K Number
K071210
Date Cleared
2007-12-19

(232 days)

Product Code
Regulation Number
866.5100
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The FIDIS™ CONNECTIVE 10* kit is a semi-quantitative homogeneous fluorescentbased microparticles immunoasay using flow cytometry. The test system is used to simultaneously detect the presence of 10 autoantibody specificities: double stranded DNA (dsDNA), SSA (60 kDa and 52 kDa), SSB, Sm, Sm/RNP, Scl70, Jo1, ribosome and centromere. (*Antibodies to dsDNA, Sm, Sm/RNP, SSA, SSB, Scl-70, Jo-1, ribosome and centromere can be reported using this assay).

Clinical utility: The results of the FIDIS™ CONNECTIVE 10* are to be used in conjunction with the clinical findings and the other laboratory tests to aid in the diagnosis of connective diseases (systemic lupus erythematosus (SLE), Sjogren's syndrome, mixed connective tissue disease (MCTD), scleroderma, dermatomyositis and CREST syndrome).

FIDIS™ CONNECTIVE 10* kit uses serum only, and is to be run on the FIDIS™ Instrument and MLX-BOOSTER™ Software.

FIDIS™ CONNECTIVE 10* kit may be used with the CARIS™ system (diluting and dispensing device).

This test is for in vitro diagnostic use.

Device Description

The FIDIS™ CONNECTIVE 10* kit is a semi-quantitative homogeneous fluorescentbased microparticles immunoasay using flow cytometry. The test system is used to simultaneously detect the presence of 10 autoantibody specificities: double stranded DNA (dsDNA), SSA (60 kDa and 52 kDa), SSB, Sm, Sm/RNP, Scl70, Jo1, ribosome and centromere. The kit includes a 96 wells microplate, vials of color-coded microsphere beads coupled with antigens and internal standard beads, sample dilution buffer, calibrator, positive control, negative control, anti-human IgG coupled to phycoerythrin, washing buffer, reconstitution buffer, package insert, microplate assay configuration worksheet, and microplate sealing films. The test is run on the FIDIS™ Instrument and MLX-BOOSTER™ Software, and may be used with the CARIS™ system (diluting and dispensing device).

AI/ML Overview

The Biomedical Diagnostics S.A. (bmd) FIDIS™ CONNECTIVE 10* assay is a semi-quantitative homogeneous fluorescent-based microparticles immunoassay that uses flow cytometry to detect 10 autoantibody specificities. The device's performance was established through analytical performance studies and comparison studies against a predicate device (K053653 FIDIS™ CONNECTIVE 10* system) and manual assay preparation steps.

1. Table of Acceptance Criteria and Reported Device Performance

The document provides acceptance criteria mainly for precision and outlines performance through agreement studies.

Precision Acceptance Criteria and Reported Performance

Sample rangeAcceptance criteria (CV%)Within-run minimal CV%Within-run maximal CV%Between-run minimal CV%Between-run maximal CV%
Less than 10 AU/mL or IU/mLNot determined5.9%12.8%7.4%13.6%
10 to 29 AU/mL or IU/mLCV ≤ 20%2.5%11.8%8.2%17.3%
29 to 800 AU/mL or IU/mLCV ≤ 15%2.1%12.7%4.5%14.3%

Comparison Study with Predicate Device (K053653 FIDIS™ CONNECTIVE 10)*

The acceptance criteria for the comparison study are implicitly that the agreement (Positive Percent Agreement, Negative Percent Agreement, and Overall Agreement) should demonstrate substantial equivalence to the predicate device. While specific percentage targets are not explicitly stated as "acceptance criteria" in the table format, the observed agreement percentages with 95% LCLs are reported and deemed sufficient for substantial equivalence.

Antigenic SpecificityPositive Percent Agreement (PPA)Negative Percent Agreement (NPA)Overall Agreement (OA)95% LCL PPA95% LCL NPA95% LCL OA
dsDNA91.84%98.51%95.69%82.29%93.11%91.15%
SSA 60 kDa100%98.57%99.15%93.95%93.40%96.04%
SSA 52 kDa95.08%96.97%96.06%87.78%90.77%91.90%
SSB100%97.14%97.85%87.79%91.28%93.39%
Sm95.83%97.33%96.97%81.71%91.84%92.35%
Sm/RNP96.88%97.44%97.27%86.02%92.15%93.10%
Scl7096.43%97.14%96.94%84.15%91.28%92.28%
Jo196.3%100%98.96%83.60%95.75%95.15%
Centromere83.33%100%95.83%65.82%95.92%90.72%
Ribosome100%100%100%84.67%95.75%96.62%

Comparison Study: Manual vs. Automated (CARIS™) Assay Preparation

Similar to the predicate comparison, the acceptance criteria here are also implicit that agreement between manual and automated methods shows substantial equivalence.

Antigenic SpecificityPositive Percent Agreement (PPA)Negative Percent Agreement (NPA)Overall Agreement (OA)95% LCL PPA95% LCL NPA95% LCL OA
dsDNA100%100%100%93.27%95.75%97.36%
SSA 60 kDa100%100%100%93.12%95.69%97.31%
SSA 52 kDa100%98.48%98.48%94.40%93.01%96.04%
SSB100%100%100%88.71%95.63%96.80%
Sm100%95.83%96.88%88.27%89.58%92.12%
Sm/RNP100%95.77%97.09%91.06%89.44%92.64%
Scl70100%100%100%90.19%95.69%96.96%
Jo1100%100%100%89.12%95.75%96.90%
Centromere100%94.52%95.70%86.09%87.90%90.43%
Ribosome100%100%100%68.77%95.69%96.13%

2. Sample Size Used for the Test Set and Data Provenance

  • Precision Study Test Set:

    • Within-run: 6 samples (5 for Jo1) tested 10 times in the same run.
    • Between-run: 6 samples (5 for Jo1) tested 3 times per run, across 6 runs.
    • The document does not specify the country of origin or whether the data was retrospective or prospective for the precision study samples.
  • Comparison Study with Predicate Device Test Set:

    • Sample Size: 264 samples.
    • Composition: 194 samples positive for one or more parameters, and 70 negative samples (including some with potential biological interferences).
    • Data Provenance: The document states "obtained from routine laboratory" for interfering substances study (part of analytical performance) and implies a similar source for the comparison study, but the specific country of origin or retrospective/prospective nature isn't explicitly stated. However, given the manufacturer is based in France, it's plausible the samples originated from Europe.
  • Comparison Study with Automated (CARIS™) system Test Set:

    • Sample Size: 264 samples.
    • Composition: 194 samples positive for one or more parameters, and 70 negative samples (including some with potential biological interferences).
    • Data Provenance: Same as above.

3. Number of Experts Used to Establish Ground Truth for the Test Set and Their Qualifications

The document does not mention the use of human experts to establish ground truth for the test sets. Instead, the "ground truth" for the comparison studies is defined by the results obtained from the predicate device (K053653 FIDIS™ CONNECTIVE 10 system)* or the manual assay preparation steps of the modified device. For the analytical performance (e.g., precision, interfering substances), the "truth" is based on the inherent characteristics of the spiked or characterized samples.


4. Adjudication Method for the Test Set

This type of in-vitro diagnostic device study typically doesn't involve human adjudication in the traditional sense (e.g., for imaging studies). For the comparison studies, discrepancies were handled as follows:

  • "All equivocal samples with predicate and modified CONNECTIVE 10* assays are considered negative for the comparison and the evaluation studies."
  • For individual specificities, the number of equivocal results is noted (e.g., "There were 3 equivocal results with the assay. For purposes of calculation, these results were considered as negative."). This method effectively converts equivocal results into negative results for agreement calculations.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, and Effect Size of AI vs. Without AI Assistance

No, a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not done. This device is an in-vitro diagnostic test system, not an AI-assisted diagnostic tool that human readers would interact with in the context of interpretation. The comparison is between different versions or methods of the assay itself, or against a predicate device.


6. If a Standalone (Algorithm Only Without Human-in-the-Loop Performance) Was Done

The device itself is a standalone in-vitro diagnostic test system. The performance studies evaluate the device's analytical characteristics and its agreement with a predicate device or different operational modes (manual vs. automated sample preparation). There isn't an "algorithm" in the sense of a standalone AI intended to perform diagnostic interpretation; rather, it's a test system that provides semi-quantitative results. The system operates without human interpretive intervention beyond running the assay and reading the numerical output.


7. The Type of Ground Truth Used

The ground truth for the comparison studies was:

  • The results generated by the predicate device (K053653 FIDIS™ CONNECTIVE 10 system)* when comparing the modified device to the predicate.
  • The results generated by the manual assay preparation steps when comparing the automated (CARIS™) assay preparation to manual preparation.

For the analytical precision and interfering substances studies, the ground truth was based on the known characteristics of the samples (e.g., characterized positive/negative samples, spiked samples, or samples with known interferences). There is no mention of pathology, expert consensus, or outcomes data being directly used as ground truth for these performance evaluations.


8. The Sample Size for the Training Set

The document does not specify a separate "training set" in the context of machine learning or algorithm development. The studies performed are for performance evaluation and comparison studies for a medical device. Therefore, no training set size is reported in the provided text.


9. How the Ground Truth for the Training Set Was Established

As no training set is indicated, this question is not applicable. The device's operational characteristics and comparisons are the focus of performance evaluation, not machine learning model training.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows a logo for Biomedical Diagnostics. The logo consists of the letters "bmd" in a stylized font, with a horizontal line underneath. Below the line, the words "biomedical diagnostics" are written in a smaller, sans-serif font. The letters "bmd" are in a bold, sans-serif font, with the "o" being a perfect circle.

Premarket Notification 510(k) Summary

DEC 1 9 2007

Assigned 510(k) Number: K071210

1. Submitted by :
Name:Biomedical Diagnostics S.A (bmd)
Contact Person:Christelle COURIVAUD
Regulatory Affairs Manager
Address:Actipole 25, 4-6 Bld de Beaubourg
77435 Marne-La-Vallée Cedex 2
FRANCE
Telephone:33 (0)1 64 62 10 12
Fax:33 (0)1 64 62 09 66
Establishment
Registration Number:3003935253
US Agent correspondent:
Hoppe Regulatory Consultants
Ms P. Ann HOPPE
2335 Massey Lane
Decatur GA 30033 USA
Phone: 404 248 0002
E-mail: Hoppe Regulatory@cs.com
2. Device Name
Trade/Proprietary Name :FIDIS™ CONNECTIVE 10* assay
Common/Usual Name :MX006 - FIDIS™ CONNECTIVE 10*: Detection test of 10autoantibody specificities: double stranded DNA (dsDNA),SSA 60kDa, SSA 52kDa, SSB, Sm, Sm/RNP, Scl70, Jo-1,Ribosome and Centromere.
Classification Name:Antinuclear antibody immunological test system
Trade/Proprietary Name :FIDIS™ Analyzer
Classification Name:Instrumentation for Chemical Multiplex Systems
Trade/Proprietary Name :CARIS™ System

·

Registered Office : Actipole 25 4-6 bd de Beaubourg 77435 Marne La Vallée cedex 2

{1}------------------------------------------------

3. Intended use of the device

The FIDIS™ CONNECTIVE 10* kit is a semi-quantitative homogeneous fluorescentbased microparticles immunoasay using flow cytometry. The test system is used to simultaneously detect the presence of 10 autoantibody specificities: double stranded DNA (dsDNA), SSA (60 kDa and 52 kDa), SSB, Sm, Sm/RNP, Scl70, Jo1, ribosome and centromere.

(*Antibodies to dsDNA, Sm, Sm/RNP, SSA, SSB, Scl-70, Jo-1, ribosome and centromere can be reported using this assay).

Clinical utility:

The results of the FIDIS™ CONNECTIVE 10* are to be used in conjunction with the clinical findings and the other laboratory tests to aid in the diagnosis of connective diseases (systemic lupus erythematosus (SLE), Sjogren's syndrome, mixed connective tissue disease (MCTD), scleroderma, dermatomyositis and CREST syndrome).

FIDIS™ CONNECTIVE 10* kit uses serum only, and is to be run on the FIDIS™ Instrument and MLX-BOOSTER™ Software.

FIDIS™ CONNECTIVE 10* kit may be used with the CARIS™ system (diluting and dispensing device).

This test is for in vitro diagnostic use.

4. Materials supplied

1 x 96 wells microplate with filtering membrane and a lid.1 plate
1 Vial (A) of 10 sets of color-coded microsphere beads coupled with dsDNA, SSA 60 kDa,SSA 52 kDa, SS-B, Sm, Sm/RNP, Scl-70, Jo-1, ribosomes, centromere antigen, plus 1 setof Internal standard beads.Lyophilized (to be diluted with the buffer named D)Sufficient quantity toobtain 6mL afterreconstitution
1 Vial (B) of sample dilution buffer (white vial)Ready to use2 x 115mL
1 Vial of calibrator titered for the specificities to be mesuredReady to useEach titer is printed on the vial label1 x 1,5mL
1 Vial of positive control concentrate. This control has a standard reactivity, that providesevidence of the proper functioning of reagents and correct assay performance.To be dilutedExpected values are printed on the vial label.1 × 250 µL
1 Vial of negative control* concentrateTo be diluted1 x 250µL

{2}------------------------------------------------

1 Vial of anti-human IgG coupled to phycoerythrinReady to use1 x 12mL
1 Vial (C) of washing buffer (black vial)Ready to use1 x 100mL
1 Vial (D) of reconstitution buffer for the microsphere setReady to use1 x 6mL
Package insert1
Microplate Assay Configuration Worksheet1
Microplate sealing films6

5. Predicate Device

510K NumberDevice Classification NameManufacturer Name
K053653FIDIST TM CONNECTIVE 10*bmd

6. Comparison with the predicate

Predicate DeviceFIDIST™ CONNECTIVE 10*K053653Modified DeviceFIDIST™ CONNECTIVE 10*
Intended useIndividual determination in human serum ofIgG antibodies against:dsDNA, SSA 60kDa, SSA 52kDa, SSB,Sm, Sm/RNP, Scl70, Jo-1, Ribosome andCentromereSame(minor text changes)
CUT-OFFNegative<30for the 10 specificities- In IU/mL fordsDNA- In AU/mL forthe 9 otherspecificitiesSame
Equivocal30-40for the 10 specificitiesSame
Positive>40for the 10 specificitiesSame
Material suppliedMicroplate with capsMicroplate with sealing films
BeadsVial of color-coded microsphere setready to use (6mL)Vial of color-coded microsphere setLyophilized (sq 6mL)
Sample dilutionPBS-Tween concentratedSample dilution buffer ready to use
Washing bufferPBS-Tween concentratedWashing buffer ready to use
Internal standardbeadsNoYes

{3}------------------------------------------------

Predicate DeviceFIDIS™ CONNECTIVE 10K053653Modified DeviceFIDIS™ CONNECTIVE 10
Assay configuration1 "reagent-blank" well1 "calibrator" well1 "negative control" well1 "positive control" well1 "reagent-blank" well1 "negative control" well1 "positive control" well2 "calibrator" wells
Diluted sample wellsSame
A second calibrator well every 48 well seriesNo
Incubation time2 x 30mn RTSame
Wash step2 x 200μL2 x 300μL
Assay protocolOptional final wash stepFinal wash step (not optional)
SoftwareBooster Version 1.35Booster Version 2.2
Detection MethodFluorescence(using Luminex 100)Fluorescence(using Luminex 200)
Sample preparationManual preparationSame
Automatic sample preparation (option)CARIS™Same

7. Performance Characteristics

1. Analytical performance

a. Precision

Precision of the assay was assessed in 53 samples. Precision was determined by calculating the within-run (intra-assay) and the between run (inter-assay).

· For within run: 6 samples (except for Jo1 only 5 samples were tested) 10 times in a same run.

· For between run: 6 samples (except for Jo1 only 5 samples were tested) in 6 runs, 3 times per run.

{4}------------------------------------------------

Sample rangeAcceptancecriteria forwithin-runandbetween-runWithin-runminimal CV%for the 10parametersWithin-runmaximalCV% for the10 parametersBetween-runminimal CV%for the 10parametersBetween-runmaximalCV% for the10 parameters
Less than 10AU/mL or IU/mLNotdetermined5.9%12.8%7.4%13.6%
10 to 29 AU/mLor IU/mLCV≤20%2.5%11.8%8.2%17.3%
29 to 800 AU/mLor IU/mLCV≤15%2.1%12.7%4.5%14.3%

Table 1: Summary of FIDIS™ CONNECTIVE 10* precision results

b. Linearity/ assay reportable range

FIDISTM CONNECTIVE 10* assay has been optimized to express the average binding capacity at the current dilution (1/200) by a flow cytometric reading resulting of the median fluorescence value obtained from 200 microspheres per parameter.

Further dilutions potentially give rise to inaccurate results because the reaction conditions and the equilibrium of the immunological reaction would be modified.

c. Interfering Substances

The study was conducted by testing 30 negative samples (for dsDNA, SSA 60kDa, SSA 52kDa, SSB, Sm, Sm/RNP, Scl70, Jo-1, Ribosome and Centromere) characterized as positive for various potential interferences obtained from routine laboratory (listed in following table).

Number of positive sample
dsDNASSA60 kDSSA52kDSSBSmSm/RNPScl70Jo1Ribo
Cryoglobulinemia N*=2000000000
Complement N*=7210111000
IgG monoclonalimmunoglobulins N*=1000000000
IgM monoclonalimmunoglobulins N*=5000000000
Rheumatoid factor N*=8122101000
Plasma N*=3000000000
Hemolyzed sera N*=3000000000
Anti-smooth muscleantibodies N*=1000000000

Table 2: Potential interferences results

*N: number of samples tested

{5}------------------------------------------------

d. Threshold values

Threshold values were estimated from the 2 selected populations:

  • 50 samples from blood donors -
  • 48 samples selected for their potential biological interferences -

The negative thresholds (30AU/mL or 30IU/mL) correspond to the 97.9% for dsDNA. SSA. Sm/RNP: 99.0% for centromere and ribosome, and 100% for SSB, Sm, Scl70 and Jo1 for the populations studied.

e. Stability of the assay results after final wash step

This assay included 3 test series with:

  • 5 positive samples for SSA 52kDa, SSA 60kDa, SSB, Sm, Sm/RNP, - Sc170, dsDNA and Centromere;
  • 4 positive samples for Jol;
  • -3 positive samples for Ribosome.

Each series of tests was washed and read after different times:

  • T= 0 hour: the first series of tests was read immediately after the final wash step.
  • ﺖ T= 4 hours: the second series of test was read after 4 hours of storage at room temperature away from direct sunlight.
  • T=18 hours: the last series was read after 18 hours of storage at room temperature away from direct sunlight.
%CV acceptance criteriaParameterSampleMean results Obtained after 18H (AU/mL)%CV Obtained after 18HMean results Obtained after 4H (AU/mL)%CV Obtained after 4H
%CV ≤ 15%SSA 52Sample 11699Not calculatedNot calculated
Sample 2855
Sample 3525
Sample 41305
Sample 518011
SSA 60Sample 6497Not calculatedNot calculated
Sample 79411
Sample 8684
Sample 9857
Sample 109513
Acceptance criteriaParameterSampleMean results Obtained after 18H (AU/ml)%CV Obtained after 18HMean results Obtained after 4H (AU/ml)%CV Obtained after 4H
%CV ≤15%SSBSample 111676Not calculatedNot calculated
Sample 12595
Sample 13857
Sample 1412111
SmSample 15235Not calculatedNot calculated
Sample 16337
Sample 1730910
Sample 18507
Sample 1910313
Sample 201473
Sm/RNPSample 212027Not calculatedNot calculated
Sample 228211
Sample 2337411
Sample 24599
%CV ≤15%Scl70Sample 2511613Not calculatedNot calculated
Sample 26574
Sample 27555
Sample 281443
Sample 293084
Jo1Sample 301887Not calculatedNot calculated
Sample 317913
Jo1Sample 322426Not calculatedNot calculated
Sample 33597
Sample 34776
Sample 359116
CentromereSample 3682117811
Sample 37817786
Sample 38229215
Sample 39195200
RibosomeSample 404013Not calculatedNot calculated
Sample 411417
Sample 42716
dsDNASample 434911Not calculatedNot calculated
Sample 441239
Sample 455110
Sample 461843
Sample 47628

Table 4: Stability of the assay results after final wash step

{6}------------------------------------------------

{7}------------------------------------------------

Based on the common laboratories practices, the time range recommended is "one hour for a plate when stored at room temperature away from direct sunlight".

2. Comparison study with predicate

bmd has compared the results obtained with modified FIDIS™ CONNECTIVE 10* versus the results obtained with predicate FIDISTM CONNECTIVE 10* K053653 in manual use.

The study was performed on 264 samples characterized with the predicate test and the result repartition is described as below:

  • 194 samples were positive for one or more parameters (see table 5). -
  • ー 70 negative samples including some samples evaluated for their potential biological interferences.
57
48
23
29
40
28
26
26
17
46

Table 5: Number of positive samples per parameter.

FIDIS™ All equivocal samples with predicate and modified CONNECTIVE 10* assays are considered negative for the comparison and the evaluation studies.

dsDNASSA 60kDa
N=116PosNegTotalN=118
Pos45146Pos
Neg46670Neg
Total4967116Total

Tables 6: Specificity performances

of calculation, these results were considered as negative. Positive percent agreement: 100% (48/48) Negative percent agreement: 98.57% (69/70) Overall agreement: 99.15% (117/118)

Pos

48

0

48

Neg

I

રેજે

70

There were 3 equivocal results with the assay. For purposes There were 2 equivocal results with the assay. For purposes of calculation, these results were considered as negative. Positive percent agreement: 91.84% (45/49) Negative percent agreement: 98.51% (66/67) Overall agreement: 95.69% (111/116)

  • 2017

Total

49

69

118

{8}------------------------------------------------

SSA 52kDa
N=127
PosNegTotal
Pos58260
Neg36467
Total6166127

There were 3 equivocal results with the assay. For purposes of calculation, these results were considered as negative.

Positive percent agreement: 95.08% (58/61) Negative percent agreement: 96.97% (64/66)

Overall agreement: 96.06% (122/127)

Sm
N=99PosNegTotal
Pos23225
Neg17374
Total247599

There were 6 equivocal results with the assay. For purposes of calculation, these results were considered as negative.

Positive percent agreement: 95.83% (23/24) Negative percent agreement: 97.33% (73/75) Overall agreement: 96.97% (96/99)

Scl70
N=98PosNegTotal
Pos27229
Neg16869
Total287098
SS-B
N=93
PosNegTotal
Pos23225
Neg06868
Total237093

There is 1 equivocal result with the assay. For purposes of calculation, these results were considered as negative. Positive percent agreement: 96.43% (27/28)

Negative percent agreement: 97.14% (68/70) Overall agreement: 96.94% (95/98)

There were I equivocal result with the assay. For purposes of calculation, these results were considered as negative.

Positive percent agreement: 100% (23/23) Negative percent agreement: 97.14% (68/70) Overall agreement: 97.85% (91/93)

Sm/RNP
N= 110
PosNegTotal
Pos31233
Neg17677
Total3278110

There were 7 equivocal results with the assay. For purposes of calculation, these results were considered as negative.

Positive percent agreement: 96.88% (31/32) Negative percent agreement: 97.44% (76/78) Overall agreement: 97.27% (107/110)

Jo1
N=96
PosNegTotal
Pos26026
Neg16970
Total276996

There is 0 equivocal result with the assay.

Positive percent agreement: 96.30% (26/27) Negative percent agreement: 100% (69/69) Overall agreement: 98.96% (95/96)

{9}------------------------------------------------

Centromere
N=96
PosNegTotal
Pos20020
Neg47276
Total247296
Ribosome
N=87PosNegTotal
Pos18018
Neg06969
Total186987

There were 7 equivocal results with the assay. For purposes of calculation, these results were considered as negative.

Positive percent agreement: 83.33% (20/24) Negative percent agreement: 100% (72/72)

Overall agreement:95.83% (92/96)

There is 1 equivocal result with the assay. For purposes of calculation, these results were considered as negative. Positive percent agreement: 100% (18/18) Negative percent agreement: 100% (69/69) Overall agreement: 100% (87/87)

AntigenicSpecificitySamplenumberPositivepercentagreementproportionNegativepercentagreementproportionOverallagreementproportion
dsDNA11691.84%98.51%95.69%
SSA 60 kDa118100%98.57%99.15%
SSA 52 kDa12795.08%96.97%96.06%
SSB93100%97.14%97.85%
Sm9995.83%97.33%96.97%
Sm/RNP11096.88%97.44%97.27%
Scl709896.43%97.14%96.94%
Jo19696.3%100%98.96%
Centromere9683.33%100%95.83%
Ribosome87100%100%100%

Table 7: Summary of performance agreement results

In addition to the analysis above, the 95% one-sided lower confidence limit in percent of proportion agreement (95% LCL (%) was calculated using the Exact Binomial Test for proportions to determine how low this proportion could be with a 95% confidence.

{10}------------------------------------------------

AntigenicSpecificityPositive percent agreementNegative percent agreementOverall percent agreement
N1R1P1(%)95%LCL (%)N2R2P2(%)95%LCL (%)NRP(%)95%LCL (%)
dsDNA494591.8482.29676698.5193.1111611195.6991.15
SSA 60 kDa484810093.95706998.5793.4011811799.1596.04
SSA 52 kDa615895.0887.78666496.9790.7712712296.0691.90
SSB232310087.79706897.1491.28939197.8593.39
Sm242395.8381.71757397.3391.84999696.9792.35
Sm/RNP323196.8886.02787697.4492.1511010797.2793.10
Scl70282796.4384.15706897.1491.28989596.9492.28
Jo1272696.383.60696910095.75969598.9695.15
Centromere242083.3365.82727210095.92969295.8390.72
Ribosome181810084.67696910095.75878710096.62

Table 8: Summary of agreements results - 95% LCL (%)

N1 = No. of positives; R1 = No. of positive agreements; P1 = R1/N1 N2 = No. of negatives; R2 = No. of negative agreements; P2 = R2/N2

N = N1 + N2; R = R1 + R2; P = R/N

All of results show that FIDIS™ CONNECTIVE 10* system can be considered substantially equivalent to the predicate K053653 FIDIS™ CONNECTIVE 10* system.

3. Comparison study with predicate a. Precision

Precision of the assay was assessed in 36 samples. Precision was determined by calculating the within-run (intra-assay) and the between run (inter-assay):

  • For within run: 4 samples 10 times in a same run. .
  • For between run: 4 samples in 6 runs, 3 times per run. .
Sample rangeAcceptancecriteria forwithin-runandbetween-runWithin-runminimal CV%for the 10parametersWithin-runmaximalCV% for the10 parametersBetween-runminimal CV%for the 10parametersBetween-runmaximalCV% for the10 parameters
Less than 10AU/mL or IU/mLNotdeterminedNot evaluatedNot evaluatedNot evaluatedNot evaluated
10 to 29 AU/mLor IU/mLCV≤20%3.8%10.3%7.3%13.9%
29 to 800 AU/mLor IU/mLCV≤15%1.7%10.8%3.7%12.5%
Table 9: Summary of CARISTM Precision results

{11}------------------------------------------------

b. Comparison study (manual versus automated assay preparation steps)

bmd has compared the results obtained with the modified FIDISTM CONNECTIVE 10* for the manual or automated (with CARIS™) assay preparation steps.

The study was performed on 264 samples characterized with the predicate test and the result repartition is described as below:

  • 194 samples were positive for one or more parameters (see Table 12) -
  • 70 negative samples including some samples evaluated for their potential biological interferences.
Pathological sample number/parameter
SSA5248
SSA6040
SSB22
Sm26
Sm/RNP33
Scl7027
JOI25
Centro23
Ribo6
dsDNA42

Table 10: Number of the pathological population per parameter.

All equivocal samples with FIDIS™ CONNECTIVE 10* assays are considered negative for the comparison and the evaluation studies.

dsDNAN=112PosNegTotal
Pos430
Neg06969
Total4369112

Tables 11: Agreement performances

SSA 60kDaN=110PosNegTotal
Pos42042
Neg06868
Total4268110

of calculation, these results were considered as negative. Positive percent agreement: 100% (43/43) Negative percent agreement: 100% (69/69) Overall agreement: 100% (112/112)

There were 4 equivocal results with the assay. For purposes There Were 1 equivocal results with the assay. For purposes of calculation, these results were considered as negative. Positive percent agreement: 100% (42/42) Negative percent agreement: 100% (68/68) Overall agreement: 100% (110/110)

{12}------------------------------------------------

PosNegTotal
dsDNAN=112Pos43043
Neg06969
Total4369112

There were 4 equivocal results with the assay. For purposes of calculation, these results were considered as negative.

Positive percent agreement: 100% (43/43) Negative percent agreement: 100% (69/69) Overall agreement: 100% (112/112)

Tables 11: Agreement performances

SSA 60kDa

N=110

Pos

Neg

08110
Positive percent agreement: 100% (42/42)Negative percent agreement: 100% (68/68)There were 1 equivocal results with the assay. For purposesof calculation, these results were considered as negative.

Pos

42

0

NUTTS

Total

42

68

Neg

0

68

Overall agreement: 100% (110/110)

SSA 52kDaN=118MANUAL
PosNegTotal
CARISTYPos52153
Neg06565
Total5266118

There were 2 equivocal results with the assay. For purposes of calculation, these results were considered as negative. Positive percent agreement: 100% (52/52)

Negative percent agreement: 98.48% (65/66)

Overall agreement: 99.15% (117/118)

SmMANUAL
N=96PosNegTotal
CARISTPos24327
Neg06969
Total247296

There were 5 equivocal results with the assay. For purposes of calculation, these results were considered as negative. Positive percent agreement: 100% (24/24) Negative percent agreement: 95.83% (69/72) Overall agreement: 96.88% (93/96)

Scl70
N=97PosNegTotal
CARISPos29029
Neg06868
Total296897

There is 0 equivocal result with the assay. Positive percent agreement: 100% (29/29) Negative percent agreement: 100% (68/68) Overall agreement: 100% (97/97)

SSBN=92MANUAL
PosNegTotal
Pos25025
CARISTNeg06767
Total256792

There is 1 equivocal result with the assay. For purposes of calculation, these results were considered as negative. Positive percent agreement: 100% (25/25) Negative percent agreement: 100% (67/67) Overall agreement: 100% (92/92)

Sm/RNP
N=103MANUAL
PosNegTotal
CARISPos32335
Neg06868
Total3271103

There were 6 borderline results with the assay. For purposes of calculation, these results were considered as negative. Positive percent agreement: 100% (32/32) Negative percent agreement: 95.77% (68/71) Overall agreement: 97.09% (100/103)

Jo1N=95MANUAL
PosNegTotal
CARISTPos26026
Neg06969
Total266995

There is 0 equivocal result with the assay. Positive percent agreement: 100% (26/26) Negative percent agreement: 100% (69/69) Overall agreement: 100% (95/95)

{13}------------------------------------------------

CentromereMANUAL
N=93PosNegTotal
Pos20424
Neg06969
Total207393
RibosomeMANUAL
N=76PosNegTotal
CARISPos808
Neg06868
Total86876

There were 6 equivocal results with the assay. For purpose There is 0 equivocal result with the assay. of calculation, these results were considered as negative.

Positive percent agreement: 100% (20/20)

Negative percent agreement: 94.52% (69/73)

Overall agreement: 95.70% (89/93)

e There is 0 equivocal result with the assay.

Positive percent agreement: 100% (8/8) Negative percent agreement: 100% (68/68) Overall agreement: 100% (76/76)

AntigenicSpecificitySamplenumberPositivepercentagreementNegativepercentagreementOverallagreement
proportionproportionproportion
dsDNA112100%100%100%
SSA 60 kDa110100%100%100%
SSA 52 kDa118100%98.48%98.48%
SSB92100%100%100%
Sm96100%95.83%95.83%
Sm/RNP13100%95.77%95.77%
Scl7097100%100%100%
Jo195100%100%100%
Centromere93100%94.52%94.52%
Ribosome76100%100%100%

Table 12: Summary of performance agreement results

{14}------------------------------------------------

In addition to the analysis above, the 95% one-sided lower confidence limit in percent of proportion agreement (95% LCL (%) was calculated using the Exact Binomial Test for proportions to determine how low this proportion could be with a 95% confidence.

AntigenicSpecificityPositive percent agreementNegative percent agreementOverall percent agreement
N₁R₁P₁(%)95% LCL (%)N₂R₂P₂(%)95% LCL (%)NRP(%)95% LCL (%)
dsDNA434310093.27696910095.7511211210097.36
SSA 60 kDa424210093.12686810095.6911011010097.31
SSA 52 kDa525210094.40666598.4893.0111811799.1596.04
SSB252510088.71676710095.63929210096.80
Sm242410088.27726995.8389.58969396.8892.12
Sm/RNP323210091.06716895.7789.4410310097.0992.64
Scl70292910090.19686810095.69979710096.96
Jo1262610089.12696910095.75959510096.90
Centromere202010086.09736994.5287.90938995.7090.43
Ribosome8810068.77686810095.69767610096.13

Table 13: Summary of performance agreements results - 95% LCL (%)

N1 = No. of positives; R1 = No. of positive agreements; P1 = R1/N1 N2 = No. of negatives; R2 = No. of negative agreements; P2 = R2/N2 N = N; + N2; R = R1 + R2; P = R/N

All previous evaluation results indicate that manual and automated (with CARIS™) assay preparation steps are considered substantially equivalents.

8. Conclusions

In conclusion, all supporting data demonstrate that the FIDIS™ CONNECTIVE 10* system can be considered substantially equivalent to the predicate device.

{15}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/15/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is circular and contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" around the perimeter. Inside the circle is an abstract symbol that resembles a stylized eagle or bird in flight.

Public Health Service

Food and Drug Administration 2098 Gaither Road Rockville MD 20850

DEC 1 9 2007

Biomedical Diagnostics S.A. (BMD) c/o Ms. Christelle Courivaud Regulatory Affairs Manager Actipole 25. 4-6 Bld de Beaubourg 77435 Marne La Vallée cedex 2 France

Re: K071210

Trade/Device Name: FIDISTM CONNECTIVE 10* Assay Regulation Number: 21 CFR 866.5100 Regulation Name: Antinuclear antibody immunological test system Regulatory Class: Class II Product Code: LLL, LKS, LKO, LKP, LSW, LJM, MQA Dated: December 5, 2007 Received: December 7, 2007

Dear Ms. Courivaud:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820). This letter will allow you to

{16}------------------------------------------------

Page 2 -

begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific information about the application of labeling requirements to your device, or questions on the promotion and advertising of your device, please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (240) 276-0450. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at 240-276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at 240-276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely yours,

Robert D. Becker

Robert L. Becker, Jr., M.D., Ph.Ø Director Division of Immunology and Hematology Devices Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health

Enclosure -

{17}------------------------------------------------

Indication for Use

510(k) Number (if known): K071210

Device Name:

FIDISTM CONNECTIVE 10*

Indication For Use:

The FIDIS™ CONNECTIVE 10* kit is a semi-quantitative homogeneous fluorescent-based microparticles immunoassay using flow cytometry. The test system is used to simulaneously detect the presence of 10 autoantibody specificities: double stranded DNA (dsDNAA), SSA (60 kDa and 52 kDa), SSB, Sm, Sm/RNP, Scl70, Jo1, ribosome and centromere,

(*Antibodies to dsDNA, Sm, Sm/RNP, SSA, SSB, Scl-70, Jo-1, ribosome and centromere can be reported using this assay).

Clinical utility:

The results of the FIDIS™ CONNECTIVE 10* are to be used in conjunction with the clinical findings and the other laboratory tests to aid in the diagnosis of connective diseases (systemic lupus erythematosus (SLE), Sjogren's syndrome, mixed connective disease (MCTD), scleroderma, dermatomyositis and CREST syndrome).

FIDIS™ CONNECTIVE 10* kit uses serum only, and is to be run on the FIDIS™ Instrument and MLX-BOOSTER™ Software.

FIDIS™ CONNECTIVE 10* kit may be used with the CARIS™ system (diluting and dispensing device).

This test is for in vitro diagnostic use.

Prescription Use __ X (21 CFR Part 801 Subpart D)

And/Or

Over the Counter Use (21 CFR Part 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE; CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of In Vitro Diagnostic Device Evaluation and Safety (OIVD)

Maria M Elan

Division Sign-Off Office of In Vitro Diagnostic Device Evaluation and Safety

§ 866.5100 Antinuclear antibody immunological test system.

(a)
Identification. An antinuclear antibody immunological test system is a device that consists of the reagents used to measure by immunochemical techniques the autoimmune antibodies in serum, other body fluids, and tissues that react with cellular nuclear constituents (molecules present in the nucleus of a cell, such as ribonucleic acid, deoxyribonucleic acid, or nuclear proteins). The measurements aid in the diagnosis of systemic lupus erythematosus (a multisystem autoimmune disease in which antibodies attack the victim's own tissues), hepatitis (a liver disease), rheumatoid arthritis, Sjögren's syndrome (arthritis with inflammation of the eye, eyelid, and salivary glands), and systemic sclerosis (chronic hardening and shrinking of many body tissues).(b)
Classification. Class II (performance standards).