K Number
K060923

Validate with FDA (Live)

Date Cleared
2006-08-15

(133 days)

Product Code
Regulation Number
876.1500
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The NaviAid ™ BGE device is an accessory to an endoscope and is intended to ensure complete positioning of a standard endoscope in the small intestine (i.e., an endoscope that is 10-13 mm in diameter and is used for standard intestinal endoscopic visualization).

Device Description

The NaviAid™ BGE affords deep access into the small intestine, while maintaining all the advantages of an endoscopic procedure, such as back-and-forth navigation, stopping propagation if needed, real-time operation, video imaging, working channel including biopsy and treatment. The NaviAid™ Balloon Guided Endoscopy (BGE) system comprises an accessory kit containing the disposable balloon system and an air supply control unit for inflating and deflating the balloon system. The role of the Balloon Guided Endoscopy (BGE) disposable accessory kit is to facilitate advancement of a standard endoscope deeper into the small intestine. The NaviAid™ BGE system "upgrades" standard endoscopes to a double balloon endoscopy system. The NaviAid™ BGE accessory kit includes two balloons - the Guiding Balloon for Small Intestine ("GBS") and the Stabilizing Balloon ("SB"). Both balloons are inflated by ambient air. The Air Supply Unit ("ASU") operates and controls the inflation and deflation of the two balloons through two foot-pedals. Each balloon is connected to a dedicated tube that runs along the endoscope, and is connected at its proximal (user) end to the ASU. The balloon tubes are attached to the endoscope with clips and silicon bands. The NaviAid™ BGE accessory kit is mounted (deflated) on the tip of the endoscope, and is inserted with the endoscope into the gastrointestinal tract of the patient. The SB is the proximal balloon that is used to anchor the endoscope (near its tip) to the intestinal wall. The GBS is the distal balloon that can be advanced ahead of the endoscope tip or pulled back through pushing/pulling action on the GBS inflation tube at its proximal side, outside the patient body. When the GBS balloon is advanced and then inflated, it functions as a distal anchor, to which the endoscope tip (now with a deflated balloon) is advanced, and the GBS inflation tube serves as a "guidewire" that leads the endoscope as it is pushed towards the GBS anchoring location. The sequence of inflation and deflation of the balloons enables "pleating the small intestine on to the endoscope" or forming a rail on which the endoscope can be guided and advanced towards the GBS anchoring location. The balloons and tubes do not compromise the endoscope's flexibility, although its field of view may be reduced by up to 8%. Additionally, the balloons do not significantly compromise the maneuverability of the endoscope's tip and do not limit the usage of any standard endoscopy tools, such as biopsy forceps, snare, needle etc. The NaviAid BGE accessory kit is disposable and intended for single use, while the ASU is re-usable.

AI/ML Overview

The provided document describes the NaviAid™ BGE device and its non-clinical performance tests to demonstrate substantial equivalence to a predicate device, the Fujinon Double Balloon Endoscopy System. There is no information regarding a study defining acceptance criteria, and then proving the device meets that criteria. The document states "Clinical Performance Data: Not Applicable".

Here's an analysis of the provided information as it relates to your questions:

1. A table of acceptance criteria and the reported device performance

The document does not explicitly present a table of acceptance criteria with corresponding performance results. Instead, it lists several non-clinical performance tests conducted without specifying quantitative acceptance criteria or detailed results for each. The conclusion drawn is that "The non-clinical tests demonstrated that the NaviAid™ BGE device meets its design and performance specifications."

2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

This information is not provided. The document focuses on non-clinical tests rather than clinical studies with test sets of patient data.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

This information is not applicable as no clinical study with a "test set" and "ground truth" established by experts is described.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

This information is not applicable as no clinical study with a "test set" requiring adjudication is described.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

No such study was done. The device is an accessory to an endoscope, not an AI-powered diagnostic tool, and the document explicitly states "Clinical Performance Data: Not Applicable".

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

This is not applicable. The NaviAid™ BGE device is a physical accessory used with a human-operated endoscope, not a standalone algorithm.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

This is not applicable as no clinical study generating "ground truth" is described. The "ground truth" for the non-clinical tests would have been the engineering specifications and test protocols designed to ensure the device functions as intended.

8. The sample size for the training set

This is not applicable as there is no mention of a "training set" for an algorithm. The device is a physical medical accessory, not an AI model.

9. How the ground truth for the training set was established

This is not applicable as there is no "training set."

Summary of available information related to acceptance criteria and performance:

The document lists the following non-clinical performance tests conducted on the NaviAid™ BGE device:

  • Electrical & Mechanical Safety Testing (IEC 60601-1)
  • Electromagnetic Compatibility Testing (IEC 60601-1-2)
  • Software Validation (IEC 60601-1-4 & FDA Guidelines)
  • Biocompatibility Testing (ISO 10993)
  • ASU Test
  • Bond Strength Test
  • Air Leakage Test
  • Balloon Burst Pressure Test
  • GBS Inflation Tube & Bending Radius Test
  • Endoscope Flexibility & Insertion Test
  • Components Degradation Test
  • Endoscope Diameter Test
  • GBS Inflation/Deflation Cycle Test
  • ASU Connections
  • Disposables Mounting Testing
  • In-Vitro Validation Test
  • Packaging Validation

The document concludes that "The non-clinical tests demonstrated that the NaviAid™ BGE device meets its design and performance specifications." It also states, "Furthermore, the tests showed that the NaviAid™ BGE device is easy to use and user friendly and does not cause damage to the intestine."

Essentially, the "acceptance criteria" here are implied to be the successful completion and passing of these various non-clinical engineering and safety standards/tests. The "study that proves the device meets the acceptance criteria" is the execution of these listed non-clinical tests, with the reported device performance being that it "meets its design and performance specifications."

{0}------------------------------------------------

A. STEIN – REGULATORY AFFAIRS CONSULTING

. .

Appendix E

510(K) SUMMARY OF SAFETY & EFFECTIVENESS

NAVIAID™ BGE DEVICE

Page 1 of 5

510(k) Number K060923

Applicant's Name:

Company name:Smart Medical Systems Ltd.
Address:10 Hayetsira Str.
Ra'anana 43663
ISRAEL
Tel.:+972-9-7444478
Fax:+972-9-7444543
e-mail:gluria@smartmedsys.com

Contact Person:

Official Correspondent: Ahava Stein
Company name:A. Stein – Regulatory Affairs Consulting
Address:Beit Hapaamon (Suite 213)
20 Hata'as Str. (Box 124)
Kfar Saba 44425
ISRAEL
Tel:+ 972-9-7670002
Fax:+972-9-7668534
e-mail:asteinra@netvision.net.il

Name of the device:

NaviAid™ BGE device

Trade or proprietary name, if applicable:

NaviAid™ BGE device

Common or usual name:

Balloon Guided Endoscope Accessory Device or Balloon Guided Endoscopy Accessory Device

{1}------------------------------------------------

Page 2 of 5

A. STEIN - REGULATORY AFFAIRS CONSULTING

Appendix E

Establishment Registration No .:

Establishment registration form (Form FDA 2891) has been submitted but no registration number has been assigned yet.

Classification Name:

Gastrointestinal Tubes (and accessories)

Classification:

FDA has classified Double Balloon Endoscope devices as a Class II medical device, with product code KNT and 21 CFR classification code 876.1500. Review by the Gastroenterology/Urology Devices Panel.

Predicate Device:

The NaviAid™ BGE device is substantially equivalent to the Double Balloon Enteroscopy System (manufactured by Fujinon Inc. and the subject of 510(k) document no. K040048). A comparison table and detailed discussion are presented in Section 12 of this application.

Device Description:

The NaviAid™ BGE affords deep access into the small intestine, while maintaining all the advantages of an endoscopic procedure, such as back-and-forth navigation, stopping propagation if needed, real-time operation, video imaging, working channel including biopsy and treatment.

The NaviAid™ Balloon Guided Endoscopy (BGE) system comprises an accessory kit containing the disposable balloon system and an air supply control unit for inflating and deflating the balloon system. The role of the Balloon Guided Endoscopy (BGE) disposable accessory kit is to facilitate advancement of a standard endoscope deeper into the small intestine. The NaviAid™ BGE system "upgrades" standard endoscopes to a double balloon endoscopy system. The NaviAid™ BGE accessory kit includes two balloons - the Guiding Balloon for Small Intestine ("GBS") and the Stabilizing Balloon ("SB"). Both balloons are inflated by ambient air. The Air Supply Unit ("ASU") operates and controls the inflation and deflation of the two balloons through two foot-pedals. Each balloon is connected to a dedicated tube that runs along the endoscope, and is connected at its proximal (user) end to the ASU. The balloon tubes are attached to the endoscope with clips and silicon bands.

{2}------------------------------------------------

Page 3 of 5

The NaviAid™ BGE accessory kit is mounted (deflated) on the tip of the endoscope, and is inserted with the endoscope into the gastrointestinal tract of the patient. The SB is the proximal balloon that is used to anchor the endoscope (near its tip) to the intestinal wall. The GBS is the distal balloon that can be advanced ahead of the endoscope tip or pulled back through pushing/pulling action on the GBS inflation tube at its proximal side, outside the patient body. When the GBS balloon is advanced and then inflated, it functions as a distal anchor, to which the endoscope tip (now with a deflated balloon) is advanced, and the GBS inflation tube serves as a "guidewire" that leads the endoscope as it is pushed towards the anchoring GBS balloon. The sequence of inflation and deflation of the balloons enables "pleating the small intestine on to the endoscope" or forming a rail on which the endoscope can be guided and advanced towards the GBS anchoring location.

The balloons and tubes do not compromise the endoscope's flexibility, although its field of view may be reduced by up to 8%. Additionally, the balloons do not significantly compromise the maneuverability of the endoscope's tip and do not limit the usage of any standard endoscopy tools, such as biopsy forceps, snare, needle etc.

The NaviAid BGE accessory kit is disposable and intended for single use, while the ASU is re-usable.

Intended Use / Indication for Use:

The NaviAid™ BGE device is an accessory to an endoscope and is intended to ensure complete positioning of a standard endoscope in the small intestine (i.e., an endoscope that is 10-13 mm in diameter and is used for standard intestinal endoscopic visualization).

Comparison of Technological Characteristics with the predicate device:

The BGE device is similar to the Fujinon Double Balloon Endoscopy System regarding intended use and regarding the technological characteristics of the device.

Both the NaviAid™ BGE device and the Fujinon System are intended to facilitate penetration of standard endoscopes deep into the small intestine, while maintaining all the advantages of an endoscopic procedure, such as back-and-forth navigation, stopping propagation if needed, real-time operation, video imaging, working channel including biopsy and treatment. Both devices are based on the same basic principles of operation, that is, utilization of specialized double balloons based on an endoscopy system, that are alternately inflated and deflated in order to progress the endoscope and ensure complete positioning of the endoscope in the small intestine.

The devices both include the same basic components for achieving complete positioning of the endoscope in the small intestine, including an air supply unit & control pump and a double balloon system for advancing the endoscope. Although, in the Fujinon system one balloon is an integral part of the endoscope device (though it is disposable and replaced

{3}------------------------------------------------

Pg 4 of 5

A. STEIN - REGULATORY AFFAIRS CONSULTING

Appendix E

after every procedure) and the second balloon is positioned on an overtube, whereas, in the NaviAid™ BGE device both balloons are separate components not connected to the endoscope, thus enabling use of the device with any standard endoscope.

The balloon pressure specifications are also similar, as are the outer diameter of the fully inflated portion of the balloon. The working length of the systems is slightly different; the working length of the Fujinon System is 1,350 mm, whereas the working length of the NaviAid™ BGE device depends on the endoscope device and is typically 2,200 mm. Both devices are supplied non-sterile, for single use only.

The patient contact materials and accessories provided with the device are different. The Fujinon system is supplied with endoscopy accessories such as a light source, biopsy forceps, etc., whereas the NaviAid" BGE device is obviously not supplied with these accessories, as such accessories are supplied with the user supplied endoscope. Similar accessories include a foot pedal for additional control of the inflation/deflation sequencing. The NaviAid™ BGE device is also supplied with additional accessories to facilitate connecting the BGE disposable balloon system to the endoscope device (e.g. silicone bands and o-ring).

Both the NaviAid™ BGE device and the Fujinon System comply with the electrical and mechanical safety testing requirements and the electromagnetic compatibility testing requirements for electronic medical devices.

Non-Clinical Performance Data

The following performance tests were conducted on the NaviAid™ BGE device:

  1. Electrical & Mechanical Safety Testing (IEC 60601-1)

  2. Electromagnetic Compatibility Testing (IEC 60601-1-2)

  3. Software Validation (IEC 60601-1-4 & FDA Guidelines)

    1. Biocompatibility Testing (ISO 10993)
    1. ASU Test
    1. Bond Strength Test
    1. Air Leakage Test
    1. Balloon Burst Pressure Test
  1. GBS Inflation Tube & Bending Radius Test
    1. Endoscope Flexibility & Insertion Test
    1. Components Degradation Test
    1. Endoscope Diameter Test
    1. GBS Inflation/Deflation Cycle Test
    1. ASU Connections
    1. Disposables Mounting Testing
    1. In-Vitro Validation Test
    1. Packaging Validation

{4}------------------------------------------------

Page 5 of 5

A. STEIN - REGULATORY AFFAIRS CONSULTING

Appendix E

Clinical Performance Data

Not Applicable

Conclusions Drawn from Non-Clinical and Clinical Tests:

The non-clinical tests demonstrated that the NaviAid™ BGE device meets its design and performance specifications. Furthermore, the tests showed that the NaviAid™ BGE device is easy to use and user friendly and does not cause damage to the intestine.

The NaviAid™ BGE device may be safely and effectively used in Balloon Guided Endoscopy procedures in order to reach depths of the intestine that may not otherwise be accessible with a standard endoscope device.

Substantial Equivalence:

In summary, the intended use and principle of operation of the NaviAid™ BGE device and the Fujinon Double Balloon Endoscopy System are the same. Furthermore, the basic technological characteristics of the devices, including basic components, design, specifications and safety requirements, are similar. The differences in the technological characteristics do not raise new questions of safety and effectiveness. Consequently, the NaviAid™ BGE device is substantially equivalent to the Fujinon Double Balloon Endoscopy System.

Performance Standards:

The NaviAid™ BGE device complies with the voluntary recognized standards:

    1. Electrical & Mechanical Safety Testing (IEC 60601-1)
    1. Electromagnetic Compatibility Testing (IEC 60601-1-2)
    1. Software Validation (IEC 60601-1-4 & FDA Guidelines)
    1. Biocompatibility Testing (ISO 10993)

{5}------------------------------------------------

Image /page/5/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three lines representing its body and wings. The eagle is positioned within a circle, and the text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" is arranged around the upper portion of the circle.

Food and Drug Administration 9200 Corporate Blvd. Rockville MD 20850

AUG 1 5 2006

Smart Medical Technologies, Ltd. % Ms. Ahava Stein Regulatory Affairs Consultant Regulatory Affairs Consulting Beit Hapa'amon (Box 124) 20 Hata'as St., 44425 Kfar Saba ISRAEL

Re: K060923

Trade/Device Name: NaviAid™ BGE Device Regulation Number: 21 CFR 876.1500 Regulation Name: Endoscope and accessories Regulatory Class: II Product Code: FDA and KOG Dated: August 4, 2006 Received: August 8, 2006

Dear Ms. Stein:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Image /page/5/Picture/10 description: The image is a circular logo with the letters "FDA" in large, bold font in the center. Above the letters, the numbers "1906-2006" are printed. Below the letters, the word "Centennial" is printed in a smaller font. There are three small stars below the word "Centennial". The logo is surrounded by a thick black border.

Protecting and Promoting Public Health

{6}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at one of the following numbers, based on the regulation number at the top of this letter:

21 CFR 876.xxx(Gastroenterology/Renal/Urology)240-276-0115
21 CFR 884.xxx(Obstetrics/Gynecology)240-276-0115
21 CFR 894.xxx(Radiology)240-276-0120
Other240-276-0100

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150

or at its Internet address http://www.fda.gov/cdrl/industry/support/index.html.

Sincerely yours,

Nancy C. Hogdon

Nancy C. B rogdon Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{7}------------------------------------------------

INDICATIONS FOR USE STATEMENT

510(k) Number (if known):K060923
Device Name:NaviAid TM BGE Device
Indications for use:The NaviAid TM BGE device is an accessory to an endoscope and is intended to ensure complete positioning of a standard endoscope in the small intestine (i.e., an endoscope that is 10-13 mm in diameter and is used for standard intestinal endoscopic visualization).

V Prescription Use (Per 21 C.F.R. 801 Subpart D) OR

Over-The-Counter Use (Optional Format Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)


Concurrence of CDRH, Office of Device Evaluation (ODE)

Nancy L. Brogdon
(Division Sign Off)

(Division Sign-Off) Division of Reproductive, Abdominal. and Radiological Devic 510(k) Number_

Page 1 of 1

§ 876.1500 Endoscope and accessories.

(a)
Identification. An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.(b)
Classification —(1)Class II (special controls). The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.