K Number
K072283
Date Cleared
2007-09-19

(34 days)

Product Code
Regulation Number
870.1250
Reference & Predicate Devices
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Dorado™ PTA Balloon Dilatation Catheter is recommended for Percutaneous Transluminal Angioplasty (PTA) of the renal, iliac, femoral, popliteal, tibial, peroneal, and subclavian arteries and for the treatment of obstructive lesions of native or synthetic arteriovenous dialysis fistulae. This device is also recommended for post-dilatation of balloon expandable and self-expanding stents in the peripheral vasculature. This catheter is not for use in the coronary arteries.

Device Description

The subject device, the Dorado™ PTA Balloon Dilatation Catheter, is composed of a multi-lumen catheter with 5 or 5.8 French shaft outer diameter, depending on balloon size. A composite balloon is mounted onto the distal tip of the catheter. During use, the balloon position in the vessel is identified by two radiopaque marker bands that indicate the working length of the balloon and aid in placement.

AI/ML Overview

The provided text is a 510(k) summary for a medical device (Dorado™ PTA Balloon Dilatation Catheter), which focuses on demonstrating substantial equivalence to predicate devices, rather than an AI/ML device study. Therefore, much of the requested information regarding AI/ML device performance, such as sample sizes for test/training sets, expert qualifications, adjudication methods, MRMC studies, and ground truth types, is not applicable or available in this document.

However, I can extract the general acceptance criteria and the conclusion regarding the device's performance based on the provided text.

Acceptance Criteria and Device Performance for Dorado™ PTA Balloon Dilatation Catheter

Acceptance CriteriaReported Device Performance
Pre-determined acceptance criteria of design verification and validation as specified by applicable standards, guidance, test protocols, and/or customer inputs.The Dorado™ PTA Balloon Dilatation Catheter met all the predetermined acceptance criteria of design verification and validation as specified by applicable standards, guidance, test protocols and/or customer inputs.

Study Information (Based on provided text):

  • Sample Size for Test Set and Data Provenance: Not applicable. This document describes a device submission based on design verification and validation to demonstrate substantial equivalence to predicate devices, not a clinical study with a test set in the conventional sense of AI/ML performance evaluation.
  • Number of Experts and Qualifications for Ground Truth: Not applicable. Ground truth as typically defined for AI/ML models is not relevant here. The evaluation is against engineering specifications, standards, and comparison to legally marketed predicate devices.
  • Adjudication Method: Not applicable. No adjudication method in the context of expert review for AI/ML performance is described. The "adjudication" is essentially the FDA's review and determination of substantial equivalence based on the provided data.
  • Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study: No. This is not an AI/ML device, and no MRMC study is mentioned.
  • Standalone Performance Study (Algorithm Only): No. This information is not relevant for a traditional medical device like a balloon catheter.
  • Type of Ground Truth Used: Not applicable in the context of AI/ML. The "ground truth" for this device's acceptance would be its adherence to established engineering specifications, performance standards, and comparison to the documented performance and design of the predicate devices.
  • Sample Size for Training Set: Not applicable. This is not an AI/ML device, so there is no "training set."
  • How Ground Truth for Training Set was Established: Not applicable.

Summary of the Device's Evaluation:

The Dorado™ PTA Balloon Dilatation Catheter underwent a design verification and validation process. The conclusion states that the device "met all the predetermined acceptance criteria of design verification and validation as specified by applicable standards, guidance, test protocols and/or customer inputs." This implies that the device manufacturing, materials, and performance (e.g., balloon inflation/deflation, catheter trackability, radiopacity, etc.) were tested against established specifications and found to be compliant.

The 510(k) submission primarily focuses on demonstrating "substantial equivalence" to existing, legally marketed predicate devices rather than proving novel clinical efficacy or safety through a large-scale clinical trial. The comparison of indications for use, technological characteristics, and fundamental scientific technology to predicate devices formed the basis of this substantial equivalence claim.

{0}------------------------------------------------

510(k) Summary of Safety and Effectiveness 21 CFR 807.92

As required by the Safe Medical Devices Act of 1990, coded under Section 513, Part (I)(3)(A) of the Food, Drug and Cosmetic Act, a summary of the safety and effectiveness information upon which substantial equivalence determination is based as follows:

Submitter Information:

Applicant:Bard Peripheral Vascular, Inc1625 West 3rd StreetP.O. Box 1740Tempe, Arizona 85280
Phone:480-303-2524
Fax:480-449-2546
Contact:Genevieve Balutowski, Senior Regulatory Affairs Specialist
DateJuly 18, 2007

Subject Device Name:

Device Trade Name:Dorado™ PTA Balloon Dilatation Catheter
Common or Usual Name:Percutaneous Catheter (21 CFR 870.1250,Product Code DQY& LIT)
Classification:Class II
Classification Panel:Cardiovascular

Predicate Devices:

  • . Conquest™ PTA Balloon Dilatation Catheter (K014212, cleared January 17, 2002), manufactured by Bard Peripheral Vascular, Inc.
  • Ultraverse® Small Vessel PTA Balloon Dilatation Catheter (K012913, . cleared September 28, 2001), manufactured by Bard Peripheral Vascular, Inc.

Image /page/0/Picture/12 description: The image shows the word "BARD" in a bold, outlined font. The letters are blocky and geometric, giving the word a strong and distinct appearance. The outline emphasizes the shape of each letter, making the word stand out.

SEP 1 9 2007

{1}------------------------------------------------

  • Sterling™ Over-The-Wire PTA Balloon Dilatation Catheter (K053116, . cleared December 16, 2005), manufactured by Boston Scientific Corporation.
  • PolarCath™ Peripheral Dilatation System (K060572, cleared March 15, . 2006), manufactured by Boston Scientific Corporation.

Device Description:

The subject device, the Dorado™ PTA Balloon Dilatation Catheter, is composed of a multi-lumen catheter with 5 or 5.8 French shaft outer diameter, depending on balloon size. A composite balloon is mounted onto the distal tip of the catheter. During use, the balloon position in the vessel is identified by two radiopaque marker bands that indicate the working length of the balloon and aid in placement.

Intended Use of Device:

Dilatation of stenosis in the peripheral vascular, for the treatment of obstructive lesions of native or synthetic arteriovenous dialysis fistulae and post-deployed stent expansion of self-expanding and balloon expandable peripheral vascular stents.

Indications for Use of Device:

The Dorado™ PTA Balloon Dilatation Catheter is recommended for Percutaneous Transluminal Angioplasty (PTA) of the renal, iliac, femoral, popliteal, tibial, peroneal, and subclavian arteries and for the treatment of obstructive lesions of native or synthetic arteriovenous dialysis fistulae. This device is also recommended for post-dilatation of balloon expandable and self-expanding stents in the peripheral vasculature. This catheter is not for use in the coronary arteries.

Comparison of Indications for Use to Predicate Devices:

The Indications for Use of the Dorado™ PTA Balloon Dilatation Catheter is a combination of the Indications for Use of the predicate devices. Therefore, while the indications statement of the subject and predicate devices are different, the differences do not affect the safety and effectiveness of the device when used as labeled as all the indications of the subject device are cleared for similar legally marketed devices.

TRADE SECRET/CONFIDENTIAL INFORMATION Notify CR Bard Before Releasing this Document.

Image /page/1/Picture/14 description: The image shows the word "BARD" in a bold, outlined font. The letters are all capitalized and evenly spaced. The font is simple and geometric, with straight lines and sharp angles. The image is in black and white.

{2}------------------------------------------------

Technological Comparison to Predicate Devices:

The Dorado™ PTA Balloon Dilatation Catheter has the following similarities to the predicate devices:

  • Similar intended use (all predicates) .
  • Similar indications for use (all predicates) .
  • Same target population (all predicates) .
  • Similar fundamental scientific technology (all predicates) .
  • t Similar operating principle (all predicates)
  • . Similar packaging materials (Conquest™ PTA Balloon Dilatation Catheter and Ultraverse® Small Vessel PTA Balloon Dilatation Catheter)
  • Same sterility assurance level and method of sterilization (Conquest™ . PTA Balloon Dilatation Catheter and Ultraverse® Small Vessel PTA Balloon Dilatation Catheter)

Conclusions:

The Dorado™ PTA Balloon Dilatation Catheter met all the predetermined acceptance criteria of design verification and validation as specified by applicable standards, guidance, test protocols and/or customer inputs. The Dorado™ PTA Balloon Dilatation Catheter is substantially equivalent to the legally marketed predicate devices, the Conquest™ PTA Balloon Dilatation Catheter, the Bard Ultraverse® Small Vessel Balloon Dilatation Catheter, the Sterling™ Over-The-Wire PTA Balloon Dilatation Catheter, and the PolarCath™ Peripheral Dilatation System.

Bard Peripheral Vascular, Inc.

TRADE SECRET/CONFIDENTIAL INFORMATION Notify CR Bard Before Releasing this Document.

Image /page/2/Picture/16 description: The image shows the word "BARD" in a bold, sans-serif font. The letters are outlined in black, giving them a distinct and modern appearance. The font style is consistent throughout the word, creating a unified and recognizable logo or brand name.

{3}------------------------------------------------

Image /page/3/Picture/1 description: The image shows the logo for the Department of Health and Human Services. The logo consists of a circular seal with the words "DEPARTMENT OF HEALTH AND HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract image of an eagle with three lines representing its wings and a stylized body.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

SEP 1 9 2007

Bard Peripheral Vascular, Inc. c/o Mr. Genevieve Balutowski Senior Regulatory Affairs Specialist 1625 West 3rd Street P.O. Box 1740 Tempe, AZ 85280-1740

Re: K072283

Dorado™ PTA Balloon Dilatation Catheter Regulation Number: 21 CFR 870.1250 Regulation Name: Percutaneous Catheter Regulatory Class: Class II Product Code: DQY, LIT Dated: September 5, 2007 Received: September 6, 2007

Dear Mr. Balutowski:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{4}------------------------------------------------

Page 2 - Mr. Genevieve Balutowski

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at 240-276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at 240-276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely yours,

R. Vuhner

ABram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

Indications for Use

510(k) Number (if known): Ko72283

Device Name: Dorado™ PTA Balloon Dilatation Catheter

Indications for Use: The Dorado™ PTA Balloon Dilatation Catheter is recommended for Percutaneous Transluminal Angioplasty (PTA) of the renal, iliac, femoral, popliteal, tibial, peroneal, and subclavian arteries and for the treatment of obstructive lesions of native or synthetic arteriovenous dialysis fistulae. This device is also recommended for postdilatation of balloon expandable and self-expanding stents in the peripheral vasculature. This catheter is not for use in the coronary arteries.

Prescription Use X (Part21 CFR 801 Subpart D) AND/OR

Over-The-Counter Use (21CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Dumua R. Vachmes

Division Sign-C Olvision of Cardiovascular Devices

510(k) Number K07228

Page 1 of 1

Bard Peripheral Vascular, Inc.

TRADE SECRET/CONFIDENTIAL INFORMATION Notify CR Bard Before Releasing this Document.

§ 870.1250 Percutaneous catheter.

(a)
Identification. A percutaneous catheter is a device that is introduced into a vein or artery through the skin using a dilator and a sheath (introducer) or guide wire.(b)
Classification. Class II (performance standards).