K Number
K040048
Manufacturer
Date Cleared
2004-06-07

(147 days)

Product Code
Regulation Number
876.1500
Reference & Predicate Devices
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The device is intended for the Optical visualization of the upper gastrointestinal tract. This includes the esophagus, stomach, duodenum, and small bowel. It is intended for the observation, diagnosis, and endoscopic treatment.

Device Description

The small intestine is one of the most difficult organs to access in gastro-intestinal (GI) tract endoscopy. The Fujinon double balloon enteroscopy system facilitates diagnosis and treatment of the upper GI tract including the small intestine. The Fujinon double balloon enteroscopy system utilized balloons and over-tube to ensure complete positioning of the enteroscope in the small intestine. The tip of the scope can be smoothly inserted to reach the area of diagnosis. The Fujinon double balloon enteroscopy system includes the following: G5 endoscope / enteroscope (EN-450P5/20), Over-Tube (TS-12140), Balloon (BS-1), Balloon Pump Controller (PB-10). The subject device is used with the Fujinon Series 400 processor, the same processor used with the predicate device. The Fujinon double balloon enteroscopy system is used in conjunction with other peripherals specified in the Operation Manual such as: Light Source, Processor, Cart, Data Keyboard, Foot Switch, Monitor, Video Printer, Camera and Hard Copy Unit, VCR, ElectroSurgical Tastruments.

AI/ML Overview

The provided text is a 510(k) summary for the Fujinon Double Balloon Enteroscopy System. It focuses on demonstrating substantial equivalence to a predicate device rather than presenting a standalone study with detailed performance acceptance criteria and results.

Therefore, much of the requested information (such as specific performance criteria, detailed study design, sample sizes for test and training sets, expert qualifications, and adjudication methods) is not present in the provided document. The document emphasizes comparison to a predicate device and safety testing.

Here's what can be extracted and what is missing:


1. Table of Acceptance Criteria and Reported Device Performance

Acceptance Criteria: The document does not explicitly state quantitative acceptance criteria for device performance. Instead, it relies on demonstrating substantial equivalence to a predicate device and passing general safety tests. The implicit acceptance criterion is that the new device performs at least as well as the predicate device and is safe for its intended use.

Reported Device Performance: The primary "performance" reported is compliance with regulatory standards and the device's physical specifications.

CharacteristicReported Performance (Fujinon Double Balloon Enteroscopy System EN-450P5/20)
Viewing DirectionForward
Observation Range5~100mm
Field of View120°
Image Size130% (super)
Distal End Diameter8.5mm
Flexible Portion Diameter8.5mm
Bending Capability (Up)180°
Bending Capability (Down)180°
Bending Capability (Left)160°
Bending Capability (Right)160°
Forceps Channel Diameter2.2mm
Working Length2000mm
Total Length2300mm
Lens Diameter2.1mm
# of Pixels411,988 pixels
Pixels per Square mm83,892
Pixel Size (H x V)0.0032 x 0.003725
Active area of CCD (H x V)2.458 x 1.840
Type of CCD ChipInterline color chip
Balloon Pressure (PB-10)3.6 kpa - 7.6 kpa (5.6 kpa +/- 2 kpa)
Max Pump Flow Rate (PB-10)170 ml/10 sec
BiocompatibilityPassed (for balloon and overtube materials)
Electrical SafetyPassed
Thermal TestingPassed
EMC TestingPassed

2. Sample Size Used for the Test Set and Data Provenance

  • Sample Size: Not specified. The document refers to "electrical safety, thermal, and EMC testing requirements" and "biocompatibility testing" but does not give sample sizes for these tests. There is no mention of a clinical test set to evaluate diagnostic or treatment performance compared to the predicate device in terms of image quality or efficacy in patient outcomes.
  • Data Provenance: Not specified. The tests are general safety and performance tests typically conducted in a laboratory or engineering setting, not on patient data.

3. Number of Experts Used to Establish Ground Truth for the Test Set and Their Qualifications

  • Not Applicable. The document describes engineering and biocompatibility testing, not a study requiring expert readers to establish ground truth for a diagnostic outcome. The device's substantial equivalence is primarily based on technical specifications and safety profile compared to the predicate.

4. Adjudication Method for the Test Set

  • Not Applicable. As no expert reading or diagnostic performance evaluation study is described, no adjudication method would be relevant.

5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study Was Done

  • No. The document does not describe an MRMC comparative effectiveness study. The focus is on demonstrating substantial equivalence through technical specifications and safety testing, not on measuring human reader improvement with or without AI assistance. The device itself is an endoscope system, not an AI-assisted diagnostic tool.

6. If a Standalone (Algorithm Only Without Human-in-the-Loop Performance) Was Done

  • Not Applicable. The Fujinon Double Balloon Enteroscopy System is a medical device (endoscope), not a standalone algorithm. Its performance is intrinsically linked to human operation.

7. The Type of Ground Truth Used (Expert Consensus, Pathology, Outcomes Data, etc.)

  • Not Applicable / Implicit. For the engineering tests mentioned (electrical safety, thermal, EMC), the "ground truth" would be established by passing industry standards and internal specifications. For biocompatibility, the ground truth would be established by complying with recognized biocompatibility standards. There is no mention of ground truth related to diagnostic accuracy (e.g., pathology, outcomes data) because such a study is not described.

8. The Sample Size for the Training Set

  • Not Applicable. This device is an endoscope system, not an AI model. Therefore, there is no "training set" in the context of machine learning.

9. How the Ground Truth for the Training Set Was Established

  • Not Applicable. As there is no training set for an AI model, this question is not relevant to the provided 510(k) summary.

{0}------------------------------------------------

K 040048 pg ' of 4

510 (k) Summary

JUN - 7 2004

Date Prepared [21 CFR 807.92(a)(1)]

January 9, 2004

Submitter's Information [21 CFR 807.92(a)(1)]

Joseph M. Azary C/o Fujinon Inc. 543 Long Hill Avenue Shelton, CT. 06484

Azary Technologies has received authorization to submit this 510(k) on behalf of the sponsor Fujinon Inc. 10 High Point Drive, Wayne, NJ 07470, Establishment Registration# 2431293.

Trade Name, Common Name, Classification [21 CFR 807.92(a)(2)]

The device trade names are: Fujinon Double Balloon Enteroscopy System Common Name: Enteroscope Classification: Class II, 21 CFR 876.1500, FDA

Predicate Device [21 CFR 807.92(a)(3)]

  • Fujinon Enteroscope EN-410WM K993704
    The subject device have the same indications for use, material composition, viewing direction, image size, bending, reprocessing/sterilization method, and working dimensions as the predicate. The subject device uses the same processor and peripherals as the predicate device.

The main differences between the subject device and predicate device are as follows:

  • Minor differences with observation range, field of view, diameter, and length. .
  • Subject device includes an overtube, balloons, and a balloon pump. ●
  • The subject device includes the G5 upgrade, which is characterized by the following minor differ-. ences:
    • The L-Port has been eliminated. The L-port functioned as a lens wash port . Doctors had o the option to take a syringe to inject a fluid to use it as a high pressure wash for the lens. This function was eliminated because demand was low and it was rarely used by the surgeons.
    • The J-Port was repositioned. The J-Port is used as a jost water wash port. The J-Port was 0 repositioned based on doctor feedback. The port was moved from the bottom part to the top (end) of the scope. There was also a desire to eligainate check valves to facilitate reprocessing and cleaning, as well as prevent clogging.

{1}------------------------------------------------

K040.48 pg 2 of 4

  • A G5 forceps inlet port was modified. The new port is smaller and comes with a rubber O cap. The smaller port and rubber cap help increase suction and reduce leakage.
  • The jet wash line check valve was removed. Internal check valves were removed to elimio nate the potential for clogging and to facilitate cleaning, disinfection, and sterilization. The valves are now external and removable.
  • The suction and air/water cylinders and valves were upgraded. They were updated to ac-0 commodate the new valves. The function of the valves is the same.
  • Addition of the FOV, which is the rubber forceps inlet valve cover. This helps create a wa-O tertight seal when the endoscope is used.
  • Upgrade to CA-500 cleaning adaptor. The cleaning adaptor allows the scope to be con-0 nected to tubes for cleaning.

Description of the Device [21 CFR 807.92(a)(4)}

The small intestine is one of the most difficult organs to access in gastro-intestinal (GI) tract endoscopy. The Fujinon double balloon enteroscopy system facilitates diagnosis and treatment of the upper GI tract including the small intestine.

The Fujinon double balloon enteroscopy system utilized balloons and over-tube to ensure complete positioning of the enteroscope in the small intestine. The tip of the scope can be smoothly inserted to reach the area of diagnosis.

The Fujinon double balloon enteroscopy system includes the following:

  • G5 endoscope / enteroscope (EN-450P5/20) .
  • Over-Tube (TS-12140) .
  • Balloon (BS-1) .
  • Balloon Pump Controller (PB-10) .

The subject device is used with the Fujinon Series 400 processor, the same processor used with the predicate device.

The Fujinon double balloon enteroscopy system is used in conjunction with other peripherals specified in the Operation Manual such as:

  • Light Source .
  • . Processor
  • Cart .
  • Data Keyboard
  • Foot Switch .
  • Monitor .
  • Video Printer .
  • Camera and Hard Copy Unit .
  • VCR .
  • ElectroSurgical Tastruments .

{2}------------------------------------------------

$\qquad K^{v}4\quad 0048\quad B_{y}\quad ^{3}g\quad 4$

Specifications Chart:

EN-450P5/20
Viewing DirectionForward
Observation Range5~100mm
Field of view120
Image size130% (super)
Distal end diameter8.5mm
Flexible portion diameter8.5mm
Bending capabilitySee below
Up180
Down180
Left160
Right160
Forceps channel diameter2.2mm
Working length2000mm
Total length2300mm
Lens diameter2.1mm
# of pixels411,988 pixels
Pixels per square mm83, 892
Pixel size (H x V).0032 x .003725
Active area of CCD (H x V)2.458 x 1.840
Type of CCD chipInterline color chip
------------------------------------------------------------------------------------------------------------------------------------------------------------------------------.-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------NAL R BLOGARIADoor------------------------------------------------------------------------------------------------------------------------------------------------------------------------------Book Book Child Child Child Children Children Children Children Children Children Children Children Children Children Children Children Children Children Children Children ChA-4-4-4-4-4Section of the country of the closed on the controlled to the contribution of the comments of the comments of the comments of the comments of the comments of the comments ofA more. I cost and consisted com accession ofFEE AREER CRIMINATION COLUMN AN-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------Access the contributed to the contributed to the contributed to the contributed to the contributed to the contributed to the contributed to the contributed to the contributed------------------------------------------------------------------------------------------------------------------------------------------------------------------------------and the collection of the desires the first the first the first the first theAcres of the Council A children and consider and any and------------------------------------------------------------------------------------------------------------------------------------------------------------------------------Advertising the.------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
SELL CLEAN I LAND OR BALL BOOK OFand the most of the property of the continued any and changes and the considered and the contribution of the concessesCAR COLOR COLLECT OF COLLECTIONCompany Controller Comments ofCompany of the call of the collection
------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------Company of Canada Cases of Children Company Company Company Company Comments of Children Comments of Children..AND LI & A R ANNUA MOND SHOULD STATUS CONSULER
------------------------------------------------------------------------------------------------------------------------------------------------------------------------------AND LA BREAD LE LANDRED--------------------------------------------------------------------------------------------------------------FIR IN IN MINNERALING A.AN LAND CANDREAL CONNECT COLLEGION------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------A has and a comments a che all can and a comments of the first andCompany of Concession Company Company Company Company Company Company Company Company Company Company Company Company Company Company Company Company Company Company Company

Over-Tube Specifications

.------------------------------------------------------------------------------------------------------------------------------------------------------------------------------1------------------------------------------------------------------------------------------------------------------------------------------------------------------------------Specificant program and a che consisted and a set a superior contract------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------.ARE E ME RE LEBERT----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------LE MILE II DAMEAND AND A COLLEGION FOR- Status Support For Children Comments a.------------------------------------------------------------------------------------------------------------------------------------------------------------------------------.------------------------------------------------------------------------------------------------------------------------------------------------------------------------------Company of the county of the comments of the promote on
CO CALL COLLECTION1Carl Property Concession ComeCan Beach and Child A company a...CON EL LAND - NA AN AND AND AND AND------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------...

Balloon Pump Controller with Remote Switch PB-10 Specifications

Set Pressure of Balloon3.6 kpa - 7.6 kpa (5.6 kpa +/- 2 kpa)
Maximum flow rate of pump170 ml/10 sec
Power Supply120V / 230V
Dimensions300 (W) x 200(H) x 300(D) mm
Weight8.7 kg

The balloon is composed of natural rubber, which has been subjected to and passed biocompatibility I his banding vertube is composed of polyvinylpyrolidone hydromer coating, which has been subjected to and passed biocompatibility testing.

Fujinon Inc. Double Balloon Enteroscopy 510(k) Page 5-3

{3}------------------------------------------------

K cttco48 Pg 4 of 4

Intended Use [21 CFR 807.92(a)(5)]

The device is intended for the Optical visualization of the upper gastrointestinal tract. This includes the esophagus, stomach, duodenum, and small bowel. It is intended for the observation, diagnosis, and endoscopic treatment.

Technological Characteristics [21 CFR 807.92(a)(6)]

Fujinon, Inc. believes that the subject device is substantially equivalent to the predicate device. The subject device have the same indications for use, material composition, image size, bending, reprocessing/sterilization method, and working dimensions as the predicate. The subject device uses the same processor and peripherals as the predicate device.

The main technological difference is the balloon and overtube as a way to facilitate diagnosis and treatment of the upper GI tract including the small intestine, as well as the minor changes associated with the G5 upgrade.

Performance Data [21 CFR 807.92(b)(1)}

The subject device has been subjected to and passed electrical safety, thermal, and EMC testing requirements. The materials in the endoscope are identical to the materials used in the predicate device. The balloon and overtube materials have been subjected to and passed biocompatibility testing.

Conclusion [21 CFR 807.92(b)(3)]

We believe the changes are minor and conclude that the subject devices are as safe and effective as the predicate devices.

Fujinon Inc. Double Balloon Enteroscopy 510(k) Page 5-4

{4}------------------------------------------------

Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized depiction of an eagle with three horizontal lines above it, representing the department's commitment to health and well-being. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the eagle symbol.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

JUN - 7 2004

Fujinon, Inc. c/o Mr. Joseph M. Azary Azary Technologies LLC 543 Long Hill Avenue SHELTON CT 06484

Re: K040048

Trade/Device Náme: Fujinon Double Balloon Enteroscopy System Regulation Number: 21 CFR §876.1500 Regulation Name: Endoscope and accessories Product Code: 78 FDA Regulation Number: 21 CFR §876.5980 Regulation Name: Gastrointestinal tube and accessories Product Code: 78 KNT Regulatory Class: II Dated: May 5, 2004 Received: May 7, 2004

Dear Mr. Azary:

We have reviewed your Section 510(k) premarket notification of intent to market the device we nave reviewed your becamined the device is substantially equivalent (for the indications ferenced above and nave acterimes ally marketed predicate devices marketed in interstate for use slated in the enolosure) is regars and ment date of the Medical Device Amendments, or to commerce prior to way 20, 1976, the exaccidance with the provisions of the Federal Food, Drug, devices that have occh receired in asse equire approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The r ou may, therefore, mance are as not include requirements for annual registration, listing of general controls provisions of the 1100 ling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), II your device is classified (see above) into entire regulations affecting your device can be it may be subject to additional controller Life 21, Parts 800 to 898. In addition, FDA may round in the Code of Possients concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean i Tease oc advised that FDF i robation a vour device complies with other requirements of the Act that I DA has made a decommandin an administered by other Federal agencies. You must of any I ederal Statutes and regulations, but not limited to: registration and listing comply with an the rice I requirements, in 201); good manufacturing practice requirements as set (21 CFR Part 807), labemg (21 CFR Part 820); and if applicable, the electronic forth in the quality systems (QS) . og (sections 531-542 of the Act); 21 CFR 1000-1050.

{5}------------------------------------------------

This letter will allow you to begin marketing your device as described in your Section 510(k) This letter will allow you to begin marteenig your ential equivalence of your device to a legally premarket notification. The PDA inding of backantal epical on thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please If you desire specific advice tor your device on one of the following numbers, based on the regulation number at the top of the letter:

8xx.1xxx(301) 594-4591
876.2xxx, 3xxx, 4xxx, 5xxx(301) 594-4616
884.2xxx, 3xxx, 4xxx, 5xxx, 6xxx(301) 594-4616
892.2xxx, 3xxx, 4xxx, 5xxx(301) 594-4654
Other(301) 594-4692

Additionally, for questions on the promotion and advertising of your device, please contact the Additionally, 101 questions on the promote of the regulation entitled, "Misbranding
Office of Compliance at (301) 594-4639. Othice of Compliance at (501) 594-4057. Fab.) production of Super Bear the Division of Same by reference to premarket notification (21 OF Act may be obtained from the Division of Small
information on your responsibilities under the Act may be sitted from the 1800) 6 information on your responsionalities unec tales tales tales toll-free number (800) 638-2041 or
Manufacturers, International and Consumer Assistance at its toll-fremation htt Manufacturers, International and Colladmor i assistement of the main.html.

Sincerely yours,

Nancy C. Brogdon

Nancy C. Brogdon Director, Division of Reproductive, Abdominal and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{6}------------------------------------------------

5 10(k) Number (if known): // 04 00 48

Device Name: Fujinon Inc. Double Balloon Enteroscopy System

Indications For Use: The device is intended for the Optical visualization of the upper gastrointestinal tract. This includes the esophagus, stomach, duodenum, and small bowel. It is intended for the observation, diagnosis, and endoscopic treatment.

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH Office of Device Evaluation (ODE)

OR

Prescription Use
(Per 21 CFR 801.109)

Allin (Division Sign-Off)

Division of Reproductive, Abdominal, and Radiological Devices 510(k) Number.

Over-The-Counter Use _________________________________________________________________________________________________________________________________________________________

(Optional Format 1-2-96)

Fujinon Inc. Double Balloon Enteroscopy 510(k) Page 4

§ 876.1500 Endoscope and accessories.

(a)
Identification. An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.(b)
Classification —(1)Class II (special controls). The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.