Search Filters

Search Results

Found 2 results

510(k) Data Aggregation

    Why did this record match?
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Mitek VAPR VUE Radiofrequency System is intended for resection, ablation and excision of soft tissue, and hemostasis of blood vessels in patients requiring arthroscopic surgery of the knee, shoulder, ankle, elbow and wrist.

    The P50 and P90 Electrodes when used with Mitek VAPR Electrosurgical Systems are intended for resection, ablation and excision of soft tissue, and hemostasis of blood vessels in patients requiring arthroscopic surgery of the knee, shoulder, ankle, elbow and wrist.

    The VAPR CP90 Electrodes when used with Mitek VAPR VUE Electrosurgical System is intended for resection, ablation and excision of soft tissue, and hemostasis of blood vessels in patients requiring arthroscopic surgery of the knee, shoulder, ankle, elbow and wrist.

    The DePuy Mitek VAPR Electrodes for use with all VAPR Electrosurgical Systems are intended for resection, ablation and excision of soft tissue, hemostasis of blood vessels and coagulation of soft tissue in patients requiring arthroscopic surgery of the knee, shoulder, ankle, elbow, and wrist.

    The DePuy Mitek VAPR Electrodes for use with the VAPR VUE Radiofrequency System are intended for resection, ablation and excision of soft tissue, hemostasis of blood vessels and coagulation of soft tissue in patients requiring arthroscopic surgery of the knee, shoulder, ankle, elbow, and wrist.

    Device Description

    The VAPR VUE Radiofrequency System is DePuy Mitek's next generation of VAPR Radiofrequency Systems. The system consists of a generator, wired or wireless footswitch and electrodes. The VAPR VUE generator was designed to provide comparable RF performance with legacy VAPR electrodes.

    As with its predicate devices including the VAPR 3 generator, this electrosurgical system utilizes bipolar technology specifically designed to provide a range of arthroscopic surgical treatments including soft tissue ablation, contouring, cutting and coagulation and temperature control.

    The VAPR VUE system offers five output types of operation: Vaporization (ablation), Coagulation, CP Vaporization (ablation), Blended Vaporization (ablation and coagulation) and temperature control coagulation. All types are present in the predicate device (VAPR3 generator) except the CP Vaporization (ablation).

    The VAPR P50 and P90 Electrodes are an addition to the family of LPS Electrodes. These Suction Electrodes are soft tissue ablation and coagulation devices intended for use with the VAPR Electrosurgical Systems. They extend the utility of the system by removing bubbles created during activation from the operating site.

    The VAPR CP90 Electrodes are identical in technological characteristics as the P90 electrode only it is intended to be run only off the VAPR VUE platform at pre-determined default settings specific for the device. The CP90 with Handcontrols offers the integration of handcontrol capabilities via three buttons molded on the existing one-piece handles. Those buttons control ablation, coagulation and the generator mode functions.

    AI/ML Overview

    The information provided describes the DePuy Mitek VAPR VUE Radiofrequency System and associated electrodes, which is a medical device for electrosurgical cutting and coagulation. The submission is a 510(k) premarket notification, aiming to demonstrate substantial equivalence to legally marketed predicate devices.

    Here's an analysis of the provided text in relation to your request:

    1. Table of Acceptance Criteria and Reported Device Performance

    The submission does not provide specific quantitative acceptance criteria or detailed reported device performance in terms of clinical metrics (e.g., success rates, complication rates, accuracy). Instead, it relies on demonstrating compliance with recognized electrical safety and software standards and asserting comparable RF performance.

    The device's performance is generally stated as:

    • RF Performance: "The VAPR VUE generator was designed to provide comparable RF performance with legacy VAPR electrodes."
    • Safety and Performance Testing: "In addition safety and performance testing have been done to validate the performance and safety of the device. It has been demonstrated that these device modifications will not affect safety and effectiveness of the subject devices."
    • Verification Testing: "Verification of the VAPR VUE System and Electrodes includes electrical, Safety and software and performance tests to show that the device meets its product Performance specifications over a range of operating conditions."

    Acceptance Criteria (Implied by Standards & Equivalence):

    Acceptance Criteria CategorySpecifics (Implied by Standards/Assertions)Reported Device Performance
    Electrical SafetyCompliance with EN 60601-1: 1995 (General requirements for safety), EN60601-2-2:2007 (Particular requirements for the safety of high-frequency surgical equipment), EN60601-1-2: 2006 (Electromagnetic compatibility)."Verification of the VAPR VUE System and Electrodes includes electrical, Safety... tests to show that the device meets its product Performance specifications over a range of operating conditions." "Results of performance and safety testing have demonstrated that the modified device is suitable for its intended use."
    Software Safety/QualityCompliance with EN 60601-1-4/A1:2005 (Programmable electrical medical systems), ISO 62304 (Medical device software-software life cycle process), FDA Guidance for Premarket Submission for Software, General Principals of Software Validation, Guidance Off-The-Shelf Software Use."Verification of the VAPR VUE System and Electrodes includes ... software and performance tests to show that the device meets its product Performance specifications over a range of operating conditions." "Results of performance and safety testing have demonstrated that the modified device is suitable for its intended use."
    RF Performance ComparabilityExpected to provide comparable RF performance to predicate devices (VAPR IIITM)."The VAPR VUE generator was designed to provide comparable RF performance with legacy VAPR electrodes."
    Functional PerformanceDevice meets product performance specifications over a range of operating conditions."Verification of the VAPR VUE System and Electrodes includes electrical, Safety and software and performance tests to show that the device meets its product Performance specifications over a range of operating conditions." "Validation testing for the VAPR VUE includes testing to show the device meets user needs."
    Substantial EquivalenceDemonstrates equivalence in intended use, essential components, materials, performance specifications, and technology characteristics to predicate devices.The conclusion states: "Results of performance and safety testing have demonstrated that the modified device is suitable for its intended use. Based on the indications for use, fundamental scientific technology, and comparison to the predicate devices, the VAPR VUE Radiofrequency System, along with the Premiere50, Premiere90 and CP90 Electrodes are shown to be substantially equivalent to the predicate devices..."

    2. Sample Size Used for the Test Set and the Data Provenance

    The document does not specify sample sizes for "test sets" in the context of clinical performance or accuracy in a medical imaging or diagnostic sense. The testing mentioned (electrical, safety, software, performance) would typically involve engineering verification and validation on the device itself and its components, rather than patient data test sets.

    • Test Set Sample Size: Not applicable/not specified in the provided text for clinical performance. The testing mentioned is for engineering verification and validation.
    • Data Provenance: Not applicable, as no clinical/patient data test sets are described. The testing is device-centric.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and the Qualifications of Those Experts

    This information is not provided because the submission explicitly states:

    • "No clinical studies are required to demonstrate safety and efficacy of the device in support of an application for premarket clearance."
    • Therefore, there were no "ground truth" derived from expert review of clinical cases.

    4. Adjudication Method for the Test Set

    This information is not applicable as there were no clinical test sets requiring adjudication by experts.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • No MRMC study was done. This device is an electrosurgical tool, not an AI-assisted diagnostic or imaging device. The concept of "human readers improve with AI vs without AI assistance" is not relevant to this type of device.

    6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) was done

    • Not applicable. This is a hardware-based electrosurgical system, not an algorithm, and does not operate in a standalone diagnostic capacity.

    7. The Type of Ground Truth Used

    • No clinical grand truth was used. The "ground truth" for the device's performance is its compliance with engineering specifications, electrical safety standards, software standards, and its ability to achieve its intended physical effects (resection, ablation, coagulation) in laboratory or simulated environments, as demonstrated through non-clinical performance and safety testing.

    8. The Sample Size for the Training Set

    • Not applicable. This device does not employ machine learning or AI that would require a "training set" of data.

    9. How the Ground Truth for the Training Set was Established

    • Not applicable. As no training set was used, no ground truth needed to be established for it.
    Ask a Question

    Ask a specific question about this device

    K Number
    K062074
    Manufacturer
    Date Cleared
    2006-11-14

    (116 days)

    Product Code
    Regulation Number
    886.4390
    Why did this record match?
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The IRIDEX Wireless Footswitch is intended for use with compatible IRIDEX Laser Systems operated in hospital or outpatient facilities.

    The IRIDEX Wireless Footswitch is indicated for use as an accessory to provide input control of laser emission to compatible IRIDEX Laser Systems. This accessory includes a wireless footswitch and a receiver. It is cleared for use for the particular indications of the laser system to which it is attached.

    Device Description

    The Wireless Footswitch is an alternate option to the existing cabled footswitch for users of IRIDEX laser consoles. It consists of two discreet parts - a footswitch/transmitter and receiver. The wireless footswitch receiver connects to the laser through the same connector as the current wired footswitch, and operates with the console by directly emulating the functionality of the current wired footswitch.

    AI/ML Overview

    This document describes the IRIDEX Wireless Footswitch, seeking clearance as an accessory for existing IRIDEX laser systems. As an accessory that emulates the functionality of a wired footswitch, the submission primarily focuses on demonstrating substantial equivalence to predicate devices rather than independent performance studies with detailed acceptance criteria.

    Therefore, the requested information fields related to clinical efficacy studies, expert ground truth, and training data are largely not applicable in the context of this 510(k) submission. The FDA clearance is based on the device's functional similarity to predicate devices and its adherence to general controls for safety and effectiveness.

    Here's an attempt to populate the table and answer the questions based on the provided text, indicating "Not Applicable" where the information is not present or relevant to a substantial equivalence submission for an accessory device:


    1. Table of Acceptance Criteria and Reported Device Performance

    Given that this is a 510(k) for an accessory (wireless footswitch) demonstrating substantial equivalence to wired and other wireless footswitches, the "acceptance criteria" are not framed in terms of clinical performance metrics like sensitivity or specificity. Instead, acceptance is based on functional equivalence and safety to predicate devices. The document implies functional equivalence to existing wired and wireless footswitches as the primary criterion.

    Acceptance Criteria CategorySpecific CriteriaReported Device Performance
    Functional EquivalenceThe wireless footswitch must directly emulate the functionality of the current wired footswitch of IRIDEX laser consoles. It must provide input control of laser emission to compatible IRIDEX Laser Systems in the same manner as the wired footswitch. Its use should be cleared for the particular indications of the laser system to which it is attached."The wireless footswitch receiver connects to the laser through the same connector as the current wired footswitch, and operates with the console by directly emulating the functionality of the current wired footswitch."
    Indications for UseThe device's indications for use must be similar to or the same as predicate devices and compatible with the existing IRIDEX laser systems."The IRIDEX Wireless Footswitch shares the same indications for use... with, and thus is substantially equivalent to, the predicate devices."
    Safety and EffectivenessThe device must be demonstrated to be substantially equivalent to predicate devices in terms of technical characteristics and operation, implying it meets the same safety and effectiveness standards as those devices. This includes general controls provisions of the Act (e.g., annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration). Compliance with applicable standards is usually demonstrated."The review of the indications for use and technical characteristics provided demonstrates that the IRIDEX Wireless Footswitch is substantially equivalent to the predicate devices."
    Design FeaturesThe design features (e.g., wireless transmission) should be similar to other legally marketed wireless footswitches."The IRIDEX Wireless Footswitch shares... similar design features, and functional features with, and thus is substantially equivalent to, the predicate devices."

    Study Details:

    1. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

      • Not Applicable. The submission is for substantial equivalence of an accessory device (wireless footswitch) based on functional emulation and comparison to predicate devices, not requiring a clinical performance study with a test set of patient data. The documentation does not mention such a study.
    2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

      • Not Applicable. No clinical test set or ground truth validation by experts is mentioned or required for this type of 510(k) submission.
    3. Adjudication method (e.g. 2+1, 3+1, none) for the test set

      • Not Applicable. No clinical test set.
    4. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

      • Not Applicable. This device is a hardware accessory (wireless footswitch), not an AI-powered diagnostic tool. MRMC studies are irrelevant.
    5. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

      • Not Applicable. This is a hardware accessory, not an algorithm. Performance is assessed based on its ability to emulate the wired footswitch function and comply with relevant standards (though not explicitly detailed in this summary).
    6. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

      • Not Applicable. Not a clinical diagnostic device requiring ground truth for clinical performance. The "ground truth" for this device's functionality would be the successful and consistent control of the laser system, as performed by the wired equivalent.
    7. The sample size for the training set

      • Not Applicable. This is a hardware accessory, not a machine learning algorithm requiring a training set.
    8. How the ground truth for the training set was established

      • Not Applicable. No training set for an algorithm.
    Ask a Question

    Ask a specific question about this device

    Page 1 of 1