating it is a hardware medical device, not software-only.

No, this device is not an IVD (In Vitro Diagnostic).

Here's why:

  • IVD Definition: In Vitro Diagnostic devices are used to examine specimens taken from the human body (like blood, urine, or tissue) to provide information about a person's health. This testing is done outside of the body.
  • Device Function: The Viatrac 14 PLUS Peripheral Dilatation Catheter is a medical device used inside the body to physically dilate narrowed blood vessels. It is a therapeutic device, not a diagnostic one.
  • Intended Use: The intended use clearly describes a procedure performed directly on the patient's vasculature, not the analysis of a biological sample.

The information provided about the device's function, intended use, and description all point to it being an interventional medical device used for treatment, not an IVD used for diagnosis through laboratory testing.

N/A

Intended Use / Indications for Use

The Viatrac 14 PLUS Peripheral Dilatation Catheter is indicated to dilate stenosis in the peripheral vasculature (iliac, femoral, ilio-femoral, popliteal, infra popliteal, renal and carotid arteries) and for the treatment of obstructive lesions of native or synthetic arteriovenous dialysis fistulae. This device is also indicated for post-dilatation of balloon-expandable and self-expanding stents in the peripheral vasculature.

Product codes

DQY, LIT

Device Description

The Viatrac 14 PLUS Pcripheral Dilatation Catheter has an integrated shaft system and a balloon near the distal tip. The shaft has a combination of single lumen and dual lumen tubing. One lumen is used for inflation of the balloon with contrast medium. The second lumen in the distal shaft permits the use of a guide wire to facilitate the advancement of the catheter to and through the stenosis to be dilated.

The balloon has radiopaque marker(s) to aid in positioning the balloon in the stenosis, and is designed to provide an expandable segment of known diameter and length at a specific pressure.

On the 135 cm catheter length, there are two proximal shaft markers (95 cm and 105 cm from the distal tip). On the 80 cm catheter length, there is a single proximal marker (55 cm from the distal tip). Both indicate the relative position of the catheter to the end of a brachial, femoral or renal guiding catheter. An additional marker is located at the guide wire exit notch and aids in locating the guide wire exit notch.

The design of this catheter does not incorporate a lumen for distal dye injections and distal pressure measurements.

Mentions image processing

Not Found

Mentions AI, DNN, or ML

Not Found

Input Imaging Modality

Not Found

Anatomical Site

peripheral vasculature (iliac, femoral, ilio-femoral, popliteal, infra popliteal, renal and carotid arteries), arteriovenous dialysis fistulae

Indicated Patient Age Range

Not Found

Intended User / Care Setting

Not Found

Description of the training set, sample size, data source, and annotation protocol

Not Found

Description of the test set, sample size, data source, and annotation protocol

Not Found

Summary of Performance Studies (study type, sample size, AUC, MRMC, standalone performance, key results)

The substantial equivalence of the Viatrac 14 PLUS Peripheral Dilatation Catheter has been demonstrated through data collected from in vitro bench tests and retrospective analyses of clinical studies.

Key Metrics (Sensitivity, Specificity, PPV, NPV, etc.)

Not Found

Predicate Device(s)

K012050, K053118, K071189

Reference Device(s)

Not Found

Predetermined Change Control Plan (PCCP) - All Relevant Information

Not Found

§ 870.1250 Percutaneous catheter.

(a)
Identification. A percutaneous catheter is a device that is introduced into a vein or artery through the skin using a dilator and a sheath (introducer) or guide wire.(b)
Classification. Class II (performance standards).

0

K072798 page 143

JAN 31 2002

510(k) Summary

Submitter's NameAbbott Vascular
Submitter's Address:3200 Lakeside Drive
Santa Clara, CA 95054
FDA Registration
Number:3004635528
Telephone:408-845-0869
Fax:408-845-0921
Contact Person:Cherece L. Jones, Regulatory Affairs
Associate
Date Prepared:September 24, 2007
Device Trade Name:Viatrac® 14 PLUS Peripheral Dilatation
Catheter
Device Common Name:Catheter, Percutaneous
Product Code :DQY
Device Classification:Class II

The 510(k) Summary is submitted in accordance with 21 CFR Part 807, Section 807.92

Summary of Substantial Equivalence:

The Viatrac 14 PLUS Peripheral Dilatation Catheter with modified indications is substantially equivalent to the Viatrac 14 PLUS Peripheral Dilatation Catheter (K012050, cleared 09/06/2001) in addition to the currently marketed predicate devices, the Sterling Monorail PTA Balloon Dilatation Catheter, K053118 and the Cordis Aviator PLUS Percutaneous Transluminal Angioplasty (PTA) Balloon Catheter, K071189. Changes made to the Viatrac 14 PLUS Peripheral Dilatation Catheter will more closely align the indications with Clinical use. There are no changes to the device included in this proposal or as a result of this change.

1

Device Description:

The Viatrac 14 PLUS Pcripheral Dilatation Catheter has an integrated shaft system and a balloon near the distal tip. The shaft has a combination of single lumen and dual lumen tubing. One lumen is used for inflation of the balloon with contrast medium. The second lumen in the distal shaft permits the use of a guide wire to facilitate the advancement of the catheter to and through the stenosis to be dilated.

The balloon has radiopaque marker(s) to aid in positioning the balloon in the stenosis, and is designed to provide an expandable segment of known diameter and length at a specific pressure.

On the 135 cm catheter length, there are two proximal shaft markers (95 cm and 105 cm from the distal tip). On the 80 cm catheter length, there is a single proximal marker (55 cm from the distal tip). Both indicate the relative position of the catheter to the end of a brachial, femoral or renal guiding catheter. An additional marker is located at the guide wire exit notch and aids in locating the guide wire exit notch.

The design of this catheter does not incorporate a lumen for distal dye injections and distal pressure measurements.

Intended Use:

The Viatrac 14 PLUS Peripheral Dilatation Catheter is indicated to dilate stenosis in the peripheral vasculature (iliac, femoral, ilio-femoral, popliteal, infra popliteal, renal and carotid arteries) and for the treatment of obstructive lesions of native or synthetic arteriovenous dialysis fistulae. This device is also indicated for post-dilatation of balloon-expandable and self-expanding stents in the peripheral vasculature.

Technological Characteristics:

There have been no changes to the technological characteristics such as materials, biocompatibility, performance properties, sterilization and packaging therefore they are substantially equivalent to the currently marketed predicate devices.

Performance Data:

The substantial equivalence of the Viatrac 14 PLUS Peripheral Dilatation Catheter has been demonstrated through data collected from in vitro bench tests and retrospective analyses of clinical studies.

2

DEPARTMENT OF HEALTH & HUMAN SERVICES

Public Health Service

Image /page/2/Picture/2 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three flowing ribbons extending from its body. The eagle faces to the right. Encircling the eagle is the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" in a circular arrangement.

JAN 31 2008

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

Abbott Vascular, Inc. c/o Ms. Cherece Jones Regulatory Affairs Associate 3200 Lakeside Drive Santa Clara, CA 95054

Re: K072798

Trade/Device Name: Viatrac 14 PLUS Peripheral Dilatation Catheter Regulation Number: 21 CFR 870.1250 Regulation Name: Percutaneous Catheter Regulatory Class: Class II (two) Product Code: LIT Dated: September 24, 2007 Received: October 1, 2007

Dear Ms. Jones:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

3

Page 2 - Ms. Cherece Jones

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at 240-276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at 240-276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely yours,

una R. Vachner

A Bram D. Zuckerman, M.D. - Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

4

Indications for Use Statement

510(k) Number (if known)K072798
Device NameViatrac® 14 PLUS Peripheral Dilatation Catheter
Indications for UseThe Viatrac® 14 PLUS Peripheral Dilatation Catheter is indicated to dilate stenosis in the peripheral vasculature (iliac, femoral, ilio-femoral, popliteal, infra-popliteal, renal arteries, and carotid arties) and for the treatment of obstructive lesions of native or synthetic arteriovenous dialysis fistulae. This device is also indicated for post-dilatation of balloon-expandable and self-expanding stents in the peripheral vasculature.
Prescription Use (Part 21 CFR 801 Subpart D)X
OR
Over-The-Counter Use (Part 21 CFR 801 Subpart C)

PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED

Concurrence of CDRH, Office of Device Evaluation (ODE)

Donna R. Vachner
(Div: Dir-Off)

ेal diovascular Devices

... . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .