K Number
K060359
Device Name
ARCHIECT INSULIN AND CONTROLS, MODEL 8K41-01, 8K41-10
Date Cleared
2006-04-14

(60 days)

Product Code
Regulation Number
862.1150
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP Authorized
Intended Use
The ARCHITECT® Insulin Calibrators are for the calibration of the ARCHTECT® i System when used for the quantitative determination of human insulin in human serum and plasma. The ARCHITECT® Insulin Controls are for the verification of the accuracy and precision of the ARCHITECT® i System when used for the quantitative determination of human insulin in human serum or plasma.
Device Description
The ARCHITECT® Insulin Calibrators are for the calibration of the ARCHITECT® i System when used for the quantitative determination of human insulin in human serum and plasma. The ARCHITECT® Insulin Controls are for the verification of the accuracy and precision of the ARCHITECT® i System when used for the quantitative determination of human insulin in human serum and plasma.
More Information

Not Found

No
The document describes calibrators and controls for a laboratory system, not the system itself, and there is no mention of AI/ML in the provided text.

No
This device is for the calibration and verification of a system that measures human insulin, not for therapeutic intervention. It is an in-vitro diagnostic (IVD) product that helps ensure the accuracy of a diagnostic test.

No

The device consists of calibrators and controls for an existing system, used for calibrating and verifying the accuracy and precision of quantitative determination of human insulin. These are components used in a diagnostic process, but they are not the diagnostic device itself which performs the determination. They are used to ensure the proper functioning of the diagnostic device.

No

The device description refers to "Calibrators" and "Controls," which are typically physical reagents or materials used in laboratory testing, not software. The intended use also specifies calibration and verification of a physical system ("ARCHITECT® i System").

Yes, this device is an IVD (In Vitro Diagnostic).

Here's why:

  • Intended Use: The intended use clearly states that the calibrators and controls are for the "quantitative determination of human insulin in human serum and plasma." This involves testing biological samples (serum and plasma) outside of the body to obtain diagnostic information.
  • Device Description: The description reiterates the use of the calibrators and controls for the "quantitative determination of human insulin in human serum and plasma" and for "verification of the accuracy and precision" of the system used for this determination. This further confirms its role in an in vitro diagnostic process.
  • Predicate Devices: The listed predicate devices (ADIVA Centaur and ACS: 180 Insulin Calibrators; Bayer Ligand Plus 1, 2, 3 Controls) are also IVD devices used in clinical laboratories for similar purposes (insulin testing and quality control).

The core function of these calibrators and controls is to support the accurate and reliable performance of an in vitro diagnostic test (the quantitative determination of insulin in biological fluids). This directly aligns with the definition of an IVD.

N/A

Intended Use / Indications for Use

The ARCHITECT® Insulin Calibrators are for the calibration of the ARCHTECT® i System when used for the quantitative determination of human insulin in human serum and plasma.

The ARCHITECT® Insulin Controls are for the verification of the accuracy and precision of the ARCHITECT® i System when used for the quantitative determination of human insulin in human serum or plasma.

Product codes

JIT, JJX

Device Description

Calibrators

The ARCHITECT® Insulin Calibrators are for the calibration of the ARCHITECT® i System when used for the quantitative determination of human insulin in human serum and plasma.

Controls

The ARCHITECT® Insulin Controls are for the verification of the accuracy and precision of the ARCHITECT® i System when used for the quantitative determination of human insulin in human serum and plasma.

Mentions image processing

Not Found

Mentions AI, DNN, or ML

Not Found

Input Imaging Modality

Not Found

Anatomical Site

Not Found

Indicated Patient Age Range

Not Found

Intended User / Care Setting

Not Found

Description of the training set, sample size, data source, and annotation protocol

Not Found

Description of the test set, sample size, data source, and annotation protocol

Not Found

Summary of Performance Studies (study type, sample size, AUC, MRMC, standalone performance, key results)

Not Found

Key Metrics (Sensitivity, Specificity, PPV, NPV, etc.)

Not Found

Predicate Device(s)

K021535, K030452

Reference Device(s)

Not Found

Predetermined Change Control Plan (PCCP) - All Relevant Information

Not Found

§ 862.1150 Calibrator.

(a)
Identification. A calibrator is a device intended for medical purposes for use in a test system to establish points of reference that are used in the determination of values in the measurement of substances in human specimens. (See also § 862.2 in this part.)(b)
Classification. Class II (special controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9.

0

APR 1 4 2006

K060359

This is a Summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of CFR.

The assigned 510(k) Number is: Company/Contact Person Hideo G. Noda Denka Seiken Co., Ltd. 3-4-2 Nihonbashi Kayabacho, Chuo-ku, Tokyo, Japan 103-0025 Operator Number: 9053049 Establishment Registration Number: 3003871639 Date Prepared: Dec 19, 2005

Device Name

Calibrators

Trade Name: ARCHITECT® Insulin Calibrators (A-F) Common Name: Calibrator Device Classification: 21 CFR 862:1150 Class II Classification Panel: Clinical Chemistry Product Code: JIT

Controls

Trade Name: ARCHITECT® Insulin Controls (Low, Medium, and High) Common name: Quality Control Material (Assayed) Single (Specified) analyte Device Classification: 1 CFR 862:1660 Class I Classification Panel: Clinical Chemistry Product Code: JJX

Legally marketed device to which equivalency is claimed: ADIVA Centaure and ACS: 180 Insulin

1

Indications for Use

Koco359 510(k) Number (if known) All

Device Name: ARCHITECT® Insulin Calibrators and Controls

Indications For Use:

Intended Use and Indications for use

The ARCHITECT® Insulin Calibrators are for the calibration of the ARCHTECT® i System when used for the quantitative determination of human insulin in human serum and plasma.

The ARCHITECT® Insulin Controls are for the verification of the accuracy and precision of the ARCHITECT® i System when used for the quantitative determination of human insulin in human serum or plasma.

______________________________________________________________________________________________________________________________________________________________________________ Prescription Use YES (Part 21 CFR 801 Subpart D)

AND/OR

Over-The-Counter Use NO (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD)

2

Description of the Device

Calibrators

The ARCHITECT® Insulin Calibrators are for the calibration of the ARCHITECT® i System when used for the quantitative determination of human insulin in human serum and plasma.

Controls

The ARCHITECT® Insulin Controls are for the verification of the accuracy and precision of the ARCHITECT® i System when used for the quantitative determination of human insulin in human serum and plasma.

Comparison of Technological Characteristics:

The ARCHITECT® Insulin Calibrators (A-F) is substantially equivalent to the ADIVA Centaur and ACS 180 Insulin Calibrators (K021535).

Comparison with predicate: Calibrators

Similarities:

CalibratorDevicePredicate
Intended UseThe ARCHITECT® Insulin
calibrators are for
calibration of the
ARCHITECT® i System
when used for the
quantitative determination of
human insulin in human
serum and plasma.For in vitro diagnostic use in
the calibrating the ADIVA
Centaur® or ACS 180®
Insulin assays.
MethodologyCMIA (Chemiluminescent
Microparticle Immunoassay)Chemiluminescent
Microparticle Immunoassay
Binding ProteinInsulinInsulin
Assay ProtocolsDirect, quantitative
immunoassayDirect, quantitative
immunoassay
Traceability
/StandardizationRelative Light Unit (RLU)
matched to Primary
Calibrators. The
calibrators of the
ARCHITECT Insulin are
referenced to the World
Health Organization (WHO)Referenced to the World
Health Organization (WHO)
Insulin 1st. International
Reference Preparation,
66/304.
Assigned values of

3

| Insulin 1st. International
Reference Preparation,
66/304 | calibrators are traceable to
this standardization |

----------------------------------------------------------------------------------------------------------------------

Comparison with predicate: Controls

The ARCHITECT® Insulin Controls (Low, Medium and High) are substantially equivalent to the Bayer Ligand Plus 1, 2, 3 Controls (K030452).

Comparison with predicate: Controls

Similarities:

ControlsDevicePredicate
Intended UseThe ARCHITECT® Insulin
Controls are for the
verification of the accuracy
and precision of the
ARCHITECT® i System
when used for the
quantitative determination of
human insulin in human
serum and plasma.For the in vitro diagnostic
use to monitor the precision
and accuracy of
immunochemistry and
procedures for ADIVA
Centaur® and ACS 180®
Systems.
MethodologyChemiluminescent
Microparticle Immunoassay
(CMIA)Chemiluminescent
Microparticle Immunoassay
Binding ProteinInsulinInsulin
Assay ProtocolDirect, quantitative
immunoassayDirect, quantitative
immunoassay
Levels3 levels Low, Medium and
High: targets: 8, 40,
120 μU/mL.3 levels (Ligand 1, 2, 3).

Comparison with predicate: Calibrators

Differences:

CalibratorsDevicePredicate
PlatformARCHITECT® i SystemADIVA Centaur® or ACS
180®
MatrixAcetate buffer with sodiumBuffered saline with

4

| | azide and preservatives | casein, potassium
thiocyanate (3.89%),
sodium azide and
preservatives |
|--------------------------|------------------------------------------------|--------------------------------------------------------------------------------|
| Calibration Range/Levels | 6 levels: 0, 3, 10, 30, 100,
and 300 µU/mL, | High and Low level, per
assigned value card |
| Assay Sample Type | Serum and plasma | Serum |

Comparison with predicate: Controls

Differences:

ControlDevicePredicate
PlatformARCHITECT® i SystemADIVA Centaur® or ACS
180°
MatrixAcetate buffer with
preservativesLypholized, Multi
Constituent Controls.
Human serum with
nonhuman contents added,
no preservatives or
stabilizers
TraceabilityPrimary ControlsNot given
Value AssignmentRelative Light Unit (RLU)
matched to Primary
Controls.Adjusted to the level listed
in expected values of assay
package insert.
Assay Sample TypeSerum and plasmaSerum

Conclusion:

As summarized above the ARCHITECT® Insulin Calibrators (A-F) and Controls (Low, Medium, and High) are substantially equivalent to the ADIVA Centaur® and ACS: 180® Insulin Calibrators (K021535) and Bayer Ligand Plus 1, 2, 3 Controls (K030452).

Substantial equivalence for the calibrators has been demonstrated as recommended by the FDA Guidance for Industry "Abbreviated 510(k) Submission for In Vitro Calibrators" (issued on: Feb 22, 1999) and for controls as recommended by the FDA Guidance for Industry "Points To Consider Document On Assayed and Unassayed Quality Control Material" (Draft Guidance Released for comment on February 3, 1999).

5

DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/5/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo features a stylized eagle with three curved lines representing its wings. The eagle is positioned to the right of the department's name, which is written in a circular pattern around the logo. The text is in all capital letters and is written in a sans-serif font.

Public Health Service

Food and Druq Administration 2098 Gaither Road Rockville MD 20850

APR 1 4 2006

Mr. Toshimi Matsunaga Denka Seiken Co., Ltd. 1-2-2 Minamihoncho, Gosen-shi Niigata, Japan 959-1695

Re: K060359

Trade/Device Name: ARCHITECT® Insulin Calibrators and Controls Regulation Number: 21 CFR§862.1150 Regulation Name: Calibrator Regulatory Class: Class II Product Code: JIT, JJX Dated: February 2, 2006 Received: March 2, 2006

Dear Mr. Matsunaga:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820).

6

Page 2 -

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific information about the application of labeling requirements to your device, or questions on the promotion and advertising of your device, please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (240) 276-0484. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely yours,

Albert G. A.

Alberto Gutierrez, Ph.D Director Division of Chemistry and Toxicology Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health

Enclosure

7

Indications for Use

510(k) Number (if known) NA K060359

Device Name: ARCHITECT® Insulin Calibrators and Controls

Indications For Use:

Intended Use and Indications for use

The ARCHITECT® Insulin Calibrators are for the calibration of the ARCHTECT® i System when used for the quantitative determination of human insulin in human serum and plasma.

The ARCHITECT® Insulin Controls are for the verification of the accuracy and precision of the ARCHITECT® i System when used for the quantitative determination of human insulin in human serum or plasma.

| Prescription Use

(Part 21 CFR 801 Subpart D)YES
AND/OR
Over-The-Counter Use
(21 CFR 801 Subpart C)NO

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD)

Office of In Vitro Diagnostic Device
Evaluation and Safety

K060359