(94 days)
The Linemaster wireless foot switch is an accessory that is indicated for use to provide foot switch input control to any medical device that uses a wired foot switch with a switch closure (on/off), to activate said device.
The system is intended for use in hospitals, outpatient clinics and physicians offices.
The Linemaster wireless foot switch will be an accessory to and provide foot switch input control to any medical device that uses a foot switch to control on/off inputs. The system includes a wireless foot switch and a receiver. The wireless foot switches share the same transmission schemes, however each manufacturer will have a unique identification code that is programmed into the transmitter and receiver. This identification code will only allow communication of systems that have been programmed with the same identification code. Systems from different manufacturers will not communicate with each other.
When coupled with a system that uses a wired foot switch the wireless system will function in the same way as the wired version. Apart from the software, the technological characteristics are the same or similar to, those of the predicate wired device where LED-based systems and infrared signals are used to operate the device.
The elimination of wires on the floor of the operating room will improve the safety and efficiency by uncluttering the OR floor and reducing set-up and cleanup time.
The provided text describes the Linemaster IR Wireless Foot Switch and its conformity to various safety and performance standards. However, it does not detail specific acceptance criteria or a study proving the device meets those criteria in the way typically expected for an AI/ML device.
It's crucial to understand that this document is a 510(k) Summary for a non-AI/ML medical device (a wireless foot switch). Therefore, many of the requested categories for AI/ML device evaluation (like sample size for test/training sets, expert ground truth, MRMC studies, standalone performance, etc.) are not applicable and thus not present in this type of submission.
Here's a breakdown of the available information based on your request, highlighting where the document's content differs from what would be expected for an AI/ML device:
1. Table of Acceptance Criteria and Reported Device Performance
| Acceptance Criteria (Standard Compliance) | Reported Device Performance |
|---|---|
| Electrical Safety | |
| IEC/UL 60601.1 (Fire, Shock, Mechanical Hazards) | Meets standard |
| Can/CSA C22.2 No. 60601.1 (Electric Shock, Fire, Mechanical, other specified hazards) | Meets standard |
| EN60601-1-2: 2001 (Electromagnetic Compatibility - EMC) | Meets standard |
| Electromagnetic Immunity | |
| IEC 1000-4-2: 1995 (Electrostatic Immunity) | Meets standard |
| IEC 1000-4-3: 1995 (Radiated Electromagnetic Field Immunity @ 10uv/m) | Meets standard |
| IEC 1000-4-4: 1995 (Electrical Fast Transients Immunity) | Meets standard |
| IEC 1000-4-5: 1995 (Surge Immunity) | Meets standard |
| IEC 1000-4-6: 1996 (Conducted RF Immunity) | Meets standard |
| IEC 1000-4-8: 1993 (Power Frequency Magnetic Field Immunity) | Meets standard |
| IEC 1000-4-11: 1994 (Voltage Dips and Variations) | Meets standard |
| Electromagnetic Emissions | |
| EN 55011: 1998 (Radiated and Conducted Emissions, Group 1 Class B) | Meets standard |
| FCC Part 15 (Radiated and Conducted Emissions, Class B) | Meets standard |
| Power Quality | |
| IEC 61000-3-2: 2000 (Power Harmonics Class A) | Meets standard |
| IEC 61000-3-3: 1995 + A1: 2001 (Voltage Fluctuation, Section 5) | Meets standard |
| Overall Effectiveness | |
| Safety and Effectiveness of wireless footswitch | Demonstrated through risk analysis, verification, and validation testing |
Explanation of "Study":
For this device, the "study" is the verification and validation testing performed to ensure compliance with the listed national and international consensus standards for electrical medical equipment and electromagnetic compatibility. This is a standard process for hardware devices, not an AI/ML performance study.
The following questions are not applicable or cannot be answered from the provided text, as they pertain to AI/ML device evaluation criteria which are not relevant to a physical wireless foot switch submission:
2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)
Not applicable. There isn't a "test set" in the context of data for an AI/ML model. The testing involved physical device performance against engineering and safety standards.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)
Not applicable. "Ground truth" for an AI/ML model is not relevant here. Compliance with standards is typically assessed by trained engineers and certified testing laboratories.
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set
Not applicable. There is no "adjudication method" in the context of data labeling for AI/ML.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
Not applicable. This is a hardware device; it does not "assist human readers" in interpreting data or images.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
Not applicable. This is a physical device that functions as an accessory. Its performance is inherent to its design and manufacturing, not an algorithm's output.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)
Not applicable. "Ground truth" in the AI/ML sense is not relevant. The "truth" here is adherence to specified engineering and safety parameters as defined by the standards.
8. The sample size for the training set
Not applicable. There is no "training set" for an AI/ML model.
9. How the ground truth for the training set was established
Not applicable. There is no "training set" or corresponding ground truth establishment process for a physical foot switch.
In summary, the provided document is a 510(k) summary for a traditional medical device, demonstrating its safety and effectiveness through compliance with established engineering and safety standards. It does not involve AI/ML technology, and therefore, the specific criteria for evaluating AI/ML models are not addressed.
{0}------------------------------------------------
0535/0
Image /page/0/Picture/1 description: The image shows the logo for Linemaster, a company that makes foot switches. The logo features the word "LINEMASTER" in large, bold letters, with a stylized globe behind it. Below the company name is the text "AMERICA'S FOOT SWITCH LEADER".
MAR 2 0 2006
510K SUMMARY OF SAFETY AND EFFECTIVENESS DATA
General Information
Manufacturer:
Linemaster Switch Corporation 29 Plaine Hill Road Woodstock, CT 06281 860-974-1000 Phone 860-974-1533 Fax
Establishment Registration Number:
????????????
Contact Person:
Michael W. Szostek Components/Safety Engineer mszostek@linemaster.com
Date Prepared:
December 1, 2005
Classification Name:
Endoscope and Accessories Bone Cutting Instruments and Accessories. Electrosurgical Cutting and Coagulation Device and Accessories. General and Plastic Surgery Any class II device that uses a foot switch as an accessory.
Common and Usual Name:
Wireless Foot switch
1
$\rho i+\sqrt{3}$
{1}------------------------------------------------
Image /page/1/Picture/0 description: The image shows the logo for "LINEMASTER", which is written in large, bold, black letters. Above and below the text is a black and white drawing of the earth. Below the logo is the text "AMERICA'S FOOT SWITCH LEADER".
Proprietary Name:
IR Wireless Foot Switch
Predicate Devices:
The Linemaster infrared wireless footswitch is substantially equivalent in function and intended use to the following legally marketed devices: Applied Surgical, LLC, Gemini Surgical and Display System (K051831), Stryker wireless universal footswitch (K033135), SurgASSIST system with wireless remote control (K020343), Megadyne Mega power electrosurgical generator accessories (K050579) and Valleylab force FXc accessories (K944602).
Device Description:
The Linemaster wireless foot switch will be an accessory to and provide foot switch input control to any medical device that uses a foot switch to control on/off inputs. The system includes a wireless foot switch and a receiver. The wireless foot switches share the same transmission schemes, however each manufacturer will have a unique identification code that is programmed into the transmitter and receiver. This identification code will only allow communication of systems that have been programmed with the same identification code. Systems from different manufacturers will not communicate with each other.
When coupled with a system that uses a wired foot switch the wireless system will function in the same way as the wired version. Apart from the software, the technological characteristics are the same or similar to, those of the predicate wired device where LED-based systems and infrared signals are used to operate the device.
The elimination of wires on the floor of the operating room will improve the safety and efficiency by uncluttering the OR floor and reducing set-up and cleanup time.
Performance Data:
The infrared wireless foot switch system meets the following standards:
- Classified to IEC/UL 60601.1 Medical electrical equipment by Underwriter's . Laboratories, Inc with respect to fire, shock, and mechanical hazards in accordance with IEC/UL 60601.1.
- Classified with respect to electric shock, fire, mechanical, and other specified . hazards only, in accordance with Can/CSA C22.2 No. 60601.1
{2}------------------------------------------------
Image /page/2/Picture/0 description: The image shows the logo for Linemaster, which is placed over a graphic of the earth. The text "LINEMASTER" is in bold, black letters. Below the logo is the text "AMERICA'S FOOT SWITCH LEADER".
- EN60601-1-2: 2001 ●
- IEC 1000-4-2: 1995 Electrostatic Immunity
- IEC 1000-4-3: 1995 Radiated Electromagnetic Field Immunity @ 10uv/m .
- IEC 1000-4-4: 1995 Electrical Fast Transients Immunity ●
- . IEC 1000-4-5: 1995 Surge Immunity
- . IEC 1000-4-6: 1996 Conducted RF Immunity
- . IEC 1000-4-8: 1993 Power Frequency Magnetic Field Immunity
- . IEC 1000-4-11: 1994 Voltage Dips and Variations
- . EN 55011: 1998 Radiated and Conducted Emissions, Group 1 Class B
- FCC Part 15 Radiated and Conducted Emissions, Class B .
- . IEC 61000-3-2: 2000 Power Harmonics Class A
- IEC 61000-3-3: 1995 + A1: 2001 Voltage Fluctuation, Section 5 .
The safety and effectiveness of the wireless footswitch described in this submission has been demonstrated through risk analysis, verification, and validation testing.
Indications for Use:
The Linemaster wireless foot switch will be an accessory to and provide foot switch input control to any medical device that uses a switch closure (on 1001), to activate said device.
Warnings:
Plasma video displays can seriously affect the operation of infrared devices. The plasma devices can increase switch activation times significantly or gerreest the switch operation altogether.
Do not use in the vicinity of plasma video displays.
Summary of Technological Characteristics:
The infrared wireless foot switch in this submission is similar with regard to design, operation, and materials and intended use to the predicate devices indicated above. Therefore, no new safety or efficacy issues are created and the wireless foot switch is substantially equivalent to the predicate device
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image contains the text 'Public Health Service'. The text is in a simple, bold font. The words are stacked on top of each other. The background is white.
Image /page/3/Picture/2 description: The image shows the seal of the Department of Health and Human Services (HHS). The seal features a stylized eagle with its wings spread, symbolizing protection and service. The words "DEPARTMENT OF HEALTH AND HUMAN SERVICES - USA" are arranged in a circular pattern around the eagle.
MAR 2 0 2006
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Linemaster Switch Corporation c/o Mr. Michael W. Szostek Component/Safety Engineer 29 Plaine Hill Road Woodstock, Connecticut 06281
Re: K053510
Trade/Device Name: IR Wireless Foot Switch Regulation Number: 21 CFR 878.4400 Regulation Name: Electrosurgical cutting and coagulation device and accessories Regulatory Class: II Product Code: GEI Dated: March 10, 2006 Received: March 13, 2006
Dear Mr. Szostek:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set
{4}------------------------------------------------
Page 2 - Mr. Michael W. Szostek
forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html
Sincerely yours,
egree
Mark N. Melkerson Acting Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
Indications for Use
510(k) Number : (K053510)
Device Name: IR Wireless Foot Switch
Indications For Use:
The Linemaster wireless foot switch is an accessory that is indicated for use to provide foot switch input control to any medical device that uses a wired foot switch with a switch closure (on/off), to activate said device.
The system is intended for use in hospitals, outpatient clinics and physicians offices.
Prescription Use × (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use _ (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
elgmdl
(Divis Division of General, Restorative, and Neurological Devices
Page 1 of 1
510(k) Number_ Kos350
§ 878.4400 Electrosurgical cutting and coagulation device and accessories.
(a)
Identification. An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.(b)
Classification. Class II.