(104 days)
The PM-50 Pulse Oximeter is a non-invasive, spot-check, oxygen saturation and pulse rate monitor. It operates only on battery power using existing PM-50 disposable and reusable sensors labeled for patients ranging from neonates to adults.
The PM-50 is a flexible, portable, battery powered Pulse Oximeter. The PM-50 Pulse Oximeter acquires the physiological signals
The provided 510(k) summary for the PM-50 Pulse Oximeter does not include specific acceptance criteria or details of a study that quantifies device performance against those criteria.
The submission focuses primarily on demonstrating substantial equivalence to a legally marketed predicate device (K052693 PM-50 Pulse Oximeter) through modifications involving the addition of new SpO2 probes.
Here's an analysis based on the information provided, highlighting what is present and what is missing based on your requested information:
1. A table of acceptance criteria and the reported device performance
- Not provided. The document states that "Laboratory testing and clinical study was conducted to validate and verify that the PM-50 Pulse Oximeter met all design specifications and was substantially equivalent to predicate device." However, no specific performance metrics, acceptance criteria, or quantitative results from these tests are included in this summary.
2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)
- Not provided. The summary mentions "clinical study" but offers no details about the sample size (number of patients, number of readings), data provenance, or study design (retrospective or prospective).
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)
- Not applicable/Not provided. For a pulse oximeter, the "ground truth" for oxygen saturation is typically established by arterial blood gas analysis, not by expert interpretation of images. The document does not describe how ground truth was established, beyond implying standard clinical methods would be used in a "clinical study."
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set
- Not applicable/Not provided. Adjudication methods like 2+1 or 3+1 are typically used for medical image interpretation where multiple experts assess the same case. This is not relevant for a pulse oximeter's core function of measuring physiological parameters.
5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
- Not applicable. An MRMC study is relevant for AI systems that assist human readers in interpreting complex medical data (like imaging studies). The PM-50 Pulse Oximeter is a standalone measurement device, not an AI-assisted diagnostic tool for human interpretation.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
- Yes, implicitly. The device itself is a standalone algorithm/device. Its performance is measured directly, not in conjunction with human interpretation for its primary function. The "clinical study" mentioned would assess the device's accuracy in measuring SpO2 and pulse rate. However, the details of how this standalone performance was measured and quantified are not provided.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)
- Not explicitly stated, but typically a gold standard like arterial blood gas analysis. For pulse oximeters, the accepted gold standard for determining actual oxygen saturation (SpO2) is typically arterial blood gas (ABG) analysis. The document only vaguely refers to a "clinical study" without detailing the ground truth methodology.
8. The sample size for the training set
- Not applicable/Not provided. This device is a physiological measurement device, not an AI/machine learning model that undergoes a "training" phase with a dataset in the conventional sense. Its "training" would be more akin to calibration and design optimization based on engineering principles and physiological models.
9. How the ground truth for the training set was established
- Not applicable. As above, this is not an AI/ML development pipeline.
Summary of what the document does state regarding validation:
- Testing Conducted: "Laboratory testing and clinical study was conducted to validate and verify that the PM-50 Pulse Oximeter met all design specifications and was substantially equivalent to predicate device."
- Quality Assurance Measures:
- Risk Analysis
- Environmental Testing
- Function and Performance Testing
- Clinical Study
- Conclusion: "The conclusions drawn from clinical and laboratory testing of the PM-50 Pulse Oximeter demonstrates that the device is as safe, as effective, and performs as well as or better than the legally marketed predicate device, the PM-50 Pulse Oximeter numbered K#052693."
In essence, the 510(k) summary asserts that testing was performed and concluded substantial equivalence, but it does not provide the specific data, acceptance criteria, or detailed methodologies typically sought when evaluating device performance against established clinical metrics. This level of detail is often found in the full 510(k) submission, not necessarily in the brief summary provided for public access. The core claim here is one of "substantial equivalence" based on similar technology and unspecified testing results matching a predicate device.
{0}------------------------------------------------
510(K) SUMMARY
SEP - 5 2006
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR § 807.92.
The assigned 510(k) number is: Ko61442
Submitter:
Shenzhen Mindray Bio-medical Electronics Co., LTD Mindray Building, Keji 12th Road South, Hi-tech Industrial Park, Nanshan, Shenzhen, 518057, P. R. China
Tel: +86 755 2658 2888
Fax: +86 755 2658 2680
- Contact Person: .
- Li Dongling
Shenzhen Mindray Bio-medical Electronics Co., LTD Mindray Building, Keji 12th Road South, Hi-tech Industrial Park, Nanshan, Shenzhen, 518057, P. R. China
- . Date Prepared:
May 12, 2006
Name of the device:
- Trade/Proprietary Name: PM-50 Pulse Oximeter .
- 0 Common Name: Pulse Oximeter
- e Classification
21 CFR 870.2700 Oximeter, Pulse Class II 21 CFR 870.2710 Ear Oximeter, Pulse Class II
Legally Marketed Predicate Device:
K052693 PM-50 Pulse Oximeter
MINDRAY PROPRIETARY
0026
{1}------------------------------------------------
Description:
The PM-50 is a flexible, portable, battery powered Pulse Oximeter. The PM-50 Pulse Oximeter acquires the physiological signals - oxygen saturation (SpO2) and pulse rate (PR). The signals are converted into digital data and processed, and the SpO2 and pulse rate values are calculated and displayed on LCD screen.
PM-50 uses a two-wavelength pulsatile system - red and infrared light – to obtain SpO2 based on the different light absorption of oxygenated and reduced hemoglobin. The light source in the finger sensor emits red and infrared light, which are partially absorbed and modulated by the arterial blood pulsation at the sensor site. The photodetector in finger sensor collects and converts the light into electronic signal which is proportional to the light intensity. The electronic signals are sent to the oximeter and processed by the oximeter's circuitry. Thereafter, the SpO₂ and pulse rate are obtained and indicated on the LCD screen.
Statement of intended Use:
The PM-50 Pulse Oximeter is a non-invasive, spot-check, oxygen saturation and pulse rate monitor. It operates only on battery power using existing PM-50 disposable and reusable sensors labeled for patients ranging from neonates to adults.
Comparison of Technological Characteristics:
The PM-50 Pulse Oximeter employs the same functional technology as the predicate device.
Testing:
The device modification involved the addition of SpO2 probes for the PM-50 Pulse Oximeter. The purpose of adding these added probes is a business decision. The subject new probes are selectable by customers; we will not discontinue marketing the previously cleared probes.
Laboratory testing and clinical study was conducted to validate and verify that the PM-50 Pulse Oximeter met all design specifications and was substantially equivalent to predicate device.
The following quality assurance measures were applied to the development of the
{2}------------------------------------------------
device:
- . Risk Analysis
- . Environmental Testing
- Function and Performance Testing
- Clinical Study .
Conclusion:
The conclusions drawn from clinical and laboratory testing of the PM-50 Pulse Oximeter demonstrates that the device is as safe, as effective, and performs as well as or better than the legally marketed predicate device, the PM-50 Pulse Oximeter numbered K#052693.
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a stylized caduceus symbol, which is a staff with two snakes entwined around it. The symbol is placed within a circle that contains the text "DEPARTMENT OF HEALTH AND HUMAN SERVICES - USA". The text is arranged around the perimeter of the circle.
SEP - 5 2006
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Shenzhen Mindray Bio-Medical Electronics Company, Limited C/O Ms. Susan D. Goldstein-Falk MDI Consultants, Incorporated 55 Northern Boulevard, Suite 200 Great Neck, New York 11021
Re: K061442
Trade/Device Name: PM-50 Pulse Oximeter Regulation Number: 21 CFR 870.2700 Regulation Name: Oximeter Regulatory Class: II Product Code: DQA Dated: August 8, 2006 Received: August 9, 2006
Dear Ms. Goldstein-Falk:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{4}------------------------------------------------
Page 2 - Ms. Goldstein-Falk
,
3.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
Chiu-Lian, Ph.D.
Chiu Lin, Ph.D. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
Exhibit #E
Page 1 of
510(k) Number (if known):
Device Name: PM-50 Pulse Oximeter
Indications For Use:
The PM-50 Pulse Oximeter is a non-invasive, spot-check, oxygen saturation and pulse rate monitor. It operates only on battery power using existing PM-50 disposable and reusable sensors labeled for patients ranging from neonates to adults.
Prescription Use × (Per 21 CFR 801 Subpart D)
OR
Over-The Counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Hice Suliom
Jon Sign-Off) ിഗ്ന Sign-Om)
Sign of Anesthesiology, General Hospital, I .. Jotion Control, Dental D o (0 (k) Number:
0024
§ 870.2700 Oximeter.
(a)
Identification. An oximeter is a device used to transmit radiation at a known wavelength(s) through blood and to measure the blood oxygen saturation based on the amount of reflected or scattered radiation. It may be used alone or in conjunction with a fiberoptic oximeter catheter.(b)
Classification. Class II (performance standards).