K Number
K071210
Device Name
MODIFICATION TO FIDIS CONNECTIVE 10*, MODEL MX006
Date Cleared
2007-12-19

(232 days)

Regulation Number
866.5100
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP Authorized
Intended Use
The FIDIS™ CONNECTIVE 10* kit is a semi-quantitative homogeneous fluorescentbased microparticles immunoasay using flow cytometry. The test system is used to simultaneously detect the presence of 10 autoantibody specificities: double stranded DNA (dsDNA), SSA (60 kDa and 52 kDa), SSB, Sm, Sm/RNP, Scl70, Jo1, ribosome and centromere. (*Antibodies to dsDNA, Sm, Sm/RNP, SSA, SSB, Scl-70, Jo-1, ribosome and centromere can be reported using this assay). Clinical utility: The results of the FIDIS™ CONNECTIVE 10* are to be used in conjunction with the clinical findings and the other laboratory tests to aid in the diagnosis of connective diseases (systemic lupus erythematosus (SLE), Sjogren's syndrome, mixed connective tissue disease (MCTD), scleroderma, dermatomyositis and CREST syndrome). FIDIS™ CONNECTIVE 10* kit uses serum only, and is to be run on the FIDIS™ Instrument and MLX-BOOSTER™ Software. FIDIS™ CONNECTIVE 10* kit may be used with the CARIS™ system (diluting and dispensing device). This test is for in vitro diagnostic use.
Device Description
The FIDIS™ CONNECTIVE 10* kit is a semi-quantitative homogeneous fluorescentbased microparticles immunoasay using flow cytometry. The test system is used to simultaneously detect the presence of 10 autoantibody specificities: double stranded DNA (dsDNA), SSA (60 kDa and 52 kDa), SSB, Sm, Sm/RNP, Scl70, Jo1, ribosome and centromere. The kit includes a 96 wells microplate, vials of color-coded microsphere beads coupled with antigens and internal standard beads, sample dilution buffer, calibrator, positive control, negative control, anti-human IgG coupled to phycoerythrin, washing buffer, reconstitution buffer, package insert, microplate assay configuration worksheet, and microplate sealing films. The test is run on the FIDIS™ Instrument and MLX-BOOSTER™ Software, and may be used with the CARIS™ system (diluting and dispensing device).
More Information

Not Found

No
The summary describes a standard immunoassay using flow cytometry and mentions software for running the test, but there is no indication of AI or ML being used for data analysis, interpretation, or any other function. The performance studies focus on traditional analytical and comparison metrics.

No
Explanation: The device is explicitly stated to be for "in vitro diagnostic use" and its clinical utility is to "aid in the diagnosis of connective diseases," not to treat or provide therapy.

Yes

The device's 'Intended Use / Indications for Use' explicitly states: "The results of the FIDIS™ CONNECTIVE 10* are to be used in conjunction with the clinical findings and the other laboratory tests to aid in the diagnosis of connective diseases" and "This test is for in vitro diagnostic use." This clearly indicates its role in aiding diagnosis.

No

The device description clearly states that the kit includes physical components such as a microplate, vials of beads, buffers, controls, and other reagents. While software (MLX-BOOSTER™ Software) is mentioned as part of the system, the core of the device is a physical immunoassay kit.

Yes, this device is an IVD (In Vitro Diagnostic).

Here's why:

  • Explicit Statement: The "Intended Use / Indications for Use" section explicitly states: "This test is for in vitro diagnostic use."
  • Nature of the Test: The device is a "semi-quantitative homogeneous fluorescentbased microparticles immunoasay using flow cytometry" used to detect autoantibodies in serum. This is a classic example of an in vitro diagnostic test, as it analyzes biological samples (serum) outside of the body to provide information for diagnosis.
  • Clinical Utility: The stated clinical utility is to "aid in the diagnosis of connective diseases." This directly aligns with the purpose of IVD devices, which are used to diagnose, monitor, or treat diseases or conditions.
  • Intended User/Care Setting: The "Intended User / Care Setting" is listed as "in vitro diagnostic use," further confirming its classification.

N/A

Intended Use / Indications for Use

The FIDIS™ CONNECTIVE 10* kit is a semi-quantitative homogeneous fluorescentbased microparticles immunoasay using flow cytometry. The test system is used to simultaneously detect the presence of 10 autoantibody specificities: double stranded DNA (dsDNA), SSA (60 kDa and 52 kDa), SSB, Sm, Sm/RNP, Scl70, Jo1, ribosome and centromere.

(*Antibodies to dsDNA, Sm, Sm/RNP, SSA, SSB, Scl-70, Jo-1, ribosome and centromere can be reported using this assay).

Clinical utility:
The results of the FIDIS™ CONNECTIVE 10* are to be used in conjunction with the clinical findings and the other laboratory tests to aid in the diagnosis of connective diseases (systemic lupus erythematosus (SLE), Sjogren's syndrome, mixed connective tissue disease (MCTD), scleroderma, dermatomyositis and CREST syndrome).

FIDIS™ CONNECTIVE 10* kit uses serum only, and is to be run on the FIDIS™ Instrument and MLX-BOOSTER™ Software.

FIDIS™ CONNECTIVE 10* kit may be used with the CARIS™ system (diluting and dispensing device).

This test is for in vitro diagnostic use.

Product codes (comma separated list FDA assigned to the subject device)

LLL, LKS, LKO, LKP, LSW, LJM, MQA

Device Description

The FIDIS™ CONNECTIVE 10* kit is a semi-quantitative homogeneous fluorescent-based microparticles immunoassay using flow cytometry. The test system is used to simultaneously detect the presence of 10 autoantibody specificities: double stranded DNA (dsDNA), SSA (60kDa and 52kDa), SSB, Sm, Sm/RNP, Scl70, Jo-1, ribosome and centromere. The kit includes a 96-well microplate, vials of color-coded microsphere beads coupled with various antigens and internal standard beads, sample dilution buffer, calibrator, positive control, negative control, anti-human IgG coupled to phycoerythrin, washing buffer, and reconstitution buffer for the microsphere set. It is designed to be run on the FIDIS™ Instrument and MLX-BOOSTER™ Software, and optionally with the CARIS™ system (diluting and dispensing device).

Mentions image processing

Not Found

Mentions AI, DNN, or ML

Not Found

Input Imaging Modality

Not Found

Anatomical Site

Not Found

Indicated Patient Age Range

Not Found

Intended User / Care Setting

Not Found

Description of the training set, sample size, data source, and annotation protocol

Not Found

Description of the test set, sample size, data source, and annotation protocol

Comparison with predicate:
The study was performed on 264 samples characterized with the predicate test, with 194 samples positive for one or more parameters and 70 negative samples (including some with potential biological interferences). All equivocal samples with predicate and modified CONNECTIVE 10* assays were considered negative for comparison and evaluation.

Comparison (manual versus automated assay preparation):
The study was performed on 264 samples characterized with the predicate test, with 194 samples positive for one or more parameters and 70 negative samples (including some with potential biological interferences). All equivocal samples with FIDIS™ CONNECTIVE 10* assays were considered negative for the comparison and evaluation studies.

Summary of Performance Studies (study type, sample size, AUC, MRMC, standalone performance, key results)

  1. Analytical performance:
    a. Precision: Assessed in 53 samples.
    - Within-run (intra-assay): 6 samples (5 for Jo1) tested 10 times in a same run.
    - Between-run (inter-assay): 6 samples (5 for Jo1) tested in 6 runs, 3 times per run.
    - Key results: CVs were within acceptance criteria (CV

§ 866.5100 Antinuclear antibody immunological test system.

(a)
Identification. An antinuclear antibody immunological test system is a device that consists of the reagents used to measure by immunochemical techniques the autoimmune antibodies in serum, other body fluids, and tissues that react with cellular nuclear constituents (molecules present in the nucleus of a cell, such as ribonucleic acid, deoxyribonucleic acid, or nuclear proteins). The measurements aid in the diagnosis of systemic lupus erythematosus (a multisystem autoimmune disease in which antibodies attack the victim's own tissues), hepatitis (a liver disease), rheumatoid arthritis, Sjögren's syndrome (arthritis with inflammation of the eye, eyelid, and salivary glands), and systemic sclerosis (chronic hardening and shrinking of many body tissues).(b)
Classification. Class II (performance standards).

0

Image /page/0/Picture/0 description: The image shows a logo for Biomedical Diagnostics. The logo consists of the letters "bmd" in a stylized font, with a horizontal line underneath. Below the line, the words "biomedical diagnostics" are written in a smaller, sans-serif font. The letters "bmd" are in a bold, sans-serif font, with the "o" being a perfect circle.

Premarket Notification 510(k) Summary

DEC 1 9 2007

Assigned 510(k) Number: K071210

1. Submitted by :
Name:Biomedical Diagnostics S.A (bmd)
Contact Person:Christelle COURIVAUD
Regulatory Affairs Manager
Address:Actipole 25, 4-6 Bld de Beaubourg
77435 Marne-La-Vallée Cedex 2
FRANCE
Telephone:33 (0)1 64 62 10 12
Fax:33 (0)1 64 62 09 66
Establishment
Registration Number:3003935253
US Agent correspondent:
Hoppe Regulatory Consultants
Ms P. Ann HOPPE
2335 Massey Lane
Decatur GA 30033 USA
Phone: 404 248 0002
E-mail: Hoppe Regulatory@cs.com
2. Device Name
Trade/Proprietary Name :FIDIS™ CONNECTIVE 10* assay
Common/Usual Name :MX006 - FIDIS™ CONNECTIVE 10*: Detection test of 10
autoantibody specificities: double stranded DNA (dsDNA),
SSA 60kDa, SSA 52kDa, SSB, Sm, Sm/RNP, Scl70, Jo-1,
Ribosome and Centromere.
Classification Name:Antinuclear antibody immunological test system
Trade/Proprietary Name :FIDIS™ Analyzer
Classification Name:Instrumentation for Chemical Multiplex Systems
Trade/Proprietary Name :CARIS™ System

·

Registered Office : Actipole 25 4-6 bd de Beaubourg 77435 Marne La Vallée cedex 2

1

3. Intended use of the device

The FIDIS™ CONNECTIVE 10* kit is a semi-quantitative homogeneous fluorescentbased microparticles immunoasay using flow cytometry. The test system is used to simultaneously detect the presence of 10 autoantibody specificities: double stranded DNA (dsDNA), SSA (60 kDa and 52 kDa), SSB, Sm, Sm/RNP, Scl70, Jo1, ribosome and centromere.

(*Antibodies to dsDNA, Sm, Sm/RNP, SSA, SSB, Scl-70, Jo-1, ribosome and centromere can be reported using this assay).

Clinical utility:

The results of the FIDIS™ CONNECTIVE 10* are to be used in conjunction with the clinical findings and the other laboratory tests to aid in the diagnosis of connective diseases (systemic lupus erythematosus (SLE), Sjogren's syndrome, mixed connective tissue disease (MCTD), scleroderma, dermatomyositis and CREST syndrome).

FIDIS™ CONNECTIVE 10* kit uses serum only, and is to be run on the FIDIS™ Instrument and MLX-BOOSTER™ Software.

FIDIS™ CONNECTIVE 10* kit may be used with the CARIS™ system (diluting and dispensing device).

This test is for in vitro diagnostic use.

4. Materials supplied

1 x 96 wells microplate with filtering membrane and a lid.1 plate
1 Vial (A) of 10 sets of color-coded microsphere beads coupled with dsDNA, SSA 60 kDa,
SSA 52 kDa, SS-B, Sm, Sm/RNP, Scl-70, Jo-1, ribosomes, centromere antigen, plus 1 set
of Internal standard beads.
Lyophilized (to be diluted with the buffer named D)Sufficient quantity to
obtain 6mL after
reconstitution
1 Vial (B) of sample dilution buffer (white vial)
Ready to use2 x 115mL
1 Vial of calibrator titered for the specificities to be mesured
Ready to use
Each titer is printed on the vial label1 x 1,5mL
1 Vial of positive control concentrate. This control has a standard reactivity, that provides
evidence of the proper functioning of reagents and correct assay performance.
To be diluted
Expected values are printed on the vial label.1 × 250 µL
1 Vial of negative control* concentrate
To be diluted1 x 250µL

2

| 1 Vial of anti-human IgG coupled to phycoerythrin

Ready to use1 x 12mL
1 Vial (C) of washing buffer (black vial)
Ready to use1 x 100mL
1 Vial (D) of reconstitution buffer for the microsphere set
Ready to use1 x 6mL
Package insert1
Microplate Assay Configuration Worksheet1
Microplate sealing films6

5. Predicate Device

510K NumberDevice Classification NameManufacturer Name
K053653FIDIST TM CONNECTIVE 10*bmd

6. Comparison with the predicate

| | | Predicate Device
FIDIST™ CONNECTIVE 10*
K053653 | | Modified Device
FIDIST™ CONNECTIVE 10* |
|-------------------|----------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------|-------------------------------------------------------------|
| Intended use | | Individual determination in human serum of
IgG antibodies against:
dsDNA, SSA 60kDa, SSA 52kDa, SSB,
Sm, Sm/RNP, Scl70, Jo-1, Ribosome and
Centromere | | Same
(minor text changes) |
| CUT-OFF | Negative | 40
for the 10 specificities | | Same |
| Material supplied | | Microplate with caps | | Microplate with sealing films |
| | Beads | Vial of color-coded microsphere set
ready to use (6mL) | | Vial of color-coded microsphere set
Lyophilized (sq 6mL) |
| | Sample dilution | PBS-Tween concentrated | | Sample dilution buffer ready to use |
| Washing buffer | | PBS-Tween concentrated | | Washing buffer ready to use |
| | Internal standard
beads | No | | Yes |

3

| | Predicate Device
FIDIS™ CONNECTIVE 10
K053653 | Modified Device
FIDIS™ CONNECTIVE 10 |
|---------------------------------------|---------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------|
| Assay configuration | 1 "reagent-blank" well
1 "calibrator" well
1 "negative control" well
1 "positive control" well | 1 "reagent-blank" well
1 "negative control" well
1 "positive control" well
2 "calibrator" wells |
| | Diluted sample wells | Same |
| | A second calibrator well every 48 well series | No |
| Incubation time | 2 x 30mn RT | Same |
| Wash step | 2 x 200μL | 2 x 300μL |
| Assay protocol | Optional final wash step | Final wash step (not optional) |
| Software | Booster Version 1.35 | Booster Version 2.2 |
| Detection Method | Fluorescence
(using Luminex 100) | Fluorescence
(using Luminex 200) |
| Sample preparation | Manual preparation | Same |
| Automatic sample preparation (option) | CARIS™ | Same |

7. Performance Characteristics

1. Analytical performance

a. Precision

Precision of the assay was assessed in 53 samples. Precision was determined by calculating the within-run (intra-assay) and the between run (inter-assay).

· For within run: 6 samples (except for Jo1 only 5 samples were tested) 10 times in a same run.

· For between run: 6 samples (except for Jo1 only 5 samples were tested) in 6 runs, 3 times per run.

4

| Sample range | Acceptance
criteria for
within-run
and
between-
run | Within-run
minimal CV%
for the 10
parameters | Within-run
maximal
CV% for the
10 parameters | Between-run
minimal CV%
for the 10
parameters | Between-run
maximal
CV% for the
10 parameters |
|--------------------------------|--------------------------------------------------------------------|-------------------------------------------------------|-------------------------------------------------------|--------------------------------------------------------|--------------------------------------------------------|
| Less than 10
AU/mL or IU/mL | Not
determined | 5.9% | 12.8% | 7.4% | 13.6% |
| 10 to 29 AU/mL
or IU/mL | CV≤20% | 2.5% | 11.8% | 8.2% | 17.3% |
| 29 to 800 AU/mL
or IU/mL | CV≤15% | 2.1% | 12.7% | 4.5% | 14.3% |

Table 1: Summary of FIDIS™ CONNECTIVE 10* precision results

b. Linearity/ assay reportable range

FIDISTM CONNECTIVE 10* assay has been optimized to express the average binding capacity at the current dilution (1/200) by a flow cytometric reading resulting of the median fluorescence value obtained from 200 microspheres per parameter.

Further dilutions potentially give rise to inaccurate results because the reaction conditions and the equilibrium of the immunological reaction would be modified.

c. Interfering Substances

The study was conducted by testing 30 negative samples (for dsDNA, SSA 60kDa, SSA 52kDa, SSB, Sm, Sm/RNP, Scl70, Jo-1, Ribosome and Centromere) characterized as positive for various potential interferences obtained from routine laboratory (listed in following table).

Number of positive sample
dsDNASSA
60 kDSSA
52kDSSBSmSm/RNPScl70Jo1Ribo
Cryoglobulinemia N*=2000000000
Complement N*=7210111000
IgG monoclonal
immunoglobulins N*=1000000000
IgM monoclonal
immunoglobulins N*=5000000000
Rheumatoid factor N*=8122101000
Plasma N*=3000000000
Hemolyzed sera N*=3000000000
Anti-smooth muscle
antibodies N*=1000000000

Table 2: Potential interferences results

*N: number of samples tested

5

d. Threshold values

Threshold values were estimated from the 2 selected populations:

  • 50 samples from blood donors -
  • 48 samples selected for their potential biological interferences -

The negative thresholds (30AU/mL or 30IU/mL) correspond to the 97.9% for dsDNA. SSA. Sm/RNP: 99.0% for centromere and ribosome, and 100% for SSB, Sm, Scl70 and Jo1 for the populations studied.

e. Stability of the assay results after final wash step

This assay included 3 test series with:

  • 5 positive samples for SSA 52kDa, SSA 60kDa, SSB, Sm, Sm/RNP, - Sc170, dsDNA and Centromere;
  • 4 positive samples for Jol;
  • -3 positive samples for Ribosome.

Each series of tests was washed and read after different times:

  • T= 0 hour: the first series of tests was read immediately after the final wash step.
  • ﺖ T= 4 hours: the second series of test was read after 4 hours of storage at room temperature away from direct sunlight.
  • T=18 hours: the last series was read after 18 hours of storage at room temperature away from direct sunlight.
%CV acceptance criteriaParameterSampleMean results Obtained after 18H (AU/mL)%CV Obtained after 18HMean results Obtained after 4H (AU/mL)%CV Obtained after 4H
%CV ≤ 15%SSA 52Sample 11699Not calculatedNot calculated
Sample 2855
Sample 3525
Sample 41305
Sample 518011
SSA 60Sample 6497Not calculatedNot calculated
Sample 79411
Sample 8684
Sample 9857
Sample 109513
Acceptance criteriaParameterSampleMean results Obtained after 18H (AU/ml)%CV Obtained after 18HMean results Obtained after 4H (AU/ml)%CV Obtained after 4H
%CV ≤
15%SSBSample 111676Not calculatedNot calculated
Sample 12595
Sample 13857
Sample 1412111
SmSample 15235Not calculatedNot calculated
Sample 16337
Sample 1730910
Sample 18507
Sample 1910313
Sample 201473
Sm/RNPSample 212027Not calculatedNot calculated
Sample 228211
Sample 2337411
Sample 24599
%CV ≤
15%Scl70Sample 2511613Not calculatedNot calculated
Sample 26574
Sample 27555
Sample 281443
Sample 293084
Jo1Sample 301887Not calculatedNot calculated
Sample 317913
Jo1Sample 322426Not calculatedNot calculated
Sample 33597
Sample 34776
Sample 359116
CentromereSample 3682117811
Sample 37817786
Sample 38229215
Sample 39195200
RibosomeSample 404013Not calculatedNot calculated
Sample 411417
Sample 42716
dsDNASample 434911Not calculatedNot calculated
Sample 441239
Sample 455110
Sample 461843
Sample 47628

Table 4: Stability of the assay results after final wash step

6

7

Based on the common laboratories practices, the time range recommended is "one hour for a plate when stored at room temperature away from direct sunlight".

2. Comparison study with predicate

bmd has compared the results obtained with modified FIDIS™ CONNECTIVE 10* versus the results obtained with predicate FIDISTM CONNECTIVE 10* K053653 in manual use.

The study was performed on 264 samples characterized with the predicate test and the result repartition is described as below:

  • 194 samples were positive for one or more parameters (see table 5). -
  • ー 70 negative samples including some samples evaluated for their potential biological interferences.
57
48
23
29
40
28
26
26
17
46

Table 5: Number of positive samples per parameter.

FIDIS™ All equivocal samples with predicate and modified CONNECTIVE 10* assays are considered negative for the comparison and the evaluation studies.

dsDNASSA 60kDa
N=116PosNegTotalN=118
Pos45146Pos
Neg46670Neg
Total4967116Total

Tables 6: Specificity performances

of calculation, these results were considered as negative. Positive percent agreement: 100% (48/48) Negative percent agreement: 98.57% (69/70) Overall agreement: 99.15% (117/118)

Pos

48

0

48

Neg

I

રેજે

70

There were 3 equivocal results with the assay. For purposes There were 2 equivocal results with the assay. For purposes of calculation, these results were considered as negative. Positive percent agreement: 91.84% (45/49) Negative percent agreement: 98.51% (66/67) Overall agreement: 95.69% (111/116)

  • 2017

Total

49

69

118

8

SSA 52kDa
N=127
PosNegTotal
Pos58260
Neg36467
Total6166127

There were 3 equivocal results with the assay. For purposes of calculation, these results were considered as negative.

Positive percent agreement: 95.08% (58/61) Negative percent agreement: 96.97% (64/66)

Overall agreement: 96.06% (122/127)

Sm
N=99PosNegTotal
Pos23225
Neg17374
Total247599

There were 6 equivocal results with the assay. For purposes of calculation, these results were considered as negative.

Positive percent agreement: 95.83% (23/24) Negative percent agreement: 97.33% (73/75) Overall agreement: 96.97% (96/99)

Scl70
N=98PosNegTotal
Pos27229
Neg16869
Total287098
SS-B
N=93
PosNegTotal
Pos23225
Neg06868
Total237093

There is 1 equivocal result with the assay. For purposes of calculation, these results were considered as negative. Positive percent agreement: 96.43% (27/28)

Negative percent agreement: 97.14% (68/70) Overall agreement: 96.94% (95/98)

There were I equivocal result with the assay. For purposes of calculation, these results were considered as negative.

Positive percent agreement: 100% (23/23) Negative percent agreement: 97.14% (68/70) Overall agreement: 97.85% (91/93)

Sm/RNP
N= 110
PosNegTotal
Pos31233
Neg17677
Total3278110

There were 7 equivocal results with the assay. For purposes of calculation, these results were considered as negative.

Positive percent agreement: 96.88% (31/32) Negative percent agreement: 97.44% (76/78) Overall agreement: 97.27% (107/110)

Jo1
N=96
PosNegTotal
Pos26026
Neg16970
Total276996

There is 0 equivocal result with the assay.

Positive percent agreement: 96.30% (26/27) Negative percent agreement: 100% (69/69) Overall agreement: 98.96% (95/96)

9

Centromere
N=96
PosNegTotal
Pos20020
Neg47276
Total247296
Ribosome
N=87PosNegTotal
Pos18018
Neg06969
Total186987

There were 7 equivocal results with the assay. For purposes of calculation, these results were considered as negative.

Positive percent agreement: 83.33% (20/24) Negative percent agreement: 100% (72/72)

Overall agreement:95.83% (92/96)

There is 1 equivocal result with the assay. For purposes of calculation, these results were considered as negative. Positive percent agreement: 100% (18/18) Negative percent agreement: 100% (69/69) Overall agreement: 100% (87/87)

| Antigenic
Specificity | Sample
number | Positive
percent
agreement
proportion | Negative
percent
agreement
proportion | Overall
agreement
proportion |
|--------------------------|------------------|------------------------------------------------|------------------------------------------------|------------------------------------|
| dsDNA | 116 | 91.84% | 98.51% | 95.69% |
| SSA 60 kDa | 118 | 100% | 98.57% | 99.15% |
| SSA 52 kDa | 127 | 95.08% | 96.97% | 96.06% |
| SSB | 93 | 100% | 97.14% | 97.85% |
| Sm | 99 | 95.83% | 97.33% | 96.97% |
| Sm/RNP | 110 | 96.88% | 97.44% | 97.27% |
| Scl70 | 98 | 96.43% | 97.14% | 96.94% |
| Jo1 | 96 | 96.3% | 100% | 98.96% |
| Centromere | 96 | 83.33% | 100% | 95.83% |
| Ribosome | 87 | 100% | 100% | 100% |

Table 7: Summary of performance agreement results

In addition to the analysis above, the 95% one-sided lower confidence limit in percent of proportion agreement (95% LCL (%) was calculated using the Exact Binomial Test for proportions to determine how low this proportion could be with a 95% confidence.

10

| Antigenic

SpecificityPositive percent agreementNegative percent agreementOverall percent agreement
N1R1P1
(%)95%
LCL (%)N2R2P2
(%)95%
LCL (%)NRP
(%)95%
LCL (%)
dsDNA494591.8482.29676698.5193.1111611195.6991.15
SSA 60 kDa484810093.95706998.5793.4011811799.1596.04
SSA 52 kDa615895.0887.78666496.9790.7712712296.0691.90
SSB232310087.79706897.1491.28939197.8593.39
Sm242395.8381.71757397.3391.84999696.9792.35
Sm/RNP323196.8886.02787697.4492.1511010797.2793.10
Scl70282796.4384.15706897.1491.28989596.9492.28
Jo1272696.383.60696910095.75969598.9695.15
Centromere242083.3365.82727210095.92969295.8390.72
Ribosome181810084.67696910095.75878710096.62

Table 8: Summary of agreements results - 95% LCL (%)

N1 = No. of positives; R1 = No. of positive agreements; P1 = R1/N1 N2 = No. of negatives; R2 = No. of negative agreements; P2 = R2/N2

N = N1 + N2; R = R1 + R2; P = R/N

All of results show that FIDIS™ CONNECTIVE 10* system can be considered substantially equivalent to the predicate K053653 FIDIS™ CONNECTIVE 10* system.

3. Comparison study with predicate a. Precision

Precision of the assay was assessed in 36 samples. Precision was determined by calculating the within-run (intra-assay) and the between run (inter-assay):

  • For within run: 4 samples 10 times in a same run. .
  • For between run: 4 samples in 6 runs, 3 times per run. .

| Sample range | Acceptance
criteria for
within-run
and
between-
run | Within-run
minimal CV%
for the 10
parameters | Within-run
maximal
CV% for the
10 parameters | Between-run
minimal CV%
for the 10
parameters | Between-run
maximal
CV% for the
10 parameters |
|--------------------------------|--------------------------------------------------------------------|-------------------------------------------------------|-------------------------------------------------------|--------------------------------------------------------|--------------------------------------------------------|
| Less than 10
AU/mL or IU/mL | Not
determined | Not evaluated | Not evaluated | Not evaluated | Not evaluated |
| 10 to 29 AU/mL
or IU/mL | CV≤20% | 3.8% | 10.3% | 7.3% | 13.9% |
| 29 to 800 AU/mL
or IU/mL | CV≤15% | 1.7% | 10.8% | 3.7% | 12.5% |

Table 9: Summary of CARISTM Precision results

11

b. Comparison study (manual versus automated assay preparation steps)

bmd has compared the results obtained with the modified FIDISTM CONNECTIVE 10* for the manual or automated (with CARIS™) assay preparation steps.

The study was performed on 264 samples characterized with the predicate test and the result repartition is described as below:

  • 194 samples were positive for one or more parameters (see Table 12) -
  • 70 negative samples including some samples evaluated for their potential biological interferences.
Pathological sample number/parameter
SSA5248
SSA6040
SSB22
Sm26
Sm/RNP33
Scl7027
JOI25
Centro23
Ribo6
dsDNA42

Table 10: Number of the pathological population per parameter.

All equivocal samples with FIDIS™ CONNECTIVE 10* assays are considered negative for the comparison and the evaluation studies.

| dsDNA

N=112PosNegTotal
Pos430
Neg06969
Total4369112

Tables 11: Agreement performances

| SSA 60kDa

N=110PosNegTotal
Pos42042
Neg06868
Total4268110

of calculation, these results were considered as negative. Positive percent agreement: 100% (43/43) Negative percent agreement: 100% (69/69) Overall agreement: 100% (112/112)

There were 4 equivocal results with the assay. For purposes There Were 1 equivocal results with the assay. For purposes of calculation, these results were considered as negative. Positive percent agreement: 100% (42/42) Negative percent agreement: 100% (68/68) Overall agreement: 100% (110/110)

12

PosNegTotal
dsDNA
N=112Pos43043
Neg06969
Total4369112

There were 4 equivocal results with the assay. For purposes of calculation, these results were considered as negative.

Positive percent agreement: 100% (43/43) Negative percent agreement: 100% (69/69) Overall agreement: 100% (112/112)

Tables 11: Agreement performances

SSA 60kDa

N=110

Pos

Neg

08110
Positive percent agreement: 100% (42/42)
Negative percent agreement: 100% (68/68)There were 1 equivocal results with the assay. For purposes
of calculation, these results were considered as negative.

Pos

42

0

NUTTS

Total

42

68

Neg

0

68

Overall agreement: 100% (110/110)

| SSA 52kDa

N=118MANUAL
PosNegTotal
CARISTYPos52153
Neg06565
Total5266118

There were 2 equivocal results with the assay. For purposes of calculation, these results were considered as negative. Positive percent agreement: 100% (52/52)

Negative percent agreement: 98.48% (65/66)

Overall agreement: 99.15% (117/118)

SmMANUAL
N=96PosNegTotal
CARISTPos24327
Neg06969
Total247296

There were 5 equivocal results with the assay. For purposes of calculation, these results were considered as negative. Positive percent agreement: 100% (24/24) Negative percent agreement: 95.83% (69/72) Overall agreement: 96.88% (93/96)

Scl70
N=97PosNegTotal
CARISPos29029
Neg06868
Total296897

There is 0 equivocal result with the assay. Positive percent agreement: 100% (29/29) Negative percent agreement: 100% (68/68) Overall agreement: 100% (97/97)

| SSB

N=92MANUAL
PosNegTotal
Pos25025
CARISTNeg06767
Total256792

There is 1 equivocal result with the assay. For purposes of calculation, these results were considered as negative. Positive percent agreement: 100% (25/25) Negative percent agreement: 100% (67/67) Overall agreement: 100% (92/92)

Sm/RNP
N=103MANUAL
PosNegTotal
CARISPos32335
Neg06868
Total3271103

There were 6 borderline results with the assay. For purposes of calculation, these results were considered as negative. Positive percent agreement: 100% (32/32) Negative percent agreement: 95.77% (68/71) Overall agreement: 97.09% (100/103)

| Jo1

N=95MANUAL
PosNegTotal
CARISTPos26026
Neg06969
Total266995

There is 0 equivocal result with the assay. Positive percent agreement: 100% (26/26) Negative percent agreement: 100% (69/69) Overall agreement: 100% (95/95)

13

CentromereMANUAL
N=93PosNegTotal
Pos20424
Neg06969
Total207393
RibosomeMANUAL
N=76PosNegTotal
CARISPos808
Neg06868
Total86876

There were 6 equivocal results with the assay. For purpose There is 0 equivocal result with the assay. of calculation, these results were considered as negative.

Positive percent agreement: 100% (20/20)

Negative percent agreement: 94.52% (69/73)

Overall agreement: 95.70% (89/93)

e There is 0 equivocal result with the assay.

Positive percent agreement: 100% (8/8) Negative percent agreement: 100% (68/68) Overall agreement: 100% (76/76)

| Antigenic
Specificity | Sample
number | Positive
percent
agreement | Negative
percent
agreement | Overall
agreement |
|--------------------------|------------------|----------------------------------|----------------------------------|----------------------|
| | | proportion | proportion | proportion |
| dsDNA | 112 | 100% | 100% | 100% |
| SSA 60 kDa | 110 | 100% | 100% | 100% |
| SSA 52 kDa | 118 | 100% | 98.48% | 98.48% |
| SSB | 92 | 100% | 100% | 100% |
| Sm | 96 | 100% | 95.83% | 95.83% |
| Sm/RNP | 13 | 100% | 95.77% | 95.77% |
| Scl70 | 97 | 100% | 100% | 100% |
| Jo1 | 95 | 100% | 100% | 100% |
| Centromere | 93 | 100% | 94.52% | 94.52% |
| Ribosome | 76 | 100% | 100% | 100% |

Table 12: Summary of performance agreement results

14

In addition to the analysis above, the 95% one-sided lower confidence limit in percent of proportion agreement (95% LCL (%) was calculated using the Exact Binomial Test for proportions to determine how low this proportion could be with a 95% confidence.

| Antigenic

SpecificityPositive percent agreementNegative percent agreementOverall percent agreement
N₁R₁P₁
(%)95% LCL (%)N₂R₂P₂
(%)95% LCL (%)NRP
(%)95% LCL (%)
dsDNA434310093.27696910095.7511211210097.36
SSA 60 kDa424210093.12686810095.6911011010097.31
SSA 52 kDa525210094.40666598.4893.0111811799.1596.04
SSB252510088.71676710095.63929210096.80
Sm242410088.27726995.8389.58969396.8892.12
Sm/RNP323210091.06716895.7789.4410310097.0992.64
Scl70292910090.19686810095.69979710096.96
Jo1262610089.12696910095.75959510096.90
Centromere202010086.09736994.5287.90938995.7090.43
Ribosome8810068.77686810095.69767610096.13

Table 13: Summary of performance agreements results - 95% LCL (%)

N1 = No. of positives; R1 = No. of positive agreements; P1 = R1/N1 N2 = No. of negatives; R2 = No. of negative agreements; P2 = R2/N2 N = N; + N2; R = R1 + R2; P = R/N

All previous evaluation results indicate that manual and automated (with CARIS™) assay preparation steps are considered substantially equivalents.

8. Conclusions

In conclusion, all supporting data demonstrate that the FIDIS™ CONNECTIVE 10* system can be considered substantially equivalent to the predicate device.

15

DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/15/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is circular and contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" around the perimeter. Inside the circle is an abstract symbol that resembles a stylized eagle or bird in flight.

Public Health Service

Food and Drug Administration 2098 Gaither Road Rockville MD 20850

DEC 1 9 2007

Biomedical Diagnostics S.A. (BMD) c/o Ms. Christelle Courivaud Regulatory Affairs Manager Actipole 25. 4-6 Bld de Beaubourg 77435 Marne La Vallée cedex 2 France

Re: K071210

Trade/Device Name: FIDISTM CONNECTIVE 10* Assay Regulation Number: 21 CFR 866.5100 Regulation Name: Antinuclear antibody immunological test system Regulatory Class: Class II Product Code: LLL, LKS, LKO, LKP, LSW, LJM, MQA Dated: December 5, 2007 Received: December 7, 2007

Dear Ms. Courivaud:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820). This letter will allow you to

16

Page 2 -

begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific information about the application of labeling requirements to your device, or questions on the promotion and advertising of your device, please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (240) 276-0450. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at 240-276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at 240-276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely yours,

Robert D. Becker

Robert L. Becker, Jr., M.D., Ph.Ø Director Division of Immunology and Hematology Devices Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health

Enclosure -

17

Indication for Use

510(k) Number (if known): K071210

Device Name:

FIDISTM CONNECTIVE 10*

Indication For Use:

The FIDIS™ CONNECTIVE 10* kit is a semi-quantitative homogeneous fluorescent-based microparticles immunoassay using flow cytometry. The test system is used to simulaneously detect the presence of 10 autoantibody specificities: double stranded DNA (dsDNAA), SSA (60 kDa and 52 kDa), SSB, Sm, Sm/RNP, Scl70, Jo1, ribosome and centromere,

(*Antibodies to dsDNA, Sm, Sm/RNP, SSA, SSB, Scl-70, Jo-1, ribosome and centromere can be reported using this assay).

Clinical utility:

The results of the FIDIS™ CONNECTIVE 10* are to be used in conjunction with the clinical findings and the other laboratory tests to aid in the diagnosis of connective diseases (systemic lupus erythematosus (SLE), Sjogren's syndrome, mixed connective disease (MCTD), scleroderma, dermatomyositis and CREST syndrome).

FIDIS™ CONNECTIVE 10* kit uses serum only, and is to be run on the FIDIS™ Instrument and MLX-BOOSTER™ Software.

FIDIS™ CONNECTIVE 10* kit may be used with the CARIS™ system (diluting and dispensing device).

This test is for in vitro diagnostic use.

Prescription Use __ X (21 CFR Part 801 Subpart D)

And/Or

Over the Counter Use (21 CFR Part 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE; CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of In Vitro Diagnostic Device Evaluation and Safety (OIVD)

Maria M Elan

Division Sign-Off Office of In Vitro Diagnostic Device Evaluation and Safety