Search Results
Found 4 results
510(k) Data Aggregation
(376 days)
MQA
EliA Rib-P is intended for the in vitro semi-quantitative measurement of IgG antibodies directed to Rib-P in human serum as an aid in the diagnosis of systemic lupus erythematosus (SLE) in conjunction with other laboratory and clinical findings. EliA Rib-P uses the EliA IgG method.
EliA Rib-P is a semi-quantitative solid-phase fluoroenzymeimmunoassay, for the determination of autoantibodies against Rib-P. The EliA Rib-P test System is fully integrated and automated system which comprises of assay-specific reagents, EliA method-specific reagents, and general reagents.
Assav-Specific Reagents include:
- EliA Rib-P Wells: coated with human recombinant ribosomal P-proteins P0, P1 . and P2 - 2 carriers (12 wells each), ready to use;
- . EliA ANA 3 Positive Control 250 or 2500/5000: Human monoclonal antibodies in Tris buffer containing IgG antibodies to Ro52, Rib-P and RNA Pol III – 6 single use vials, 0.3 mL each, ready to use;
- . EliA IgG/IgM/IgA Negative Control 250 or 2500/5000: Human blood preparation from healthy donors in PBS containing BSA, detergent and 0.095% sodium azide - 6 single-use vials, 0.3 mL each, ready to use;
EliA Method-Specific Reagents include:
- EliA Sample Diluent: PBS containing BSA, detergent and 0.095% sodium azide . - 6 bottles, 48 mL each, ready to use; or 6 bottles, 400 mL each, ready to use;
- I EliA IqG Conjuqate 50 or 200: ß-Galactosidase labeled anti-lgG (mouse monoclonal antibodies) in PBS containing BSA and 0.06% sodium azide - 6 wedge shaped bottles, 5 mL each, ready to use; or 6 wedge shaped bottles, 19 mL each, ready to use:
- . EliA IgG Calibrator Strips: Human IqG (0, 4, 10, 20, 100, 600 uq/L) in PBS containing BSA, detergent and 0.095% sodium azide - 5 strips, 6 single-use vials per strip, 0.3 mL each, ready to use;
- EliA IgG Curve Control Strips: Human IgG (20 ug/L) in PBS containing BSA, . detergent and 0.095% sodium azide - 5 strips, 6 single-use vials per strip, 0.3 mL each, ready to use:
- EliA IgG Calibrator Well: coated with mouse monoclonal antibodies 4 carriers . (12 wells each), ready to use;
General Reagents include:
- Development Solution: 0.01% 4-Methylumbelliferyl-β-D-galactoside, 110, 6x >170, or 6x >1165 determinations;
- I Stop Solution: 4% Sodium Carbonate - 6 bottles (65 mL, 119 mL, or 2800 mL each), sufficient for 6x >292, 6x >560, or 6x >13100 determinations;
- 트 Washing Solution Additive: detergent, preservative
The provided document is a 510(k) Summary for the EliA Rib-P device, an in vitro semi-quantitative immunoassay for detecting IgG antibodies directed to Rib-P as an aid in diagnosing Systemic Lupus Erythematosus (SLE). This document does not describe a study involving an AI-enabled device or human-in-the-loop performance. Instead, it focuses on the analytical and clinical performance of a standalone laboratory diagnostic device and its substantial equivalence to a predicate device.
However, I can extract and structure the information related to acceptance criteria and the performance study of this specific diagnostic device in the format requested, even without AI-specific parameters.
Acceptance Criteria and Device Performance Study for EliA Rib-P
The EliA Rib-P device is an automated semi-quantitative solid phase fluoroenzymeimmunoassay for the measurement of IgG antibodies directed to Rib-P in human serum. This summary outlines the performance characteristics tested to demonstrate its substantial equivalence to a legally marketed predicate device, the Quanta Lite Ribosome P ELISA.
1. Table of Acceptance Criteria and the Reported Device Performance
Parameter | Acceptance Criteria (Implied by Study Design) | Reported Device Performance (EliA Rib-P) |
---|---|---|
Analytical Performance | ||
Precision/Reproducibility | Inter-run, inter-instrument, and lot-to-lot variability to be within acceptable limits (typically low %CV). CLSI EP05-A3 guidelines followed. | Phadia 250: |
- Range of Total Imprecision %CV across 5 samples: 4.0% - 13.3%
- Within-lab Imprecision %CV across 4 samples: 3.5% - 13.8%
Phadia 2500/5000: - Range of Total Imprecision %CV across 5 samples: 5.7% - 18.4% (one sample had high CV, 18.4%) |
| Linearity/Reportable Range| Coefficient of determination (R²) close to 1.00, and slopes close to 1.00, across the measuring range. CLSI EP6-A guidelines followed. | Phadia 250: - R² values: 1.00 for all three dilution ranges.
- Slopes: 0.95, 1.00, 1.00.
Phadia 2500E: - R² values: 0.99, 1.00, 0.99 for the three dilution ranges.
- Slopes: 1.03, 1.00, 1.02.
"Linearity was shown for the entire measuring range." |
| Detection Limit | LoB, LoD, and LoQ to be established according to CLSI EP17-A2 guidelines (false positives and false negatives 10 EliA U/mL |
| Comparison Studies | | |
| Method Comparison (vs. Predicate Device) | High Positive Percent Agreement (PPA), Negative Percent Agreement (NPA), and Total Agreement with the predicate device. CLSI EP09c-ED3 followed. | EliA Rib-P (equivocal considered negative): - PPA: 94.7% (95% CI: 82.3% - 99.4%)
- NPA: 98.6% (95% CI: 96.4% - 99.6%)
- Total Agreement: 98.1% (95% CI: 96.0% - 99.3%)
EliA Rib-P (equivocal considered positive): - PPA: 100% (95% CI: 90.7% - 100%)
- NPA: 88.1% (95% CI: 83.7% - 91.6%)
- Total Agreement: 89.5% (95% CI: 85.6% - 92.6%) |
| Instrument Comparison | Strong correlation and agreement between different Phadia instrument series. | Regression analysis showed a slope of 0.94 (95% CI: 0.93 - 0.98) and an intercept of -0.76 (95% CI: -1.20 - -0.48) between Phadia 250 and Phadia 2500E. |
| Clinical Studies | | |
| Clinical Sensitivity and Specificity | Acceptable sensitivity for SLE and high specificity against other autoimmune and infectious diseases. | EliA Rib-P (equivocal considered positive): - Sensitivity (for SLE): 34.9% (95% CI: 27.2% - 43.3%)
- Specificity (against disease controls): 99.3% (95% CI: 97.9% - 99.9%)
EliA Rib-P (equivocal considered negative): - Sensitivity (for SLE): 28.1% (95% CI: 21.0% - 36.1%)
- Specificity (against disease controls): 99.8% (95% CI: 98.7% - 100%) |
2. Sample Size Used for the Test Set and Data Provenance
- Precision/Reproducibility:
- Phadia 250: 5 samples, tested in 252 replicates each across 3 lots and 3 instruments over 7 days.
- Within-Lab Imprecision: 4 samples, tested in 80 replicates each on 1 instrument over 20 days.
- Phadia 2500/5000 (E-module): 5 samples, tested in 84 replicates each on 3 instruments over 7 days.
- Provenance: Not explicitly stated, typically laboratory-prepared controls or banked human serum samples.
- Linearity/Assay Reportable Range: 3 serum samples (diluted). Provenance not explicitly stated.
- Detection Limit: 4 blank and 4 low-level samples (depleted IgG sera and prepared low-level samples) tested in 5-fold determination across 3 runs on Phadia 250 and Phadia 2500E.
- Analytical Specificity (Interference): 3 serum samples (one negative, one equivocal, one high positive) spiked with interfering substances. Provenance not explicitly stated.
- Assay Cut-Off: A cohort of 70 apparently healthy blood donors and 30 samples from SLE patients. Provenance not explicitly stated.
- Method Comparison with Predicate Device: 323 patient samples. Provenance not explicitly stated, but implies diverse clinical samples covering the measuring range.
- Instrument Comparison: 47 positive, 10 equivocal, and 28 negative samples. Provenance not explicitly stated.
- Clinical Sensitivity and Specificity: 560 clinically defined serum samples from various diagnostic groups (146 SLE, 414 disease controls). Provenance not explicitly stated, but these are patient samples with confirmed diagnoses.
- Expected Values/Reference Range: 638 apparently healthy subjects, equally distributed by age and gender from Caucasian, African American, Hispanic, and Asian populations obtained from a blood bank.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications
The document does not mention the use of experts to establish ground truth for the analytical performance characteristics. For clinical studies, the "ground truth" for patient samples was based on "clinically defined serum samples with a diagnosis" (e.g., systemic lupus erythematosus, Celiac disease, etc.). This typically implies diagnosis by medical professionals (e.g., rheumatologists, gastroenterologists, etc.) based on established diagnostic criteria, but the number and specific qualifications of these experts are not explicitly stated in this summary.
4. Adjudication Method (e.g. 2+1, 3+1, none) for the Test Set
No explicit adjudication method is mentioned. The ground truth for clinical samples appears to be based on pre-existing clinical diagnoses.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, If So, What Was the Effect Size of How Much Human Readers Improve With AI vs Without AI Assistance
This section is not applicable as the device is a standalone in vitro diagnostic immunoassay, not an AI-enabled device or an assist device for human readers. No MRMC study was conducted.
6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) Was Done
This device is a standalone algorithm/device in the sense that it performs automated semi-quantitative measurement of antibodies without real-time human interpretation impacting the measurement result. The assay directly measures the amount of antibody via fluorescence. The interpretation of results (negative, equivocal, positive) is based on predefined cut-offs.
7. The Type of Ground Truth Used (expert consensus, pathology, outcomes data, etc)
- Analytical ground truth: Based on reference materials, spiked samples, and dilution series with known concentrations or characteristics.
- Clinical ground truth: "Clinically defined serum samples with a diagnosis" (e.g., SLE patients, various disease controls). This implies established clinical diagnoses, likely based on standard diagnostic criteria, which could involve expert clinical assessment, pathology, and other laboratory findings.
8. The Sample Size for the Training Set
This document does not specify a separate "training set" in the context of machine learning or AI. For a traditional immunoassay, method development and optimization would involve various experiments and sample sets, but these are not explicitly termed "training sets" here. The "Assay Cut-Off" study used 70 healthy blood donors and 30 SLE patients to define the cut-off, which could be considered a form of "training" for the interpretive criteria.
9. How the Ground Truth for the Training Set Was Established
As noted above, a formal "training set" in the AI sense is not described. For the cut-off determination, the ground truth was established by using "apparently healthy blood donors" and "samples from SLE patients," indicating that these samples had known clinical statuses (healthy or diagnosed with SLE).
Ask a specific question about this device
(62 days)
MQA
The EL-ANA Profiles™ is intended to measure autoantibodies directed against the following autoantigens: single-stranded DNA (ssDNA), doublestranded DNA (dsDNA), Sm, RNP/Sm, SSA (Ro), SSB (La), Scl-70, Histones, Jo-1, Ribosomal Protein P , and Centromere. This system is intended as an aid in diagnosis of systemic lupus erythematosus and related rheumatic diseases.
Not Found
This is an FDA premarket notification (510(k)) letter for the EL-ANA Profiles™ device. The letter states that the device is substantially equivalent to legally marketed predicate devices and can be marketed subject to general controls.
Unfortunately, this document does not contain the detailed acceptance criteria or the study that proves the device meets those criteria. The letter is an official approval document, but it does not include the technical data and results of the clinical studies or performance validation studies. Such information is typically found in the 510(k) summary or the full 510(k) submission, which are separate documents.
Therefore, I cannot provide the requested information based solely on the provided text.
To address your request fully, I would need to review the actual 510(k) summary or the entire 510(k) submission for K024151.
Ask a specific question about this device
(63 days)
MQA
An enzyme linked immunosorbant assay (ELISA) for the semiquantitative detection of ribosome antibody in human serum to aid in the diagnosis of systemic lupus erythematosus (SLE) and other related connective tissue diseases.
Not Found
The provided text is a 510(k) clearance letter from the FDA for a diagnostic device. It does not contain information about acceptance criteria, device performance, study details (sample size, data provenance, expert qualifications, adjudication, MRMC studies, standalone performance), or ground truth establishment. This document only states that the device is substantially equivalent to legally marketed predicate devices and can be marketed.
Therefore, I cannot provide the requested information from the given text.
Ask a specific question about this device
(492 days)
MQA
The assay is to be used to detect antibodies in a single serum specimen. The results of the assay are to be used as an aid in the diagnosis of Systemic Lupus Erythematosus (SLE). FOR IN VITRO DIAGNOSTIC USE.
The Ribosomal P ELISA test is an enzyme-linked immunosorbent assay (ELISA) for the detection and semi-quantitation of IgG antibodies to Ribosomal P in human sera. The Ribosomal P ELISA test is an enzyme linked immunosorbent assay to detect IgG, M, A, antibodies to Ribosomal P. Purified Ribosomal P antigen is attached to a solid phase microtiter well. Diluted test sera is added to each well. If the antibodies are present that recognize the antigen, they will bind to the antigen in the well. After incubation the wells are washed to remove unbound antibody. An enzyme labeled antihuman IgG, M. A is added to each well. If antibody is present it will bind to the antibody attached to the antigen on well. After incubation the wells are washed to remove unbound conjugate. A substrate solution is added to each well. If enzyme is present the substrate will undergo a color change. After an incubation period the reaction is stopped and the color intensity is measured photometrically, producing an indirect measurement of specific antibody in the patient specimen.
Here's an analysis of the provided information regarding the acceptance criteria and study for the Ribosomal P ELISA Test Kit:
Acceptance Criteria and Reported Device Performance
Acceptance Criteria Category | Acceptance Criteria (Implicit/Explicit) | Reported Device Performance |
---|---|---|
Sensitivity | Acceptable relative sensitivity to Ouchterlony method. While no specific numeric target is given, the high value suggests an expectation of strong agreement. | Relative Sensitivity = 100% (95% CI: 93.5%-100%) (46/46 samples positive by Ouchterlony were positive by the device). |
Specificity | Acceptable relative specificity to Ouchterlony method. While no specific numeric target is given, the high value suggests an expectation of strong agreement. | Relative Specificity = 99.3% (95% CI: 97.8%-100%) (136/137 samples negative by Ouchterlony were negative by the device). |
Agreement | Acceptable overall agreement with the Ouchterlony method. | Relative Agreement = 99.5% (95% CI: 98.4%-100%) (182/183 samples showed agreement). |
Precision (C.V.) | Coefficient of Variation (C.V.) should be less than 15% with proper technique. | Intra-Assay and Inter-Assay C.V.s: Most reported C.V.s for Sera 1-4 are well below 15%. Serum 5 has an inter-assay C.V. of 25.6% and some individual assay C.V.'s of 19.6%. Serum 6 has even higher C.V.s (46.3%-48.0%). This suggests the device meets the criteria for some serum concentrations but not others, particularly at low values. |
Linearity | The assay should be semi-quantitative, meaning there should be a strong correlation between index values and log2 of dilution. | Linearity (r-value): 0.991 to 0.997 for five positive sera when comparing the Ribosomal P Index Value to log2 of dilution. This indicates excellent linearity and confirms its semi-quantitative nature. |
Cross-Reactivity | Antibodies to alternate autoimmune antigens (Ro, La, Scl-70, Jo-1, Sm, RNP, DNA) should not cross-react with the Ribosomal P ELISA kit. | Zero cross-reactivity observed: All 19 sera tested with high levels of antibodies to other autoimmune antigens produced negative interpretations (Immunoprobe Index Value |
Ask a specific question about this device
Page 1 of 1